Standard dilution analysis (SDA) is a recently described calibration method that combines internal standardization and standard additions to improve accuracy and precision. By combining these two traditional methods, SDA offers corrections for sample matrix effects while also minimizing signal fluctuations due to changes in the excitation source and the sampling process. In its most basic form, SDA is performed by introducing a constant amount of sample into an instrument while varying the amount of standard solution added to the sample. The procedure is simple to perform by hand, but automation of the process is challenging because the dilution of the standard solution by a blank occurs on-line while the measurement is taking place. The experiments presented here are aimed at improving the automation of the SDA process. Inductively coupled plasma optical emission spectrometry (ICP OES) is used as a model for SDA applications, although the calibration method is applicable to any technique that accepts samples as a liquid flow. The end goal of automation is for successive samples to be analyzed quickly by using an automatic sampler to switch from sample to sample, as it is the case in routine analytical methods. The required dilution of the standard by a blank mixture is achieved through the use of a two-channel pinch valve, with dilution occurring by diffusion as the solution is pulled from the valve toward the plasma by a second channel in the instrument's peristaltic pump. The proposed automated SDA method is effective, with analyte recoveries within a few percent of 100 for the elements evaluated (i.e. Al, Cd, Co, Cr, Cu, Fe, Ni and Pb) and nine different sample matrices. The method also provides promising results when analyzing real-world samples. The limits of detection obtained with the automated SDA method are similar to those typically found for external standard calibration and ICP OES, falling within the single-digit μg L−1 range.
In practice, SDA offers significantly improved accuracies and precisions when compared to traditional EC and IS methods, while offering slight improvements over SA.8,13 Two reviews are referenced that incorporate SDA into summaries of other commonly used calibration techniques in atomic spectrometry.18,19 The most significant benefit that SDA provides over traditional SA is its improved sample throughput due to a simpler sample preparation procedure. In SDA, a full calibration is performed with only two solutions. On the other hand, an important challenge for implementing SDA in routine analysis is associated with its automation. In its earliest iterations, SDA was performed manually, by pouring the blank-containing solution into the standard-containing solution. Recent advances in the method have been aimed at the automation of solution mixing, namely the use of a three-port mixing chamber directly coupled to the sample introduction system of an ICP OES.13 This setup allowed for SDA to be performed in its entirety using only the commercial instrument's own automatic sampler to dilute the standard-containing solution with the blank. However, it involved an elongated rinsing step to completely wash out residual sample from the mixing chamber before proceeding to the next cycle. In another work, a multi-channel nebulizer was used to perform SDA inside the spray chamber of an ICP OES instrument. However, automation was also limited in this case, as the instrument's automatic sampler was not used and the sample had to be fed manually into the system.15
The method presented here uses a new version of SDA solution preparation in order to speed up the automation process. These experiments use ICP OES as an application model for the calibration strategy. However, it is essential to reiterate that SDA is not limited to ICP OES, and it may be applied to any technique capable of accepting liquid samples and simultaneously monitoring multiple analytical signals. As mentioned previously, SDA requires a constant amount of sample to be introduced into the analytical system, while the amount of standard is varied through dilution with a blank. To simplify the automation of the method, the two calibration solutions are introduced into the ICP using two separate channels on the instrument's peristaltic pump. One channel is connected to the instrument's automatic sampler, which moves between individual samples as would be the case in any typical analysis. The second channel introduces the standard and blank solutions in an alternating fashion. Consistently alternating between standard and blank solutions eliminates the need for an elongated cleaning step beyond the typical rinse time adopted between samples. This is possible because the dilution occurs in a separate channel from the sample introduction itself.
StotalA = SstdA + SsamA | (1) |
To simplify the math, a constant K is defined in eqn (2) as the ratio of analyte signal in the standard solution alone to internal standard signal, where I represents the internal standard. This definition of K allows for eqn (1) to be rewritten as shown in eqn (3).
(2) |
StotalA = KSI + SsamA | (3) |
Eqn (3) is the equation for a line when the analyte signal is plotted against the internal standard signal, with a slope of K and a y-intercept that is equal to the analyte signal that arises from the sample alone. Now consider the ratio of this intercept and slope, as shown in eqn (4).
(4) |
In general, a measured signal is equal to the concentration of analyte, C, multiplied by the calibration sensitivity, m. Substituting this in to eqn (4), the concentration of the analyte in the sample is now present in the calculation, as shown in eqn (5).
(5) |
This equation can be rearranged to allow for direct calculation of the analyte concentration in the sample, as shown in eqn (6).
(6) |
When the analyte in the standard is undiluted, the internal standard must also be undiluted, thus the internal standard signal is at its maximum. The maximum internal standard signal can be estimated by first introducing the standard into the experimental system and allowing the signal to stabilize before dilution by the blank,8 or it can be calculated by introducing a second internal standard into the blank solution, providing an inverse relationship between the two internal standards. The internal standard present with the analytes in the standard solution is referred to as the primary internal standard, while the internal standard present in the blank is referred to as the secondary internal standard. The two internal standards are plotted against each other, and the maximum internal standard signal is defined as the y-intercept of this second plot. The use of a secondary internal standard improves sample throughput, as the maximum primary internal standard signal is calculated from the same points in time that provide the full calibration.13
Note that the final calculation presented in eqn (6) assumes that the sample solution and the standard plus blank solution mix are present in the analytical system in equal parts. The addition of a simple dilution factor to eqn (6) corrects for an experiment in which the sample to standard plus blank ratio is not one to one.
Nine different sample matrices were spiked with the stock solution to have the same added analyte concentration as the diluted standard, in order to determine how well the automated SDA process corrected for matrix effects. The examined matrices included DDI, tap water, 50% v/v ACS grade nitric acid, 40% v/v ethanol, a soft drink, a zero-sugar soft drink, a red wine, 50% v/v mouthwash, and 1% m/v Ca.
Several reference materials were analyzed in order to evaluate the accuracy of the automated SDA method. A water pollution standard (VHG Labs, Manchester, NH, USA) was diluted in 1% v/v HNO3 before performing SDA. A standard reference material of Oyster Tissue (NIST 1566B) from the National Institute of Standards and Technology (NIST, Gaithersburg, MD, USA) and two standards of metals in hydrocarbon oil (300 and 900 μg g−1, Agilent Technologies, Wilmington, DE, USA) were subjected to microwave-assisted digestion before analysis. The digestion program used with an Ethos Up system (Milestone, Sorisole, Italy) consisted of a 15 min ramp to 200 °C, a 15 min hold at 200 °C, and a 10 min cooling step. Aliquots of 1.5 mL trace-metal grade HNO3, 6.5 mL DDI and 2 mL 30% v/v H2O2 were used in the digestion process.
A second stock solution was prepared for the analysis of the reference materials and for determination of the method's limits of detection (LODs). A multi-element standard containing all of the 8 analytes and Sc as an internal standard at 100 mg L−1 (Teledyne Leeman Labs, Hudson, NH, USA) was diluted 1:100 in DDI with added 1% v/v HNO3. A second blank was prepared in 1% v/v HNO3, with 0.5 mg L−1 Y as an internal standard and 100 mg L−1 Li to help balance the ionic strength of the two solutions.
The analyte and internal standard emission lines used here are provided in Table 1. Emission intensities of each analyte and internal standard were monitored at two different wavelengths, but only results from one line are presented here. The use of different emission lines for a given analyte changes the calculations to a minimal extent, and the use of multiple internal standard emission lines is even more inconsequential. Unless otherwise noted, all results discussed here use the measured intensity of the lines provided in Table 1.
Element | Species | Wavelength (nm) |
---|---|---|
Al | I | 396.152 |
Cd | I | 228.802 |
Co | II | 237.863 |
Cr | II | 267.716 |
Cu | I | 324.754 |
Fe | II | 259.940 |
Ni | I | 341.476 |
Pb | I | 405.781 |
Sc | II | 361.383 |
Y | II | 371.029 |
The novelty of the automation technique presented here lies in how the sample and standard plus blank solutions are introduced into the plasma simultaneously, with the standard being diluted from a maximum to blank level in an alternating fashion. Sample solutions are placed in an automatic sampler and drawn up through the tubing by the peristaltic pump. A second channel on the peristaltic pump creates the standard plus blank mixture through the use of a two channel pinch valve (Cole-Parmer, Vernon Hills, IL, USA). Tubing from one channel in the valve draws the prepared standard solution, while the second channel draws from the prepared blank solution. One channel of the valve is open by default. When 12 V are applied to the valve, the first channel closes and the second one opens. A digital timer outlet (NEARPOW, TX, USA) is used to switch a low cost, simple 12 V DC power supply (120 VAC to 12 VDC) adapter, from off to on, and vice versa, every 60 s. This results in an alternating flow of pure standard or blank solution once per minute. The two channels are joined together into a single piece of tubing by a plastic Y junction. The pinch valve allows for the standard and blank solutions to be interchanged without the introduction of air bubbles into the tubing. As the liquid flow remains constant, the standard solution is diluted by the blank solution on-line inside of the tubing as it travels from the valve through the peristaltic pump. The sample and the standard plus blank flows are joined together on the back end of the peristaltic pump by a second Y junction before introduction into the nebulizer and into the plasma.
All instrumental parameters used in the present work are provided in Table 2. All of the ICP OES operating parameters were kept at their default values. As the pinch valve setup provided dilution of the standard plus blank mixture independent of the ICP OES′ automatic sampler, the uptake delay and stabilization time relate to the introduction of the sample itself into the plasma. These parameters were also kept at their default values. Only minimal changes were required to the software in order to run automated SDA. The pump speed was kept constant during the rinse step instead of speeding it up in order to preserve the nature of the standard plus blank dilution. The read time and number of replicates are the two most important parameters when SDA is performed. The read time was kept short (to the software minimum of 1 s) to provide the highest possible time resolution of the signal change provided by the dilution of the standard by the blank, and vice versa. The number of replicates was kept high (to the software maximum of 100) to provide additional points in the SDA calibration curve (as each point in time provides one point on the curve).
Instrumental parameter | Operating condition |
---|---|
Plasma | |
RF applied power | 1.20 kW |
Viewing mode | Axial |
Nebulizer gas flow rate | 0.7 L min−1 |
Plasma gas flow rate | 12.0 L min−1 |
Auxillary gas flow rate | 1.0 L min−1 |
General | |
Pump speed | 12 rpm |
Uptake delay | 90 s |
Stabilization time | 30 s |
Rinse time | 30 s |
Replicates | 100 |
Read time | 1 s |
As the valve was switching channels every 60 s, the signal intensity traces switch from increasing to decreasing every 60 s. Although the channels in the pinch valve operate in an on/off logic fashion, the intensities changed rapidly over a time of approximately 30 s before stabilizing instead of changing immediately. This means that the dilution of the standard by the blank happened by diffusion as the solution travels from the valve through the peristaltic pump and into the plasma. That explains why one observes a ramp rather than a jump in the signal profiles, as shown in Fig. 2. The gray boxes in Fig. 2 show representative times for which a single automated SDA run is performed (100 time points), with the rinse, sample uptake, and stabilization times accounting for the gap in between. A 100-point collection consisted of a full cycle of signal change from minimum to maximum and back to minimum again, but not necessarily beginning and ending with a maximum or minimum.
Fig. 3 provides an example of the plots needed to perform SDA calculations. The sample was DDI spiked with 2.27 mg L−1 Cr, which is the same concentration of Cr present in the prepared standard. Recall that Y is present in the standard solution as an internal standard, with Sc in the blank solution as the secondary internal standard. The Cr versus Y signal plot has a y-intercept of 16304 and a slope of 0.437. The Y vs. Sc signal plot has a y-intercept of 36988. Eqn (7) provides an example SDA calculation using the data shown in Fig. 3, which return a result of a 2.29 mg L−1 Cr, only a + 0.9% error from the true spike value of 2.27 mg L−1.
(7) |
Fig. 4 SDA plots obtained for Cr spiked into 9 different matrices. The topmost curve (black diamonds) corresponds to spiked deionized water. |
Note that the slope of the line was preserved between sample matrices. This is predicted by the theory, as the slope is defined as the signal ratio of the analyte present in the standard alone to the internal standard. If this ratio was not preserved between samples, the internal standard reacts differently to sample matrices than the analyte, making it a poor choice of internal standard (in a traditional sense) when examining multiple sample types. However, SDA would still work if the lines have different slopes, as the sample matrix for a given sample is preserved between the standard and blank solutions for an individual sample run.
Full investigations comparing SDA to traditional calibration methods have been performed and published previously.8,13 The focus of this research was simplifying the automation of SDA, and thus a comparison with a traditional calibration method was not performed. Nevertheless, previous results can be summarized as follows:8,13 SDA provides improved accuracy and precision when compared to a simple external standard calibration or internal standard method, while providing similar or slightly improved results compared to the traditional standard additions method. The major benefit provided by SDA over standard additions is the simplified solution preparation, and thus improved sample throughput.
Each of the 9 matrices was initially run in triplicate. One additional measurement was taken for the tap water, soda, and wine matrices, resulting in a total number of 30 measurements for each element. This additional measurement was to assess the effect of small bubbles passing through the peristaltic pump tubing, as shown in Fig. 5.
Traditional ICP OES methods involve averaging emission signals over extended lengths of time in order to improve signal precision. However, this is not possible in the case of automated SDA, as the signals need to be collected as quickly as possible in order to achieve a wide range of signal coverage. If too long of a time period is averaged, the dilution of the standard by the blank is lost in the averaging, resulting in no signal change over time. Collecting data points quickly provides many points on individual calibration curves. Although short measurement times are more subject to noise than longer integrations, the additional points collected in a single calibration averages the noise out in the end.
When a bubble comes through the pump tubing, discontinuities appear in the time resolved signals, as shown by the Cr emission intensities on the left-hand side of Fig. 5. The dominant, slowly increasing and decreasing signals are created by the dilution of the standard by the blank, and vice versa, that allow SDA to function. The sharp deviations in signal that occur over two or three data points were caused by bubbles. SDA plots of these data sets, bubbles included, are provided on the right-hand side of Fig. 5. The linear fits depict all 100 data points, including the bubble points that fall far away from the line. The sheer number of points in the calibration curve allows for the method to correct for a bubble or two passing through the sample introduction line over the course of the measurement.
Recall that the sample matrices were spiked with 2.27 mg L−1 Cr. When including all 100 data points, the data sets shown in Fig. 5 return a calculated spiked Cr concentration of 2.31, 2.07, and 2.32 mg L−1 in the wine, 1% m/v Ca, and tap water matrices, respectively. When removing the points caused by bubbles, these calculated concentrations are basically unchanged in wine and 1% m/v Ca, and only slightly changed in tap water, returning values of 2.30, 2.06, and 2.20 mg L−1, respectively, showing that this method is capable of handling bubbles passing through the pump tubing, even when collecting data points quickly.
Element | LOD (mg L−1) | |
---|---|---|
1 mg L−1 standard | 10 mg L−1 standard | |
Al | 0.006 | 0.05 |
Cd | 0.005 | 0.03 |
Co | 0.005 | 0.04 |
Cr | 0.002 | 0.04 |
Cu | 0.003 | 0.05 |
Fe | 0.006 | 0.05 |
Ni | 0.008 | 0.05 |
Pb | 0.1 | 0.06 |
The LODs are largely similar for each element, with lead being the lone exception. When using a 1 mg L−1 standard, the LOD for each element (except lead) is on the order of single digit parts per billion. When using a 10 mg L−1 standard the LODs increase similarly, up an order of magnitude, to tens of parts per billion. The LOD for lead is on the order of 100 μg L−1 for both standard concentrations. These are typical LODs for ICP OES when using traditional calibration methods. The LODs differ when using different concentration standards because a full calibration is collected every time SDA is performed.
When performing a calibration by SDA, using a higher concentration standard results in more uncertainty in values near zero. If the standard concentration is the only parameter that is changed between two SDA calibrations, the total number of points in the changing signal profile is unchanged. Thus using lower concentrations improves estimates of the noise level as the detection limit is approached, as there are more points in the calibration at low standard concentrations because the maximum standard concentration is lower. As a full calibration is required in order to return a blank signal using SDA, using a lower concentration standard provides slightly lower LODs.
Table 4 displays the average percent recovery calculated for each analyte element across the 9 sample matrices. The automated SDA method accurately corrects for matrix effects regardless of the analyte selected, returning a recovery of near 100% for each element. Table 5 displays the same data in a different way, showing the average percent recovery calculated for each sample matrix across the 8 spiked analytes. Again, SDA is able to provide accurate results across a wide range of matrix types. Even a 1% m/v Ca solution, which is a particularly difficult matrix, returns an accurate average recovery of 92%. The repeatability of the automated SDA method was examined by running spiked DDI, 1% m/v Ca, and 50% v/v mouthwash matrices in 10 successive replicates. The results are shown in Table 6. As expected from the previous results, both DDI and 50% v/v mouthwash return concentrations within a few percent of the spiked value. The 1% m/v Ca matrix again returns results that are below the spiked value, but still accurate especially considering the difficulty of the matrix itself. Regarding the precision of the automated SDA technique, the RSDs were on the order of 3% irrespective of the matrix examined. It is important to note that accuracy and precision need to be combined in order to provide better estimates of uncertainty between different measurement techniques.20 As the goal of this work was to simplify the automation of SDA as a calibration technique, a full comparison of SDA to other traditional calibration methods was not performed, but has been previously reported. SDA has been shown to offer similar results to traditional standard additions methods, while improving upon simple external calibration or internal standardization methods.8,13
Element | Recovery (%) | n |
---|---|---|
Al | 98 ± 3 | 30 |
Cd | 99 ± 3 | 30 |
Co | 98 ± 3 | 30 |
Cr | 100 ± 3 | 30 |
Cu | 99 ± 7 | 30 |
Fe | 100 ± 3 | 30 |
Ni | 97 ± 3 | 30 |
Pb | 96 ± 3 | 30 |
Matrix | Recovery (%) | n |
---|---|---|
DDI | 100 ± 2 | 24 |
Tap water | 102 ± 6 | 32 |
50% v/v nitric acid | 98 ± 2 | 24 |
40% v/v ethanol | 99 ± 3 | 24 |
Soda | 99 ± 1 | 32 |
0 sugar soda | 98 ± 4 | 24 |
Wine | 100 ± 3 | 32 |
1% m/v Ca | 92 ± 2 | 24 |
50% v/v mouthwash | 99 ± 2 | 24 |
Element | Spike | DDI matrix | 1% Ca m/v matrix | 50% v/v mouthwash matrix | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
mg L−1 | Avg | StDev | RSD (%) | Accuracy (%) | Avg | StDev | RSD (%) | Accuracy (%) | Avg | StDev | RSD (%) | Accuracy (%) | |
Al | 1.36 | 1.34 | 0.04 | 2.7 | −1.4 | 1.42 | 0.09 | 6.3 | 4.6 | 1.44 | 0.05 | 3.6 | 5.9 |
Cd | 4.54 | 4.5 | 0.1 | 2.5 | −1.4 | 4.2 | 0.1 | 3.0 | −8.4 | 4.5 | 0.1 | 3.2 | 0.06 |
Co | 4.54 | 4.5 | 0.1 | 2.6 | −1.9 | 4.1 | 0.1 | 3.2 | −9.5 | 4.4 | 0.2 | 3.7 | −2.1 |
Cr | 2.27 | 2.27 | 0.06 | 2.7 | −0.1 | 2.09 | 0.08 | 3.7 | −7.8 | 2.27 | 0.09 | 3.9 | −0.04 |
Cu | 1.36 | 1.34 | 0.03 | 2.5 | −1.9 | 1.25 | 0.03 | 2.7 | −8.3 | 1.34 | 0.05 | 3.4 | −1.3 |
Fe | 4.54 | 4.5 | 0.1 | 2.6 | 0.2 | 4.2 | 0.2 | 3.8 | −7.5 | 4.5 | 0.2 | 3.7 | 0.07 |
Ni | 4.54 | 4.4 | 0.1 | 2.4 | −2.6 | 4.0 | 0.1 | 2.9 | −11.6 | 4.5 | 0.2 | 3.4 | −1.9 |
Pb | 13.6 | 13.0 | 0.3 | 2.4 | −4.9 | 12.1 | 0.3 | 2.6 | −10.9 | 13.2 | 0.5 | 3.7 | −3.3 |
Element | Water pollution standarda | Oyster tissueb | 300 μg g−1 metals in oilc | 900 μg g−1 metals in oilc | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Concentration (mg L−1) | Recovery (%) | n | Concentration (mg L−1) | Recovery (%) | n | Concentration (mg L−1) | Recovery (%) | n | Concentration (mg L−1) | Recovery (%) | n | |
a VHG Labs, Manchester, NH, USA. b NIST, Gaithersburg, ND, USA. c Agilent Technologies, Wilmington, DE, USA. | ||||||||||||
Al | 4.53 | 91 ± 8 | 6 | 0.38 | 93 ± 3 | 3 | 0.40 | 110 ± 6 | 6 | 1.10 | 103 ± 3 | 6 |
Cd | 0.27 | 110 ± 2 | 6 | 0.34 | 94 ± 2 | 6 | 0.99 | 93 ± 2 | 6 | |||
Co | 0.97 | 98 ± 3 | 6 | |||||||||
Cr | 1.01 | 102 ± 3 | 6 | 0.25 | 67 ± 1 | 6 | 0.84 | 79 ± 2 | 6 | |||
Cu | 1.12 | 113 ± 3 | 6 | 0.16 | 107 ± 4 | 3 | 0.34 | 93 ± 2 | 6 | 1.00 | 94 ± 3 | 6 |
Fe | 0.96 | 97 ± 5 | 6 | 0.41 | 95.3 ± 0.8 | 3 | 0.32 | 88 ± 2 | 6 | 0.90 | 85 ± 3 | 6 |
Ni | 1.17 | 118 ± 4 | 6 | 0.37 | 102 ± 2 | 6 | 1.13 | 106 ± 3 | 6 | |||
Pb | 1.08 | 109 ± 4 | 6 | 0.32 | 89 ± 4 | 6 | 1.04 | 97 ± 5 | 6 |
Overall the results obtained using automated SDA for real samples were accurate, with average recoveries largely falling between 90 and 110%. The lone exception is Cr in the hydrocarbon oil samples, which provided recoveries significantly below 90%. However, the results for the EC method applied to the same sample were worse than those obtained by SDA, which suggests not a problem with the SDA method in the determination of Cr, but more likely a problem during the sample preparation.
The results obtained for Al in the water pollution standard and oil standards used a secondary emission line at 167.019 nm. This was necessary due to an interfering emission that was observed in the sample causing a falsely high recovery value.
Experiments presented here were aimed at improving the automation of the SDA process, allowing for successive samples to be analyzed quickly by simply using an automatic sampler to switch from sample to sample, as is the case in most routine analytical methods. The automation of the dilution of the standard plus blank mixture was achieved through the use of a two-channel pinch valve, with the dilution occurring by diffusion as the solution is pulled from the valve to the plasma by a second channel in the instrument's peristaltic pump. The proposed automated SDA method proved to be effective over various sample matrices, and also provided promising results when analyzing real-world samples.