Efficient micromixing in enclosed microchannels is essential for reliable lab-on-a-chip operation but typically requires planar or multilayer soft-lithographic fabrication, limiting geometric freedom and increasing production complexity. Here, we report a fully monolithic split-and-recombine (SAR) micromixer fabricated in a single step using stereolithography digital light processing (SLA-DLP), eliminating molds, bonding, and cleanroom processing. The three-dimensional SAR architecture systematically divides, reorients, and recombines fluid streams, enabling high mixing performance over a wide range of operating conditions. Computational fluid dynamics simulations and experimental validation show excellent agreement, achieving mixing efficiencies above 0.90 across Reynolds numbers from 0.1 to 100. Leveraging its compact and robust performance, the micromixer was integrated into a five-output microfluidic concentration gradient generator, which produced stable and reproducible concentration profiles for both fluorescent tracers and protein solutions. The complete device, including microchannels and functional fluidic features, was printed in under 1.5 h using a standard desktop SLA-DLP system. These results demonstrate that additive manufacturing can deliver high-performance micromixing capabilities, establishing a rapid, accessible, and fully digital route for the fabrication of advanced microfluidic systems.
To address the limitations of purely diffusive transport, a wide range of micromixer strategies has been developed over the past two decades. Passive micromixers, in particular, have attracted sustained interest because they rely exclusively on geometric features rather than external energy sources, making them simple, robust, and readily integrable into larger lab-on-a-chip systems.9,10 Representative passive designs include the herringbone micromixer introduced by Stroock et al., which induces chaotic advection through transverse secondary flows,11 obstacle-based micromixers that perturb streamlines to enhance interfacial stretching,12,13 and serpentine and zig-zag channels that increase residence time and elongate fluid interfaces.14,15 Among these approaches split-and-recombine (SAR) architectures are particularly notable, as they systematically divide and reorient fluid streams to repeatedly reduce diffusion length scales, offering a scalable rout toward efficient mixing over a broad range of Re and channel geometries.16,17
Numerous SAR geometries have been proposed to minimize the diffusion length and mixing time under laminar flow conditions. Early planar or double-layer SAR designs, typically fabricated using soft lithography or deep reactive ion etching, demonstrated efficient mixing at low Re (<10),.18,19 Subsequent variations including D-shaped sub-channels,20 double-layer Y-type SAR structures,19 and crossing or periodically interconnected junctions21,22 improved interfacial stretching while reducing pressure drop, typically achieving complete mixing (ME > 0.9) within channel lengths of 5–20 mm for Re > 30. Multi-layer PDMS and silicon–glass hybrid devices18 extended the operational Re range up to 100, while further geometric optimization, such as zigzag-angle tuning, broadened the effective operating window.23 More recent studies have emphasized modular and scalable architectures, including a stackable SAR micromixer,24 a “twisted” three-dimensional design,25 and the Y–Y micromixer,19 a CNC-milled polycarbonate device employing alternating vertical ducts and planar Y-junctions, which achieve stable mixing (ME > 0.9) across wide Re range (1–100) with moderate pressure drop.
A comparative overview of SAR micromixers reported over the past decade is provided in Table 1, highlighting their materials, fabrication methods, characteristic dimensions, and operational ranges. The majority of reported SAR micromixers continue to rely on traditional photolithography and soft-lithography routes. Alternative manufacturing approaches, such as polymer lamination and xurography, have been explored to accelerate prototyping,26 but these methods remain largely constrained to quasi-two-dimensional layouts. Consequently, fabrication complexity continues to limit both geometric freedom and scalability, motivating the exploration of more accessible and versatile manufacturing techniques for passive micromixers.27,28 Additive manufacturing has recently emerged as a promising alternative for microfluidic device fabrication. More sophisticated three-dimensional structures have been realized using two-photon polymerization (2PP), particularly for printing molds for PDMS replication.17 While such approaches can achieve excellent mixing performance in microchannels as small as 100 × 100 μm2, they still rely on multistep soft-lithography workflows and indirect fabrication. In contrast stereolithography (SLA) 3D printing offers a combination of relatively high resolution, affordability, and the ability to directly fabricate enclosed internal channels with arbitrary three-dimensional configurations.27,28 Beyond rapid prototyping, SLA 3D printing enables the integration of macro- and micro-scale features, such as ports, connectors, and reservoirs, within a single print, facilitating rapid iteration and increased functional complexity directly from digital models.
| Fabrication method | Material | Re range | Channel dimensions (width × height) | Channel length | Reference |
|---|---|---|---|---|---|
| CNC micromilling + assembly | Polycarbonate | 1 to 100 | [0.4 to 0.6 mm] × 0.4 mm | 17.6 mm | 19 |
| Soft-lithography + multilayer bonding | PDMS | 0.5 to 100 | n/r | 8 mm | 36 |
| Soft-lithography | PDMS | 1 to 80 | 200 μm × n/r | 10.2 mm | 20 |
| DRIE + glass bonding | Si/glass | 0.5 to 100 | 200 μm × 200 μm | n/r | 18 |
| Soft-lithography | PDMS | 1 to 220 | 300 μm × 150 μm | 8.5 mm | 21 |
| Soft-lithography | PDMS | n/r (200 mm s−1) | 100 μm × 30 μm | 8.75 mm | 22 |
| Soft-lithography | PDMS | 0.309 to 309 | 100 μm × 100 μm | 40 mm | 23 |
| Lamination/Xurography | PVC film | 0.1 to 100 | 700 μm × 70 μm | >13 mm | 26 |
| Soft-lithography | PDMS | 3.4 to 23.8 | 500 μm × 150 μm | 17 mm | 24 |
| Soft-lithography | PDMS | n/r (1.0 to 1000.0 μl min−1) | 200 μm × 200 μm | 6.3 mm | 25 |
| Two-photon polymerization + PDMS molding | PDMS | 0.01 to 10 | 100 μm × 100 μm | 2 mm | 17 |
| SLA 3D printing | Polymer resin | n/r | 500 μm diameter | 22 mm | 37 |
| Soft-lithography | PDMS | 0.1 to 45 | 100 μm × 100 μm | 10.51 mm | 38 |
| Soft-lithography | PDMS | 3.3 to 40 | 50 μm × n/r | 1.5 mm | 39 |
| CNC milling and engraving machine | Polycarbonate | 1 to 100 | 400 μm × 400 μm | n/r | 40 |
| CNC milling and engraving machine | PMMA | 0.5 to 100 | 200 μm × 200 μm | 10.4 mm | 41 |
| Layer by layer milling | PMMA, PTFE | 25 to 100 | 800 μm × 800 μm | 31.5 mm | 42 |
| FDM 3D printing + PDMS molding | PDMS | n/r (60 to 960 μl min−1) | 200 μm × 300 μm | 30 mm | 43 |
| Multijet 3D printer | Polymer resin | 3.7 to 37.04 | 500 μm × 400 μm | 20 mm | 44 |
| Multijet 3D printer | Polymer resin | 0.1 to 50 | 750 μm diameter | 20 mm | 45 |
| SLA DLP 3D printing | Polymer resin | 0.1 to 100 | 300 μm × 300 μm | 10.4 mm | This study |
Despite the geometric freedom offered by SLA-based 3D printing, truly monolithic SAR micromixers fabricated in a single additive manufacturing step and comprehensively characterized over broad flow regimes remain largely unexplored. In particular, the incorporation of such micromixers into system-level microfluidic architectures has received limited experimental attention.
In this study, we present the design, numerical simulation, and experimental characterization of a fully 3D-printed SAR micromixer operating across a broad range of Re, along with its direct incorporation into a five-output microfluidic concentration gradient generator (MCGG) as a representative application. By embedding SAR micromixers within conventional “Christmas tree” and Y-junction network architectures, we fabricated a monolithic MCGG capable of producing five discrete, stable, and reproducible outlet concentrations over a wide range of flow conditions, which is particularly advantageous for high-throughput systems. The performance of the SAR micromixer and MCGG was evaluated through computational fluid dynamics (CFD) simulations and experimental measurements, confirming mixing uniformity and gradient reproducibility across Re ranging from 0.1 to 100. In addition, the fabrication process proved efficient and accessible, enabling complete device fabrication in under 1.5 hours using a mid-range SLA-DLP printer.
![]() | (1) |
is the expected concentration across the channel cross section. The value of ME ranges between 0 and 1, with 1 indicating homogeneous mixing across the channel cross-section.
![]() | (2) |
Another parameter used to characterize the micromixer is the mixing time at various tested flow rates (corresponding to different Re). We defined complete mixing as ME greater than 0.90, which is consistent with the criterion adopted in previous studies.17 We then determined the location along the micromixer where this criterion was achieved for each flow rate. The mixing time was then calculated using the following expression
![]() | (3) |
![]() | (4) |
This relationship highlights that the time required for molecular diffusion increases quadratically with the diffusion distance. Split-and-recombine (SAR) micromixers reduce mixing time by repeatedly splitting and recombining flow streams, effectively shortening the length scale over which diffusion must occur. Fig. 1 shows the three-dimensional (3D) model of the designed SAR micromixer, which has been optimized for SLA DLP 3D printing, along with the 3D printed device, more information about the optimization process can be found in the SI. The micromixer comprises eight SAR mixing units, enabling a mixing efficiency above 0.90 for Re ranging from 0.1 to 100. The cross-sectional dimensions of the inlet and outlet channels are 300 × 300 μm2. The narrowing sections of the SAR units were designed with a minimum width or height of 150 μm. This dimension represents a practical fabrication limit for our SLA DLP setup. Although the printer's XY resolution is 40 μm, we found that attempting to print enclosed channels with dimensions smaller than 150 μm frequently led to clogging or the retention of excess cured resin that could not be effectively removed during the cleaning process. For the dimensions reported in this manuscript, the cleaning procedure was consistently successful in clearing the internal volumes. However, as documented in literature, achieving smaller enclosed features using commercial apparatus remains a significant challenge in 3D-printed microfluidics due to the difficulty of extracting uncured resin from narrow channels.33
At each stage, the fluid streams split along the channel height into two sub-streams that flow symmetrically upward and downward. These sub-streams are then geometrically stretched vertically and narrowed horizontally to reach cross-sectional dimensions of 150 μm in width and 300 μm in height. Finally, they are recombined, doubling the number of fluid lamellae. This iterative sequence progressively increases the interfacial area between fluids and reduces the characteristic diffusion length. If l0 represents the initial diffusion length, the effective diffusion length after n SAR mixing stages becomes
![]() | (5) |
Consecutively, the required mixing time is reduced to
![]() | (6) |
This exponential reduction demonstrates the theoretical basis for the SAR strategy's mixing performance. Even a small number of stages can reduce the diffusion time by several orders of magnitude.
The final geometry of the SAR micromixer was refined through a systematic investigation of the fabrication constraints associated with the SLA-DLP process to ensure successful monolithic printing and consistent removal of uncured resin. Internal channel widths were selected to allow for the effective clearing of trapped material from the narrowing sections of the SAR units, thereby avoiding internal defects. A specific offset was integrated at the splitting junctions to maintain the structural separation of the channels and prevent them from merging during the printing process. Additionally, a unidirectional arrangement of the units was chosen over an alternating design to encourage a consistent rotational flow that enhances the split-and-recombine mechanism, particularly at higher Re. The architecture also incorporates straight transition sections rather than smooth curves. This geometry promotes stronger reorientation of fluid layers, improving mixing efficiency across a broad range of Reynolds numbers while remaining compatible with the manufacturing capabilities of the printer.
Computational fluid dynamics simulations were conducted using COMSOL Multiphysics, as described in section 2.1, for Re ranging from 0.1 to 100. The mixing efficiency was evaluated at multiple positions along the length of the micromixer to quantify the mixing performance after each SAR mixing unit. A baseline simulation of a Y-shaped straight channel, identical in cross-sectional dimensions but without SAR mixing units, was used as a reference to quantify the enhancement in mixing achieved through SAR geometric modulation.
Fig. 2(A) illustrates the concentration distribution along the micromixer and the progressive homogenization after each SAR unit at an inlet flow rate of 9.027 μL min−1 corresponding to Re = 1. It is evident that the mixing efficiency increases consistently from unit to unit, highlighting the role of split-and-recombine geometry in enhancing interfacial contact and reducing diffusion length scale. Without the SAR structures, the species travelled the full 10.4 mm with no significant mixing. The quantified simulation results for flow rates from 0.9 to 900 μL min−1 (Re 0.1 to 100) are shown in Fig. 2(B). These results indicate that the SAR micromixer achieves complete mixing (ME > 0.90) after eight mixing units, corresponding to a total channel length of 10.4 mm. In contrast, the straight channel exhibited negligible mixing, with the mixing efficiency remaining below 0.25 along its entire length, even at low Re.
To experimentally validate the SAR micromixer, we conducted two independent sets of experiments using a phenolphthalein/NaOH colorimetric assay and fluorescence confocal microscopy of fluorescein isothiocyanate–dextran (FITC-Dextran 10 kDa) in Milli-Q water, as described in section 2.3. Phenolphthalein is a transparent, colorless solution that turns pink in the presence of alkaline conditions, providing a direct visual indicator of the mixing process. Numerical simulations were also performed for each experimental condition, and the corresponding diffusion coefficients of the mixing agents were determined.
Fig. 3(A) shows a side-by-side comparison of the experimental (left column) and simulated (right column) mixing profiles of phenolphthalein and NaOH in ethanol at a flow velocity corresponding to Re = 1. A stable laminar interface is observed from the inlet up to the first SAR unit, indicating minimal initial mixing. As the fluid progresses through the SAR units, the two streams gradually mix, and the resulting pink coloration fills the channel uniformly, with no discernible unmixed streams. This visual uniformity confirms the adequate mixing achieved by the SAR design. The simulated concentration field closely matches the experimental result, supporting the accuracy of the numerical model. Experiments and simulations were performed at flow rates corresponding to Re = 0.1, 0.5, 1, 5, 10, 30, 50, and 100 and the mixing efficiency after the eighth SAR unit was computed using eqn (2). The results presented in Fig. 3(B) show the mixing efficiency (ME) as a function of Re for both the 4th and 8th SAR units. Error bars in Fig. 3(B) represent the standard deviation calculated from three independent images, with three randomly selected transects per image analyzed along the channel, as described in section 2.3. The experimental and simulated data exhibit strong agreement across the entire range of tested Re (0.1–100). Complete mixing (ME > 0.90) is achieved after the final SAR unit (8th) for all conditions. The ME after the 4th unit across the investigated Re range exhibits a non-monotonic relationship between ME and Re, a phenomenon documented in various passive SAR micromixers. In the diffusion-dominated regime (Re < 1), the initial decrease in ME occurs because the residence time within the device is reduced as the flow rate increases, limiting the opportunity for molecular species to diffuse across the fluid interface. This behavior has been previously observed in similar SAR geometries.17 As Re increases from 1 to 50, the system transitions into an advection-dominated regime. Here, the specific 3D geometry of the mixer forces the fluid to split and recombine while changing planes, which induces centrifugal forces and a rotational component to the flow.34,35 The temporary decrease in ME observed for Re > 50 is attributed to these rotational flow components counterbalancing the split-and-recombination effect. However, with eight SAR units, the high number of fluid lamellae significantly reduces the diffusion length scale. This ensures that the counteracting balance no longer impacts the overall performance, allowing for rapid homogenization and maintaining a high final ME across the entire Re range. The specific Re values where these transitions occur are specific to our geometry and dimensions. Minor deviations between simulation and experiment are attributed to fabrication imperfections and internal surface roughness, which can affect optical clarity and the accuracy of image-based quantification.
While mixing efficiency offers insight into final mixing quality, it does not reflect how rapidly that state is reached along the length of the micromixer. To capture this, we evaluated the mixing time, defined as the time required for the flow to achieve a mixing efficiency of 0.90 or higher. This was calculated using eqn (3). As shown in Fig. 3(C), the mixing time decreases significantly with increasing Re, from more than 14 seconds at Re = 0.1 to less than 25 milliseconds at Re = 100. Importantly, not all Re require the same number of SAR units to achieve complete mixing. Fewer units are needed at lower Re as the slower flow velocity allows more time for molecular diffusion to occur. In contrast, at higher Re, where residence time is limited and diffusion alone is insufficient, more SAR units are necessary to actively enhance mixing through repeated splitting and recombination.
The internal pressure distribution of the device was calculated through numerical simulations employing the same parameters used for our experimental studies. As detailed in the SI, the simulations predict a maximum pressure drop of approximately 7.5 kPa at Re = 100 across the SAR micromixer under the same conditions used in the experiments. The monolithic 3D-printed architecture accommodated these pressures without leakage or structural failure.
Fig. 3(D) shows the mixing dynamics of FITC-Dextran 10 kDa in Milli-Q water across nine channel cross-sections, beginning immediately before the first mixing unit and ending after the eighth, as measured by confocal fluorescence microscopy. The left panel images are reconstructed from acquired z-stacks along the channel height and depict the fluorescence intensity distribution across the cross-sections. The lower halves of these images, corresponding to the bottom half of the 300 μm-deep channels, do not show a fluorescent signal due to attenuation through the channel depth and overlying resin, combined with the limited sensitivity of the confocal system. Therefore, these results are used primarily for qualitative mechanistic insight, while quantitative validation of homogenization is provided by the rest of experimental analysis and simulations. Based on the top halves, alternating dark and fluorescent streams are observed across the channel width, demonstrating successful split–recombination and progressive homogenization as flow proceeds through the SAR units. In these experiments, the inlet flow rate was set to 9.027 μL min−1 (total flow rate of 18.054 μL min−1 in the mixing channel, corresponding to Re = 1). The right panel of Fig. 3(D) shows cross-sectional images from computational fluid dynamics simulations using COMSOL Multiphysics. Comparison of the experimental results (top halves of the channels) with the simulations shows good agreement in the dynamics of split–recombination and in the locations of the combined sub-streams.
The results presented in this section confirm that the SAR micromixer achieves high performance across a wide range of flow conditions, with strong agreement between simulations and experimental observations. The ability to achieve complete mixing (ME > 0.90) within eight mixing units for a wide range of Re (0.1 to 100) highlights the effectiveness of the SAR architecture. When compared with previously reported SAR micromixers summarized in Table 1, the present device achieves comparable or higher mixing efficiency while maintaining performance across a broader Re range. Notably, the micromixer was fabricated entirely using desktop SLA 3D printing, demonstrating the feasibility of producing geometrically complex internal structures without the need for cleanroom facilities or multistep microfabrication processes. Despite its three-dimensional design, the micromixer can be easily and rapidly fabricated, tested, and iteratively refined within a single working day, enabling fast prototyping and performance optimization. These capabilities make it ideally suited for integration into more advanced microfluidic systems. In the following section, we demonstrate the incorporation of this 3D-printed SAR micromixer into a multi-output microfluidic concentration gradient generator (MCGG), designed to produce five distinct, well-defined concentrations at discrete outlets for chemical and biological applications.
In addition to the microfluidic network, the system includes 3D-printed functional modules such as threaded fluidic ports, Luer-lock adapters, and a sample collector compatible with standard microcentrifuge tubes. These components were integrated seamlessly during the design phase and fabricated simultaneously using SLA DLP 3D printing (Fig. 5), demonstrating the versatility of the approach in producing both micro- and macro-scale features for plug-and-play microfluidic systems.
Simulations were performed using the Laminar flow and transport of diluted species modules in COMSOL Multiphysics. The model predicted progressive concentration division at key internal nodes and complete mixing at the outlet, even under low Re regimes. When two input solutions with relative concentrations of 0 and 1 were introduced, the simulated outlet values matched the target profile of 0, 0.25, 0.5, 0.75, and 1.0. To assess the accuracy of intermediate mixing, normalized concentrations were evaluated at three critical nodes, located within the primary and secondary branches where splitting and recombination occur. While complete mixing is not required at the internal nodes, the incorporation of two SAR units is essential to achieve the correct average concentration delivered to the subsequent stage. By introducing two SAR units, the fluid layers are geometrically split and recombined, providing a symmetrical average concentration between the right and left halves of the cross-section, independent of local mixing efficiency. This ensures that the two split subsections deliver the intended intermediate concentrations to the next division channels, therefore maintaining the gradient accuracy. A simple straight channel cannot achieve this behavior, as the absence of transverse transport would cause the left and right halves of the cross-section to retain their initial unmixed concentrations (approximately 0 and 1). As shown in Fig. 6(A), the average concentration of the right and left halves of each cross-sectional plane at these nodes (nodes 1, 2, and 3) closely followed the theoretical values of 0.33, 0.50, and 0.66, respectively, across a range of Re. This confirmed that the designed network reliably achieved predictable intermediate concentrations prior to final mixing.
Fig. 6(B) presents the simulated cross-sectional profiles at the same three critical nodes at a representative flow rate of 90.27 μL min−1 (corresponding to Re = 5), illustrating the lamellar distribution and progressive interfacial development at internal mixing points. The final outlet concentrations, evaluated as the average cross-sectional concentration at each outlet, are shown in Fig. 6(C). The results exhibit consistent output gradients of 0, 0.25, 0.5, 0.75, and 1.0 across all simulated flow rates, confirming the validity of the MCGG design under low and high Re conditions. This stability highlights a key advantage of the integrated SAR nodes: the resulting gradient is governed by the device architecture and volumetric balancing rather than the diffusion time.
To experimentally evaluate the performance of the designed MCCG, FITC-Dextran 10 kDa was dissolved in ultrapure water and introduced through one inlet, while Milli-Q water was introduced through the other. The experimental results, shown in Fig. 7(A, left), captured the dynamic formation of concentration lamellae and progressive dilution within the device, closely resembling the simulated profiles (Fig. 7(A), right). The SAR structures effectively split and recombined flow streams, confirming their functional role in gradient formation. Quantification of fluorescent intensity at the outlets was analyzed through the image analysis procedure described in section 2.4. Normalized outlet concentrations were obtained from the averaged fluorescence intensities measured at each outlet channel. The error bars in Fig. 7(B) represent the standard deviation derived from three random transects analyzed along the channel, as described in section 2.4. The fluorescence-based analysis yielded a coefficient of determination (R2) of 0.9984 relative to the simulation data, confirming the high reproducibility and accuracy of the experimental gradient generation.
A separate evaluation using bovine serum albumin (BSA) as a model macromolecule was performed to assess the generator's performance with biomolecules of lower diffusivity. As described in section 2.4, outlet samples were collected and analyzed using the bicinchoninic acid (BCA) assay. Each BSA concentration value represents the average of triplicate measurements obtained by spectrophotometric analysis, with error bars indicating the corresponding standard deviation. The quantified BSA concentrations (Fig. 7(B)) followed the expected stepwise gradient, showing excellent correlation with the simulated values, with a coefficient of determination (R2) of 0.9972. These results confirm that the MCGG can reliably generate well-defined and reproducible concentration gradients for both fluorescent tracers and protein solutions.
Fig. 7(C) represents the experimentally measured normalized concentrations, obtained from fluorescence intensities at each outlet, across Re from 0.1 to 100. The results show five consistent output levels that closely match the expected gradient values (0, 0.25, 0.5, 0.75, and 1.0), confirming that the MCGG maintains stable and reproducible gradients over the full range of tested flow conditions. These results validate the SAR-integrated MCGG as a reliable platform for rapid generation of discrete concentration gradients under laminar flow conditions. The system consistently delivers target outputs, at various flow rates, enabled by both the hydraulic resistor network and the enhanced interfacial mixing provided by the SAR architecture.
Compared with existing microfluidic gradient generators, our design emphasizes a compact 3D-printed architecture, SLA DLP-enabled rapid prototyping, and precision output control through embedded SAR units. Together, these results highlight the potential of low-cost, cleanroom-free SLA DLP 3D printing to enable next-generation microfluidic platforms for the reproducible and scalable generation of chemical gradients.
The proposed cleanroom-free process enables the rapid production of optically transparent, enclosed microchannels and integrated ports in a single fabrication step, reducing total production time to under 1.5 h using a desktop SLA DLP printer. This workflow eliminates molds and bonding steps, simplifying the transition from digital design to functional microfluidic devices. The demonstrated mixing performance and gradient reproducibility highlight the potential of SLA DLP additive manufacturing as an accessible method for fabrication of such microfluidic systems.
Overall, this work establishes an efficient and versatile route for the rapid prototyping of lab-on-a-chip devices that integrate advanced micromixing strategies and gradient generation.