An implantable ionic liquid-gel microelectrode for in vivo monitoring of K+ levels in the living rat brain†
Received
28th February 2025
, Accepted 30th March 2025
First published on 31st March 2025
Abstract
The exploration of new interfaces for in vivo analysis holds great promise for electrochemical acquisition of chemical signals involved in brain events. In this study, we designed and created an implantable electrode using a liquid/liquid (L/L) interface concept to monitor in vivo variations of K+ in the living brain. Poly(1-butyl-3-vinylimidazolium bis(trifluoro-methylsulfonyl)imide) (PB) and an ionic liquid of 1-decyl-3-methylimidazolium bis(trifluoromethanesulfonyl)-imide (C10M) were optimized to form a uniform ionic liquid gel (ILG), which exhibits a wide potential window and remarkably enhances interfacial mechanical stability. Furthermore, the specific ionophore [2.2.3]-triazacryptand (TAC) was optimized and incorporated into the ILG (ILG-TAC) to molecularly tailor the micro-interface between the gel phase and water phase. The developed implantable ILG-TAC electrode demonstrated high selectivity for K+, and good anti-biofouling capability with a signal deviation less than 8.5% over 50 days of continuous implantation. This ion-transfer-based sensing strategy introduces a novel approach for in vivo analysis, especially beneficial for detecting redox inactive species.
Introduction
The electrochemical technique using implantable microelectrodes has emerged as a pivotal and promising approach for acquiring in vivo chemical signals in the brain, aiding in the exploration of molecular mechanisms underlying brain events.1,2 Widely used sensing strategies rely on solid electrode/electrolyte interfaces, such as carbon fiber and noble-metal microelectrodes, which generate faradaic current signals via electron transfer processes.3–6 Nonetheless, these sensing models frequently encounter limitations in detecting electrochemically inactive species, as these species are incapable of undergoing redox reactions within the water-splitting potential window.6
The interface between two immiscible electrolyte solutions, referred to as the liquid/liquid (L/L) interface supported by a pipette, offers significant advantages for the direct analysis of ionic species.7–10 This interface operates based on the difference in solvation energies of ions in two adjacent phases, generating ionic current signals and characteristic transfer potentials for discrimination.11–14 Unlike conventional electrochemical sensors, L/L interfacial systems depend on ion transfer rather than electron transfer, enabling the identification and quantification of non-electroactive species. Moreover, the L/L interface supported by a pipette allows for high spatial resolution for in vivo analysis by reducing the tip size to the micrometer scale.15–19 Despite its conceptual robustness, the application of the L/L interface for in vivo analysis is still in its early stages.16,20 For instance, Zhang et al. employed a 1,2-dichloroethane (1,2-DCE)/water interface sensor to detect choline in the brain, demonstrating its potential feasibility for in vivo analysis.21 However, the inherent mechanical instability of the L/L interface often poses a risk of organic solvent leakage, which can induce biological toxicity and impair brain function.22 To address this, Gu et al. recently developed an enhanced L/L interfacial iontronics by using polyvinylchloride (PVC) to improve the interfacial stability of the organic phase.23 A potential shortcoming is that the low conductivity of PVC may increase the resistance of the organic phase, resulting in elevated background current signals.22,24 Additionally, in the complex brain environment where numerous ionic interferences coexist,25 the analytical model solely based on the inherent ion transfer potentials at the L/L interface faces significant limitations.
Motivated by these challenges, we designed and developed an implantable L/L interfacial sensor supported by a micropipette filled with an ionic liquid-gel (ILG) for in vivo monitoring of extracellular K+ in the living rat brain (Scheme 1). A hydrophobic ionic liquid, 1-decyl-3-methylimidazolium bis(trifluoromethanesulfonyl)imide (C10M), was optimized as the organic phase. Simultaneously, conductive poly(1-butyl-3-vinylimidazolium bis(trifluoro-methylsulfonyl)imide) (PB), instead of PVC, was used to enhance the mechanical stability of the C10M/W interface by forming an organic gel. It is important to note that the traditional L/L interfaces require the addition of an organic supporting electrolyte at relatively high concentrations (∼10−3 M to 10−2 M) to ensure sufficient electrical conductivity of the organic phase.13 The use of conductive ILs as the organic phase eliminates the requirement for supplementary organic supporting electrolytes, thus simplifying the fabrication process of interfacial sensors. By molecularly tailoring the ILG/water interface using the specific K+ ionophore, [2.2.3]-triazacryptand (TAC)(ILG-TAC/W), we achieved selective facilitation of K+ transfer from the aqueous phase to the ILG phase. The developed ILG-TAC electrode exhibited remarkable anti-fouling capability, maintaining a current signal decrease of less than 8.5% even after 50 days of immersion in bovine serum albumin solution. Eventually, the resulting ILG-TAC/W interfacial sensor was successfully implanted into the cerebral cortex of rat brains, and in vivo variation of K+ in rat brains was selectively determined with hyperkalemia models.
 |
| Scheme 1 (a) Working principle of the molecularly tailored ILG-TAC/W interface supported by a micropipette for facilitating the K+ transfer process. (b) In vivo determination of K+ in the living rat brain with acute and chronic hyperkalemia models using the developed ILG-TAC electrode. | |
Results and discussion
The ionophores of [2.2.3]-triazacryptand (TAC), C2, and dibenzo-18-crown-6 (C3, a commonly used ionophore towards K+) were taken as specific recognition molecules for K+. Among them, TAC and C2 were synthesized as reported previously (Scheme S1, ESI†),26,27 and characterized by 1H NMR and 13C NMR spectroscopy (Fig. S1†). Moreover, PB was prepared and characterized by 1H NMR and FT-IR spectroscopy (Fig. S2†).28 The primary criterion for selecting the IL was its high hydrophobicity, as this property enables the formation of a well-separated L/L interface with aqueous solution. We selected three hydrophobic ILs with the bis(trifluoromethanesulfonyl)imide anion, 1-ethyl-3-methylimidazolium bis(trifluoromethyl-sulfonyl)imide (C2M), 1-butyl-3-methylimidazolium bis(trifluoromethyl-sulfonyl)-imide (C4M), and 1-decyl-3-methylimidazolium bis(trifluoromethanesulfonyl)imide (C10M) as the organic phase to form C2M/W, C4M/W, and C10M/W interfaces with aqueous solution, respectively. The C10M/W micro-interface exhibited the widest polarized potential window (∼800 mV) as determined by cyclic voltammetry (CV) (Fig. S3†), and hence was chosen to fabricate the uniform ILG with PB at the mass ratio of 1
:
1 (Fig. 1a and b). When the pipette filled with ILG was left standing vertically for 30 min, the ILG level within the pipette showed no noticeable change (Fig. 1d). In contrast, the liquid level of pure C10M in the pipette moved about 5 mm within 10 s, resulting in droplet formation at the pipette tip (Fig. 1c). These results indicate that the incorporation of PB efficiently enhanced the mechanical stability of the ILG interface, which is critical for subsequent implantation testing in the brain. Then, the micropipette was prepared and characterized by scanning electron microscopy (SEM), exhibiting a round orifice with a radius of ∼2.5 μm (Fig. 1e(i)). The ionophore of TAC was dissolved in the organic gel to tailor the ILG/W micro-interface for K+ detection via specific chemical binding. In order to ensure the position of the L/L interface, fluorescein was added to the gel phase. As shown in Fig. S4,† the fluorescence signal was clearly observed along the micropipette, indicating that the gel filled the micropipette uniformly. Furthermore, the ILG-TAC was further demonstrated to completely fill the micropipette tip by microscopy (Fig. 1e(ii)). The polarized potential window remained almost unchanged after TAC was incorporated into ILG in blank aqueous solution, indicating that the addition of TAC had negligible effect on the initial potential window (Fig. 1f).
 |
| Fig. 1 (a and b) Typical photographs of C10M and ILG formed by C10M and PB. (c and d) (i) Upright and (ii) inverted positions of C10M and ILG in a pipette. (e) (i) SEM image of a micropipette, and (ii) confocal image of the micropipette filled with ILG-TAC. (f) CV curves of ILG (i) and ILG-TAC (ii) electrodes using cell 1: Ag|AgCl|ILG‖5 mM NaCl|AgCl|Ag. (g) CV curves recorded at the three ionophores tailored ILG/W micro-interface ((i) ILG-TAC/W, (ii) ILG-C3/W, (iii) ILG-C2/W and (iv) ILG/W) using electrochemical cell 2: Ag|AgCl|ILG + TAC‖x mM K+|AgCl|Ag ((i–iii) x = 5; (iv) x = 80), scan rate: 10 mV s−1. (h) Gibbs free energies of the complexation reaction between C2, C3 and TAC and K+ obtained by DFT simulations. (i) DPV curves obtained at the ILG-TAC electrode in the presence of different K+ concentrations (CK+: 0, 0.8, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50 and 60 mM). (j) Ip values obtained at the ILG-TAC electrode at different K+ concentration. (k) Comparison of Ip values at the ILG-TAC and the PVC-TAC electrodes under different K+ concentrations. | |
To verify the K+ recognition capability of the ILG-TAC/W interface, the voltammetric behaviors were recorded using electrochemical cells 1–2. As illustrated in Fig. 1g, a well-defined positive current with a half-wave potential of ∼0.26 V was observed at the ILG-TAC electrode in aqueous solution containing 5 mM K+ (curve i), while no significant current response was observed at the ILG/W interface, even though the K+ concentration was enhanced to 80 mM (curve iv). Thus, such current response at the ILG-TAC electrode was ascribed to the transfer of K+ facilitated by TAC. The transfer potentials of facilitated ion transfer were calibrated using the tetramethylammonium cation (TMA+) as an internal reference ion. In this case, the amount of TAC (200 mM) exceeds that of K+, making the diffusion of K+ the rate-determining step of facilitated ion transfer. The forward steady-state wave seen in the voltammetric curve is controlled by the hemispherical diffusion of K+ from the aqueous phase to the ILG-TAC phase by the following chemical reaction:
K+ (W) + TAC (ILG) → K-TAC+ (ILG) |
During the reverse scan, the voltammetric curve indicates the hemispherical diffusion of the [K-TAC]+ complex away from the ILG-TAC/W interface. From a thermodynamic perspective, the half-wave potential
for the reaction between K+ and TAC in the case of a 1
:
1 (n = 1) mass ratio can be expressed as:
where

represents the formal potential of K
+ transfer across the ILG/W interface,
z denotes the transferred charge of K
+,
βTAC is the association constant in the ILG phase, and [K
+] indicates the concentration of K
+ in the aqueous phase. Moreover, the electrochemical performance of the ILG-TAC is less affected by the silanization of the micropipette (Fig. S5
†). By comparison, a lower current response was observed at more positive

(∼0.31 V) when using C3 as ionophore in ILG (ILG-C3), but no obvious current response was obtained at the C2 tailored ILG/W interface (ILG-C2/W) (curves ii–iii,
Fig. 1g). These results suggested that TAC is a suitable ionophore to facilitate the transfer of K
+ at the ILG-TAC/W interface. Upon incorporating the ionophore into the ILG, density functional theory (DFT) calculations were performed to estimate the Gibbs free energy (Δ
G) values of three ionophores reacted with K
+. As shown in
Fig. 1h, the Δ
Gs of TAC, C2, and C3 were ∼−46.0 kcal mol
−1, ∼−15.0 kcal mol
−1, and ∼−30.6 kcal mol
−1, respectively. The most negative binding energy of TAC suggests that it has a stronger binding affinity with K
+ than C2 and C3. Furthermore, the molecular cavity size of TAC is estimated to be 2.90 Å, which aligns more closely with the ionic diameter of K
+ (∼2.66 Å) than those of C2 (∼4.32 Å) and C3 (∼3.2 Å). Coupled with additional oxygen binding sites in TAC, the selective facilitation of K
+ transfer was achieved at the ILC-TAC/W interface. By testing the transfer current at the ILG-TAC/W interface over different times, 4 min was optimized as the reaction time for K
+ determination (Fig. S6
†). Owing to the high sensitivity, differential pulse voltammetry (DPV) was conducted to study the variation of current responses in the aqueous phase containing different concentrations of K
+. The peak current (
Ip) increased with increasing K
+ concentration, exhibiting a linear relationship ranging from 0.8 to 60 mM, with a detection limit of 0.13 mM (
Fig. 1i and j). Moreover, the mass ratio of C
10M to PB was changed to 1
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
:
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
2 and 2
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
:
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
1, but the much narrower linear range was obtained under these conditions (Fig. S7
†). As a result, the 1
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
:
![[thin space (1/6-em)]](https://www.rsc.org/images/entities/char_2009.gif)
1 mass ratio of C
10M to PB was optimized to prepare the organic phase for K
+ determination. In fact, a popular approach to overcome the L/L interfacial stability is using polyvinyl chloride (PVC) as a gelling agent to solidify the organic phase.
29,30 We incorporated PVC into the 1,2-DCE phase with TAC, resulting in a PVC-TAC gel electrode. The electrochemical behavior of the PVC-TAC electrode was studied by DPV (Fig. S8
†). The same amount of K
+ was found to generate a lower current response at the PVC-TAC microelectrode than the ILG-TAC microelectrode (
Fig. 1k). These observations indicate that the ILG-TAC electrodes possess a wider linear range and higher sensitivity to K
+.
Electrochemical detection of chemical substances in the complex biological brain environment is widely used for the early diagnosis, prevention and treatment of diseases.31–33 However, a significant challenge in electrochemical sensing based on the current output arises when coexisting species (e.g. biomacromolecules and proteins) nonspecifically adsorb onto the polarized interface, resulting in the loss of current signals.34–36 Therefore, the anti-biofouling capability of the developed ILG-TAC/W interface is crucial for reliable in vivo analysis. Bovine serum albumin (BSA) was selected as a model protein to simulate the surface biofouling in rat brains.27,37 Surprisingly, no obvious fluorescence was observed at the interface after the ILG-TAC electrode was immersed in the fluorescein isothiocyanate(FITC)-labeled BSA (FITC-BSA) solution for 50 days (Fig. 2a). Moreover, the peak current exhibited a decrease of less than 8.5% (Fig. 2c and d(i)). Nevertheless, fluorescence was clearly observed at the 1,2-DCE/W interface supported by a micropipette after immersion in BSA solution for 15 h (Fig. 2b). Meanwhile, the current response to K+ was rapidly decreased by 44% (Fig. 2d(ii)). These results demonstrate that the ILG-TAC electrodes possess high anti-biofouling capability, suitable for in vivo analysis in the brain.
 |
| Fig. 2 (a) Fluorescence images (upper) and relative fluorescence intensity (lower) tested at the ILG-TAC electrode immersed in 5 mg per mL FITC-BSA solution for 0 d, 5 d, 10 d, 25 d, 50 d and 60 d. (b) Fluorescence images (upper) and relative fluorescence intensity (lower) obtained at the micropipette filled with 1,2-DCE which was immersed in 5 mg per mL FITC-BSA solution for 0 h, 2 h, 5 h, 10 h, 15 h and 20 h. (c) DPV curves of the ILG-TAC electrode for 5 mM K+ after immersion in 5 mg per mL BSA for 0 to 60 d. (d) Ip/I0 variations at the ILG-TAC electrode during immersion in BSA for 60 days (i). Ip/I0 at a micropipette filled with 1,2-DCE during immersion in BSA for 20 hours (ii). I0 represents the initial peak current obtained at the ILG-TAC electrode. | |
Selectivity is another important factor for in vivo analysis in the practical brain environment. To assess the selective detection capability of the ILG-TAC microelectrode for K+, selectivity tests were conducted against metal ions, anions, amino acids, and biological species that coexist in the rat brain. The amount of interference is decided referring to their concentrations in the practical brain as reported previously.11,12 Na+ and Ca2+ are identified as the main interferences due to their similar transfer potentials to that of K+ and high concentrations in the brain (Na+ at ∼100 mM and Ca2+ at ∼1 mM). As depicted in Fig. 3a, the interferences from Na+ and Ca2+ were less than 5.7% and 4.5%, respectively. For other species, including anions, amino acids and biologically relevant reagents, the interference was also negligible (<4.1%) (Fig. S9†). Moreover, slight changes (<6.1%) were observed in competition tests for K+ in the presence of other interferences (Fig. S10†), indicating a high selectivity of the ILG-TAC microelectrode for the detection of K+. DFT simulations further estimated the binding constant of TAC with K+ to be 5.8 × 1021, which is 3.4 × 1010 times greater than that of Na+, and 5.3 × 1015 times greater than that of Ca2+ (Fig. 3b). This strong binding affinity of TAC with K+ explains the high selectivity of the ILG-TAC electrode towards K+.
 |
| Fig. 3 (a) Selectivity tests of the ILG-TAC electrode for metal anions (Na+, Ca2+, Mg2+, Cu2+, Fe3+, Zn2+, Co2+, and Ni2+ at concentrations of 10 mM for Na+, 1 mM for Ca2+ and Mg2+, and 10 μM for other ions). (b) Binding constants between TAC and K+, Na+ and Ca2+ obtained from DFT simulations. (c) A schematic diagram of the implantation of the 1, 2-DCE-TAC electrode and ILG-TAC electrode filled with fluorescein in rat brains. (d and e) Typical CV curves obtained at the 1,2-DCE-TAC electrode (d) and ILG-TAC electrode (e) in aCSF solution before (i) and after (ii) implantation in the rat brain for 20 min. The fluorescence image of brain slices after implantation of the electrode for 5 h (iii). (Inset) Photographs of the interfaces filled in capillaries under conditions of (d) and (e), respectively. (f) GFAP staining images of coronal brain sections from the rat in the group implanted with the ILG-TAC electrode for 7 days. | |
The mechanical stability of the L/L interface supported by a micropipette is vital for in vivo acquisition of K+ in the brain (Fig. 3c). To address this, PB, a polymer of ionic liquid, was incorporated into C10M to fabricate gel microelectrodes. As expected, the ILG-TAC/W interface exhibited a stable polarized potential window even after being implanted in the brain for 24 h without obvious gel interfacial movement (Fig. 3e(i) and (ii)). At the same time, we added fluorescein in the gel phase to observe the moving of the gel interface. Once the organic gel flows out of the micropipette and contaminates the brain tissue, fluorescence signals will be observed on the brain slices. As displayed in Fig. 3e(iii), no fluorescence signal was observed in the brain slice, indicating no leakage of gel from the micropipette. These observations demonstrate that the ILG-TAC electrodes possess excellent interfacial stability, which is highly suitable for implantable determination. In contrast, the conventional 1,2-DCE/W interface supported by a micropipette exhibited a significant change of polarized potential window after 20 min of implantation in the brain (Fig. 3d). The liquid level of 1,2-DCE in the micropipette also shifted toward the orifice, reducing the liquid length within the micropipette from 17 mm to 13 mm (Fig. 3d(i) and (ii)). Fluorescein was used to label the 1,2-DCE phase, and fluorescence was observed at the border of the brain slice using confocal microscopy, revealing the leakage of 1,2-DCE into the brain tissue (Fig. 3d(iii)).
After mechanical injury to the brain, astrocytes proliferate abnormally, leading to increased expression of glial fibrillary acidic protein (GFAP).38 Thus, GFAP expression in the rat brain serves as a reliable indicator of tissue damage caused by implanted microelectrodes. To assess the inflammatory response induced by the implantation of the ILG-TAC microelectrodes, GFAP expression in the rat brain was evaluated. As displayed in Fig. 3f, only a little fluorescence response of GFAP expression was observed at the implantation site after 7 days compared to the initial situation (Fig. S11†). In combination with the results of 2,3,5-triphenyltetrazolium chloride (TTC) staining of brain tissue slices (Fig. S12†), the ILG-TAC microelectrodes demonstrated slight tissue damage and excellent biocompatibility in vivo. Therefore, the ILG-TAC electrode, with better mechanical stability and sensitivity for K+, was established for implantable K+ analysis in the brain.
In a state of hyperkalemia, the elevated potassium ion levels might interfere with the normal electrophysiological processes of neurons, leading to heightened neuronal excitability and potentially disrupting neural function39,40 Based on the promising electrochemical performance, we implanted the ILG-TAC microelectrodes in the cerebral cortex to in vivo track K+ in rat brains (Fig. 4a). In this study, acute and chronic hyperkalemia models were constructed. We constructed a rat model of acute hyperkalemia by continuous tail vein infusion of saline containing 2% KCl. Compared to the electrocardiogram (ECG) of normal rats, the QRS complex peak voltage in the ECG of acute hyperkalemia model rats sharply decreased from 0.78 mV to 0.37 mV, while the peak voltage of the T-wave increased from 0.19 mV to 0.54 mV (Fig. S13†). This significant rise in T-wave amplitude is indicative of the severe electrolyte imbalance associated with acute hyperkalemia. The plasma K+ concentration in the acute hyperkalemia model rats was further measured using the ILG-TAC microelectrode. The current response was found to be greater in the brain of an acute hyperkalemia rat compared to that of a normal rat (Fig. 4d), indicating that the plasma K+ concentration in the terminal blood increased from 4.82 mM to 7.56 mM (Fig. 4e). These results indicate that the acute hyperkalemia model rats were successfully established. Subsequently, we used the ILG-TAC microelectrodes to track the K+ concentration in rat brains. After stopping the injection of 2% KCl solution for 35 min, the K+ concentration in the cerebral cortex began to rise slowly, increasing by approximately 23.6% within one hour (Fig. 4c). Simultaneously, we recorded local field potential (LFP) signals from the brains of acute hyperkalemia model rats. The intensity of neuronal LFP signals started to decrease significantly, which may be due to the increase in extracellular K+ concentration leading to suppression of neuronal activity (Fig. 4c). In contrast, in control rats injected only with saline, there was no significant change in K+ concentration in the cerebral cortex, and the corresponding LFP signals indicated that neuronal activity remained almost unchanged (Fig. 4b). These results suggest that the extracellular K+ amount in the rat brain is susceptible to acute changes in plasma K+ levels.41,42
 |
| Fig. 4 (a) Schematic diagram of the ILG-TAC electrode implanted in the cerebral cortex for in vivo detection of K+. (b and c) LFP recording (upper) and in vivo measurements of K+ levels (lower) in (b) normal and (c) acute hyperkalemia rat brains. (d) DPV curves of the ILG-TAC electrode in the cortex of normal and acute hyperkalemia rats. (e) The concentrations of K+ in the plasma of normal and acute hyperkalemia rats. (f) LFP recording in (i) normal and (ii) chronic hyperkalemia rats. (g) In vivo measurements of K+ levels in the cortex of the rat brain with chronic hyperkalemia for 14 days. (h) DPV curves of the ILG-TAC electrode in the cortex of normal and chronic hyperkalemia rats. (i) The concentrations of K+ in the plasma of normal and chronic hyperkalemia rats. | |
On the other hand, the ILG-TAC microelectrode was employed to study the variations of K+ levels in rat brains followed by chronic hyperkalemia. The chronic hyperkalemia model was achieved through unilateral nephrectomy and a diet supplemented with 20% KCl over a 14-day period. The ECG revealed a decrease in the QRS peak voltage from 0.78 to 0.49 mV (Fig. S14†). Additionally, peaked T-waves were observed in the ECG of chronic hyperkalemia rats, with the peak voltage increasing from 0.19 to 0.28 mV. Despite a 2.51 mM increase in plasma K+ concentration in rats with chronic hyperkalemia (Fig. 4i), only minimal changes were noted in both cortical K+ levels and spontaneous local field potential (LFP) activity (Fig. 4f–h). These results suggest that chronic alterations in plasma K+ levels do not lead to significant fluctuations in K+ concentrations within the brain. This regulatory mechanism may be mediated by adaptive modulation of K+ influx or extrusion processes at the blood–brain barrier in response to chronic changes in plasma K+ levels. However, it appears that the extent of such adaptive responses is markedly reduced during acute changes in plasma potassium.43
Conclusions
In summary, a novel ionic liquid gel electrode was designed and developed based on the principles of L/L electrochemistry. This innovative approach enables current measurements through an ion transfer mechanism, providing a promising platform for detecting various ionic species, including electrochemically inactive species. More importantly, the developed ILG electrode demonstrated remarkable interfacial stability and excellent anti-biofouling properties, both of which are critical for in vivo applications. Furthermore, the successful functionalization of the ILG electrode with specific ionophores for K+ recognition significantly enhanced its selectivity. This advancement fulfills the stringent requirements for selective K+ measurement in complex brain environments. These features allow the application of the L/L interface concept to implantable sensors within the real brain system. This study introduces a new strategy for the in vivo acquisition of chemical expressions within complex brain systems. It is worth noting that the present methodology can be easily extended to detect other chemical species, particularly non-electroactive ions, by incorporating appropriate ionophores into the organic phase. This flexibility not only broadens the applicability of the L/L electrode but also paves the way for future research in the field of biosensing and neurochemistry.
Ethical statement
All animal experiments were performed according to the guidelines of the Care and Use of Laboratory Animals formulated by the Ministry of Science and Technology of China and were approved by the Animal Care and Use Committee of East China Normal University (Shanghai, China).
Data availability
The authors confirm that the data supporting the findings of this study are available within the article and its ESI.†
Author contributions
L. M. Zhang and T. Liu conceived and designed the experiments, while Z. H. Zhang and Z. H. Chen performed the experiments and analyzed the data. All authors co-wrote the manuscript.
Conflicts of interest
There are no conflicts to declare.
Acknowledgements
This work was supported by the Natural Science Foundation of China (22022402, 21974051 for L. M. Zhang), the National Key Research and Development Program of China (2022YFF0710000), and the Postdoctoral Fellowship Program of CPSF under Grant Number GZC20240477. We also gratefully acknowledge the support from the East China Normal University Multifunctional Platform for Innovation (004).
Notes and references
- X. Chai, X. Zhou, A. Zhu, L. Zhang, Y. Qin, G. Shi and Y. Tian, Angew. Chem., Int. Ed., 2013, 52, 8129–8133 CrossRef CAS PubMed.
- H. Dong, Q. Zhou, L. Zhang and Y. Tian, Angew. Chem., Int. Ed., 2019, 5, 13948–13953 CrossRef PubMed.
- Y. Wang, Y. Qian, L. Zhang, Z. Zhang, S. Chen, J. Liu, X. He and Y. Tian, J. Am. Chem. Soc., 2023, 145, 2118–2126 CrossRef CAS PubMed.
- T. S. Davis, H. A. C. Wark, D. T. Hutchinson, D. J. Warren, K. O'Neill, T. Scheinblum, G. A. Clark, R. A. Normann and B. Greger, J. Neural. Eng., 2016, 13, 36001 CrossRef CAS PubMed.
- H. Wei, L. Li, J. Jin, F. Wu, P. Yu, F. Ma and L. Mao, Anal. Chem., 2020, 14, 10177–10182 CrossRef PubMed.
- A. A. Muhammad, N. R. Logan, G. R. J. John and P. I. Pedro, IEEE Trans. Neural Syst. Rehabil. Eng., 2019, 27, 1724–1731 Search PubMed.
- D. Zhan, S. Mao, Q. Zhao, Z. Chen, H. Hu, P. Jing, M. Zhang, Z. Zhu and Y. Shao, Anal. Chem., 2004, 76, 4128–4136 CrossRef CAS PubMed.
- N. T. Iwai, M. Kramaric, D. Crabbe, Y. Wei, R. Chen and M. Shen, Anal. Chem., 2018, 90, 3067–3072 CrossRef CAS PubMed.
- S. Wilke, R. Schürz and H. Wang, Anal. Chem., 2001, 73, 1146–1154 CrossRef CAS.
- X. Huang, L. Xie, X. Lin and B. Su, Anal. Chem., 2016, 88, 6563–6569 CAS.
- Q. Li, S. Xie, Z. Liang, X. Meng, S. Liu, H. H. Girault and Y. Shao, Angew. Chem., Int. Ed., 2009, 48, 8010–8013 CAS.
- M. Welle, K. Alanis, M. L. Colombo, J. V. Sweedler and M. Shen, Chem. Sci., 2018, 9, 4937–4941 RSC.
- P. J. Rodgers, P. Jing, Y. Kim and S. Amemiya, J. Am. Chem. Soc., 2008, 130, 7436–7442 CrossRef CAS PubMed.
- L. Xie, X. Huang and B. Su, ACS Sens., 2017, 2, 803–809 CrossRef CAS PubMed.
- M. L. Colombo, J. V. Sweedler and M. Shen, Anal. Chem., 2015, 87, 5095–5100 Search PubMed.
- S. Liu, Q. Li and Y. Shao, Chem. Soc. Rev., 2011, 40, 2236–2253 RSC.
- P. He, Y. Shao, Z. Yu, X. Liang, J. Liu, Y. Bian, Z. Zhu, M. Li, C. M. Pereira and Y. Shao, Anal. Chem., 2022, 94, 9801–9810 CrossRef CAS PubMed.
- H. Wang, Y. Ruan, L. Zhu, X. Shi, W. Zhao, H. Chen and J. Xu, Angew. Chem., Int. Ed., 2021, 133, 13352–13358 CrossRef.
- W. Ma, W. Xie, R. Tian, X. Zeng, L. Liang, C. Hou, D. Huo and D. Wang, Sens. Actuators, B, 2023, 377, 133075 Search PubMed.
- S. H. Huang, M. Parandhaman, S. Farnia, J. Kim and S. Amemiya, Chem. Commun., 2023, 59, 9575–9590 RSC.
- X. Wang, T. Xu, Y. Zhang, N. Gao, T. Feng, S. Wang and M. Zhang, ACS Sens., 2021, 6, 2757–2762 CrossRef CAS PubMed.
- J. D. Mann, A. B. Butler, J. E. Rosenthal, C. J. Maffeo, R. N. Johnson and N. H. Bass, Ann. Neurol., 1978, 3, 156–165 CrossRef CAS PubMed.
- C. Gu, F. Kong, S. Liang, X. Zhao, B. Kong, T. Jiang, J. Yu, Q. Li, Y. Lin, S. Bai and Y. Shao, Sci. Adv., 2024, 10, eadr7218 CrossRef CAS.
- C. Gu, X. Nie, J. Jiang, Z. Chen, Y. Dong, X. Zhang, J. Liu, Z. Yu, Z. Zhu, J. Liu, X. Liu and Y. Shao, J. Am. Chem. Soc., 2019, 141, 13212–13221 Search PubMed.
- L. Zhang and Y. Tian, Acc. Chem. Res., 2018, 51, 688–696 Search PubMed.
- H. He, M. A. Mortellaro, M. J. P. Leiner, R. J. Fraatz and J. K. Tusa, J. Am. Chem. Soc., 2003, 125, 1468–1469 CrossRef CAS.
- F. Zhao, Y. Liu, H. Dong, S. Feng, G. Shi, L. Lin and Y. Tian, Angew. Chem., Int. Ed., 2020, 59, 10426–10430 CrossRef CAS.
- J. Zhao, X. Shen, F. Yan, L. Qiu, S. Leec and B. Sun, J. Mater. Chem., 2011, 21, 7326 RSC.
- C. I. Cámara, C. A. Bornancini, J. L. Cabrera, M. G. Ortega and L. M. Yudi, Talanta, 2010, 83, 623–630 Search PubMed.
- R. P. Buck and E. Lindner, Acc. Chem. Res., 1998, 31, 257 CrossRef CAS.
- J. S. Río, O. Y. F. Henry, P. Jolly and D. E. Ingber, Nat. Nanotechnol., 2019, 14, 1143–1149 Search PubMed.
- S. S. Timilsina, P. Jolly, N. Durr, M. Yafia and D. E. Ingber, Acc. Chem. Res., 2021, 18, 3529–3539 CrossRef PubMed.
- M. Karaboğ and M. K. Sezgintürk, J. Pharm. Biomed. Anal., 2022, 209, 114479 CrossRef PubMed.
- Y. Liu, Z. Liu, F. Zhao and Y. Tian, Angew. Chem., Int. Ed., 2021, 60, 14429–14437 CrossRef CAS PubMed.
- C. Zhang, Z. Liu, L. Zhang, A. Zhu, F. Liao, J. Wan, J. Zhou and Y. Tian, Angew. Chem., Int. Ed., 2020, 59, 20499–20507 CAS.
- L. Liu, F. Zhao, W. Liu, T. Zhu, J. Zhang, C. Chen, Z. Dai, H. Peng, J. Huang, Q. Hu, W. Bu and Y. Tian, Angew. Chem., Int. Ed., 2017, 56, 10471–10475 Search PubMed.
- T. Feng, W. Ji, Y. Zhang, F. Wu, Q. Tang, H. Wei, L. Mao and M. Zhang, Angew. Chem., Int. Ed., 2020, 59, 23445–23449 CAS.
- A. Golabchi, B. Wu, B. Cao, C. J. Bettinger and X. T. Cui, Biomaterials, 2019, 225, 119519 CrossRef CAS PubMed.
- M. Douglas, D. Rizzolo and D. Kruger, Clin. Rev., 2017, 27, 40–49 Search PubMed.
- A. F. Andrew, C. T. Robert, Y. G. Eugene and A. F. Robert, World Neurosurg., 2016, 88, 32–35 Search PubMed.
- R. Rasmussen, E. Nicholas, N. C. Petersen, A. G. Dietz, Q. Xu, Q. Sun and M. Nedergaard, Cell Rep., 2019, 28, 1182–1194 CAS.
- H. Monai, S. Koketsu, Y. Shinohara, T. Ueki, P. Kusk, N. L. Hauglund and A. J. Samson, Sci. Rep., 2021, 11, 8150 CAS.
- W. Stummer, A. L. Betz and R. F. Keep, J. Cereb. Blood Flow Metab., 1995, 15, 336–344 CrossRef CAS.
|
This journal is © The Royal Society of Chemistry 2025 |
Click here to see how this site uses Cookies. View our privacy policy here.