Laura
Suárez-Criado
a,
Eduardo
Bolea-Fernandez
bc,
Lana
Abou-Zeid
b,
Mathias
Vandermeiren
b,
Pablo
Rodríguez-González
a,
Jose Ignacio
Garcia Alonso
a and
Frank
Vanhaecke
*b
aUniversity of Oviedo, Department of Physical and Analytical Chemistry, Julián Clavería 8, 33006-Oviedo, Spain
bGhent University, Department of Chemistry, Atomic & Mass Spectrometry – A&MS Research Unit, Campus Sterre, Krijgslaan 281 – S12, 9000 Ghent, Belgium. E-mail: Frank.Vanhaecke@UGent.be
cUniversity of Zaragoza, Aragón Institute of Engineering Research (I3A), Department of Analytical Chemistry, Zaragoza, Spain
First published on 15th January 2024
High-precision determination of the isotopic composition of mercury (Hg) is of paramount importance for unraveling its biogeochemical cycle and for identifying the origin of Hg in environmental compartments. Cold vapor generation multi-collector inductively coupled plasma-mass spectrometry (CVG-MC-ICP-MS) is the standard approach for such application. Cold vapor generation provides a high Hg introduction efficiency into the ICP, while chromatographic Hg isolation is not required as a result of the selective reaction between Hg2+ and SnCl2. For environmental or biota samples with low Hg concentrations, however, this approach still presents challenges and reliable measurements typically require a Hg concentration ≥1 μg L−1 in the solution analyzed. Recent improvements of MC-ICP-MS instrumentation, including the introduction of the so-called Jet interface and 1013 Ω Faraday cup amplifiers, enhance the signal-to-noise ratio. In this study, it was investigated to what extent this allows Hg isotopic analysis at lower concentration. Performance in Hg isotopic analysis was compared using two different sets of cones (standard vs. Jet), two plasma conditions (wet vs. dry) and two amplifier types (1011 Ω vs. 1013 Ω). Satisfactory accuracy and precision were achieved at a Hg concentration down to 0.1 μg L−1 in the solution measured when using Jet cones, dry plasma conditions, and the four available 1013 Ω amplifiers. The uncertainty expressed as 2SD for the δ202Hg values measured for the in-house standard solution was ±0.2‰ at 0.25 μg Hg L−1 and ± 0.3‰ at 0.1 μg Hg L−1. The method was subsequently applied to the analysis of real surface water samples contaminated with toxic metals.
Traditionally, high-precision Hg isotopic analysis is carried out using multi-collector inductively coupled plasma-mass spectrometry (MC-ICP-MS) using cold vapor generation (CVG) for sample introduction.6 CVG provides a very high analyte introduction efficiency owing to the highly efficient reaction between Hg2+ and the reductant (typically SnCl2 or NaBH4), theoretically enabling an analyte introduction efficiency of (nearly) 100% into the ICP ion source.7 Moreover, chromatographic isolation of Hg prior to isotopic analysis using MC-ICP-MS is not necessary when using CVG with SnCl2 as separation of Hg from the concomitant matrix is accomplished in the CVG device. In general, previous studies reporting Hg isotopic compositions were limited to samples with a Hg concentration >1 μg L−1 due to insufficient precision at lower analyte concentration, although the measurement of the Hg isotopic composition of marine samples at a concentration level of 0.25 μg L−1 using 1012 Ω Faraday cup amplifiers has been reported on.8
CVG-MC-ICP-MS enables accurate and precise Hg isotopic analysis at concentration levels down to 1 μg L−1 Hg, but the analysis of samples with an even lower Hg concentration level remains a challenging task. To overcome these issues, some studies have used preconcentration strategies using a chromatographic technique,9,10 a gold trap,11 or activated carbon trap.12 However, these preconcentration procedures make Hg isotopic analysis even more labor-intensive and enhance the risk of potential Hg isotope fractionation during the sample preparation protocol. Alternatively, also modification of the CVG unit has been reported to lead to higher sensitivity, thus enabling analysis of samples at a concentration level of 0.1 μg L−1 of Hg.13
Recent instrumental developments have enhanced the sensitivity of MC-ICP-MS instrumentation. First, the ion transmission efficiency was improved by enhancing the ion extraction efficiency, which is influenced by extraction and acceleration voltages, geometries of sampling and skimmer cones and interface pressure. For example, Thermo Fisher Scientific equipped their MC-ICP-MS instruments with a so-called ‘Jet interface’, that uses a high-performance pump to improve the interface vacuum, and geometrically modified sampling and skimmer cones (Jet sampler and X-type skimmer).14 Secondly, precision and accuracy of isotope ratios can be compromised by low-intensity ion beams, with the signal-to-noise ratio being affected by electronic baseline noise from the resistor in the amplifier feedback loop.15 Previously, for monitoring low-intensity ion beams, secondary electron multipliers (SEM) operated in ion counting mode were preferred over Faraday cup detectors, as SEMs are not affected by noise issues. However, SEMs suffer from variability in detection efficiency over time and beam intensity,15 while also requiring correction for detector dead time. This correction introduces an additional source of uncertainty.16,17 More recently however, 1013 Ω Faraday cup amplifiers instead of standard 1011 Ω ones were introduced, theoretically resulting in a 100-fold increase in output voltage and a 10-fold increase in the noise only, enhancing the signal-to-noise ratio by an order of magnitude, or an experimentally documented factor of 5–10.18,19
To the best of our knowledge, the combination of a Jet interface and 1013 Ω amplifiers has not been previously evaluated for measuring Hg isotope ratios in samples with low Hg concentrations (<1 μg L−1). In this study, the accuracy and precision attainable in Hg isotopic analysis using CVG-MC-ICP-MS were evaluated with different combinations of standard or Jet cones, with 1011 Ω or 1013 Ω amplifiers, and with wet (achieved through the continuous nebulization of an aqueous Tl isotopic standard for mass bias correction) and dry plasma conditions. An optimized protocol for analyzing the Hg isotopic composition in samples with low Hg content (<1 μg L−1) is proposed. The methodology was validated by analyzing the secondary standard reference material UM-Almaden and applied to surface water samples from three locations contaminated with toxic metals.
Standard solutions of Hg (NIST SRM 3133) and Tl (NIST SRM 997) from the National Institute of Standards and Technology were used for external and internal correction for the mass bias caused by instrumental mass discrimination, respectively. A previously characterized in-house (IH) standard solution of Hg (Inorganic Ventures, The Netherlands, Lot: F2-HG02105) was used for optimization, method development and validation purposes.6 The UM-Almaden secondary reference material (NIST RM 8610) was also used for assessing the accuracy and precision in Hg isotope ratio measurements.
A combination of a100 μL min−1 PFA concentric nebulizer and a spray chamber with cyclonic and Scott-type sub-units was used for continuous nebulization of a 5 μg L−1 Tl standard solution relied on for internal correction for instrumental mass discrimination. Fig. 1 provides a diagram of the instrumental set-up used in this work, illustrating the mixing of the sample flows obtained by CVG (Hg) and pneumatic nebulization (Tl) in a ‘T piece’.
The operating conditions of the Neptune XT MC-ICP-MS unit are summarized in Table 1. The Faraday cups L3, L2, L1, C and H1 were used to monitor the signals of the Hg isotopes 198Hg, 199Hg, 200Hg, 201Hg and 202Hg, respectively, while the Faraday cups H2 and H3 were used to monitor the signals for 203Tl and 205Tl (under wet plasma conditions only). Different amplifier configurations were evaluated as indicated in Table 1: (i) 1011 Ω amplifiers for each Faraday cup, (ii) 1011 Ω amplifiers for Faraday cups L2, L1, H1, H2 and H3 and 1013 Ω ones for Faraday cups L3 and C, and (iii) 1011 Ω amplifiers for Faraday cups L1, H2 and H3 and 1013 Ω ones for Faraday cups L3, L2, C and H1. During the course of this work, two additional 1013 Ω amplifiers (resulting in a total of 4) were installed, which explains why conditions (ii) and (iii) were evaluated. Prior to every measurement session, the software automatically performed a routine gain calibration and baseline integration during approximately 630 s. The use of Tau correction was also explored to address amplifier decay disparities since this could be more relevant in the case of the 1013 Ω amplifiers.20 The correction was carried out automatically using the MC-ICP-MS software.
Neptune XT MC-ICP-MS | ||||
---|---|---|---|---|
Instrument settings | Wet plasma conditions | Dry plasma conditions | ||
RF power (W) | 1250–1300 | 1250–1300 | ||
Cool gas flow rate (L min−1) | 15 | 15 | ||
Auxiliary gas flow rate (L min−1) | 1 | 1 | ||
Nebulizer gas flow rate (L min−1) | 0.70–0.75 | 0.97–0.98 | ||
Carrier gas flow rate (L min−1) | 0.25–0.30 | 0.14–0.15 | ||
Additional gas flow rate (L min−1) | 0.03–0.04 | — | ||
Sample uptake rate (mL min−1) | 0.7 | 0.7 | ||
Tl standard solution uptake Tl rate (mL min−1) | 0.17 | — | ||
SnCl2 solution uptake rate (mL min−1) | 0.7 | 0.7 | ||
Uptake time (s) | 100 | 100 | ||
Wash time (s) | 150 | 150 | ||
Mass resolution | Low resolution | Low resolution | ||
Mode | Static mode | Static mode | ||
Sampling cone | Standard | Jet cone | Standard | Jet cone |
Skimmer cone | H-type | X-type | H-type | X-type |
Total mercury (THg) concentrations were determined using an Agilent 8800 tandem ICP-mass spectrometer (ICP-MS/MS, Agilent Technologies, Japan). The sample introduction system consisted of a 400 μL min−1 MicroMist nebulizer, mounted onto a Peltier-cooled (2 °C) Scott-type spray chamber. The ICP-MS/MS instrument was operated in single-quadrupole (SQ) mode, whereby Q1 was used as an ion guide, and the collision/reaction cell (CRC) in “no gas” or “vented” mode. Quantification was accomplished via external calibration (calibration curve) with Tl as an internal standard to correct for potential matrix effects, instrument drift, and/or plasma instability.
Data acquisition parameters | |
---|---|
Integration time (s) | 4.194 |
Blocks | 5 |
Cycles/block | 10 |
Total cycles | 50 |
Faraday cup configuration | |||||||
---|---|---|---|---|---|---|---|
L3 (198Hg) | L2 (199Hg) | L1 (200Hg) | C (201Hg) | H1 (202Hg) | H2 (203Tl) | H3 (205Tl) | |
(i) | 1011 Ω | 1011 Ω | 1011 Ω | 1011 Ω | 1011 Ω | 1011 Ω | 1011 Ω |
(ii) | 1013 Ω | 1011 Ω | 1011 Ω | 1013 Ω | 1011 Ω | 1011 Ω | 1011 Ω |
(iii) | 1013 Ω | 1013 Ω | 1011 Ω | 1013 Ω | 1013 Ω | 1011 Ω | 1011 Ω |
MDF was reported in delta notation (δXXXHg‰) relative to the NIST SRM 3133 standard4 using eqn (1):
![]() | (1) |
MIF was reported in capital delta notation (ΔXXXHg‰) and was calculated from the corresponding measured δXXXHg values using eqn (2)–(4):4
Δ201Hg = δ201Hg − (δ202Hg × 0.752) | (2) |
Δ200Hg = δ200Hg − (δ202Hg × 0.502) | (3) |
Δ199Hg = δ199Hg − (δ202Hg × 0.252) | (4) |
Fig. 2 illustrates the average δXXXHg (‰) values (n = 5) for the IH standard obtained with both cone configurations at concentrations of 5.0, 2.5, 1.0 and 0.5 μg L−1 Hg. The values presented in Fig. 2 were acquired under wet plasma conditions using 1011 Ω amplifiers and are provided as an example. Similar observations were made with the different settings evaluated, as observed in Tables S1–S4† for δ202Hg, δ201Hg, δ200Hg and δ199Hg, respectively. As can be observed in Fig. 2, the average values obtained using both configurations did not show significant differences at the different concentration levels evaluated. However, for the δ201Hg value, better agreement with the IH standard reference value was achieved with the Jet cones configuration for two measurement sessions. In general, using Jet cones resulted in a better precision. Measurement results after blank subtraction can be found in Table S5.† Broadly speaking, there is minimal difference between values with and without blank subtraction, which indicates that the effect of the blank can be considered negligible within the experimental precision.
![]() | ||
Fig. 2 Average δXXXHg (‰) values for the IH standard (n = 5) obtained at different concentration levels with standard and jet cones under wet plasma conditions and using 1011 Ω amplifiers. Red dashed lines represent the average ± 2SD for previously obtained IH values.6 Error bars correspond to 2SD and are obtained from n = 5 CVG-MC-ICP-MS measurements. |
Table 2 shows the internal precision, expressed as average relative standard deviation (RSD, %) and SD for the Hg isotope ratios obtained from six measurements of a 1.0 μg L−1 NIST SRM 3133 solution under various configurations. As can be seen, better internal pression was achieved for all Hg isotope ratios when measurements were conducted with the Jet cones (RSD% values ranging from 0.0048 to 0.0055% in comparison to the RSD% values obtained with the standard cones ranging from 0.0074 to 0.082%). The average RSD% values for the isotope ratios obtained from measurements at decreasing concentrations can be found in Table S6.† As the concentration level decreases, the average RSD% values increase.
Cones | Plasma | Amplifiers | 199Hg/198Hg | 200Hg/198Hg | 201Hg/198Hg | 202Hg/198Hg | 205Tl/203Tl | |
---|---|---|---|---|---|---|---|---|
Standard | Wet | (i) | Average (%) | 0.0077 | 0.0080 | 0.0082 | 0.0074 | 0.0019 |
SD | 0.0002 | 0.0005 | 0.0004 | 0.0005 | 0.0009 | |||
Standard | Wet | (ii) | Average (%) | 0.0108 | 0.0100 | 0.0052 | 0.0094 | 0.0016 |
SD | 0.0020 | 0.0012 | 0.0006 | 0.0016 | 0.0001 | |||
Jet | Wet | (i) | Average (%) | 0.0055 | 0.0050 | 0.0055 | 0.0048 | 0.0011 |
SD | 0.0004 | 0.0003 | 0.0006 | 0.0004 | 0.0001 | |||
Jet | Wet | (ii) | Average (%) | 0.0075 | 0.0068 | 0.0041 | 0.0071 | 0.0014 |
SD | 0.0006 | 0.0007 | 0.0004 | 0.0007 | 0.0001 | |||
Jet | Dry | (i) | Average (%) | 0.0043 | 0.0042 | 0.0046 | 0.0041 | |
SD | 0.0002 | 0.0005 | 0.0003 | 0.0005 | ||||
Jet | Dry | (ii) | Average (%) | 0.0076 | 0.0076 | 0.0034 | 0.0078 | |
SD | 0.0017 | 0.0016 | 0.0002 | 0.0016 | ||||
Jet | Wet | (iii) | Average (%) | 0.0040 | 0.0115 | 0.0044 | 0.0044 | 0.0030 |
SD | 0.0005 | 0.0015 | 0.0010 | 0.0003 | 0.0002 | |||
Jet | Dry | (iii) | Average (%) | 0.0035 | 0.0132 | 0.0042 | 0.0045 | |
SD | 0.0003 | 0.0037 | 0.0008 | 0.0007 |
The average δXXXHg (‰) values (n = 5) for the IH standard obtained under both dry and wet plasma conditions at concentration levels ranging from 0.25 to 5.0 μg L−1 are shown in Fig. 3 and in Tables S1–S4.†Fig. 3 displays the values acquired with 1011 Ω amplifiers and Jet cones under wet and dry plasma conditions. As expected, the 2SD values increase for both wet and dry plasma as the Hg concentration decreases. For most δXXXHg (‰) values and concentration levels, a better agreement with the reference value is established in dry plasma conditions. According to the results shown in Table 2, the lowest RSD values were obtained in dry plasma conditions using the Jet cones configuration and 1011 Ω amplifiers. Nonetheless, no significant differences in internal precision were found between the different configurations at a concentration of 1 μg L−1 Hg, even though signal intensities were higher in dry plasma conditions (Table 2). Table S6† also illustrates that the RSD values for the isotope ratio data increase as the concentration level decreases.
![]() | ||
Fig. 3 Average δXXXHg (‰) values for the IH solution (n = 5) obtained at different concentration levels under wet and dry plasma conditions using 1011 Ω amplifiers and jet cones. Red dashed lines represent the average ± 2SD of the previously obtained IH values.6 Error bars correspond to 2SD and are obtained from n = 5 CVG-MC-ICP-MS measurements. |
Fig. 4 provides a comparison of the average δXXXHg (‰) values (n = 5) for the IH standard obtained with the three different amplifier setups under dry plasma conditions using the Jet cones configuration. It was observed that the 1013 Ω amplifiers were saturated at concentration levels exceeding 1.0 μg L−1, and thus, the values obtained at concentration levels of 2.5 and 5.0 μg L−1 Hg are shown for the first configuration only. It needs to be noted that the average δ200Hg and δ201Hg values for the IH standard obtained with the second and the third amplifier setup were otherwise collected under exactly the same instrumental conditions, and thus, the differences need to be attributed to variability between different measurement sessions. Notably, the accuracy and precision improved when more 1013 Ω amplifiers were used, as seen in the δ199Hg and δ202Hg values. The average δXXXHg (‰) values for the IH solution obtained under different instrument configurations are summarized in Tables S1–S4† for δ199Hg, δ200Hg, δ201Hg and δ202Hg, respectively. With the third amplifier configuration (using four 1013 Ω amplifiers) under dry plasma conditions and with Jet cones, the average isotope ratio results for the IH standard solution obtained at a concentration level of 0.5 μg L−1 are similar to the values from another study obtained at a concentration of 10 μg L−1.6 This represents a significant improvement, since it allows for the determination of the Hg isotopic composition of samples containing 20-fold lower Hg concentrations with the same level of precision.
![]() | ||
Fig. 4 Average δXXXHg (‰) values for the IH solution (n = 5) obtained at different concentration levels under dry plasma conditions, with Jet cones and using the three different amplifier configurations (i), (ii), and (iii) described in Table 1. Red dashed lines represent the average ± 2SD of the previously obtained IH values.6 Error bars correspond to 2SD and are obtained from n = 5 CVG-MC-ICPMS measurements. |
The average RSD (%) of the isotope ratio measurements increased significantly when two different amplifiers were combined for the collection of the 198Hg and XXXHg isotopes, respectively, compared to the average RSD (%) values when only one type of amplifier was used (Table 2). Please note that, in amplifier configuration (ii), 199Hg/198Hg, 200Hg/198Hg and 202Hg/198Hg isotope ratios are calculated on the basis of data obtained combining two amplifier types, while in amplifier configuration (iii), only for the 200Hg/198Hg isotope ratio data obtained using two amplifier types were used.
The effect of tau correction was evaluated to address differences in the detector response time when a combination of both amplifier types was used. However, no significant differences were observed, nor for the δXXXHg (‰) values neither for the RSD (%) of the Hg isotope ratios, as shown in Table S7.†
Method | δ 199Hg | δ 200Hg | δ 201Hg | δ 202Hg | 199Hg/198Hg RSD (%) | 200Hg/198Hg RSD (%) | 201Hg/198Hg RSD (%) | 202Hg/198Hg RSD (%) | |
---|---|---|---|---|---|---|---|---|---|
IH (literature values)6 | Average | −0.14 | −0.30 | −0.45 | −0.59 | — | — | — | — |
SD | 0.05 | 0.06 | 0.06 | 0.08 | — | — | — | — | |
Method 1 | Average | −0.13 | −0.34 | −0.44 | −0.56 | 0.0028 | 0.0126 | 0.0039 | 0.0037 |
SD | 0.03 | 0.05 | 0.04 | 0.03 | 0.0005 | 0.0012 | 0.0005 | 0.0003 | |
Method 2 | Average | −0.10 | −0.20 | −0.36 | −0.51 | 0.0029 | 0.0202 | 0.0047 | 0.0046 |
SD | 0.05 | 0.07 | 0.05 | 0.06 | 0.0001 | 0.0015 | 0.0005 | 0.0004 |
![]() | ||
Fig. 5 Average δXXXHg (‰) values for the IH standard (n = 5) obtained at different extents of Hg concentration mismatch between the IH standard solution and the NIST SRM 3133 standard solution at a concentration level of 0.25 μg L−1 Hg. The red dashed lines represent the average ± 2SD of the previously obtained IH values.6 Error bars correspond to 2SD and are obtained from n = 5 CVG-MC-ICPMS measurements. Blue circles represent 100% matching, orange circles represent 95% matching, yellow circles represent 106% matching, and grey circles 80% matching. |
The accuracy and precision of the δXXXHg (‰) values (n = 20) for the IH standard at concentration levels ranging from 1.0 to 0.1 μg L−1 were evaluated. Table 4 shows that the highest 2SD values were obtained for δ200Hg, especially at lower concentration levels (i.e., 0.25 and 0.10 μg L−1). This can be attributed to the monitoring of the 200Hg ion beam using a Faraday cup connected to a 1011 Ω amplifier. Self-evidently, the 2SD values increase as the Hg concentration levels decrease. However, at the 0.25 μg L−1 concentration level, the δXXXHg values remain similar to those obtained at 0.50 and 1.0 μg L−1 Hg, as well as those reported in an earlier study6 carried out at 10 μg L−1 Hg concentration level and ranging from 0.13 to 0.27‰. At the 0.10 μg L−1 concentration level, the 2SD values are below 0.50, which is consistent with results obtained from relatively short transient signals during species-specific Hg isotopic analysis via gas chromatography (GC) MC-ICP-MS analysis.24 Considering the SD, the average δXXXHg results were found to be in good agreement with the reference values6 at Hg concentration levels as low as 0.1 μg L−1.
δ 199Hg ± 2SD | δ 200Hg ± 2SD | δ 201Hg ± 2SD | δ 202Hg ± 2SD | |
---|---|---|---|---|
IH (literature values)6 | −0.14 ± 0.09 | −0.30 ± 0.12 | −0.45 ± 0.12 | −0.59 ± 0.15 |
1 μg L−1 | −0.12 ± 0.10 | −0.25 ± 0.10 | −0.41 ± 0.10 | −0.54 ± 0.09 |
0.5 μg L−1 | −0.10 ± 0.12 | −0.22 ± 0.16 | −0.37 ± 0.13 | −0.47 ± 0.14 |
0.25 μg L−1 | −0.11 ± 0.14 | −0.20 ± 0.27 | −0.39 ± 0.13 | −0.50 ± 0.18 |
0.1 μg L−1 | −0.09 ± 0.31 | −0.17 ± 0.49 | −0.34 ± 0.31 | −0.54 ± 0.34 |
0.25 μg L−1 (n = 51) | −0.13 ± 0.15 | −0.25 ± 0.21 | −0.40 ± 0.20 | −0.54 ± 0.20 |
Additionally, a long-term stability study (7 months) was carried out using the IH mercury standard measured at 0.25 μg L−1 (n = 51). The values found are also included in Table 4 and agree with the literature values.6
Additionally, three surface water samples diluted to achieve solutions with a Hg concentration of 0.25 μg L−1 were analyzed as a proof-of-concept application. The resulting δXXXHg (‰) and ΔXXXHg (‰) values obtained for each sample are presented in Table 6. In general, higher 2SD values were obtained for δ200Hg, consistent with the trend observed throughout this study. Despite the geographical proximity of the sources, significant differences were found; especially the δ-values for the first water source (A1) were significantly lower than those for the other two samples. The δ202Hg (‰) and Δ199Hg (‰) values obtained for the three water sources are in agreement with values for freshwaters containing Hg of mixed sources as reported in a review by Blum, Sherman and Johnson.3 Based on the delta values of the three water sources, we could hypothesize that the Hg contamination in water samples from A2 (La Piquera) and A3 (El Muselín) comes from the same source, while in the case of the water samples from A1 (La Parra), the source would be different. Accurate knowledge of the source of Hg polluting the water sources could solve the problem by making them safe for human consumption again.
Water source | δ 199Hg | δ 200Hg | δ 201Hg | δ 202Hg | Δ 199Hg | Δ 201Hg |
---|---|---|---|---|---|---|
A 1.1. | −0.53 | −0.81 | −1.30 | −1.43 | −0.16 | −0.22 |
A 1.2. | −0.52 | −0.74 | −1.24 | −1.42 | −0.16 | −0.18 |
A 1.3. | −0.46 | −0.63 | −1.35 | −1.39 | −0.11 | −0.30 |
Average | −0.50 | −0.72 | −1.30 | −1.42 | −0.14 | −0.23 |
2SD | 0.07 | 0.18 | 0.11 | 0.04 | 0.06 | 0.13 |
A 2.1. | −0.20 | −0.16 | −0.42 | −0.51 | −0.07 | −0.04 |
A 2.2. | −0.20 | −0.13 | −0.47 | −0.53 | −0.06 | −0.07 |
A 2.3. | −0.23 | −0.25 | −0.49 | −0.58 | −0.08 | −0.05 |
Average | −0.21 | −0.18 | −0.46 | −0.54 | −0.07 | −0.05 |
2SD | 0.03 | 0.12 | 0.07 | 0.07 | 0.02 | 0.04 |
A 3.1. | −0.33 | −0.01 | −0.52 | −0.66 | −0.16 | −0.03 |
A 3.2. | −0.33 | −0.16 | −0.44 | −0.60 | −0.18 | 0.01 |
A 3.3. | −0.32 | −0.23 | −0.68 | −0.66 | −0.15 | −0.19 |
Average | −0.33 | −0.14 | −0.55 | −0.64 | −0.16 | −0.07 |
2SD | 0.02 | 0.22 | 0.24 | 0.06 | 0.03 | 0.21 |
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ja00414g |
This journal is © The Royal Society of Chemistry 2024 |