Open Access Article
Sijing Yi
*,
Huanhuan Li and
Xiaoxia Liu
Department of Basic Science, Shanxi Agricultural University, Taigu, 030801, China. E-mail: yisijing@139.com; Tel: +86 354 6288341
First published on 7th April 2022
p-Nitrophenol (PNP) is a toxic contaminant in water, the detection of which has attracted considerable attention. Since ionic liquids (ILs) have been widely used as popular solvents in both extraction and catalysts for PNP, the remediation of PNP is not limited to water and traditional organic solvents. Thus, it is significant to develop approaches for the detection of PNP in ILs. Accordingly, the present work is focused on the detection of PNP in a series of imidazolium-based ILs, 1-hexyl-3-methyl-imidazolium bromide ([Hmim]Br), 1-butyl-3-methyl-imidazolium tetrafluoroborate ([Bmim]BF4), 1-butyl-3-methyl-imidazolium trifluoromesulfonate ([Bmim]TfO), 1-butyl-3-methyl-imidazolium trifluoroacetate ([Bmim]TA), and 1-butyl-3-methyl-imidazolium nitrate ([Bmim]NO3), using a europium-based fluorescent probe, Na3[Eu(DPA)3] (DPA = 2, 6-pyridinedicarboxylic acid). This fluorescent probe showed excellent selectivity and sensitivity toward [Bmim]NO3 in aqueous solution. Further studies showed that not only the fluorescence performance of the europium complex was enhanced in the other four ILs compared with that in water, but also the detecting capability for PNP was improved. The order of the quenching efficiency in different solvents was: [Bmim]BF4 > [Bmim]TfO > [Hmim]Br > [Bmim]TA > water. The higher sensitivity for PNP in ILs was proven to be related to the efficient energy transfer of the europium complex and lower solvent polarity of the ILs. The quenching mechanism for the detection of PNP was established as being due to the ground state electrostatic interactions between the fluorescent probe and analyte, photoinduced electron transfer (PET) and inner filter effect (IFE).
Recently, ionic liquids (ILs) as innovative fluids have been widely acclaimed due to their excellent physicochemical properties, including nonflammability, excellent thermal stability, low combustibility, and high ionic conductivity.18,19 They have shown great promise to be an alternative to conventional solvents in the fields of catalysis,20 electrochemical analysis,21 and wastewater treatment,22–24 especially in the extraction of organic compounds. The progress in IL science inevitably means that the exposure of PNP and their derivatives is not limited to water and traditional organic solvents. For instance, the removal efficiency of PNP in wastewater has been shown to be improved significantly using ILs,25–29 and metal nanoparticles used for the catalytic reduction of nitrophenol have been shown to be stabilized and size-controlled by ILs.30–35 Consequently, the detection of nitro compounds in ILs has become a reality, and it is thus significant to investigate the potential influence of ILs on the fluorescence and response performance of fluorescent probes. At the same time, ILs have been recognized as an ideal media for the study of lanthanide luminescence properties owing to their favorable spectral properties, such as no interference in visible spectral regions and the obviously enhanced fluorescence properties of lanthanide compounds in ILs.36–39 Benefiting from the potential hydrogen bond formation between ILs and ligands of lanthanide complexes, improvements in both luminescence efficiency and stability have been achieved.40,41 It is expected that the use of ILs as solvents might bring about improved fluorescence and detection ability of lanthanide-based fluorescent probes. However, to the best of our knowledge, related studies have been scarcely reported.
Inspired by this, here we selected a series of imidazolium ILs (Fig. 1a): 1-butyl-3-methyl-imidazolium tetrafluoroborate ([Bmim]BF4), 1-hexyl-3-methyl-imidazolium bromide ([Hmim]Br), 1-butyl-3-methyl-imidazolium trifluoromethansulfonate ([Bmim]TfO), 1-butyl-3-methyl-imidazolium trifluoroacetate ([Bmim]TA), and 1-butyl-3-methyl-imidazolium nitrate ([Bmim]NO3) to clarify the specific solvent effect on the detection performance of the europium-based fluorescent probe. ILs with imidazolium as cations were chosen here not only because they have been extensively applied in the extraction, reduction and analysis of PNP,23,25–35 but also because they provide an ideal solvent environment to improve the fluorescence properties of europium complexes.42–45 In the course of better detection performance, we synthesized the lanthanide complex Na3[Eu(DPA)3] (abbreviated as Eu[III], DPA = 2,6-pyridinedicarboxylic acid; Fig. 1b) as a fluorescent probe. A desirable fluorescence intensity was obtained as a result of saturated coordination of DPA ligands to the Eu3+ ion.46–50 Enhanced detection capability of this fluorescent probe was achieved using ILs as a solvent, and the underlying mechanism of sensing behavior towards PNP was also explored. It is hoped that our research results will provide guidelines for scientific research on the detection of nitro compounds in ILs and promote the further application of ILs.
![]() | ||
| Fig. 1 (a) Chemical structure of cations and anions of the selected ionic liquids and (b) the europium complex Na3[Eu(DPA)3]. | ||
| Q = (I0 – I)/I0 × 100% | (1) |
![]() | ||
| Fig. 2 (a) Excitation and (b) emission spectra of Eu[III] in water and different ILs. The spectra were normalized to the most intense peak and 5D0 → 7F2 transition, respectively. | ||
As shown in the emission spectra of Eu[III] in different solvents upon excitation at their corresponding excitation wavelengths (Fig. 2b), characteristic emission bands associated with the 5D0 → 7FJ (J = 0–4) transition of the central Eu3+ ion could be observed in the range of 570–720 nm.53 It should be noted that not only the relative intensity of the different transitions changed, but also the peak shapes of the bands were changed in different solvents, especially the emission bands corresponding to the 5D0 → 7F1 transition. Such variations indicated that the coordination sphere of the Eu3+ ion was affected by the ILs and different symmetry classes occupied by Eu3+ in the europium complexes.54 As is well known, the intensity ratio of 5D0 → 7F2 to 5D0 → 7F1 transitions is closely related to the chemical environment around the Eu3+ ion.55 The values of the intensity ratio, η, can thus be used to measure the degree of asymmetry in the Eu3+ coordination sphere.56 A higher intensity ratio usually implies a lower symmetry of the coordination sphere. The η values of Eu[III] in different solvents were calculated and are listed in Table 1. As shown in Table 1, all the η values of Eu[III] dispersed in ILs decreased compared to that in water, indicating the higher coordination symmetry of the Eu3+ complex. In addition, the luminescence decay curves recorded for the 5D0 → 7F2 emission of Eu[III] dispersed in different solvents could be well fitted to a single exponential function (Fig. S1†). This indicated no dissociation of the europium complexes occurred and that only one europium site was present in water or the ILs.57 Furthermore, the lifetime values (τ) of Eu[III] obtained in ILs were all relatively longer than that in water, reflecting a more efficient energy transfer between the DPA ligands and Eu3+ ions. This could also be proved from the decreased nonradiative rate constant (knr) value of Eu[III] in the ILs (Table 1), which was calculated using eqn (S1) and (S2) listed in the ESI.† The lowest knr value was obtained in the IL [Bmim]BF4, indicating that a comparatively efficient radiative relaxation process occurred in this IL compared to in the other solvents.
| Types of solvent | Water | [Bmim]BF4 | [Hmim]Br | [Bmim]TfO | [Bmim]TA |
|---|---|---|---|---|---|
| η | 4.23 | 3.19 | 3.84 | 3.86 | 3.98 |
| τ/ms | 1.50 | 2.38 | 1.98 | 1.87 | 1.86 |
| knr/ms−1 | 0.299 | 0.100 | 0.159 | 0.183 | 0.186 |
The influence of ILs on the coordination sphere was mainly attributed to the hydrogen bond interactions of the imidazolium ions in the ILs with Eu[III].40,58 As ideal H-bond donors, the hydrogen atoms of the C–H bonds on the aromatic rings of the imidazolium cations form hydrogen bonds with the carboxyl groups of the DPA ligands. This is beneficial to decreasing the nonradiative transition deactivation of the europium complexes and results in a more effective energy transfer efficiency.
Additionally, the fluorescence intensities of Eu[III] were found to also greatly depend on the types of ILs. As shown in Fig. 3a, the emission intensity recorded at 616 nm in different solvents at the same Eu[III] concentration varied as follows: [Bmim]TA > water > [Bmim]BF4 > Bmim]TfO > [Hmim]Br > [Bmim]NO3. It was noteworthy that the fluorescence intensity was found to be almost completely quenched when the Eu[III] complex was dispersed in [Bmim]NO3. Such solvent-dependent luminescent properties can be applied to the sensing of ILs containing nitro groups. Therefore, the emissive response was first investigated by gradually titrating an aqueous solution of Eu[III] with [Bmim]NO3. As shown in Fig. 3b, the fluorescence intensity of Eu[III] decreased significantly with the addition of [Bmim]NO3, and was almost undetectable when the [Bmim]NO3 content reached 15 vol%. The decay dynamics of the emission band at 616 nm could be well fitted using an exponential equation (y = 548.34
e−49.32x, R2 = 0.999, see the inset in Fig. 3). Moreover, the luminescence lifetimes of Eu[III] remained almost unchanged in the absence and presence of [Bmim]NO3 (Fig. S2†), suggesting that the quenching of the emission intensity can be attributed to the non-emission ground state formed between the nitrate ions of [Bmim]NO3 and the DPA ligands of Eu[III] rather than the central Eu3+ ions.59,60 Such luminescent quenching properties imply that the europium complexes can be applied sense molecules that contain nitro groups.
![]() | ||
| Fig. 4 Quenching efficiency of the fluorescence of Eu[III] dispersed in water and different types of ILs after the addition of PNP (0.18 mM). | ||
To examine the PNP-detecting sensitivity of Eu[III] in different solvents in detail, fluorescence quenching titrations of PNP were carried out to monitor the emissive response (Fig. 5a and S3†). By gradually increasing the concentration of the analyte, the detected fluorescence intensity decreased continuously. The quenching efficiency was quantitatively analyzed using a Stern–Volmer equation (eqn (2)):
| (I0/I) = Ksv[A] + 1 | (2) |
| Types of solvent | Fitting equation | R2 | Ksv (M−1) | LOD (M) |
|---|---|---|---|---|
| Water | I0/I = 10 795[A] + 1 |
0.992 | 1.08 × 104 | 8.43 × 10−6 |
| [Bmim]BF4 | I0/I = 19 726[A] + 1 |
0.998 | 1.97 × 104 | 4.48 × 10−6 |
| [Hmim]Br | I0/I = 17 431[A] +1 |
0.999 | 1.78 × 104 | 3.29 × 10−6 |
| [Bmim]TfO | I0/I = 18 537[A] +1 |
0.995 | 1.84 × 104 | 6.62 × 10−6 |
| [Bmim]TA | I0/I = 12 179[A] +1 |
0.987 | 1.22 × 104 | 9.90 × 10−6 |
Furthermore, the LOD was also calculated using eqn (3), which can be used to ascertain the lowest concentration of the analyte that can be reliably measured:
| LOD = 3 × σ/Ksv | (3) |
Based on the above experimental results, the detection performance of the fluorescent probe was improved when Eu[III] was dispersed in the ILs, and the quenching efficiencies were enhanced in the same order as that of the energy transfer efficiency of Eu[III]. The highest quenching percentage and Ksv value obtained in [Bmim]BF4 corresponded to the longest lifetime and lowest knr value of Eu[III]. This indicated that the energy transfer effect of the europium complexes played an important role in the fluorescence detection efficiency. To confirm our speculation, the effect of the appropriate addition of ILs on the detection performance was further investigated. Since the peak shape and position were found to broaden and right shift in the fluorescence spectra of Eu[III] when the IL content was higher than 10%, aqueous solutions with 10% IL content were selected to study the influence of the IL type on the fluorescence detection performance.
Fig. S4† shows the fluorescence excitation and emission spectra of Eu[III] dispersed in water and aqueous solutions containing 10% ILs, wherein no obvious changes could be observed in the shape and position of the excitation and emission bands. This indicated that the coordination environment of Eu3+ remained unchanged when the europium complexes were dispersed in aqueous solutions with an ionic liquid content of 10%. All the η values in 10% ILs obtained from 5D0 → 7F2 to 5D0 → 7F1 transitions were around 4.2, indicating the similar asymmetry degree of the Eu3+ coordination sphere. In addition, the τ values of Eu[III] dispersed in 10% ILs were very close to that in water, which were 1.50, 1.46, 1.44 and 1.59 ms for 10%[Bmim]TA, 10%[Bmim]TfO, 10%[Bmim]BF4, and 10%[Hmim]Br, respectively (Fig. S5†). These results also reflected the similar energy transfer process of Eu[III].
With the addition of 0.18 mM of PNP, all the fluorescence intensity values were significantly decreased, and the quenching degree differed according to the types of the ILs. As shown in Fig. 6, the quenching percentages were all unexpectedly decreased when Eu[III] was dispersed in aqueous solutions containing 10% ILs. The fluorescence quenching percentage of Eu[III] in the different solvents was in a descending order as follows: water (72.8%), 10%[Bmim]BF4 (72.0%), 10%[Hmim]Br (71.2%), 10%[Bmim]TfO (64.9%), 10%[Bmim]TA (47.3%). Furthermore, the europium complex dispersed in an aqueous solution containing 10%[Bmim]TA showed the lowest quenching percentage.
![]() | ||
| Fig. 6 Quenching efficiency of the fluorescence intensity of Eu[III] dispersed in water and aqueous solutions containing different types of ILs after the addition of PNP (0.18 mM). | ||
The sensing sensitivities towards PNP of Eu[III] in different solvents was also investigated by conducting in-depth fluorescence quenching titrations (Fig. 7a). Upon gradually increasing the PNP content, the detected fluorescence intensity decreased correspondingly. The Stern–Volmer curves given by the plots of I0/I versus PNP concentration are shown in Fig. 7b. Except for a global linear fitting of [Bmim]TA, all the other Stern–Volmer curves of aqueous solutions containing ILs exhibit a good linear relationship in the range of 0–0.10 mM PNP concentrations, and exhibit nonlinear curves when the PNP concentration is higher than 0.10 mM (Fig. S6†). This also reflected that the fluorescence quenching mechanism could be attributed to the presence of both static and dynamic quenching at high PNP concentrations ([A] > 0.10 mM). In addition, it can also be speculated that static quenching was dominant in the [Bmim]TA solvent system because linear fitting was achieved across the entire PNP concentration range. The calculated Ksv and LOD values for PNP in aqueous solutions containing the different ILs are also listed in Table 3. The Ksv values were all decreased somewhat in aqueous solutions containing ILs in comparison to in water (1.08 × 104 M−1). The highest Ksv value (9.92 × 103 M−1) and the lowest LOD (6.56 × 10−6 M) were obtained using an aqueous solution containing 10%[Bmim]BF4. The decreased detection performance of Eu[III] dispersed in aqueous solutions containing the 10% ILs demonstrated that the changed coordination sphere and enhanced energy transfer efficiency of Eu[III] in the pure ILs played an important role in the quenching process. Furthermore, the lower quenching efficiencies obtained in 10% ILs implied that other factors also potentially affect the quenching process.
| Types of solvent | Fitting equation | R2 | Ksv (M−1) | LOD (M) |
|---|---|---|---|---|
| 10%[Bmim]BF4 | I0/I = 9917.4[A] + 1 | 0.995 | 9.92 × 103 | 6.56 × 10−6 |
| 10%[Hmim]Br | I0/I = 8569.4[A] + 1 | 0.993 | 8.57 × 103 | 7.47 × 10−6 |
| 10%[Bmim]TfO | I0/I = 8091.2[A] + 1 | 0.991 | 8.09 × 103 | 8.86 × 10−6 |
| 10%[Bmim]TA | I0/I = 4710.1[A] + 1 | 0.991 | 4.71 × 103 | 1.08 × 10−5 |
![]() | ||
| Fig. 8 (a) Schematic presentation of the multi quenching mechanism of Eu[III] by PNP. (b) UV-visible spectra of PNP dispersed in water and different ILs and (c) aqueous solutions containing 10% IL. | ||
Moreover, the competitive absorption of the excitation and emission radiation of the DPA ligands by PNP could also lead to fluorescence quenching, which is well-known as an inner filter effect (IFE).12,61–63 As is known, the DPA ligand transfers energy to the central Eu3+ ion, subsequently resulting in the enhanced fluorescence emission of the europium complexes. The emission intensities of Eu[III] decrease due to the light interception by PNP. In the UV-visible spectra, a higher absorbance of PNP at the maximum excitation wavelength of Eu[III] indicates a more intense competition of energy absorption, and thus a stronger quenching effect on the fluorescence intensity of Eu[III] can be expected.17,59 As shown in Fig. 8b, the optical densities of PNP dispersed in ILs recorded at the corresponding excitation wavelength were all higher than that in water (0.640 at 288 nm). Hence, a more efficient quenching efficiency was achieved when Eu[III] was dispersed in ILs. Moreover, the absorption frequencies of n → π* electronic transitions of nitroaromatic compounds are known to increase with the polarity of the solvent.64 Accordingly, the peak position of PNP dispersed in ILs was observed to red-shift as a consequence of the lower solvent polarity of the ILs compared to water.65 Furthermore, this variation also facilitated more of a spectral overlap with the emission spectrum of Eu[III], which led to the re-absorption of the emitted light of Eu[III]. Therefore, a decrease in the emission intensity was observed. As verified from Fig. S7,† the area of spectral overlap for PNP and Eu[III] dispersed in ILs is mostly larger than that in water, thus a more efficient quenching efficiency and a higher sensitivity were achieved during the titration of Eu[III] dispersed in the ILs. In summary, the quenching process is a synergetic result of ground state electrostatic interactions and a favorable PET process, as well as a strong IFE.
Furthermore, the absorption spectra of PNP in aqueous solutions containing 10% ILs were also investigated and are shown in Fig. 8c. The optical densities of PNP recorded at 288 nm were in the same order as the quenching efficiencies obtained from the fluorescence quenching titrations. A relatively lower absorbance was observed for an aqueous solution containing 10%[Bmim]BF4 and 10%[Hmim]Br, followed by 10%[Bmim]TfO, and particularly 10%[Bmim]TA. So, the lower quenching efficiency detected in aqueous solution containing 10% ILs could be explained as being the result of a poor IFE. Interestingly, a new absorption band was found to appear at 402 nm after the addition of 10%[Bmim]TfO and 10%[Bmim]TA, which could be ascribed to the decreased intermolecular H-bond interactions of PNP. As shown in the inset of Fig. 8c, the H-bonds were mainly formed between the nitro groups of PNP and the hydroxyl groups of water in aqueous solution. After the addition of the ILs, H-bonds were also formed between water molecules and the carboxyl and sulfonate groups of TA− and TFO−, thereby resulting in competition with the PNP. As a result of the relatively stronger hydrogen bonding ability of the ILs, the H-bond interaction between PNP and water decreased and almost disappeared when PNP was dispersed in aqueous solution containing ILs. Hence, the absorbance bands of PNP shifted to a higher wavelength. It is worth noting that the reported hydrogen bonding abilities of TA− and TFO− are relatively stronger than those of Br− and BF4−,59 therefore meaning that the hydrogen bond interaction has less influence after the addition of [Bmim]BF4 and [Hmim]Br. Therefore, no new bands were observed in aqueous solutions containing [Bmim]BF4 and [Hmim]Br. Furthermore, the UV-vis absorption spectra and their corresponding emission spectra (Fig. S8†) showed that aqueous solutions containing 10% ILs had comparatively less overlap in their absorption bands with the ligand centered emission band of Eu[III].
Footnote |
| † Electronic supplementary information (ESI) available: Details on the calculation method of the luminescence parameters and luminescence decay curves of Eu[III] in different solvents. Emission spectra of fluorescence titration and UV-vis absorption spectra of PNP in different solvents. See DOI: 10.1039/d2ra00251e |
| This journal is © The Royal Society of Chemistry 2022 |