Jingyun Jianga,
Jinsheng Zhana,
Wanqing Yuec,
Mengsu Yangc,
Changqing Yi*b and
Cheuk-Wing Li*a
aState Key Laboratory of Quality Research in Chinese Medicine, Institute of Chinese Medical Sciences, University of Macau, Macau SAR, China. E-mail: cheukwli@umac.mo; Fax: +853 2884 1358; Tel: +853 88228513
bKey Laboratory of Sensing Technology and Biomedical Instruments, School of Engineering, Sun Yat-Sen University, Guangzhou, Guangdong Province, China
cKey Laboratory of Biochip Technology, Biotech and Health Centre, Shenzhen Research Institutes of City University of Hong Kong, Shenzhen, China
First published on 14th April 2015
Although polydimethylsiloxane (PDMS) is the popular choice for prototyping microdevices, it suffers from non-specific surface adsorption, small molecule partitioning and drying issues. This paper presents a single microfabrication approach to prepare different thermoplastics that improved the issues inherent to PDMS microdevices. This approach involved transferring microfluidic features from a low-cost printed circuit board (PCB) by hot embossing followed by chloroform vapour assisted bonding. The micro features on the PCB master were comprised of copper metal which maintained structural integrity under pressure and high temperature, making this single microfabrication approach compatible with polycarbonate (PC), polymethylmethacrylate (PMMA), polystyrene (PS) and polysulfone (PSU) thermoplastics. We have modified the vaporized solvent bonding method so that lower temperature and pressure were sufficient for device bonding while the risk of microchannel deformation and clogging was minimized. Various fluorescent dyes and nanoparticles were used to demonstrate the better surface adsorption and partition properties of thermoplastics when compared with the PDMS counterpart. In addition, drying issues were markedly improved in the thermoplastic microdevices. This microfabrication method provided a single and low-cost approach to prepare various thermoplastic microdevices and the entire process (from PCB master to molding and bonding) was completed within 2 hours.
Various techniques are available for molding thermoplastic device, includes hot embossing,12 solvent imprinting,13 injection molding14 and laser ablation.15 Among these techniques, hot embossing is widely employed because it is cost effective, easy to operate and suitable for the replication of various microstructures. However, the major challenge of hot embossing is the preparation of a master that resists high temperature and pressure. Masters have been prepared by nickel,8 silicon,16 SU8 (ref. 17) and plastic material.18 Silicon and SU8 masters have limited reusability due to cracking and distortion under high temperature and pressure.16,18,19 Nickel master is very durable but expensive to prepare by micromachining.8 Plastic master often requires the preparation of a primary SU8 master18 thus the fabrication process is indirect.
Effective bonding techniques are of equal importance in the fabrication of thermoplastics microdevices. Adhesive,20–22 thermal,9,23–30 solvent,18 microwave,31 ultrasonic32 and surface modification bonding33 methods have been reported but thermal bonding is the most commonly used bonding method. However, high temperature and pressure involved in thermal bonding may lead to channel deformation.34 Another popular choice for the bonding of thermoplastic device is solvent bonding. While solvent can be introduced on the substrate through sprinkling,35 capillary action36 and solvent bath dipping,18 it is difficult to avoid excessive solvent absorption that leads to collapse of microchannels. To tackle this issue, sacrificial techniques have been introduced by filling up the microchannels with ice34 or wax.37 Nevertheless, these bonding procedures require subsequent removal of residual sacrificial layer, which can be difficult for complex designs and inevitably complicated the fabrication process. Vaporized solvent bonding allows a controllable dose of solvent deposition so as to prevent excessive solvent absorption. A majority of vaporized solvent bonding methods are demonstrated by cyclic olefin copolymers38–43 that can be handled at room temperature.
In this paper, we described a single approach to prepare four different thermoplastic based microdevices: PMMA, PC, PS and PSU were selected due to their differences in glass transition temperatures as well as their suitability for different reported applications: PMMA provides better electrophoretic performance,9 PC is a good choice for PCR analysis due to its high glass transition temperature and low thermal conductivity.9 PS is a standard material for cell culture10 and PSU exhibits outstanding hemocompatibility with blood.11 Printed circuit board (PCB) master can be used directly for hot embossing and it is durable after a hundred times of repeatedly embossment with negligible deformation of microstructures. We have fine-tuned the vaporized solvent bonding technique so that a lower temperature and pressure were applied for device bonding in order to minimize the risk of channel collapse/deformation.
Parameters used for the embossing and bonding different thermoplastics were tabulated in Tables 1 and 2 respectively. In molding processes, embossing temperature was set to be 20 °C above the glass transition temperature of the corresponding thermoplastic material, except for PSU where a 2-step embossing temperature was used to ensure successful molding due to the minor difference between the glass transition and melting temperature of PSU.
Substrate | Tg | Hot embossing | ||
---|---|---|---|---|
Temp | Time | Pressure | ||
PS | 105 °C | 125 °C | 10 min | 92 psi |
PMMA | 120 °C | 140 °C | 5 min | 92 psi |
PC | 150 °C | 170 °C | 15 min | 92 psi |
PSU | 190 °C | Step 1 220 °C | Step 1 8 min | 92 psi |
Step 2 220 → 190 °C | Step 2 5 min |
Substrate | CHCl3 saturation time | Gap distance | CHCl3 vapour time | Evaporation time | Bonding parameters | ||
---|---|---|---|---|---|---|---|
Temp | Time | Pressure | |||||
PS | 4 min | 12.5 mm | 30 s | 2 min | 85 °C | 30 s | 70 psi |
PMMA | 10 min | 7.5 mm | 60 s | 5 min | 100 °C | 45 s | 80 psi |
PC | 10 min | 10.0 mm | 35 s | 3 min | 100 °C | 60 s | 70 psi |
PSU | 15 min | 7.5 mm | 60 s | 2 min | 140 °C | 60 s | 80 psi |
Conventional vaporized solvent bonding methods consider only the solvent vapour time, i.e., the duration that a substrate is treated with the vapour of bonding solvent. Here, we further defined two waiting period to increase bonding reproducibility (Table 2). (i) Before each bonding process, chloroform was sealed in the Petri dish for a certain period of time, i.e. the “CHCl3 saturation time” before incubating the embossed substrate (ii) after the vaporized substrate was gently brought into contact with a blank substrate prior to thermal bonding, another waiting period was introduced, i.e., the “ evaporation time”.
The distance between chloroform liquid surface and embossed substrate was critical to bonding performance and reported as the gap distance in Table 2. The CHCl3 saturation time and gap distance determined the vapour environment inside the Petri dish before substrate exposure. While increasing the saturation time and decreasing the gap distance facilitated bonding, too much vapour being deposited on the substrate would lead to clogged microchannels and increased the overall fabrication time. The CHCl3 vapour time was the duration when substrate was exposed to preconditioned vapour environment. After vaporized substrate was brought into contact with another piece of substrate, the sandwich was transferred onto the stage of hot compressor for thermal bonding. We found that another source of bonding irreproducibility was originated from the duration of this handling time so the “evaporation time” was defined to ensure a consistent period of time was given before embossing. If evaporation time exceeded the designated period, the escape of chloroform vapour from the sandwich would lead to incomplete bonding. During the fine-tuning of parameters in Table 2, when some vaporized conditions leaded to too short a evaporation time, CHCl3 saturation time and gap distanced were adjusted accordingly in order to increase vapour exposure on substrates.
The final bonding step, vaporized solvent bonding, was dependent to temperature, pressure and the duration of bonding. By comparing our modified vaporized solvent bonding process with other thermal bonding processes found in literatures (Table 3), our method required a significantly shorter compression time of about 1 minute and the bonding temperature used were below the glass transition temperatures of corresponding substrates to minimize channel deformation. It is also worth noting that a majority of reports using conventional vaporized solvent bonding have only demonstrated their feasibility by one substrate material (mainly cyclic olefins copolymer, Table 4). Here, we have successfully produced microchips based on PC, PMMA, PS and PSU materials by using relatively lower pressure and shorter bonding time, which was helpful to minimize channel deformation/collapse. The entire fabrication process (from PCB preparation, molding to bonding) could be completed within 2 hours, which is suitable for the purpose of rapid prototyping microdevices. Although vaporized solvent bonding involved only minute amount of CHCl3 should have permeated after the final bonding step with elevated temperature,18 copious rinsing of mold plastics is recommended to remove any residual chloroform which may do harm to cells or interfere with biological experiments.
Author | Substrate | Bonding parameters | ||
---|---|---|---|---|
Temp | Time | Pressure | ||
Chen9 | PC | 150 °C | 20 min | (No pressure) |
Sun28 | PMMA | 165 °C | 30 min | 20 kPa (3 psi) |
Tan29 | PMMA | 160 °C | 30 min | (No pressure) |
Locascio26 | PS | 105 °C | 10 min | (No pressure) |
Shadpour27 | PSU | 185 °C | 30 min | (No pressure) |
PS | 110 °C | 15 min | (No pressure) | |
PC | 150 °C | 20 min | (No pressure) | |
PMMA | 107 °C | 20 min | (No pressure) | |
Yang30 | PC | 133 °C | 2 h | 1400 psi |
Li24 | PC | 138 °C | 10 min | 900 psi |
Buch23 | PC | 140 °C | 10 min | 900 psi |
Liu25 | PMMA | 85 °C | Step 1: 5 min | Step 1: 44 psi |
Step 2: 15 min | Step 2: 500 psi |
Author | Substrate | Solvent | Vapour time | Bonding parameters | ||
---|---|---|---|---|---|---|
Temp | Time | Pressure | ||||
Ogilvie46 | PMMA | Chloroform | 4 min (25 °C) | 65 °C | 20 min | 203 psi (140 N cm−1) |
COC | Cyclohexane | 4 min (25 °C) | 65 °C | 20 min | 203 psi (140 N cm−1) | |
Rahmanian38 | COC | Cyclohexane | 15 min (30 °C) | RT | (Not mentioned) | 500 psi |
Chen39 | COP | Cyclohexane | 7.5 min (30 °C) | RT | 1 min | 2000 psi (13.79 MPa) |
Chen40 | COC | Cyclohexane | 7 min (30 °C) | RT | 10 min | 500 psi (3.45 MPa) |
Liu41 | COC | Methylcyclohexane | 1 min, RT | 40 °C | 10 min | 200 psi |
Ro42 | COC | Methylcyclohexane | 3 min, RT | 70 °C | 10 min | 100 psi |
Sauer43 | COC | Toluene | 1.5 h, RT | 133 °C | 2 min | 1125 psi |
Ogonczyk48 | PC | Dichloromethane | 30 min (25 °C) | 125–135 °C | 30 min | 58 psi (0.4 MPa) |
The top view and cross-section of the bonded chips were illustrated in Fig. 3. The cross sections of the microchannels were obtained by mechanical cutting. Fig. 3a and c compared the top-view intersecting channel of different thermoplastics with PDMS and all these moulds were prepared from an identical PCB master. Among all thermoplastics, PMMA demonstrated the best fidelity in reproducing the feature from its master, as shown by the more well-defined corners of the intersecting channel. Details about the widths and depths of microchannels after embossing and bonding processes were plotted in Fig. 4. Fig. 3d showed the cross section of the thermoplastics and channel shapes were resemblance to the PDMS counterpart (Fig. 3b). Moreover, in each of these cross sections image, the two layers of substrates were completely merged together, an indicative of good bonding quality.
A quantitative comparison of channel dimension before and after bonding was plotted in Fig. 4. Each column in the plot represented an averaged result of 5 replicas. The small standard deviations in each thermoplastics suggested the high reproducibility in the embossing and bonding processes. PDMS replicas were used as a reference for the actual dimension of the PCB master. Compared with PDMS, the embossed channel depth and width of PMMA and PS were essentially the same while PC and PSU were slightly decreased. In addition, the dimension loss percentage (the channel dimension difference between embossed substrate and bonded chip over the dimension of embossed substrate) were showed in dotted lines. It was clear that PMMA demonstrated the smallest deviation in channel dimensions among all thermoplastics. The average depth of the bonded channels was 24.5 μm with the loss percentage of 4.38%. Similarly, the average width measured at the bottom of inverted trapezoidal shaped cross section (Fig. 3d) was 198 μm and the loss percentage was less than 3%. However, PS and PSU were subjected to a relatively greater loss in dimension after the bonding process. Overall, the bonding quality was satisfactory with dimension loss percentage below 10% among all thermoplastics, which should be acceptable for many microfluidic applications.
In order to evaluate the quality and usability of these thermoplastic chips, a series of adsorption and partition tests were performed. The prerequisite of these tests was leak-free bonding among the thermoplastic devices. Except for PDMS microdevices, epoxy adhesive was used to fix Teflon tubing inserted into the access ports of thermoplastics to ensure no solution leaking occurred at the reservoirs throughout the experiments. All types of thermoplastic chips have passed this leakage test by withstanding a flow speed of 325 μL h−1 for at least half an hour.
High adsorption (on surface) and partition (into bulk) of certain small molecules and proteins are notorious to PDMS based microfluidic devices due to its porous and hydrophobic nature,5,7 which compromised the accuracy of quantitative analysis in certain assays.4 Thus, adsorption and partition of dye and particles were examined in thermoplastic microchips and PDMS microchips for comparison. We have selected carboxylate-modified, fluorescent microspheres and three fluorescent dyes excitable by commonly used filter sets (DAPI, FITC and TRITC) in fluorescence microscopy. Fig. 5 showed a PDMS based microchannel filled with Rhodamine B (RhoB) dye followed by buffer rinsing, and a significant residual fluorescent signal could be detected. The percentage of species adsorption was determined by the normalization of background subtracted residual fluorescence signal by the background subtracted dye filled signal. In addition to surface adsorption, some species tended to enter the PDMS porous matrix. Such species partition was quantified by the evaluation of the partition distance as showed in Fig. 5, i.e., the distance where fluorophores travelled into the bulk of the substrate. Partition rate (μm min−1) of each species was calculated by the partition distance over the duration of dye fill time. The quantitative analysis of species adsorption and partition in varies thermoplastics with PDMS elastomer were shown in Fig. 6. Referring to the left y-axis that represented the percentage adsorption of species on microchannel surface, all 4 thermoplastics and PDMS exhibited considerable surface adsorption (20% or above) to the 30 nm, carboxylate-modified polystyrene (L5155) after microchannels being continuously rinsed with buffer at 325 μL h−1 for 20 min. While Rhodamine B (RhoB) is well-known to be severely adsorbed on PDMS surface,5 this commonly used dye was compatible with other thermoplastics and worked particularly well with PMMA. 7-Amino-4-methyl-coumarin (AMC) and fluorescein (FLUO) were compatible with all thermoplastics and the elastomer while these dyes worked particularly well with PS with negligible adsorption. By referring the right y-axis that represented the rate of species entering the bulk of materials in Fig. 6, PDMS elastomer was clearly inferior to other thermoplastics and severely partitioned by 30 nm nanoparticles, RhoB and AMC. In contrast, all other thermoplastics exhibited negligible partition of the species used in this study. Accordingly, AMC, FLUO and RhoB, which worked in DAPI, FITC, TRITC filter sets respectively, have performed well by having low adsorption and partition in various thermoplastics and they were suitable candidates for on-chip quantitative analysis.
While the porous nature of PDMS attributed to the uptake of small molecules as shown in the partition of dye species, it also facilitated the diffusion of water vapour into the bulk polymer. Our second challenge to these thermoplastic devices was their ability to tackle the notorious drying issues associated with PDMS microdevices.6 We leverage the capability of microfluidic droplet formation to prepare well-defined water droplets followed by monitoring the change in droplet size during the course of on-chip incubation to study the extent of water drying among all substrates.
The microchip design layout was shown in Fig. 7. This widely adopted droplet formation design was comprised of two inlets, one for dispersed phase (water) and the other for continuous phase (oil). The main channel dimensions were 20 μm in depth and 90 μm in width. After uniform droplet formation was achieved, the tubings of oil and water inlets were removed such that individual droplets were retained within the microchannel at even spacing as illustrated in the inset of Fig. 7. Fig. 8 showed the relationship between the normalized droplet size (by normalizing the residual droplet volume at a particular time point over the original droplet size at zero time point) and their incubation time inside the microchannels made of different substrates. Since PS was incompatible with Aquapel coating (severe cracking resulted once the chemical filled the microchannel), the substrate was excluded from this test. For the remaining candidates, images captured at 10 minutes intervals were used for the evaluation of droplet size change as a function of time.
Fig. 7 Design layout of the droplet formation microdevice with inset showing the micrograph of droplets maintained steadily within a PC-based microchannel for incubation. |
The on-chip experiments for different substrates were performed side-by-side so variation in ambient temperature and humidity were negligible. During the entire one-hour observation period, the evenly spaced droplets remained static within the microchannels, however, ∼80% droplet size shrinkage was observed in PDMS based microdevices. PC has demonstrated the minimal size shrinkage by retaining almost 90% of its original droplet size. All the tested thermoplastics have performed significantly better than the PDMS elastomer in terms of maintaining droplet size and the results were consistent with their low species partition. The results also suggested it is straightforward to adopt thermoplastic microdevices to conduct droplet-based chemical reactions so as to minimize the impact of volume lost accompanying with PDMS substrates.
This journal is © The Royal Society of Chemistry 2015 |