Anna K.
Boardman
a,
Sandra
Allison
a,
Andre
Sharon
ab and
Alexis F.
Sauer-Budge
*ac
aCenter for Manufacturing Innovation, Fraunhofer USA, Brookline, MA 02446, USA. E-mail: asauerbudge@fraunhofer.org; Fax: +1 617 353 1896; Tel: +1 617 353 1888
bDepartment of Mechanical Engineering, Boston University, Boston, MA 02215, USA
cDepartment of Biomedical Engineering, Boston University, Boston, MA 02215, USA
First published on 11th October 2012
To accurately diagnose microbial infections in blood, it is essential to recover as many microorganisms from a sample as possible. Unfortunately, recovery of such microorganisms depends significantly on their adhesion to the surfaces of diagnostic devices. Consequently, we sought to minimize the adhesion of methicillin-sensitive Staphylococcus aureus (MSSA) to the surface of polypropylene- and acrylic-based bacteria concentration devices. These devices were treated with 11 different coatings having various charges and hydrophobicities. Some coatings promoted bacterial adhesion under centrifugation, whereas others were more likely to prevent it. Experiments were run using a simple buffer system and lysed blood, both inoculated with MSSA. Under both conditions, Hydromer's 7-TS-13 and Aqua 65JL were most effective in reducing bacterial adhesion.
The prompt diagnosis and treatment of bacteremia is of significant interest to health care professionals. This condition is often the result of a severe infection introduced to the body by an infected catheter or other device. When left undiagnosed, bacteremia can lead to systemic inflammatory response syndrome (SIRS) and ultimately sepsis.6 The identification of SIRS is based upon the recognition of two or more of the following symptoms: fever (or hypothermia), accelerated heart rate or respiratory rate and abnormal white blood cell count.7 Symptom identification can often be slow and inaccurate, and once SIRS is correctly diagnosed, patients are often at high risk for developing sepsis, the 13th leading cause of death in the US.8 Consequently, it is necessary to develop quick and accurate diagnostics for detecting bacteria in blood, urine, and other normally sterile fluids.
When developing diagnostics for detecting bacteria, adhesion to the diagnostic device's surface should be considered.9–11 Bacterial adhesion is a complex process that is affected by many factors.12,13 Both specific and non-specific interactions affect the bacteria's ability to attach to the surface, as well as surface properties (chemical composition, charge, roughness) and the associated flow conditions.14–18
The chemical composition of the surface can influence bacterial adherence to a surface.19–23 Materials with different functional groups change bacterial adhesion depending on material hydrophobicity and charge.24 For example, Chu and Williams studied the effects of physical configurations of suture materials on bacterial adhesion.25 They showed that polydioxanone sutures exhibited slight affinity towards the adherence of Escherichia coli and Staphylococcus aureus; however, Dexon sutures had much higher affinity towards the two bacteria. Additionally, if the surface chemistry of the material is modified or changed, bacterial adhesion can be affected. James and Jayakrishnan proved that surface thiocyanation of PVC decreased adhesion of S. aureus and Staphylococcus epidermidis because of the change in hydrophilicity of the native PVC (as measured by contact angle) from 72° to 50°.26
Surface charge can affect the bacterial adhesion. At a neutral pH, bacteria are commonly negatively charged; consequently, a slight repulsion to negatively charged surfaces is expected. For example, Kiremitci-Gumusderelioglu and Pesmen showed that bacterial adhesion was reduced on negatively charged PMMA/AA, but it increased on positively charged PMMA/DMAEMA.27 Similarly, Terada et al. found that E. coli's adhesion to modified PE sheets, where the functional groups were positively charged, was significantly higher than that of negatively charged functional groups on the modified PE sheets.28,29
The material's surface roughness can also play a large role in bacterial adhesion. Surface irregularities promote bacterial adhesion whereas very smooth surfaces do not.30,31 One reason could be that rough surfaces have a greater surface area and the depressions in the roughened surfaces provide a favorable location for colonization. Taylor et al. produced a range of roughness on PMMA and tested bacterial adhesion. Large augmentations in roughness produced by silicone carbide paper (grades P400 and P120) had no significant effect on adhesion compared to the smooth surface. But, a small increase in surface roughness using silicon carbide paper P1200 resulted in a noteworthy increase in bacterial adhesion.32 In a comparison of polished, unpolished, and abraded stainless steel, it was shown that on the roughest surface (abraded), most of the S. aureus remained on the surface whereas on the smoothest surface (polished), significantly less bacteria were present.33 Due to the multiplicity of factors that can affect the adhesion of bacteria, a survey studying the bacterial adhesion to a variety of surface coatings would be a useful resource to researchers developing bacterial diagnostics and other medical devices.
In this paper, we describe machined polypropylene and acrylic devices for concentrating bacteria and compare how various surface coatings impact the recovery of low numbers of bacteria. We investigated the surface coatings in two model matrices: a simple buffer and lysed whole blood. Ideally, the surface of the device would prevent bacterial adhesion, and yet not be anti-microbial to facilitate a wide-range of downstream bacterial diagnostics (e.g. both genetic-based and culture-based methods). Additionally, the surface of the device would enable the recovery of very low numbers of bacteria in a complex matrix (e.g. blood) so as to be useful for the development of in vitro bacteremia diagnostics. Bacterial adhesion to commercially available and in-house coatings was tested.
![]() | ||
| Fig. 1 Images of the blunt-nosed devices. Left: angled top view. Right: top view. | ||
Polypropylene was the material of most interest due to its common use in laboratory consumables and all coatings were applied to it. However, one of the coating vendors suggested using acrylic to produce a more effective coating. As a result, some acrylic devices were also tested.
| Company | Coatings | Hydrophilic or hydrophobic | Charged | Main structure | General coating process |
|---|---|---|---|---|---|
| a ND, not disclosed. | |||||
| AST | HydroLAST | Slightly hydrophilic | Positive (weak) | ND | Dip coated |
| Negative (weak) | ND | Dip coated | |||
| Neutral | ND | Dip coated | |||
| BioCoat | Hydak B-23KX2/L-578 | Hydrophilic | Negative | Hyaluronic acid | Dip coated |
| Hydromer | 7-TS-13 | Hydrophilic | Neutral | Polyvinylpyrrolidone | Dip coated |
| Aqua65JL | Hydrophilic | Neutral | Polyvinylpyrrolidone | Dip coated | |
| IST | Philix | Hydrophilic | ND | SiO2 | Vacuum vapor delivery |
| Hydrophilic | ND | PEG | Vacuum vapor delivery | ||
| Hydrophilic | ND | Al2O3 | Vacuum vapor delivery | ||
| Repellix | Super-hydrophobic | ND | ND | Vacuum vapor delivery | |
All the devices treated with experimental coatings were compared to identical devices treated with the triblock copolymer Pluronic F127 (Sigma-Aldrich). This material has been used to reduce bacterial adhesion in other studies and could be easily applied in our laboratory.34 The blunt-nosed devices were coated by submerging them in a beaker of Pluronic solution (0.5 g L−1) and placing the beaker in a sonic bath for 10 minutes.
Also, Pluronic F127 was used to coat the inside of 15 mL polypropylene centrifuge tubes (Fisher Scientific). It was determined that almost 100% of bacteria were recovered when the 15 mL conical tubes were coated with Pluronic F127, so these tubes were used as standards.
A solution was made containing 1 part of 0.05% BSA, 2 parts of 1× PBS and 1 part of 103 cfu mL−1 MSSA bacteria diluted with water. The final solution of BSA, PBS, and bacteria contained approximately 100 cfu per 400 μL. This concentration was verified by plating 200 μL of the solution on an LB agar plate, incubating overnight at 37 °C, and counting the number of resulting colonies.
Each of the coated blunt-nosed devices and centrifuge tubes were filled with 400 μL of the above solution. The samples were spun in a swinging bucket centrifuge (Eppendorf 5810 R) at 3200 RCF for 5 minutes to concentrate the bacteria to the bottom of the devices/tubes. Next, 350 μL of the supernatant were drawn off and discarded. (In initial studies, the supernatant was also interrogated by quantitative plating, but few to no bacteria were ever found in the supernatant. To save on materials, the supernatant was not plated further). The remaining 50 μL (the “pellet”) was aspirated with a pipette and deposited on an LB agar plate. After adding 70 μL of sterile deionized water to each device/tube, the water was pipetted up and down to remove any loosely adhered bacteria (the “wash”) and deposited on a separate LB agar plate.
These plates were incubated at 37 °C overnight and the resulting colonies were counted the next day. Those numbers were compared to the number of colony forming units originally added to the devices/tubes to determine how many bacteria could be removed with just aspiration and how many needed an additional rinse to be removed.
Fig. 2 and 3 show the percentage of bacteria recovered from our coated devices and tubes when tests were run using the simple buffer system (Fig. 2) and lysed whole blood (Fig. 3). In these figures, each bar represents the average recovery observed for a single type of coating. The dark gray portions of the bars correspond to the amount of bacteria recovered from the pellet of each device or tube while the light gray portions of the bars show the additional residual bacteria recovered during the subsequent wash step. Since removal of the pellet involved a single aspiration and was less mechanically rigorous than the wash, higher pellet recovery rates were expected for coatings with greater resistances to bacterial adhesion.
![]() | ||
| Fig. 2 Comparison of coatings in a simple buffer system (PBS–BSA). The bottom (dark shading) bars represent the bacterial recovery from the pellet at the bottom of the disposable. The top (light shading) bars represent the bacterial recovery after 70 μL of deionized water were added to the chamber and pipetted up and down. Each bar represents the mean of at least three replicate experiments with a single standard deviation shown. | ||
![]() | ||
| Fig. 3 Comparison of coatings in pooled whole blood. The bottom (dark shading) bars represent the bacterial recovery from the pellet at the bottom of the disposable. The top (light shading) bars represent the bacterial recovery after 70 μL of deionized water were added to the chamber and pipetted up and down. Each bar represents the mean of at least three replicate experiments with a single standard deviation shown. | ||
When assessing the performance of each coating, two main factors were considered: the overall recovery of bacteria as well as how many bacteria were recovered from the pellet versus from the wash. We rationalized that the bacteria collected in the pellet were less adhered to the surface than those recovered in the subsequent more vigorous wash step. Accordingly, the pellet recovery was an important factor to bear in mind and it was observed that the recovery varied considerably between coatings.
To determine whether the results were statistically significant, Student t-tests were performed for each of the tested coatings against the controls (Pluronic-coated devices and Pluronic-coated commercial tubes). Summaries of the statistical analysis are shown in Tables 2 and 3.
| Pellet recoveries | Total recoveries | |||
|---|---|---|---|---|
| Pluronic coated devices | Pluronic coated tubes | Pluronic coated devices | Pluronic coated tubes | |
| AST – weakly positive | − | − | − | − |
| AST – weakly negative | − | − | 0 | − |
| AST – neutral | − | − | 0 | − |
| BioCoat – PP | − | − | 0 | − |
| BioCoat – acrylic | − | − | − | − |
| Hydromer – 7-TS-13 | + | 0 | + | 0 |
| Hydromer – Aqua 65JL | + | 0 | 0 | − |
| IST – Al2O3 | + | 0 | + | − |
| IST – Al2O3 + SiO2 | 0 | − | 0 | − |
| IST – super-hydrophobic | 0 | − | 0 | − |
| IST – PEG | 0 | − | 0 | − |
| Pellet recoveries | Total recoveries | |||
|---|---|---|---|---|
| Pluronic coated devices | Pluronic coated tubes | Pluronic coated devices | Pluronic coated tubes | |
| AST – weakly positive | − | − | − | − |
| AST – weakly negative | 0 | − | 0 | 0 |
| AST – neutral | 0 | − | − | − |
| BioCoat – PP | 0 | 0 | 0 | 0 |
| BioCoat – acrylic | 0 | 0 | 0 | 0 |
| Hydromer – 7-TS-13 | + | 0 | + | 0 |
| Hydromer – Aqua 65JL | + | 0 | + | 0 |
| IST – Al2O3 | 0 | − | 0 | 0 |
| IST – Al2O3 + SiO2 | 0 | − | 0 | 0 |
| IST – super-hydrophobic | 0 | − | 0 | 0 |
| IST – PEG | 0 | − | 0 | 0 |
The bars to the far left in each graph show the recovery obtained from 15 mL commercial polypropylene centrifuge tubes coated with Pluronic. Previous tests in our laboratory showed that these conditions yielded almost 100% recovery of MSSA, and subsequently we used these values as the standards to which all the coatings would be compared. We found that approximately 100% of the bacteria were recovered from the control tubes when the simple buffer system was used (79% in the pellet and 28% in the wash). When lysed blood was used about 80% of the MSSA was recovered (59% in the pellet and 23% in the wash).
In addition to the Pluronic-coated polypropylene tubes, we also coated polypropylene and acrylic blunt-nosed devices with Pluronic. These devices were machined, not molded like the commercial tubes, and were used as a second control condition to which we compared the coated devices' recoveries. While the commercial tube recovery rates were considered to be the maximum possible under perfect conditions, the recovery rates from the Pluronic-coated blunt-nosed devices were in fact a more realistic control condition to which to compare the coatings because the surfaces were manufactured using the same process. The commercial tubes may have additional surface coatings or treatments that influence bacterial adhesion; our efforts at researching the manufacturing method were unproductive. Therefore, the machined blunt-nosed devices were used as our benchmarks.
When the simple buffer system was used with the polypropylene devices coated with Pluronic, we observed total recovery rates of approximately 80% with nearly equal amounts of bacteria coming from the pellet and wash. These figures remained roughly the same when blood was used, once again with similar recoveries obtained from the pellet and wash.
When acrylic blunt-nosed devices were treated with Pluronic, recovery rates were comparable to those seen in the commercial tubes when both the simple buffer system and lysed blood were used. For the simple buffer system, recoveries were found to be 85% in the pellet and 28% in the wash. For lysed blood, recoveries were found to be 38% in the pellet and 49% in the wash. Overall, the Pluronic-coated acrylic devices performed better than the equivalent polypropylene devices. However, polypropylene is a much more common material in biomedical devices due to its ability to be autoclaved, its strength to withstand high centrifugation, and ease in molding. Subsequently, the polypropylene devices were of greater interest to us than the acrylic devices.
AST's weakly positive coating rendered the lowest recovery rate of all the coatings when lysed blood was used. This coating had only a 17% total recovery, found entirely in the pellet. AST's other coatings also resulted in low overall recoveries when compared to Pluronic-coated devices using lysed blood. On the other hand, the neutral and negative coatings resulted in pellet recoveries that were statistically similar to those seen with the Pluronic-coated devices. More bacteria were collected in the pellet using lysed blood as compared to the simple buffer system presumably because of the additional components present in blood (such as cells, cell fragments, proteins, fat, etc). These components can interact with the bacteria and block the walls of the devices, which can interfere with the electrostatic association. All of these factors could help to concentrate bacteria under centrifugation.
In the blood and lysis buffer test, little difference was observed between the polypropylene and acrylic devices. Both of the two materials had almost 80% recovery between the pellet and the wash. The acrylic device performed slightly better in the pellet recovery than the polypropylene device with 61% and 43%, respectively, confirming predictions of a better coating on an acrylic material.
Overall, regardless of the material or the experimental conditions, devices treated with BioCoat's coatings did not prevent bacterial adhesion better than similar devices treated with Pluronic. Under the best conditions, BioCoat's coatings performed as well as Pluronic-coated controls. When comparing BioCoat’s coatings to the other negatively charged AST coating, BioCoat achieved slightly higher recoveries. However, because Biocoat’s coating performed better on acrylic than the more common polypropylene, it is not as good of a candidate for devices that are required to be made of polypropylene.
Almost 100% total recovery was obtained from Hydromer's two coatings when whole lysed blood was used. The pellet recovery for 7-TS-13 was 61% and Aqua 65JL was 49%. Both the pellet and total recoveries were significantly better than the Pluronic-coated devices and comparable to the commercial tubes treated with Pluronic.
In the lysed blood tests, IST's PEG coating had the best result of the four IST coatings tested, which was more consistent with our hypothesis based on PEG's exceptional resistance to protein adsorption. The PEG coated devices had almost 90% total recovery; however the pellet recovery was quite low at 39%, indicating that most of the bacteria were weakly adhered to the surface and became dislodged during the wash step. The other three coatings (Al2O3, SiO2, super-hydrophobic) each had less than 80% total recovery and none of them had over 50% of the bacteria present in the pellet. The Al2O3 coating was tested because the devices could be coated quickly, even though Al2O3 has been found to attract bacteria.45 The SiO2 coating is essentially a glass coating on the polypropylene blunt-nosed device.46 Untreated glass can adhere biological organisms, suggesting that the bacteria could have adhered to the coating. The low recovery of the super-hydrophobic coating was unexpected since the aqueous solution should have very minimal interactions with the coating wall due to its high hydrophobicity. Super-hydrophobic coatings have contact angles greater than 150° but can have a low or high adhesive force.47 On these devices, the aqueous solutions had a low adhesive force, and adopted a non-wet-contact mode, so it was expected that the bacterial solutions would have negligible contact with the surface. Some evidence exists to support the hypothesis that super-hydrophobic coatings can prevent biofilm formation and bacterial colonization.48,49 However, with 67% total recovery and, of that, only 28% in the pellet, the performance of the super-hydrophobic coating was lackluster. It may be due to the fact that in the conditions of our experiments, the centrifugal forces overcome the repulsive hydrophobic forces resulting in poorer performance than that may be achieved with other experimental parameters (e.g. implantable devices).
Of all the coatings tested, two in particular showed the most promise: Hydromer's 7-TS-13 and Aqua 65JL. The 7-TS-13 coating resulted in total recoveries comparable to those seen with Pluronic-coated commercial tubes or Pluronic-coated devices, when both lysed blood and the simple buffer system were used. This was also true for the Aqua 65JL coating except that its total recoveries were slightly lower than those corresponding to Pluronic-coated commercial tubes when the lysed blood model was used.
When just the recovery rates from the pellets were considered, the Hydromer coatings still outperformed the others tested. When either the simple buffer or lysed blood tests were used, both Hydromer coatings showed pellet recoveries similar to those of the coated commercial tubes and better than those of the Pluronic-coated blunt-nosed devices. Although the 7-TS-13 coating was not statistically better than the 65JL coating, its average recoveries were higher and it was the only coating of all those tested that out-performed the Pluronic-coated devices under the simple buffer and lysed blood conditions for both the total recovery and pellet recovery. These coatings were applied via dip coating, which worked well with the size and shape of our device. Future testing will need to be performed for coating smaller dimensions.
By studying this array of coatings, we were able to gain insight into the best coatings to minimize MSSA adhesion to coated polypropylene and acrylic devices. Based on the obtained data, we were able to determine that Hydromer's 7-TS-13 coating was the most effective in resisting bacterial adhesion, even in devices in which the bacterial interactions with the surface were driven via centrifugation. In the development of in vitro diagnostic devices for the detection of bacteremia, coatings such as Hydromer's 7-TS-13 are critical so the low numbers of bacteria found in solution are not lost to non-specific binding to the surface. With the increasing focus on miniaturization and microfluidic devices that have high surface area to volume ratios, coatings that reduce bacterial adhesion are of even greater importance.
Footnote |
| † Electronic supplementary information (ESI) available: Details on the fabrication of the blunt-nosed devices are specified. See DOI: 10.1039/c2ay25662b |
| This journal is © The Royal Society of Chemistry 2013 |