Hengchong
Shi
a,
Dean
Shi
*b,
Zhanhai
Yao
a,
Shifang
Luan
a,
Jing
Jin
a,
Jie
Zhao
a,
Huawei
Yang
a,
Paola
Stagnaro
c and
Jinghua
Yin
*a
aState Key Laboratory of Polymer Physics and Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun, 130022, P. R. China. E-mail: yinjh@ciac.jl.cn
bMinistry-of-Education Key Laboratory for the Green Preparation and Application of Functional Materials, Faculty of Materials Science and Engineering, Hubei University, Wuhan, 430062, P. R. China. E-mail: deanshi2001@yahoo.com
cIstituto per Io Studio delle Macromolecole, Consiglio Nazionale delle Ricerche, Via de Marini 6, Genova, 16149, Italy
First published on 14th December 2010
Macromonomer cyclooctene-poly(ethylene glycol) (cyclooctene-PEG) was first synthesized before being copolymerized with cyclooctene by ring opening metathesis polymerization (ROMP) to obtain an amphiphilic graft copolymer (poly(cyclooctene)-g-PEG) with polycyclooctene as the hydrophobic trunk chain and PEG as hydrophilic side chains. The structure of poly(cyclooctene)-g-PEG copolymer was characterized by FTIR and 1H-NMR. The surface properties of poly(cyclooctene)-g-PEG film were evaluated through water contact angle and X-ray photoelectron spectroscopy (XPS). Water contact angle decreased from 87.7° to 65.8° along with increasing the content of PEG. Protein adsorption results showed that poly(cyclooctene)-g-PEG copolymers had significant effect on preventing bovine serum albumin (BSA) from absorbing onto the polymer surface.
There are three paths to synthesize amphiphilic graft copolymers. Path I, the “grafting from” approach:14,15 the side chains are grown from the polymeric backbone pending initiating groups by a variety of controlled polymerization methods such as atom transfer radical polymerization (ATRP), nitroxide-mediated polymerization (NMP) or reversible addition fragmentation chain transfer (RAFT). Path II, the “grafting through” approach:16,17 macromonomers are synthesized first and then homopolymerized or copolymerized with other monomers by ATRP, NMP and RAFT. Path III, the “grafting onto” approach:18–20 two kinds of homopolymers are first prepared separately. One has a lot of reactive sites along with the backbone and the other has functional group at the end of molecule chain. Copolymers are formed via coupling reaction between them.
In Path I, the grafting point can be easily controlled and the purification of copolymers from the reaction system is easy. Although the chain length of the side chain is adjustable in Path III, the grafting site and grafting degree cannot be controlled. However, both the chain length and position of the grafted chains can be easily controlled in method Path II. So in this article, Path II was used to prepare the poly(cyclooctene)-g-PEG amphiphilic graft copolymer. Ring opening metathesis polymerization (ROMP) was adopted with a cyclic monomer (cyclooctene) as the initial point. The surface properties were characterized by water contact angle analysis and X-ray photoelectron spectroscopy (XPS). Finally, protein adsorption experiments were performed on the surfaces of the amphiphilic poly(cyclooctene)-g-PEG copolymer and polycyclooctene, respectively. The results showed that poly(cyclooctene)-g-PEG had a significant effect on the reduction of the adsorption of bovine serum albumin (BSA). These amphiphilic polyclooctene-g-PEG copolymers have potential applications in drug delivery, implant, biomedical coating and so on.
Scheme 1 Synthesis route of cyclooctene-PEG macromonomer. |
The NMR chemical shifts of cyclooctene-PEG macromonomer were assigned as follows:
1H-NMR δ (ppm) with CDCl3 as solvent: 7.0 to 7.5 (aromatic-H), 5.5–5.7 (CH2–CH in cyclooctene), 4.6–4.7 (–CH–O in cyclooctene), 3.6 (–O–CH2–CH2– in PEG), 3.38 (CH3–O–PEG).
The Mn and PDI of cyclooctene-PEG macromonomer are 1085 and 1.02, respectively, according to MALDI-TOF-MS result.
Scheme 2 Synthesis route of poly(cyclooctene)-g-PEG. |
The NMR chemical shifts of poly(cyclooctene)-g-PEG were assigned as follows:
1H-NMR δ (ppm) with CDCl3 as solvent: 5.37 (CH2–CH in polymer chain), 4.84 (–CH–O in polymer chain), 4.33 (–COO–CH2–CH2–O), 3.6 (–O–CH2–CH2– in PEG), 3.38 (CH3–O–PEG).
Molecular weights and molecular weight distributions were measured on a Waters-2414 gel permeation chromatography (GPC) instrument. The measurements were carried out at 30 °C using CHCl3 as the eluent with a flow rate of 1.0 mL min−1. The system was calibrated with polystyrene standards.
FTIR spectra of both monomers and polymers were recorded using a Bruker Vertex 70 FTIR spectrometer from 4000 to 400 cm−1 at a resolution of 2 cm−1 for 32 scans.
Surfaces for XPS spectroscopy, contact angle measurement, BSA protein adsorption testing were prepared on glass slide by spin-coating 0.5% (w/w) solutions of poly(cyclooctene)-g-PEG graft copolymers and polycyclooctene (virgin) in toluene at 2000 rpm for 30 s using a KW-4A spin coater. Then the surfaces of the samples prepared for measurement were dried in a vacuum oven at reduced pressure at room temperature for at least 12 h to remove solvent completely.
The determination of contact angle was performed on a Drop Shape Analyzer DSA100 (KRÜSS company). A droplet of water (2 μL) was put on the surface of a film and the contact angle was measured. Ten measurements were carried out for a single sample and the values obtained were averaged.
The X-ray photoelectron spectroscopy (XPS) was measured with VG ESCALAB MK at room temperature by using an Mg-Kα X-ray source (hν = 1253.6 eV) at 14 kV and 20 mA. The sample analysis chamber of the XPS instrument was maintained at a pressure of 1 × 10−7 Pa. The takeoff-angle (the angle between the analyzer and the surface normal) was kept at 30° for all samples analyzed. All the C1s peaks were calibrated to the standard binding energy of 284.6 eV for neutral carbon in order to correct the charging energy shifts. A linear-background method removed the XPS background and the peaks analysis carried out by using curve-fitting software.
Fig. 1 FTIR spectrum of cylooctene–PEG and cyclooctene–TDI. |
Fig. 2 is a 1H-NMR spectrum of cyclooctene–PEG macromonomer with CDCl3 as solvent. The peak area ratio between chemical shifts at 5.5–5.7 and 3.38 were attributed to CH2–CH in cyclooctene and CH3– in PEG group and their peak area ratio was 2:3, which was in accordance with the theoretical value.
Fig. 2 1H-NMR spectrum of cyclooctene–PEG macromonomer. |
Graft copolymers 1#–4# were synthesized by ROMP with the appropriate ratio of cyclooctene and cyclooctene–PEG macromonomer in dichloromethane (Table 1), using Grubbs' generation II catalyst. Gel permeation chromatography was used to estimate the molecular weights and molecular weight polydispersity of the graft copolymers by using CHCl3 as eluent. The results are also presented in Table 1. As expected, the polydispersity (Mw/Mn) of all these samples was estimated (by GPC) to be about 2. Although the cyclooctene-PEG macromonomer content in the copolymer calculated by 1H-NMR was a little bit less than its original addition amount, the contents of PEG in these copolymers are still tunable by changing the cyclooctene-PEG macromonomer incorporations.
Samples | [M]/[Cat] | m cyclooctene-PEG mol (%) | M n × 10−4/g mol−1 | PDI | M cyclooctene –PEG mol (%) |
---|---|---|---|---|---|
1# | 250:1 | 5 | 2.69 | 2.54 | 4.70 |
2# | 250:1 | 10 | 1.87 | 2.25 | 9.23 |
3# | 250:1 | 15 | 1.21 | 1.76 | 13.87 |
4# | 250:1 | 20 | 1.15 | 1.64 | 16.18 |
Fig. 3 displayed the FTIR spectrum of poly(cyclooctene)-g-PEG copolymer. The peak at 1110 cm−1 was attributed to C–O stretching vibration in the poly(ethylene glycol) chains. The peaks at 2925 and 2854 cm−1 were ascribed to the methylene stretch vibration in graft copolymer trunk chain.
Fig. 3 FTIR spectrum of poly(cyclooctene)-g-PEG. |
Fig. 4 and Fig. 5 were XPS wide scan spectra of poly(cyclooctene)-g-PEG film surfaces and their corresponding oxygen contents. The PEG graft contents on the film surface, which can be represented by the oxygen contents increased with the increase of the addition amount of cyclooctene-PEG macromonomer. This phenomenon was in accordance with the result that a higher content of PEG grafts in the surface led to a lower contact angle.
Fig. 5 The oxygen content of the poly(cyclooctene)-g-PEG surface. |
The water contact angle test is an effective approach for measuring the hydrophility of a polymer surface.15 Lower contact angle values represent higher hydrophilicity. As shown in Fig. 6, the contact angles gradually decreased from 87.7° for polycyclooctene to 65.8° for poly(cyclooctene)-g-PEG copolymer with increasing the PEG contents. The decrease of contact angles should be ascribed to the hydrophilic PEG molecule chains in poly(cyclooctene)-g-PEG copolymers.
Fig. 6 The contact angles of polycyclooctene (virgin) and poly(cyclooctene)-g-PEG (1#–4#). |
Bovine serum albumin (BSA) adsorption was one of the methods used to evaluate the biocompatibility of the materials.24 The relative amount of the protein adsorbing on each surface was measured according to XPS measurements. The N1s signal from the peptide bonds was used to mark the relative amount of protein adsorbing on the sample surface. Fig. 7 displayed the XPS N1s core-level spectra of the surfaces of polycyclooctene and poly(cyclooctene)-g-PEG after the protein adsorption in 10 mg mL−1 of BSA buffer solution for 24 h. As shown in Fig. 7, the intensity of the N1s peak component (at the BE of about 400 eV) for the virgin polycyclooctene film surface is much higher than those of poly(cyclooctene)-g-PEG copolymers with different PEG grafting degrees. This strong affinity of the polycyclooctene for BSA is probably due to the hydrophobic interaction of the protein molecules with the highly hydrophobic polycyclooctene surface.25
Fig. 7 XPS N1s core-level spectra of polycyclooctene (virgin) and poly(cyclooctene)-g-PEG (1#–4#), after exposure to a PBS buffer solution containing 10 mg ml−1 BSA for 24 h. |
There are some explanations for reduction of proteins adsorption of poly(cyclooctene)-g-PEG copolymers. The PEG molecules can be highly stretched in water because of its minimum interfacial free energy, hydrophilicity, high surface mobility, steric stabilization effects, unique solution properties and molecular conformation in water.26 These highly stretched PEG molecule chains would prevent protein molecules from approaching the film surface with strong steric exclusion force.27–30 This steric repulsive exclusion force between PEG molecule chains and protein molecules not only comes from the loss of conformation entropy of PEG molecule chains but also relates to the osmotic interaction between them when the protein molecules were approaching to the polycyclooctene surface.29,31,32
This journal is © The Royal Society of Chemistry 2011 |