Megan K.
Dorris
and
Craig E.
Lunte
*
R. N. Adams Institute for Bioanalytical Chemistry, Department of Chemistry, University of Kansas, 2030 Becker Drive, Lawrence, Kansas 66047, USA
First published on 26th October 2010
Co-Q10 is a lipid-soluble benzoquinone that is an important factor in free radical scavenging, mitochondrial membrane stability and ATP synthesis. Dietary Co-Q10 is a powerful antioxidant that has been useful in lessening the damage associated with ischemia-reperfusion injuries and aiding in the recovery of myocardial function after myocardial infarction. However, the role of dietary Co-Q10 in oxidative damage and repair is not well understood. Previous LC-EC methods have used packed carbon bed electrodes with high overpotentials that were sufficient to oxidize and reduce several biological compounds, thereby decreasing the selectivity that can be achieved with EC detection. Thin-layer cell dual electrode detection enables monitoring of reduced and oxidized forms of Co-Q10 simultaneously and selectively. The oxidation (+0.45 V vs.Ag/AgCl) and reduction (−0.4 V vs.Ag/AgCl) electrode potentials were optimized to oxidize and reduce the electroactive quinone moiety. The reduced form of Co-Q10 was prepared from the commercially available oxidized form using a Jones reductor. Confirmation of its formation was determined using the current ratios of the peak and half wave potentials from previously generated hydrodynamic voltammograms, using the oxidized form with electrodes in a series configuration. This analytical system was successfully applied to determine basal concentrations of oxidized (510 nM) and reduced (500 nM) Co-Q10 in human plasma. Peak identity of oxidized and reduced Co-Q10 was confirmed by two orthogonal methods: by the current ratios at +0.45 V and +0.25 V and −0.4 V and −0.2 V (vs.Ag/AgCl) as well as by retention time. Detection limits were determined to be 5 nM, with a linear range of three orders of magnitude.
Fig. 1 shows the mechanism for the reversible redox chemistry that occurs between ubiquinone-10 and ubiquinol-10.10,11 Due to the rapid redox cycling between ubiquinol-10 and ubiqinone-10 within the mitochondrial respiratory chain during oxidative stress, a method that can selectively and simultaneously detect ubiquinone-10 and ubiquinol-10 is necessary to establish the roles of this antioxidant in the maintenance of tissue homeostasis under stressful conditions.
Fig. 1 Mechanism of the electron-proton transfer of ubiquinone-10 and ubiquinol-10. |
Several methods have been developed for the simultaneous detection of ubiquinone-10 and ubiquinol-10 using liquid chromatography coupled with UV absorbance, fluorescence or electrochemical detection (EC). UV detection offers a simple and robust means to quantify analytes of interest, but requires significant sample pretreatment prior to analysis. In addition, ubiquinone and ubiquinol absorb at non-selective wavelengths (e.g. 254 nm) ,12–14 which in complex biological matrices, such as plasma, may result in interferences decreasing method selectivity. While UV detection is typically associated with low micromolar detection limits (∼0.5 μM); however, as Finckh et al. demonstrated, low nanomolar detection limits (∼4 nM) using UV absorbance can be obtained provided a sufficiently large sample volume is available.15 Alternatively, fluorescence detection offers a selective means to quantify ubiquinone-10 and ubiquinol-10 with low nanomolar detection limits (∼3.0 nM).16 However, sample derivatization prior to analysis is required. It has been shown that oxidation of ubiquinol-10 begins to occur within 1–2 h; therefore, any pre-treatment of the sample may result in conversion of ubiquinol-10 to ubiquinone-10, making the quantification of ubiquinol-10 inaccurate.17
EC detection provides direct and selective detection of ubiquinone-10 and ubiquinol-10. This detection technique requires no sample derivatization and is selective for analytes which are inherently electrochemically active. Previous techniques have used packed carbon bed electrodes for detection, in which a generator electrode was placed upstream from the analytical cell at a potential sufficient to completely oxidize (+0.7 V vs.Ag/AgCl) or reduce (−0.7 V vs.Ag/AgCl) the analyte of interest prior to the subsequent re-oxidation or reduction within the analytical cell10,11,15,17–28 In addition, this detection scheme contributes to increased dead volumes and residence times, resulting in increased retention times and band broadening.29 Furthermore, this detection scheme requires large sample volumes in order to overcome the large background noise associated with the increased surface area of these packed bed electrodes to obtain the low nano-molar to pico-molar detection limits typically associated with EC detection. Reduction and oxidation potentials frequently employed by these methods are known to be sufficient for the reduction of various solvents, while simultaneously oxidizing several other biologically relevant compounds often found in plasma or tissue samples decreasing the method selectivity and sensitivity. Alternatively thin layer flow cells improve mass transport efficiencies due to decreased electrode surface area and channel thickness.30 Therefore, the development of an analytical method with enhanced selectivity to qualitatively assess analytes of interest is necessary to further investigate the role of Co-Q10 in its ability to maintain tissue homeostasis during oxidative stress.
This report describes the development of an analytical method using liquid chromatography with amperometric detection for the simultaneous detection of ubiquinone-10 and ubiquinol-10 in human plasma. The method discussed offers decreased total analysis time when compared to previous methods, while maintaining low nanomolar detection limits.
Standard solutions were prepared by dissolving 0.4–1.5 mg of ubiquinone-10 in 1.0 ml, 1-propanol:H2O (9:1 v/v). Standard solutions of ubiquinol-10 were prepared from the corresponding ubiquinone-10 using a Jones reductor. The Jones reductor consisted of zinc dust activated with dilute sulfuric acid (0.5 M). Zinc dust (1.2–1.5 g) was placed into a 3 ml BD syringe (BD Inc., Franklin Lakes, NJ, USA) with a Millex GP 0.22 μm filter with a PES membrane (Carrigtwohill, Co. Cork, Ireland). Dilute sulfuric acid (2.0 ml) was flushed through the syringe to activate the zinc, followed by rinsing with an equal volume of water. An additional 1.0 ml of acid was added and flushed through immediately before the addition of the ubiquinone-10 standard (700 μM-1.2 mM) to ensure the reaction went to completion. Any residual ubiquinol-10 was rinsed with 250 μl 1-propanol:H2O (9:1 v/v) and was collected in a 1.5 ml microcentrifuge tube. Standards were stored in the dark at −20 °C through out the day. Reductors were rinsed with 1.0 ml of water and then reactivated using the same procedure described above for multiple uses. A consistent decrease in peak height of ubiquinol-10 (∼10%) was observed when compared to the peak height of the same concentration of ubiquinone-10. It was determined that the decrease in peak height of ubiquinol-10 was due to loss of product within the reductor and not due to the formation of ubiquinone-10 as a result of oxidation. The reuse of each reductor, for multiple reactions, was determined as well as the reproducibility of the reaction from reductor to reductor was evaluated. Deviation observed from reductor to reductor was selectively eliminated based on two criteria. First, solution color was taken into account since this has been reported previously as a means to determine formation of the quinol.17 Secondly, chromatographic verification was used to ensure that there was no quinone present. Only those reductors which produced a clear solution, as well as no chromatographic evidence of the quinone were used for future method development. The relative standard deviations, based on peak heights between reductors, as well as multiple uses of a single reductor, were determined to be 6.6% and 6.2%, respectively. It was concluded that no significant variation occurred in the preparation of each reductor to notably affect the reduction of ubquinone-10 to ubiquinol-10.
Separation was achieved using a Zorbax Bonus RP C18 column (2.1 × 100 mm, 3.5 μm) (Agilent Technologies, Santa Clara, CA, USA). The mobile phase consisted of MeOH/ACN/glacial acetic acid (59/39/2% v/v) with 55 mM sodium acetate at a flow rate of 0.75 ml/min. Oxygen was removed from the system by continuously purging the mobile phase with argon.
Previously developed methods have used mobile phases consisting of a wide variety of solvents (e.g.MeOH, EtOH, 1-propanol, IPA, etc.); however, these solvents often contain several trace elements that generally result in increased background noise and correspondingly poorer limits of detection. Initially, for the separation of ubiquinone-10 and ubiquinol-10 a mobile phase consisting of ACN:MeOH (75:25 v/v) with 50 mM lithium perchlorate, (utilizing a Phenomenex Gemini RP-C18, 150 × 2.0 mm, 5 μm) was employed as described by Tang et al.10 Under these conditions a total analysis time of 25 min was obtained with peak widths of over a minute. In order to improve efficiencies and decrease band broadening different ratios of ACN and MeOH were investigated. It was determined that varying the amount ACN and MeOH in the mobile phase had no significant effect on retention time or peak efficiencies. Therefore, the stationary phase was investigated to determine if changing the hydrophobicity of the column packing would decrease retention times and improve band broadening. The RP-C18 Zorbax Bonus-RP (3.5 μm, 2.1 × 100 mm) provided a triply end-capped stationary phase, minimizing non-polar interactions with the incorporation of a polar amide group in the alkyl chain, modifying the interaction of ubiquinone-10 with the stationary phase, when compared to traditional RP-C18 columns, which expose long chain hydrocarbons resulting in increased interactions with the poly-unsaturated backbone of ubiquinone-10. Using this column reduced total analysis times from 25 min to 8 min, while also decreasing peak width from ∼1.5 min to 0.62 min, respectively. The ratios of ACN and MeOH were subsequently varied to further decrease peak width and improve retention times. It was determined that a mobile phase consisting of 60:40 (% v/v) MeOH:ACN proved to give the optimal separation conditions by further reducing the retention times of ubiquinone-10 and ubiquinol-10 to 5 and 4 min, respectively. Table 1 summarizes the key parameters determined for this method. Although, there is a decrease in peak efficiencies from the initial conditions using a Gemini RP C-18 (N = ∼6600) to the Zorbax Bonus RP (N = ∼1600), column interactions were minimized resulting in decreased retention times and band broadening was reduced.
Although lithium perchlorate is a typical inorganic salt used in non-aqueous mobile phases, it was observed that over time precipitation of the salt from the mobile phase was accumulating at the surface of the electrodes resulting in cross-talk. Substitution of lithium perchlorate by sodium acetate (55 mM) eliminated the precipitation and resulting cross-talk while providing sufficient conductance to minimize noise.
As discussed previously, ubiquinol-10 is readily oxidized to ubiquinone-10 when exposed to light at room temperature. In addition, oxidation to the respective quinone moiety also occurs in the presence of oxygen. In order to minimize loss of target analyte, glacial acetic acid (2% v/v) was added to the mobile phase described above reducing the pH to ∼6. The addition of acetic acid in the mobile phase provided supplementary protons in solution reducing any potential oxidation of ubiquinol-10 to ubiquinone-10.
Fig. 2 Comparison of HDV's obtained directly and in the series configuration for ubiquionone-10 (■,□) and ubiquinol-10 (▲,△). |
In the series configuration selectivity is enhanced due to the ability to monitor chemically reversible processes at the downstream electrode. This is advantageous as chemically reversible redox couples, such as ubiquinone-10 and ubiquinol-10, can be easily distinguished at the downstream electrode from the chemically irreversible species, which more easily oxidize or reduce at the upstream electrode. This configuration was used to confirm the formation of ubiquinol-10 from ubiquinone-10 using the Jones reductor by comparing the current ratios at the peak and half wave potentials for a standard of ubiquinol-10versus endogenous ubiquinol-10. The same experiment was performed, as mentioned above, using endogenous samples and standards to demonstrate the selectivity of this electrode configuration. Ubiquonone-10 was used as a control for this experiment. The current ratios determined for ubiquionol-10 (i+0.3/+0.45) and ubiquinone-10 (i−0.25/−0.4) standards and endogenous ubiquinol-10 and ubiquinone-10 were 0.77, 0.42 and 0.72, 0.47, respectively. This shows that the Jones reductor produced only ubiquinol-10 and further demonstrates the selectivity of this electrode configuration. A representative chromatogram for the separation of ubiquinone-10 and ubiquiol-10, with the electrodes in series configuration, is shown in Fig. 3. Good baseline resolution was obtained under these conditions, and both compounds were detected in less than 5 min, decreasing retention times previously reported10,15,21,23,24
Fig. 3 Chromatograms showing the response of ubiquinone-10 and ubiquinol-10 using the series dual-electrode configuration. 3a, reduction electrode (−0.40 V vs.Ag/AgCl) is upstream and oxidation electrode (+0.45 V vs.Ag/AgCl) is downstream; 3b, oxidation electrode (+0.45 V vs.Ag/AgCl) is upstream and reduction electrode (−0.40 V vs.Ag/AgCl) is downstream. The separation was carried out using Zorbax Bonus RP C18 column (2.1 × 100 mm, 3.5 μm) with MeOH/ACN/gacial acetic acid (59/39/2% v/v), 55 mM sodium acetate at a flow rate of 0.75 ml/min. |
In this configuration two different potentials can be monitored for simultaneous detection of reducible and oxidizable species directly. Fig. 4 shows the response for ubiquinone-10 and ubiquinol-10versus the response of the sample matrix with the electrodes in a parallel configuration. In this configuration the analyte of interest is seen at both the oxidation and reduction electrodes simultaneously; therefore, no subsequent re-oxidation or reduction can occur, as shown in Fig. 4a and 4b. In addition, in this configuration qualitative information can be easily obtained. For the purpose of this method, this configuration qualitatively and simultaneously shows the redox state of endogenous Co-Q10, as shown in Fig. 4c, with increased sensitivity compared to previously published methods. Based on this reasoning the electrodes were used in the parallel configuration for further method validation.
Fig. 4 Chromatograms showing the response of ubiquinone-10 and ubiquinol-10 using the parallel configuration dual-electrode detection. Figure 4a injection of ubiquinone-10 and 4b injection of ubiquinol-10. Figure 4c injection of ubiquinone-10 and ubiquinol-10 using the parallel electrode configuration. Chromatographic conditions as listed in Fig. 3. |
The stability of ubiquinol-10 in samples was determined over the course of a 24 h period. Injections of the same standard concentration (1.0 μM) were performed every hour for 8 h. Peak heights were compared to determine any relative changes in ubiquinol-10. From this analysis, it was determined that there was no significant change in peak height over the course of 8 h. The standard was then stored at −20 °C overnight in the dark and the same standard was subsequently injected the following day. From this it was determined that after a 24 h period there was significant loss in the response of ubiquinol-10 and the formation of ubiquinone-10 was observed. The loss in response of ubiquinol-10 was determined to be a result of time and not due to any freeze/thaw cycle that occurred. Therefore, standards were prepared fresh daily.
Fig. 5 Detection of ubiquinone-10 and ubiquinol-10 in pooled human plasma using a parallel electrode configuration. Conditions as in Fig. 3. |
Spiked concentration (nM) | Extraction efficiency ubiquinone-10 | Extraction efficiency ubiquinol-10 |
---|---|---|
a n = 6 extractions/concentration. | ||
500 | 99.9% ± 4.3% | 100% ± 11.6% |
750 | 93.9% ± 7.1% | 89.0% ± 8.0% |
1000 | 87.8% ± 11.7% | 85.9% ± 9.7% |
This journal is © The Royal Society of Chemistry 2011 |