Liang
Qiao
a,
Hongyan
Bi
a,
Jean-Marc
Busnel
a,
Mohamad
Hojeij
a,
Manuel
Mendez
a,
Baohong
Liu
b and
Hubert H.
Girault
*a
aLaboratoire d'Electrochimie Physique et Analytique, Ecole Polytechnique Fédérale de Lausanne, Station 6, CH-1015, Lausanne, Switzerland. E-mail: hubert.girault@epfl.ch
bDepartment of Chemistry, Institute of Biomedical Sciences, Fudan University, Shanghai, 200433, P.R. China
First published on 25th June 2010
A specific multi-phosphopeptide enrichment method is presented to complement current phosphopeptide isolation strategies that normally bias mass spectrometry analysis towards the detection of mono-phosphopeptides. An adsorption model is developed that shows how the physical parameters of adsorption such as specific surface area, adsorption time, etc. control the extraction of mono- and multi-phosphorylated peptides on oxide materials such as alumina. Commercial alumina particles, alumina supported on aluminium particles and alumina on aluminium foils were used to enrich multi-phosphopeptides. The synthesized phosphopeptides and tryptic digests of casein and milk were employed as samples to validate the adsorption simulation. Both experimental and theoretical results show how the selective enrichment of multi-phosphopeptides can be achieved when using extractors with a high specific affinity for phosphate groups but with a relatively small loading capacity or specific surface area. We also show that a commercial aluminium foil represents an ideal substrate to enrich multi-phosphorylated peptides for laser desorption ionisation mass spectrometry.
During the past years, many techniques for isolating phosphorylated proteins or peptides from cell lysates and protein digests have been developed to simplify the analysis of protein phosphorylation. Among various chemical and affinity-based strategies,12–17 solid extraction methods have drawn wide interest due to their effectiveness as well as simplicity. One widely used solid extraction approach, namely immobilized metal affinity chromatography (IMAC),18–24 was introduced in 1975 by Porath et al.,24 relying on the interactions between metal–ligand complexes and phosphate groups. Another popular solid extraction approach is based on the special affinities between phosphate groups and metal oxides, such as ZrO2,5,6 TiO2,3,7 Nb2O5,25 and Al2O3.4,26 With these two principles, many studies have been reported to enrich phosphopeptides by using on-column,27 in-solution7 or on-target28,29 strategies.
For most of the extraction methods reported, both mono- and multi-phosphorylated peptides are collected together. Multi-phosphopeptides being more difficult to ionize than mono-phosphorylated ones, MS analysis normally detects rather exclusively mono-phosphorylated species from the enriched mixtures, neglecting many multi-phosphopeptides that are, however, very important for understanding biological pathways.30,31 To date, there are only a few reports of the specific analysis of multi-phosphorylated peptides.30,32–36
Here, we present a very simple yet very selective enrichment method for multi-phosphopeptides on commercial aluminium foils. Multi-phosphopeptides hold a stronger affinity toward solid extracting phases such as oxides compared with mono-phosphorylated ones due to the larger number of phosphate groups and therefore one should expect them to be preferentially enriched. However, the final enrichment is also regulated by adsorption/desorption kinetics and the loading capacity of extracting phases. Using a classic Langmuir isotherm to model the solid phase extraction processes, the physical factors influencing the enrichment results were studied. Commercial porous alumina particles, alumina supported on aluminium particles and alumina on aluminium foils were used as solid phase extractors with different specific surface areas. The synthesized phosphopeptides with different phosphorylation states were used to estimate the extraction rate and loading capacity of each extractor. Alumina on Al particles and foils, which have a relatively small specific surface area, were further used to selectively extract multi-phosphopeptides from the tryptic digests of α- and β-casein and drinking milk (Table 1). As a result, the obtained selectivity is very good and comparable with the best known methods using more complex functional materials.35
| Peptide sequence | Modifications | (M + H)+ |
|---|---|---|
| a PHOS stands for phosphorylation, and MSO stands for methionine sulfoxide. b The peptide sequences were confirmed by both mass measurements and literature reports except the MSO modified ones which were confirmed by mass measurements only. | ||
| TVDMESTEVFTKKTK, α-S2-casein (153–167) | 1PHOS, 1MSO | 1839.8 |
| DIGSESTEDQAMEDIK, α-S1-casein (58–73) | 2PHOS | 1927.7 |
| DIGSESTEDQAMEDIK, α-S1-casein (58–73) | 2PHOS, 1MSO | 1943.7 |
| YKVPQLEIVPNSAEER, α-S1-casein (119–134) | 1PHOS | 1951.9 |
| FQSEEQQQTEDELQDK, β-casein (48–63) | 1PHOS | 2061.8 |
| NVPGEIVESLSSSEESITR, β-casein (22–40) | 4PHOS | 2352.9 |
| QMEAESISSSEEIVPNSVEQK, α-S1-casein (74–94) | 3PHOS, 1MSO | 2577.0 |
| NTMEHVSSSEESIISQETYK, α-S2-casein (17–36) | 4PHOS | 2619.0 |
| NTMEHVSSSEESIISQETYK, α-S2-casein (17–36) | 4PHOS, 1MSO | 2634.9 |
| QMEAESISSSEEIVPNSVEQK, α-S1-casein (74–94) | 4PHOS | 2640.9 |
| VNELSKDIGSESTEDQAMEDIK, α-S1-casein (52–73) | 3PHOS | 2678.0 |
| VVRNANEEEYSIGSSSEESAEV, α-S2-casein (58–79) | 4 PHOS | 2704.9 |
| QMEAESISSSEEIVPNSVEQK, α-S1-casein (74–94) | 5PHOS | 2720.9 |
| QMEAESISSSEEIVPNSVEQK, α-S1-casein (74–94) | 5PHOS, 1MSO | 2736.9 |
| KNTMEHVSSSEESIISQETYK, α-S2-casein (16–36) | 4PHOS | 2747.1 |
| EKVNELSKDIGSESTEDQAMEDIK, α-S1-casein (50–73) | 3PHOS | 2935.2 |
| EKVNELSKDIGSESTEDQAMEDIK, α-S1-casein (50–73) | 3PHOS, 1MSO | 2951.2 |
| ELEELNVPGEIVESLSSSEESITR, β-casein (17–40) | 4PHOS | 2966.2 |
| NANEEEYSIGSSSEESAEVATEEVK, α-S2-casein (61–85) | 4PHOS | 3008.0 |
| NANEEEYSIGSSSEESAEVATEEVK, α-S2-casein (61–85) | 5PHOS | 3088.0 |
| RELEELNVPGEIVESLSSSEESITR, β-casein (16–40) | 4PHOS | 3122.3 |
| KNTMEHVSSSEESIISQETYKQEK, α-S2-casein (16–39) | 4PHOS | 3132.2 |
| QMEAESISSSEEIVPNSVEQKHIQK, α-S1-casein (74–98) | 4PHOS | 3147.2 |
800 Da, bovine pancreas) was obtained from AppliChem GmbH (Darmstadt, Germany). Aluminium (99%, powder) was obtained from Acros (Chemie Brunschwig AG). Aluminium oxide (99.99%; >95% α-form) was purchased from Fluka AG. All these reagents were used as received without further purification. Fresh milk and aluminium foil were purchased from a local grocery store. Deionized water (18.2 MΩ cm) was produced by an alpha Q-Millipore System (Zug, Switzerland).
:
30 (w/w), in 25 mM ammonium bicarbonate buffer). Digestion of milk was performed according to a reported method.28 30 μl fresh milk was diluted in 900 μl 25 mM NH4HCO3 aqueous solution. The solution was then centrifugated at 16
000 rpm for 15 min. The supernatant was saved for tryptic digestion. After denaturation at 100 °C for 5 min, the supernatant was incubated with 30 μg trypsin for 12 h at 37 °C for proteolysis.
![]() | ||
| Scheme 1 Schematic illustration of phosphopeptide adsorption/desorption processes on a solid extractor. | ||
The kinetic laws for such a model for a mixture of mono- and bi-phosphopeptides read:
![]() | (1) |
![]() | (2) |
Because the affinity between the phosphate group and alumina is strong, it is reasonable to assume kf ≫ kb and kf ≫ koff. The equations can then be simplified as:
![]() | (3) |
Similarly, the kinetic law for a mixture of mono-, bi-, tri- and tetra-phosphopeptides reads:
![]() | (4) |
The relationship between cb, the peptide surface coverage ratio P and the initial peptide concentration c0 is:
![]() | (5) |
According to eqn (5), the kinetics eqn (4) can be written as:
![]() | (6) |
![]() | (7) |
A simulation shown in Fig. 1a illustrates that specific extraction of multi-phosphopeptides from a given peptide-mixture can be realized by adjusting the amount of the extractor. The constant condition for Fig. 1a is Γmax = 10−5 mol m−2, A = 0.1 m2 g−1, V = 10−5 l, kon/koff = 106 M, ci0 = 10−6 M, and kb/kf = 0.01. However, the equilibrium enrichment result is also influenced by the assumed constants. Because kf should be much larger than kb, we keep the assumption that kb/kf = 0.01. The influence of AΓmax can be observed by comparing Fig. 1a and 1b. Keeping all constants the same except increasing A to 1 m2 g−1, the amount of extractor must be decreased by 10 times in order to selectively enrich only multi-phosphopeptides from the same sample. kon/koff can influence the extracted amounts of mono-phosphopeptides when the extractor is in excess, whereas the largest amount of extractors being employed for selective extraction of multi-phosphopeptides is always the same under various kon/koff ratios, as shown in ESI SI-1.† Similarly, selective extraction of only tetra-phosphopeptides or tri- and tetra-phosphopeptides should also be achieved by adjusting the amounts of extractors and influenced by Γmax, A, and kon/koff in a similar way.
![]() | ||
| Fig. 1 (a, b) Simulated phosphopeptide adsorption equilibrium results as functions of extractor-amounts, (a): extractors with small specific surface area, (b): extractors with large specific surface area; (c) Simulated phosphopeptide adsorption kinetics with zooming-out from left to right. In figure c left, P1 is overlapped with P2. In figure c right, P3 is overlapped with P4. | ||
Extraction kinetics are simulated in Fig. 1c to verify the feasibility of approaching the enrichment equilibrium. The constant condition is kon = 106 min−1 M−1, koff = 1 min−1, kb/kf = 0.01, Γmax = 10−5 mol m−2, A = 0.1 m2 g−1, ci0 = 10−6 M, V = 10−5 l and ms = 10−5 g. According to the equilibrium simulation in Fig. 1a, under such a constant condition, tetra-phosphopeptides should be selectively extracted. In Fig. 1c, it is found that adsorption rates are very fast for all the phosphopeptides, while desorption rates for bi-, tri- and tetra-phosphopeptides are much slower than that of mono-phosphorylated peptides due to the very small value of kb/kf. Therefore, from a kinetic viewpoint, selective enrichment of multi-phosphopeptides can be easily achieved while the separation of bi- and tri-phosphopeptides from tetra-phosphopeptides is rather impossible when there is a moderate value for koff. In the following experiments, it will be shown that bi-, tri- and tetra-phosphopeptides were always enriched together from α-casein and milk digests, validating the present kinetics simulation.
Combining the equilibrium and kinetics simulations, we can infer the following conclusions:
1. Selective adsorption of multi-phosphopeptides from a given sample can be realized by using any solid extractors having special affinities for the phosphate group via adjusting the employed amounts of extractors and incubation time.
2. Particles with large specific surface area and loading capacity should show bad selectivity while big particles with small specific surface area should be suitable for the selective enrichment of multi-phosphopeptides.
3. Separation of tetra-phosphopeptides from other multi-phosphopeptides is possible at equilibrium but not feasible in a time resolved manner, and all multi-phosphopeptides should be extracted with a similar rate during experiments.
4. Fast or moderate kinetics are preferred for mono-phosphopeptide desorption. Then, porous materials are not suitable for the selective enrichment of multi-phosphopeptides as their 3-D structures may limit the mass transport.
![]() | ||
| Fig. 2 TEM images of the commercial (a) Al@Al2O3 powder and (b) alumina powder. | ||
| Extractor | BET surface area/m2 g−1 | Average particle diameter/μm |
|---|---|---|
| Al2O3@Al | 0.11 | 21 |
| Al2O3 | 4.49 | 0.33 |
A simple standard peptide-mixture was firstly used to validate this model. The mixture was composed of four peptides in equal molar amounts, including angiotensin I, oxidized β-insulin, α_syn(107–140)_pY125 (P1) and α_syn(121–140)_pY125pS129 (P2). We chose here α_syn(121–140)_pY125pS129 instead of α_syn(107–140)_ pY125pS129 as the multi-phosphopeptide, because the dephosphorylated fragment, i.e. the mono-phosphorylated α_syn(107–140), can be generated from bi-phosphorylated α_syn(107–140) during the MALDI-TOF analysis, thereby influencing the detection of α_syn(107–140)pY125. Fig. 3a shows a mass spectrum of the peptide-mixture (900 nM each peptide) without any pre-treatment, where angiotensin I dominates the spectrum and the peaks of the phosphopeptides are very weak. Fig. 3b and 3c show mass spectra of peptides enriched from the mixture (900 nM each peptide, 20 μl) by 20 μg Al@Al2O3 and 20 μg alumina powders respectively. The incubation time was 1 h to make sure that enrichment equilibrium of P1 had been approached. On the spectrum, almost only doubly phosphorylated peptide α_syn(121–140)_pY125pS129 (m/z: 2527.4) is observed in the case of using Al@Al2O3 as the extractor. In contrast, several intensive peaks of mono-phosphopeptides also appear on the spectrum in the case of using alumina powder as the extractor. The peak with m/z of 3940.4 corresponds to the singly charged α_syn(107–140)_pY125. The peaks with m/z of 2284.2 and 1594.5 correspond to by-products generated during the synthesis of α_syn(121–140)_pY125pS129, which are indeed mono-phosphopeptides of DNEAEMPpSEEGYQDYEPEA and PpSEEGYQDYEPEA respectively. The experimental results are consistent with the estimates made previously by considering the simulation and BET results, proving that it is reasonable to predict the performance of extractors in phosphopeptide enrichment by using the current model.
![]() | ||
| Fig. 3 (a) Mass spectrum of 900 nM four-peptide-mixture without any pre-treatment, and mass spectra of peptides extracted from 20 μl 900 nM four-peptide-mixture by (b) ∼20 μg Al@Al2O3 powder and (c) ∼20 μg Al2O3 powder; *: multi-phosphopeptide related peaks; #: single-phosphopeptide related peaks. | ||
Phosphopeptide enrichments were also performed using various extractor-amounts to further confirm the deductions obtained from simulation. The concentrations of peptide mixture were fixed at 900 nM, the volume was fixed at 20 μl and the amounts of Al@Al2O3 were varied from 4 μg to 200 μg. The incubation time was always 1 h to make sure that the enrichment equilibrium of mono-phosphopeptides had been approached. As shown in the ESI SI-2,† nearly only multi-phosphopeptide α_syn(121–140)_pY125pS129 was observed in the mass spectra when the amount of Al@Al2O3 was less than 40 μg. A weak signal corresponding to α_syn(107–140)_pY125 was obtained when the amount of Al@Al2O3 was 100 μg. Whereas the enrichment had no specificity towards multi-phosphopeptides when the amount of Al@Al2O3 was as high as 200 μg. This result agrees with the simulation shown in Fig. 1a.
Adsorption–desorption kinetics were evaluated and it was found that the equilibrium for mono-phosphopeptides could be approached in 6 min in the case of using the 4-peptide-mixture as a sample and Al@Al2O3 as the extractor (ESI SI-4†). When the incubation time was only 1 min, a weak peak of α_syn(107–140)_pY125 could be observed. However, when the incubation time was longer than 6 min, almost only signals corresponding to α_syn(121–140)_pY125pS129 were obtained, agreeing with the simulation in Fig. 1c.
It should also be possible to selectively enrich multi-phosphopeptides by using alumina powders according to the extraction model. Considering that the specific surface area of alumina powder is ∼45 times larger than that of Al@Al2O3, very low amounts of Al2O3 particles should be used for specifically extracting multi-phosphopeptides. Indeed, as shown in Fig. SI-3,† selective isolation of multi-phosphopeptides from 20 μl of 900 nM four-peptide-mixture can be partially achieved when the quantity of alumina powder is lowered to 2 μg, agreeing with the simulation in Fig. 1b. However, such a small extractor amount is inconvenient to handle. Therefore, only Al@Al2O3 was further used to enrich multi-phosphopeptides from protein tryptic digests in the following parts.
![]() | ||
| Fig. 4 (a) Mass spectrum of peptides extracted from 20 μl 850 nM β-casein digest by ∼20 μg Al@Al2O3 powder; and (b) mass spectrum of peptides extracted from 20 μl 850 nM α-casein digest by ∼20 μg Al@Al2O3 powder; *: multi-phosphopeptide related peaks. | ||
Similar specific multi-phosphopeptide enrichment was also achieved when using α-casein digest as the sample. As shown in Fig. 4b, from 20 μl 850 nM tryptic digest, 10 multi-phosphopeptides were enriched by 20 μg Al@Al2O3 and identified by mass spectrometry, while no mono-phosphopeptide was observed, Fig. 4b, Table 1. Indeed, it is rare to find so many multi-phosphorylated peptides from α-casein digest with a single MS run and no pre-separation. Considering that the total concentration of multi-phosphopeptides here should be much larger than 900 nM, some samples were wasted in this case.
The specific multi-phosphopeptide enrichment strategy was also applied to milk protein digests (∼1 μM protein concentration, 20 μl). As shown in Fig. 5, 11 multi-phosphorylated peptides from milk digests were enriched by 20 μg Al@Al2O3 and identified, while no mono-phosphopeptides were obtained, further proving the unique specificity. This result is very different from that obtained by using a strategy developed previously for analyzing milk phosphoproteins: only 3 multi-phosphopeptides and 5 mono-phosphorylated ones were detected.28
![]() | ||
| Fig. 5 Mass spectrum of peptides extracted from 20 μl tryptic milk digest (∼1 μM protein concentration) by ∼20 μg Al@Al2O3 powder; *: multi-phosphopeptide related peaks. | ||
In this section, it has been proven that the equilibrium/kinetics controlled multi-phosphopeptide enrichment strategy is also successful in analyzing tryptic digests of standard protein and simple protein mixtures. With these results, it is logical to predict that this approach can also be used in the study of protein phosphorylation of complex samples, such as cell lysates. In order to extract as many multi-phosphopeptides as possible and avoid the co-enrichment of mono-phosphopeptides, the total amount of multi-phosphopeptides should be estimated and suitable quantities of extractors should be calculated. If the amount of multi-phosphopeptide is really low in a sample, extractors with very small specific surface area can be used.
Indeed, we have reported recently that commercial aluminium foil can be used as a disposable layer for the MALDI target plate with the advantages of convenience and low cost.42 Specifically, we printed the aluminium foil with TiO2 nanoparticles (20 nm) for on-plate enrichment of mono-phosphopeptides.42 Here, we use the bare Al@Al2O3 foil to on-plate enrich multi-phosphorylated peptides without any modification. As shown in Fig. 6a and b, the on-plate enrichment method is quite successful. 1 Multi-phosphopeptide from 85 fmol β-casein digest and 14 multi-phosphopeptides from 85 fmol α-casein digest were enriched and identified separately, while only 1 weak peak corresponding to single phosphorylated peptide was observed in each spectrum, illustrating the high specificity of the strategy. The detection limit influenced by both extraction efficiency and MS sensitivity is about 20 fmol, as shown in Fig. 6c and 6d, where 1 multi-phosphorylated peptide from β-casein digest and 5 multi-phosphorylated peptides from α-casein digest were identified.
![]() | ||
| Fig. 6 Mass spectra from (a) 1 μl 85 nM tryptic β-casein digest, (b) 1 μl 85 nM tryptic α-casein digest, (c) 1 μl 20 nM tryptic β-casein digest, and (d) 1 μl 20 nM tryptic α-casein digest after 30 min in situ phosphopeptide enrichment on Al@Al2O3 foil; *: multi-phosphopeptide related peaks; #: single-phosphopeptide related peaks. | ||
We have recently reported that the rotogravure-printed TiO2 spots on Al foil substrate are better at adsorbing multi-phosphopeptides compared to screen-printed TiO2 spots.42 Now with the present model, we can attribute this phenomenon to the difference in specific surface area. On rotograved TiO2 spots, we have a low coverage of TiO2 nanoparticles with part of the aluminium foil substrate still partially exposed, whereas the screen printed spots are a few microns thick forming a mesoporous structure with a very high binding capacity. The present model corroborates the previous observations that rotograved films resulted in a co-enrichment of multi- and mono-phosphopeptides, and that screen-printed resulted mainly in the enrichment of mono-phosphopeptides.
In spite of the advantages, such as facility and high selectivity, the present multi-phosphopeptide analysis method is still limited by the inherent drawbacks of MALDI-MS. Because of its poor reproducibility and limited compatibility with various tandem MS techniques, MALDI-MS is not suitable for complex biosample analysis or quantification analysis, and therefore we did not evaluate the adsorption percentages of multi-phosphopeptides. HPLC-ESI-MS combined with the current multi-phosphopeptide enrichment method could be more interesting with a view to proteome research. However, compared to ESI, MALDI holds the advantage in performing in situ ionization. With the Al foil or the previously reported TiO2–Al foil, it is possible to extract in situ mono- and multi-phosphopeptides from a piece of tissue, respectively, and to run MALDI-MS-imaging on these Al foils to investigate the surface phosphorylated species distribution.
Footnote |
| † Electronic supplementary information (ESI) available: Mathematical simulation and adsorption equilibrium and kinetics investigations. See DOI: 10.1039/c0sc00224k |
| This journal is © The Royal Society of Chemistry 2010 |