Abbas
Alshehabi
*,
Shinsuke
Kunimura
and
Jun
Kawai
Department of Materials Science and Engineering, Kyoto University, Sakyo-ku, Kyoto, 606-8501, Japan. E-mail: shehabi@material.mbox.media.kyoto-u.ac.jp
First published on 12th August 2010
A low-power (1.5 W) portable Bremsstrahlung X-ray reflectometer (XRR) has been designed, realised and tested. The purpose of this apparatus is to measure thicknesses of multilayers within an X-ray beam energy range of 1–9.5 keV in industrial and research environments. Experiments have been carried out in this range with a measurement time of 10 min. The reflectometer apparatus was set up aligning the X-ray tube, sample holder and Si-PIN detector in one plane. A Mo/Si (9.98 nm) multilayer sample was used in the measurement. The direct beam intensity at (0.00°) was measured. Intensity was measured at several glancing angles and reflectivity was calculated. Although one measurement is sufficient in a dispersive energy X-ray reflectometer (XRR), measurement was taken at 0.45°, 0.60° and 0.80°. The sample was tilted at an angle θ and the detector was linearly elevated corresponding to 2θ at each measurement. A calibration equation was proposed to fit the apparatus geometry. Experimental reflectivity was calculated and compared to theoretical results. The portable X-ray reflectometer (XRR) was proved feasible in multilayer nano-thickness measurement.
In the angle dispersive method; a monochromatic X-ray beam is used and the angular scan is carried out either mechanically or by means of a spatially extended detector such as an image plate. In the energy dispersive method; XRR makes use of a polychromatic X-ray beam and measures the energy by means of an energy sensitive detector, making the fixed geometry during data collection an advantage over the angular dispersive method. An attenuator is used to make X-rays reach the detector less intense. However using the energy dispersive method involves dealing with the simultaneous presence of photons having different energies, which interact in different ways with the sample material; it has been verified that continuum X-rays from a low power X-ray tube can achieve a 1 ng detection limit using a portable total reflection X-ray fluorescence (TXRF) spectrometer.9 There is still no available portable apparatus using XRR measuring the thickness of multilayers in the nano-scale range with a low power source. In this paper we prove that nano-thickness measurement is possible using a low power continuum X-ray source by a portable X-ray reflectometer.
In the present paper we designed and developed a portable X-ray reflectometer. Continuum X-rays are used in the measurement. For continuum X-rays, the spectrum is measured as a function of X-ray energy while it is measured as a function of glancing angle in case of monochromatic X-rays.10–12 The power of the X-ray tube used is weak (1.5 watts) compared with previous works in related papers e.g. in synchrotron and rotating anode X-ray tube.13
![]() | ||
Fig. 1 a) A schematic view of the portable X-ray reflectometer; fixed X-ray tube, sampled radially elevated θ and detector heightened at a position corresponding to 2θ; and b) A photograph of the portable X-ray reflectometer. |
The weight of the device, including the power supply amplifier, computer and other accessories, is less than 7 kg. The X-ray tube has a maximum voltage of 9.5 kV, a maximum current of 150 μA and a spot size of ∼50 μm (V) × 10 mm (H) (Manufacturer, Hamamatsu Photonics, model L9491 with W anode). The detector is an energy dispersive Si-PIN detector.14–19
A slit (150 μm) placed in front of the detector window, served to cut off the angular and spectroscopic tails of the continuum X-ray beam incident on the mirror. A Mo/Si multilayer sample (50 layers, d = 9.98 nm, dMo/d(Mo + Si) = 0.4) deposited by magnetron sputtering was measured.20 Elevating the sample, a white X-ray incident beam was fixed at a glancing angle θ. The reflected beam intensity was measured at an appropriate detector height corresponding to angle 2θ and reflectivity was calculated.
![]() | (1) |
![]() | ||
Fig. 2 Intensity versus X-ray energy at (0.00°, 0.45°, 0.60°, 0.80°). Intensity at 0.00° is the direct beam. |
![]() | ||
Fig. 3 Reflectivity versus X-ray energy at (0.45°, 0.60°, 0.80°). The decay of the total reflection is followed by the first Bragg peak. |
cos2θ and sin2θ are expressed in terms of intensity as illustrated as follows:
![]() | (2) |
![]() | (3) |
A(θ) is an empirical value supposedly dependent on the angular deviation, the experimental and theoretical curves are in good agreement when A(0.45°) = 0, A(0.60°) = 1.2 and A(0.80°) = 4. Fitting with the experimental results leads to a nonlinear response curve that can be fitted with good approximation with the following calibration equation:
![]() | (4) |
![]() | ||
Fig. 4 Reflectivity versus energy by the calibration equation at 0.60°. (a) Measured Reflectivity of experimental X-ray intensity versus X-ray energy compared with the net calculated reflectivity versus X-ray energy at 0.60°; ![]() ![]() ![]() ![]() |
The sharp peak at around 3 keV in the experimental spectrum is due to the Ar Kα peak. The experimental curves are less corresponding to the theoretical curves as the region of X-ray total reflection gets sharper. This can be to some extent due to instrumental effects like imperfect collimation and the degree of inaccuracy. The Bragg peak in the theoretical and experimental spectra is more corresponding at 0.45° than at 0.60°, which in turn is more corresponding than at 0.80°.
From Fig. 3 thickness is directly obtained by the Bragg condition; e.g. the Bragg peak at 0.45° (0.007rad) corresponds to 8.70 keV (0.14 nm); applying the Bragg condition (2d sinθ = nλ), thickness will be 9.53 ± 0.45 nm. The intensity difference calibrated in the suggested problem does not contribute to this value since thickness is deducted from the Bragg peak position.
This slight difference and the difference in theoretical and experimental spectra in all the measurements were due to some problems concerning the accuracy of analysis. Any slight misalignment of the detector height could result in a less accurate detection angle of the reflected X-rays. As a result; reflectivity distribution might have changed and the X-ray flux might not be strong enough at 2θ. A shift of peaks with increasing angle might have been due to this reason. This might have resulted in lower reflectivities at 0.60° and 0.80°. The intensity was very weak in the range below 2 keV because of the Be window. The mode of operation was a θ/2θ mode (where θ is the glancing angle and 2θ is the detection angle) which assures the incident angle is always half of the angle of diffraction. The detector is heightened linearly with an instrument error of a micrometre. This can overall be a reasonable measurement since 1–10% variation of thickness and density happens after multilayer fabrication due to surface effects. Bulk density is always assumed in the calculated curves.
Although the detection angle is moved linearly, the X-ray portable reflectometer is adequate for detecting the reflecting beams due to small glancing angles and a calibration eqn (4) has been proposed. Since polarisation makes negligible contribution at small angles; unpolarised X-ray beam was used in the experiment. Because the reflection at the surface and interface is due to the different electron densities in the different layers, which correspond to different refractive indices, an error is always present in a sample. The average density near the surface in a multilayer is less than the normal bulk value and presumably increases with depth. Oxidation can also occur during and after multilayer preparation; that might contribute to the error.
Since the (Mo/Si) multilayer surface was glossy and white X-rays were used, specular reflection might not have been maintained and a combination of specular and diffuse reflection might have reflected off the multilayer surface.
In the present paper we proved that continuum X-rays can be used for X-ray reflectivity measurement with a weak (1.5 watts) X-ray source. It provides the advantages of a continuum X-ray source in X-ray reflectivity; in terms of easiness of measurement; not associated with its usual problems related to strong intensity like sample damage, attenuation and provides less polychromatic effects. It also provides a portable apparatus for X-ray reflectivity. The portable apparatus presented is verified applicable in X-ray reflectivity measurement.
This journal is © The Royal Society of Chemistry 2010 |