Fernando
Cortés-Salazar
a,
Andreas
Lesch
b,
Dmitry
Momotenko
a,
Jean-Marc
Busnel
a,
Gunther
Wittstock
b and
Hubert H.
Girault
*a
aLaboratoire d'Electrochimie Physique et Analytique, Ecole Polytechnique Fédérale de Lausanne, Station 6, CH-1015, Lausanne, Switzerland. E-mail: hubert.girault@epfl.ch; Fax: +41 21-693 3667; Tel: +41 21-693 3145
bDepartment of Pure and Applied Chemistry, Center of Interface Science (CIS), Faculty of Mathematics and Natural Sciences, Carl von Ossietzky University of Oldenburg, D-26111, Oldenburg, Germany
First published on 3rd June 2010
A fountain pen probe has been developed to extend the scope of scanning electrochemical microscopy (SECM) experiments to dry surfaces. The fountain pen is fabricated by UV-photoablation of a polyethylene terephthalate (PET) film and consists on one side of one microchannel for flowing an electrolyte solution to the open tip, integrating a reference/counter electrode and on the other side a carbon track. The exposed tip of the track forms the working electrode located close to the microchannel outlet. The fountain pen can operate in a pointillist mode where a nanolitre droplet at the bottom of the probe connects it to a well-defined surface area to study locally the substrate, but can also operate in a scanning mode leaving a linear wet track of solution behind it to monitor the surface activity. The electrochemical characterization of the proposed fountain pen probe was performed by cyclic voltammetry, approach curves and lateral line scans over insulating and conductive substrates, showing that the flow rate and the probe-substrate distance have a major influence on its electrochemical behavior. An SECM image of a gold on glass EPFL logo is presented as a proof-of-concept that fountain pen probes can be employed for the detection of surface activity when scanning in a contact regime.
Additionally, since the SECM positioning system and the micrometre sized probe provide localized electrochemical characterization within areas ranging from nm2 to μm2, the study of electrochemical processes occurring at microenvironments can also be performed by SECM.30–32 In order to extract reliable information, different constraints have to be considered. On the one hand, the electrochemical cell has to be contained within a microscopic volume, which requires not only the use of small electrodes (i.e. working, reference and counter electrodes), but also a special arrangement of a miniaturized electrochemical cell in such a way that electrode collision does not occur upon scanning. On the other hand, considering that solute concentrations in low volume systems can vary drastically upon solvent evaporation, experimental artifacts may arise and surface modifications have to be minimized (e.g. to avoid precipitation processes).
Micrometer sized electrodes are readily prepared by different techniques, such as by sealing a microwire (e.g. platinum wire or a carbon fiber) either in a glass pipette, in a resin such as epoxy5,33 or in an insulating polymer film.34–36 Additionally, integrated three37 or two31,32 electrode cells have been introduced to simplify the electrochemical cell design, to impede collisions between electrodes and to allow an easier positioning of the electrochemical cell. The scanning droplet cell introduced by Lohrengel et al. is based on a three-electrode cell setup, where a capillary containing the reference and counter electrode delivers an electrolyte droplet on a conductive surface. The working electrode is thus defined by the wetted area.37 Despite common potentiostatic techniques can be employed with this setup for surface activity studies, it is restricted to the study of conductive substrates. Spaine and Baur reported a positionable microcell supported in a theta glass capillary.31 The working electrode located at one of the channels was composed of a carbon fiber sealed with epoxy, while a Ag/AgCl wire was positioned as the reference electrode at the other compartment within a concentrated solution of NaCl. This device was completed by a salt bridge at the bottom of the pulled glass capillary, which allows the electrical connection and a physical separation between the reference electrode and the analyte solution. Both voltammetric and SECM measurements were performed successfully at subnanolitre volumes. In spite of that, the preparation of this dual probe is cumbersome and a high expertise is required. Additionally, the mechanical stability of the probe is easily compromised by any probe-substrate crash. Another procedure for preparing integrated two electrode microcells was recently introduced by Turcu et al. where a common glass-encapsulated microelectrode was chemically coated with an Ag film and used as reference electrode.32 The main advantage of this dual system is the simplicity of the device and its preparation method. Steady-state voltammetry and SECM images of Pt bands in a nanolitre droplet were obtained as a proof-of-concept of the functionality of the proposed dual system. In order to decrease the influence of solution evaporation on the SECM imaging, Turcu et al. covered a nanolitre droplet of the redox mediator solution with a layer of paraffin oil, but with adsorption of this oil onto the microelectrode only between 3 to 4% of the initial current changes were observed after 50 min of continuous scanning on a Pt band microarray. With the purpose of avoiding any contamination, special attention has to be paid to the positioning of the redox mediator droplet, the microelectrode and then the covering mineral oil layer. The latter obliges the use of a physical barrier that precludes the paraffin leakage, limiting at the same time the scanned area. The use of a humidity chamber to provide a controlled water-saturated atmosphere or the addition of glycerin to the supporting electrolyte have been also employed with the purpose of decreasing the effect of evaporation on microsystems.31,38–41 Recently, Juncker et al. introduces a multipurpose microfluidic probe (MFP) which result from the coupling of microfluidics and scanning probes.42 The MFP is based on two apertures placed at the bottom of the device from where solution is injected and aspirated, creating a locally microfluidic flow that covers the sample surface. Taking advantage of microfluidic probes, evaporation problems could be overcome by flowing constantly solution to the gap in between the probe and the sample surface. A similar concept has been already applied in atomic force microscopy (AFM) for surface patterning, where a fluidic channel is introduced into the AFM cantilever, allowing the fluid transport from a reservoir to the end of the cantilever.43–46
The present work introduces a fountain pen probe for SECM, fabricated by ablating two parallel and aligned microchannels on opposite sides of a polyethylene terephthalate (PET) sheet. One of the microchannels is filled with carbon ink before coating and the carbon track tip is used as working electrode, while the other is covered by lamination and is used for flowing solution to the tip. An integrated two-electrode cell is completed by a third microchannel filled with Ag/AgCl ink and that is in physical contact with the solution flowing through the microchannel. This new probe is based on the same preparation method of the recently developed soft stylus probe,47 but introduces a fluidic microchannel and a counter/reference electrode (CE/RE) that may extend the scope of SECM to the scanning of dry surfaces in a contact regime. As a proof-of-concept a gold on glass EFPL logo was imaged by using the fountain pen probe in scanning contact regime.
For using the new probe in the contact regime, the probe was mounted with a slight tilt with respect to the sample surface thanks to the tapering and the stiffness of the probe. When the probe was pressed against the sample the polymeric probe bends in a way that the cross-section of the carbon track is closer to the sample than the opening of the microfluidic channel (inset of Fig. 3a, later). In order to enable imaging, a new routine was included in the SECMx software that allows examining a surface in contact and contact-less regime during the forward and the backward scanning, respectively. During the forward scan the probe is pressed against the sample, bent, and dragged over it. This is possible without damaging the studied surface, if the sample material is harder than the PET sheet from which the probe was made. After completion of the forward scan, the probe is lift off the sample surface by a preset stroke height so that the probe is freely suspended. It is then moved horizontally back to the start position of the line scan followed by an incremental step in the low frequency scan direction. The probe is brought again into mechanical contact with the sample by a vertical translation equal to the negative value of the stroke height and the next line scan in the contact mode commences. Images are constructed from the data acquisition of forward line scans in contact mode. By doing this, the probe bending degree and bending direction are always the same within one image frame when scanning in contact regime.
![]() | ||
Fig. 1 a) Schematic representation of the fountain pen probe used as working electrode for SECM; b) optical image of the carbon electrode exposed by mechanical polishing. |
The electrochemical characterization of the fountain pen probes was performed by cyclic voltammetry, approach curves and horizontal line scans over conductive and insulating substrates. Before each experiment, an aqueous solution of FcCH2OH in KNO3 was allowed to flow out until a drop of the solution covered completely the tip of the electrode and made a physical contact with the substrate (see inset in Fig. 2a and Fig. S1, ESI†). Fig. 2a presents the cyclic voltammogram of FcCH2OH at a fountain pen probe at a flow rate of 25 μL h−1 and 0 μL h−1. Since the electrode was placed around 500 μm above the sample, a steady state bulk situation can be assumed. The steady state currents obtained for both cases results from the pseudo-hemispherical diffusion present at the microelectrode. Despite the very low charging current observed, the slope of the rising part of the curve is lower than expected for a reversible redox couple due to a high ohmic drop between the CE/RE and the WE. This can be overcome by increasing the concentration of the supporting electrolyte, increasing the area of the CE/RE electrode and by reducing the actual separation distance between these two electrodes (i.e. 4 cm). According to Fig. 2a, the flow rate has little influence on the cyclic voltammetry when the electrode is located far away from the substrate, the small increase observed at 25 μL h−1 being most likely due to an enhanced mass transport.
![]() | ||
Fig. 2 a) Cyclic voltammetry at a fountain pen probe in 3.14 mM FcCH2OH and 0.1 M KNO3 at a flow rate of 25 μL h−1 (dashed lines) and 0 μL h−1 (continuous line). Scan rate 10 mV s−1. b) Experimental approach curve of a fountain pen probe (continuous line) and a soft stylus probe (dashed line) over unbiased gold electrode (1) and insulating glass (2). ET = 0.40 V, translation rate = 0.5 μm s−1 and flow rate = 25 μL h−1. The insets in Fig. 2a and 2b show the conditions in which the experiments were performed. |
Approach curves at a flow rate equal to 25 μL h−1 over an insulating glass and a conductive gold film were performed resulting in the curves shown in Fig. 2b (continuous lines), and compared qualitatively to approach curves (dashed lines) obtained in the same conditions, but using the carbon working electrode immersed classically in the electrolyte above the substrate. The latter case corresponds to the recently developed soft stylus probe.47 As expected a decrease of the current is recorded when the soft stylus probe is approached to an insulating substrate, while a current increase is observed when it is approached towards a conductive surface. In contrast, the fountain pen probe gives a continuous increase of the current profile in approach curves both, over glass or over gold. This is due to the fact that the blocking of the redox mediator diffusion is counterbalanced by the solution flux out of the microfluidic channel. Moreover, it has to be taking into account that the lateral flux to the electrode surface will become faster as the probe is located closer to the surface. As a consequence, an enhanced positive feedback behavior is observed. When the soft stylus probe touches the substrate surface, the probe is bended and the current changes as a function of the bending. It should be noted that the probe-sample distance change starts to differ distinctly from the vertical position change of the positioning system when the electrode physically touches the substrate. For this reason, the distance measured from the surface to the lowest point of the fountain pen (hP = hA − lT) was defined as the difference between the height of the attachment point of the fountain pen with respect to the sample surface (hA) and the length of the probe in the unbent state (lT). When the probe is pressed against the surface a tilt angle (α) between the sample surface and the probe will be established. Thus d is completely defined by d = hP + tL sin (α), where tL is the thickness of the polymeric film that is between the sample surface and the carbon electrode (see Fig. S2, ESI†).47 When the fountain pen probe is closer to the insulating glass, a more efficient blocking of the redox mediator diffusion is achieved. As a result, the redox mediator flux that is coming from the open microchannel is overcome and a decrease of the current is observed until the probe starts to be bended and the current does not depend anymore strongly on the vertical position of the positioning system. Conversely, an increase on the current takes place when the fountain pen probe is approached to the gold film as expected from positive feedback in conventional SECM setups. It should be noted that the fountain pen can cause positive feedback although the sample potential is not externally controlled. This can be expected if the area on the sample wetted by the electrolyte is substantially larger than the active electrode area of the fountain pen probe. In this situation a concentration cell can be established between the sample areas in the vicinity of the probe and more distant wetted regions on the conducting sample regions.52 The fact that the fountain pen probe presents a stronger positive and negative feedback than the soft stylus probe47 is explained by a more parallel alignment between the active electrode area and the substrate.
From Fig. 2b, it can be observed that for detecting of surface activity, the fountain pen probe must be used in a contact regime, where the highest current contrast is achieved. This is confirmed by performing lateral line scans over a glass–gold boundary at different hp values (Fig. 3a, all the scans were aligned according to the observed position of the glass–gold boundary). When the probe is at 10 μm from the substrate, no current difference is observed between the insulating and the conductive substrates. At hP equal to −10 μm and −20 μm, a current contrast between the insulating and conductive regions is observed as the blocking of the redox mediator diffusion or its recycling can not be shielded by the flowing solution. It is expected that the fountain pen probe is not limited to work in contact regime, since it might be also employed in a contact-less regime for micro-structuring and patterning of dry surfaces.
In contrast to the bulk-like situation (Fig. 2a), the flow rate has an important effect on the steady state currents, when the probe is in close contact with the otherwise dry substrate (Fig. 3). When the probe is approached to a conductive substrate, a slightly higher current increase is observed for lower flow rates (Fig. 3b). The latter might be explained by a lower dilution factor of the redox mediator concentration when lower flow rates are employed. In contrast, when approaching the probe to an insulating substrate, lower currents are recorded as lower flow rates are employed. This phenomenon is a consequence of a faster depletion on the redox species concentration, when less species are delivered to the gap in between the probe and the sample surface. Therefore, higher current contrast between active and non-active surfaces should be obtained when scanning with the fountain pen probe at lower flow rates, as it is confirmed in Fig. 3c. Fig. 3c shows several lateral line scans over a glass-gold boundary with a fountain pen probe working in contact regime (hP = −20 μm) for three different flow rates. Despite a better current contrast with a flow rate of 5 μL h−1, evaporation processes started to become crucial, since microchannel blocking by electrolyte precipitation took place. Flow rates values higher than 25 μL h−1 showed delamination of the PE/PET covering films due to the higher pressure required to achieve this flow rate. For this reason, flow rates between 15 to 25 μL h−1 are suggested as appropriate flow rates for SECM imaging with the fountain pen.
![]() | ||
Fig. 3 a) Lateral SECM line scans in feedback mode with a polished fountain pen probe over a glass partially covered by gold at a hP value of 10 μm (pointed line), −10 μm (dashed line) and −20 μm (continuous line). ET = 0.4 V, flow rate = 25 μL h−1, step size = 5 μm and translation rate vT = 25 μm s−1. 2.1 mM FcCH2OH with 0.1 M KNO3. b) Experimental approach curves of a fountain pen probe over unbiased gold electrode (1) and insulating glass (2) at a flow rate of 25 μL h−1 (continuous line), 15 μL h−1 (dashed line) and 5 μL h−1 (pointed line). ET = 0.4 V, translation rate 0.5 μm s−1. 2.1 mM FcCH2OH with 0.1 M KNO3. c) Lateral SECM line scans in feedback mode with a fountain pen probe over a glass partially covered by gold at a flow rate of 25 μL h−1 (continuous line), 15 μL h−1 (dashed line) and 5 μL h−1 (pointed line). hP = −20 μm, ET = 0.4 V, step size = 5 μm and translation rate vT = 25 μm s−1. 2.5 mM FcCH2OH with 0.1 M KNO3. The inset in Fig. 3a shows the orientation of the probe during forward scanning. |
It is important to recall that evaporation processes are a major issue when working in microvolumes of liquids. To avoid this major drawback, the use of a humidity chamber, the addition of glycerin to the supporting electrolyte and the covering of the aqueous microsystem with a mineral oil, have been proposed.31,32,38–41 A different approach can be proposed by using the fountain pen probe, since evaporation processes could be partially counterbalanced by the constant flow of solution from the open microchannel. With the aim of assessing this possibility, an EPFL logo prepared by evaporation through a mask was imaged under dry conditions by using a fountain pen probe in a contact regime (Fig. 4). This image was build by using a new routine included in the SECMx software, that allows to examine a surface in contact and contact-less regime during the forward and the backward scanning, respectively (see experimental part). The experimental time employed for this image was 2.24 h, during which no apparent changes of the recorded current due to evaporation process (i.e. drastically increase on the redox mediator concentration) were observed. The only problem arising from the evaporation of the supporting electrolyte is the deposited material on the sample that actually could affect the degree by which the fountain pen is bended. The latter is most likely the reason for the small shift of the recorded position of the letters “P”, “F” and “L” in the bottom part of the image, as compared with the optical picture of EPFL logo taken after the SECM image was obtained from this sample (Fig. 4b). Future work will be focused on the removal of the solution that is flowing out from the open microchannel using a push–pull system. Only small scratches on the gold film are observed in Fig. 4b. They were most likely present on the sample before imaging and may arise from handling the samples but not from the scanning movement because they are not parallel to the scanning movement and do not occur in a repetitive manner. This is not unexpected because the gold and other substrate materials are harder than the polymeric material employed for the preparation of the fountain pen probe. In Fig. 4a, the EPFL logo can be clearly observed due to the recycling of the redox mediator obtained over the conductive gold film (positive feedback) and hindered diffusion of the redox mediator over the insulating areas (negative feedback). The SECM image of the EPFL logo shows the feasibility of applying the fountain pen probe to surface activity characterization of dry surfaces. Additionally, the contact regime allows the fountain pen probe to study moderately tilted surfaces without observing any topographical artifacts, as the gold micro EPFL logo was mounted without any attempt of leveling the surface and a tilt would otherwise clearly visible on scan length of 1400 μm.
![]() | ||
Fig. 4 a) SECM image of a gold micro EPFL logo taken with a polished fountain pen in feedback mode. The fountain pen was lift off after each forward scan by a stroke height of 200 μm and placed back at the original vertical position of the scanning but shifted by 20 μm in y direction. hP = 30 μm, ET = 0.4 V, flow rate = 20 μL h−1, step size = 20 μm and translation rate vT = 20 μm s−1. 2.6 mM FcCH2OH with 0.1 M KNO3. b) Optical picture of the micro EPFL logo employed for this experiment. |
Footnote |
† Electronic supplementary information (ESI) available: Details of probe geometry and mounting and further explanation of the probe-substrate distance employed. See DOI: 10.1039/c0ay00096e |
This journal is © The Royal Society of Chemistry 2010 |