Xiuqian
Jiang
ab,
Mark
Waterland
ab,
Len
Blackwell
b and
Ashton
Partridge
*ab
aMacDiamid Institute for Advanced Materials and Nanotechnology, Private Bag, 11222, Palmerston North, New Zealand. E-mail: A.Partridge@massey.ac.nz; Fax: +64-6-350-5682; Tel: +64-6-350-5918
bInstitute of Fundamental Science, Massey University, Private Bag, 11222, Palmerston North, New Zealand
First published on 8th February 2010
A rapid quantitative immunoassay for estriol-16-glucuronide by Surface Plasmon Resonance (SPR) sensing has been developed and applied to urine samples from non-pregnant and pregnant subjects. The assay was based on a partially-purified polyclonal antibody (pAb) raised in sheep, which showed negligible cross-reactivity with estrone-3-glucuronide and estriol-17-glucuronide. Colloidal gold coated by the pAb was used as the signal generator in the SPR-based inhibition immunoassay. An estriol-16-glucuronide-ovalbumin conjugate with an oligoethylene glycol (OEG) as linker was used to immobilize the steroid on the biosensor chip surface. The SPR assay had a limit of detection of 0.016 ng/mL, and could be performed rapidly giving results in two minutes. The assay can be carried out directly on any urine samples without complicated sample pretreatment. A one-step lateral flow strip test was also developed using the same pAb nanogold conjugates and bovine serum albumin estriol-16-glucuronide conjugates as the capture agent spotted onto a nitrocellulose membrane as the test line. A sensitive and repeatable lateral flow assay was achieved with a limit of detection of 0.49 ng/mL in time-diluted urine using a low coating concentration of the polyclonal antibody. Despite the strip sensor displaying adequate sensitivity in a standard curve generated by exposure to estriol-16-glucuronide in a spiked urine blank, the application of the strip sensors to real urine samples was not so successful due to matrix effects.
Estriol-16-glucuronide (E3-16G) is a steroid derived from ovarian estradiol by hepatic hydroxylation and conjugation and has been proposed as an alternative urinary metabolite to estrone glucuronide for the monitoring of ovarian function.1 Metabolism of estradiol produces three estriol glucuronides, one conjugated at position 3, one at position 17 and one at position 16. Of the three, E3-16G is the preferred metabolite for monitoring fertility, since it is excreted rapidly whereas the estriol-3-glucuronide undergoes a complex entero-hepatic re-circulation before excretion and hence is delayed with respect to E3-16G.2 E3-16G is also produced by the feto-placental unit during pregnancy and is the predominant estrogen in the urine of pregnant women increasing in parallel with the healthy growth of the foetus.3 It has also been associated with breast cancer.4 Much work has been done on development of assays for the measurement of estriols in the urine of pregnant women,5 but there are few direct assays with the sensitivity to measure E3-16G levels in the urine of non-pregnant women. Detection methods for E3-16G have traditionally been based on radioimmunoassay (RIA),6,7 high pressure liquid chromatography (HPLC),8 and liquid chromatography (LC) coupled with fluorescence measurement,9 mass spectrometry (MS) or UV spectrophotometry.10,11 However, these analytical methods require many clean-up steps, are time-consuming and may be expensive.
An accurate nanoparticle-enhanced SPR biosensor-based assay suitable for measurement of E3-16G in liquid samples has been described and validated in a previous study with a sensitivity close to 0.1 nmol/24 h (14 pg/mL).12 Baker (1979) showed that the mean value of the mid cycle peak of E3-16G excretion using a direct radioimmunoassay method was approximately 75 nmol/24 h.13 Therefore, the sensitivity of the SPR assay is more than sufficient to measure the low levels of E3-16G encountered in urine samples during the human menstrual cycle. A direct assay such as the SPR assay and LFIA is simpler and cheaper to perform than methods which need extraction or derivation but it introduces the possibility of matrix effects. All immunoassays are subject to matrix effects which can interfere with antibody binding reactions and hence give rise to false values. Measurement of a physiological marker such as E3-16G, which is present in low levels in a bodily fluid such as urine, makes the acquisition of a blank sample very difficult to obtain since there is always some analyte present, even in a child's urine.14 However, the sensitivity of the enhanced SPR assay is sufficient to allow for dilution of the urine samples and hence minimization of matrix effects.15
Despite widespread use, a LFIA for E3-16G suitable for use with urinary samples obtained during the normal menstrual cycle has not been reported. The aim of this work is to apply the SPR assay to detect low levels of E3-16G in clinical samples for use as a reference assay. This paper describes the validation and application of a SPR assay and attempts to develop a LFIA, with the use of a common colloidal gold-labeled anti-E3-16G pAb reagent, for the determination of E3-16G in human female urine.
A BIAcore X-100 system (GE Healthcare Uppsala, Sweden) was used to monitor the real-time binding performance of polyclonal anti-E3-16G antibodies and gold conjugates of a polyclonal anti-E3-16G antibodies to an E3-16G-OEG-OVA coated surface. Reagents for SPR work, including the CM5 chip, amine coupling kit (0.1 M NHS, 0.4 M EDC and 1 M ethanolamine) and HBS-EP running buffer (0.01 M HEPES pH = 7.4, 0.15 M NaCl, 3 mM EDTA and 0.005% surfactant P20) were purchased from GE healthcare (Uppsala, Sweden). An IsoFlow™ Reagent Dispenser (Imagenetechnology, Hanover, USA) was used to dispense control and test lines on a nitrocellulose membrane. The strips were cut with an M-70 cutter (advanced Sensor System Ltd, Ambala, India). An HP scanjet 3500c photo scanner and dedicated software were used to obtain images of the strips.
:
5 dilution of 10 mM TM buffer, pH = 7.10), E3-17G or E1-3G standards, and pAb (15 μg/mL) according to a volume ratio of 1
:
1
:
1 and incubating for 5 min at room temperature. Then the samples (15 μl) were injected for 90 s. The surface was regenerated using a mixture of 100 mM NaOH and 10% acetonitrile in Milli.Q water (5 μl for 30 s). Each sample was performed in triplicate and the calculations of the IC50 values were performed using Graph Pad Prism.
:
5, 1
:
10, and 1
:
30) of TM buffer (10 mM) or time-diluted blank urine were pre-incubated with an equivalent volume of a zero standard (HBS-EP buffer) and Au-pAb conjugates (1
:
100 dilution in HBS-EP buffer) at room temperature and the SPR response measured (20 μl/120 s at a flow rate of 10 μl/min) in triplicate. The chip surface was regenerated using NaOH (100 mM) and 20% acetonitrile in Milli.Q water (10 μl for 60 s).
:
30 dilution of each spiked urine specimen was incubated with an equivalent volume of a 1
:
30 diluted TM buffer (10 mM) and a 1
:
100 dilution of Au-pAb conjugates for which the nominal pAb concentration was 1.87 μg/mL (187 μg/mL was divided by the dilution factor of 100). The SPR response for the mixture (20 μl for 120 s) was tested in triplicate at a flow rate of 10 μl/min. Between exposures the biosensor surface was regenerated with NaOH of 100 mM and 20% acetonitrile in Milli.Q water (10 μl for 60 s).
:
30 from volunteer A or further diluted as shown in Table 1 from volunteer B), a 1
:
30 dilution of 10 mM TM buffer and a 1
:
100 dilution of the gold conjugates (CpAb = 1.87 μg/mL) flowed through the chip in 120 s at a flow rate of 10 μl/min. The chip surface was cleaned with 100 mM NaOH and 20% acetonitrile in Milli.Q water (10 μl for 60 s).
| Time-diluted sample | DF in SPR format |
|---|---|
| Week 4 | 1 : 300 |
| Week 6 | 1 : 60 |
| Week 8 | 1 : 300 |
| Week 10 | 1 : 1500 |
| Week 12 | 1 : 4000 |
| Week 14 | 1 : 12000 |
| Week 18 | 1 : 12000 |
| Week 24 | 1 : 25000 |
| Compounds | IC50 (ng/mL) | Cross-reactivity (%) |
|---|---|---|
| E3-16G | 0.93 12 | 100 |
| E3-17G | 5.15E4 | 1.80E-2 |
| E1-3G | 1.75E9 | 5.30E-8 |
In order to enhance the sensitivity of the SPR assay for E3-16G, Au nanoparticle-pAb conjugate (Au-pAb) was used according to the method of Jiang et al.12 The matrix effect of the urine samples was established in the SPR by measuring the binding response of E3-16G in different dilutions of time-diluted blank urine (Table 3) compared with buffer controls. The results for the 1
:
5 dilution showed a decrease in binding of approximately 50% when compared to the control, and indicated the extent of interference by the matrix. Further dilution to 1
:
10 and 1
:
30 showed the lowering of the matrix effects down to 36.9% and 4.2%, respectively. Subsequent assays were carried out using a dilution factor of at least 1
:
30.
| Samples | Response of samples (RU) | ||
|---|---|---|---|
1 : 5 dilution |
1 : 10 dilution |
1 : 30 dilution |
|
| Control | 115.2 ± 4.4 | 129.1 ± 2.7 | 144.3 ± 3.0 |
| Urine sample | 57.7 ± 2.8 | 81.5 ± 0.4 | 138.3 ± 0.5 |
A reference calibration curve (standard curve, Fig. 1) was generated using the Au-pAb conjugates (1
:
100 dilution), time-diluted blank urine (1
:
30 dilution) and a series of E3-16G standards prepared in HEP-BS buffer. The sensitivity (LOD) of the curve was 16 pg/mL and the IC50 value was 49 pg/mL. These were similar to the values reported previously for the buffer-only standard curve (LOD of 14 pg/mL)12 and the intra-assay coefficient of variation (CV) was ≤ 3.8%. Thus, there was no discernible matrix effect in the presence of the diluted blank urine standard.
![]() | ||
| Fig. 1 Normalized SPR reference calibration curves performed in the presence of the time-diluted blank urine (■) and TM buffer (▼). All values are mean ± SD from triplicate measurements of each sample. | ||
To determine the accuracy of the assay, E3-16G standards of 1.29, 3.87, and 6.45 ng/mL (10, 30, and 50 nmol/24 h) spiked in blank urine were diluted 1
:
30 with HBS-EP buffer, and the recoveries were measured in triplicate based on the reference curve (Fig. 1). The recoveries ranged from 92% to 108% with the highest CV being 0.71% (Table 4). The acceptable recovery of the standard samples spiked in blank urine indicated the high level of repeatability and accuracy of the SPR assay.
| Spiked E3-16G (ng/mL or nmol/24 h) | SPR | Lateral flow test | ||||
|---|---|---|---|---|---|---|
| Mean ± SD (ng/mL) | Recovery (%) | Intra-assay CV (%) | Mean ± SD (nmol/24 h) | Recovery (%) | Intra-assay CV (%) | |
| 1.29 ng/mL (10 nmol/24 h) | 1.191 ± 0.008 | 92 | 0.71 | 12.04 ± 1.91 | 120 | 15.88 |
| 3.87 ng/mL (30 nmol/24 h) | 4.172 ± 0.014 | 108 | 0.34 | 29.23 ± 3.33 | 97 | 11.40 |
| 6.45 ng/mL (50 nmol/24 h) | 6.960 ± 0.039 | 108 | 0.56 | 49.16 ± 3.14 | 98 | 6.39 |
Fig. 2 shows two standard curves performed with spiked E3-16G standards in 10 mM TM buffer (pH = 7.10) and blank urine (pH = 6.77). Both curves were plotted with the mean ± SD for duplicate measurements. Surprisingly, the presence of the blank urine led to a higher intensity on the test strips. For the buffer curve, the working range was from 5.2 nmol/24 h to 153.7 nmol/24 h (R2 = 0.99) with an IC50 value of 33.10 nmol/24 h. For the urine curve, the working range was between 11.1 nmol/24 h and 266.81 nmol/24 h (R2 = 0.99) with an IC50 of 54.22 nmol/24 h. The intra-assay CVs were lower than 15% for the buffer curve, and 13.7% for the blank urine curve. To investigate the observed results, a series of experiments were carried out over a range of pH values indicated that the higher intensity of the blank urine samples was independent of the pH.
![]() | ||
| Fig. 2 Buffer (▼) and urine blank (■) standard curves obtained from test strips, calibrated with duplicate measurements. Error bars represent the SD of the measurements. | ||
To evaluate the reproducibility of the test strips and reliability of the synthesis of the Au-pAb conjugates, two sets of conjugates were prepared in triplicate and deposited with different concentration of pAb (Set A (CpAb = 187 μg/mL) and Set B (CpAb = 73 μg/mL)). Standard curves were generated for each set and the LOD, IC50 and the inter-assay CVs were determined as shown in Table 5. For the conjugates of set A, the highest intra-assay CV (not shown) was 9.6%, the inter-assay CV of the LOD (40.7 ± 2.8 nmol/24 h) and IC50 (119.7 ± 3.2 nmol/24 h) were 2.7% and 6.9%, respectively. For the tests carried out with the conjugates of set B, the intra-assay CV was in the range of 0.18% to 11.9% (not shown), the inter-assay CV of the LOD (3.8 ± 1.3 nmol/24 h) and IC50 (26.1 ± 2.4 nmol/24 h) were 9.00% and 34.89%. The results demonstrated the high stability and reproducibility of the LFIA format using the Au-pAb conjugates as signal generators. The lower LOD and IC50 for set B showed that the assay sensitivity was improved by applying a lower concentration of the pAb coating. Thus subsequent LFIA analysis of the urine samples used the lower concentration of pAb in the conjugates.
| Conjugates | Set A (CpAb = 187 μg/mL) | Set B (CpAb = 73 μg/mL) | ||||||
|---|---|---|---|---|---|---|---|---|
| Conj A1 | Conj A2 | Conj A3 | Inter-CV | Conj B1 | Conj B2 | Conj B3 | Inter-CV | |
| IC50 (nmol/24 h) | 121.9 | 121.1 | 116.0 | 2.67% | 24.37 | 25.12 | 28.78 | 9.04% |
| LOD (nmol/24 h) | 43.44 | 37.87 | 40.83 | 6.85% | 2.55 | 3.69 | 5.20 | 34.86% |
The accuracy of the strip assay was determined by measuring the recoveries of the same three control samples (10, 30, and 50 nmol/24 h E3-16G spiked in blank urine). The recoveries ranged from 97% to 120% (Table 4). The higher recovery (120%) and variance (16%) values for the 10 nmol/24 h control was thought to correspond to the absorption value being close to the lower end of the straight-line portion of the calibration curve.
![]() | ||
| Fig. 3 A) E3-16G concentration obtained using the SPR biosensor (▲, left Y axis) and LFIA strips (■, right Y axis). The mean intra-assay CV of strip pattern was 11.7 ± 8.8%. All values are mean ± SD from triplicate measurements of each sample. B) Excretion pattern of PdG (▼, right Y axis) compared with E3-16G concentration values obtained from the SPR biosensor (▲, left Y axis). | ||
The measured range of E3-16G excretion rates (from 27 nmol/24 h (day 5) to 90 nmol/24 h (day 17)) shown by the SPR assay was consistent with the results reported by Baker et al.13 However the excretion rates determined using the test strips were much lower than anticipated, being in the range of 0.2 nmol/24 h to 9.2 nmol/24 h. This was presumed to be due to the matrix effects of the adult urine sample. Although the absolute values of E3-16G obtained using the two techniques were in a different range, the profiles were remarkably similar. Both patterns showed that the E3-16G excretion rate reached a maximum (around 90 nmol/24 h in SPR pattern and 10 nmol/24 h in LFIA pattern) on day 17. Moreover, several common smaller peaks occurred after the main peak. The rise in E3-16G excretion rate on day 12 that led to the pre-ovulatory peak on day 17 identified the beginning of the fertile window.13 The PdG excretion rate exceeded the threshold value of 7 μmol/24 h identifying the end of the fertile window on day 23 giving a fertile period of 11 days.19 This threshold also confirms that the E3-16G peak on day 17 is the ovulatory estrogen peak. The smaller post-ovulatory peaks are consistent with the fact that the excretion levels of E3-16G in the luteal phase are usually lower than that in the follicular phase.13 The decline in E3-16G levels for the first 3 days is consistent with the observation that in some cycles the levels are still declining after bleeding commences as a continuation of the previous luteal phase.13 The luteal phase length defined as the number of days from the mid-cycle estrogen peak to the day before the next bleed was 14 days, and is in agreement with the fact that the mean luteal phase length in normal menstrual cycles is 14 ± 2 days.20 The likely day of ovulation is the day after the E3-16G peak, thus the direct assay for E3-16G identifies the key cycle parameters equally as well as measurements of E1-3G.21
The excretion rates for a second menstrual cycle (not shown in Fig. 3) were also measured with the strip assay and gave a maximum excretion rate on the peak day of 11 nmol/24 h. The excretion rates obtained with the strip assay were mainly out of the linear regression range (4.2 to 198.2 nmol/24 h) of the calibration curve, and therefore the intra-assay CV was high. The comparison of the SPR and strip results reveal that the matrix effect of the adult urine sample is different from that in the “blank” child's urine, and highlights the difficulties associated with acquiring a truly blank adult urine sample. The question of why the matrix effects were more predominant in the strip assay resulted from the high sensitivity of the SPR assay giving a reliable response at a greater dilution factor. As a consequence, the matrix effect of the adult urine in LFIA format was not removed.
Besides the application of the direct E3-16G assay to urinary samples from a normal menstrual cycle, eight weekly urine samples from the first two trimesters of a pregnant woman (volunteer B) were tested using the SPR biosensor. The daily production of E3-16G before delivery is some 2000 fold greater than normal cycle luteal-phase excretion.13 Thus the analyte concentration of urine samples from pregnant women vastly exceeds the working range of the calibration curve shown in Fig. 1 (0.12 to 1.13 nmol/24 h). The determination of E3-16G in these samples necessitated pre-dilution (see Table 1) and Fig. 4 shows the patterns of E3-16G excretion throughout the first and second trimester of the pregnancy. The insert in Fig. 4 displays the patterns for weeks 4 to week 12, since the output of E3-16G in the first trimester was much lower than that in the second trimester and hence a separate scale is necessary for these data. The data clearly show how the output of E3-16G rises steadily from near the end of the first trimester (week 8) to almost doubling every 2 weeks.
![]() | ||
| Fig. 4 Excretion of E3-16G during the first and second trimester determined using the SPR biosensor (▲). All values are mean ± SD from triplicate measurements of each sample. The intra-assay CV of 3 measurements was lower than 4.9%. | ||
The present work demonstrates that the SPR biosensor-based assay can be applied to the measurement of E3-16G in urine, to give a rapid, precise and accurate analytical tool that requires minimal sample preparation. It has the potential to be used as a reference assay for E3-16G, however, a larger sample set of menstrual cycles would need to be analysed for complete validation of the system. No estriol assay for the menstrual cycle levels of E3-16G is available for a comparison of the present results with a reference assay.
| This journal is © The Royal Society of Chemistry 2010 |