aElemental Bio-imaging Facility, Department of Chemistry, Materials and Forensic Science, University of Technology Sydney, Sydney NSW 2007, Australia. E-mail: philip.doble@uts.edu.au; Fax: +61 2 9514 1460; Tel: +61 2 9514 1792
bDivision of Immunology and Rheumatology, Stanford University School of Medicine, 300 Pasteur Drive, Stanford, California 94305-5166, USA
cAgilent Technologies, Inc., Integrated Biology Solutions Unit, Santa Clara, California, USA
First published on 13th February 2009
Laser ablation inductively coupled plasma mass spectrometry (LA ICP-MS) was employed to image deposits of calcium phosphate based crystals in knee cartilage and synovial fluid from arthritic patients. A reaction/collision cell containing hydrogen minimised plasma interferences on calcium and also improved the image quality without significant sensitivity reduction. Areas of high calcium and phosphorus intensities consistent with crystal deposits were observed for both the cartilage and synovial fluid samples. These areas were also characterised by high magnesium and strontium intensities. Distribution patterns of other elements such as copper and sulfur did not correlate with the crystal deposits. Filtered and non-filtered solutions of calcium phosphate crystals grown in synthetic synovial fluid were also imaged as further evidence of crystal deposits. The crystal deposits were detected in the unfiltered solution, and were absent from the filtered solutions.
The identification of specific crystal types in synovial fluid is currently the most popular method of diagnosing all common forms of crystal-associated arthritis.3 In general, crystal arthritides cannot be distinguished from other causes of arthritis based on history or physical examination.5 Despite the multitude of techniques that have been applied to BCP and CPPD crystal detection, clinical diagnosis almost exclusively relies on light microscopy (LM). However, LM lacks the resolution required to image small crystals (<1 μm), is non-specific (though CPPD crystals can generally be distinguished under polarised LM) and accuracy is operator dependent.3,5–7 Sensitivity can be improved by staining the sample with calcium specific dyes, particularly for BCP crystals, which are often too small to be seen by LM. However these are often subject to false positives.
Other techniques for the determination of BCP crystals include scanning electron microscopy (SEM) and transmission electron microscopy (TEM). These methods are expensive, complex and not widely available for routine analysis in a clinical setting. Techniques recently applied to crystal detection and identification include atomic force microscopy (AFM), Fourier-transform infrared spectroscopy (FTIR) and Raman spectroscopy. AFM and Raman show great potential for this application. Additionally, FTIR and Raman spectrometers may be made compact, making the technology capable of point-of-care operation.3 A comprehensive review of the analytical tools available for the detection and identification of calcium phosphate crystals was recently published by Yavorskyy et al.3 The review made it evident that a simple, inexpensive technique which is capable of detecting and identifying calcium phosphate-based crystals accurately and with a high sensitivity would be invaluable in the diagnosis of crystal associated arthritides.
Laser ablation inductively coupled plasma mass spectrometry (LA ICP-MS) is increasingly applied to the study of metals in biological samples.8,9LA ICP-MS offers direct multielemental analysis of solids and liquids with trace and ultra trace detection capabilities, and a linear dynamic range of seven or more orders of magnitude. Though its application to imaging of soft tissues is emerging, it is considered a mature technique for the analysis of geological and metallurgical samples.10 In LA ICP-MS systems, a focused laser beam is used to mobilise sample material as droplets or vapour from the sample surface.9 The material is then transported to the plasma (usually by argon carrier gas) where the material is ionised and carried through to the mass spectrometer, which selectively detects ions at a given mass-to-charge ratio.
LA ICP-MS may be utilised to construct maps that provide elemental spatial information at trace and bulk levels of biological samples such as small tumours in rat brains,11 β-amyloid plaques in mouse models of Alzheimer’s disease,12 and the distribution of platinum in mouse kidneys treated with cis-platin.13 The environmental sciences have also utilised the potential of this technique to analyse the environmental changes or pollutant accumulation evident from the changing metal concentrations in increment growth layers of many types of organisms such as scales, shells, otoliths and tree xylem.14
Recently, Chaudhri et al.,15 determined the elemental distribution of bladder and kidney stones by LA ICP-MS. LA ICP-MS has also been employed to study element time profiles in teeth16 and bones,17 and to determine strontium ratios in calcium phosphates.18 Quantitative analysis of heavy metals in bone employing reference standards and spiking hydroxapatite standards has also been reported.17,19,20 However, LA ICP-MS has not been used to image crystal deposits in biological samples. Element concentrations in synovial fluids from osteoarthritic patients have been determined by ICP-MS.21
This study demonstrates the potential of LA ICP-MS for crystal detection in biological samples. Cartilage sections were obtained from knee or hip arthroplasty and were imaged by LA ICP-MS for the presence of crystal-associated elements. Additionally, eight synovial fluid samples, six from patients diagnosed with osteoarthritis (OA) and two from patients diagnosed with rheumatoid arthritis (RA), were analysed by the same method. Results from each sample type are presented here. We propose that this method of analysis be termed “Elemental Bioimaging”.
Laser ablation of biological samples was performed using a frequency quintupled Nd:YAG laser with ablated material transported to the ICP-MSvia a stream of argon carrier gas. The ICP-MS was operated in reaction gas mode to reduce the interferences on calcium and strontium isotopes. The optimised experimental parameters for LA ICP-MS analysis are recorded in Table 1. Cartilage samples (generally less than 10 × 10 mm) were imaged in approximately 20 h, whilst synovial fluid took approximately 2.5 h to image.
Agilent 7500ce ICP-MS | New wave UP213 laser ablation | ||
---|---|---|---|
Rf power | 1370 W | Wavelength | 213 nm |
Cooling gas flow rate | 15 l min−1 | Repetition frequency | 20 Hz |
Carrier gas flow rate | 1.25 l min−1 | Laser energy density | 0.2 J cm−2 (at 30%) |
Hydrogen flow rate | 2.5 ml min−1 | Cartilage samples | |
Sample depth | 5.0 mm | Spot size | 55 μm |
QP bias | −16 V | Scan rate | 30 μm s−1 |
OctP bias | −20 V | Synovial fluid | |
Scan mode | Peak hopping | Spot size | 100 μm |
Dwell time | 0.1 s, 0.2 s for P | Scan rate | 40 μm s−1 |
Synthetic synovial fluid | |||
Spot size | 55 μm | ||
Scan rate | 50 μm s−1 |
Imaging was performed using the software package ENVI v4.2 (Research Systems Inc., Boulder, CO, USA). The samples were systematically scanned across the entire area of the sample. For cartilage samples, the distance between adjacent scan lines was 29 μm, which ensured complete ablation of the tissue sample. For the synovial spots, the line scans were placed 50 μm apart. The data files generated by ChemStation (Agilent Technologies) were then collected and imported into ENVI. As no quantification data was collected, images only show relative signal intensities.
Synthetic synovial fluid was prepared from human serum (Sigma, Germany) by mixing equal volumes (30%) of laboratory grade (Ajax, Sydney) 10 mM calcium chloride, 50 mM tetrasodium pyrophosphate and 7 mM magnesium sulfate according to a method reported by Cheng et al. for crystal formation of CPPD in aqueous solutions.22,23Tris(hydroxymethyl)methylamine (Ajax, Sydney) was added to maintain the pH around physiological levels. The calcium, magnesium and pyrophosphate salts were chosen based on the assumption that their counter ions would not form precipitates in the synthetic synovial solution. The synthetic synovial fluid was then left in the oven at 37 °C for 2 weeks to encourage crystal formation. A disposable 0.22 μm syringe filter was used to remove any large particles from solution. Both filtered and non-filtered solutions were dropped (5 μl) onto a clean quartz slide and air dried before analysis by LA ICP-MS.
Initial experiments attempted to generate images with the reaction cell operating only as an ion guide. The calcium image signal was poor in quality due to a high background signal. Investigations into methods of background attenuation for calcium lead to experimentation with the octopole collision/reaction cell with hydrogen and helium, as well as utilising the cool plasma technique.24–27 Experiments were performed using an in-house prepared plastic standard that contained all the elements of interest.
Use of a cool plasma and helium collision gas decreased the background signal for calcium. However, sensitivity was significantly reduced for other elements. Hydrogen reaction gas significantly minimised the m/z 40 and m/z 44 background on calcium and interferences on strontium without significant loss of sensitivity. A hydrogen flow rate of 2.5 ml min−1 gave the best compromise of attenuation of background and minimal matrix interference whilst maintaining signal sensitivity.
Elemental distribution maps of calcium, phosphorus, magnesium and strontium in a cartilage sample are presented in Fig. 1. Element signals were normalised to 13C as an internal standard to compensate for changes in mass ablated or signal drift.28,29 Many of the tissue samples showed areas where it had folded or twisted over itself, i.e. the tissue was thicker in those areas. The use of 13C as an internal standard compensated for the difference in tissue thickness.
![]() | ||
Fig. 1 Elemental distribution maps of calcium (43Ca), phosphorus (31P), magnesium (24Mg), strontium (88Sr) and zinc (66Zn) in a knee cartilage section from a patient diagnosed with osteoarthritis. |
Fig. 1 shows corresponding regions of relatively high calcium and phosphorus intensities in the cartilage section taken from a patient with osteoarthritis. These common areas may be representative of calcium phosphate-based crystal deposits. A high frequency of CPPD crystals in articular tissues removed from OA hips and knees has been reported in the literature.30 Other elements including copper, iron, and selenium did not follow the same trend in distribution.
The areas are significantly larger than the average crystal size (up to 20 μm).3 This may be due to aggregation of the crystals within the cartilage matrix. An alternative explanation was that the relatively larger laser spot size used for imaging of the smaller crystals swamped the ICP-MS signal. That is, the crystal sizes were magnified by the use of a spot size larger than the average crystal size. This may be overcome by higher resolution imaging.
In addition, several regions of relatively high calcium and phosphate intensities correlated with relatively high magnesium and strontium intensities. Magnesium whitlockite, a calcium orthophosphate crystal in which calcium is partly substituted with magnesium, has been reported in both osteoarthritic and normal articular cartilage and recent data suggest it may play a pathological role in arthritis.31–33 Our results were consistent with the association of magnesium whitlockite and BCP crystals as areas of relatively high magnesium intensities consistently overlapped with areas of high calcium intensities.
Fig. 1 also shows an area of relatively high strontium intensity which corresponded to areas of high calcium, phosphorus and magnesium. To our knowledge the association of strontium with BCP crystals has not been reported before, however it has been detected in osteoarthritic synovial fluid.21 Possible interferences of 88Sr include double charged rare earth elements, calcium dimers and argides, and calcium phosphates.18,34 When tuning the ICP-MS on NIST 612, the formation of diatomic interferences was monitored and found to be minimal. The reaction cell was shown to further reduce this. The signal collected from m/z 88 was identified as strontium as the Sr88/Sr86 isotope ratio agreed within the accepted range of natural abundances.35 Therefore, the strontium distribution maps appear to be genuine.
An interesting trend in the distribution of zinc was also observed in some samples. Areas of high calcium, phosphorus and magnesium corresponded to regions of low zinc intensity.
This method also has the potential for ratio imaging. More objective and definitive detection of BCP crystals could be achieved by generating a 2D image of the Ca : P ratio, whereby areas with a ratio corresponding to that of BCP or CPPD crystals are selectively coloured against the remaining tissue. This may allow CPPD crystals to be distinguished from BCP crystals.
Synovial fluid was collected from patients suffering from OA and rheumatoid arthritis. A small volume (12 μl) was then dropped onto a quartz coverslip and air dried. The laser experimental parameters ensured that the centre of the drop was completely ablated. The thicker outer-rim was not always ablated completely. No sample preparation was necessary, an advantage against other methods such as XRF, which requires the crystals to be isolated first.3
BCP crystals are uniquely associated with OA and have been reported to occur in up to 60% of patient synovial samples.3,30CPPD crystals have been associated with 30% of OA synovial fluids.30 Images consistent with the presence of crystal deposits were observed for one of the OA samples and both of the RA samples. Though BCP crystals are uniquely associated with OA,3 the presence of CPPD crystals in RA joint fluids is not uncommon as these crystals are often found in many arthritides.36
Fig. 2 shows the distribution of several elements in a drop of synovial fluid from an osteoarthritic joint. Similar trends of common regions of relatively high calcium, phosphorus, magnesium and strontium were observed. An image of copper was included to illustrate the different distribution of elements not associated with calcium phosphate based crystals.
![]() | ||
Fig. 2 Elemental distribution maps of calcium (43Ca), phosphorus (31P), magnesium (24Mg), strontium (88Sr) and copper (63Cu) in a drop of osteoarthritic synovial fluid. |
The solution was filtered to remove the solid matter and drops of both filtered and non-filtered solution were left to air dry on a clean slide. As the fluid was not as viscous as synovial fluid, the spot was not as thick and a faster scan rate could be applied whilst still ensuring complete ablation.
Images of the elemental distribution of calcium, phosphorus, magnesium and strontium are presented in Fig. 3a. Again, results consistent with the presence of crystal deposits were observed in these images by common regions of relatively high intensities of crystal-associated elements.
![]() | ||
Fig. 3 (Upper) Elemental distribution maps of calcium (43Ca), phosphorus (31P), magnesium (24Mg) and strontium (88Sr) in a drop of synthetic synovial fluid. (Lower) Elemental distribution maps of calcium (43Ca), phosphorus (31P), magnesium (24Mg) and strontium (88Sr) in a drop of filtered synthetic synovial fluid. |
In distinct contrast, the images produced from the ablation of the filtered solution do not show the small common regions that are consistent with crystal deposits (Fig. 3b). These images instead show a more uniform distribution of the elements with increasing concentration radiating out from the centre of the spot, similar to that seen for non-crystal associated elements in the synovial fluid sample. Considering the common size of CPPD crystals, filtration at 0.22 μm should remove any crystals or solid matter that formed in the solution.
The presence of CPPD or BCP crystals in the cartilage or synovial fluid samples was not confirmed by other methods currently used for this purpose. However, this preliminary study gives a good indication of the potential of LA ICP-MS to detect BCP and CPPD crystals in both cartilage and synovial fluid samples. Additionally, the calcium-to-phosphorus ratio may be used to identify the crystals present. Further studies are currently being pursued to confirm crystal presence by SEM prior to laser ablation. This would validate the method in terms of false positives/negatives and sensitivity limits compared to other methods.
Additionally, quantification data may be obtained using standard serum solutions for spot calibration and matrix matched standardised tissue for cartilage samples.