Paul I.
Okagbare
a and
Steven A.
Soper
*abc
aDepartment of Chemistry, Louisiana State University, Baton Rouge, LA 70803, USA
bDepartment of Mechanical Engineering, Louisiana State University, Baton Rouge, LA 70803, USA
cCenter for BioModular Multi-Scale Systems, Louisiana State University, Baton Rouge, LA 70803, USA
First published on 24th November 2008
The design, performance and application of a novel optical system for high throughput single molecule detection (SMD) configured in a continuous flow format using microfluidics is reported. The system consisted of a microfabricated polymer-based multi-channel fluidic network situated within the optical path of a laser source (λex = 660 nm) with photon transduction accomplished using an electron-multiplying charge coupled device (EMCCD) operated in a frame transfer mode that allowed tracking single molecules as they passed through a large field-of-view (FoV) illumination zone. The microfluidic device consisted of 30 microchannels possessing dimensions of 30 µm (width) × 20 µm (depth) with a 25 µm pitch. Individual molecules were electrokinetically driven through the fluidic network and excited within the wide-field illumination area with the resulting fluorescence collected via an objective and imaged onto the EMCCD camera. The detection system demonstrated sufficient sensitivity to detect single DNA molecules labeled with a fluorescent tag (AlexaFluor 660) identified through their characteristic emission wavelength and the burst of photons produced during their transit through the excitation volume. In its present configuration and fluidic architecture, the sample processing throughput was ∼4.02 × 105 molecules s−1, but could be increased dramatically through the use of narrower channels and a smaller pitch. The system was further evaluated using a single molecule-based fluorescence quenching assay for measuring the population differences between duplexed and single-stranded DNA molecules as a function of temperature for determining the duplex melting temperature, Tm.
An approach toward increasing the data production rate for processing biochemical reactions is to increase the number of reactions run in parallel, which allows for analyzing a large number of targets simultaneously. This parallel processing is commonly carried out in titer plates, with increases in the well density used to increase data throughput. For example, titer plates can be configured to run 96–9600 reactions in parallel on a 3″ × 5″ plate.7,8 With increases in well density also comes reductions in well volume, placing severe demands on the readout hardware to provide parallel detection to further increase throughput while at the same time providing high detection sensitivity and favorable limits of detection in spite of the reduced well volume. Other applications that require the ability to transduce large numbers of targets quickly include capillary array electrophoresis,9 DNA microarrays,10 small-molecule arrays,11,12 and protein microarrays.13 In addition, micro-Total Analysis Systems (µ-TAS) have been employed as a tool for increasing throughput in many applications due to their exquisite characteristics conducive to high sampling rates brought about by reductions in analysis times and the parallelization of processing units.14–17
Another analytical technology, which facilitates the realization of high throughput biochemical data processing, is single molecule detection (SMD). SMD provides the ability to eliminate many sample processing steps associated with multi-step assays, significantly reducing assay time. This has recently been demonstrated by Wabuyele et al., where the overhead processing time associated with PCR amplification of target DNAs was eliminated by detecting single nucleotide polymorphisms (SNPs) in genomic DNA using single-pair FRET (spFRET).18 Using spFRET, the authors were able to detect SNPs in a processing time less than 5 min. Another advantage of SMD is the detailed information revealed when monitoring biochemical reactions. For example, it can yield insight into complex fluctuation phenomena and hidden heterogeneity that cannot be observed in ensemble techniques.19
However, the challenge associated with many SMD assays is their intrinsic low throughput, especially for flow-based SMD experiments where a laser focused to a diffraction-limited spot is used to interrogate material as it traverses through the excitation volume.20,21 To realize the full processing potential of SMD for high throughput applications, the readout hardware needs to be configured in a parallel format without sacrificing sensitivity.
The common approach for realizing high single molecule sensitivity is, as stated above, the utilization of ultra-small probe volumes, which reduces background interferences resulting from the sample matrix emanating from autofluorescence or scattering.22,23 A common method for achieving ultra-small probe volumes is the use of confocal optics,24–27 in which the laser beam is focused to a diffraction-limited area producing sampling volumes in the sub-picoliter regime. However, these confocal sampling volumes are usually designed with single element detectors and, thus, generate small sample processing throughput rates.
This limitation can be addressed by operating the readout system in a multi-channel configuration, such that single molecules are transported through a series of fluidic processors that are interrogated using wide-field illumination with the resulting fluorescence imaged onto an array detector for parallel readout. An array detector that can be employed for fluorescence imaging appropriate for single molecule high throughput processing is the charge coupled device (CCD).28 Advancements in CCD technology has made available ultra-sensitive cameras appropriate for realizing detection of single fluorescence molecules in flow-based measurements where the signal is transitory in nature.29–32 For example, Emory and Soper33 recently described a CCD-based single molecule analysis system in which a flowing sample composed of double-stranded DNA molecules labeled with an intercalating dye (TOTO-3) was imaged onto a 1340 × 100 pixel CCD camera operated in a time-delayed integration (TDI) mode.33 The TDI mode was used to collect photo-generated signal from single multiply-labeled DNA molecules into a potential well of the CCD array as the molecules traversed the field-of-view (FoV), thereby enhancing the signal-to-noise ratio (SNR) and at the same time increasing the duty cycle of the detection system. Van Orden et al. also demonstrated the ability to image single DNA molecules labeled with an intercalating dye using a CCD operated in a snap-shot mode.34
While the aforementioned works are excellent examples of increasing the sample throughput for flow-based SMD experiments, the limitations are that the single molecules were double-stranded DNAs loaded with multiple fluorophores. In the case of λ-DNA (∼48500 bp), each DNA molecule had a load of approximately 9700 dye molecules. Therefore, high throughput flow-based SMD systems have yet to demonstrate the ability to detect single fluorophore molecules.
Another viable operational modality for a CCD camera directed toward SMD is the frame transfer mode (FTM) when the CCD contains an electron-multiplying stage (EMCCD), which allows for high speed imaging.35,36 For example, Li and Yeung utilized a frame transfer EMCCD to monitor dynamic conformational changes of flavin adenine dinucleotides (FADs) in different media on a cover slip in which the molecules were immobile. Their results demonstrated the ability to monitor the real-time dynamics of the photoproduct formation between FAD molecules at the single molecule level.37
The availability of FTM EMCCDs has created the capability to study biochemical reactions at the single molecule level utilizing flow-based measurements in which the single molecules are dynamically transported through the illumination zone. These EMCCDs are equipped with a full frame storage region and a charge multiplication register that generates secondary electrons via an impact-ionization process. The net gain on these cameras can be greater than 1000-fold. Although, the resulting gain has a complex relationship with the applied voltage, it can be estimated using the expression;
G = (1 + g)N | (1) |
St = S × QE × G | (2) |
In this paper, we combined the advantages of microfluidics with a wide-field fluorescence imaging system and present, for the first time, the ability to detect single chromophore molecules in a multi-channel flow-through configuration and monitor biochemical reactions at the single molecule/fluorophore level. The system reported herein provides high throughput single molecule processing capabilities using an ultra-sensitive epi-fluorescence imaging system consisting of an FTM EMCCD. The system possessed a large FoV (200 µm, diameter, generated by a 40×/0.75 microscope objective) capable of imaging a series of parallel fluidic channels simultaneously with an optical irradiance that was near saturation as opposed to our previous report using a TDI mode of operation for the CCD.33 As an example of the utility of this system to monitor biomolecular reactions at the single molecule level at relatively high processing rates, a melt analysis of double-stranded DNA (dsDNA) using a resonance energy transfer reporting system was used. The dsDNA contained a construct labeled on the 5′-end of the sense strand with a dye and a quencher on the 3′-end of the anti-sense strand. The number of photon burst events arising from melted duplexes at different temperatures could be used to determine the melt profile of this double-stranded complex.
![]() | ||
Fig. 1 (a) Optical set-up of the imaging system using epi-illumination and employing a large FoV. The beam from the diode laser (λ = 660 nm) was isolated using a laser line filter (F1) and then shaped with a plano-convex lens (L1), which was used to focus the laser beam behind the input aperture of the microscope objective (OBJ). Following beam shaping, it was directed by a dichroic filter (DF) into the OBJ with the collimated laser beam impinging upon a multi-channel microchip (MC). The fluorescence signal generated from the chip was collected by this same objective, passed through the DF and spectrally selected using a long pass filter (F2) and an interference band pass filter (F3). A mirror (M) was used to steer the fluorescence signal onto a CCD after passing it through a lens (L2), which focused the radiation onto the photoactive area of the CCD. The total magnification of the system was 40×. (b) An optical micrograph showing a section of the multi-channel microfluidic chip; the fluidic network consisted of microchannels with dimensions of 30 µm wide × 20 µm deep and a pitch of 25 µm. All channels had a common sample input reservoir. The driving electric field was applied at this reservoir and another on the opposite end of the fluidic network (not shown) to drive the sample electrokinetically through the imaging area. (c) Diagram showing the operation of the CCD in a frame transfer mode with image accumulation occurring during single molecule travel within the FoV of the microscope. The images of single molecules produced streaks on the CCD due to the molecular transit time being greater than the CCD exposure time and multiple frames summed to produce the final image. |
The EMCCD camera (Cascade 1K, Photometrics, Tuscon, AZ, USA) was a frame transfer EMCCD equipped with signal enhancement via on-chip gain and possessed 8 µm × 8 µm pixels configured in a 1004 × 1002 front illumination format. The camera had a 10 MHz digital converter and a 16-bit digitization range. The frame transfer capability of the EMCCD allowed near 100% duty cycle when single molecules were tracked along their flow path within the illumination field. The CCD camera was thermoelectrically cooled to −30 °C to minimize dark noise and operated at the optimal gain to maximize the SNR in the measurement.
Stock solutions of λ-DNA molecules were labeled with an intercalating dye (Syto-63) and diluted in 1× TTE buffer to 10 pM. This concentration was selected to keep the single molecule occupancy probability low to reduce the double occupancy probability. With the system possessing a probe volume of 0.8 fL (CCD pixel size = 8 µm × 8 µm), a sample concentration of 100 pM produced a single molecule occupancy probability of 0.01 (probability of double occupancy = 0.0001).
The single fluorophore detection capability of the system was evaluated using DNA molecules that were end-labeled with a single AlexaFluor 660 dye molecule. In this case, only a single image frame was used for the presented data due to the short bleaching lifetimes associated with the single chromophore assemblies.
The 40× objective generated a FoV of 200 µm, which allowed imaging five microchannels with the current fluidic architecture. Changing the imaging objective to a lower magnification would allow for the processing of more fluidic channels (see Table 1). For example, the use of a 5× objective would provide the ability to image 30 channels with the channel architecture used herein due to its larger FoV. However, the cost for this higher throughput is reductions in the collection efficiency of the resulting fluorescence emission, which could potentially reduce single molecule sensitivity. Clearly, reductions in the channel width and pitch could significantly improve the number of channels we could image. For example, using a microfluidic chip with a 1 µm wide channel and a pitch of 1 µm, which can be fabricated using optical lithography, would permit the ability to image 100 such channels for a FoV of 200 µm.
Total magnification (magnification/numerical aperture) | Effective field-of-view (FoV) | Number of channels within the FoV (30 µm × 20 µm; 25 µm pitch) | Single molecule SNR for λ-DNA stained with Syto-63 | Theoretical single molecule throughput |
---|---|---|---|---|
40×/0.75 | 200 µm | 5 | 810 | 6.7 × 104 |
10×/0.5 | 800 µm | 17 | 510 | 2.28 × 105 |
5×/0.25 | 1.6 mm | 30 | 314 | 4.02 × 105 |
![]() | ||
Fig. 2 (a) Plot of the signal-to-noise ratio (SNR) at different field strengths used to transport single molecules through the irradiation zone of the imaging system. The SNR values were obtained from an average of 10 different measurements. (b) Fluorescence image from the CCD for a single molecule tracked along several pixels of the CCD, which was accumulated from two image frames that were acquired using 100 ms exposure time per frame with the fluorescence collected using a 40× objective (NA = 0.75). |
To demonstrate the ability to track single dsDNA molecules traveling electrokinetically through the irradiation zone, experiments were next carried out using single and multiple microchannels. Fig. 2b shows a fluorescence image of a single DNA molecule tracked along the FoV of the imaging system by accumulating two image frames with a 100 ms exposure time per frame. From the length of the streak (∼60 µm) and the image accumulation time (2 frames at 100 ms per frame), we estimated the linear velocity to be 0.03 cm/s, very close to that determined above at a field strength of 300 V/cm.
To evaluate the sample throughput (ST) capabilities of the system, a 100 pM solution of λ-DNA was electrokinetically driven through the system at a field strength of 300 V/cm and the fluorescence image was collected using the 10× objective (NA = 0.5); the FoV of the system was 800 µm, which allowed imaging 17 channels. With the system operated in a frame transfer mode, the SMD sensitivity was improved by using 3× binning (3 × 3) with the multiplication gain of the EMCCD set at 3700. Fig. 3a shows one image frame from a series of images acquired for single λ-DNA molecules stained with Syto-63 traveling through the illumination zone. Each fluorescent streak represents an individual DNA molecule migrating through the chip due to the low probability of double occupancy per imaging pixel at this DNA concentration. The intensity distribution of this image was constructed by transforming the image in Fig. 3a to a 3-D image (see Fig. 3b).
![]() | ||
Fig. 3 (a) Image of single λ-DNA molecules migrating through a series of microchannels (17 channels are shown). The image was acquired using a 10× (NA = 0.5) objective at an exposure time of 100 ms with the CCD multiplication gain set at 3700 (controller gain set at 1). The sample solution (100 pM) was electrokinetically pumped through the fluidic channels with a field strength of 300 V/cm (linear velocity = 0.037 cm/s). Each fluorescent spot represents a single λ-DNA molecule. (b) 3-D image showing the intensity distribution of λ-DNA molecules shown in (a). (c) Single λ-DNA molecules migrating through series of 25 microchannels. Image was acquired using a 5× objective (NA = 0.25). In all cases, the λ-DNA was stained with Syto-63 at a dye to base pair ratio of 5 : 1. |
To further demonstrate the ability to increase the throughput of the multi-channel fluidic device, a 5× objective (NA = 0.25) was incorporated into the imaging system and the same sample concentration of λ-DNA as that use above (100 pM) was analyzed. Fig. 3c presents a single image frame showing photon bursts from single λ-DNA molecules in 25 different microchannels. This objective generated a FoV of 1.6 mm, which allowed imaging 30 channels using our present fluidic architecture, but only 25 are shown in Fig. 3c.
For flow-based single molecule detection experiments in which the single molecule events are dynamically transported through a fixed interrogation (sampling) zone, the sample throughput (ST) can be calculated from:33
ST = SE × DE × DR × DC | (3) |
DR = Fv × Cb | (4) |
As noted above, scaling down the microchannel width and pitch can be used to further increase the ST of the present system. For example, reducing the channel width to 3 µm with a pitch of 3 µm and using the 10× objective to produce a reasonable collection efficiency (NA = 0.5) will increase the number of channels that can be imaged to 133 with a corresponding ST of 1.78 × 106. A further reduction in channel width to 1 µm with an inter-channel spacing of ∼400 nm (diffraction-limited resolution is ∼λ/2) will allow the system to image 571 microchannels using the 10× objective increasing ST to 7.65 × 106 molecules/s. While this represents a significant increase in ST over our previous work,33 it also indicates exceptional increase in ST over previous research reports.34,39–46 For example, Van Orden et al. reported an ST of 2000 DNA fragments s−1 using a CCD operated in a snap-shot mode,34 while Ma et al. reported an ST of ∼8500 molecules s−1 from their single molecule spectroscopy studies.43 Recently, Chansin et al. speculated a throughput of ∼1500 molecules s−1 using their synthetic nanopores to measure DNA translocation events.46
![]() | ||
Fig. 4 (a) Detection of individual dye molecules (AlexaFluor 660) end-labeled to a 60 base ssDNA distributed in different microfluidic channels (five channels are shown) for a sample concentration of 750 pM. The dye/ssDNA conjugate was pumped through the fluidic channels using a field strength of 80 V/cm (linear velocity = 0.01 cm/s). To further visualize the fluorescent photon bursts in this image, a sub-frame (25 × 42 pixels) was expanded and quantitative information extracted by applying a threshold condition to this sub-frame (see text for details). (b) Image acquired when the fluidic channels were filled with buffer only, and the processed sub-frame with application of the threshold condition as described in the text. |
From careful inspection of the sub-frame image shown in Fig. 4a, we were able to count 64 photon bursts generated from single ssDNA-AlexaFluor 660 conjugates. From the size of the image taken (5 µm × 8.4 µm × 1.4 µm, Pv) and the concentration of the conjugates used herein (750 pM, Cb), we could calculate the number of expected events (Nev) from:
Nev = CbPv | (5) |
Different concentrations of ssDNA-AlexaFluor 660 conjugates were next evaluated using our multi-channel single molecule detector system and processed with the same threshold value as noted above. An average of eight images per dye concentration was plotted with the calibration plot constructed from the number of photon burst events versus the conjugate concentration. This analysis was performed on a sub-frame image with a size of 25 × 42 CCD pixels. The number of spots (single molecule events) observed was linear with the sample concentration yielding a correlation factor of 0.98 (data not shown).
![]() | ||
Fig. 5 (a) Fluorescence single molecule quenching assay for determining duplex Tm using a 15-base pair oligonucleotide labeled with AlexaFluor 660 on its 5′-end and its complement labeled with the Black Hole Quencher-3 (BHQ-3) on its 3′-end. (b) Fluorescence single molecule image of the 15-bp oligonucleotide when the sample was heated to 67 °C. Fluorescence from single DNA/AlexaFluor 660 molecules were visible as shown in this image due to duplex melting, producing ssDNAs. Also shown is the sub-frame of this image (37 × 67 pixels) that was processed as described in Fig. 4. (c) Plot of number of single molecule events counted over a fixed interval of time at different sample temperatures (■, concentration = 750 pM). Also shown is a plot of the normalized fluorescence intensity of the duplexed DNA at different temperatures using a bulk measurement (dsDNA concentration = 40 nM, ●; and 1 nM, ▲) measured using a conventional fluorometer (λex = 675 nm). |
For these experiments, the duplexed samples (750 pM) were prepared in Tris/HCl buffer containing 20 mM Tris-HCl, 10 mM MgCl2 and 10 mM KCl at pH 7.5. The microfluidic device was equipped with a strip heater, which was placed 200 µm upstream from the imaging area. The duplexes were electrokinetically driven through the fluidic device at the desired temperature using an 80 V/cm field strength. When operated at room temperature, there was no fluorescence signal registered on the detector. The absence of fluorescence photon bursts in the EMCCD image clearly indicated that the reporter dye was sufficiently quenched by BHQ-3 and no single-stranded constructs could be seen (data not shown). This is in agreement with previous studies using this fluorophore–quencher pair with a reported 96% quenching efficiency.47
To generate ssDNA in the flowing sample, the strip heater was then turned on with the temperature set to 67 °C. Fig. 5b shows a single molecule fluorescence image of photon bursts generated from the ssDNA constructs when imaged at this temperature. To clearly visualize these spots and extract quantitative information by counting the single molecule events, the threshold condition was applied as discussed earlier. A sub-frame (37 × 67 pixels) was selected and processed (see Fig. 5b). The presence of fluorescent photon bursts in the 67 °C image versus that taken at room temperature is expected; the thermal energy introduced into the system produced a sufficient population of ssDNAs due to thermal denaturation of the duplexed DNA and thus, the energy transfer interaction between the two labels is eliminated because of proximity considerations.
A melt analysis was then conducted on the oligonucleotide duplex by taking images of single molecules traversing through the interrogation zone at different temperatures and counting the number of events (i.e., ssDNAs) at each temperature investigated. This analysis generated a sigmoidally-shaped plot (see Fig. 5c) with a sharp rise in the number of single molecule events between 30 and 50 °C with a midpoint at ∼40 °C. This midpoint represents the melting temperature (Tm) of the duplex, which is defined as the temperature in which 50% of the DNA is in its double-stranded form and 50% in its single-stranded form. This value agrees favorably with that obtained from a theoretical prediction (Tm = 41.7 °C) using oligonucleotide analysis software.48,49
The melting temperature of the duplex was also estimated by measuring fluorescence from a bulk sample of the duplex (1 nM) as a function of temperature using a standard fluorometer (λex = 675 nm). Fig. 5c shows the resulting plot of fluorescence intensity obtained at various temperatures, which generated a Tm value of 43 °C. The slight deviation of this value (43 °C) compared to the single molecule result could be attributed to the slightly higher concentration of the duplex used in the bulk solution since the inter-strand duplex Tm depends on its concentration.48 This supposition was further supported by the experimental Tm (50 °C) obtained for a 40 nM solution of this same duplex (see Fig. 5c).
As an example of the necessity for high throughput processing required to perform this measurement, we were able to reduce the processing time by a factor directly related to the number of channels imaged during the single molecule measurement, which in this case yielded a 5-fold reduction in processing time compared to a single point SMD flow-based scenario. Implementation of a smaller microchannel width and pitch producing a higher packing density of channels would significantly increase the data processing rate. For 1 µm wide channels and a pitch of 400 nm, the number of channels that could be effectively imaged using a 200 µm FoV would be 143 channels, reducing the processing time by 143-fold.
The results obtained from a fluorescence quenching assay associated with DNA duplex melting demonstrated the ability of our system to generate high throughput data processing appropriate for evaluating the fate of a biochemical reaction by monitoring population differences between reactants and/or products in a single molecule approach. The appealing aspect of the single molecule approach is the ability to use low concentrations of precious biological reactants and enzymes, the exquisite sensitivity associated with single molecule detection to record subtle changes in the reaction profile and the ability to monitor reactions efficiently in ultra-small volumes. This will be particularly attractive for high throughput screening applications, where the effects of elements from a large combinatorial library must be screened against a therapeutic target (i.e., bioenzyme). Owing to the large size of most combinatorial libraries (>106 elements), high throughput processing is critical.
This journal is © The Royal Society of Chemistry 2009 |