Stephan T.
Koev
ab,
Michael A.
Powers
ab,
Hyunmin
Yi
c,
Li-Qun
Wu
d,
William E.
Bentley
def,
Gary W.
Rubloff
bfg,
Gregory F.
Payne
df and
Reza
Ghodssi
*abf
aDepartment of Electrical and Computer Engineering, University of Maryland, College Park, MD 20742, USA. E-mail: ghodssi@eng.umd.edu; Fax: 1-301-314-9218; Tel: 1-301-405-8158
bInstitute for Systems Research, College Park, MD 20742, USA
cDepartment of Chemical and Biological Engineering, Tufts University, Medford, MA 02155, USA
dUniversity of Maryland Biotechnology Institute, College Park, MD 20742, USA
eDepartment of Chemical Engineering, College Park, MD 20742, USA
fBioengineering Graduate Program, University of Maryland, College Park, MD 20742, USA
gDepartment of Materials Science and Engineering, College Park, MD 20742, USA
First published on 17th October 2006
While microcantilevers offer exciting opportunities for mechano-detection, they often suffer from limitations in either sensitivity or selectivity. To address these limitations, we electrodeposited a chitosan film onto a cantilever surface and mechano-transduced detection events through the chitosan network. Our first demonstration was the detection of nucleic acid hybridization. In this instance, we electrodeposited the chitosan film onto the cantilever, biofunctionalized the film with oligonucleotide probe, and detected target DNA hybridization by cantilever bending in solution (static mode) or resonant frequency shifts in air (dynamic mode). In both detection modes, we observed a two-order of magnitude increase in sensitivity compared to values reported in literature for DNA immobilized on self-assembled monolayers. In our second demonstration, we coupled electrochemical and mechanical modes to selectively detect the neurotransmitter dopamine. A chitosan-coated cantilever was biased to electrochemically oxidize dopamine solution. Dopamine's oxidation products react with the chitosan film and create a tensile stress of approximately 1.7 MPa, causing substantial cantilever bending. A control experiment was performed with ascorbic acid solution. It was shown that the electrochemical oxidation of ascorbic acid does not lead to reactions with chitosan and does not change cantilever bending. These results suggest that chitosan can confer increased sensitivity and selectivity to microcantilever sensors.
Chitosan has three important properties that make it uniquely suited as a bio-device interface material.17,18 Chitosan is stimuli-responsive and undergoes a soluble-to-insoluble transition in response to a pH switch from 6 to 6.5. Chitosan is also film-forming such that this pH switch can generate a stable hydrogel network. These two properties—pH-responsive and film-forming—allow stable chitosan films to be electrodeposited at cathode surfaces.19,20 The pH-switching at the cathode surface is easily controlled, and chitosan can be electrodeposited with high spatial21 and temporal22 control. Interestingly, due to the stability of the chitosan film, various components can be co-deposited with chitosan and entrapped within its hydrogel network.23–25 Chitosan's third unique property is that it is nucleophilic. Chemically, chitosan has primary amines at nearly every repeating unit of its polymeric structure. These amines are nucleophilic and readily react with electrophiles. For instance, we have used standard glutaraldehyde-based coupling chemistries to bio-functionalize electrodeposited films with biological components (e.g., proteins,26 nucleic acids,27 and virus particles28). In sum, previous results have shown that chitosan can perform two functions at the interface—it can be directed to assemble (i.e., electrodeposit) in response to localized signals, and it can be functionalized through standard chemistries. Here, we report a third function: chitosan networks can transmit mechanical information of chemical and biological events to the cantilever surface.
We examined two model experimental systems to demonstrate chitosan's mechano-transduction function. Our first demonstration is the detection of DNA hybridization, which is a “conventional” application for microcantilever mechano-sensors.5,12,14,15,29,30 For this demonstration, we employ chitosan's stimuli-responsive and film-forming properties for electrodeposition on the cantilever, after which we employ chitosan's nucleophilic properties to functionalize the film with an oligonucleotide-probe. Hybridization of this probe with a complementary target DNA is detected as a differential deflection when measured in solution (static mode), or a resonant frequency shift when measured in air (dynamic mode).
Our second demonstration of chitosan's mechano-transduction function is a novel one, and it couples electrochemical and mechanical modes to selectively detect the neurotransmitter dopamine. Dopamine and related catechols are readily oxidized; interestingly, oxidized catechols are electrophilic and react with chitosan.31 In some cases, oxidized catechols can covalently crosslink chitosan and substantially alter its mechanical properties.32 For our second demonstration, this change in mechanical properties of chitosan films due to dopamine oxidation is detected by the differential deflection of microcantilevers (static mode). Biological samples often contain components such as ascorbic acid that are also easily oxidized and interfere with the dopamine detection.33,34 We show that the mechano-detection method can successfully discriminate between dopamine and ascorbic acid solutions.
The microcantilever sensor consists of layers of chitosan (varying thickness), Au (80 nm), Cr (20 nm), and Si3N4 (500 nm), on a Si substrate with a SiO2 layer (500 nm). The length of the released beam is 100 µm, and the width is 40 µm. The air gap (distance between cantilever bottom and substrate) is 20 µm. A cross sectional schematic of the device is shown in Fig. 1A. In both static and dynamic modes, the cantilever displacement is measured using a Veeco NT1100 optical interferometer. In the dynamic mode, the resonant frequency of the cantilever is measured before and after the biomolecular detection event, and the presence of the analyte is inferred from the resonant frequency shift. In this mode, measurements are performed in air after drying the cantilever chip to maximize the resonance quality factor. In the static mode, the bending of the cantilever is measured before and after the detection event, and the presence of the analyte is inferred from the change in bending. In this mode, measurements are taken with the chip immersed in solution.
Fig. 1 (A) Cross sectional schematic of microcantilever with chitosan used for detection of DNA hybridization and for detection of dopamine electrochemical oxidation. (B) SEM of fabricated cantilever. The (111) silicon crystallographic planes are sloped and visible near the cantilever base. (C) Optical micrograph of cantilever after chitosan electrodeposition. The chitosan is deposited everywhere except at the electrically isolated tip. (D) Contact profiler scan of chitosan film along dashed line in C. Higher electric field near the edges increases the deposition rate and the resulting chitosan thickness there. |
(1) |
(2) |
(3) |
Assuming a small resonant frequency shift Δfres, the sensitivity of the cantilevers to uniformly distributed target loading with surface density σt can be shown to be:
(4) |
Here, the factor C accounts for incomplete coverage of the cantilever with chitosan and target biomolecules, and it becomes 1 if the whole cantilever is covered. Note from Fig. 1A that the tip of the cantilever has an electrically isolated metal rectangle with an area approximately 20% of the cantilever surface. This rectangle does not experience chitosan deposition and is used for displacement measurements in dynamic mode. Chitosan has a rough surface and typically increases optical measurement error. The factor C in eqn (4) due to the chitosan-free rectangle in this design is 0.31. To maximize sensitivity, the rectangle can be reduced in size or moved to a different location on the cantilever.
Note from eqn (4) that the sensitivity increases with increasing resonant frequency and with decreasing cantilever thickness. However, decreasing the thickness generally leads to a reduction of the quality factor because the ratio of elastic and inertial forces to air damping forces is reduced. This in turn hurts the accuracy of resonant frequency measurements. An optimal design for a cantilever would require empirical knowledge of the Q factor as a function of resonant frequency and thickness. The present design is not optimized for sensitivity but rather for ease of measurement and experimentation. Based on the material properties commonly reported in literature and tabulated in the electronic supplementary information (ESI)† entry for this paper, the calculated resonant frequency is 60 kHz and the sensitivity is 63 Hz cm2 µg−1. The contribution of chitosan is ignored in this first-order analysis since the mechanical properties of chitosan and its surface topography vary considerably with deposition conditions.
(5) |
Here, σs is the surface stress, ν is the Poisson ratio, L is the beam length, and H is the thickness, and E is the effective Young's modulus as defined previously. The Stoney equation is inaccurate for short cantilevers, and several modified equations have been derived.37,38 Sader38 showed that the error in cantilever tip displacement given by the Stoney equation is approximately 10% for cantilevers with a length to width ratio of 2.5 (the cantilevers in our work are 100 µm long and 40 µm wide). This error is considered acceptable here and the original Stoney equation is therefore used.
The amount of bending for a given stress increases with length of the cantilever and with reducing thickness. However, as in the case of the dynamic mode operation, other factors such as stiction, cantilever fragility, and ease of measurement were taken into consideration when determining device dimensions. As a result, the cantilevers used for static operation were designed with the same dimensions as those for dynamic operation. They do not, however, have the isolated metal rectangle at the tip and are entirely covered by chitosan. In static mode, a narrowband illumination source is used for the interferometer, enabling accurate measurements to be taken even with chitosan on the surface.
After KOH etching, the released cantilevers consist of layers of SiO2, Si3N4, and Cr/Au. The SiO2 is removed in concentrated HF because its residual stress causes the cantilever to bend out of plane considerably. This bending is undesirable, and it interferes with optical displacement measurements. The residual stress of the metal layer also causes some cantilever bending. However, that results in only 1 µm upward displacement of the tip and does not impact the optical measurements.
Finally, the wafer is diced in 3 mm by 25 mm chips to facilitate handling during the multiple biochemical reaction steps. Prior to dicing, the wafer is covered in photoresist (without spinning) to protect the released cantilevers. After dicing, the protective photoresist is removed in acetone, and the majority of cantilevers are observed to be intact.
Medium molecular weight chitosan flakes from Sigma Aldrich were used to make an aqueous solution with 0.5% w/v chitosan concentration and pH = 5. The deposition conditions were chosen experimentally to obtain the desired chitosan thickness. In a typical experiment, a voltage of −0.9 V applied for 30 s results in an average film thickness of 100 nm as measured by contact profilometry. The chitosan coverage on the cantilever is not uniform because the electric field and the deposition rate near the electrode edges are higher than in the middle of the electrode. Fig. 1C shows a cantilever after deposition. A representative profile of the chitosan film is given in Fig. 1D. Better uniformity of the film could be obtained by slowing down the deposition but is not necessary for this application. Here, the only role of the chitosan is as an interface for attaching the biomolecules, and its uniformity is not a major concern. The deposited chitosan film has tensile residual stress that causes out-of-plane bending of the cantilever when the sample is dried. In practice, the amount of residual stress depends on the exact deposition conditions and varies between experiments. Based on an average measured displacement of the cantilever tip of 2 µm, the calculated chitosan residual stress is approximately 60 MPa tensile. This out-of-plane bending limits the maximum chitosan thickness that can be deposited on the cantilevers. For bending of more than about 10 µm it becomes impossible to measure the cantilever position with the optical interferometer because of depth-of-focus limitations.
After the deposition, the sample is rinsed and immersed in 1 M NaOH solution for 5 min to neutralize the chitosan film. Finally, the sample is equilibrated in SSC buffer (sodium saline citrate) to bring the pH to 7.1.
The resonant frequency of bare cantilevers without chitosan was measured to be approximately 58 kHz. The calculated resonant frequency for such cantilevers using eqn (1) is approximately 60 kHz. This close agreement suggests that the material properties, cantilever dimensions and assumptions used in the calculation are approximately valid and allows us to use the sensitivity expression eqn (4). As discussed previously, the mechanical properties of the chitosan film vary considerably with the deposition conditions and a theoretical prediction of the resonant frequency of a cantilever with chitosan was not attempted. However, the measured resonant frequency of the chitosan-coated cantilever in practice is only a few kHz above the resonant frequency of the uncoated cantilever. Therefore, eqn (1) and (4) can be used as approximations for cantilevers with chitosan as well. The quality factor of the cantilever's resonance was measured by dividing the displacement at the resonant frequency by the displacement at DC using the same voltage. The average measured Q factor is 20.
Multiple measurements of a single sample are taken to check for repeatability and determine the error bars on the resonant frequency. In the absence of chitosan, the standard deviation of resonant frequency measurements is 5 Hz. This error is caused mainly by ambient vibrations of the interferometer's stage, which corrupt the frequency response peak. Once chitosan is added to the cantilever, the resonant frequency uncertainty increases considerably because the film absorbs humidity. The standard deviation of resonant frequency measurements taken on 5 different days with chitosan was approximately 100 Hz. Since the shifts due to biological binding events in our experiments are on the order of kHz, this is an acceptable error. However, to improve the detection limit it would be necessary to reduce the humidity-induced variation of the resonant frequency. For example, a reference cantilever can be used to track the instantaneous humidity, and all resonant frequency measurements can be subtracted from that of the reference cantilever.39 Alternatively, the sensor can be isolated from the environment during measurement to provide more repeatable conditions.
Oligonucleotide | Sequence and end modifications | Total bases | Concentration in solution |
---|---|---|---|
Surface prode (dnaK) | 20 | 20 µg mL−1 | |
Surface probe (6xHis) | 18 | 20 µg mL−1 | |
Target (dnaK) | 70 | 1.5 µM | |
Sandwich probe | 20 | 6 µM |
Fig. 2 (A) Vertical profile of cantilever with chitosan and complementary probe DNA in response to hybridization and denaturation (measured in solution). The hybridization causes upward differential bending of ∼500 nm at the tip. The bending is reversed by denaturation. (B) Response of cantilever with chitosan and noncomplementary probe DNA to hybridization and denaturation (measured in solution). The differential bending is <30 nm. |
A “control” cantilever was also tested in which the electrodeposited chitosan was functionalized with a DNA probe that is not complementary to the target. Specifically, we functionalized the film with a DNA probe to the common sequence for the hexa-histidine fusion tag (6xHis) and then tested this functionalized cantilever against the non-complementary dnaK target. No hybridization is expected for this mis-matched probe-target pair. Consistent with this expectation, Fig. 2B shows little or no change in cantilever bending upon “hybridization” and “denaturation”. The small variations in this case are caused by measurement error and possibly to nonspecific DNA binding. These observations further support the conclusion that the 500 nm differential bending observed in Fig. 2A is the result of specific bio-molecular interactions.
The mechanism of surface stress generation by DNA hybridization is generally not well understood. It has been explained by electrostatic and steric interactions between the DNA molecules as well as maximization of their configurational entropy,5,12,29,30 but comprehensive models are lacking. The generated surface stress in this study significantly exceeds values reported in literature for DNA with similar concentration immobilized by self- assembled monolayer (SAM) techniques instead of on chitosan. We measured maximum stresses of approximately 1.5 N m−1, while others report 0.02 N m−1 for SAMs immobilized DNA.14 The reason for this significant increase is presumably the large effective surface area of chitosan due to its 3D hydrogel structure and its high density of amine groups. We believe that micromechanical sensors can significantly benefit in sensitivity by the use of chitosan to immobilize the probe molecules.
Fig. 3 (A) Frequency response of cantilever with chitosan and complementary probe DNA after hybridization and denaturation (measured in air). Points are raw data and curves are fitted Lorentzian functions. Hybridization reduces the resonant frequency by 2.5 kHz, and denaturation reverses the shift. (B) Frequency response of cantilever with chitosan and noncomplementary probe DNA after hybridization and denaturation (measured in air). Frequency shifts are <500 Hz and are caused by humidity variation between measurements. |
A control cantilever is functionalized with the noncomplementary probe (6xHis) and subjected to the same steps described above. Since the target and probe are mismatched in this case, minimal hybridization is expected. Fig. 3B shows the measured frequency response of the control cantilever at each step. The initial resonant frequency is 61.6 kHz. Upon hybridization, the resonant frequency becomes 61.0 kHz and upon denaturation 60.7 kHz. These differential changes in resonant frequency are small compared to the matching DNA case and are caused mainly by air humidity variations. This explanation is consistent with Fig. 3A, in which the difference between pre-hybridization and post-denaturation measurements is 0.7 kHz (ideally, it should be 0). In Fig. 3B, the corresponding difference is 0.9 kHz. Note that the complementary and noncomplementary measurements are taken within a 10 min period of each other are subject to similar environmental humidity. These results suggest that using a reference cantilever in parallel with the measurements to track instantaneous humidity variations could considerably improve the detection limit.39
The DNA hybridization can affect the cantilever resonant frequency by three different mechanisms: change in the spring constant, increase in the mass, or change in the damping. Additional characterization is needed to determine the contributions of each effect to the observed resonant frequency shifts. In studies with SAMs immobilized DNA on resonators, it is typically assumed that the frequency shift is caused by mass changes.4 If we assume that the mass increase effect dominates and the DNA is distributed uniformly on the cantilever, the calculated target DNA mass is approximately 16 µg cm−2 based on a frequency shift of 1 kHz. This is equivalent to 2.8 × 1014 target molecules cm−2, two orders of magnitude over what is reported for studies using self assembled monolayers.14,40 This estimate is not rigorous because it has not been verified that the resonant frequency shift is caused by mass change alone and that the target DNA is uniformly distributed on the cantilever. The important observation, however, is that the frequency shift caused by hybridization of chitosan-bound DNA far exceeds that caused by self-assembled DNA. The nature of the mechanism that causes the shift is not significant for this application.
Fig. 4 (A) Response of cantilever with chitosan to ascorbic acid electrochemical oxidation (measured in solution). The chitosan does not react chemically. The small bending at the tip is due to measurement error and nonspecific interactions. (B) Response of cantilever with chitosan to dopamine electrochemical oxidation (measured in solution). The chitosan film is crosslinked and generates a tensile stress. The cantilever bends up considerably (∼800 nm at the tip). |
The oxidation procedure is repeated with the same chitosan-coated cantilever in a dopamine solution. Fig. 4B shows the vertical profile of the cantilever before and after the dopamine oxidation. The cantilever bends up by approximately 800 nm due to the crosslinking of chitosan by the dopamine oxidation products. The estimated stress in the chitosan film generated by the crosslinking is 1.7 MPa. These results show that the device successfully discriminates between ascorbic acid and dopamine oxidation.
It should be noted that real biological samples often contain both dopamine and ascorbic acid simultaneously, with the concentration of ascorbic acid being higher.33,34 Currently, our method is not capable of detecting dopamine mixed with appreciable amounts of ascorbic acid because the dopamine oxidation products are reduced by the ascorbic acid and are not allowed to react with the chitosan. In experiments where the concentration of ascorbic acid was 10 times lower than that of dopamine, the chitosan crosslinking still occurred. However, when the two concentrations were comparable, no chitosan crosslinking was observed. Well-established electrochemical methods for dopamine detection such as fast-scan cyclic voltammetry are also impacted by the interference of ascorbic acid when it is present in high concentration in dopamine samples.33
Note also that the concentration of dopamine in our experiments (100 mM) is substantially higher than what is found in biological samples (several µM) and the detection limit has not been explored yet. Advanced cyclic voltammetry techniques are actually capable of detecting low µM concentration of dopamine.41 We envision three different strategies to enhance the sensitivity and detection limit of our sensor. First, the degree of chitosan crosslinking and the cantilever response in this work depend on the dopamine concentration and the oxidation time. For low dopamine concentrations the cantilever response can be enhanced by longer reaction times, although this would impact the temporal resolution of the detector. Second, the sensitivity of the cantilever can be further increased by reducing its spring constant to increase the bending for a given chitosan stress. Third, the interferometric measurement system in this work had large experimental error (∼30 nm) since it was not originally designed to measure through liquid. Cantilever displacement measurements in liquid with accuracy of 0.1 nm are routinely performed using the optical lever method.4 This method would considerably improve the sensitivity of dopamine mechano-detection.
Our second demonstration was the mechano-transduction of the electrochemical oxidation of dopamine. The reaction alters the chitosan film's mechanical properties,31,32 and this “mechanical information” is transmitted to the cantilever. More research is needed to provide quantitative explanations for this mechano-transduction. Fortunately, chitosan does not react with oxidation products from ascorbic acid, the common interfering species in biological samples. This differential reactivity confers selectivity to the mechanical detection of dopamine, although the detection method fails if the dopamine sample has a high concentration of ascorbic acid. Potentially, the coupling of electrochemical and mechanical measurements could provide independent and complementary information that could enhance the reliability of dopamine detection in the presence of interfering chemicals.
In a broader sense, we believe chitosan is a promising material that can perform important interfacial functions: assembly (electrodeposition), functionalization (conjugation)18 and now, mechano-transduction. It is interesting to contrast the capabilities of chitosan with better-studied interfaces created through self-assembled monolayers (SAMs). SAMs self-assemble at the interface (typically through gold-thiol bonds), while chitosan is directed-to-assemble onto electrodes (often gold) in response to applied signals. SAMs form monolayers at the surface, while chitosan forms 3-D gel networks. These differences may provide unique advantages for chitosan for mechano-sensor applications. Specifically, directed assembly may allow sensor addresses to be independently functionalized, while the 3-D structure may allow the generation of larger signals because more ligand can be functionalized onto the cantilever. Clearly, more work is necessary to demonstrate these potential advantages.
Footnote |
† Electronic supplementary information (ESI) available: Additional chemical evidence, UV/vis spectra and a table of properties. See DOI: 10.1039/b609149k |