Jer-Shing
Huang
and
King-Chuen
Lin
*
Department of Chemistry, National Taiwan University, Taipei 106 and Institute of Atomic and Molecular Sciences, Academia Sinica, Taipei 106, Taiwan. E-mail: kclin@ccms.ntu.edu.tw; Fax: 886-2-23621483
First published on 9th December 2004
We have developed a current normalization method for laser-induced breakdown spectroscopy (LIBS) for analyzing liquid droplets. An electrospray ionization device is employed to generate a stream of microdroplets. The spray needle serves as the anode, through which the analyte solution is spread toward the other metal base which is the cathode. Upon laser irradiation at the liquid droplets, the time-resolved laser-induced breakdown (LIB) emission and plasma-induced current signals are acquired concurrently on a single-shot basis. The plot of LIB emission intensity against the current intensity yields a straight line. The slopes in the correlation plots increase with the sample concentration, whereby the calibration curve is obtained. The resultant limit of detection (LOD) of Na sample may reach 0.6 ± 0.1 mg l−1, about 20 times better than that obtained by LIB/background normalization. The impinging laser energy dependence of both normalization methods is also investigated. The correlation linearity for the background normalization is found to be restricted within a small range of laser energy. When a two-line ratio is involved to account for the plasma temperature, its linear feature is markedly improved. According to the Boltzmann equation, the plasma temperature as a function of delay time relative to the onset of continuum background is determined to gain insight into the temperature effect on the LIB/background method. In contrast, the LIB/current method, which has taken into account the ablated amount of microdroplets and the plasma excitation temperature, retains correlation linearity over a much wider range of laser energy.
To compensate for this signal fluctuation, the average method over a number of laser shots is usually used, but the improvement of the signal-to-noise (S/N) ratio is offset by the nonlinear optical ablation effect.14,15 Internal standard method is an alternative solution via the measurement of the intensity ratio of analyte and reference element.16–19 The application of internal standardization to the chemical analysis of the multi-component samples requires knowledge of the internal standard concentrations. The reference element should also be assumed to have the same fluctuation pattern as that of the analyzed sample.15,19,20 To overcome this problem, Winefordner and co-workers developed a calibration approach based on the linear regression coefficient calculated between spectra of a range of certified standards and the spectrum of a reference sample.21 Without using any normalization, this method was demonstrated in the analysis of binary alloys using LIBS to give rise to results accurate to approximately 5% for major components. Schechter and co-workers adopted a correlation plot of analyte signal versus background to determine the sample concentration.15 They measured a sequence of spectra from single breakdown events in detecting trace metals in soil and aerosols. The line peak of analyte was found to correlate well with the baseline, which had the same multiplicative fluctuation as the sample signal. This method does not depend on a constant constituent of a reference element involved.
As with the method of internal standardization, external reference normalization is similarly reported to be useful in reducing the laser-induced breakdown (LIB) signal fluctuation. For instance, Cheung and Young, by applying LIBS in a liquid jet, found that the single-shot LIB signal with the corresponding acoustic normalization showed a linear relationship with the impinging pulse energy.14 The slope obtained by the plot of normalized LIB emission versus pulse energy may yield information of the sample concentration. Chaleard et al. quantized the LIB emission signals in air at atmospheric pressure by taking intensities of the emission lines as a function of the vaporized mass and the plasma excitation temperature.22 The ablated mass was accounted for by using an acoustic signal and excitation temperature was measured by a two-line method. Normalization of the LIB emission signals by these two parameters allowed for an accurate determination of Cu and Mn in various alloy matrices. Hakkanen et al. analyzed inhomogeneous paper coatings and found linear correlations between the LIB emission signals and the coat weight and the binder contents of the coatings.23
Applications of LIBS in liquid state are sparse compared with those in solid state. In recent years, we have developed a correlation method based on plasma-induced current normalization for analysis of the sample droplets.24 Instead of using bulk or jet as the liquid sampling methods,6,14,15 we have used an electrospray ionization needle to generate microdroplets of metal salt solutions in the study of matrix effects on the LIBS.25 The LIB and plasma-induced current signals were detected simultaneously on a single-shot basis.24 We found that the intensities of single-shot time-resolved LIB emission may linearly correlate with the corresponding current intensities. Given the calibration curve obtained from the slopes of the correlation plots, the limit of detection (LOD) for the Na analysis may reach 1 mg l−1, even with the matrix salts add up to 2000 mg l−1.25 The current normalization method is capable of suppressing the signal fluctuation, improving the LOD determination and concurrently correcting the matrix effect. Although LIBS applied in liquid state cannot be more sensitive than the conventional detection methods such as flame emission or furnace absorption, it resists matrix interference without losing much sensitivity. LIBS may also be operated in a hostile environment, which becomes difficult for conventional optical methods. In this work, we take a step further to characterize both normalization methods of LIB/current and LIB/background as a function of the laser energy and compare the resultant LOD of Na sample solution. We also discuss in detail the factor of plasma temperature, which should play an important role especially in the LIB/background normalization.
![]() | ||
Fig. 1 Schematic diagram of LIBS setup. |
The time-resolved LIB emission obtained was fed into either a transient digitizer (Model 9450A, LeCroy) for single-shot profile recording or a boxcar integrator (Model SR250, Stanford Research System) for signal processing. The time-resolved spectra of the analyte were integrated within a gate, which was opened at 900 ns delay relative to the onset of the continuum background emission. The LIB emission could then be the least interfered with by the intense continuum background. The gate width was optimized to 12 μs. The adjustment of the gate width and position may help suppress the LIB signal fluctuation to improve the signal-to-noise ratio.
Meanwhile, the correlation between the LIB signal and the continuum background was also analyzed from the same time-resolved spectrum. The background emission was integrated immediately following its onset within a 100 ns gate, while the LIB emission signal was integrated within a 12 μs gate in 900 ns delay. Although the plasma composition including continuum background, molecular emission, and atomic emission is hardly temporally isolated, the behavior of continuum background emission can be traced by substitution of the blank solution.24 Its maximum emission lasts for hundreds of ns and then decays rapidly close to the baseline. The gate width and position for the sample LIB signal are so selected as to gain the maximum intensity of LIB emission, but minimum interference from the continuum background.
For the experiment of plasma temperature correction, a two-line ratio was adopted, comprising two emission lines of Ca+ ion in the 4p2P1/2,3/2 → 4s2S1/2 transitions at 396.85 and 393.37 nm, respectively. These two lines were detected with an optical multichannel analyzer (OMA) (OMA III, Model 1460, EG&G PARC) mounted behind a monochromator centered at 395 nm. The OMA detector with temperature at −20 °C was triggered by a pulse generator (Model 1304, EG&G PARC), acquiring the spectral data through a scan control card (Model 1463, EG&G PARC).
![]() | ||
Fig. 2 Shot-to-shot correlation plot of LIB emission versus plasma-induced current as a function of Na concentration. The regression coefficients are 0.982, 0.980 and 0.984 for the concentration 140, 80 and 40 mg l−1, respectively. |
As was mentioned under Introduction, the continuum background emission in the frequency domain has been widely adopted to correlate with the LIB emission to diminish the shot-to-shot signal fluctuation. The continuum emission originates from a complicated superposition of various continuum spectra including the radiation from free-free transition, recombination of free electron and ion, free-bound transition and pseudo-continuum of strongly broadened lines.6,31 Similar to the analysis used in the frequency domain,15 in this work the signal intensities of LIB and background emission may be obtained from the integration of a time-resolved profile within an individual gate. Fig. 3 shows a linear plot of the LIB emission against the background emission on a single-shot basis, yielding a slope sensitive to the concentration variation.
![]() | ||
Fig. 3 Shot-to-shot correlation plot (with regression coefficient 0.953) of Na LIB emission versus continuum background emission at laser energy 21 mJ. |
The slopes obtained in either LIB/current or LIB/background plots as a function of the sample concentration give rise to a calibration curve, as shown in Fig. 4. Note that the calibration curves do not pass through the origin. This is caused by the fact that the slope of the correlation plot for the blank solution is not zero.24 It suggests that the LIBS signal and continuum background emission may not be completely temporally demarcated. Since the same magnitude of background is contributed to all other measurements for different concentrations of sample, the small, non-zero background does not affect the determination of sensitivity of the calibration curve. In the case of current normalization, given the standard deviation of the blank solution, σ, obtained from the slope of the correlation plot and the slope of the calibration curve, m, the limit of detection (LOD), as defined by 2σ/m, reaches 0.6 ± 0.1 mg l−1 for the Na sample at a pulse energy 23 mJ. The LOD and the linear dynamic range have been substantially improved in comparison with the treatment by averaging spectral intensities over the same number of laser pulses.24 In contrast, for the case of background normalization, the sensitivity obtained by the calibration curve along with the standard deviation of the blank solution may result in a LOD of 11.6 mg l−1 for the Na sample at the same laser energy 23 mJ, which is about 20 times worse than that determined in the LIB/current correlation method. The standard deviation of the blank solution determined in the background normalization method is five to six times larger than that with the current normalization. The gate width and position over the background emission have been varied and turn out to have similar results.
![]() | ||
Fig. 4 Calibration curves for the Na concentration obtained with the correlation methods of LIB/current (●) and LIB/background (○), respectively. The laser energy is at 23 mJ. |
![]() | ||
Fig. 5 Correlation plots of LIB emission of Na sample at 200 mg l−1versus current intensity obtained at the laser energy of (a) 15 (regression coefficient R = 0.942), (b) 21(R = 0.954) and (c) 42 mJ (R = 0.977). |
In contrast, the linearity for the Na LIB/background correlation holds only within a small range of laser energy about 21 mJ in this experiment. As shown in Fig. 6, when the energy is reduced to 15 mJ, the LIB emission signal begins to rise almost exponentially with the background emission. At reduced laser energy, the background emission signal is sensitive to the energy fluctuation, while the LIB emission rises only when the energy exceeds a threshold. Our work shows that the correlation linearity may not hold unless the threshold energy of air breakdown is reached. Nevertheless, the background emission can be easily saturated as the energy is larger than 25 mJ (Fig. 6).
![]() | ||
Fig. 6 Correlation plots of LIB emission of Na sample at 200 mg l−1versus continuum background emission obtained at the laser energy of (a) 15, (b) 21 and (c) 42 mJ. |
To verify this point, we adopt a two-line ratio method to monitor the plasma temperature. The Na sample is changed to Ca for a better selection of the two emission lines. In this case, the Ca+ ion emitting at 396.8 and 393.4 nm for the 4p2P1/2 → 4s2S1/2 and 4p2P3/2 → 4s2S1/2 transitions, respectively, are detected. As depicted in Fig. 1, an additional OMA detector is employed to acquire these two emission lines, while the other PMT is used to detect the time-resolved LIB emission of Ca atom in the 4s4p1P → 4s2 1S transition at 422.7 nm. The obtained background emission is first normalized to an area ratio of the two emission lines acquired without any time delay on a single-shot basis. Then the LIB intensity is plotted against the normalized background intensity, as in the case of the Na sample. After correction for the plasma temperature, Fig. 7(a) shows that deviation from linearity in the Ca LIB/background plot observed at low energy, 13 mJ, has been markedly reduced. When the energy is increased to 22 mJ, the LIB/background correlation may hold a straight line without any correction, as mentioned in the preceding section. At this energy, the linear feature remains almost the same before and after the temperature correction, as shown in Fig. 7(b). The accuracy of the two-line method depends on the energy difference of the upper states with respect to kT (k is Boltzmann constant) and the related oscillator strength.33 In spite of small energy difference, 222 cm−1, between the upper states, the selected two-line ratio of Ca+ fine-structure doublets seems appropriate to serve as a temperature correction factor. For cooler plasma conditions, like this work with low breakdown energy, the shot-to-shot variation of two-line ratio may reflect sensitively the fluctuation of background emission due to the temperature factor.
![]() | ||
Fig. 7 Correlation plot of LIB emission of Ca versus continuum background emission obtained at the laser energies of (a) 13 mJ and (b) 22 mJ. (○) Denotes the background emission without plasma temperature correction. (●) Denotes the background emission normalized by an area ratio of two emission lines, which are acquired without any time delay relative to the background emission. These two emission spectra acquired by an OMA detector are displayed in the inset. |
In the above treatment, we simply correct the factor of background plasma temperature. If the temperature gradient is non-zero, the continuum background and LIB signals, which are acquired in 0 and 900 ns delay, respectively, should both take into account the plasma temperature to improve the correlation linearity. Why can linearity of LIB/background be improved significantly even though the LIB signal is not corrected with the factor of plasma temperature? To answer this question, one should know how the plasma temperature changes in time delay of 0 and 900 ns relative to the onset of continuum background emission. Thus, we take a step further to determine temperature of the plasma plume at the times to acquire background and LIB signals by using multiple-line ratios based on the Boltzmann equation. According to the equation which assumes existence of local thermal equilibrium in the plasma plume, the intensity of a spectral line Iki emitting from the upper level k to the lower level i may be expressed as6,33
![]() | (1) |
For the Ca case, as listed in Table 1, six transition lines selected in the Boltzmann plot comprise three pairs of fine-structure doublets as the upper states emitting in the range 315–400 nm. The emission lines obtained at the laser energy 28 mJ are fitted with Lorentzian function to resolve the spectral overlap of the fine-structure doublets and then integrated to obtain individual line intensity Iki. As is shown in Fig. 8, the slope obtained in the Boltzmann plot yields temperature of 12900 ± 900 and 11
700 ± 700 K for a delay of 0 and 900 ns, respectively. The temperature turns cool by about 1200 ± 1100 K. Note that the plasma temperature behaves differently from that detected in metal aerosols by Yalcin et al. with pulse energy of 40–150 mJ at 532 nm.34 At least two aspects may explain the difference. First, a much smaller breakdown energy at 28 mJ was used in this work such that a lower temperature may result in the plasma. Second, a sampling method with microdroplet generation was adopted, different from the way used to form the metal aerosols. Thus, it is not surprising to find different temperature-cooling behavior. While the uncertainty range is considered, the temperature decay difference becomes smaller.
![]() | ||
Fig. 8 Boltzmann plot of Ln(Iλ/gA)
versus the upper state energy (see text for the notation). The Ca+ ion spectra of multiple emission lines are acquired in 0 (○, ---) ns and 900 (▲, ![]() |
λ/nm | E k/cm−1 | g k | Transition | A ki/(108) s−1 |
---|---|---|---|---|
315.886 | 56![]() |
4 | 4d2D3/2 → 4p2P1/2 | 3.1 |
317.933 | 56![]() |
6 | 4d2D5/2 → 4p2P3/2 | 3.6 |
370.602 | 52![]() |
2 | 5s2S1/2 → 4p2P1/2 | 0.88 |
373.690 | 52![]() |
2 | 5s2S1/2 → 4p2P3/2 | 1.7 |
393.366 | 25![]() |
4 | 4p2P3/2 → 4s2S1/2 | 1.47 |
396.846 | 25![]() |
2 | 4p2P1/2 → 4s2S1/2 | 1.4 |
The temperature change at these two delay times can alternatively be monitored by measuring the two-line ratios. The two emission lines acquired at different delay times are averaged over 2000 laser shots. Fig. 9 shows the time delay dependence of the area ratio of the two emission lines from the fine-structure doublets of the Ca+ ion in the 4p2P1/2,3/2 → 4s2S1/2 transitions at 396.85 and 393.37 nm, respectively. The decrease tendency of the ratio reflects the temperature decrease with the time delay. One should note that the temperature fluctuation is markedly reduced with increase in the time delay. Since the standard deviation of the ratio at the time to acquire the LIB signal is relatively small, one can realize why the correlation linearity may be markedly improved even though the temperature correction is ignored for the LIB signal. For this reason, when the background emission is first normalized by the two-line ratio acquired in 900 ns delay, the correlation plot of Ca LIB/background shows insignificant difference before and after plasma temperature correction at the laser energy 13 mJ (Fig. 10).
![]() | ||
Fig. 9 Two-line ratios obtained from Ca+ ion emission at 393.4 and 396.8 nm as a function of delay time relative to the onset of background emission. |
![]() | ||
Fig. 10 Correlation plot of LIB emission of Ca versus continuum background emission obtained at a laser energy of 13 mJ. (●) Denotes the background emission without plasma temperature correction. (△) Denotes the background emission normalized by an area ratio of two emission lines, which are acquired in 900 ns delay relative to the background emission. |