Neil W. Barnett*, Benjamin J. Hindson and Simon W. Lewis
School of Biological and Chemical Sciences, Deakin University, Geelong, Victoria 3217, Australia.. E-mail: barnie@deakin.edu.au
First published on UnassignedUnassigned14th January 2000
A simple and robust capillary electrophoresis chemiluminescence detection system for the determination of morphine, oripavine and pseudomorphine is described, based upon the reaction of these analytes with acidic potassium permanganate in the presence of sodium polyphosphate. The reagent solution was contained in a quartz detection cell which also held both the capillary and the anode. The resultant chemiluminescence was monitored directly using a photomultiplier tube mounted flush against the base of the detection cell. To ensure that no migration of the permanganate anion occurred, the anode was placed at the detector end whilst the electroosmotic flow was reversed by the addition of hexadimethrine bromide (0.001% m/v) to the electrolyte. The three analytes were separated counter to the electroosmotic flow via their interaction with α-cyclodextrin. The methodology realised detection limits (3 × S/N) of 2.5 × 10−7 M for both morphine and oripavine and 5 × 10−7 M for pseudomorphine. The relative standard deviations of the migration times and the peak heights for the three analytes ranged from 0.6 up to 0.8% and from 1.5 up to 2.1%, respectively.
Morphine, oripavine and pseudomorphine are alkaloids (see Fig. 1) which are extracted from commercially grown opium poppies (Papaver somniferum) and both flow analysis37–39 and high performance liquid chromatography,40 with acidic potassium permanganate chemiluminescence detection, have been used to monitor these analytes in a variety of process streams. Notwithstanding this success, it was decided to investigate the feasibility of coupling this type of chemiluminescence detection to capillary electrophoresis with a view to eliminating the concomitant interferences sometimes observed with flow analysis whilst reducing the time taken per analysis. Recently, Bjørnsdottir and Hansen41,42 reported a capillary electrophoresis procedure for the determination of six Papaver somniferum alkaloids present in opium and pharmaceutical preparations. This methodology41,42 employed an aqueous electrolyte containing cyclodextrins and UV detection at 214 nm, which realised a detection limit of 1 × 10−6 M for morphine.
![]() | ||
Fig. 1 Structures of (A) morphine, (B) oripavine and (C) pseudomorphine. |
This paper describes some preliminary results on the determination of morphine, oripavine and pseudomorphine using a modified version of the separation previously described by Bjørnsdottir and Hansen41,42 coupled with a simple and robust acidic potassium permanganate chemiluminescence detection system.
![]() | ||
Fig. 2 Schematic diagram of the capillary electrophoresis instrumentation coupled to the chemiluminescence detection system. |
Chemiluminescence was monitored by mounting the detection cell flush against the end window of a photomultiplier tube (R9798SB, Thorn EMI, Ruislip, Middlesex, UK) operated at 1100 V provided by a stable power supply (PM28BN, Thorn EMI). The signal from the photomultiplier tube was connected to a transimpedance amplifier (Type A1, Thorn EMI) and the output recorded using a chart recorder (Type 3066, Yokogawa Hokushin Electric, Tokyo, Japan).
The separation buffer consisted of 50 mM 6-aminocaproic acid prepared in 0.001% m/v hexadimethrine bromide (diluted from a 0.1% m/v stock solution). The pH of this solution was adjusted to 4.5 with dilute acetic acid (1 + 4). Once the pH had been adjusted, this solution was used to dissolve the appropriate amount of α-cyclodextrin. The chemiluminescence reagent was prepared by dissolving the desired quantity of potassium permanganate and sodium polyphosphate followed by adjustment to the required pH with concentrated sulfuric acid. The process liquor was diluted with water (1 + 49) prior to analysis.
Resolution (Rs) was calculated using the equation Rs = 2[(t2 − t1)]/W1 + W2, where t is the migration time in minutes, W is the baseline peak width in minutes and the subscripts 1 and 2 refer to morphine and oripavine, respectively.
Bjørnsdottir and Hansen42 reported the electrophoretic separation of six opiate alkaloids using the following conditions: 64 cm × 50 μm id fused silica capillary, an applied voltage of 30 kV with a 50 mM 6-aminocaproic acid buffer adjusted to pH 4.0 containing heptakis(2,6-di-O-methyl)-β-cyclodextrin. In our preliminary experiments, based upon this study,42 we utilised a similar electrolyte (including hexadimethrine bromide at 0.001% m/v) with a 32 cm × 75 μm id fused silica capillary and an applied voltage of −20 kV, which resulted in currents ranging from 90 to 120 μA. Adjustment of the electrolyte to pH 4.5 reduced the current to an acceptable level (48 μA). The lower current minimised the undesirable effects which Joule heating can have on the separation.47 The unavailability of heptakis(2,6-di-O-methyl)-β-cyclodextrin led us to explore the feasibility of separating our three analytes with and without α-cyclodextrin. In the absence of α-cyclodextrin, pseudomorphine was separated from morphine and oripavine, but the latter two analytes were not resolved from each other. The inability to separate morphine from oripavine is not surprising considering the structural similarities of these two compounds (see Fig. 1). Furthermore, this result is consistent with that reported by Bjørnsdottir and Hansen,41 who showed that without any cyclodextrins present, the separation of six Papaver somniferum alkaloids yielded only two peaks. Subsequently, the effect of the addition of α-cyclodextrin to the electrolyte over the concentration range 30–80 mM was examined with complete separation of the three analytes being achieved at 80 mM.
Our previous experience with acidic potassium permanganate detection chemistry37–39,48,49 has shown that performing a series of univariate searches, with respect to certain parameters, may increase the chemiluminescence intensity. The variables investigated in this study were the concentrations of potassium permanganate and sodium polyphosphate and the pH. The effect of potassium permanganate concentration over the range 0.01 to 1 mM on the chemiluminescence emission is shown in Fig. 3(a). Not only did the emission intensity increase with increasing potassium permanganate concentrations up to 0.5 mM but a significant improvement in apparent resolution was also observed. The resolution between oripavine and morphine was calculated, using the equation given earlier, to be 0.9 and 1.5 at 0.075 and 0.75 mM potassium permanganate, respectively. This result was probably due to more rapid kinetics of the chemiluminescent reaction as the ratio of reagent to analyte was increased, thereby reducing band broadening as a consequence of dilution. The concentration of sodium polyphosphate had a marked effect on the sensitivity [see Fig. 3(b)], particularly in the region from 0.0025 up to 0.075% m/v, where a steep rise in signal intensity occurred. Increasing the concentration of sodium polyphosphate, from 0.25 up to 1% m/v, produced only small changes in the chemiluminescence response for the three analytes. Although the mechanism of this detection chemistry has, as yet, not been elucidated, the sodium polyphosphate clearly plays an essential role in the production of analytically useful chemiluminescence under these and similar conditions.37–39,48,49 The chemiluminescence response was also affected by the reaction pH [see Fig. 3(c)], fortuitously with the three analytes each exhibiting maximum emission intensity at approximately pH 1.9. Based on these observations, we chose the chemiluminescence reagent conditions 0.5 mM potassium permanganate and 0.5% m/v sodium polyphosphate at pH 1.9 for all further experiments.
![]() | ||
Fig. 3 (a) Effect of potassium permanganate concentration on the chemiluminescence response of (A) morphine, (B) oripavine and (C) pseudomorphine (1× 10−5 M). The sodium polyphosphate concentration was 0.1% m/v, adjusted to pH 2.0 with concentrated sulfuric acid. The electrolyte consisted of 50 mM 6-aminocaproic acid, pH 4.5, 0.001% m/v hexadimethrine bromide with 80 mM α-cyclodextrin added. Other conditions were applied voltage –20 kV, injection time 10 s, current 49–56 μA, 20 °C, capillary 32 cm × 75 μm id. (b) Effect of sodium polyphosphate concentration on the chemiluminescence response for (A) morphine, (B) oripavine and (C) pseudomorphine (1 × 10−5 M). The potassium permanganate concentration was 0.5 mM, each solution contained the appropriate concentration of sodium polyphosphate and was adjusted to pH 2.0 with concentrated sulfuric acid. All other conditions as in (a). (c) Effect of pH on the chemiluminescence response for (A) morphine, (B) oripavine and (C) pseudomorphine (1 ×10−5 M). The potassium permanganate concentration was 0.5 mM and the sodium polyphosphate concentration was 0.5% m/v, adjusted to the required pH with concentrated sulfuric acid. All other conditions as in (a). |
In addition to the change in the apparent resolution with permanganate concentration mentioned previously, a flowing as opposed to a static reagent was also observed to have a profound effect on the separation (see Fig. 4). Contrary to what was expected, a static reagent yielded both poorer apparent resolution and sensitivity when compared with a flowing reagent. Under the static conditions, the reagent may be diluted or displaced from the tip of the capillary by the electrolyte. Consequently, dilution and dispersion of the analyte zones may occur prior to their chemiluminescent detection. In contrast, pumping fresh potassium permanganate into the detection cell resulted in a continuous supply of fresh reagent to the region surrounding the tip of the capillary and as such the analytes reacted immediately upon elution. As with our earlier observations, when using increased reagent concentrations, the explanation for this unexpected behaviour probably stems from the rapid kinetics of this detection chemistry, which are clearly faster than dilution (see Fig. 4).
![]() | ||
Fig. 4 Electropherograms showing the effect of reagent flow on the separation of (A) morphine, (B) oripavine and (C) pseudomorphine (1 × 10−5 M), using (a) a static reagent and (b) a flowing reagent. The potassium permanganate concentration was 0.5 mM, the sodium polyphosphate concentration was 0.5% m/v and the solution was adjusted to pH 1.9 with concentrated sulfuric acid. All other conditions as described in Fig. 3(a). |
Calibration functions were obtained for oripavine, morphine and pseudomorphine using a series of 12 mixed standard solutions over the concentration range 2.5 × 10−7–1 ×10–4 M and the analytical figures of merit are summarised in Table 1. The responses observed for morphine, oripavine and pseudomorphine, over the entire concentration range, closely approximated linearity and this is reflected by the slopes of the log–log calibration functions (see Table 1). Although the detection limits attained were superior to those achieved by Lee and Whang,36 such a direct comparison of performance would not be strictly valid owing to the different nature of the analytes. Likewise, it is similarly inappropriate to compare the peak height precision achieved during the present study (see Table 1) with the peak area precision reported by Lee and Whang36 (relative standard deviations from 4.9 up to 8.3%). However, it is our contention that as we achieved superior precision values at a lower analyte concentration, using peak height, then our methodology would appear to be more reproducible than that described by Lee and Whang.36 As with our earlier report,34 using tris(2,2′-bipyridyl)ruthenium(III), the satisfactory precision attained during the current investigation probably reflected the simple and robust nature of the detection device. It should be borne in mind that the results described here are of a preliminary nature, as a thorough optimisation of the detector configuration has not yet been completed. Notwithstanding this, the approach described above appears to offer certain advantages for the determination of these and other analytes which elicit analytically useful chemiluminescence upon reaction with acidic potassium permanganate.48–50 Specifically, the determination of these three analytes in process liquors. As can be seen from Fig. 5, the UV absorbance electropherogram exhibited inferior sensitivity for oripavine and pseudomorphine compared with that achieved with chemiluminescence, as would be expected from the UV absorbance detection limits shown in Table 1. As can also be seen from Fig. 5(a), the analysis time was 4 min, which compared favourably with a run time of 10.2 min using the current Glaxo Wellcome standard HPLC methodology.40 Furthermore, the major possible interferences (oripavine and pseudomorphine) with respect to flow analytical methodology are clearly separated from morphine [Fig. 5(a)]. Whereas in certain process samples the morphine to oripavine and pseudomorphine ratio is high [see Fig 5(a)], in other liquors this ratio is significantly lower, hence the contribution of the interferences is much larger with respect to flow analysis. It is also evident from Fig. 5 that a peak corresponding to the electroosmotic flow (1.4 min) was observed in the UV absorbance electropherogram, but, such a peak was not visible when chemiluminescence detection was employed. In order to monitor the electroosmotic flow when using this type of detection, a suitable neutral marker species which also elicits chemiluminescence upon reaction with acidic potassium permanganate would be required. Upon completion of the optimisation, the methodology described above will be validated against the current standard HPLC method,40 and these results will be reported in due course.
Relative standard deviation (%)d | |||||||
---|---|---|---|---|---|---|---|
Analyte | Detection limita/M | Calibration functionb | Correlation coefficient (r2) | Log–log calibration functionc | Correlation coefficient (r2) | Migration time | Peak height |
a Calculated as 3 × S/N, values in parentheses were estimated using UV absorbance detection at 214 nm.b y = peak height/mV, x = concentration/M.c y = log (peak height/mV), x = log (concentration/M).d n = 10 at 2.5 × 10−5 M. | |||||||
Morphine | 2.5 × 10−7 (5 × 10−6) | y = 2.08 × 106x | 0.9985 | y = 0.99x + 6.23 | 0.9988 | 0.6 | 1.5 |
Oripavine | 2.5 × 10−7 (5 × 10−6) | y = 1.84 × 106x | 0.9910 | y = 1.04x + 6.38 | 0.9963 | 0.7 | 1.5 |
Pseudomorphine | 5 × 10−7 (5 × 10−6) | y = 1.15 × 106x | 0.9976 | y = 0.97x + 5.95 | 0.9978 | 0.8 | 2.1 |
![]() | ||
Fig. 5 Electropherograms showing the separation of (A) morphine, (B) oripavine and (C) pseudomorphine in a dilute process liquor (1 + 49) using (a) chemiluminescence detection and (b) UV absorbance detection at 214 nm. All other conditions as in Fig. 4(b). |