David J. Kinahan*ab,
Sinéad M. Kearneyab,
Olivier P. Faneuilabc,
Macdara T. Glynnab,
Nikolay Dimov‡
ab and
Jens Ducrée*ab
aBiomedical Diagnostics Institute, National Centre of Sensor Research, Dublin City University, Glasnevin, Dublin 9, Ireland. E-mail: david.kinahan@dcu.ie; jens.ducree@dcu.ie
bSchool of Physical Sciences, Dublin City University, Glasnevin, Dublin 9, Ireland
cTelecom Physique Strasbourg, Université de Strasbourgh, Strasbourg, France
First published on 1st December 2014
Rotational microfluidic platforms have attracted swiftly growing interest over the last decade due to their suitability for integration and automation of sample preparation and detection. Valving is of pivotal importance on these compact “Lab-on-a-Disc” (LoaD) platforms as all liquids are exposed to the same centrifugal field. A number of valving technologies have been developed to coordinate timing of serial and/or parallel multi-step/multi-liquid assay protocols comprising of laboratory unit operations (LUOs) such as the release, metering and mixing of sample and reagents. So far these valving techniques could be broadly categorised into rotationally controlled or externally actuated schemes. Only recently a new, “event-triggered” flow control has been introduced. In this approach, a valve is opened upon arrival of a liquid at a defined destination on the disc; this innovative mechanism for the first time permits the cascading of LUOs independent of the spin rate. In one technology, dissolvable films (DFs) are configured with a pneumatic chamber to offer function akin to an electrical relay. Dissolving one DF, termed the control film (CF), results in the release of liquid at a distal location through a so-called load film (LF). In this paper, a new method for temporal control of actuating DF-based, event-triggered CFs which are serially aligned at defined distances along a paper strip is introduced. Liquids are transported through the paper strip at a given velocity, thus setting well-defined intervals between subsequent LUOs, e.g. incubation steps. As a proof-of-concept, we present a disc with integrated metering and mixing which can perform a prototypical, 4-fold serial dilution; a common function in bioanalytical protocols. Imbibition of the paper strip sequentially opens five valves for serial dilution and mixing. To illustrate an unprecedented level of on-disc automation, this is followed by a branched cascade of 17 event-triggered valves (for a total of 22 liquid handling steps) which completes the serial dilution protocol.
As all liquid residing on a disc is exposed to the same centrifugal field, valving technologies are critical to implement and coordinate a sequence of LUOs7 including metering, mixing, and reagent release typical of most assay protocols.
Traditional flow control techniques for LoaD platforms could be categorized into externally actuated and rotationally controlled schemes. In the former, a peripheral instrument provides power to ablate sacrificial material or induce a phase-change,8–12 to pressurize5 or to physically manipulate the disc.13 This instrument focussed approach expands the number and flexibility of LUOs which can be performed along a single process chain on a cartridge. But this advanced control tends to compromise the innate simplicity of both the instrument and the cartridge at the core of the LoaD system.
The more ubiquitous, rotationally actuated valves are triggered by a change in the disc spin rate which unbalances the equilibrium between centrifugally induced hydrostatic pressure and, for instance, the pressure induced by capillary action or compression of gas pockets. These valves might be categorised into high-pass and low-pass valves. High-pass valves, which are actuated through an increase in spin rate, include capillary burst valves,14–18 centrifugo-pneumatic dissolvable-film (DF) valves,19 burstable foils,20 elastomeric membranes21 and dead-end pneumatic chambers.22 Conversely, low-pass valves are actuated through a reduction in the rotational frequency. They are often based on siphons4,23,24 where a decrease of the spin rate lets the capillary force prevail the hydrostatic pressure and so liquid can move radially inbound to prime a siphon channel. As this technology demands a hydrophilic channel which is difficult to stabilize on common polymer surfaces, centrifugo-pneumatic mechanisms based on centrifugally controlled compression of gas have been used to prime even slightly hydrophobic microchannels innate to most polymers.25,26
The chief drawback of common, rotationally-actuated capillary burst valves is that their operational range is limited by the minimum achievable feature sizes and tolerances associated with their manufacturing techniques. This caps the maximum spin rate available during centrifugal liquid handling and also smears the definition of burst frequencies; both factors thus restrict the number of discrete LUOs which can be automated in a single process. While this limitation has been, to some extent, circumvented through combining low-pass and high-pass valves,4,23 this strategy can be unreliable and consumes precious disc real-estate.
Lately, a new paradigm in rotational flow control has been introduced. In this event-triggered technique,27 the arrival of liquid at defined locations on the disc prompts the opening of another, distant valve through pneumatic coupling. This technology constitutes an advancement of previously described centrifugo-pneumatic DF valving19,28–30 and operates akin to an electrical relay. These valves27 are composed of a pneumatic chamber sealed by two DFs; called the Load Film (LF) and the Control Film (CF). The pneumatic structure is shaped such that, at typical spin rates, the restrained liquid cannot penetrate into the chamber sufficiently to wet and dissolve either film. However, when the CF is dissolved by an (ancillary) liquid, the restrained liquid enters the pneumatic chamber, wets and dissolves the LF and thus exits through this outlet.
Based on these event-triggered valves a number of LUOs including release of wash reagents, liquid metering and selective liquid routing were demonstrated.27 In addition, advanced process control such as parallelization and logical operations such as AND-condition triggering27 was implemented. As a proof-of-concept, these valves were employed for realizing a multi-liquid/multi-step solid-phase RNA extraction protocol on a LoaD platform.27 These event-triggered valves offer a number of advantages, including operation without peripheral instrumentation and their quasi independence of the spin rate.27 Nevertheless, a constraint of these valves rests in temporal process control; the interval between subsequent valve actuations depends on the time taken for a liquid to dissolve a DF and the time for the released liquid to be pumped to the CF of the next valve in the cascade. This interval, which is typically of order 120 s, may not suffice for some LUOs such as extended mixing or biological incubation steps.
In this manuscript a new method for actuating event-triggered DF valving is introduced which is based on transporting liquid to the CF of a valve through a low-cost and ubiquitous paper strip31,32 integrated on a polymeric disc.33–35 Combined with the material-specific speed of imbibition, the spacing between the CFs defines the (extended) time span between valve openings. Indeed, imbibition has been used previously to as a valving technology in (non-centrifugal) paper microfluidics,36–38 while the addition of a sucrose solution to paper has been harnessed to modulate the speed of liquid movement.39 Wicking through paper has also been utilized to trigger electrical/magnetic valves integrated with lateral flow devices.40
In this paper we first describe the valving concept and the manufacture of the disc. We then present a simplified disc to illustrate the sequential opening of three valves with timing controlled through paper imbibition. Towards application in common assays to determine protein concentrations,41,42 we then demonstrate a four-fold serial dilution (of dyed water) along with the on-disc preparation of calibration controls and the further diluted samples with additional reagents (DI water). This platform orchestrates the opening of a series of five valves by paper imbibition, hence providing sufficient time for mixing LUOs to take place. Furthermore, to underpin the potential of event-triggered valves for large-scale on-disc automation, a further, 17-valves are triggered to run several post-processing steps.
In the technology introduced in this work, the actuation mechanism of the DF-based event-triggered valves is fundamentally changed to time longer running LUOs. This is implemented through paper imbibition advancing from a reservoir to a sequence of CFs. For DF-based event-triggered valves, a minimum centrifugal field is required to pump the liquid through the pneumatic chamber (typically shaped like a microchannel) to the LF. At given hydraulic diameters and lengths of the microchannels, it then can be assumed that wicking of liquid through paper will be slower than the corresponding microchannel flow. Thus, the speed of the ancillary liquid travelling to the CF is reduced to extend the time span for opening a single valve.
This mechanism is illustrated in Fig. 1. Initially, the DF-based, event-triggered valves will not release liquid from any of the reservoirs, even at high spin rates. Following loading of the ancillary liquid, the paper strip is wetted at one end and the liquid front begins to progress along the paper strip. Upon reaching the CF of a valve, it dissolves, thus venting the pneumatic chamber and permitting the flow of the restrained liquid into the pneumatic chamber. This, in turn, wets and dissolves the LF, consequently opening an exit route through the valve.
It can be inferred from Fig. 1 that as the liquid front progresses further along the paper strip, the CFs for downstream valves will also be wetted and thus opened in sequence. This permits the addition/release of a single liquid on the disc to control a series of valves and so significantly simplify the disc architecture. The sequence of valve opening is defined by the position of the CFs along the paper strip, while the time between valve actuations is defined by the space between the CFs. This interval for the liquid front to advance along the paper strip is also influenced by the grade/thickness of paper and can also be influenced through pre-treatment of the paper strips.39
The disc is manufactured using broadly the same layer configurations described previously.27 In sequence, the outer layer (Layer 1) is manufactured from PMMA and contains loading vents. Layer 2 is cut from PSA and consists of microchannels for pneumatic venting and liquid transport. Note that, for the disc cartridge in the initial study presented in Fig. 2, a single layer for PSA was used. For the disc in the serial dilution study, the top-level PSA layer was doubled by rolling two sheets of PSA together before placing it on the knife-cutter. Due to the complexity and density of structures on this disc, it was found that this thicker layer of PSA improved sealing of the disc. For this disc, these identical layers are referred to together as Layer 2 and separately as Layers 2a and 2b. The large reservoirs for holding liquid are made of voids in the PMMA Layer 3. Layer 4, machined in PSA, acts as a cover supporting the DF tabs. The material cut from this layer also limits the amount of DF material exposed. Note that, during assembly, the paper strip is placed into a reservoir (created by Layer 3) and is adhered to Layer 4 while applying gentle pressure. It should also be noted that the reservoir holding the paper strip is extended through Layers 3–6 at the point where the ancillary liquid contacts the paper strip. This increases the volume available and, as this location is radially farther outwards than the paper strip, ensures the DFs are only wetted through imbibition.
![]() | ||
Fig. 2 Serial actuation of valves through paper imbibition. (a) Schematic of this disc architecture. Control films are indicated by CF and the corresponding Load Film by LF. (b) Reservoirs are loaded with pseudo-reagents (dyed water) for visualisation purposes. The valves will not burst at typical spin rates (0–60 Hz) until the paper trip is wetted. The locations of the CFs show as dark shadows beneath the paper strip and are also indicated by red arrows. During the experiment the disc rotates at a constant 30 Hz. The direction of centrifugal force is shown by green block arrows. (c) The ancillary liquid is loaded and wets the paper strip. For the duration of the experiment, the spin rate is maintained at a 30 Hz. (d) Liquid wicks along the paper strip (indicated by light green arrows). (e) The liquid front contacts and dissolves CF1, thus venting the pneumatic chamber. As the gas is escapes, Reagent 1 enters the valve, wets LF1 and is released (indicated by light green arrows). The active connecting pneumatic channel is shown by a blue line. (f–g) Valves 2 and 3 are opened using the same mechanism. (h) Burst times for the valves presented here (n = 3). The intervals are measured at the time the CF is observed to be vented (through observation of liquid reaching the LF in the corresponding valve) (see Movie 1 and 2†). |
The DF tabs are placed into, and supported by, Layer 5 (PSA). Layer 6 (PMMA) contains through holes to connect microchannels located in the different disc layers, while Layer 7 (PSA) displays the lower-level microchannels for venting and liquid flow. Finally, Layer 8 (PMMA) acts as the floor of these channels. This multi-level architecture isolates microchannels on Layer 2 and Layer 7 and thus permits the channels to cross over.
The paper strip used in the simplified disc was manufactured using a slower wicking paper strip (Whatman CHR Grade 20 (Sigma-Aldrich P/N 3020-917), 170 μm thick, flow rate: 85 mm/30 min) while a faster grade, (Whatman CHR Grade 1 (Sigma-Aldrich P/N 3001-917), 180 μm thick, flow rate: 130 mm/30 min) was used for the serial dilution disc. The DFs used in this study are non-adhesive and thus were mounted onto double-sided PSA tabs as described previously.19,27 The DF (Barnyarns, Rippon, UK), here termed E-film, is a low-cost component based on polyvinyl alcohol which is primarily used for embroidery. This film has been characterised previously27 and typically dissolve in the presence of DI water in less than 10 s.
A specialised experimental test and development instrument,27,44,45 generates images of the disc while in motion. Aside from where otherwise specified, all discs are tested at a spin rate of 30 Hz. The discs are accelerated and slowed down at 12.5 Hz s−1. For the serial-dilution discs, a custom program (written in LabVIEW) varies the rotational frequency between 20 Hz and 30 Hz at set time intervals (10 s). This induces ‘shake-mode’ mixing46 by angular acceleration (i.e. the so-called Euler force) and, where valves are present, pneumatically induced mixing.47
In order to determine the concentration of food dye present on-disc, the samples were taken off-disc and pipetted onto the micro-titre plate in duplicate volumes of 40 μl. Absorbance was measured at 505 nm wavelength and the dye concentrations were interpolated using standard curves generated from the reference chambers also present on-disc. Absorbance measurements were not made ‘on-disc’ as such techniques are widely published4,30,48 and are not the innovation focus of this work.
These samples are then mixed with two known liquid volumes (DI water, called Reagent 1 (20 μl) and Reagent 2 (160 μl)). This representative experiment permits demonstrates our multi-step automated liquid handling technology while using volumes commensurate with the geometries we can achieve with our prototyping technology. The on-disc work-flow is shown in Fig. 3.
For the purposes of demonstrating the capabilities of the valves, it is considered that Reagent 2 represents the active agent in a kinetic assay (common in many analyte estimation assays requiring a serial-dilution based standard curve) and thus should be added to both diluted samples and references in as synchronous a manner as possible.
The protocol may be divided into four separate sections:
1. Dilution, mixing and metering of the sample controlled by paper imbibition
2. Transfer of diluted samples to read chambers located on the periphery of the disc and mixing with Reagent 1
3. Addition of Reagent 2 to the read-chambers
4. Addition of Reagent 2 to the reference chambers.
![]() | ||
Fig. 4 The serial dilution disc according to the workflow described in Fig. 3. (a) An image of the disc directly after the start of the experiment. Green dyed water wets the base of the paper strip to commence wicking and also as an ancillary liquid to open Valves 9–11. Note that Mixing Chamber (MC) 1 and MCs 2–5 have been loaded with 70 μl of dilute food dye (considered stock, 100% concentration) and 35 μl of DI water, respectively. Each of the read chambers (RCs) has been pre-loaded with 20 μl of DI water and each Diluent Chamber ‘X’ has been loaded with 160 μl of DI water. Reference RCs 1–6 are pre-loaded with 35 μl of dyed water (of known concentration) and 20 μl of DI water. These concentrations are 100%, 80%, 60%, 40%, 20% and 0%, respectively. (b) Same disc following completion of the liquid handling workflow. Note the different intensity of food dye in the RC and REF chambers (indicating dilution has taken place; nominal dilution in each reservoir is indicated in red text). The overlay boxes indicated the regions shown in the subsequent subfigures (c-i to c-vi) These images show the wetting of CF1 through paper imbibition and the release and transfer of a metered volume of dyed water from MC11 into MC2. (d-i to d-iii) Transfer of liquid from MC5 into the overflow chamber. This leaves a metered volume of liquid present at the base of each MC. (e-i to e-iii) These images show the addition of diluted sample to pre-loaded DI water resulting in the wetting and dissolving of CFs. In all RCs, the dissolution of the co-located CFs/LFs results in the release of the diluent into the RC. In addition, in RC1 (shown) a second CF is dissolved. This triggers the release of diluent into the first reference read chamber (REF) and triggers a cascade of six valve actuations to add diluent to all REFs. Note also in these images that Euler mixing can be observed which was induced by the spin profile (see Movie 3 and 4†). |
The event-triggered valves can function broadly independent of the spin rate. With valve actuation controlled by paper imbibition (thus extending the time between actuations to ∼150 s, Fig. 5a), the disc can be subjected to a robust spin profile (alternating the rotational frequency every 10 s between 20 Hz and 30 Hz with acceleration/deceleration at 12.5 Hz s−1) without opening the valves prematurely. This decoupling of the spin rate from valve actuation also enables the afore-mentioned mixing strategies.
The convective mixing is particularly important as it can be seen that, at the high centrifugal field, small density changes can result in the stratification of liquids (see pneumatic mixing in Fig. 4c-iv and c-vi and Euler mixing in Fig. 4e-iii).
Due to the distance between the overflow structure and the MCs on the disc, the overflow released by V5 triggers V7 and V8 but an intermediary ancillary liquid (through V6) triggers V9–11 (Fig. 3). The overflow from opening of V5 actuates V6–8 (Fig. 3 and 4d). The samples then flow into their respective RC chamber where they can mix with pre-loaded Reagent 1 (20 μl).
Note it was found that in most cases the intermediate ancillary liquid was not released through Valve 6 in a timely manner (Fig. 5b). This is primarily a result of poor manufacturing fidelity, the radially inward location of the ancillary liquid, and the small fluid volume (leading to a small hydrostatic pressure head). Nevertheless, in the cases where this valve opened within an expected time frame all diluted samples were added to their respective read chambers within an ∼30 s window.
Similar to the step described in Section 5.4, a second CF is located in RC1. Upon addition of sample to RC1 (as described above), a valve is actuated (wetting of CF17) and cascades the release of Valves 17–22. This releases 160 μl of Reagent 2 into each reference chamber to mix with the pre-loaded sample/Reagent 1 mixture (Fig. 3 and 4e).
The actuation times for V17–22 (REF1–6), relative to the addition of sample to RC1, are shown in Fig. 5c. It can be seen that these valves triggered rapidly (∼15 s per valve); all six valves were actuated in 71 s.
From the experimental data the ratio of each dilution to the next is 2.1, 1.9, 1.5, 1.6, meaning the first chamber dilutes the sample more than expected by design while the subsequent chambers dilute it less than by design. This effect is most likely a result of losses to dead volumes in the valves and microchannels (as observed during experiments) and/or variation in system performance due to the differing radial positions of the mixing chambers (i.e. different flow properties due to the different local centrifugal force).
While these issues bear further investigation, it should be noted that, from disc to disc/chamber to chamber (Fig. 6), the variation in the dilution ratio was quite small. Therefore, it is entirely feasible that through system characterisation, iterative design and better manufacture a highly automated and repeatable system might be created based on this technology.
Along with increasing the scope of temporal process control, this technology offers further benefits. The ability to place CFs at known spaces along a paper strip permits an extended sequence of process steps to be initiated through the movement of a single liquid element. This has particular advantages for complex processes, as the use of a single paper strip to trigger multiple valves can simplify disc architectures relative to the use of conventional, DF-based event-triggered valves. Nevertheless, a challenge of this approach is the strict dependency between valve actuations is lost; should a CF not be wetted/dissolve fully, but a downstream CF is, the valve actuation sequence will be disarrayed.
Due to their independence of the spin rate, event-triggered valves (actuated by both paper imbibition and bulk flow) enable LUOs which are governed by the spin profile. In this case, it enabled aggressive Euler and pneumatic mixing through the use of a ‘shake-mode’ spin profile. In addition, aside from a single valve which was affected by manufacturing issues, it was found that both the paper imbibition and bulk flow triggered valves actuated within reliable time frames. This is very important both, to permit sufficient time for mixing and to enable kinetic assays on this platform. In fact, it may be inferred that, with more robust design and better manufacturing techniques, kinetic reagents can be delivered to all eleven read chambers in less than 100 s.
Beyond timing of DF-based event-triggered flow control, paper wicking can endow further capabilities of the LoaD platform. For example, the already established modulation of the speed of paper imbibition via the centrifugal force33 may leverage novel schemes for high- and low-pass valving.
In conclusion, we demonstrated paper-triggered DF valves can be combined with the previously introduced, event-triggered valves to coordinate the timing of a complex series of LUOs on a disc widely independent of the spin rate. In this work, an unprecedented level of automation has been introduced through the implementation, in parallel and in series, of 22 discrete liquid handling steps.
Footnotes |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c4ra14887h |
‡ Current Affiliation: Dept. of Biochemical Engineering, University College London, United Kingdom. |