Jo-Han
Ting
a,
Po-Chuan
Lin
b,
Shivam
Gupta
a,
Ching-Hao
Liu
a,
Tzuhsiung
Yang
b,
Chi-Young
Lee
a,
Yi-Ting
Lai
*cde and
Nyan-Hwa
Tai
*a
aDepartment of Materials Science and Engineering, National Tsing Hua University, Hsinchu, 300, Taiwan, ROC. E-mail: nhtai@mx.nthu.edu.tw
bDepartment of Chemistry, National Tsing Hua University, Hsinchu, 300, Taiwan, ROC
cDepartment of Materials Engineering, Ming Chi University of Technology, New Taipei City, 24301, Taiwan, ROC. E-mail: laieating@mail.mcut.edu.tw
dCenter for Plasma and Thin Film Technologies, Ming Chi University of Technology, New Taipei City, 24301, Taiwan, ROC
eBiochemical Technology R&D Center, Ming Chi University of Technology, New Taipei City, 24301, Taiwan, ROC
First published on 7th August 2023
Non-invasive methods for sensing glucose levels are highly desirable due to the comfortableness, simplicity, and lack of infection risk. However, the insufficient accuracy and ease of interference limit their practical medical applications. Here, we develop a non-invasive salivary glucose biosensor based on a ferrocene-chitosan (Fc-Chit) modified carbon nanotube (CNT) electrode through a simple drop-casting method. Compared with previous studies that relied mainly on trial and error for evaluation, this is the first time that dipole moment was proposed to optimize the electron-mediated Fc-Chit, demonstrating sturdy immobilization of glucose oxidase (GOx) on the electrode and improving the electron transfer process. Thus, the superior sensing sensitivity of the biosensor can achieve 119.97 μA mM−1 cm−2 in phosphate buffered saline (PBS) solution over a wide sensing range of 20–800 μM. Additionally, the biosensor exhibited high stability (retaining 95.0% after three weeks) and high specificity toward glucose in the presence of various interferents, attributed to the specific sites enabling GOx to be sturdily immobilized on the electrode. The results not only provide a facile solution for accurate and regular screening of blood glucose levels via saliva tests but also pave the way for designing enzymatic biosensors with specific enzyme immobilization through fundamental quantum calculations.
Recent pioneering efforts have been made to exploit non-invasive blood glucose analysis using optical methods such as Raman,10 fluorescence,11 and near/mid-infrared range spectroscopy,12 or electrochemical methods such as voltammetric and amperometric detection,13–15 to analyze glucose in biofluids that are correlated with blood glucose concentration. Among all non-invasive blood glucose monitoring approaches, saliva has been shown to reflect blood glucose concentration and provide a simple, inexpensive, and fast measurement method.16
Multiple electronic glucose biosensors have been proposed for the selective detection of glucose in various biological fluids using carbon nanotubes (CNTs) due to their high electrical conductivity, corrosion resistance, and superior mechanical properties.14,17 However, the hydrophobicity and poor dispersibility of CNTs result in deficiencies in accuracy and bonding sites to enzymes (typically glucose oxidase, GOx), making CNT-based biosensors inappropriate for low glucose concentrations (particularly in saliva, which has a lower glucose concentration than blood).18
To realize a biosensor with high sensitivity and a wide-sensing range of glucose concentration, we demonstrate a salivary glucose biosensor based on ferrocene-chitosan (Fc-Chit) modified CNTs on a screen-printed electrode (GOx/Fc-Chit/CNTs/SPE) through a simple drop-casting process. Although the mechanism of enzymatically electric communication by Fc-Chit for biosensors is challenging due to the flexibility and diversification of chitosan, detailed investigations of optimizing Fc-Chit are scarce. Herein, we performed quantum chemical calculations and proposed dipole moment as an underlying mechanism to confirm our experimental data. The superior sensing sensitivity of 119.97 μA mM−1 cm−2 with high specificity and stability and human saliva glucose measurement demonstrate not only the feasibility of the non-invasive glucose biosensor but also a design strategy for other enzymatic biosensors.
FTIR spectroscopy was conducted to confirm the formation of Fc-Chit, as shown in Fig. S1.† The two characteristic peaks at 1642 and 1550 cm−1 can be attributed to the amide groups and the overlap of amide and amine groups on chitosan, respectively.24 In Fc-Chit, the intensity of the peak at 1550 cm−1 for primary amine N–H bending decreases, while a new peak is observed at 823 cm−1, which can be assigned to the cyclopentadiene of ferrocene.25 The result indicates that N-alkylation of the glucosamine unit on chitosan with ferrocenecarboxaldehyde occurs, resulting in the formation of Fc-Chit. To investigate the mechanism of ferrocene attachment to chitosan, X-ray photoelectron spectroscopy (XPS) was further conducted to analyze the electronic structures of chitosan and Fc-Chit. The additional signal of Fe 2p in the Fc-Chit spectrum confirms that ferrocene is successfully grafted onto chitosan (Fig. S2†). In Fig. 2c, the N 1s photoelectron spectrum of chitosan shows a dominant peak at 399.4 eV, attributed to the NH2 group.26 The lower intensity peak (atomic ratio of 23.5%) at around 401.0 eV for N–CO implies the presence of the acetylamine group, which is due to the incomplete deacetylation of chitin and is in agreement with the product report from Sigma-Aldrich (around 75% deacetylation).27 It is worth noting that after introducing ferrocene, the ratio of NH2 decreases from 73.5% to 25.1%. In contrast, a dominant peak at approximately 400.0 eV representing C–NH–C appears in Fc-Chit,28 indicating that the formation process consumes the amino groups of chitosan and produces secondary amines between the ferrocene molecule and chitosan.29 In addition, the increased intensity of protonated NH3+ groups at 401.5 eV indicates a proton transfer process from the covalent linking of C–NH–C to unreacted amino groups in Fc-Chit. Based on the above results, the aldehyde group of ferrocene can react with amine through the formation of Schiff bases, which are subsequently reduced to C–NH–C with the assistance of ethanol,30,31 as illustrated in Fig. 2b.
To ensure that each layer in GOx/Fc-Chit/CNTs/SPE was successively constructed using the facile drop-casting method, AFM and SEM were used to characterize the variation of surface morphologies. Fig. 2c and d display the distinct particles on bare SPE, which are attributed to the carbonized polypropylene. Upon drop-casting the CNT dispersion, the CNTs are uniformly distributed, thereby reducing the roughness of the electrode surface (Table 1) (the optimization of CNT layers is provided in Fig. S3†). In Fig. 2d, multiple nano-sized pores can be observed, which are constructed by entangled CNT networks. The nanoporous structure of CNTs/SPE not only provides spaces for the subsequent Fc-Chit to penetrate but also facilitates ion transfer for glucose sensing.32 As the Fc-Chit is introduced, the diameter of CNTs slightly increases while the pore size and number decrease (as depicted in Fig. S4†), indicating that CNTs are covered and wrapped by self-assembled Fc-Chit film. After the drop-casting of the enzyme, the surface of GOx/Fc-Chit/CNTs/SPE is entirely covered by GOx, and the morphology of CNTs can no longer be observed. As shown in Table 1, the decreased surface area can confirm that the gelatinous GOx is successfully immobilized on the electrode.
Electrode | Average roughness (Ra, nm) | Root-mean-square roughness (Rq, nm) | Surface area (μm2) | R s (Ω) | R ct (Ω) |
---|---|---|---|---|---|
SPE | 420 | 525 | 554 | 133.3 | 338.9 |
CNTs/SPE | 376 | 488 | 492 | 129.8 | 1.8 |
Fc-Chit/CNTs/SPE | 362 | 474 | 490 | 135.2 | 2.2 |
GOx/Fc-Chit/CNTs/SPE | 389 | 472 | 426 | 141.6 | 4.4 |
To evaluate the ion diffusion behavior and sensing response of the electrodes, EIS was conducted in K3Fe(CN)6 solution to reveal the ion and charge transport dynamics (Fig. 3a). The Nyquist diagrams of each electrode exhibit an incomplete semicircle in the high-frequency region, followed by a 45°-incline caused by Warburg diffusion impedance (W0).33 In the equivalent circuit model, the equivalent series resistance (Rs) represents the combination of the solution resistance and the internal electrode resistance, which can be reflected by the intercept at the real part of impedance (X-axis).34 The double-layer capacitance (CPE) is placed in parallel to a charge transfer resistance (Rct). Rct represents the interfacial resistance of the charge transfer between the electrolyte and electrode and can be described by the diameter of the semicircle. Here, both Rs and Rct decrease after the drop-casting of the CNT dispersion, and in particular, Rct decreases from 338.9 to 1.8 ohm (Table 1). This indicates that uniformly distributed conductive CNTs with numerous pores are beneficial for ion diffusion, hence effectively reducing the resistance of electron transport. In the subsequent modification using Fc-Chit and GOx, both Rs and Rct increase (as shown in Table 1). This can be attributed to the fact that the low-conductive Fc-Chit and GOx films tend to clog the nanopores, hence hindering the access of ions to the electrode.
Fig. 3b shows the CV curves of GOx/Fc-Chit/CNTs/SPE at various scan rates in the potential range of −0.30 to 0.60 V in PBS solution. The redox peak current increases linearly with the scan rate from 20 to 300 mV s−1. The linear relationship between the redox peak current and scan rate indicates a surface-controlled process, which can be described by the following equation:35
(1) |
GOx(FADH2) + O2 ⇌ GOx(FAD) + H2O2 | (2) |
To evaluate the impacts of each layer on the electrocatalytic redox of GOx, GOx was drop-cast on SPE with different modified layers for CV measurement in PBS solution. As shown in Fig. 3c, GOx/Fc-Chit/SPE displayed minimal response in the potential window ranging from −0.80 to 0.30 V. Conversely, GOx/CNTs/SPE exhibited apparent catalytic oxidation (−0.56 V) and reduction (−0.69 V) peaks. The shifting of the redox pair potentials is attributed to the overlap with the potential of the consumption of oxygen, which agrees with the results in previous works.39,40 These results indicate that for the electrode with Fc-Chit alone, the redox center of GOx is inadequate in electrical communication with the SPE surface due to the insufficient electrical conductivity, as confirmed by EIS (Table 1). In contrast, the presence of CNTs can effectively facilitate charge transfer in the redox reaction at the CNT-coated SPE. The redox peaks around −0.2 and 0 V, which only appeared in CV curves in the presence of CNTs, may result from the redox reaction of hydroxyl ions on the carbon surface, as described in other literature reports.41 Note that the highest redox peak currents and decreased peak-to-peak separation (ΔEp) of 0.12 V can be observed for GOx/Fc-Chit/CNTs/SPE, suggesting an efficient electrocatalytic process, however, which contradicts the higher Rct observed by EIS (Table 1). These results aroused our interest in the effect of Fc-Chit on the electrocatalytic activity of GOx. Although previous works have reported that ferrocene-branched chitosan can serve as a redox mediator and improve the attachment of GOx to the electrode,42,43 there is a lack of direct evidence to support this speculation. Therefore, an adherence test was performed by attaching 3M Scotch tape to the electrode surface and then peeling it off, as shown in Fig. 3d. Optical microscope images revealed that a significant amount of GOx on GOx/Fc-Chit/SPE and GOx/CNTs/SPE was transferred to the 3M tape. However, with Fc-Chit as the intermediate layer, almost no GOx was transferred from the electrode to the tape. The enhanced adhesion results from the strong electrostatic interactions between negatively charged acid-treated CNTs and positively charged Fc-Chit, allowing the latter to penetrate and cement the CNT layer (confirmed by Fig. 3d), thereby forming a robust film with numerous active sites to immobilize negative enzyme molecules for fluid glucose sensing.24
To optimize the sensing performance of GOx/Fc-Chit/CNTs/SPE, we evaluated its glucose-responsive capability with different ratios of ferrocene-to-chitosan. Herein, predetermined amounts of ferrocenecarboxaldehyde were grafted to chitosan at branching ratios of 50, 75, 100, and 125% w/w and drop-cast onto the SPE to measure the sensitivity to fluid glucose at different concentrations in PBS. As shown in Fig. 4a, the cathodic current response decreases with an increase in glucose concentration, consistent with our previous works.44 All GOx/Fc-Chit/CNTs/SPEs with different ferrocene-to-chitosan ratios exhibit a reliable linear detection range of glucose concentration from 20 to 800 μM (Fig. S6†), covering the salivary glucose levels from 94 to 456 μM for healthy individuals and patients with diabetes,45 respectively. Notably, GOx/Fc-Chit/CNTs/SPE with a 75% ferrocene-to-chitosan ratio (0.75Fc-Chit) exhibits the highest sensitivity of 119.97 μA mM−1 cm−2 with a correlation coefficient of 0.9941 (Fig. 4b), elucidating the feasibility of the proposed biosensor for a non-invasive glucose sensing and management technique.
To gain a better understanding of our experimental data on the sensing performance with various ferrocene concentrations on an atomic scale, we performed quantum chemical calculations. Specifically, we calculated the dipole moments of the optimized geometries of Fc-Chit with one, two, or three condensations of FcCOOH on the 75% deacetylated chitosan backbone (which was obtained from Sigma-Aldrich) via the amine groups (abbreviated to Fc1-Chit, Fc2-Chit, and Fc3-Chit, respectively). As shown in Fig. 4c, the 75% deacetylated chitosan backbone (referred to as the backbone) is composed of three chitosan and one chitin in the order of chitosan–chitin–chitosan–chitosan from C1 to C4. Additionally, we investigated three different regiomers for Fc1-Chit and Fc2-Chit, which we denoted as 1, 2, 3 and 4, 5, 6, respectively. The results of these calculations are shown in Table 2.
Molecule | 1 | 2 | 3 | 4 | 5 | 6 | 7 |
---|---|---|---|---|---|---|---|
Total dipole moment | 2.75 | 2.24 | 2.30 | 2.53 | 3.05 | 1.88 | 2.15 |
Dipole moment (x) | 0.24 | 0.42 | 0.19 | 0.19 | 0.41 | 0.58 | 0.70 |
Dipole moment (y) | 1.94 | 0.63 | −0.20 | 0.90 | 0.56 | 0.84 | 0.92 |
Dipole moment (z) | −1.94 | −2.11 | −2.29 | −2.36 | −2.97 | −1.59 | −1.82 |
Table 2 shows that among the Fc-Chit molecules investigated, Fc2-Chit regiomer 5 exhibits the largest dipole moment, followed by Fc1-Chit regiomer 1. This finding is consistent with our experimental observation that 0.75Fc-Chit, which contains approximately 2.5 ferrocenes attached to the backbone by mass, exhibits the highest glucose sensing sensitivity of 119.97 μA mM−1 cm−2. Interestingly, as the amount of ferrocene increases to 1.00Fc-Chit, which translates to 3.33 ferrocenes attached to the backbone (7), the glucose sensitivity decreases to 108.78 μA mM−1 cm−2, in line with the lower calculated dipole moment of 7 compared to 5. Initially, it was hypothesized that the sensing performance would improve with the number of ferrocenes attached to the backbone. However, the increased number of mobile electrons from ferrocene is counteracted by the reduced dipole moment, resulting in weaker binding to the acidic CNTs responsible for charge transfer to the electrode.
A closer look at the optimized structures of Fc-Chit regiomers 5 and 7 sheds light on the unexpected change in dipole moments upon increasing the number of ferrocenes attached to the backbone. Fig. 5c shows that the backbone in 5 retains the whole hydrogen bonding network as in 1, Fc1-Chit. However, upon the attachment of the third Fc at the third saccharide, as occurred in 7, the bulky Fc substituent forces the fourth saccharide to rotate away from its optimal conformation, removing all the hydrogen bonding networks between the third and fourth saccharide units (boxed regions in Fig. 5c). This rotation hinders the alignment of the partial negative charges of the backbone, as seen in the dipole moment in the z direction in Table 2, reducing the total dipole moment of 7 (from −2.97 a.u. in 5 to −1.82 a.u. in 7). This effect also transpires in 2, 4, and 6, which are regiomers of Fc1-Chit and Fc2-Chit that have the third saccharide attached to a CH2Fc unit (see the Density Functional Theory (DFT) calculation section in the ESI†). A larger effect is observed when both the third and fourth saccharides are functionalized (6 and 7), due to the steric clash between the Fc on the third saccharide and both the fourth saccharide backbone and its Fc substituent. According to the experimental and simulated results, while ferrocene can facilitate electrochemical communication between the immobilized GOx and the electrode surface, leading to more sensitive and stable biosensors, excessive ferrocene may obstruct the active sites of chitosan for the biocatalyst due to steric hindrance and prevent GOx from anchoring on the electrode surface.46 These findings not only provide a molecular-level understanding of the optimization of the proposed biosensor for glucose detection but also pave the way for designing other enzymatic biosensors through fundamental quantum calculations.
The pH-dependent behavior of the enzymatic redox current peaks is important to consider since enzymatic catalysis involves the transfer of protons and electrons,47,48 which may impact the sensitivity and accuracy of glucose measurements. Therefore, it is crucial to ensure that the biosensor is stable and reliable under various pH conditions. In Fig. 5a, as the pH value increases, the half-wave potential (E1/2) corresponding to FAD moves toward a more negative potential (the E1/2 is −0.66 V at pH 7.0 and is −0.78 at pH 9.0). In contrast, the cathodic peak potentials for lower pH values move toward less negative ones (the E1/2 is −0.47 V at pH 4.0). Furthermore, the enzymatic reduction peak potential is linearly proportional to the pH value in the pH range from 3 to 8 with the slope of −61.7 mV pH−1 (Fig. 5b), which is in accordance with the literature for a reversible redox process with two-proton/electron activity.49 Using the Nernst equation can further validate the actual number of electrons involved in the reduction of FAD:50
(3) |
To evaluate the applicability of GOx/Fc-Chit/CNTs/SPE in human salivary glucose sensing, actual glucose concentrations were measured in commercial artificial saliva. However, wide-ranging fluid viscosity is a vital issue in saliva samples, which can pose a challenge for practical glucose biosensor applications.51 CV curves show a lower cathodic peak current with a negative-shifted potential for artificial saliva compared to that for PBS solution (Fig. S7a†), indicating an inefficient electron transfer process. To address this, the artificial saliva samples were diluted with PBS to 75% to demonstrate the sensing performance, as shown in Fig. 5c and S7b.† After dilution, the ΔEp decreased to nearly 0.12 V, similar to that in the PBS solution, and exhibited a linear response in the range of 20–800 μM with high sensitivity of 108.21 μA mM−1 cm−2. To assess the specificity of GOx/Fc-Chit/CNTs/SPE for practical salivary glucose monitoring, various common interfering molecules, such as uric acid (UA), ascorbic acid (AA), lactic acid (LA), sucrose, and fructose were tested. The current difference of enzyme reduction peak was measured using 100 μM glucose with the same concentration of each interferent in PBS solution (Fig. 5d and S8†). Though a 15.2% increment of current difference can be observed for glucose solution with 100 μM sucrose, it was reported that the salivary sucrose clearance time could be completed within 20 min,52,53 indicating little influence on salivary glucose measurement by the fabricated biosensor. Almost no distinction can be observed in the enzyme reduction peak current in the presence and absence of other interferents (<5%), indicating that the immobilized GOx and glucose-specific electrocatalytic reaction allow for a highly specific and stable biosensor for fluid glucose.
The feasibility of using the GOx/Fc-Chit/CNTs/SPE for human salivary glucose monitoring was demonstrated by measuring dynamic changes in salivary glucose concentration relating to diet. A healthy individual's saliva sample was tested, and the glucose concentration was calculated using standard glucose calibration (Fig. S9†). The results indicate that GOx/Fc-Chit/CNTs/SPE can clearly reveal changes in salivary glucose concentration between postprandial and fasting conditions in the normal range.54 The storage stability of GOx/Fc-Chit/CNTs/SPE was investigated by measuring the average reduction peak current of 500 μM glucose after a predetermined number of days of storage in a vacuum at 4 °C. After being stored for three weeks, the GOx/Fc-Chit/CNTs/SPE exhibited good stability, with the relative peak current remaining at 95.0% compared to the first day (Fig. 5e), and excellent operational stability with a standard deviation lower than 2.0%. The GOx/Fc-Chit/CNTs/SPE shows better storage and operational stability than previous works using metal nanoparticles, polymerized ionic liquids, or metal oxide composites as biosensors.55–57 The improved stability may result from the electrostatic interactions between the positively charged Fc-Chit and negatively charged GOx and acid-treated CNT supporting layer, leading to a robust and rigid structure.
Moreover, the high biocompatibility of chitosan is reported to be an ideal enzyme fixative that can maintain protease activity during enzyme immobilization.58 The porous structure of the entangled CNT layer also allows for the enzyme to form a 3D enzyme structure, thus enhancing the tolerance of the fabricated biosensor to environmental changes. Table 3 summarizes the sensing performance of the selected state-of-the-art chitosan or ferrocene-modified glucose biosensors. While some biosensors possess a wider linear sensing range than GOx/Fc-Chit/CNTs/SPE, they require the use of metal/metal oxide components, leading to high costs and complicated fabrication processes. On the other hand, the proposed biosensor exhibits superior sensitivity that is one order of magnitude higher than most published results. Besides, this study showcases the integration of experimental and theoretical approaches to optimize biosensor design, in contrast to previous literature that relied mainly on data-driven evaluation through trial and error. These findings can lead to more efficient and effective biosensor development and advance practical applications in various fields.
Materials | Detection target | Linear range (mM) | Sensitivity (μA mM−1 cm−2) | LODa (μM) | Publishing year and reference |
---|---|---|---|---|---|
a LOD: limit of detection. | |||||
Fc-Chit/CNT@Cu/GCE | 0.1 M NaOH solution | 0.2–22 | 1.26 | 13.52 | 2020 (ref. 59) |
Fe3O4-chitosan-β-cyclodextrin/MWCNTs | Human serum | 0.04–1.04 | 23.59 | 19.30 | 2021 (ref. 60) |
MPC-CHT-GOx | Saliva, serum, and urine | 0.25–3 | 56.12 | 4.1 | 2019 (ref. 61) |
Chit/ZnO NPs/GCE | Grape juice | 10–900 | 41.53 | 0.9 | 2020 (ref. 62) |
AgNWs/CS/GOx | PBS solution | 1–15 | 16.72 | 2.1 | 2016 (ref. 63) |
CS-GO | Human dermal fibroblast cell | 0.01–20 | 0.17 | 1.0 | 2019 (ref. 64) |
CHIT(GOx)/AuLr-TiND | Saliva, serum, and sweat | 0.04–15.05 | 23.47 | 1.75 | 2021 (ref. 65) |
GOx/chitosan immobilized LIGE | PBS solution | 0–8 | 43.15 | 0.43 | 2021 (ref. 66) |
P(SNS-Fc-co-EDOT)/CNT | PBS solution | 0.01–1.0 | 11.58 | 1.63 | 2020 (ref. 67) |
GO x /Fc-Chit/CNTs/SPE | Saliva | 0.02–0.8 | 119.97 | 20.0 | This work |
Footnote |
† Electronic supplementary information (ESI) available: Electroactive surface area calculation, Density Functional Theory (DFT) calculations, human saliva test, and related characterizations. See DOI: https://doi.org/10.1039/d3na00340j |
This journal is © The Royal Society of Chemistry 2023 |