This review covers developments in ‘Atomic Spectrometry’. It covers atomic emission, absorption, fluorescence and mass spectrometry, but excludes material on speciation and coupled techniques which is included in a separate review. It should be read in conjunction with the other related reviews in the series.1–6 A critical approach to the selection of material has been adopted, with only novel developments in instrumentation, techniques and methodology being included. A number of themes have emerged from this year's review. Novel developments in sample preparation include the use of magnetic nanoparticles for solid phase extraction and an increase in the number of viable methods for use in quantitation of peptides, proteins and DNA by elemental tagging, particularly the development of multiplexed techniques. There were also some advances in electrochemical vapour generation, such as the use of reducing reagents immobilized on electrodes and development of electrolytic membrane diffusion cells. New instrumentation has been to the fore, with research into some novel and interesting miniaturized plasma sources being of particular interest, plus the development of a new distance of flight (DOF) mass spectrometer. There have been some insightful fundamental studies of particle atomization and ionization in ICP-MS, and subsequent space-charge effects after ion sampling, which should enable practical steps to solving some of these problems. One of the main drivers of research in isotope ratio analysis has been to increase detection power to enable smaller and smaller quantities of analyte to be measured, which yields advantages in geochronology and nuclear forensics. Laser ablation has played a large part in many types of analysis where high resolution is required, including analysis of single particles, small inclusions or elemental mapping of tissues, so the development of LA methods with extremely small spot sizes and efficient sample transport to the source has been key. In addition, several novel methods have been developed, such as laser ablation molecular isotopic spectrometry (LAMIS) for isotope ratio analysis, and stand-off LIBS techniques such as ‘underwater LIBS’. In contrast, very little research into the use of chemometrics has been reported, and fundamental studies are generally confined to characterization of new plasma sources.
Cui et al.10 have coated Fe3O4@Si–OH MNPs with the ionic liquid methyltrioctylammonium chloride (MTOA+ Cl−) for use in extracting Cd, Cu and Zn from biological samples. The target analytes were determined in the final eluate solution of 0.8 mL of 1.0 mol L−1 HNO3, with LODs of 0.33, 0.54 and 0.56 μg L−1, respectively. The method was found to be tolerant of concentrations of other transition and heavy elements such as Co, Cr, Fe, Mn, Ni and Pb up to between 1 and 4 mg L−1, but also much higher concentrations (ca. 1000×) of polyanions, alkaline and alkaline earth elements. The method was validated by analysis of the CRMs GBW09103 human urine and GBW07601 human hair, and spiked samples, with recoveries between 89 and 120%.
Selectivity in SPE has long been a goal of the analyst, and ion-imprinted polymers (IIPs) are a class of adsorbent that holds some promise. Ebrahimzadeh et al.11 have developed a novel IIP for Au preconcentration, by incorporating a dipyridyl ligand in ethylene glycol dimethacrylate, then coating it onto porous silica particles. They achieved an LOD of <0.2 ng mL−1 using FAAS and selectivity of preconcentration in the presence of palladium and platinum. Another novel IIP has been reported by de Lima et al.12 who formed the polymer by reduction of SeIV to Se0 prior to reaction with 4-vinylpyridine and vinylimidazole, hence they designated the product as an atom imprinted polymer (AIP). Preconcentration of Se was achieved in FI mode by first reducing the analyte to Se0, using 0.5% (w/v) citric acid, then preconcentration on the AIP which was immobilised in a mini-column. The Se0 was then eluted by oxidation to SeIV using hypochlorite, subsequently reduced to SeH2 using NaBH4 and analysed by HG-FAAS. The method was optimized using a Doehlert design to give an LOD of 53 ng L−1 and preconcentration factor of 232. Validation was by the determination of Se in DORM-2 dogfish muscle, with 98% recovery. Selenium was also successfully determined in Brazil nuts, apricot, white beans, rice flour and milk powder with recoveries in the range 100 to 109%. It is worth noting that significant matrix interferences (>10%) were observed in the presence of less than 10-fold excess of Cr3+ and Cu2+, so the authors recommended that the method of standard additions be used.
Bauer et al.13 reported a novel SPE method whereby trace metals were preconcentrated from environmental waters using silica NPs. The NPs were functionalised with strong cation exchange ligands based on 3-mercaptopropyltriethoxysilane modified with acetonitrile, trifluoroacetic anhydride and urea hydrogen peroxide. These particles had diameters in the range from 300 to 700 nm so could be suspended as a slurry and nebulized directly into ICP-OES without the need for elution of the analytes. The authors found that the analytical signal initially decreased by 10% 2.5 min after initial ultrasonic suspension of the slurry, but stayed constant thereafter up to 12.5 min. They attributed the initial decrease to settling out of larger particles. An average preconcentration factor of 10.4 was achieved, yielding LODs between 0.03 μg L−1 for Be to 0.48 μg L−1 for Fe, and the method was validated by the determination of Al, Be, Cd, Co, Cr, Cu, Fe, Mn, Ni and Pb in NIST1643e fresh water.
Ozdemir et al.14 have reviewed (160 references) the use of microbacteria as SPE media, with particular attention paid to preconcentration of metals using bacteria immobilised on resins.
Two of the key requirements for DLLME are the initial dispersion of the extracting solvent and subsequent centrifugation of the mixed phases in order to separate the highly dispersed organic droplets from the aqueous phase. This can also be achieved using an ionic liquid (IL) in place of an organic solvent. The role of ILs in extraction methods has been reviewed (113 references) by Han et al.,18 covering liquid and solid phase extraction utilising hollow fibres, single drop and dispersive methods for both organic and inorganic analytes.
There have been some novel variations on the standard procedure. Naeemullah et al.19 extracted a vanadium-8-HQ complex (0.1–1.0 mL of a 0.1% w/v solution of 8-HQ) from waters using 1-butyl-3-methylimidazolium hexafluorophosphate (20–70 mL of a 1:1 mixture of the IL and chloroform). In this case dispersion of the IL was aided by using chloroform in the mixture and high pressure argon (15 kPa for 5 min). The sample was centrifuged for 10 min at 3000 rpm and dissolved in 200 mL of 0.1 mol L−1 HNO3 in ethanol–water (1:1 v/v) before analysis using ETAAS, to yield an LOD and enrichment factor of 20 ng L−1 and 100, respectively. The use of a hollow fibre, in conjunction with room temperature ILs, has been reported by Zeng and Tang.20 They used butyl- octyl- or hexyl- 3-methylimidazoliumhexafluorophosphate (or tetraborate) ILs to aid complexation of Au with dithizone, with subsequent extraction into 9 μL of 1-octanol trapped inside a hollow fibre. This solution was then flushed out with a microsyringe, mixed with 65 μL of 95% (v/v) alcoholic solution with 1% (v/v) HNO3 and siphoned into a FAA spectrometer for analysis. They found that the use of the ILs improved extraction efficiency by five-fold to give an enrichment factor of 130 and LOD of 0.9 ng mL−1 for Au. Shah et al.21 performed a microextraction of Pb using an IL inside a 10 mL syringe. The pH buffered sample was placed in the syringe with 0.01% (w/v) dithizone, then 500 μL of 0.075% (v/v) Triton X-114 and 100 μL of 1-Butyl-3-methyl-imidazolium hexafluorophosphate added to extract a Pb–dithizone complex. Separation of the IL phase was effected by decanting from the syringe and diluting in 0.5 mL of methanol, before injecting 100 μL aliquots into the sample uptake tube of FAAS. The method yielded an enrichment factor of 82.6 and LOD of 0.281 μg L−1.
Normally, absolute quantitation of peptides and proteins by IDA requires the use of an isotopically enriched isotopomer of the target peptide, which will likely have to be synthesised for the purpose. This goes some way to explaining why the technique has not been widely adopted. Esteban-Fernandez et al.26 have used the metal coded affinity tag (MeCAT) approach for absolute quantitation of proteins using ID-ICP-MS. The MeCAt reagent was loaded with (separately) natural and isotopically enriched Yb then tagged to various proteins via an iodoacetamide (IA) cysteine-reactive group on the reagent to form a MeCAT(Yb)-IA-Protein complex. Proteins were separated by gel electrophoresis (GE) and analysed either directly by LA-ICP-MS or after excision and digestion of the gel spots. Absolute quantitation was performed by IDA after spiking with standard proteins labelled with enriched Yb. Crucially, both natural and enriched MeCAT(Yb)-IA-Protein complexes migrated equally during the GE separation, and a stable 172Yb:171Yb ratio was observed along the spot when using LA-ICP-MS. The same group27 used the same labelling strategy and compared this method with LC-ESI-MS, LC-ESI-MS-MS and HPLC-ICP-MS for the relative quantitation of alpha-lactalbumin, using Eu, Ho, Lu, Tb and Tm in a tetraplex labelling approach, and SDS-PAGE separation prior to analysis. They found that all three methods yielded quantitative results.
Liu et al.28 reported a non-species-specific IDA method, using LC-ICP-MS, for absolute quantitation of bradykinin. They used commercially available 2′,2′′-(10-(2-((2,5-dioxopyrrolidin-1-yl)oxy)-2-oxoethyl)-1,4,7,10-tetraazacyclododecane-1,4,7-triyl) tri-acetic acid (DOTA–NHS–ester) to react with bradykinin, then labelled the resultant conjugate by forming a bradykinin-DOTA-Eu chelate at pH6, by using a two-fold molar excess of Eu to DOTA. Non-species-specific IDA was achieved by post-column perfusion of a 151Eu spike solution (4.27 ng g−1) via a three-way connector. The method was validated by comparison with a species-specific GC-MS method, in which the bradykinin sample was spiked with 13C-labelled phenylalanine, then hydrolysed to yield free phenylalanine before derivatization with methyl chloroformate. The two methods agreed to within 4%, with 5% RSD at concentrations in the range from 20 to 500 pmol. An LOD of 7.3 fmol was estimated for the species-non-specific LC-ICP-MS method. Wang et al.29 developed a different approach based on labelling standard peptides with Tb to act as an internal standard. They also used a DOTA–NHS–ester to form peptide -DOTA conjugates with synthesised standard peptides and the equivalent endogenous target peptide, which were then chelated with either Tb or Tm, respectively. Analysis was performed by LC-MRM-MS using external calibration, but also by using the Tb-DOTA-peptide:Tm-DOTA-peptide ratios ascertained from the widely separated mass fragments in co-eluting peaks as an internal standardisation method. They achieved an LOD of 0.8 fmol μL−1, so the technique might well benefit from using ICP-MS as the detection method.
Yet another solution to this problem has been reported by Konz et al.30 who utilised the intrinsic ability of the hormone hepcidin-25 to complex with CuII ions. They used two approaches, both involving IDA and LC-ICP-MS: post-column addition of a non-species-specific 65Cu spike; and synthesis of a labelled 65Cu-hepcidin-25 complex to use as a species-specific spike. The stability of the Cu-hepcidin-25 complex was checked using ESI-MS, and found to be stable under physiological conditions and to exist as an equimolar complex, however, they did note that spiking with the 65Cu-hepcidin-25 complex should occur directly prior to species specific IDA because Cu-exchange reactions were possible; a problem that did not occur with post-column addition. They used both methods to determine hepcidin-25 in human serum samples and found an average concentration of 21.6 ng mL−1, which agreed well with those reported in the literature of 19.2 ng mL−1. In an unrelated but relevant paper Arnquist and Holcombe31 studied metal-protein binding equilibria between CuII ions and bovine serum albumin (BSA) using preparative ultracentrifugation to separate bound and unbound Cu, with ICP-MS detection. In this way they determined the apparent (Kapp) and intrinsic (Kint) binding affinities under different conditions of pH; information which could be used to predict the type of exchange reactions reported by Konz et al.30
Absolute DNA quantitation by elemental tagging poses a similar challenge to protein quantitation. Han et al.32 have described a novel and elegant multiplexing approach to this problem. They utilised 15 DNA hybridisation probes for separate DNA targets, of 25–30 bases length, related to clinical diseases. The probes were 3′ end-functionalized with thiol, derivatized with maleimide-DOTA (MMA-DOTA), then individually labelled by complexation with a specific REE. A mixture of these DNA-MMA-DOTA-REE complexes was then added to the target DNA sequences to allow hybridisation to occur, and these were then captured by further hybridisation with capture probes which had been immobilised on MNPs. The MNPs with the attached probe-target-capture sequences were separated from the mixture containing the excess REE labelled probes. Finally, the REE labelled probes were released from the target sequences by raising the temperature to 95 °C and analysed by ICP-MS using external calibration with standard REE labelled probes. They further developed the method to allow an IDA approach to absolute quantification by using a spike-DNA-MMA-DOTA-REE probe complexed with an enriched isotope; in this case separate 161Dy-enriched and 168Er-enriched probes were tested. The method was as before, except that the MNP-immobilised capture probes contained an extra sequence which was complementary to the spike-DNA-MMA-DOTA-REE probe, and the quantitation step was performed by ID-ICP-MS. The method was applied to the analysis of B. anthracis and severe acute respiratory syndrome (SARS) targets and good agreement with a UV-vis method was obtained.
An alternative approach to DNA quantitation was reported by Leclerc et al.33 who used ICP-MS (MC and Q) to quantify the total amount of DNA extracted from cells of Legionella pneumophila by determining the mass fraction of intrinsic P. They compared the results of the ICP-MS analysis with UV-vis and quantitative polymerase chain reaction (qPCR) methods, and found good agreement between the three. The ICP-MS method could be used with about 2 mL of DNA solution at about 30 ng g−1 P per measurement and had 11× lower uncertainty compared with qPCR because of the uncertainty in the qPCR standards. Swart34 has addressed the important metrological difficulties of protein quantitation. The main problem seems to be that there are few SI-traceable reference materials for metallo-proteins, which is a serious lack when comparable results for samples of human origin are required. The author discussed the problems inherent in quantitative analysis and the production of traceable standards for this class of compounds using the examples of hemoglobin, transferrin, superoxide dismutase, ceruloplasmin, and C-reactive protein.
Determination of metals in volume limited biological samples is critical to metallomics research and microfluidic chips provide an excellent platform for the miniaturization of extraction. In this context, Wang et al.36 designed and fabricated a chip-based LPME device to combine with ETV-ICP-MS for the determination of trace Bi, Cd, Cu, Hg, Pb and Zn in cell and human serum samples. Using sodium DDTC as the chemical modifier for low-temperature ETV-ICP-MS and the extracting reagent for the chip-based LPME, the chip parameters influencing the extraction efficiency were optimized by orthogonal design experiments. Under optimized conditions, LODs were in the range from 6.6 to 89.3 pg mL−1 with the RSDs ranging from 4.23 to 8.15% (n = 7, c = 1 ng mL−1) for the target metal ions. The linear range covered three orders of magnitude with correlation coefficients higher than 0.99. The method was validated by analysis of CRMs GBW07605 human hair and GBW09152 human serum. Metal determination in HepG2 cells, Jurkat T cells and a human serum sample was achieved with the recoveries between 86.6 and 119%. This rapid and sensitive method has low sample and reagent consumption and offers potential for the analysis of quantity-limited biological samples such as cells.
The analysis of engineered nanoparticles continues to be a subject of some interest, requiring the development of new methods for characterisation of size, mass, elemental composition, and number concentration. Elzey et al.37 developed a sample introduction system to allow the separation and analysis of gold nanoparticles. An ES-differential mobility analyser (ES-DMA) was used to achieve real-time upstream size discrimination. A condensation particle counter and ICP-MS were employed as downstream detectors, providing information on number density and elemental composition, respectively, of aerosolized nanoparticles versus the upstream size selected by ES-DMA. A gas-exchange device was designed and optimized to improve the conversion of air flow (from the electrospray) to argon flow required to sustain the ICP-MS plasma, the key compatibility issue for instrumental hyphenation. This initial work demonstrates successful measurement of the key analytical factors for nanoparticle characterisation.
A similar approach, to determine the size and mass of metallic engineered nanoparticles, was reported by Gschind et al.38 using a monodisperse microdroplet generator (MDG) for sample introduction into ICP-MS. Vertical and horizontal sample introduction configurations were evaluated. The horizontal introduction system resulted in shorter measurement time by reducing temporal jitter. Quantitation of Ag, Au and CeO2 nanoparticles of different sizes and polydispersities was achieved. Results were compared to those obtained using TEM and asymmetric flow FFF, and the advantages and limitations of the proposed technique were discussed.
Finally, Pereira et al.39 have developed a multiple nebulizer based on flow blurring hydrodynamics and used it to reduce matrix effects on As and Se determination by ICP-OES. This so-called flow blurring multiple nebulizer (FBMN) is essentially an array of three pneumatic nebulizers arranged in close proximity to each other which produces droplets <53 μm diameter (Sauter mean diameter = 16.3 μm) when operated at gas and liquid flow rates of 0.6 L min−1 and 0.3 mL, respectively. They studied matrix effects caused by Ca, K, Mg and Na when using the FBMN (with and without internal standardisation using In, Ge or Y) and compared the results with a standard pneumatic nebulizer. It is not clear what advantage the FBMN nebulizer affords, given that internal standardisation is an established approach to matrix correction and the propensity for the authors to present key comparative findings in the electronic supplementary information rather than the main body of the paper.
Yilmaz et al.43 have investigated the efficacy of using K3[Mn(CN)6] as an additive for HG of Pb reduced to plumbane using NaBH4. The point being that this reagent should enhance the reaction in the same way that K3[Fe(CN)6] has been shown to do by other workers. They did indeed find that a 1% v/v solution improved sensitivity, when coupled with ICP-MS, by as much as 48-fold. The method did suffer from severe signal suppression in the presence of Cu at a concentration of 0.1 μg mL−1. This was eliminated by increasing the K3[Mn(CN)6] concentration to 1%, but does not bode well for the analysis of real samples. Kratzer et al.44 used radioactive 212Pb to investigate the HG and trapping efficiency of plumbane using a similar method which also utilised NaBH4/K3[Fe(CN)6] as an enhancing reagent. They prepared the 212Pb tracer by precipitation of PbS from Th(NO3)4 with subsequent purification by ion exchange separation to yield 212PbCl2. The efficiency of plumbane generation, trapping and subsequent volatilization was determined by measuring the activity of gamma radiation from the 212Pb tracer with a scintillation detector. The HG apparatus was made in-house from a reagent mixing manifold and gas–liquid separator, which was interfaced with an electrically heated quartz trap or charcoal column depending which part of the process was to be investigated. Results showed that the efficiency of HG was ca. 96%; trapping efficiency was 88% at 300 °C; and volatilization efficiency was 85% at 800 °C. They also found that the presence of Bi at a concentration of 400 ng mL−1 reduced the volatilization efficiency to 13%.
A novel method of CVG has been developed by Zhu et al.,45 who used a cylindrical dielectric barrier discharge (DBD) plasma reactor made from two concentric glass tubes with a copper wire in the centre, and the outer tube wrapped with copper foil to act as the other electrode. The sample solution was run down the outside surface of the inner tube to form a thin film, from which hydrides of Cd were generated by reaction with hydrogen, produced after instigation of an 18 W DBD between the electrodes. Detection was performed using FI-QF-AFS to yield LODs of 0.03 ng mL−1 and 0.008 ng mL−1, using DBDs generated with argon and helium, respectively. No interference from first row TMs at concentrations of 1 mg L−1 was observed, but signal suppression down to as low as 40% (for Ni) occurred at 10 mg L−1.
Ai et al.46 used Fe3O4 MNPs for the oxidation of Hg species prior to CVG. The system involved mixing an HPLC eluent with H2O2 prior to a column containing the MNPs, in which organo-Hg species were oxidised by OH˙ free radicals formed as a result of the catalytic breakdown of H2O2. Subsequent CVG using KBH4 and AFS detection yielded LODs of 0.7, 1.1, 0.8 and 0.9 mg L−1 (as Hg) for Hg2+, MeHg, EtHg and PhHg, respectively. The method was validated by determination of MeHg in CRM DORM-2 dogfish muscle, with 95% recovery.
Hranicek et al.48 constructed and evaluated fifteen different electrolytic cells of varying types (non-membrane, thin-layer and tubular) for electrochemical vapour generation (EVG) of H2Se and detection by QF-AAS. In all of the cells lead and platinum were used as cathode and anode, respectively, and a Nafion membrane was used to separate the chambers. After optimization, the EVG efficiency, and spatial distribution of the analyte, for the five best-performing cells was evaluated using 75Se as a radioactive tracer. In this way it was possible to use autoradiography to determine the distribution of 75Se-containing compounds in the various parts of the cells by means of an activity budget. The LODs were sub-ng mL−1 for all five cells, approaching the performance of CHG (using NaBH4) which was used as a benchmark. The tubular cell, with an LOD of 0.4 ng L−1, was deemed the best-performing cell for FI because it had the highest efficiency (78%), lowest cathode volume (43 mm3) and fastest response time. The thin layer electrolytic cell was performed similarly, with cathode volume, LOD and efficiency of 225 mm3, 0.31 ng mL−1 and 75%, respectively, but did not have such a rapid response.
An EVG cell comprising a Ti anode, separated by a cation exchange membrane between two Pb cathodes, was designed and tested by Shen-Tu et al.49 Using an electrolysis current of 3.5 A, and with anolyte (0.4 mol L−1 H2SO4) and catholyte (0.3 mol L−1 HCl) flows of 4.0 and 1.5 mL min−1, respectively, they were able to generate and determine AsH3 and SbH3 with LODs of 0.65 and 0.9 μg L−1, respectively, using AFS. Various types of cation exchange membrane were tested, including polyethylene and Nafion 324. The authors found that longevity was greatly increased to several months continuous use by including porous epoxy resin spacers either side of the membrane. The method was validated by analysis of the CRMs GBW07405 and GBW07406 soil, with recoveries of between 96 and 107%.
Sturgeon and Luong51 developed two PVG reactors constructed from arrays of UV LEDs with either 0.3 mW and 245–260 nm output, or 0.3 mW and 360 nm output. They compared the results of PVG of inorganic and methylmercury using formic acid as the reducing medium, with those obtained using a thermal reactor at 85 °C. No reduction of the species was observed using the 360 nm LEDS but the 245–260 LEDS were effective. An LOD of 0.68 ng was obtained using the 245–260 nm LEDs using a 9 min irradiation time. Analysis of PACS-2 marine sediment and DOLT-4 fish liver tissue using the method of standard additions resulted in recoveries of almost 100%.
Matusiewicz and Slachinski52 combined an ultrasonic nebulizer and UV photolysis spray chamber by inserting an 8 W, UV pen lamp directly into a cyclonic spray chamber. This enhanced the production of As, Bi, Sb, Se and Sn hydrides, Hg vapour and volatile Cd species by PVG to yield LODs of 11, 22, 16, 19, 20, 15 and 11 μg L−1, respectively. Acetic acid (40% v/v) was used as the generation medium and detection was performed by MIP-AES. The main advantages were the ability to use a very small (<30 μL) sample size by ultrasonic nebulization and the relative lack of reducing agents.
Depth profiling with pulsed GD is a promising technique that has been the subject of investigation for a number of years. The use of a pulsed voltage for sputtering reduces the sputtering rate and thermal stress, and thereby improves the analysis of thin layered and thermally fragile samples. Pulsed glow discharge has not been widely studied, however, and this has limited its practical use. Efimova et al.55 investigated the factors that must be optimized when pulsed mode is used, i.e., duty cycle, frequency and pulse length, in order to obtain the best sputtering and crater shape. Pulse length was found to be a crucial parameter for the crater shape and thermal effects. They compared pulsed DC and RF modes and found that they produced similar sputtering effects.
Combinatorial and high-throughput techniques have become well-established in materials science to aid in the accelerated discovery of novel functional materials. Thin film composition spreads are frequently used for this purpose by performing localized measurements and exploring different properties throughout the phase diagram. The thin film composition map has to be determined in order to understand the behaviour of the compound or alloy of interest. There are several established techniques that can give elemental composition maps of large areas but the total mapping time can range from several hours to tens of hours or more. Thus, faster elemental mapping techniques are needed. In a study by Gamez et al.,56 the applicability of GD-AES elemental mapping for combinatorial and high-throughput screening samples was explored by characterizing a CuNi thin film composition spread. Qualitative analysis images showed that Cu and Ni composition gradients can be obtained in a very short time period. In addition, the use of reference materials allowed quantitative elemental analysis maps to be extracted from the emission intensity images. The images of emission intensities and curve fitting parameters are discussed. Finally, the GD-AES results are compared to profiles obtained by EDXRF. The GD-AES figures-of-merit relevant to elemental mapping were compared with selected alternative techniques from the literature to put their potential impact into perspective.
An approach to the analysis of non-conducting materials using GD-MS was suggested by Qian et al.57 A thin coating of molten indium was placed on the sample surface allowing the sample to be sputtered and ionized. After optimization of the plasma conditions, the prepared samples were analyzed directly by DC-GD-MS with satisfactory detection limits, stability and reproducibility. Two NIST SRMs were used to validate the precision, accuracy and reproducibility of this method. The results showed that this method could avoid grinding, sustain the relatively stable discharge and enhance the stability of signals compared to the method of using a cathode made from an In pin rolled in the sample powder.
A method has been developed for the determination of Ba, Cu, Ni, Pb, Sr and Zn in solid geological and environmental samples by LA-ICP-MS using a spinning sample platform.60 The platform, containing a sample and a standard, is spun during the ablation, allowing the quasi-simultaneous ablation of both materials. The aerosols resulting from the ablation of sample and standard were mixed in the ablation cell allowing quantitation of the analytes by standard additions. The proportion of standard versus sample could be increased by performing the ablation further from the axis of rotation. The ablated masses were determined using a strategy based on IDA. This spinning LA method has been applied to the Allende meteorite and four powdered CRMs fused in lithium borate glasses. SRM 612 trace elements in glass was also analyzed. Deviations from the certified values were found to be less than 15% for most of the elements and samples studied, with an average precision of 10%. The results demonstrated the validity of the method for the direct analysis of solid samples using a single standard.
The advantages and possibilities of using LA-ICP-MS for IR measurements were discussed by Urgast et al.61 A comparison of quadrupole, sector-field and multi-collector systems revealed the limitations of the respective set-ups. The report focused on the possibility of using LA-ICP-MS to image isotope ratios for biological tracer studies. The authors concluded that this application has significant potential in this important field.
Isotope ratio measurements using LA-ICP-MS in air under atmospheric pressure has been performed.62 Samples of brass and two minerals (galena and zircon) were ablated in air and the laser-generated aerosol aspirated into a gas exchange device, where the air was replaced with argon and transported into an MC-ICP-MS instrument. Results were compared to those obtained using conventional LA-MC-ICP-MS in helium in a sealed cell. Comparable in-run (0.03 to 1.61%) and external (0.002 to 0.254%) precision was obtained, and the accuracy of both methods was comparable. However, the atmospheric sampling method gave up to 5-fold lower intensities and visualization of the aerosol extraction indicated that some material was lost prior to gas exchange. Nevertheless, this method is suitable for isotopic determination of bulk materials for samples which are too large to fit into an ablation cell or too valuable to be cut into smaller pieces.
The quantitation capabilities of LA-ICP-MS are known to be restricted by elemental fractionation as a result of ablation, transport, and atomization/ionization effects, which may result in inaccuracies whenever non-matrix matched standard materials are used for calibration. Flamigni et al.63 studied the vapourization processes of laser-produced aerosols and subsequent diffusion losses occurring inside the ICP. Both ICP-OES and ICP-MS detection was used. Na- and Ca-specific AES of aerosols produced by LA of silicate glasses or metals revealed variations in the onset and maximum position of atomic emission in the axial channel, which were in the range of a few mm. These shifts were found to arise from composition-dependent particle/aerosol penetration depths which determined the starting point of axial vapourization in the ICP. This was also found to influence ion extraction efficiency through the ICP-MS sampling orifice, with differences observed for oxide vs. metallic aerosols caused by differences in the rate of diffusion-driven expansion from the central channel. Analysis of individual particles by ICP-MS resulted in 1/e half-value signal durations of approximately 100 μs, which agree with modelled values if AES maxima are assumed to coincide with positions of instantaneous vapourization and starting points for atomic diffusion. Results suggested that differences of up to 15% in the relative ion extraction efficiency could be explained depending on whether analytes were metals or oxides. Implications of these findings on the accuracy achievable by state-of-the-art LA-ICP-MS systems were discussed.
Plasma sources for MS and OES continue to be the focus of much development. Jankowski and Reszke66 have published a tutorial review (133 references) of microwave plasma sources for OES and MS. They cover the generation and classification of the different types of microwave plasmas, recent advances in construction and their applications in selected fields, particularly as OES detectors for GC. They also focus on the development of mini-discharges for miniaturized spectroscopic systems and developments in elemental and molecular ionization. The review is not comprehensive and tends to concentrate on recent developments in selected laboratories.
Miniaturized plasma sources have attracted much research interest because of the desire to place as much analytical capability as possible onto a ‘chip’. Luo and Duan67 have reviewed this area (70 references), focusing on the micro-fabrication of various types of microplasma discharges used for atomic and molecular analysis by OES and MS. Most of the applications are for gaseous analysis but they also cover solids and liquids to a lesser extent, with the attendant problems of sample introduction to the fore. Frentiu et al.68 have developed a miniaturized system comprised of a Hg CVG device with a gold filament preconcentrator, and a CCP microtorch (20 W power and 200 mL min−1 argon gas flow rate), interfaced with a microspectrometer. The preconcentrator was made from an 80 μm i.d. gold filament wound into a 43-turn spiral placed inside a quartz capillary, such that preconcentrated Hg could be rapidly desorbed to achieve maximum signal after 3 s. The system was used to determine Hg in waters and polymer leachates in the concentration range 0.06 to 57.4 ng L−1, with an LOD of 0.02 ng L−1 (0.5 pg absolute). Liu et al.69 constructed a low power (2 W) microplasma made from two platinum electrodes inside an alumina chamber of 150 nL volume, for determination of P and S in gas by AES. The argon plasma was operated at atmospheric pressure, 280 V (DC) and 8 mA, and excited emission at 253.6 and 469.4 nm to yield LODs of 68 ppb and 1.0 ppm for P and S, respectively, when using a palm-sized spectrometer for detection. A liquid electrode plasma (LEP) has been characterised for use in AES by Kohara et al.70 The plasma was generated directly in solution between two platinum electrodes, by applying a pulsed (3 ms) voltage of up to 2500 V in a narrow channel (100 μm) etched in fused silica. They observed the plasma and Pb emission at 405.78 nm using an array of detection equipment, including a high speed camera, CCD interface with a microscope, and a CCD spectrometer with a fibre optic interface. They estimated Te ∼ 10000 K (using Wien's displacement), Texc = 8150 K (using a Boltzmann plot of Cu emission lines) and ne = 1.4 × 1015 cm−3 (using Stark broadening of the Hβ line). Time resolved monitoring of the emission revealed a double peak over a period of several ms. Spatial imaging showed that maximum emission occurred at the centre of the expanding plasma bubble in the confines of the 100 μm channel, but also at the anode side bubble–liquid interface where the background was much lower. An LOD of 4.0 μg L−1 was obtained for Pb. The same group71 then went on to integrate the LEP onto a PDMS chip and obtained an LOD of 0.27 ppm for Pb. The PDMS chip could be used for ‘tens’ of measurements before degradation reduced performance.
New sources for MS and OES are constantly under development. Pfeifer et al.72 developed a low consumption argon ICP-MS. The design included an air-cooled, bulb-shaped quartz torch which fitted inside a standard rf coil, and up against a modified sampling cone with an extended tip. This arrangement allowed the torch to be operated at an argon flow of 1.5 L min−1 and ions to be sampled from inside the torch, but prevented entrainment of the cooling air. The modular design allowed a variety of sample introduction systems to be interfaced, including solution nebulization, HPLC and GC, with LODs between 0.002 and 0.046 μg L−1 for Al, As, Be, Cd, Co, Cr, Hg, Mn, Ni, Pb, V and Zn for solution nebulization, compared to LODs which were between 2 and 10 times lower in conventional ICP-MS mode.
A dielectric barrier discharge (DBD) was investigated by Muller et al.73 as a possible soft ionization source. The DBD was generated in a 30 mm long glass capillary at a helium gas flow of 600 mL min−1, and from 3.5 to 6.5 kV at 21.5 kHz. The discharge was examined spectroscopically in various discharge modes to determine the populations of N2+ and N2 as indicators of the suitability of the plasma for soft ionization. They concluded that soft ionization should be favoured by operation in homogeneous mode (below 4 kV) which had the highest production of N2 metastable atoms. Another type of DBD, developed by Liu et al.,74 was used as a desorption/atomization source for AFS. The argon plasma (10 kHz, 2.5 kV, 100 mL min−1) was generated between two copper rings in a quartz capillary, and used to desorb and atomize Hg species from a TLC plate. The analytes were subsequently determined by QF-AFS with LODs of 0.51, 0.29 and 0.34 pg for Hg2+, MeHg and PhHg, respectively.
The solution-cathode GD (SC-GD) source is attractive because it can be used for the direct analysis of a liquid. Progress in this area has been reviewed (55 references) by Jamroz et al.,75 with emphasis on the design and analytical figures of merit of different types of system over the last 20 years. The review contains a table of LODs arranged alphabetically by analyte for those wishing for a quick lookup guide. The mechanism of solution-to-plasma transport and excitation in this type of source was investigated by Schwartz et al.76 by high speed imaging and laser scattering. They found that the surface of the solution deformed on plasma ignition and a number of high powered plasma nodes formed across it at timescales of 100 ns to 100 μs, which were the source of most of the analyte emission. They also concluded that these nodes only occurred at low pH which also corresponded with an increase in sample transport into the plasma. Zhu et al.77 have used a SC-GD in a novel way, as a method of generating OsO4 vapour for introduction into an ICP-OES instrument. An LOD of 0.51 ng mL−1 was obtained without the need for oxidising reagents, though there was the requirement to prepare the samples in approx 0.1 M acid solution (i.e., pH = 1). Xiao et al.78 improved the performance of an a.c. SC-GD by adding organic compounds to the electrolyte. They found that adding 3% (v/v) formic acid enhanced emission by 13-, 17- and 78-fold for Ag, Cd and Pb, respectively, to yield LODs of 1, 17 and 45 μg L−1.
There have also been new developments in spectrometers. Gundlach-Graham et al.79 coupled an ICP with a distance of flight (DOF) MS. As the name suggests, this is similar to TOF-MS in that ions must be pulsed into the spectrometer however, m/z separation was effected by using a spatially selective, commercially available microchannel plate (MCP)/phosphor detector. The MCP detector elements were stacked in a chevron arrangement and ions were selectively deflected into the plane of the detector. The instrumental setup allowed for a high-to-low mass range of 1.11:1 with m/Δm resolution (FWHM) of between 440 and 670 with an ICP source. LODs of between 2 and 30 ng L−1, LDRs of 105 and IR precision of 0.1% RSD were obtained. One of the main advantages of DOF-MS is that low-cost, commercial detectors are available which are sufficiently large (100 mm) to allow a high-to-low mass range of 1.45:1, sufficient for simultaneous detection of the lanthanide elements.
Bhatia et al.80 have developed novel zoom optics which allowed Faraday detectors with wide apertures to be fixed in place and dispersion varied using a DC quadrupole and deflection optics. This arrangement resulted in flat-topped peaks for 87Sr:86Sr IR measurement with precision of 8 ppm. A new type of ion funnel has been designed by Egorova et al.81 to replace the conventional second stage, ion optics in an ICP-MS instrument. The funnel was made from 54 ring electrodes each of 0.5 mm thickness with 0.5 mm spacing between. The electrode apertures were circular with continually decreasing inner diameters, so that the stacked arrangement formed a funnel with a 35 mm i.d. at the entrance and 2 mm i.d. at the exit. Radiofrequency and DC potentials could be applied to the electrodes and adjusted to optimize the transmission of ions of different m/z. Ion transmission was found to be lower by one order of magnitude compared with conventional ion optics, but it was possible to introduce gases such as N2 and H2 to reduce the abundance of Ar-based polyatomic ions by one and three orders of magnitude, respectively.
The use of mixed gas plasmas, or gases other than argon, in ICP-OES has been around since the first reports of the technique in the 1960s, whether for economic reasons or to modify the temperature of the plasma in some way. Scheffler and Pozebon88 found that addition of nitrogen (20 mL min−1) to the nebulizer gas flow of an axially viewed argon ICP improved its robustness by a factor of two, as evaluated by the Mg(II) 280.270 nm:Mg(I) 285.213 nm ratio. They studied matrix effects, caused by 5 g L−1 of Na or K, on ten spectral lines with excitation energies between 11.5 and 15.5 eV and found that they were generally reduced by the addition of nitrogen without degrading the LOD or linear range. In some cases (for high excitation energy) an enhancement was observed, and the matrix effect of Ca was more variable. Any number of theories can be proposed for these observations but it is generally the case that changes to the size and shape of the central channel must be considered in addition to atomization and excitation mechanisms. This influence of plasma shape was well illustrated by Miyahara et al.89 who studied the effect of nitrous oxide addition to an air ICP. Addition of 40% nitrous oxide increased the plasma volume by two-fold, but atom and ion line intensities increased, and they observed a decrease in Trot from 3200 to 2700 K when nitrous oxide was increased from 0 to 100%.
Similar in some ways to the use of mixed gases, Grindlay et al.90 reported on the effect of carbon (in the form of glycerol, citric acid, potassium citrate and ammonium carbonate) on the signal for hard-to-ionize elements: As, Au, B, Be, Cd, Hg, I, Ir, Os, P, Pd, Pt, S, Sb, Se, Te and Zn in ICP-MS. They found that some elements showed an enhancement while others were unaffected, and put this down to a combination of changes in the ion distribution in the plasma and charge transfer reactions between some of the enhanced elements and carbon-containing species in the plasma.
Single particle analysis has attracted much interest in recent years. Chan et al.91 have studied the effect of single aerosol droplets on plasma impedance by using two methods: an RF-probe coil placed inside the torch; and by tapping the phase-detector signal of the impedance matching network. They found that there was a direct correlation between temporal spikes in the impedance and Hα atomic emission, which was linearly proportional to the volume of the introduced droplet. The measurements were made over short (ca. 10 ms per pulse) timescales and the method was proposed as a trigger for single droplet introduction into the ICP with appropriate tuning of the threshold value. In addition, the authors estimated that the energy change coupled to the plasma caused by a 50 μm droplet was between 8 and 11 W (0.03 to 0.04 J).
Niemax92 published an overview of the single droplet approach to diagnostic studies in ICP-MS, with recommendations on how to improve detection efficiency. He emphasised the spatial nature of analyte atomization, ionization and diffusion processes in the plasma, and how this interacts with the position and size of the sampling orifice. He recommended the use of a narrow injector and high gas flow rate to shift the point of atomization downstream so as to optimize the position where temperature is high and analyte ions have not had much time to diffuse away from the particle origin. However, he pointed out that optimization is a complex business that also depends on analyte mass (heavier elements diffuse more slowly) and other operating conditions. In a related report, Flamigni et al.63 studied aerosols and single particles produced by LA in the context of atomization, ionization and diffusion effects on ion sampling in ICP-MS. They used LA to generate aerosols containing Na2O and CaO from NIST610 silicate glass, and elemental Na and Ca from their metals, and observed that maximum emission for the oxides occurred further along the central channel than the elemental forms, ca. +3 mm for Ca and +1.5 mm for Na. This spatial shift in vapourization resulted in a discrepancy between the relative extraction efficiencies of metallic vs. oxide aerosols of 25% for Ca/CaO to 97% in case of Na/Na2O. Modelling of penetration depth revealed that this was more pronounced for particles larger than 1 μm. The main conclusion of this interesting study was that the inherent difficulty of calibrating LA-ICP-MS without matrix matched standards will require an approach which addresses this spatial difference in vapourization of particles. Some insight into these effects might be gained by studies such as that of Aghaei et al.,93 who developed a computational fluid dynamics model to investigate the effect of varying the pressure behind the sampler, gas flow rates and applied power in an ICP-MS instrument. This allowed them to model gas pathways through the sampler cone, concluding that it is possible to control whether only the injector gas flow passes through the orifice, or if it is accompanied by some auxiliary gas flow. In a further paper94 they used this to predict the effect of sampling cone position and orifice diameter on plasma temperature, electron number density, gas density and gas flow velocity; and the implications for extraction of ions formed from vapourizing particles travelling through the plasma. They conclude that a large sampler orifice favours extraction of both injector and auxiliary gas, and consequently large ion clouds, while a smaller orifice favours extraction of small ion clouds.
As if spatial heterogeneity in the ICP itself was not enough to contend with, there is also the time-worn problem of space charge effects in the post-sampler ion beam. Bergeson et al.95 used LIF to study space charge effects on Ca, caused by the matrix species Cs, Mg, and Pb, in the second vacuum stage of a commercially available ICP-MS instrument with right-angle ion-reflecting geometry. They used a CW-Ti:Sapphire laser, frequency doubled to 393.4 nm, to excite the Ca(II) 4s2S1/2 → 4p2P3/2 transition, and monitored the subsequent fluorescence at 854.2 nm from the ion beam at the entrance aperture to the third vacuum stage using a CCD detector. Interestingly, they found that the Pb matrix (1.5 mM) caused an off-centre deflection in the Ca ion beam rather than the more commonly assumed defocusing effect. They concluded that this could be corrected for by on-the-fly adjustment of the ion optics, with the caveat that this effect may well be specific to the type of ion optics studied.
Houk's group96 have continued their studies of metal argide (MAr+) polyatomic ions of Co, Cr, Cu, Fe, Mn, Ni, Ti, V and Zn in the ion extraction interface of an ICP-MS instrument. They compared experimental measurements of MAr+/M+ ratios with theoretical ratios calculated from the expected thermodynamic equilibrium at Tgas = 6000 K, and found that the experimental ratios were far lower than the calculated ratios, with the weakest bound species exhibiting the largest discrepancy. Increasing the ion kinetic energies by increasing the plasma potential, via a homemade guard electrode between the RF coil and torch, was observed to further reduce the experimental ratios, and adjusting the skimmer cone geometry also had an effect. Taking this evidence in total they postulated that the low experimental ratios were caused by collision induced dissociation (CID) due to a shock wave at the tip of, or in the throat of, the skimmer cone, thereby yielding far fewer MAr+ ions than expected. Bouchoux et al.97 published a theoretical investigation into interferences on Se in ICP-Ms. They estimated heat of formation, ionization energy (IE), and proton affinity (PA) for Se, SeH, SeH2, ArH, ArH2, Ar2, Ar2H, Ar2H2, ArCl and ArHCl using quantum mechanical methods. Three composite methods (G3, G4, and W1) were used with accuracy better than 5 kJ mol−1 for ΔH°f, PA, and IE. They also used CCSD(T) calculations using the aug-cc-pVTZ basis set for the neutral non-covalent complexes which were thought to mainly involve dispersive forces. Hence, they were able to make predictions about the thermochemical reactions between the ions Se˙+, SeH+, SeH2˙+, SeH3+, Ar2˙+, Ar2H+, Ar2H2˙+, Ar2H3+, ArCl+, ArClH˙+, and ArClH2+ with neutral gases H2, CH4, NH3, O2, CO, CO2, NO, and N2O used in DRC-ICP-MS. Results were presented in tables of data and as thermochemical ladders, including previously unreported values for Ar and Ar2 hydrides and ArCl-containing neutral and ionized species. A full discussion of these results is beyond the scope of this review so readers interested in optimizing the performance of DRC-ICP-MS for Se determinations are advised to refer to the full paper.
The structure of the LIBS plasma was studied by Bai et al.109 using dual-wavelength differential spectroscopic imaging. This involved the use of two filters, alternatively placed in front of the CCD detector, to eliminate the background continuum signal from a specific emission line. A pure aluminum target in an argon atmosphere was used to observe the distribution of ambient gas and target evaporated species inside the ablation plume using different ablation regimes (e.g., different laser fluences and pulse durations). This study demonstrated the key role that laser fluence has on the internal structure of the LIBS plasma. Similarly, Eschlböck-Fuchs et al.110 investigated the expansion dynamics and the optical emission of LIBS as a function of the temperature of sample materials such as aluminum alloys, silicon wafers and metallurgical slag. They observed that higher temperatures lead to larger plume volumes and brighter broadband plasma emission in air, though other plasma parameters were found to be independent of temperature, at least for the slag samples.
Freeman et al.111 studied ns and fs LIBS operated in vacuum and at atmospheric pressure conditions. They investigated the dynamics of plasma evolution and characteristics of the emission spectra and observed that the fs-LIBS plasma plume expanded in a narrower angular distribution, probably due to a strong, self-generated magnetic field created at the onset of plasma expansion. They also noticed that ns-LIBS spectra showed transition lines from lower-charged ions, while fs-LIBS spectra were mainly dominated by neutral species emission. The fs-LIBS technique was employed by Banerjee et al.112 for the detection of submicron buried layers inside a semiconductor material (e.g., Si) by making use of the precise hole drilling capabilities of fs-LA. A correlation between depth resolution and processing speed was observed when carrying out these analyses. Ahmido et al.113 used fs-LIBS to detect visible to NIR signatures from molecular fragments of the nitro (NOx) group after the ablation of potassium nitrate (KNO3) and sodium nitrate (NaNO3) coatings. A potential application of this method could be the detection of nitro-group energetic materials at standoff distances.
Spectra in LIBS are sometimes complex and difficult to interpret, thus Labutin et al.116 have developed an algorithm to automatically identify emission lines. This simulates the spectra corresponding to different temperatures and electron densities, then searches for the best correlated pair of experimental and model spectra constructed using information about atomic and ionic lines, broadening mechanisms and the parameters of the spectrograph. The model then attributes lines to the peaks. The algorithm was successfully applied to interrogate the spectral range from 393.34 to 413.04 nm, identifying 41 peaks.
Time-resolved, thermodynamic characterization of small, high-jitter powerchip LIBS plasmas was performed by Merten et al.123 under a variety of atmospheres (e.g., air, Ar and He). This fast-evolving plasma was evaluated for LTE in terms of both the McWhirter and temporal criteria. The former considers the electron/heavy particle collision rate required to guarantee that the total effective population and depopulation rates are dominated by electron collisions, and the latter examines the evolution of temperature and electron number density and compares their rate of change to the rate at which electron collisions can thermalize the change. These considerations were used to conclude that LTE might be present at the early stage of the powerchip LIBS plasma generation, especially under an argon atmosphere, making possible calibration-free LIBS measurements.
Another variant of conventional LIBS, called nanoparticles enhanced LIBS (NELIBS), was investigated by Giacomo et al.125 The method is based on the deposition of a drop of noble metal colloidal solution on the surface of a solid conducting target, in order to enhance the LIBS emission intensity. It was shown that the NELIBS enhancements were related to changes in the ablation process because a significant decrease in the ablation threshold was detected for all of the investigated metal samples. However, no significant differences were observed in the temperature and electron number density for NELIBS compared with LIBS.
LIBS is a powerful analytical technique for the analysis of samples placed in a gas environment (e.g., air, Ar, He) or in vacuum, but also for the analysis of samples immersed in liquids. Lazic et al.126,127 studied LIBS plasma and vapour bubble formation after ablation of Al targets underwater, using dual pulse excitation, by fast photography and time resolved spectroscopy. Position-selective emission spectroscopy of a LIBS plasma produced on an Al target in water was investigated by Matsumoto et al.128 They observed that the density at the edge of the plasma produced by long-pulse irradiation (ca. 100 ns) was low enough to achieve well-resolved, narrow spectral lines, even without time-gated detection. These results might change the current concept of the instrumentation for underwater LIBS and open new possibilities for in situ chemical analysis of solid materials in water.
The combination of LIBS with diode-laser LAAS and LIF has been investigated by Khater129 to improve the analytical capability and figures of merit of LIBS for trace analysis of light elements (e.g., C, S, P and N) in steel samples.
Biscan et al.133 investigated the production of 23S1 metastable helium in a laser-induced plasma at low pressure. The absorption CRDS technique was employed to measure the density time evolution of metastable helium and ablated atoms, produced during the ns laser ablation process of various materials (e.g., graphite, aluminum and iron) in a low pressure (e.g., 0.03–2 mbar), helium atmosphere. They concluded that the width of ablated atom lines was mainly related to directional movement of atoms and that the appearance of metastable He might be attributed to excitation in collisions with ejected electrons and heavy excited particles.
Many light atoms (e.g., H, O, C) have LIF transitions in the VUV spectrum that are difficult to excite. Two-photon LIF has been developed to overcome this difficulty by exciting the LIF transition by absorption of two photons. One problem with this approach is that amplified spontaneous emission (ASE), which is created by the population inversion between the excited and the de-excited states, might affect the linear response of the LIF signal with respect to the number density of the target species and laser intensity. With this in mind, Li et al.135 investigated, using a Monte Carlo model, the effects of line-narrowing of ASE, which is due to the fact that ASE photons near the central frequency are amplified more effectively. Consideration of these line-narrowing effects was necessary to avoid significant errors in the interpretation of the LIF and ASE signals.
It is always appropriate to periodically re-assess the principles and knowledge-base. In this respect, a review by Baxter et al.136 was a welcome assessment of some fundamental aspects of SF-MS. Factors which influence the IR measurement, at all stages of the process from the true ratio, through the measured ratio, to the final corrected ratio, were deliberated. The effects of Poisson counting statistics and flicker noise on ratio precision were treated mathematically, and accuracy offsets via abundance sensitivity tailing were discussed. Usefully, a thorough consideration of mass discrimination was contained in this tutorial, reviewing both the correction functions and the methodology of internal and external normalisation, as well as sample standard bracketing. Finally there was a comprehensive evaluation of dead time correction including a systematic review of the differing estimation methods which will be of great use to novice and experienced users.
There have also been developments in detection systems, such as a new generation of Faraday detectors equipped with low-resistivity (1012 Ω) resistors reported by Koornneef et al.139 These detectors were designed to provide lower baseline noise compared to the conventional 1011 Ω amplifiers, and were used to measure 87Sr/86Sr and 143Nd/144Nd for 1 ng amounts of analyte. Comparison with previous studies indicated that reproducibility was improved by factors of 2 and 1.4, respectively. Combined with increases in instrument efficiency, this has the potential to benefit applications such as radiometric dating or characterising isotopic disequilibrium in geological samples.
Hu et al.140 assessed the effect on sensitivity of fast interface pumping and the X-skimmer and Jet-sampling cones in a commercial MC-ICP-MS instrument. Combinations of new and old cones were trialled for Hf, Lu and Yb with a view to optimizing Hf IR measurements in zircons by LA. Results indicated that the X- and Jet-cone combination improved Hf sensitivity by a factor of ∼5 compared to the standard cone combination. An interesting feature of the analysis was that the Jet-sampling cone resulted in a non-linear mass-dependent fractionation that could not be corrected by standard fractionation laws. However, it was found that this fractionation effect was inhibited by the addition of small quantities of nitrogen to the central argon-helium flow from the collision cell.
The effects of mass discrimination in the sampling and post-sampling regions of ICP-MS was examined by Kivel et al.141 Using tantalum or graphite targets positioned close behind the skimmer cone orifice, elements ranging in mass between Li and U were admitted under wet plasma conditions to the interface. Material deposited on the target was then analysed both in bulk by solution nebulization, and as individual points along a radial transect from the central axis by laser ablation. The bulk measurements revealed a heavy-isotope bias, indicating that extraction of ions from the plasma contributed in the order of 50% of the measured discrimination. Radial measurement of Cd isotope ratios demonstrated a heavy-isotope bias at the axis which progressed to a relatively light-isotope bias outwards. As the ion beam was subject to “clipping” downstream from the target position, this was taken by Kivel et al. to indicate that further light-ion removal completed the remaining fraction of the instrumental mass discrimination.
Sometimes major improvements in analytical efficiency may be too good to be true. This may well be the case with intracavity optogalvanic spectroscopy (ICOGS) for measuring radiocarbon isotope systematics. Initial reports of this technique (Murnick, D. E.; Dogru, O.; Ilkmen, E. Anal. Chem. 2008, 80, 4820−4824) suggested that it could match the LOD achieved by AMS for 14C/12C. At the time, this report generated significant interest because of the potential to simplify the analysis, reduce costs and increase the size of datasets. However, Persson et al.142 have produced a frank assessment of the technique based on a comprehensive dataset and they have not been able to reproduce the reported optogalvanic signal dependence on 14C concentration and wavelength. In addition, they have now estimated that the detection limit for 14C is likely to be > 2 orders of magnitude higher than originally estimated.
Instigation of the double spike or GS-IDA techniques for determining isotopic composition requires calculation of the appropriate spike composition and an understanding of the optimum sample-spike mixing proportions. John144 has developed models where the concentrations of the spike and sample are allowed to vary independently, rather than restricting spike-plus-sample to a fixed concentration mixture (within the bounds of the signal capacity of the detector). The result was a decrease of up to 30% in the theoretical error, or greater reductions where isobaric interferences were present.
Double spike systematics have also been applied to the analysis of Pt stable isotope ratios.145 Using two separate Pt double spikes, the 198Pt/194Pt ratio could be reproduced to within 10−5 amu−1. These findings indicated that, despite variable amounts of isobaric 192Os, accuracy of the Pt isotope ratio was unaffected because 192Pt was not used in the double-spike de-convolution. Application of this technique provided a redefined Pt atomic weight and indicated that Pt stable isotope fractionation is quantifiable above the reproducibility of the method.
Stable Zn isotopes have been the focus of a study by Makishima and Nakamura,146 who have developed a low-blank method for Zn separation using TEVA resin, followed by a double spike determination of δ66Zn. Blank levels were reduced from greater than 1 ng reported in earlier studies to around 50 pg, enabling δ66Zn to be reproduced to 0.004% RSD (2σ) for 100 ng of Zn. This method will prove valuable in high-resolution studies of Zn where material is limited and Zn concentrations are low.
Shalev et al.147 compared MC-ICP-MS and TIMS for measuring the non-radiogenic 88Sr/86Sr ratio using a double spike. Results of the double spike bias corrections were found to be more reproducible than sample-standard bracketing for MC-ICP-MS. The two instruments produced the same 88Sr/86Sr values, with TIMS having smaller within-run errors but needing a longer integration time to achieve this.
The need for rapid yet accurate analysis of isotopic materials has made LA a potential sampling tool for safeguards and nuclear forensic investigations. Kappel et al.149 used a nanosecond laser coupled with MC-ICP-MS to directly analyse U isotope ratios in single U-doped glass particles. Uncertainties were better that 1.4% and 6% (2σ) for the 235U/238U and 236U/238U ratios, respectively. A similar study of direct measurement of uranium isotopes in enriched uranium pellets and rods was undertaken by Marin et al.,150 using a 266 nm Nd:YAG LA–SF–ICP-MS. A combined precision and accuracy of <2% RSD (σ) for the 235U/238U ratio was achieved. Each study defined a reproducibility window between 0.5% and 2% as the state of the art for 235U/238U determinations on individual particles, setting the analytical benchmark for nuclear forensic methods in the future.
Efforts to measure 236U/238U in samples with ultra-low 236U have been advanced by Tanimizu et al.151 using tandem Q-ICP-MS/MS. A key feature of this analysis was the oxidation of U in the reaction cell between the tandem quadrupoles. In this reaction only 1% of the UH+ was converted to UOH+. The reduction in 235UOH+ production enabled 236U16O+/238U16O+ to be measured down to 10−9, which approaches the analytical LOD of TIMS.
As with many isotope systems reviewed, the common aim is to increase sensitivity, reduce background interference and improve precision of analysis, ultimately to enable IR measurement of ever smaller quantities. One requirement in the nuclear field is the monitoring of humans, particularly if exposed to anthropogenic sources of uranium in their occupation, or as a result of proximity to a nuclear installation. Arnason et al.152 took on the challenge of measuring <40 pg mL−1 of uranium isotopes in urine. Solid phase extraction of the urine matrix followed by U isolation using a double-washed UTEVA resin minimised the blanks to <0.3 pg, i.e., ∼1%. Based on a urine sample of as little as 1 mL and ≥50 pg U, the 235U/238U ratio was measured within 2% (2σ), which is a useful level of detection for monitoring.
Where the quantity of uranium is not restricted, 235U/238U can be determined by more conventional mass spectrometric techniques. Mathew et al.153 used sample loads of 1 to 2 μg U to assess the total evaporation TIMS method. One finding was that the mass bias, assessed by various U standards, was consistent within individual analytical turrets (batches), but variable between turrets. Using these standard-based ‘external’ estimates of the mass bias resulted in precision better than 0.04% (2σ) for the 235U/238U ratio, which approaches the precision attainable by U double spike analysis. However, this external method could not detect if an individual sample had incomplete matrix separation, which may influence the degree of mass bias.
Pramann et al.155 have investigated a potential source of “broadband” interference across the Si mass spectrum in MC-ICP-MS. This took the form of a broad raised background centred at 29Si–30Si, present in NaOH-containing solutions but absent from KOH solutions. It appeared that this interference was detector specific because it was not observed in the central Faraday cup, only in the high- and low-mass Faraday arrays. The logical conclusion is that the interference, which significantly degrades Si isotope ratios, was caused by ion scattering in the presence of high sodium concentration. Absence of the interference from the centre Faraday was most likely related to this detector's position in a shielded location offset behind the main high-low Faraday array.
As often happens with analytical developments, sometimes two papers on essentially the same technique emerge simultaneously. Yang et al.156 and Sanchez-Lorda et al.157 are an example of such. Elements with relatively similar ionization efficiencies can experience similar levels of instrumental mass bias. For elements with different efficiencies, such as Hf and Yb, mass bias interrelationship can be formulated during, or prior to, the measurement. Hence, it is a logical progression to examine if elemental concentration ratios and isotope ratios can be measured within an individual multicollector analysis. In both of the aforementioned papers MC-ICP-MS was trialled for the measurement of 143Nd/144Nd in the same run as 147Sm/144Nd. These ratios are of particular interest because they form the basis of Sm-Nd geochronology, and because they are preferably measured from the same dissolution and chemical separation. Conventional analysis of the Sm-Nd system is by TIMS or MC-ICP-MS and requires IDA to determine the concentration ratio and chemical separation of Sm from Nd. Both studies used Nd isotopic standards doped with variable amounts of Sm to replicate the likely natural range of Sm/Nd, and formulated a relationship between the exponential mass bias coefficients to correct the isobaric 144Sm interference. Reported errors on 143Nd/144Nd and 147Sm/144Nd were similar for the two studies, and were roughly equivalent to measurements made by conventional methods. Sanchez-Lorda et al.157 reported that 147Sm/144Nd reproducibility ranged from 0.2% to 0.7%, which is slightly higher than for IDA measurements. A significant difference in the papers is that these workers also included a new methodology to isolate the rare earths from the matrix elements prior to the combined-run 143Nd/144Nd, 147Sm/144Nd MS analysis.
Recent improvements in sample introduction and plasma sampling have produced a step-change in the sensitivity of ICP systems. Using a PFA microflow nebulizer, desolvator and X-skimmer/Jet-sampler cones, Huang et al.158 sought to precisely determine Nd isotope ratios for ng levels of analyte using MC-ICP-MS. They achieved a sensitivity of >2000 V per μg mL−1 for Nd. At this elevated level of detection, a 2.5 ng mL−1 solution produces a 1 V signal for 146Nd (∼1 × 10−11A ion current). The analysis of small volume samples (0.5 mL) containing 0.5 ng of Nd resulted in a 143Nd/144Nd ratio for the JNdi-1 standard of 0.512101 ± 0.000029 (2σ, n = 6). With 5 ng of Nd the errors were reduced to 0.512124 ± 0.000010 which is comparable to typical long-term reproducibility statistics for NdO+ measurements by TIMS. Huang et al.158 have used this advance in sensitivity to evaluate the potential for this method to tackle Nd isotope determinations on seawater. Using a 5 km seawater sampling station off Bermuda as a trial for the method, the results were comparable to previously achieved Nd isotopic profiles.
The IRA of Pb by LA-MC-ICP-MS was also tackled in a study by Darling et al.160 who compared the results from direct measurement with samples where Pb was chromatographically separated from the matrix. A suite of samples from the Sudbury impact structure, in particular the included sulfide minerals chalcophyrite, pentlandite, pyrite and pyrrhotite, were the focus of the study. Encouragingly, the Pb isotope ratios generated by the direct sampling and by chemical isolation of Pb were within analytical uncertainty, typically 0.1% (2σ) for 206Pb/204Pb. This was achieved despite bias corrections being made to non-matrix matched external standard measurements. Common problems encountered with sulfide mineral analysis are the isobaric and molecular interferences from oxides, argades and sulfides of Re, Os, Pt and the rare-earth elements. However, by restricting oxide production to <0.5%, and monitoring the 205Tl/203Tl ratio, it was concluded that these isobaric interferences did not influence the measured Pb isotope ratios. An interesting facet of the results is that analyses from each sample, including the solution analyses, formed linear arrays in Pb-isotope space (e.g., 207Pb/204Pb – 206Pb/204Pb). This suggests a isochronous relationship, and helps define the initial (at formation) isotope ratios across the spectrum of samples studied.
A major issue encountered in LA analysis is concentration calibration. This has often been addresses by normalising analyte signals to other elements of known concentration, or by the co-introduction of a calibrated spike solution. The best accuracy can be achieved using IDA, which Claverie et al.161 demonstrated using a spinning platform within the ablation cell. Sample material and an isotope spike in solid form were located adjacent to each other on the sample holder, such that fast rotation (∼2000 rpm) of the sample under the laser resulted in a quasi-simultaneous ablation of the spike and sample. Initial trials also included the addition of a Gd spike to both sample and reference materials to correct for any differences in the ablation rate caused by position or material properties.
In some instances, e.g., for large samples, it is not appropriate to undertake laser ablation in a standard ablation cell. Dorta et al.62 experimented with fs-LA of samples in air, followed by conversion to an argon carrier in a gas exchange device, and introduction to a MC-ICP-MS instrument. Results for Hf, Pb, and Zn isotopes were comparable with air sampling and a conventional He atmosphere ablation cell. Single shot analyses produced an acceptable level of precision for archaeological applications.
Dating minerals by LA-MC-ICP-MS continues to evolve. Cottle et al.162 employed a single-shot method with modified data reduction and error propagation routines for U–Pb dating of zircon and monazite. The method effectively integrated the total signal for each isotope, such that only a single isotope ratio was generated instead of a series of measured values. Key advantages of this method are that it counters any IR bias resulting from signal decay or response-time offsets caused by differences between Faraday and ion counting detectors. The method generated accurate and reproducible ages for standard zircons and monazites relatively rapidly, being around 6 times faster than conventional LA-ICP-MS. Probably the most appropriate application is where an age population is required amongst a large number of zircons. Examples of this include identification of age grouping of grains in detrital sediments and in screening zircon populations prior to more detailed age-selective analysis.
8-HQ | 8-Hydroxyquinoline |
a.c. | Alternating current |
AA | Atomic absorption |
AAS | Atomic absorption spectrometry |
AIP | Atom-imprinted polymer |
ASE | Amplified spontaneous emission |
CID | Collision induced dissociation |
CRDS | Cavity ringdown spectroscopy |
CRM | Certified reference materials |
CVG | Chemical vapour generation |
DBD | Dielectric barrier discharge |
DLLME | Dispersive liquid–liquid microextraction |
DDTC | Diethyldithiocarbamate |
DLAAS | Diode laser atomic absorption spectroscopy |
DMA | Dimethylarsenic (include oxidation state if known) |
DOF-MS | Distance-of-flight mass spectrometry |
DRC | Dynamic reaction cell |
EDXRF | Energy dispersive X-ray fluorescence |
EPA | Environmental protection agency |
ES | Electrospray |
ESI-MS | Electrospray ionization mass spectrometry |
ETV-ICP-MS | Electrothermal vapourization inductively coupled plasma mass spectrometry |
FAAS | Flame atomic absorption spectrometry |
FFF | Field flow fractionation |
FI | Flow injection |
GC-MS | Gas chromatography mass spectrometry |
GD | Glow discharge |
GD-MS | Glow discharge mass spectrometry |
GE | Gel electrophoresis |
GF | Graphite furnace |
HG | Hydride generation |
HPLC | High performance liquid chromatography |
ICP | Inductively coupled plasma |
ICP-AES | Inductively coupled plasma atomic emission spectrometry |
ICP-MS | Inductively coupled plasma mass spectrometry |
ICP-MS/MS | Inductively coupled plasma mass spectrometry/mass spectrometry |
ICP-OES | Inductively coupled plasma optical emission spectrometry |
ID | Isotope dilution |
IDA | Isotope dilution analysis |
IDMS | Isotope dilution mass spectrometry |
IIP | Ion-imprinted polymer |
IL | Ionic liquid |
IR | Isotope ratio |
IRA | Isotope ratio analysis |
LA | Laser ablation |
LA-ICP-MS | Laser ablation inductively coupled plasma mass spectrometry |
LAMIS | Laser ablation molecular isotopic spectrometry |
LC | Liquid chromatography |
LC-MRM-MS | Liquid chromatography multiple reaction monitoring mass spectrometry |
LEP | Liquid electrode plasma |
LIBS | Laser induced breakdown spectroscopy |
LLE | Liquid–liquid extraction |
LLLME | Liquid–liquid–liquid microextraction |
LOD | Limit of detection |
LOV | Lab-on-a-valve |
LPME | Liquid phase microextraction |
MC-ICP-MS | Multicollector inductively coupled plasma mass spectrometry |
MCP | Microchannel plate |
MDG | Monodisperse microdroplet generator |
MeCat | Metal coded affinity tag |
MeSeCys | Methylselenocysteine |
MIP | Microwave induced plasma |
MNP | Magnetic nano-particle |
MS | Mass spectrometry |
NE-LIBS | Nanoparticles enhanced LIBS |
NIST | National institute of standards and technology |
NP | Nano-particle |
NTV | Near torch vapourization |
OES | Optical emission spectroscopy |
PDMS | Polydimethylsiloxane |
PFA | Perfluoroalkyl |
QMS | Quadrupole mass spectrometry |
Q-ICP-MS | Quadrupole inductively couple plasma mass spectrometry |
qPCR | Quantitative polymerase chain reaction |
r.f. | Radio frequency |
REE | Rare earth element |
RSD | Relative standard deviation |
SC-GD | Solution-cathode glow discharge |
SDS-PAGE | Sodium dodecyl sulfate polyacrylamide gel electrophoresis |
SeCys2 | Selenocystine |
SeEt | Selenoethionine |
SeMet | Selenomethionine |
SF-ICP-MS | Sector field inductively coupled plasma mass spectrometry |
SF-MS | Sector field mass spectrometry |
SIMS | Secondary ion mass spectrometry |
SPE | Solid phase extraction |
SPME | Solid phase micro-extraction |
TEM | Transmission electron microscopy |
TOF-MS | Time-of-flight mass spectrometry |
UV-vis | Ultraviolet-visible spectrophometry. |
This journal is © The Royal Society of Chemistry 2014 |