A dynamic system for the generation of stable hydrogen peroxide test atmospheres was applied to the evaluation of samplers used for the determination of hydrogen peroxide in workplace air. The system is able to generate gas mixtures of between 0.1 and 10 ppm at different combinations of relative humidity (20–80%) and temperature (10–30 °C). Gaseous hydrogen peroxide is sampled on glass filters impregnated with Ti(IV) chloride and sulfuric acid and analyzed by UV spectroscopy. An interference was observed due to the acid catalyzed decomposition of peroxyacetic acid to hydrogen peroxide. This was significantly reduced by applying high sample flow rates and by lowering the concentration of sulfuric acid. The performance of the sampler and sources of uncertainties were tested according to the European Standard EN 1076.

Article information

Article type
Paper
Submitted
06 Mar 2000
Accepted
08 May 2000
First published
23 Jun 2000

J. Environ. Monit., 2000,2, 339-343

Determination of hydrogen peroxide in workplace air: interferences and method validation

C. S. Christensen, S. Brødsgaard, P. Mortensen, K. Egmose and S. A. Linde, J. Environ. Monit., 2000, 2, 339 DOI: 10.1039/B001799J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Spotlight

Advertisements