Analytical information from doublet peaks in flow injection analysis. Part I. Basic equation and applications to flow injection titrations

Abstract

An equation relating the peak separation (Δteq), between the doublet peak obtained when the reaction product of a flow injection titration is monitored, to all the basic operational variables was derived on the basis of the well stirred mixing chamber model. These variables consisted of the flow-rate, the volume injected, the volume of the mixing chamber and the concentrations of the reagent and the injection sample. The equation was tested for the predicted variation of Δteq with these variables for a system containing real mixing chambers but using slug injection. The reactants were copper(II) and EDTA and the product was monitored with a UV detector. Good agreement was obtained. The use of a gradient tube was demonstrated to give a good calibration for OH over the range 5 × 10–5–5 M. A comparison of the doublet peak method with potentiometric methods was made.

Article information

Article type
Paper

Analyst, 1987,112, 523-526

Analytical information from doublet peaks in flow injection analysis. Part I. Basic equation and applications to flow injection titrations

J. F. Tyson, Analyst, 1987, 112, 523 DOI: 10.1039/AN9871200523

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements