Open Access Article
Selma B. Belfakir
*ab,
Jonas A. Jürgensen
a,
Marcus Petersson
d,
Max Stæhr Wraaed,
Patrick Munk
c,
Suthimon Thumtecho
be,
Mateus G. Masteghin
f,
Anne Ljungars
b,
Andreas H. Laustsen
abd,
Sandra W. Thrane
d and
Georgina M. S. Ross
*a
aVenomAid Diagnostics ApS, DK-2800 Kongens Lyngby, Denmark. E-mail: gin@venomaid.com
bDepartment of Biotechnology and Biomedicine, Technical University of Denmark, DK-2800 Kongens Lyngby, Denmark. E-mail: selbel@dtu.dk
cNational Food Institute, Technical University of Denmark, DK-2800 Kongens Lyngby, Denmark
dBactolife A/S, DK-2100 Copenhagen, Denmark
eDivision of Toxicology, Department of Medicine, Chulalongkorn University, King Chulalongkorn Memorial Hospital, The Thai Red Cross Society, Bangkok, Thailand
fDTU Nanolab, Technical University of Denmark, DK-2800 Kongens Lyngby, Denmark
First published on 24th April 2026
The rapid and sensitive detection of toxigenic Vibrio cholerae and enterotoxigenic Escherichia coli remains a critical challenge for effective disease screening and environmental monitoring, which is essential for guiding clinical management and outbreak control in regions where diarrheal diseases are endemic. To address this, we developed a single-domain antibody (VHH)-based sandwich-format lateral flow assay for the rapid and sensitive detection of cholera toxin from V. cholerae and heat-labile enterotoxin from enterotoxigenic Escherichia coli in human fecal and environmental water samples. It can be challenging to directly immobilize small proteins such as VHHs onto nitrocellulose while controlling their orientation to support optimal binding. To overcome this challenge, a bifunctional fusion protein comprising a VHH and a cellulose-binding module (CBM) was developed, enabling controlled orientation and immobilization of the VHH on nitrocellulose membranes. Using this fusion protein as the test line in an LFA allowed for the capture of cholera toxin and heat-labile enterotoxin across a wide dynamic working range (1–1000 ng mL−1), with detection limits down to 12.5 ng mL−1 in human fecal samples and 1 ng mL−1 in water. The findings presented here highlight the potential of using a CBM to facilitate the immobilization of a VHH in a sandwich format lateral flow assay for the rapid and sensitive detection of bacterial toxins in complex matrices.
Typically, detection of ETEC and/or V. cholerae in the clinic relies on bacterial culture. Although this method is specific and inexpensive, it is slow (about 24 hours) and depends on laboratory infrastructure, limiting its use in many endemic regions.7 For V. cholerae, commercial tests are available that detect the presence of the bacterium, but their specificity and sensitivity are variable.8 An overview of the tests available for the detection of ETEC and V. cholerae can be seen in Table S1. Rapid tests detecting bacteria can be negatively impacted by cross-contamination, lytic bacteriophages, or antibiotics, leading to false results.9,10 Such false results can lead to unnecessary public health responses, as occurred in South Africa, where a suspected V. cholerae outbreak was ultimately traced back to a laboratory contamination.9,10 Another option well-suited to simultaneously screening many targets is metagenomics. However, this method has low sensitivity, requires deep and expensive sequencing, and downstream computation. For example, the metagenomic assemblies that revealed that non-toxigenic V. cholerae persisted in Copenhagen sewage cost thousands of USD and took several weeks to complete, making this method inappropriate for rapid screening.11 Currently, the gold standard for detecting ETEC infections is based on the polymerase chain reaction (PCR) technology. While sensitive, PCR is slow and often unavailable in endemic regions due to the need for specialized laboratories.12
Rapid and sensitive detection of toxigenic bacteria that cause diarrheal disease, such as V. cholerae and ETEC, is essential for appropriate clinical management and for guiding timely health responses.1,13 Recently, the Global Task Force on Cholera Control (GTFCC) published a Target Product Profile (TPP) pointing towards integrating CTX detection for optimal disease monitoring.14 In line with this, there is a growing need for rapid tests for bacterial pathogens that detect the primary causative agent responsible for the main pathology of the clinical disease. This approach offers the potential for improved specificity and the ability to detect cases where culture conditions are unfavourable.13 Compared to DNA-based methods, such as quantitative PCR (qPCR) or metagenomics, toxin detection measures the phenotype directly, showing that both transcription and translation of virulence factors have occurred.
Lateral flow assays (LFAs) are potential candidates for rapid, low-cost, detection of pathogenic bacteria as they fulfill the criteria of affordability (<3 USD) highlighted in the GTFCC's TPP, and can conform to the World Health Organization's (RE)ASSURED criteria (real-time connectivity, ease of specimen collection and environmental friendliness, affordable, sensitive, specific, user-friendly, rapid, equipment-free, delivered) that guides disease control strategies and define benchmark standards for point-of-care (POC) diagnostics.15 Currently, two commercially available LFAs have been approved by the GTFCC for cholera detection, but both target bacterial surface antigens, which prevents them from discriminating between toxigenic and non-toxigenic strains.16–20 In contrast, at the time of writing, no rapid tests are commercially available for the detection of ETEC, although some have been described in the literature, as summarized in Table S1.
Despite their benefits, LFAs also have drawbacks as they often lack the sensitivity and specificity of other laboratory-based methods.21 However, it should be recognized that on-site screening assays such as LFAs have different performance requirements than laboratory-based methods, and it is not strictly necessary for on-site methods to achieve the same level of performance as their laboratory counterparts. With the emergence of novel antibody formats with unique functions, such as single-domain antibodies (e.g., VHHs), there is an opportunity to implement these as immunoreagents to improve the performance of LFAs and allow the detection of targets that could be challenging to reach with full-length antibodies.22–27 Single-domain antibodies offer several advantages over conventional immunoglobulin G (IgG), including their small size, allowing them to reach cryptic epitopes, high stability, and efficient recombinant expression and genetic engineering.23,24 Furthermore, their small size can allow for a denser coating onto assay membranes, increasing the number of available binding sites for capturing the target toxin, without compromising affinity.23 However, their small size also makes it challenging to immobilize VHHs in an oriented manner on nitrocellulose membranes. Previous studies have demonstrated that cellulose-binding modules (CBMs), protein domains with high affinity for cellulose, can be genetically fused to VHHs to enable their direct immobilization onto cellulose-based materials.28,29 Crucially, CBMs can be fused to VHHs distal to their binding site, allowing VHHs to be immobilized in an optimal orientation for interacting with their target.
In this work, we have developed a sandwich-format LFA using an engineered bifunctional fusion protein consisting of a VHH with specificity towards both CTX and LT (anti-CTX/LT VHH)30 genetically fused to a CBM as the test line. The assay, which works in 15 minutes, has a wide dynamic working range with the ability to detect CTX and LT down to concentrations of 12.5 ng mL−1 in fecal samples and 1 ng mL−1 in water samples. These findings demonstrate the promising nature of leveraging CBMs to immobilize VHHs in LFAs for the rapid and precise detection of bacterial toxins in complex samples to support current diarrheal disease management practices.
:
1 in Milli-Q (MQ) water (>18.2 MΩ cm−1) from Millipore (Burlington, USA) and vortexed to mimic diarrhea consistency. Tap water was collected directly from the tap (Kongens Lyngby, Denmark; Cinque Terre and Pisa, Italy; Casablanca and Marrakech, Morocco; Amsterdam and Utrecht, The Netherlands; Bali, Indonesia; Boston, The United States of America).
:
1 (v/v) ratio, incubated on ice for 10 min, then centrifuged at 20
000 g for 30 min at 4 °C. Purification was performed by immobilized metal affinity chromatography (IMAC) using gravity-flow columns (Bio-Rad, 7321010). HisPur Ni-NTA resin (Thermo Scientific, 10038124) was added to the supernatant at 50 mg mL−1 and incubated at 4 °C and 150 rpm for 30 min. The resin-supernatant mixture was added to chromatography columns. The columns were washed twice with IMAC wash buffer (Phosphate-Buffered Saline (PBS) with 200 mM NaCl and 20 mM imidazole). Then, IMAC elution buffer (PBS with 200 mM NaCl and 250 mM imidazole) was added to the column, and the anti-CTX/LT VHH–CBM1 and anti-CTX/LT VHH–CBM2 were collected. Protein concentration was determined by a spectrophotometer (Thermo Scientific, NanoDrop One) at 280 nm.
000 g for 15 min. The supernatant was filtered through a 0.2 µm filter. Purification was performed by IMAC using gravity-flow columns. HisPur Ni-NTA resin was added to the supernatant at 50 mg mL−1 and incubated at 4 °C and 150 rpm for 30 min. The resin-supernatant mixture was added to chromatography columns and washed with IMAC wash buffer, followed by a final wash with IMAC wash buffer. The protein was eluted with IMAC elution buffer, dialyzed overnight at 4 °C against PBS using dialysis tubing with a 3.5 kDa molecular weight cut-off, and concentrated to approximately 1 mL using a centrifugal concentrator. Protein concentration was determined using a spectrophotometer at 280 nm.
The acquired spectra were processed and smoothed using SpectraManager software (JASCO), and graphs depicting the CD spectra were generated using GraphPad Prism 10.
000 ng mL−1 CTX-B) in 98 µL of ultrapure water. Lacey carbon-coated copper grids (Agar Scientific) were glow-discharged for 10 seconds (current: 10 mA, air: 3.0 × 10−1) using EM ACE600 (Leica Microsystem). 3 µL of the prepared samples were drop-casted onto the glow-discharged grids and incubated for 2 minutes before excess was blotted off with filter paper. Following, 5 µL of uranyl acetate was spotted onto the grid and incubated for 1 minute, excess was removed by blotting, and the grids were washed with 5 µL MilliQ water. STEM was carried out in a Talos F200X G2 (Thermo Fisher Scientific) using a 70 mm C2 aperture and “spot-size” 5, resulting in a probe current <100 pA and a convergence semi-angle of 10.5 mrad. Images were acquired using high-angle annular dark-field, bright-field, and dark-field detectors. Bright-field TEM was performed to confirm the crystal structure of the AuNP.
:
1 in conjugate pad buffer (5 mM borate buffer (BB) (pH 8.8), containing 100 mM sodium tetraborate, 100 mM boric acid (Sigma-Aldrich, Switzerland) with 5% or 10% sucrose (Sigma-Aldrich, USA)) and dispensed at 10 µL cm−1, and the second conjugate pad contained anti-CTX/LT VHH–AuNP diluted 1
:
1 in conjugate pad buffer and dispensed at 20 µL cm−1. Both conjugates were dispensed using the µAirjet of the XYLite BioDot platform (Irvine, USA). A 20 mm sample pad (Millipore, Merck, Germany) was added below the conjugate pads with an overlap, and a 40 mm absorbent pad (Whatman, GE Healthcare, Germany) was affixed to the opposite end of the nitrocellulose membrane with an overlap. Test and control line reagents were diluted in 5 mM BB, 1% D(+)-trehalose dihydrate (w/v) (Carl Roth GmbH, Germany) and 1% ethanol (VWR Chemicals, France) before dispensing (conditions: 1 µL cm−1 at a rate of 20 mm s−1) using the BioDot platform. The LFAs had 0.25 mg mL−1 of goat anti-mouse F(ab′)2 (Jackson Immunoresearch, UK) as a control line and 1.5 mg mL−1 or 2 mg mL−1 anti-CTX/LT VHH–CBM as the test line. After dispensing, LFAs were dried for one hour at room temperature, cut to 4 mm with a CM5000™ Guillotine Cutter (BioDot, USA), and stored at room temperature in heat-sealed foil bags with desiccant pouches until use. A schematic representation of the developed LFA can be seen in Fig. 1.
![]() | ||
| Fig. 1 Schematic representation of the lateral flow assay architecture. A sample containing antigen (cholera toxin or E. coli heat-labile enterotoxin (ETEC-LT)) is added to the sample pad and moves up the strip by capillary action. Conjugate pad 1 contains mouse immunoglobulin G (IgG) chemically coupled to gold nanoparticles (AuNPs), and conjugate pad 2 contains anti-antigen single-domain antibodies (VHHs) chemically conjugated to AuNPs. When the sample reaches the conjugate pads, the conjugates are released. When reaching conjugate pad 2, the antigens will bind to the anti-antigen VHH–AuNP. When this complex then reaches the test line, a sandwich pair is formed between the anti-antigen VHH–AuNP and the immobilized anti-antigen VHH–cellulose-binding module (CBM) with the antigen in between. Thereafter, when the sample reaches the control line, the mouse IgG-AuNP from conjugate pad 1 will bind to an immobilized goat anti-mouse F(ab′)2. Finally, the remaining sample is absorbed by the absorbent pad. A negative test is represented by a single line at the control region, and a positive test is represented by a line at the control and test regions. Antigen structures are adapted from ref. 37 and 38. | ||
First, a comparison was made between LFAs with either 1.5 mg mL−1 anti-CTX/LT VHH–CBM1 or 1.5 mg mL−1 anti-CTX/LT VHH–CBM2 as the test line (no control line) by inserting the LFAs (n = 2) into a microwell containing 20 µL of CTX-B (0, 100, or 1000 ng mL−1), 80 µL of RB (PBS with 0.3% Tween-20), and 2 µL anti-CTX/LT VHH–AuNP. Next, test line concentration was assessed by testing the LFA with the preferred VHH–CBM at two concentrations (1.5 and 2 mg mL−1) and CTX-B (0, 100, 1000, 10
000 ng mL−1) using the same assay conditions as described above.
After establishing the most appropriate CBM and concentration, a comparison was made between an LFA with a test line of 2 mg mL−1 VHH directly immobilized by passive adsorption (LFA 1) and an LFA with a test line consisting of 2 mg mL−1 VHH–CBM (LFA 2). The LFAs were run (n = 2) by inserting the LFAs into a microwell containing 20 µL of CTX-B or LT-B (0, 1, 10, 100, 1000 ng mL−1), 80 µL of RB (PBS with 0.3% Tween-20), and 2 µL anti-CTX/LT VHH–AuNP.
Thereafter, the optimal RB was selected using the 2 mg mL−1 test line LFA in RBs containing PBS with 0%, 0.1%, 0.2%, or 0.3% Tween-20. Furthermore, three different AuNP (anti-CTX/LT VHH–AuNP) volumes were tested: 2 µL, 3 µL, and 4 µL. Conjugate pad buffers containing (i) 5 mM BB pH 8.8 with 5% sucrose, (ii) PBS pH 7.4 with 5% sucrose, (iii) conjugate diluent (0.5% PBS with 0.5% Bovine Serum Albumin (BSA) (Protease and IgG-free, Jackson Immunoresearch, UK), 0.5% casein (Carl Roth GmBH + Co. KG, Germany), 1% Tween-20 (Sigma-Aldrich, USA), and 0.05% sodium azide (Ampliqon, Denmark)) or (iv) the undiluted gold AuNPs directly sprayed onto the conjugate pad, were compared. For these experiments, the LFA was tested in blank RB or RB spiked with CTX-B (10
000 ng mL−1).
Lastly, using the same conditions as established above, LFAs were tested in a total volume of 100 µL with 1%, 5%, 10%, and 20% fecal sample (sample #1) spiked with CTX-B (10 ng mL−1) in RB.
Two concentrations (0.5 mg mL−1 and 1 mg mL−1) of anti-alpaca IgG (VHH domain) were diluted in 5 mM borate buffer containing 1% trehalose and spotted as a control line on a nitrocellulose membrane. LFAs (n = 2) were tested in a blank sample containing 100 µL RB and 2 µL anti-CTX/LT VHH–AuNP.
Using the optimal conditions chosen above, the LFA (test line: 2 mg mL−1 anti-CTX/LT VHH–CBM1, control line: 0.25 mg mL−1 goat anti-mouse F(ab′)2 specific) was tested (n = 3) in 5 µL sample (RB or fecal sample) spiked with CTX-B or LT-B and 95 µL RB (PBS with 0.3% Tween-20). When spiked into RB, CTX-B and LT-B were tested at 0, 1, 10, 100, 1000, 10
000, or 100
000 ng mL−1, and when spiked into fecal samples, CTX-B and LT-B were tested at 0, 1, 10, 100, or 1000 ng mL−1.
Using the optimal LFA configuration and conditions selected above, the LFA (n = 3) was tested in 5 µL of human fecal sample (sample #1) spiked with CTX-B, LT-B, STX-B, CTX, or LT at concentrations of 0, 0.1, 1, 10, 100, and 1000 ng mL−1 or spiked with CTX at concentrations 0, 12.5, 25, 50, and 100 ng mL−1, and 95 µL RB.
:
0, 50
:
50, and 20
:
80). Thereafter, the LFA was inserted (n = 2) into microwells containing 50 µL of tap water spiked with CTX-B (0, 1, 10, 100, and 1000 ng mL−1) and 50 µL RB (PBS with 0.5% BSA and 0.3% Tween-20) and the signal was allowed to develop for 15 min. Blank tap water samples from eight different cities were tested (n = 2) using the 50
:
50 sample-to-RB ratio.
To select the most appropriate on-strip conditions for the LFA, several optimizations were carried out. First, the optimal immobilization buffer was evaluated by Ponceau S staining (Fig. S3). A buffer containing 5 mM BB and 1% trehalose resulted in the most intense staining, indicating that more protein was immobilized, as Ponceau staining can detect protein on nitrocellulose membranes.35 Next, LFAs with test lines consisting of the anti-CTX/LT VHH fused to either CBM family 1 cel6a (CBM1) or cel7a (CBM2) were compared. These CBMs differ in their binding affinities for cellulose,32 allowing identification of the most effective CBM for immobilizing the VHH. The anti-CTX/LT VHH–CBM1 (2 mg mL−1) fusion improved the assay performance as the test line signal intensities developed on this test were observably more intense at all tested concentrations of CTX-B compared with the signal intensity on the test using an anti-CTX/LT VHH–CBM2 test line (Fig. S4a & b).
To test the hypothesis that CBMs enhance the immobilization of VHHs onto (nitro)cellulose membranes, as has previously been reported,28 an LFA with VHH directly immobilized at the test line (LFA 1) was compared to an LFA in which the VHH was immobilized via a CBM (LFA 2) fused to the C-terminal (Fig. 2). When tested in RB spiked with either CTX-B or LT-B, the LFA with the bifunctional VHH–CBM (LFA 2) produced a stronger signal and a sharper test line. This suggests that the CBM enabled more controlled anchoring at the test line. In contrast, VHHs that were passively adsorbed onto the nitrocellulose membrane resulted in a slightly more diffuse test line, which likely contributed to the reduced signal intensity and could indicate less controlled immobilization. Furthermore, when spiked with LT-B, a lower visual limit of detection (LOD) was observed. Since the concentrations tested were based on 10-fold dilution steps, a lower visual LOD might have been observed for LFA 2 (VHH–CBM) if tested in RB spiked with CTX-B at smaller concentration intervals (e.g., 2-fold). This indicates that CBM-mediated immobilization improves the anchoring of VHHs at the test line, either by enhancing VHH binding to the nitrocellulose membrane or by ensuring optimal immobilization orientation, facilitating target interaction.
Thereafter, the assay RB was optimized by increasing the Tween-20 concentration to facilitate flow and minimize non-specific interactions, with 0.3% producing the most intense test line signal at 10 ng mL−1 CTX-B (Fig. S4c).
Before implementing conjugate pads into the test architecture, different volumes of anti-CTX/LT VHH–AuNP were compared to determine which volume produced the most intense specific signal in spiked RB without introducing false positives in blank RB. Using 4 µL AuNPs yielded the most intense test line signal when tested in a ½-dipstick LFA format (Fig. S4d); therefore, this volume was used to prepare the conjugate pads.
Successful conjugation of VHH to 40 nm AuNPs (space group Fm
m, as shown in Fig. S5) was confirmed by ultraviolet-visible spectroscopy (Fig. S6), showing a 4 nm red shift in the localized surface resonance peak (λmax shifted from 523 nm to 527 nm). The maintenance of a sharp, symmetrical peak profile post-conjugation indicates the resulting nanoconjugates remained monodisperse (Fig. S6). Fig. S7 and S8 shows bright-field STEM images of the AuNPs (Fig. S7a), anti-CT/LT-AuNP (Fig. S7b and S8a), and anti-CT/LT-AuNP–VHH (Fig. S7c and S8b), in which the contrast is mainly defined by mass-thickness properties and, therefore, displays the AuNPs as the darkest feature followed by the protein (or protein plus CTX-B antigen). In Fig. S7a, the AuNP lies on the carbon support film without any clear “deposits” around it, whereas the functionalised AuNPs exhibit a distinct “halo” surrounding the particles. This “halo” was attributed to the anti-CTX/LT VHH and a mixture of anti-CTX/LT VHH and antigen (CTX-B), in which the antigen doubled the diameter of this carbonaceous matter surrounding the particle. Another indication that the VHH functionalization worked is the agglomeration of AuNPs in the presence of CTX-B (Fig. S8b). This could be caused by CTX-B being bound by multiple anti-CTX/LT VHH–AuNPs simultaneously, leading to agglomeration of the AuNPs.
Moving from a wet ½-dipstick LFA (i.e., without a conjugate or sample pad) to a full-format LFA with AuNPs dried down and pre-stored in a conjugate pad often impacts assay performance and thus requires optimization. Here, three different conjugate pad buffers were compared using a 1
:
1 dilution of the conjugate (anti-CTX/LT VHH–AuNP) in each buffer. A fourth condition was also compared; this condition used anti-CTX/LT VHH–AuNP directly on the conjugate pad, without dilution in buffer. It was found that conjugate pads with a 1
:
1 dilution of conjugate in 5 mM BB (pH 8.8) were the most optimal, as these conditions allowed for complete rehydration and release of conjugate from the pad upon wetting (Fig. S4e). Low-concentration buffers, such as BB with stabilizing agents such as sucrose, are typically recommended as conjugate pad buffers, as they ensure long-term stability of the dried-down conjugate while allowing complete rehydration.39
The assays were initially optimized using 20% toxin-spiked RB diluted in 80% RB. However, when testing fecal samples, it was determined that 5% (spiked or blank) fecal sample in (95%) RB was the most appropriate for running the assay (Fig. S4f) as it allowed for the most consistent sample flow and for total conjugate pad rehydration, resulting in a clear strip background. However, reducing the sample volume from 20% to 5% also reduced the concentration of antigen available for binding (Table S3); therefore, the reduction of sample volume can be considered a compromise that limits assay sensitivity but improves specificity by limiting matrix effects in fecal samples.
The LFA's test line signal intensities decreased proportionally (resulting in a decreased cGCPI value) to the decreasing concentration of CTX-B or LT-B in the sample, reaching a visual LOD of 1 ng mL−1 for CTX-B and 10 ng mL−1 for LT-B (absolute concentrations of 0.05 ng mL−1 and 0.5 ng mL−1, respectively) when spiked into RB (Fig. 3a).
When spiked into running buffer, a high-antigen-concentration effect (also referred to as the hook effect) can be observed at concentrations above 1000 ng mL−1 for CTX-B or LT-B, leading to reduced signal intensity at the test line (Fig. 3a). This occurs when the excess antigen in the sample saturates the test line, preventing the binding of the anti-CTX/LT VHH–AuNP at this region and leading to reduced test line signal intensity. At the concentrations tested in this study, there was no complete loss of test line signal, which would be interpreted as a false negative result, however even more extreme concentrations would likely cause this, as previously reported.40 Although the test line was not completely inhibited, signal intensities (and corresponding cGCPI values) at the highest concentration tested (100
000 ng mL−1) could be misinterpreted as a much lower response. This effect was most pronounced for CTX-B, where the signal at 100
000 ng mL−1 was lower than at 10 ng mL−1. In contrast, when tested in LT-B, the LFA did not have such a strong hook effect, likely because 100
000 ng mL−1 does not exceed the assay's upper dynamic range for this toxin. Accordingly, higher LT-B concentrations would be required to induce a hook effect, as visually detectable test line signals only emerged at ≥10 ng mL−1 (Fig. 3a). In any case, the high concentrations tested in this study exceed those expected in infected patients.41
In contrast, a visual LOD of 100 ng mL−1 (5 ng mL−1: absolute concentration) was observed for both toxins when spiked into human fecal samples (Fig. 3b), indicating that some matrix interference was affecting signal development. While the sensitivity decreased in the fecal samples, assay specificity remained high, with no false positives or background staining detected (Fig. 3b). Since the overall signal was reduced in fecal samples, it could be expected that any high-antigen-concentration effect impacting the test line would occur at even higher concentrations than those shown for spiked RB in Fig. 3a. While there are no commercial LFAs for the specific detection of only LT, other LFAs have been reported in the literature for the detection of CTX and LT (see Table S1).42,43 For example, one IgG-based LFA was able to detect CTX and LT at 10 ng mL−1 in spiked running buffer,43 however, as this assay was not assessed in biological or environmental matrices, it is difficult to assess its potential real-life applicability. In another study, researchers reported a proof-of-concept aptamer-based LFA for CTX detection. In this work, both competitive and sandwich-format LFAs were developed, achieving visual LODs of 100 ng mL−1 and 1 ng mL−1, respectively. However, in both of these works, the assays were only evaluated in toxin-spiked running buffer, and not in any relevant biological matrix.42,43 Other researchers have reported extremely sensitive assays, such as the proof of concept liposome PCR that reached an LOD of 10 pg mL−1, however, this is a molecular technique which is more complex and time-consuming than LFAs, which would limit its use case as a point-of-care test outside of a centralised laboratory.44 In other interesting examples, researchers developed a label-free carbon dots-based fluorescent aptasensor that was able to detect V. cholerae with an LOD of 426 CFU mL−1,45 and a loop-mediated isothermal amplification-based assay was able to reach LODs of ∼104 CFU g−1 of human feces in under one hour.46 It should be noted that comparing LODs expressed in CFU mL−1 and ng mL−1 is challenging as they represent different biological measures that are not directly correlated. Still, to have societal impact, LFAs for CTX detection should be able to detect the toxin at approximately 100 ng mL−1 in fecal samples47 or 1 ng mL−1 in water samples.44
The conjugate pads were tested in two positions (Fig. S9b), and the LFA with the conjugate containing anti-CTX/LT VHH–AuNP closest to the nitrocellulose membrane (configuration 2) produced the most intense test line signal (Fig. S9b). Lastly, increasing the sucrose to 10% in the conjugate pad buffer improved the signal intensity at 50 and 100 ng mL−1 and resulted in a clearer background (Fig. S9c). This is likely because the increased sucrose both enabled better conjugate stability and slowed down the release of the conjugate upon rehydration, allowing more time for the anti-CTX/LT VHH–AuNP to form a complex with the toxin in the sample.47
These optimizations enabled the assay to reach visual LODs of 10 ng mL−1 for CTX-B, 12.5 ng mL−1 for CTX, and 100 ng mL−1 for LT-B and LT (absolute concentrations of: 0.5 ng mL−1, 0.625 ng mL−1, and 5 ng mL−1, respectively) (Fig. 4). The decrease in sensitivity for CTX compared to CTX-B was anticipated as CTX is one and a half times larger than CTX-B pentamers.48 The larger antigen could result in steric hindrance, blocking free VHHs at the test line. Despite the decrease in sensitivity for the detection of the whole toxin, the test is still able to detect a clinically relevant concentration for patients with acute cholera infection, which is around 100 ng mL−1.41,49,50
CTX-B and LT-B are structurally and functionally similar.51 In contrast, the AB5-type Shiga toxin produced by Shiga toxin-producing E. coli (STEC) targets a different receptor (globotriaosylceramide) on kidney and endothelial cells.52 The LFA reported in this study can detect both CTX and LT, but not STX (Fig. 4). This is clinically relevant as both cholera and ETEC infections are treated similarly with rehydration and, in some cases, with antibiotics,1,53 whereas for STEC, supportive care alone is recommended.54,55 Therefore, this test could reduce the inappropriate use of antibiotics for STEC. In addition, as the anti-CTX/LT VHH targets an epitope that is conserved among LT toxinotypes relevant to human infection and among all pandemic cholera strains to date,31,56 we expect that the assay would be able to detect CTX and LT across all relevant serovars.
Differentiating between V. cholerae and ETEC remains important, both for treatment decisions and especially for outbreak source tracking. In the clinic, simply ascertaining that a toxigenic bacterium is the causative agent of the disease can support appropriate initiation of antibiotic usage. Today, antibiotics are prescribed based on symptoms alone due to the lack of rapid tests for pathogen identification. Confirming the presence of one of the toxins, and therefore one of the two bacteria, could potentially help reduce the use of antibiotics for viral infections with symptoms that resemble cholera and ETEC, e.g., infections by rotavirus or norovirus. However, the structural similarity between CTX and LT makes it challenging to develop binders that discriminate between these two toxins. Indeed, other assays developed for detecting CTX have also been reported to cross-detect LT.43 One possible solution to overcome this challenge of cross-detection could be to isolate and use VHHs that recognize regions of CTX-A or LT-A that are less conserved. However, this may be complicated by the limited sequence conservation within a toxinotype, which could make it challenging to achieve strong and specific binding across all LT or CTX variants. Such VHHs could potentially be discovered using phage or yeast display selection strategies, including CTX or LT for deselection to ensure that the binder only binds to CTX or LT.57
Among the tested samples, five (#1, 3, 4, 8, 11) produced a test line signal down to 100 ng mL−1 (5 ng mL−1: absolute concentration), and three (#5, #6 and #7) down to 1000 ng mL−1 (50 ng mL−1: absolute concentration), while the remaining three (#2, #9 and #10) produced no test line at any of the tested concentrations. Furthermore, one sample (#11) produced a false positive signal (i.e., the appearance of a test line in a sample free from the target antigen) (Fig. 5). In this sample (#11), the test line signal at 100 ng mL−1 was still higher than the false positive (0 ng mL−1), but the intensity of the test line in the unspiked sample indicates that some interfering component present in this fecal sample was interacting with the VHHs.
Variability in protein content in fecal samples can affect the detection of target analytes. At the same time, any proteases present in the sample can further degrade proteins and impair downstream analysis.59 To mimic this variability, the samples tested here were from different individual donors with varying age, gender, ethnicity, weight, height, health conditions, medications, and fecal consistency according to the Bristol Stool Chart (Table S2). Considering the diversity of the donors, it is understandable that the performance of the LFA was considerably impacted by the different samples (Fig. 5).
The large differences in sensitivity across samples demonstrate the challenging nature of detecting toxins in complex samples such as feces. In this work, the assay conditions were initially optimized using a sample from a single donor (#1). As a result of the assay optimizations, sample #1 produced an intense test line signal at all tested concentrations with no false positives. In contrast, a weaker test line signal intensity was obtained when testing some of the samples from other donors. In future work, to address this variability, additional sample preparation would likely be required to minimize interference from the feces and to ensure effective interaction between the immunoreagents and the toxin, although standard extraction protocols optimized for single samples may be difficult to generalize.59
Measuring one specific protein in the complex mixture of proteins present in fecal matter poses challenges.59 In fact, fecal matter is a highly heterogeneous biological matrix composed of enzymes, bile salts, undigested food particles, microorganisms, and metabolic by-products, which complicates the access to intact proteins for reliable detection by immunoassays.59 Furthermore, it is documented that the presence of bile salts can impact protein structure and conformation across both VHH and protein target, complicating reliable and uniform detection.60 For protein extraction, mechanical disruption of the fecal sample is essential to break insoluble particulates and bacterial cell walls, facilitating protein release. In this study, an additional vortexing step (30 seconds) was added before adding the sample to the LFA to ensure proper protein release.59 Therefore, developing diagnostic tools for complex variable matrices, such as human fecal samples, requires substantial work to develop a universal extraction method, with the caveat that sample preparation should not detract from the usability (or affordability) of the test.
The fecal samples tested in this study were from healthy donors, ranging from 1 to 6 on the Bristol Stool Chart61 (Table S2). To mimic the consistency of diarrhea samples (Bristol Stool Chart = 7), the fecal samples were diluted 1
:
1 in water before spiking and testing. A 1
:
1 dilution was selected as it allowed for a more liquid sample without substantially diluting the other potential interferents present in the sample. However, as fecal samples from cholera and ETEC patients are characterized by their “rice water” appearance,62 it is hypothesized that real-life samples would be more fluid than was achieved here using a 1
:
1 dilution. Indeed, the commercially available cholera LFAs rely on liquid stool, which can be added to the sample pad using a disposable pipette or rectal swab samples,16 highlighting the complexity of using “normal” solid stool (Bristol Stool Chart = 3–4) for testing. For future development, processing a rectal swab for LFA analysis could be considered if the test's sensitivity permits.
Another use case of the test could be the detection of ETEC in environmental samples from pig farms, where it is documented that ETEC causes post-weaning diarrhea in piglets.63 Here, swab samples could be taken from different pen surfaces, fecal samples from the pigs, or water samples from the troughs. To assess the sensitivity of the assay if used to detect CTX in contaminated water sources, CTX-B was spiked into (tap) water. In contrast to fecal samples, water does not require such a substantial dilution into RB to flow up the strip. Three different sample-to-RB ratios were tested, 100
:
0, 50
:
50, and 20
:
80, with a 50
:
50 ratio resulting in the best flow and the most intense signal at 1 ng mL−1 (Fig. S11).
The assay was able to reach an LOD of 1 ng mL−1 (0.5 ng mL−1: absolute concentration) of CTX-B when spiked into water (Fig. 6). That is comparable to, or even an improvement over, the standard optical ELISA (1 ng mL−1) or a magnetic bead-based sandwich enzyme-linked aptamer assay in tap water (2.4 ng mL−1), while providing a more rapid and user-friendly alternative with fewer analytical steps.44 Furthermore, no false positive signals developed in blank tap water samples from eight different cities (Fig. S12). Moreover, it is expected that this LOD could be further improved by using a 100
:
0 sample-to-RB ratio. However, to ensure enough Tween-20 for an appropriate flow, Tween-20 should be added to the sample or conjugate pad.
Still, environmental testing alone cannot confirm active disease outbreaks and must be complemented with fecal testing from symptomatic patients.64 The ability of the presented test to function in both fecal and water samples is therefore particularly valuable. Environmental monitoring of water sources remains important, as identifying contaminated sources can guide outbreak control through source removal. However, for assessing if water has fecal contamination or is safe to drink, V. cholerae detection should be combined with standard fecal contamination indicators such as crAssphage,65 thermotolerant coliforms, Escherichia coli, and fecal streptococci.64 In the context of environmental water monitoring, LFAs are suitable screening tools that can be applied to sort samples requiring confirmatory testing in a reference laboratory, thereby reducing the overall number of samples that need to be tested by these more complex, expensive, and time-consuming methods.
We exemplify how the simultaneous detection of CTX and LT with the same LFA is feasible, with high sensitivity, and can be relevant due to the similarity in spread, disease manifestation, and clinical management. Nevertheless, distinguishing between the two toxins remains important for future outbreak source tracking.
Fecal samples represent a highly complex biological matrix that poses significant challenges for reliable performance in LFAs. Due to fecal inter-sample variability, it was necessary to use a sample-to-RB ratio of 5
:
95 (µL), reducing the amount of antigen for detection and thereby the overall sensitivity of the assay. Current assays utilize fecal swabs or liquid samples, which may contain less particulate matter, making them easier to process and potentially allowing a higher sample-to-RB ratio. Furthermore, given the variability of fecal samples, testing real samples from target patient groups is essential to determine optimal assay conditions, as differences in ethnicity, diet, and microbiota may affect performance.
The results presented here showcase an opportunity to aid early detection of CTX and LT for clinical and water testing. Such a test could help ease the burden on physicians by providing a rapid, accurate tool for detecting CTX and LT, thereby saving time and costs associated with cholera and ETEC detection.