Open Access Article
weighting for MRI thermometry
Janusz H. Hankiewicz
ab,
Giacomo Parigi
c,
Zbigniew J. Celinskiad,
Yu Haod,
Allan D. Anguse,
Kristen Petersena,
Dorota Lachowicz
f,
Angelika Kmita
f and
Marek Przybylski
*fg
aCenter for the BioFrontiers Institute, University of Colorado Colorado Springs, 1420 Austin Bluffs Parkway, Colorado 80918, USA
bNational Institute of Standards and Technology, 325 Broadway St, Boulder, Colorado 80305, USA
cDepartment of Chemistry “Ugo Schiff”, Magnetic Resonance Center (CERM), University of Florence, Consorzio Interuniversitario Risonanze Magnetiche Metallo Proteine (CIRMMP), Via Sacconi 6, Sesto Fiorentino, 50019, Italy
dDepartment of Physics, University of Colorado Colorado Springs, 1420 Austin Bluffs Parkway, Colorado 80918, USA
eNational Society of Professional Engineers-Colorado, PO Box 848, Franktown, CO 80116, USA
fAcademic Centre for Materials and Nanotechnology, AGH University of Krakow, A. Mickiewicza 30, 30-059 Krakow, Poland
gFaculty of Physics and Applied Computer Science, Academic Centre for Materials and Nanotechnology, AGH University of Krakow, A. Mickiewicza 30, 30-059 Krakow, Poland. E-mail: marprzyb@agh.edu.pl
First published on 20th January 2026
Magnetic nanoparticles are used to map the temperature of the human body using MRI because their presence influences proton relaxation times. Especially at low fields, proton relaxation times exhibit strong temperature dependence, enabling precise determination of local temperature within a body with high resolution. As an illustration, we report temperature-dependent NMR and MRI investigations of agar gel with embedded Mn0.48Zn0.46Fe2.06O4 ferrite nanoparticles, evaluated as potential temperature–sensitive exogenous MRI contrast agents at low magnetic fields for noninvasive MRI thermometry. The nanoparticles consisted of an 8.5 nm magnetic core that were coated with a PEG shell, yielding a hydrodynamic diameter of 20 nm. Spin-lattice relaxation time (T1) profiles were obtained using a Fast Field Cycling NMR relaxometer. The
– weighted gradient echo MR images at 0.2 T were obtained in a temperature range of 23 °C to 50.5 °C, encompassing physiologically relevant values. Simulations of MR image intensity at 0.2 T, based on experimental T1 and
values, were carried out and compared with the corresponding temperature-dependent experimental images. Analysis of our measurements indicated that the temperature was determined with 2 °C accuracy.
Recently, we published results on the fabrication technology and comprehensive characterization of polyethylene glycol (PEG) coated mixed copper–zinc (CuZn) and manganese–zinc (MnZn) ferrite nanoparticles. When embedded in tissue mimicking agar gel, CuZn and MnZn ferrite nanoparticles reduce nuclear relaxation times T1, T2 and
of water hydrogen protons at 3.0 T magnetic field. The relaxation times are temperature dependent in the range of 5 °C to 50 °C, giving a necessary temperature contrast for tMRI at 3.0 T.3,4
MRI thermometry has become a powerful tool in assistance in MRI guided thermal surgeries.5,6 tMRI delivers real-time 2D temperature maps to assess the progress of thermal treatment of the organ in removing tumours, resection of epileptogenic zones, and correction of vascular malformation, to name a few. Monitoring the temperature simultaneously prevents damage to adjacent tissue and reduces the risk of health-related quality of life outcomes for patients.7 Details on different tMRI techniques can be found in reviews by Rieke and Odéen.8,9 Currently, thermally induced water proton resonance frequency (PRF) shift is used to monitor temperature during MRI-guided procedures.10 In practice, PRF fails due to magnetic field inhomogeneity, tissue movement, or the presence of adipose tissue.11
The basic idea behind using magnetic particles as exogenous temperature–sensitive contrast agents for tMRI is to generate temperature-dependent changes in intensity of T1, T2 weighted images by spin-echo sequences, or
weighted images by gradient echo sequences. Then the intensity of these images can be converted into absolute temperature.12
In this study, we investigate the nuclear relaxation properties and explore the possibility of using PEG-coated MnZn ferrite nanoparticles in a relatively high concentration as highly specialized temperature–sensitive contrast agents for tMRI at low magnetic fields. We note that MR scanners operating at low magnetic fields usually employ permanent magnets that do not use cryogens. Such magnets are much cheaper to maintain, provide more open space critical for surgical procedures, and can be made mobile.13,14 These new directions in MRI designs create new opportunities for diagnosing and treatment.
The availability of these low-field MRI systems increases the demand for contrast agents optimized for these lower field strengths. Superparamagnetic nanoparticles are well-suited as positive (T1 weighted) MRI contrast agents, exhibiting peak relaxation enhancement at around 0.2 T. For instance 8.5 nm commercially available carboxylic acid-coated iron oxide nanoparticles show peak at the body temperature (37 °C) at 0.23 T.1 The exploitation of ferrite nanoparticles is particularly motivated by the limitations of traditional gadolinium-based agents, which are contraindicated for patients at risk with renal impairment, by the strong temperature dependence of the relaxation enhancement, and because of their possible functionalization with therapeutic agents. Mn-doped ferrite nanoparticles can enhance their magnetic responsiveness and leads to superior magnetic hyperthermia therapy performance, while maintaining small particle size allows renal clearance and good biocompatibility.15 The development of new contrast agents that allow the performance of MRI-guided thermal ablations using low-field scanners would expand patient treatment options.
The technology we develop, however, is not aiming to replace standard MRI contrast agents delineating different tissues and blood vessels at body temperature (37 °C) such as Gd-based chelates. Unlike standard contrast agents, temperature–sensitive contrast agents do not require high values of relaxivity, but rather its strong temperature changes16 within the temperature range of interest.
For the NMRD measurements from 0.2 mT to 990 mT (0.01 MHz to 40 MHz hydrogen proton Larmor frequency), a Fast Field Cycling NMR (FFC NMR) relaxometer was used (Stelar, Spinmaster FFC-2000-1T Mede (PV), Italy) applying the standard FFC NMR technique.17–19 For the 1.45 mM oxide concentration, NMRD was conducted at temperatures of 10 °C, 20 °C, 30 °C, 40 °C and 50 °C. Measurements of samples with 0.73 mM and 0.36 mM oxide concentrations were conducted only at 20 °C. The error of temperature determination during measurements was ±0.5 °C. The switching time, i.e., the time needed to change the field during cycling, was 3 ms. The T1 values were obtained with an error smaller than 1% from the fit to a mono-exponential curve of the longitudinal magnetization decay/recovery data acquired at 16 delay times after switching the magnetic field to the desired value.
of the sample with 1.45 mM of oxide was measured using a pulse NMR spectrometer operating at 0.36 T, or 15.3 MHz for hydrogen protons, in the range of 5 °C to 50 °C, with 5 °C increments (ElLab, Poznan, Poland). The sample was placed in a standard 5 mm NMR tube (ATS Life Sciences Wilmad, Vineland, NJ)§§. The temperature of the sample was controlled by the thermoelectric Peltier device with accuracy of ±0.5 °C.
Spin–spin transverse relaxation time T2 was measured using the standard CPMG sequence with the following parameters: excitation pulse length
, delay array of 25 delays of 1 ms, generating 25 exponentially decaying spin-echoes after refocusing pulse
, repetition time = 250 ms, number of accumulations = 8.
Values of
were calculated from the apparent NMR line width ν1/2 (full width at half maximum, FWHM) using the formula:
![]() | (1) |
measurements by MRI at 0.2 T were conducted using a gradient echo sequence with multiple echo times (TE). An array of 14 TE ranging from 5.75 ms to 35.0 ms was used. To avoid T1 weighing, the repetition time (TR) was set to 100 ms. Other parameters were: flip angle (FA) = 40°, matrix = 64 × 64 pixels, field of view (FOV) = 30 × 30 mm2, number of accumulation (NA) = 16, acquisition time for individual TE = 102.4 s, total acquisition time for all 14 TE times = 23 min and 54 s. Because of such a long total acquisition time and constant temperature sweep within the temperature cell,
measurements by 0.2 T MRI were conducted only at one point when the sample reached the stable thermal equilibrium of 23 °C.
Fig. 1(b), delineated by the dashed green rectangle, shows the RF parallel LC resonator for imaging small objects. This small resonator was used in the study to increase the filling factor and improve the signal-to-noise ratio (SNR). The resonator is made of a standard solenoid and tuning variable capacitor dedicated to nuclei excitement and receiving signals. This is tuned to the 1H resonance frequency of 8.5 MHz. The solenoid consists of 8 loops of 1 mm enamelled copper wire with a length of 15 mm and an inner diameter of 63 mm. The resonator was coupled and matched to 50 Ω transmission/receiving lines using a single coupling loop and a variable capacitor, respectively.
To acquire temperature-dependent images, the phantom was placed in a dedicated MRI-compatible cell. Fig. 2(a) shows a diagram of the cell. The cell comprises a cylinder made of aerogel insulation (Aspen Aerogels, Northborough, MA, USA) and a plastic vial filled with fluorinert (FC-40, 3M Science. Applied to Life, Maplewood, MN). FC-40 is a hydrogen-free electronic liquid that serves as heat energy storage without disturbing MR images. The phantom consists of three glass 10 mm tubes filled with pure agar, agar with 0.73 mM, and an agar with 1.45 mM of oxide. The phantom is inserted in the vial with FC-40. A miniature thermocouple connected to a battery-powered thermometer (Fluke, model 116 HVAC, with a type-K thermocouple, Everett, WA, USA) is immersed in the FC-40 liquid, 40 mm from the MRI axial slice. The tip of the thermocouple is sufficiently far away to avoid potential imaging interference.23
Imaging the phantom was conducted during free and uniform cooling from 51 °C to ambient room temperature (23 °C). Individual scans started when the temperature reading reached a whole number, e.g. 50 °C, 49 °C, and so on. Because the phantom takes a long time to reach thermal equilibrium (more than 2 hours), there is a gap in MRI temperature measurements at 23 °C and 24 °C points. In the course of cooling, the phantom temperature changed over the duration of the scan by ±0.4 °C for the first four scans and remained below ±0.3 °C for the other scans.
Gradient echo MR imaging was conducted with the following parameters: TR = 100 ms, TE = 5.75 ms, FA = 40°, FOV = 30 × 30 mm2, matrix of the axial slice = 64 × 64 (in-plane spatial resolution = 0.47 mm per pixel), slice thickness = 5 mm, number of accumulations = 8, acquisition time = 51 s.
Due to the long minimum available TE = 23.0 ms of the 0.2 T scanner, we were unable to conduct imaging with a standard spin-echo (SE) sequence. Contrary to gradient echo recalled sequence, the spin-echo sequence is known to be more robust and less affected by susceptibility and chemical shift artifacts.24,25 However, due to a very short relaxation time
of the agar sample doped with 1.45 mM oxide concentration (less than 8 ms, see discussion in Transverse nuclear relaxation section below), the MR signal after 90° excitation radiofrequency (RF) pulse of the sequence was very low after a 180° RF pulse. Consequently, the signal after image reconstruction within the 1.45 mM oxide sample was not distinguishable from the noise. See an example of the SE image in Appendix A section on Fig. 13.
In the practical application of nanoparticles as temperature–sensitive contrast agents for tMRI, one needs information on the thermal dependence of nuclear relaxation times at a given magnetic field. Fig. 5(a) shows the percentage of T1 changes relative to the value at 10 °C for selected magnetic fields taken from data presented in Fig. 4(b). The percentage change determines the contrast of T1 weighted images and directly affects the accuracy of the determination of the temperature. This can be used as an indicator for the optimum magnetic field magnitude. Among the fields collected in the NMRD profiles, the 0.007 T field shows the largest T1 change, which is shown more clearly in Fig. 5(b) as a local maximum. In this figure, the slope values from Fig. 5(a), normalized to 10–50 °C temperature range, are presented as a function of the magnetic field.
![]() | ||
| Fig. 5 Temperature and magnetic field dependence of T1 for water hydrogen protons in agar gel with embedded MnZn iron oxide in concentration of 1.45 mM. (a) Percentage changes of T1. At various (b) value of the slopes from Fig. 5(a) normalized to 10–50 °C temperature range. | ||
There are only a few MRI scanners available close to 0.1 T, where minimum of T1 occurs (see Fig. 4(b)), such as the 0.064 T MRI scanner (Hyperfine, Guilford, CT). Our research group had access to a 0.2 T scanner, which operates slightly above the field of a T1 minimum.
![]() | (2) |
As previously noted,31,32 the sharp decrease in relaxivity with increasing temperature is primarily ascribed to the increase in the diffusion coefficient: as the temperature rises from 20 °C to 50 °C, the diffusion coefficient almost doubles, and consequently the diffusional time τD nearly halves, leading to a corresponding reduction in the relaxation rates at frequencies near the peak position. In lower fields, the temperature dependence is also influenced by the Néel correlation time τN, which is likewise expected to decrease as the temperature increases.
For all temperatures, the relaxivity profiles shown in Fig. 10 exhibit three common features: (1) a low-field plateau, (2) a maximum around 0.15 T, and (3) a rapid decrease above 0.15 T. At very low magnetic fields, the fluctuating part of the magnetic moment of the superparamagnetic particle, µNP, aligns along an easy magnetization direction and randomly jumps from one easy direction to another with a characteristic time called Néel correlation time.33 As the applied magnetic field increases, the non-zero time average part of µNP, called the Curie-spin magnetic moment, aligns with the direction of the magnetic field and increases up to saturation. The observed maximum of r1 around 6 MHz (0.15 T) originates from two mechanisms: the gradual increase of the Curie-spin magnetic moment with the applied magnetic field and the subsequent decrease of the spectral density function at frequencies on the order of, or larger than, the inverse of the translational diffusion correlation time (τD). As the magnetic field increases, the relaxivity decreases.34–36
mapping MRI sequence and simultaneous temperature change within the phantom holding cell,
measurements at 0.2 T were conducted only at 23 °C, when the sample reached thermal equilibrium with the magnet room. The temperature dependence of
was thus observed at 0.36 T, as a guide for planning of
weighted gradient echo MRI protocols, and for the image intensity simulations. Fig. 7 shows the transverse relaxation time
measured at 0.36 T in the range between 5 °C to 50 °C and the value measured at 0.2 T and 23 °C. Additionally, the figure also presents the temperature dependence of the T2 relaxation, as useful information for possible spin-echo imaging. One can appreciate that the 23 °C value of
for the concentration of 1.45 mM at 0.2 T fits nicely into the 0.36 T temperature data trend, justifying the use of
measured at 0.36 T data for the following 0.2 T MRI experiments and signal simulation, as shown in the subsequent section.
of samples in the phantom, the images were inherently
weighted and, as a result, T1 weighted temperature-dependent studies at 0.2 T were not possible. Therefore, we focused on testing whether
weighted images provide temperature-dependent contrast useful for tMRI. We benefited from the very short T1 in another way, as it enabled fast scanning with multiple accumulations, leading to an improved SNR and better temporal resolution of the method.
Fig. 8 shows representative gradient recalled echo images at three selected temperatures. Images are T1 weighed for pure agar gel. Images of MnZn iron oxide solutions in agar are
weighted (T1 is shorter than 20 ms and
increases from 4.0 ms to 7.7 ms in the temperature range from 23 °C to 50.5 °C) and their brightness increases with temperature.
The image intensity was analyzed within a selected circular region of interest (ROI) consisting of 274 pixels. Details of this analysis are presented in Fig. 9. Fig. 9(a) shows the thermal dependence of the mean value of the image intensity. The inset in the bottom-right corner shows the positions of four ROIs corresponding to pure agar, 0.73 mM and 1.45 mM oxide concentration in agar, and the background noise.
By assumption, the measured noise shown in Fig. 9(b) originates from Johnson–Nyquist thermal noise in both the real and imaginary parts of the complex MRI signal. This noise is squared and summed, at which point the distribution is χ2 of order two, since two Gaussian distributions have been squared and added together. Once one takes the square root of this noise, one obtains a Rayleigh distribution for the noise in a non-signal part of the image.
From Cárdenas-Blanco,37 eqn (17), we have
where mR is our measured noise in the Fig. 9(b) and σG is the standard deviation of the desired input-referred Gaussian noise. We find that the measured variation in this noise is somewhat larger across the 25 °C temperature range of the experiment, assuming the expected values from
, where kB is Boltzmann's constant, R is the effective resistance, Bw is the bandwidth, and T is the temperature in Kelvin. Our working hypothesis is that this extra variation arises from a thermal detuning of the parallel resonant circuit shown in Fig. 1(b), with the center frequency shifting away from the desired signal frequency.
Fig. 9(b) also shows a linear regression fit of experimental noise with Pearson's r = 0.7784 and R2 = 0.6059 (solid black line) and estimated values of σG (solid red line). Using these estimates of σG, we computed measured average values MAVE of the vial signals in SNR form as MAVE/σG. While these SNR values were greater than 5, we proceeded to calculate actual signal-to-noise values, A/σg using Henkelman.38
These final SNR values are shown in Fig. 9(c).
Fig. 9(d) presents the results of our major objective: the thermal dependence of the SNR of water hydrogen protons
weighted images due to the presence of ferromagnetic particles. The figure includes a linear fit to experimental points as well as 95% prediction bands. The prediction bands were used for the determination of accuracy of the method at 38 °C, the temperature slightly above the human body physiologically relevant temperature of 37 °C.39 The accuracy Δt at given temperature t in degrees Celsius is defined by the range, within which the temperature determination from MR image intensity is statistically not distinguished from temperatures below (−) and above (+) using 95% prediction bands. Details of the accuracy analysis are presented in Appendix B, and the analysis results are shown in Table 1.
weighted images, of samples with 0.73 mM and 1.45 mM concentration of MnZn iron oxide in 1% (w/w) agar gel
| Concentration (mM) | Slope (°C−1) | Accuracy (°C) |
|---|---|---|
| 0.73 | 0.041 | −7.2/+7.6 |
| 1.45 | 0.113 | ±2.2 |
As shown in Table 1, the slope values directly relate to the accuracy of temperature determination. For the two concentrations studied here, the steeper slope corresponding to 1.45 mM concentration gives better accuracy. Poor accuracy of −7.5 °C/+6.4 °C renders the 0.73 mM concentration too small for practical use. Doubling the concentration to 1.45 mM delivers a much better accuracy (−2.2 °C/+2.3 °C).
![]() | (3) |
For PEG coated nanoparticles, the presence of the shell increases the distance of the closest approach, as shown in Fig. 9.
The experimental r1 profiles were fit to the Roch–Müller–Gillis model with d and single large spin (S) resulting from the coupling of all electron spins in each nanoparticle as common parameters.42 Solid lines in Fig. 6 represent the best fit with parameters shown in Table 2. Details of the model are provided in the Appendix C section. Water diffusion at 20 °C and 50 °C used for the best fit was fixed to the experimental results reported earlier.4 The best-fit value of the distance of closest approach equals 7.0 nm and is intermediate between the value of 4.75 nm of the magnetic core and 7.5 nm of the shell experimentally determined by AFM. Néel correlation times of a few nanoseconds and an S value approaching 10
000 are obtained similarly to other iron oxide nanoparticles of similar size.1,43 The good quality of the fit across all profiles acquired in this wide temperature range also validates the Roch–Müller–Gillis theory for manganese-doped ferrite nanoparticles.
| t (°C) | 10 | 20 | 30 | 40 | 50 |
|---|---|---|---|---|---|
| D (10−9 m2 s−1) | 1.43 | 1.97*& | 2.52 | 3.16 | 3.86*& |
| τN (10−9 s) | 4.54 | 3.54 | 2.92 | 2.35 | 1.83 |
| d (nm) | 7.0 (7.5)# |
| S | 8900 |
| C (mM) | 1.45#& |
weighted experimental data was analyzed against
weighted MRI signal intensity simulations. For comparison purpose we calculated the MRI signal using eqn (4).44 As explained in Material and method section, due to limitations of the MRI scanner, we were unable to obtain experimental data for T1 weighting.
![]() | (4) |
µB is Bohr's magneton for protons, kB is the Boltzmann constant, T is temperature. For the studied temperature range of 23 °C to 50.5 °C and a magnetic field of 0.2 T, a weighting factor k[H] ≅ 1 is obtained. FA is the flip angle (rad), TR is repetition time, and TE is echo time; these are MRI sequence parameters set by an operator using guidance from relaxation measurements.
Experimental relaxation values of T1 and
, and values of the TR, TE and FA parameters used in the MRI sequence to achieve the
weighting contrast are given in Table 3. Note that values of T1 are from 0.2 T NMRD, while values of
are obtained from 0.36 T NMR studies. Results for TR/TE = 35.0/5.75 ms are slightly T1 weighted at higher temperature when T1 reaches a value of 20.4 ms.
at 0.2 T. Parameters used for weighted MRI signal intensity experiment and simulations. Experimental values of T1 and
used are presented in Fig. 4c and 6, respectively
Results of the simulations, and experimental
weighted MRI, are shown in Fig. 11(a). The intensities of
weighted images obtained from the simulation and from the experiments are increasing with temperature as the particles' magnetization decreases (see Fig. 3) and are in good agreement with each other. The paired t-test shows that at the level of 0.05, the difference between signal intensity for experimental
weighting and simulated
weighting is not significant.
The values of simulated signal intensities (SI) presented in Fig. 11(a) were used for the calculations of temperature MRI contrast between 30 °C and 50 °C using eqn (5).45
| Temperature contrast = SI30 °Co − SI50 °Co | (5) |
The results of contrast values for
weighted experiments and
weighted simulations are in good agreement, −0.09 and −0.10, respectively (see Fig. 11(b)).
Since the acquisition time of the current
weighted image is relatively long, we simulated the possible shortening of TR to achieve better temporal resolution of the temperature determination. Shortening TR from 100 ms to 35 ms lowers the MRI signal by only 1% and sacrifices the contrast by 10%, while simultaneously decreasing the image acquisition time from 6.4 s to 2.2 s. Shortening acquisition time is critical because during MRI-guided procedures, temperature information must be obtained in real time.
The proposed method of temperature measurement intrinsically suffers from the dependence of MRI temperature contrast on magnetic particle concentration. This issue can be partially alleviated by using thin filaments made of hydrogen-rich polymers with embedded particles. Although more invasive, the method can provide an MRI-compatible means of measuring temperature in one direction by thermal contact between a filament and the tissue.
As mentioned in the Introduction, PRF fails in areas rich with adipose tissue due to the complexity of the proton NMR line48 and the small temperature-dependent chemical shift.49,50 There is growing evidence that patients with obesity are at high risk of developing tumours51 and will potentially require surgery using MRI-guided thermal ablations in areas of high fat content.52 As temperature monitoring by PRF in such cases will not be reliable, we envision injections of temperature sensitive particles in fat tissue at the surgery location. Then
maps will be acquired of the region of injection and compared to the temperature calibration map obtained earlier from fatty phantom with particles injected in the same amount to convert
maps to temperature maps as temperature changes.
MRI temperature contrast, permitting non-invasive temperature measurements. Signal intensity of
weighted images obtained with the standard gradient echo sequence at 0.2 T near the physiological temperature of the human body (38 °C) delivers the accuracy of temperature determination of ±2.2 °C. Since the T1 time of free water protons in the agar phantom doped with nanoparticles is very short, image acquisition can be performed in 6.4 s. A shorter acquisition time, leading to better temporal resolution for the method, is possible by reducing TR, with some sacrifice in SNR and contrast.
As seen in Fig. 8, due to the long minimum TE time (5.75 ms), the standard gradient echo recalled sequence available on a 0.2 T scanner, delivers images from an aqueous phantom with embedded high concentration of MnZn nanoparticles with relatively low value of SNR. Use of high-end gradient equipment with much shorter TE = 2.0 ms as shown with the simulation will slightly improve SNR. Imaging the tissue with high concentrations of magnetic nanoparticles would benefit from much higher SNR achieved with the newly developed sequences with minimal TE such as Ultrashort Echo Time (UTE) or Zero Echo Time (ZTE).53 Theoretical analysis shows new ways of achieving selective T2 contrast using UTE sequence.54 Positive T1 weighted contrast obtained experimentally with UTE was reported in tissues with targeted iron-oxide nanoparticles at 3.0 T magnetic field.55
relaxometry in the presence of highly concentrated iron-oxide nanoparticles (close to concentrations presented in this paper) labelled cels at 3.0 T using 3D UTE was also published.56 Its prohibitively long acquisition time of 39 minutes can be shorten by reducing acquisition space to 2D. However, the practical exploit of our observation on the maximum of the water proton r1 relaxivity near 0.15 T in a phantom with high concentration of MnZn magnetic particles, and to use it for the temperature mapping in low-field scanners is currently limited due to the absence of the implementation of very short echo-time sequences at low fields.
In the fit analysis the following parameters were optimized: the distance of the closest approach, d, the single large spin (superspin) in each nanoparticle, S, resulting from the coupling of all electron spins in a nanoparticle, the diffusion coefficient, D, and the Néel correlation time, τe.
The other parameters entering the above equation are the Avogadro's constant NA, the molar concentration of the nanoparticles M (in mol L−1), the proton gyromagnetic ratio, γI, the proton Larmor angular frequency ωI, the Bohr magnetron µB, the electron's g-factor ge, the Langevin function
, where
, the electron Larmor frequency
, the translational diffusion time
, the heuristic parameters P and Q, and the spectral density functions JA(ωI, τD) and JF(ωI,τD,τe),
![]() | (A1) |
The P and Q parameters account for anisotropic effects at low fields, and when the anisotropic energy is much larger than the Zeeman energy, for frequencies smaller than τ−1c, the values of P and Q are equal to 0 and 1, respectively. In such cases, the quantization axis of the electron magnetic moment is fixed along the easy axis of magnetization. In our analysis we set P = 0 and Q = 1.
The spectral density functions are
![]() | (A2) |
![]() | (A3) |
| This journal is © The Royal Society of Chemistry 2026 |