Studying photoluminescence (PL) is essential for understanding and engineering light-emitting nanomaterials. Heterogeneous samples, such as thin films and microstructures, require characterization of the spatial distribution of PL intensity, spectra, lifetimes, and photon correlations at the microscale under variable temperatures. Here we describe a cryo-optical setup for micro-PL imaging, spectroscopy, lifetime, and photon bunching measurements that addresses this need. The system combines a home-built wide-field microscope with a commercial closed-loop helium cryostat, a spectrograph equipped with an sCMOS camera, and a Hanbury-Brown and Twiss interferometer (HBT-interferometer). Optical layout and components of the setup are described, and its performance is tested on samples of CsPbBr3 nanocrystal superlattices. This work aims to demonstrate that it is feasible to develop an advanced microscopy and spectroscopy setup suitable for material scientists focused on developing novel materials, even if they initially lack extensive experience in advanced optical methods.
Parallel to these developments, the buzz of quantum technology has stimulated research on colloidal quantum dots as solid-state light sources for single-photon emission and lasing, among other functions.13,14 Colloidal quantum dots have significantly broadened the expertise required in materials science research, motivating chemists to move beyond synthesis toward detailed spectroscopic characterizations.15 This created a demand for methods, such as cryogenic PL microscopy and spectroscopy, typically used in solid-state physics and quantum optics,16 to be applied to materials synthesized in the wet lab, such as colloids, thin films, and various microstructures. Access to specialized setups for studying quantum emitters often relies on collaborations, which can limit the throughput of screening the samples and the discovery of interesting properties. Recent advances in commercially available equipment, such as closed-cycle cryostats with optical access and user-friendly control interfaces, have made advanced low-temperature spectroscopy more accessible. This enables synthetic and materials-focused research groups to expand characterization capabilities in-house, bridging the gap between synthesis, structural analysis, and optoelectronic measurements. These developments have created an intersection between chemistry, physics, and biology, as new PL materials obtained by chemical synthesis are studied using microscopy techniques from the life sciences and quantum optical tools from physics.
One example that is particularly interesting to us is metal halide perovskite nanocrystals, characterized by their bright emission and unusual electronic structure.17–19 The characterization of perovskite nanocrystals as single-photon emitters has become increasingly common.20–25 Besides single particles, their self-assembly into ordered microscopic structures (variously referred to as superlattices, supercrystals, or mesocrystals) and resulting collective phenomena such as superfluorescence26,27 and superradiance,28,29 are of high scientific interest. Investigations of these curious properties and correlation of their origin with the sample morphology require PL studies of individual microscopic objects at cryogenic temperatures. Thus, we describe the layout and implementation of a cryo-optical setup for PL microscopy and spectroscopy, suitable for studying questions like these and generally applicable to the characterization of emerging luminescent nanomaterials in the ca. 4–320 K temperature range. The setup is built around a commercial cryostat, spectrograph, and camera, incorporating an HBT-interferometer. It allows spatial resolution of steady-state photoluminescence, spectrum- and time-resolved photoluminescence and spectrum-resolved correlation measurements within a single cooling cycle. The performance of the setup is tested on a sample of 8 nm CsPbBr3 perovskite nanocrystal superlattices, a dilute nanocrystal film, and a solution of Rhodamine.
![]() | ||
| Fig. 1 Schematic of the cryo-optical setup with components indicated: (1) cryostat with a breadboard and a shroud hosting a sample and a low-temperature objective (O), (2) microscope optical path (IP – image plane), (3) spectrograph and camera for spectroscopy, imaging, and spectral selection, and (4) HBT-interferometer. For abbreviations of optical components and variables, see Table 1 and eqn (1)–(3), respectively. | ||
![]() | ||
| Fig. 2 (a) Photograph of the setup with overlay graphics highlighting the modules (1–4, as in Fig. 1), beam path, and optical elements (L5–L7 are specified in Table S1 in the SI). IP are image planes, MM is the moving mirror to select the exit port, IP1 – intermediate image plane, IP2 is the slip plane which is conjugated to IP3 (exit slit), IP4 – the camera chip plane, IP5 and IP6 are image planes at the SPADs. (b) Side view of the cryostat shroud enclosing the microscope objective, (c) the close-up photograph of the objective with 405 nm laser (not visible) exciting a green PL from a thin film sample, (d) excitation spot imaged as a PL of from a uniform film of dried yellow highlighter inks on a thin glass slide, (e) image of 2.11 µm SPHERO™ Fluorescent Particles used for length scale calibration. | ||
| Component | Part (supplier) | Selected specifications |
|---|---|---|
| Excitation light source | LDH-D-C-405 (PicoQuant) | Pulsed diode laser, 405 nm, nominal minimum pulse width 49 ps |
| Fiber | SM300 (ThorLabs) | Single-mode, 320–430 nm, ca. 2.4 µm mode-field diameter at 350 nm, 4 m long |
| Fiber coupler | PAF2S-5A (ThorLabs) | f = 4.6 mm, 350–700 nm |
| Lens (L1) | LA4236-A (ThorLabs) | f = 125.4 mm, 1″ diam., uncoated, UV fused silica (UVFS), plano-convex lens |
| Dichroic mirror (DM) | DMLP425R (ThorLabs) | 25 mm × 36 mm, 425 nm cut-on |
| Longpass filter (LPF) | FELH0425 (ThorLabs) | 25.0 mm diam., 425 nm cut-on |
| Cryostat | attoDRY800xs (Attocube) | Closed-cycle helium cryostat, RT & LT Objective transmission shroud |
| Objective (O) | LT-APO (Attocube) | Clear aperture 4.7 mm, f = 2.87 mm, working distance 0.65 mm, NA = 0.82, apochromatic range 465–590 nm |
| L2 | LA4579-AB (ThorLabs) | f = 300 mm, 1″ diam., UVFS plano-convex lens, anti-reflection coated 400–1100 nm |
| L3 | LA4148-AB (ThorLabs) | f = 50 mm or 100 mm, (see text), 1″ diam., UVFS plano-convex lens, anti-reflection coated 400–1100 nm |
| L4 | LA4102-AB (ThorLabs) | f = 200 mm or 100 mm (see text), 1″ diam., UVFS plano-convex lens, anti-reflection coated 400–1100 nm |
| L5, L6 | LA4936-AB (ThorLabs) | f = 30 mm, 0.5″ diam., UVFS plano-convex lens, anti-reflection coated 400–1100 nm |
| L7, L8 | LA4647-AB (ThorLabs) | f = 20 mm, 0.5″ diam., UVFS plano-convex lens, anti-reflection coated 400–1100 nm |
| Spectrograph | Kymera 328i (Andor) | 2 exit ports (to camera and intensity interferometer), 4-place turret with 1 mirror and 3 gratings, Solis software |
| Camera | ZL41 Wave 4.2 (Andor) | sCMOS camera, 4.2 megapixel sensor (6.5 µm pixel size) with 2048 × 2048 active pixels, 13.3 mm × 13.3 mm |
| Beam splitter (BS) | BSW10R (ThorLabs) | Non-polarizing, 50 : 50, 25 mm × 36 mm, 1 mm thick, UVFS, coating 400–700 nm |
| Single-photon avalanche diode (SPAD) | MPD-050-CTC (MPD/PicoQuant) | 50 µm active area, <100 dark counts, 27 ps jitter |
| Time-tagger (TT) | Time Tagger Ultra (Swabian Instruments) | 4 channels, 8 ps time resolution (RMS jitter), 2.1 ns dead time |
The conditions for constructing this setup were determined by the targeted size of the illumination spot on the sample, as well as practical limitations such as the available light source and space on the cryostat breadboard and the optical table. Neglecting light diffraction and using the thin lens approximation, the main equations describing the optical parameters for the excitation part of the setup from fiber to objective are:30
![]() | (1) |
![]() | (2) |
![]() | (3) |
The variables are shown in Fig. 1. Briefly, eqn (1) is the lens equation for the objective (O), eqn (2) is a geometric relation between focusing angles of the objective (α) and L1 (ψ), and eqn (3) defines the diameter of the wide-field illumination spot on the sample (dspot) as a function of the incident beam spot size on the L1 (dL1) and focal distances of the objective (fObj) and L1 (fL1). The final equality in the eqn (3) is obtained by solving eqn (1) for a or b (where a is the distance from image to the rear focal plane of the objective, b is the distance between fL1 and the front focal plane of the objective) and substituting it in eqn (2).
The light source was a 405 nm pulsed laser diode. Its output was coupled into the optical fiber at one end of the laboratory and then brought onto the cryostat breadboard through a 4-meter-long optical fiber. From there, light propagated in free space. The choice of using a fiber over free space was motivated by space constraints (laser positioned away from the setup) and the desire to achieve a uniform beam profile at the sample. The spot size dL1 incident on the focusing lens L1 could be estimated using the specifications of the fiber coupler (outcoupling beam size and the divergence angle, under the assumption that the fiber is in the focus of the lens) and the distance between the fiber coupler and L1, or measured directly with a beam profiler or calibrated camera, if available.
One needs to ensure that the beam is collimated to use all the equations. However, in practice, it is not necessary. The result is judged by the size of the excitation spot at the sample. For the focusing lens L1, we chose a lens with a focal length of 125.4 mm, mounted on a flip holder approximately 9.5 bolt holes before the tip of the objective (Fig. 2, ca. 240 mm on a metric breadboard). The flip holder of L1 allows it to be retracted from the collimated beam so that the setup could be operated with a focused excitation beam. The collimated beam exiting the fiber coupler has a circular profile with a waist diameter of around 600 µm, as measured with a beam profiler (Fig. S2 in the SI). Using these values and the eqn (3), the calculated dspot = 13.7 µm. The size of the excitation spot on the sample was quantified by imaging fluorescence from a uniform dried yellow highlighter ink film on a thin glass slide (Fig. 2d). Fitting the image with a 2D Gaussian yielded a nearly circular intensity distribution, with a spot size at the sample of 16.5 µm × 18 µm (waist diameter, Fig. S3), in reasonable agreement with the calculation. The scale bar calibration was performed with Nile red SPHERO™ Fluorescent Particles of mean diameter 2.11 ± 0.2 µm, which are shown in Fig. 2e. Error propagation led to a scalebar uncertainty of 11%. Upon close inspection of the image of the excitation spot (Fig. 2d), some structure in the intensity distribution can be observed. It is good to remember that due to the high coherence of the laser light, even minor imperfections in the optics and the presence of dust produce interference and diffraction-related structures in the excitation spot. The origin of that structure was attributed to the optics after the fiber, i.e., the fiber coupler and L1, based on a visual inspection of the 405 nm beam projected onto paper. Without changing the objective, the size of dspot can be changed by either changing dL1 (for example, by replacing a fiber coupler with an alternative configuration to collimate the output of the fiber) or fL1 (by replacing the lens). In either case, one needs to check that the beam enters the objective without clipping its clear aperture.
In the detection path, the light emitted by the sample is collected through the same objective, which serves as a collimating lens for PL (Fig. 1, 2b and c). The collimated PL is then focused with a tube lens L2 to form an image on the first image plane (IP1 at the fL2). The magnification at IP1 is MObj+L2 = fL2/fObj = 104.5. An iris diaphragm is placed at IP1 to select a region of interest from the entire image, as well as to provide a diagnostic for the correct focusing of the second image plane (IP2) on the camera. The light passing through the iris is then collected by a relay lens doublet (L3 and L4). L3 is a collimating lens placed at a distance fL3 from IP1, followed by the focusing lens L4 to obtain a further magnified image on IP2, an image plane at the entrance slit of the spectrograph. The magnification by relay lenses provides an additional factor of ML3+L4 = fL4/fL3, which was varied between 4× and 1×, depending on the choice of lenses (see Table 1), for a total magnification at IP2 of approximately 400× or 100×, respectively.
IP2 is imaged by the spectrograph with unity magnification on the plane of the exit slit as well as on the camera chip (IP4). Here, it is important to consider that the spatial resolution of the microscope is fundamentally limited by light diffraction and optical aberrations. So, depending on the camera pixel size, after a certain limit, increasing the magnification does not lead to an increase in resolution, but rather leads to useless spreading of the image over too many pixels. Due to this spread, the PL signal in each pixel goes down as the inverse square of the magnification. A compromise between having high resolution and a high number of counts per pixel (based on Nyquist criteria) is to have around 3 pixels per the full width of half maximum (FWHM) of the instrumental response function (IRF) of the microscope (the width of spatial distribution produced by a point source of light, e.g., individual nanocrystal). Then the magnification in the IP4 plane is M = 3 × camera pixel size/IRFfwhm. Diffraction-limited FWHM of IRF is approximately 0.61λ/NA or, in our case, ≈0.4 μm for NA = 0.82 and λ = 550 nm. Thus, the optimal magnification at the camera plane should be around M = 3 × 6.3 µm/0.4 µm ≈ 50×. It is important to note that there are theoretical considerations and practical implementations, each with its own limitations, and while they do not always coincide, understanding the theory remains essential.
To couple light into the HBT interferometer for photon statistics and PL lifetime measurements, we used lenses L5–L8. L5 was mounted on a flipping holder in front of the spectrograph entrance slit to focus the light and record the spectrum of the whole image (as further elaborated in the Performance section), so that when the spectrograph directed the beam toward the HBT interferometer, a focused output was delivered onto L6. L6 collimated the beam, which then passed through a beamsplitter, while L7 and L8 refocused it onto the SPAD sensors with an estimated spot size 30–40 µm. It should also be noted that the SPAD pixel size must be equal to or larger than the diffraction limit, otherwise the signal is lost. Initial alignment of the beam through the spectrograph, beamsplitter, and SPADs was performed with a red HeNe laser. The SPADs were mounted on XYZ translation stages to allow fine alignment of the focused beam from the sample during measurements.
Several programming efforts supported the setup preparation. Before purchasing optical components (fiber, fiber coupler, lenses L1–L4), calculations were first carried out on paper and then verified by encoding the geometrical optics parameters in Python (v3.8.10 on Spyder v5.5.5); the script is available on GitHub.31 During setup construction, we also used FRED Optimum and Winlens Basic to practice optical engineering and to verify ray tracing. Once the setup was completed and operational (see Performance section), the Andor spectrograph and Swabian time tagger were integrated into a single Python-based workflow to streamline data acquisition and processing.32 This code, available in the SI, relies on manufacturer APIs and automates the sequential control of grating movement, initiation of TCSPC measurements, and data saving. All resulting data can be analyzed using in-house Python scripts (also available in the SI and on GitHub), which provide immediate graphical visualization, as demonstrated in the following section.
As stated above, we used a 405 nm light source for the excitation. The 405 nm excitation is a common choice for lead halide perovskite such as CsPbBr3, as it matches or falls above the absorption band edge across virtually an entire quantum-confinement range for this compound in the form of colloidal nanomaterial (from ca. 400 nm lowest excited state in <1 nm clusters,39 to ca. 428 nm in 2-monolayer nanoplatelets,40 486 nm in ca. 5–6 nm nanocubes,41 502 nm in 7–8 nm nanocubes,36 508 nm for ca. 20 nm nanocubes,42 and eventually to ca. 550 nm of the bulk CsPbBr3
43).
With a ca. 20 µm excitation spot, the maximum excitation fluence at the sample was estimated to be 1 µJ cm−2 (losses include ca. 30% transmission of the incident laser power by the single-mode fiber), with the lower limit tunable down to ca. 0.01 µJ cm−2 by a combination of the laser's power settings and neutral density filters. The laser power loss arises primarily at the initial fiber-coupling step, and depends on the laser mode quality, the optical components, and the stability of the alignment. In the setup, the laser mode is not perfectly circular, which limits the coupling efficiency into a single-mode fiber. We used the single-mode fiber to improve the beam profile at the cost of power loss. If the beam profile is not critical, a multimode fiber could be employed to reduce these losses.
The 0.01–1 µJ cm−2 fluence range allows PL experiments on 8 nm CsPbBr3 nanocubes with an excitation in the range of ca. 0.02–5 × 10−4 excitations per nanocrystal, using a size-dependent absorption cross-section of 8 × 10−15 cm2 at 405 nm.44 Compared with excitation conditions reported in the literature, the 0.01–1 µJ cm−2 range is suitable for studies of single-excitation superradiance29,45–47 as well as routine PL imaging and spectroscopy. For studies of light emission at very high excitation intensities, ca. 10–105 µJ cm−2, a regime reported in the studies of superfluorescence, the fluence can be increased by reducing dspot (for example, focusing it by a factor of 3.3 will increase the excitation fluence by a factor of 10) or by using a more powerful laser source. It is worth keeping in mind that, for studies of superlattices sized at several microns, intense and tightly-focused dspot below the dimensions of the superlattice may result in the highly non-uniform excitation of nanocrystals. One possibility to mitigate that could be the implementation of a flat top beam.48
Fig. 3 summarizes the steady-state µ-PL properties of a single ca. 6.7 × 7.3 µm superlattice at 5.6 K (the temperature at the top of the piezo-stack, near the sample) collected with the described setup at the excitation fluence of 0.3 µJ cm−2. Fig. 3a and b shows µ-PL images of the superlattice at room and cryogenic temperatures, selected as a region of interest with an iris. Fig. 3b, c, e and f illustrates spectra from the region of interest as imaged on the camera and obtained from the lines 1 pixel wide corresponding to the location in the sample marked with coloured dots in Fig. 3a and d. This superlattice shows uniform PL spectra across the entire structure, both at room and cryogenic temperatures. The narrowing and shift of the PL spectra to longer wavelengths with decreasing temperature is typical for CsPbBr3 semiconductor, both in bulk49 and nanocrystals, and is not necessarily a signature of collective effects.37
The switching between imaging and spectral modes of the spectrograph was done by switching between a mirror and a grating (400 l mm−1, blaze wavelength 550 nm) in the spectrograph. To collect PL spectra with a wavelength resolution it is necessary to close the spectrograph slits. This operation transforms the analyzed area of the image into a vertical line that is dispersed by the grating into its spectral components, enabling spectro-spatial correlation across the superlattice (in “spectral imaging mode”). To measure the spectrum from the entire image instead of the spatially resolved line, lens L5 was mounted on a flipping holder before the spectrograph (Fig. 2), allowing the sample image to be focused on the slit (“focused mode”).
The collected data consists of .tif files for the PL sample images (Fig. 3a and d) and .asc files containing the spectral information (Fig. 3b and e), including the metadata about the measurement conditions, as recorded by the spectrograph software. These files have been processed using a custom-written Python script to create a convenient spatio-spectral representation, i.e., the entirety of Fig. 3. The data files and scripts are provided in the SI.
The capability to focus the image on the spectrograph slits is further necessary to collect spectrally resolved time-resolved PL measurements, as a spot that fits the detector's active area is required, as discussed earlier. As shown in Fig. 2, the HBT-interferometer for time-resolved PL and correlation measurements consists of two SPADs positioned at 90° to each other after a 50
:
50 beamsplitter placed at a 45° incident angle to the entering light beam. This detection configuration is commonly known as the Hanbury Brown and Twiss (HBT) interferometer.50,51 It enables single photon detection and timing for time-correlated single photon counting (TCSPC), together with second-order correlation (g(2)) measurements. The photons arrival signal generated by the SPADs is handled by a time-tagger (a fast electronic event counting device), allowing for a temporal instrument response function of the apparatus limited by the width of the laser pulse (approximately 200 ps). This configuration is compatible with further extensions such as the integration of Hong–Ou–Mandel (HOM) interferometry to study the indistinguishability of photons.42,52 The optical path of the potential HOM interferometer is sketched in Fig. S4 in the SI. The performance of the TCSPC setup has been tested on a Rhodamine 6G solution in ethanol and a dilute film of CsPbBr3 colloidal nanocrystals. The PL lifetime of Rhodamine was measured as 3.87 ns, in agreement with a reference value of 3.99 ns (Fig. S5).53 An isolated emitter in the dilute film of weakly-confined CsPbBr3 nanocrystals showed an expected spectrum and photon anti-bunching (g(2) ≈ 0.4, Fig. S6), verifying the functioning of the setup.
Fig. 4 shows the results of the cryogenic temperature measurement of the spectral dynamics of a sub-5 µm CsPbBr3 nanocrystal superlattice, which exhibits multiple PL peaks (Fig. 4a and b) and was prepared from a different batch of nanocrystals than the superlattice discussed in Fig. 3. The setup enables recording of the PL intensity time decays over the entire spectral window (Fig. 4c) with decays for selected wavelengths across the spectrum are summarized in Fig. 4d. Fig. 4e and f shows the second-order correlation function of the different spectral components at the excitation fluence of 0.71 µJ cm−2. The results show expected changes in the PL decay with the wavelength (i.e., faster decay at 522 nm than at 530 nm due to the size-distribution and energy funnelling in superlattices),37,54 and a constant g(2)(0) value of approximately 1 across the PL spectrum, as expected from the Poissonian statistic of light coming from an ensemble of emitters. This capability of the setup to resolve dynamics and correlation at different emitted wavelengths is fundamental to investigate collective phenomena such as superfluorescence and superradiance that are reportedly characterized as narrower, red-shifted, and accelerated PL compared to spontaneous emission, as well as with a bunched (g(2)(0) > 1) photon statistics. The wavelength bandwidth of the measurements in Fig. 4c–f can be tuned by adjusting the width of the output slit or the grating in the spectrograph, with the narrowest being 0.2 nm. As with the µ-PL measurement (Fig. 3), the data files and the script to analyze time-resolved PL data are provided in the SI.
Supplementary information: excitation spot characterization, complete list of the setup components, sketch of HOM setup, validation measurements of Rhodamine lifetime in solution and an isolated emitter in a sample of individual perovskite nanocrystals; original data and scripts. See DOI: https://doi.org/10.1039/d5nr04127a.