Ratiometric luminescence nanothermometry carries the potential to measure temperature in situations for which established methods are unsuitable. The precision of nanothermometry depends on the excitation power, so calibration and monitoring of the optical power is mandatory—a requirement that complicates optical setups and limits nanothermometry in scenarios where precise power control or measurement is impractical or unfeasible. Here, we use Er3+-activated nanothermometers and, besides the well-known 525/545 nm ratio, define a second luminescence intensity ratio involving the emission at 660 nm to achieve a power-calibration-free nanothermometry. The intensity of this emission is strongly correlated with the power and is available anyways when using standard spectroscopic instrumentation. We apply symbolic regression to find an unambiguous mathematical expression that describes the experimental data. From this mathematical expression, we determine the mean temperature deviation resulting from the fitting error to be 0.16 K and a maximum temperature precision as small as 6 mK (0.22 K on average). In summary, our approach makes excitation power measurements in ratiometric luminescent nanothermometry superfluous.
New conceptsWe demonstrate how optical power calibration in ratiometric nanothermometry can be made superfluous. Ratiometric nanothermometers such as the prominent NaYF4:Er,Yb allow to measure temperature optically and remotely at the nanoscale by comparing the luminescence intensity ratio of two thermally coupled energy levels in lanthanide ions – but only if properly calibrated. If the calibration is not carried out carefully, readout errors of several 100 K can occur. Therefore, every conceivable effect must be considered in addition to the temperature dependence of the emission. This also includes the dependence of the emission on the optical excitation power. Here we found that, by considering a third emission line strongly correlated with the excitation power, we can unambiguously assign the correct temperature to the measured luminescence intensity ratios for different excitation power levels without even knowing the actual power. For this purpose, we applied a machine learning algorithm, namely symbolic regression, to directly fit the power dependency without any optical power measurement. This drastically simplifies the calibration of nanothermometers and, in particular, enables nanothermometry in locations where optical power measurement is not possible, such as highly integrated devices or biological in vivo samples. |
As the field of nanothermometry grows and matures, the community expands its research focus from material development to physical limitations when performing nanothermometry in real-life applications. Van Swieten and co-workers investigated the impact of noise and background on nanothermometric measurement uncertainties.9 Vonk et al. found that photonic artifacts can result in temperature read-out errors of up to 250 K.10 Furthermore, lanthanide excitation dynamics involves various nonlinear processes such as excited state absorption or energy transfer upconversion that introduce an excitation power dependence in nanothermometry.11 Pickel and co-workers tackled this dependence with rate-equation modeling, which is a complex endeavor.12–14 Recently, Jia and co-workers found that the dependence of excitation-power in lanthanide-based ratiometric nanothermometers is insensitive to temperature changes and, therefore, can be calibrated.15
However, proper calibration of power dependence in nanothermometry is only possible if one can measure the optical power that reaches the lanthanide ions at the application site. This is rarely the case in real-life applications such as biomedical in vivo applications16,17 or highly integrated devices, e.g. batteries, that involve optical fibers for excitation power delivery to the nanothermometers.17–19 Even if measuring the power is possible, it is highly undesirable because additional measurement periphery impairs the optical setups.
Here, we introduce an approach that makes additional optical power measurements superfluous. We used the popular NaYF4:Er3+,Yb3+ as a model nanothermometer on the tip of an optical fiber using the common ratio between the Erbium-emission at 525 nm (2H11/2 level) and 545 nm (4S3/2 level). Instead of measuring the optical power at the tip of the fiber - which is an impossible endeavor if the fiber is employed as an endoscopic sensor - we consider a second luminescence intensity ratio that involves a third emission line at 660 nm originating from the 4F9/2 level. This emission is strongly excitation power-dependent and its spectroscopic information available anyway. We apply a machine learning algorithm, specifically, symbolic regression, to fit an analytical expression to the experimental data. The resulting expression allows us to determine the temperature of the nanothermometers without knowing the optical power delivered through the fiber with a mean precision of 0.22 K. Our approach can be applied to any material that provides a predominantly power-dependent third emission. Our work showcases how to unambiguously calibrate the power dependency in ratiometric luminescent nanothermometers only through spectroscopic information. Hence, power dependence is no longer a challenges in nanothermometry.
Nanocrystals activated with Erbium ions are the most prominent and, strikingly, also the most versatile nanothermometers.20 We synthesized ellipsoidal β-NaYF4:Er3+,Yb3+@NaYF4 core@shell nanocrystals ((16.3 ± 0.5) nm × (26.8 ± 0.55) nm, c.f.Fig. 1a and b) that predominantly emit green luminescence under infrared excitation at 980 nm (Fig. 1c) and attached them to an optical multimode fiber with a core diameter of 200 μm (see details in the SI). Similar fiber tip sensor concepts were reported before21 and it allowed us to conveniently perform temperature-dependent spectroscopy experiments with the nanothermometers inside a furnace (c.f.Fig. 1d). We characterized the luminescence intensity ratio for different excitation powers during a cool-down period of the furnace from about 100 °C to room temperature (c.f.Fig. 1e).
Ratiometric luminescent nanothermometers are typically calibrated via the ratio
![]() | (1) |
![]() | (2) |
Here, we make use of the fact that nonlinear effects such as energy transfer upconversion or excited state absorption that result in occupation of the green-emitting energy levels also lead to efficient occupation of the red-emitting energy level at high excitation energies. At low excitation powers, the number of infrared photons is only sufficient to excite the green emissions (approximately two infrared photons generate one green photon). The ratio of the dominant green emissions is predominantly temperature-dependent (c.f.Fig. 2a). At high excitation powers, the nonlinearity is strong enough to even excite higher energy levels that first relax by emitting one photon in the green spectral range and as such slightly influence the luminescence intensity ratio as defined in eqn (1). The final energy level of this green transition is the red-emitting level that is efficiently occupied by this mechanism and as a result emits a second, red photon (c.f.Fig. 2b). As a consequence, this emission is predominantly excitation power-dependent (approximately three infrared photons generate one green photon and one red photon, c.f. Fig. S4).
We exploit the different excitation pathways of the green and red emissions to unambiguously calibrate the nanothermometers.23 For that, we define a first luminescence intensity ratio as the ratio between the intensities at 525 nm and 545 nm (Rgg) and a second ratio between the intensities at 545 nm and 660 nm (Rgr, c.f.Fig. 2c). To analyze the data, we applied a machine learning algorithm. Recently, machine learning approaches have been successfully employed to analyze nanothermometry data, e.g. multiparameter regression.24,25 Here, we employ symbolic regression for the first time which is an established machine learning approach based on evolutionary computation for searching mathematical expressions while minimizing various error metrics. In contrast to other machine learning approaches, symbolic regression yields an interpretable result and allows researchers to find natural laws from experimental data.26–28
Fig. 3a shows the cross-validation we performed for an increasing dataset size to verify that our maximum dataset size is sufficiently large. The loss, i.e. the mean squared error is very high for small dataset sizes because the amount of data is insufficient for high quality regression. It converges at larger dataset sizes which corresponds to a decrease of the mean temperature difference
to ≈0.5 K (c.f. inset in Fig. 3a).
We tuned the hyperparameters of the modelling via Bayesian optimization by minimizing |ΔTmax| (c.f.Fig. 3b and SI for details on the optimization) and fitted the model to our data. The difference between the temperature calculated from the fitted model Tcalc and the measured temperature Tmeas is depicted in Fig. 3c to e for the models obtained from the three best hyperparameters. The resulting mathematical expressions significantly differ for each model which is why the position of the maximum absolute error |ΔTmax| (green circle in Fig. 3) varies for each model. However, its value changes only in the second decimal. From the physical plausibility of the respective mathematical expressions, we decided to continue with the first model whose expression reads:
![]() | (3) |
Note that this expression is not so much a correct description of the physics involved but the equation that is found by the symbolic regression to deviate the least from the measured data. While the model yielded by the symbolic regression allows for accurate interpolation within the constrained parameter space it certainly lacks accuracy towards extrapolation. Yet, eqn (3) strikingly includes power-law expressions for Rgr which is in accordance with Jia et al..15 Eventually, the found expression allows us to determine the temperature for given Rgg and Rgr. Furthermore, we can calculate the sensor's sensitivity analytically which is possible through the interpretable result obtained from the symbolic regression. The relative sensitivity of nanothermometers is given by the inverse partial derivative normalized to the luminescence intensity ratio (c.f.eqn (2)). Thus, we can calculate said sensitivity from the root mean square of eqn (3)'s gradient vector field. Using this sensitivity, our approach allows us to analytically calculate the absolute temperature precision via
![]() | (4) |
The relative sensitivity and temperature precision is depicted in Fig. 4. We show the temperature and optical power in the fiber alongside the luminescence intensity ratios to demonstrate that the latter are coupled to real physical quantities. However, measuring these quantities is not necessary to determine the characteristic properties of the nanothermometer. Instead, knowledge of the luminescence intensity ratios is sufficient.
Both metrics have their maximum at high powers and low temperatures because the respective single sensitivities are highest for high powers and low temperatures (c.f.Fig. 1e). Since the calculated sensitivity is a root mean square of both, the temperature and power sensitivity, it is is high compared to typical values reported for nanothermometers.29
The temperature precision is about 6 mK at maximum which is very good and a result of the high sensitivity. However, the total temperature deviation is
| δTtotal = Tcalc − Tmeas + δT | (5) |
In summary, we attached β-NaYF4:Er3+,Yb3+ nanocrystals on a fiber facet and used their temperature-dependent luminescence to measure temperature. By defining a second luminescence intensity ratio that includes the red emission at 660 nm we introduced a measurable quantity that is strongly correlated with the optical excitation power. We employed a symbolic regression algorithm to yield an analytical expression that describes the temperature for the two different intensity ratios. Our approach exhibits a maximum relative root mean square sensitivity of 35% K−1 and a mean temperature precision of 0.22 K. It allows unambiguous calibration of the power dependency in Erbium-activated ratiometric luminescent nanothermometers and can be extended to any activator such as Tm3 + or Ho3 + as long as it provides predominantly power-dependent optical transitions. Furthermore, the approach enables direct analytical determination of a nanothermometer's characteristic properties such as its sensitivity. With our approach, the excitation power of ratiometric luminescent nanothermometers no longer needs to be measured, which drastically simplifies their calibration and respective optical setups. Towards applications, this is a huge benefit, in particular in minimally-invasive scenarios such as in vivo applications.