Mohsin Aliac,
Kousar Jahanb,
Jitendra Singh*c,
Ratnesh Kumar Singhd,
Sudhir Kumar Shoorae,
Xu Fengf and
Yanfeng Yue
*b
aDepartment of Applied Sciences, Modern Institute of Technology and Research Centre, Alwar 301001, India
bDepartment of Chemistry, Delaware State University, Dover, Delaware 19901, USA. E-mail: yyue@desu.edu
cDepartment of Chemistry, Hemvati Nandan Bahuguna Garhwal University, Srinagar-247174, India. E-mail: jkfcy13@hnbgu.ac.in
dDepartment of Chemistry, Government Engineering College, Bhojpur, 802301, India
eMinistry of Environment, Forest and Climate Change, Government of India, Indira Paryavaran Bhawan, Jorbagh Road, New Delhi, India
fSurface Analysis Facility, University of Delaware, Newark, DE 19716, USA
First published on 1st August 2025
Co(II) and Cr(III) salicylidene Schiff base-based complexes as novel ionophores were evaluated for the fabrication of bromide-selective electrodes. By incorporating a cation excluder along with various plasticizers (dibutyl phthalate, dioctyl phthalate, 1-chloronapthalene), optimized sensors (CoC7 and CrC7) exhibiting near-Nernstian slopes being 59.4 ± 0.07 and 59.2 ± 0.04 mV decade−1, with a broad linear range (1 × 10−2 to 6.0 × 10−7 and 1 × 10−2 to 8.7 × 10−7 mol L−1), with low detection limits (5.5 ± 0.13 × 10−7 and 6.5 ± 0.07 × 10−7 mol L−1) respectively, were successfully designed. Selectivity coefficient values of order 10−1 or less indicate that the proposed electrodes have superior selectivity for bromide ions over various interfering anions. The developed bromide electrodes demonstrated robust performance within a pH range of 4.0 to 9.0, as well as showing a sufficient shelf life (4 and 5 weeks) with up to 20% (v/v) non-aqueous tolerance and quick response times (12 and 16 s). These electrodes also served as indicator electrodes in the potentiometric titration of bromide ions against AgNO3 and were used in the determination of bromide ion concentration in water samples.
Ion-selective electrodes (ISEs) have emerged as promising tools for ion determination due to their simplicity, portability, and low cost. These electrodes exploit the selective interaction between the target ion and a specific sensing element, allowing for direct measurement without extensive sample preparation or sophisticated instrumentation.14 Fig. 1 illustrates the components and the selective mechanism for bromide ion detection using the potentiometric method. In this context, attempts were made to develop new bromide (Br−) ion-selective PVC membranes with the combination of transition metal complexes. The framework of the ligand and the characteristics of the metal ions in the coordination complex play an important role in affecting the Br− ion selectivity of these electrodes. Given the significance of the determination of Br− ions, many bromide ion-selective polymeric membrane electrodes have been studied, using different ionophores such as metal complexes HgS/Hg2Br2,15 chalcogenide glassy–crystalline AgBr-Ag2S-As2S3,16 14-phenyldibenzo[a,j]xanthene,17 bis(4-hydroxyphenyl)-1,4-diaza-1,3-butadiene mercury(II) complex,18 iron(III)-salen,19 meso-tetraphenylporphyrin manganese(III) chloride complex and 4,5-dimethyl-3,6-dioctyloxy-o-phenylene-bis(mercury trifluoroacetate),20 mercury(II) complex of a pyridine,21 graphene oxide–aluminium fumarate metal–organic framework (BGO/AlFu MOF),22 Pt(II) 5,10,15,20-tetra(4-methoxy-phenyl)-porphyrin(PtTMeOPP),23 1,3-dihexadecylimidazolium bromide,24 and an azapyrylium ion derivative.25 Isildak et al. have used a cobyrinic acid derivative macrocyclic molecule as an ionophore for selective detection of Br− ions with concentration ranges of 1 × 10−2–6.0 × 10−7 and 1.0 × 10−1–1.0 × 10−4 mol L−1 and with detection limits of 5.5 × 10−7 and 2.2 × 10−5 mol L−1.26 While the Br− ion sensors that were previously reported had strong sensitivity and selectivity towards bromide, many of them were limited by non-Nernstian slope, poor detection limit, and short working concentration range. In the context of existing bromide sensors, there is an urgent need to develop a robust bromide-selective sensor with enhanced potentiometric characteristics for the detection of low bromide concentrations. In this work, Co(II) and Cr(III) salicylidene Schiff base complexes were investigated as ionophores for Br− ion detection in aqueous media based on a potentiometric method. Interestingly, due to the high stability of the Co(II) and Cr(III) complexes and specific interactions among the metal center and different anions, an improved potentiometric performance for bromide ion detection in terms of high sensitivity, wide linear range, lower detection limit, and high stability in a wide pH range was recorded. Meanwhile, both cobalt and chromium complexes often have flexible coordination numbers and geometries, which can be customized to fit the size and charge of Br− ions. This flexibility can improve the binding efficiency and stability of halides. Additionally, the facile synthesis methodology and low cost of all precursors make these complexes excellent alternatives to expensive commercial bromide exchangers. The advancement made here will benefit future sensor development to detect other anions, such as environmentally important nutrients including NO3− and HPO42−. The components and the selective mechanism for bromide ion detection using the potentiometric method are illustrated in Fig. 1. In this context, attempts were made to develop new bromide (Br−) ion-selective PVC membranes with different combinations of these transition metal complexes. The framework of the ligand and the characteristics of the metal ions in the coordination complex play an important role in affecting the Br− ion selectivity of these electrodes.
The membranes were allowed to acclimate in a 1.0 × 10−1 mol L−1 NaBr solution for four days. Potentials were recorded by varying the concentration of the NaBr test solution within the range of 1.0 × 10−8–1.0 × 10−1 mol L−1. Each solution was agitated and the potential was recorded once it reached a stable point. The logarithmic function of the Br− ion activity was then plotted. A saturated calomel electrode (SCE) was used as the reference electrode for the potential measurements, which were performed at 25 ± 0.1 °C using an Orion 4-star pH metre with the following cell assembly:
Hg/Hg2Cl2 |KCl(satd)| 0.1 M NaBr ||PVC membrane|| test solution |Hg/Hg2Cl2| KCl(satd).
The activities of Br− ions were calculated according to the following modified Debye–Huckel approximation (eqn (1)):
![]() | (1) |
Further, UV-visible (UV-vis) spectroscopy is employed to analyze the interaction between CoC and CrC ionophores and bromide ions due to its capability to detect molecular interactions based on changes in absorbance. UV-vis absorption spectra of CoC and CrC were recorded, using an equimolar quantity of NaBr solution each having a 1.0 × 10−3 mol L−1 concentration in methanol. UV-vis spectra, as shown in Fig. 2(A and B), can discern the interlinkage between the metal chelates and bromide ions. The considerable changes in the absorbance of CoC at 245 nm and CrC at 246 and 255 nm were noted in the absorption spectra with an equimolar amount of bromide ion solution.29 Here, the results strongly imply that ionophores form 1:
1 complexes with the bromide ion. On the other hand, there were negligible changes in the UV-vis spectra of CoC and CrC ionophores when recorded in the presence of other anions. The perceived spectral shifts, coupled with the significant increase in absorbance peaks in the CoC and CrC spectra following the contact with the bromide ion-containing solution, indicate the favorable coordination of the bromide ion with the ionophore. The change in the UV-vis spectra indicates the interaction between the ionophores and bromide ions.
![]() | ||
Fig. 2 The absorption spectra of (A) CoC and (B) CrC with bromide ion (NaBr 1.0 × 10−3 mol L−1) solution in MeOH. |
To understand the coordination behavior of both CoC and CrC with different anions, a conductometric titration method was carried out. Conductometric study was performed by titrating 20 mL 1.0 × 10−4 mol L−1 of anion solution against 1.0 × 10−2 mol L−1 solution of CoC and CrC. The conductance (S cm−1) of the solution was precisely noted for each addition of ionophores. The results showed that among several anions studied, the conductance of the Br− ion solution was greatly affected with the addition of CoC and CrC. The observed variation in the conductance of the Br− ion solution with the metal complexes is plotted in Fig. 3, which explains that the conductance of the bromide ion solution starts decreasing quickly with the addition of the metal complexes. Once all bromide ions are consumed, no further change in conductance occurs, resulting in the conductometric titration curve reaching a nearly straight line at the end point. An exact stoichiometry ratio of 1:
1 is indicated for the final product of the reaction at the end of the conductometric titration curve.
One of the most critical parameters that sheds light on an ion electrode's functional selectivity is the ion–ionophore complexation occurring within the membrane. The sandwich membrane method was employed to calculate the formation constants for the 1:
1 ion–ionophore complex. A concentration-polarized sandwich membrane was formed by fusing together two membrane segments, of which only one contained the ionophore. The formation constants were then calculated by the following eqn (2):30
![]() | (2) |
Anion (Xn–) | (Log![]() |
(Log![]() |
---|---|---|
Br− | 7.09 ± 0.16 | 6.90 ± 0.06 |
I− | 5.02 ± 0.02 | 4.91 ± 0.22 |
Cl− | 4.95 ± 0.35 | 4.80 ± 0.25 |
CO32− | 4.91 ± 0.07 | 4.75 ± 0.08 |
SO42− | 4.83 ± 0.66 | 4.70 ± 0.55 |
CN− | 4.57 ± 0.36 | 4.45 ± 0.32 |
F− | 4.43 ± 0.26 | 4.31 ± 0.12 |
SCN− | 3.72 ± 0.27 | 3.66 ± 0.10 |
NO2− | 3.71 ± 0.16 | 3.55 ± 0.61 |
NO3− | 3.02 ± 0.72 | 3.04 ± 0.07 |
Sal− | 2.92 ± 0.26 | 2.75 ± 0.14 |
The compositions of potentiometric membrane electrodes fabricated by incorporating CoC and CrC were optimized by adjusting the amounts of other membrane ingredients.31 All the electrodes were calibrated in 10−1 M NaBr solution before potentiometric studies. The results show the potentiometric characteristics of bromide selective electrodes based on ionophores CoC (CoC1 to CoC10) and CrC (CrC1 to CrC10) which were evaluated as a function of bromide ion concentration in the range of 10−8 to 10−1 mol L−1 (Table 2). The electrodes with 6 mg of CoC and 94 mg PVC (CoC1) and 5 mg of CrC and 95 mg of PVC (CrC1) exhibited potentiometric characteristics demonstrating sub-Nernstian slopes (36.2 mV per decade for CoC and 38.1 mV per decade for CrC of Br−), high detection limits (2.5 × 10−4 mol L−1 for CoC and 2.0 × 10−4 mol L−1 for CrC) with narrow working concentration ranges (5.0 × 10−4–1.0 × 10−3 mol L−1 and 5.5 × 10−4–1.0 ×10−2 mol L−1). The sensitivity, linearity, and selectivity of given ionophores are significantly influenced by the membrane composition, particularly the type of plasticizer used.32 Therefore, numerous combinations of membrane constituents (as listed in Tables 2 and 3) were tested to identify the membrane that provides high performance and reproducible results. Subsequently, the membrane electrode with the highest performance was chosen for thorough potentiometric analyses. According to previously published studies, a cation excluder enhances the potentiometric properties of an anion-selective electrode and increases its selectivity.33 To improve electrode performance, a cation exchanger (CTAB) was mixed with the membrane composition. Potential studies given in Tables 2 and 3 revealed that the electrodes (CoC2 and CrC2) having 3 mg CTAB exhibited some improvement in the slope, detection limit and working concentration range.
Electrode no. | Membrane ingredients (mg) | Slope (mV per decade of [Br−]) | Linear range (mol L−1) | Detection limit (mol L−1) | |||
---|---|---|---|---|---|---|---|
Ionophore (CoC) | CTAB | Plasticizer | PVC | ||||
CoC1 | 6 | 0 | 0 | 94 | 36.2 | 5.0 × 10−4–1.0 × 10−3 | 2.5 × 10−4 |
CoC2 | 6 | 3 | 0 | 91 | 45.4 | 6.9 × 10−5–1.0 × 10−2 | 3.0 × 10−5 |
CoC3 | 6 | 3 | 55, DOP | 36 | 51.7 | 7.1 × 10−6–1.0 × 10−2 | 3.4 × 10−6 |
CoC4 | 6 | 3 | 55, NPOE | 36 | 62.4 | 7.3 × 10−6–1.0 × 10−2 | 3.7 × 10−6 |
CoC5 | 6 | 3 | 55, CN | 36 | 63.0 | 5.3 × 10−6–1.0 × 10−2 | 2.5 × 10−6 |
CoC6 | 6 | 3 | 55, TBP | 36 | 56.8 | 1.5 × 10−6–1.0 × 10−2 | 9.2 × 10−6 |
CoC7 | 6 | 3 | 55, DBP | 36 | 59.4 | 6.0 × 10−7–1.0 × 10−2 | 5.5 × 10−7 |
CoC8 | 5 | 3 | 55, DBP | 37 | 57.9 | 5.7 × 10−6–1.0 × 10−2 | 3.6 × 10−6 |
CoC9 | 7 | 3 | 55, DBP | 35 | 59.1 | 6.2 × 10−6–1.0 × 10−2 | 5.6 × 10−6 |
CoC10 | 6 | 2 | 55, DBP | 37 | 57.6 | 3.7 × 10−6–1.0 × 10−2 | 1.8 × 10−6 |
Electrode no. | Membrane ingredients (mg) | Slope (mV per decade of [Br−]) | Linear range (mol L−1) | Detection limit (mol L−1) | |||
---|---|---|---|---|---|---|---|
Ionophore (CrC) | CTAB | Plasticizer | PVC | ||||
CrC1 | 5 | 0 | 0 | 95 | 38.1 | 5.5 × 10−4 –1.0 × 10−2 | 2.0 × 10−4 |
CrC2 | 5 | 3 | 0 | 92 | 47.2 | 5.0 × 10−5–1.0 × 10−2 | 2.5 × 10−5 |
CrC3 | 5 | 3 | 55, CN | 37 | 66.5 | 7.5 × 10−6–1.0 × 10−2 | 3.9 × 10−6 |
CrC4 | 5 | 3 | 55, NPOE | 37 | 61.1 | 4.3 × 10−6–1.0 × 10−2 | 2.6 × 10−6 |
CrC5 | 5 | 3 | 55, TBP | 37 | 53.3 | 1.5 × 10−6–1.0 × 10−2 | 1.1 × 10−6 |
CrC6 | 5 | 3 | 55, DOP | 37 | 58.0 | 5.9 × 10−6–1.0 × 10−2 | 2.1 × 10−6 |
CrC7 | 5 | 3 | 55, DBP | 37 | 59.2 | 8.7 × 10−7–1.0 × 10−2 | 6.5 × 10−7 |
CrC8 | 4 | 3 | 55, DBP | 38 | 51.4 | 4.3 × 10−6–1.0 × 10−2 | 3.6 × 10−6 |
CrC9 | 3 | 3 | 55, DBP | 39 | 56.8 | 9.0 × 10−5–1.0 × 10−2 | 7.0 × 10−5 |
CrC10 | 5 | 4 | 55, DBP | 36 | 55.4 | 4.1 × 10−6–1.0 × 10−2 | 2.2 × 10−6 |
It was further seen from Table 2 that the membranes of CoC with different types of plasticizers viz., CN, NPOE, TBP, DOP, and DBP (electrode no. CoC3–CoC7) performed better than the membrane without a plasticizer showcasing wider working concentration ranges and improved slopes. DBP plasticized membrane electrode CoC7 showed the best performance among all the plasticized membranes, exhibited the widest working concentration range of 6.0 × 10−7–1.0 × 10−2 mol L−1 with a Nernstian slope of 59.4 mV per decade of [Br−] and a low detection limit of 5.5 × 10−7 mol L−1. In Table 3, similar results were also obtained with the membranes of CrC with different types of plasticizers viz., CN, NPOE, TBP, DOP, and DBP (electrode no. CrC3–CrC7), which performed better than the membrane without a plasticizer showcasing wider working concentration ranges and improved slopes. DBP plasticized membrane electrode CrC7 showed the best performance among all the plasticized membranes, exhibited the widest working concentration range of 8.7 × 10−7–1.0 × 10−2 mol L−1 with a Nernstian slope of 59.2 mV per decade of [Br−] and a low detection limit of 6.5 × 10−7 mol L−1. Due to the very low polarity and great mobility of DBP compared to other plasticizers explored, DBP is supposed to provide sufficient conditions for the integration of Br− ions into the membrane before coordinating with the metal in complexes.33
The effect of varying amounts of ionophores (CoC7, CoC8, and CoC9; CrC7, CrC8, and CrC9) was also studied, and it was observed that reducing the ionophore concentration in the membrane phase significantly changed the potentiometric characteristics of the electrodes (CoC8 and CoC9), and CrC8 and CrC9 did not show improved performance. The potentiometric responses similarly decreased for CoC10 and CrC10 which had an increased concentration of CTAB, leading to increased interference from foreign cations in the solution. This is evident from the performance comparison of all these electrodes shown in Tables 2 and 3, indicating that electrodes CoC7 and CrC7 are the best-performing electrodes, and the optimum composition of CoC7 is determined to be CoC:
DBP
:
CTAB
:
PVC in a ratio of 6
:
55
:
3
:
36 (mg) and that of CrC7 is determined to be CrC
:
DBP
:
CTAB
:
PVC in a ratio of 5
:
55
:
3
:
37 (mg). The potential responses of the CoC and CrC7 electrodes were evaluated by contentiously changing the internal solution from 1.0 × 10−1 to 1.0 × 10−8 mol L−1 and their potentiometric data were observed from their calibration curve. The study found minor differences in the Nernstian slope of the electrodes in dilute solutions (1.0 × 10−3 and 1.0 × 10−4 mol L−1), affecting their operating range and detection limitations. Therefore, the suitable concentration of internal Br− solution was determined to be 1.0 × 10−2 mol L−1 for further studies of CoC7 and CrC7.
The equilibration time of ISEs is an important parameter affecting their performance. For the activation of electrodes, the potential response of the electrodes was measured after soaking them in a 1.0 × 10−2 mol L−1 bromide solution for distinct time periods. Stable and reproducible results were obtained after conditioning the electrodes for a period of 12 hours. However, extending the equilibration time beyond 12 hours showed no improvement in the performance of CoC7 and CrC7. The best performing electrodes, CoC7 and CrC7, were used under the above-explained state. The calibration plots of these electrodes shown in Fig. 4 show that they can operate over broad working concentration ranges (6.0 × 10−7–1.0 × 10−2 mol L−1 and 8.7 × 10−7–1.0 × 10−2 mol L−1) showing Nernstian responses (59.4 and 59.2 mV per decade of [Br−]) with lower detection limits of 5.5 × 10−7 and 6.5 × 10−6 mol L−1, respectively.
For the determination of a suitable pH range for the considered electrode, the effect of pH on the performance of CoC7 and CrC7 was assessed in 1.0 × 10−2 M mol L−1 bromide ion solution. The pH range between 1.0 and 12.0 was adjusted by the addition of a desired amount of dilute HCl and NaOH solutions. The potential response for this study is shown in Fig. 5, showing that the potentials remained unchanged within the pH range of 4.0 to 9.0, making this working pH range suitable for further studies. The deviations in potentials beyond this pH range may be attributed to the oxidation of bromide in the acidic solution (below pH 4.0) and as a result of hydrolysis (above pH 9.0) of metal complexes. To determine the response time of the suggested electrodes, the concentration of the Br− solution was sequentially changed from 1.0 × 10−2 mol L−1 to 1.0 × 10−6 mol L−1. The potential responses are depicted in Fig. 6(A and B), and the results suggested that the time required for obtaining the stable potential value after immersing the electrodes successively in a series of bromide ion solutions is 12 and 16 seconds for CoC7 and CrC7, respectively. This study indicates rapid exchange kinetics between bromide ions and ionophores at the solution–membrane interface. To examine the reversibility of these electrodes, the process was reversed by measuring the potential from high to low concentrations and the response time of these electrodes remained unchanged. Furthermore, the reproducibility of the electrodes was also checked with a set of three duplicate electrodes of CoC7 and CrC7, and their potentiometric investigations were performed under identical conditions. The results were found to be in good agreement (±0.06 mV), indicating their reliable and reproducible performance.
![]() | ||
Fig. 5 Effects of pH on the performance of proposed electrode solutions with bromide at a concentration of 1.0 × 10−2 mol L−1. |
The lifetime of an ISE is an important feature that describes its active lifespan for accurately estimating primary ions in the test solution. To evaluate the lifetime of the proposed bromide selective electrodes, ee, the proposed electrodes were used daily over a period of 1 hour each day for 2 months (8 weeks). The slopes and detection limits were measured from the derived calibration plots, as shown in Table 4. The experimental results showed that there was no significant change in the potentiometric characteristics of electrodes CoC7 and CrC7 up to 5 weeks. However, the potential responses of these electrodes gradually deteriorated over time, showing sub-Nernstian slopes and elevated detection limits. It is noteworthy that the electrodes were stored in 0.1 M NaBr solution when not in use.
Time (weeks) | Electrode CoC | Electrode CrC | ||
---|---|---|---|---|
Slope (mV) | Detection limit (mol L−1) | Slope (mV) | Detection limit (mol L−1) | |
1 | 59.4 ± 0.2 | 5.5 × 10−7 | 59.2 ± 0.2 | 6.5 × 10−7 |
2 | 59.4 ± 0.04 | 5.5 × 10−7 | 59.2 ± 0.3 | 6.5 × 10−7 |
3 | 59.4 ± 0.5 | 5.8 × 10−7 | 59.1 ± 0.4 | 6.8 × 10−7 |
4 | 59.2 ± 0.3 | 6.5 × 10−7 | 59.1 ± 0.4 | 6.8 × 10−7 |
5 | 59.1 ± 0.4 | 4.1 × 10−6 | 59.0 ± 0.6 | 7.5 × 10−7 |
6 | 58.2 ± 0.7 | 5.5 × 10−6 | 58.8 ± 0.7 | 3.5 × 10−6 |
7 | 57.6 ± 0.8 | 9.4 × 10−5 | 58.2 ± 0.9 | 4.5 × 10−6 |
8 | 56.9 ± 0.5 | 1.5 × 10−5 | 57.1 ± 0.5 | 1.7 × 10−5 |
The functions of both CoC7 and CrC7 electrodes were also evaluated in partially non-aqueous solutions as shown in Table 5 with methanol–water, ethanol–water and acetonitrile–water solutions of different concentrations, and calibration curves were recorded. It was observed that that the electrodes could tolerate up to 20% (v/v) non-aqueous content as there was negligible change in their slope and linear working range at this concentration. However, above 20% non-aqueous content, the slope and working range are considerably reduced, which could be due to membrane degradation from leaching of ionophores or other ingredients from the PVC matrix.
Non-aqueous content (%, v/v) | Electrode CoC7 | Electrode CrC7 | ||
---|---|---|---|---|
Working concentration range (mol L−1) | Slope (mV per decade [Br]) | Working concentration range (mol L−1) | Slope (mV per decade [Br]) | |
Nil | 6.0 × 10−7–1.0 × 10−2 | 59.4 | 8.7 × 10−7–1.0 × 10−2 | 59.2 |
Methanol | ||||
10 | 6.0 × 10−7–1.0×10−2 | 59.4 | 8.7 × 10−7–1.0 × 10−2 | 59.2 |
20 | 6.0 × 10−7–1.0 × 10−2 | 59.4 | 8.7 × 10−7–1.0 × 10−2 | 59.2 |
30 | 2.6 × 10−6–1.0 × 10−2 | 58.5 | 3.7 × 10−6–1.0 × 10−2 | 58.0 |
35 | 5.0 × 10−6–1.0 × 10−2 | 57.9 | 8.2 × 10−6–1.0 × 10−2 | 57.5 |
Ethanol | ||||
10 | 6.0 × 10−7–1.0 × 10−2 | 59.4 | 8.7 × 10−7–1.0 × 10−2 | 59.2 |
20 | 6.0 × 10−7–1.0 × 10−2 | 59.4 | 8.9 × 10−7–1.0 × 10−2 | 59.1 |
30 | 8.5 × 10−6–1.0 × 10−2 | 58.9 | 2.7 × 10−6–1.0 × 10−2 | 58.4 |
35 | 9.2 × 10−6–1.0 × 10−2 | 57.9 | 7.5 × 10−6–1.0 × 10−2 | 57.6 |
Acetonitrile | ||||
10 | 6.0 × 10−7–1.0 × 10−2 | 59.4 | 8.7 × 10−7–1.0 × 10−2 | 59.2 |
20 | 6.1 × 10−7–1.0 × 10−2 | 59.3 | 8.9 × 10−7–1.0 × 10−2 | 59.1 |
30 | 4.5 × 10−6–1.0 × 10−2 | 58.5 | 4.7 × 10−6–1.0 × 10−2 | 58.3 |
35 | 3.1 × 10−6–1.0 × 10−2 | 57.9 | 5.5 × 10−6–1.0 × 10−2 | 58.0 |
Selectivity is a crucial response characteristic of an ion selective electrode; it permits the prediction of how an ISE will perform in the analysis of real-life samples. The IUPAC recommended fixed interference method (FIM) was used to determine the potentiometric selectivity coefficient (SC) for the assumed electrode in the presence of various foreign anions (An−). The SC in this method was determined using potential measurements of Br− ion solutions of different concentrations ranging from 10−2 to 10−8 M and containing a fixed concentration of interfering ions (1.0 × 10−2 M). The SC values for several anions are given in Table 6. Moreover, the SC values for all the interfering ions are less than 1.0; it can be concluded that the proposed Co(II) and Cr(III) complex-based electrodes are highly selective to Br− ions over the interfering ions studied.34,35
Interfering anions (B) | Selectivity coefficient (−log![]() |
|
---|---|---|
Electrode CoC7 | Electrode CrC7 | |
CN− | 1.38 | 1.27 |
SCN− | 1.50 | 1.31 |
Sal- | 1.84 | 1.62 |
F− | 1.43 | 1.22 |
NO2− | 1.62 | 1.39 |
Cl− | 1.38 | 1.18 |
I− | 1.43 | 1.17 |
NO3− | 1.56 | 1.26 |
SO42− | 2.20 | 2.30 |
CO32− | 1.03 | 1.05 |
The proposed electrodes CoC7 and CrC7 exhibited good performance as indicator electrodes in the potentiometric titration of Br− ions with AgNO3 and proved their analytical applicability. For this purpose, a 20 mL solution of 1.0 × 10−2 mol L−1 NaBr was titrated against 1.0 × 10−1 mol L−1 AgNO3 solution at pH 6.0, and the potential response observed using CoC7 and CrC7 are presented in Fig. 7. The resultant titration plot has a conventional sigmoid shape, with a sharp endpoint that matches the 1:
1 stoichiometry of the precipitated silver bromide. As a result, potentiometric titration may be used to approximate Br− ions using the recommended electrodes.
![]() | ||
Fig. 7 Potentiometric titration of 20 mL solution of 1.0 × 10−2 mol L−1 NaBr against 1.0 × 10−1 mol L−1 AgNO3 solution. |
Naturally occurring bromide ions can contaminate drinking water due to excess use of fertilizers and percolation of industrial wastewater in the groundwater. We used the proposed Co(II) complex-based CoC7 and Cr(III) complex-based CrC7 electrodes to assess the Br− ion concentration in tap water samples. In this study, the tap water samples were initially spiked with Br− ions at concentrations of 25 mg L−1, 50 mg L−1 and 100 mg L−1. Subsequently, 100 mL of every spiked water sample was collected in distinct beakers, and 2 mL of NaNO3 (0.1 M) solution was added as an ionic strength adjuster (ISA). The solution was stirred vigorously before the potentiometric analysis. The results for the estimated Br− ion concentrations in the tap water samples are given in Table 7. The concentrations determined by these electrodes are slightly higher than the amount of Br− ions added to the test samples, indicating that the bromide ions were already present in tap water samples before the addition. The results of atomic absorption spectrometry (AAS) and potentiometric analysis of the bromide ions in the tap water samples using the proposed electrodes were compared. The results presented in Table 7 demonstrate a significant level of agreement between the two methods, thereby confirming the suggested electrodes' ability for rapid measurement.
Sample no. | Added bromide concentration (mg L−1) | Found bromide concentration (mg L−1) | ||
---|---|---|---|---|
Sensor CoC7 | Sensor CrC7 | AAS | ||
1 | 25 | 25.3 ± 0.15 | 25.3 ± 0.12 | 25.7 ± 0.06 |
2 | 50 | 50.5 ± 0.05 | 50.4 ± 0.06 | 50.7 ± 0.12 |
3 | 100 | 101.3 ± 0.05 | 101.1 ± 0.0.1 | 101.5 ± 0.06 |
The potentiometric characteristics of the developed bromide ion electrodes CoC7 and CrC7 based on Co(II) and Cr(III) salicylidene Schiff base complexes were compared with the previously reported Br− ion electrodes (Table 8). This comparison revealed that the CoC7 and CrC7 electrodes operate over a wider linear concentration range and exhibit lower detection limits and faster response times than other electrodes. This superior performance is likely due to the strong and durable nature of both complexes, which is important for the longevity and reliability of ion-selective electrodes. In addition, these electrodes exhibited an ideal Nernstian behavior with a very low detection limit, further enhancing their utility and reliability in Br− ion determination under different environmental conditions.
Ref. no | Slope (mV per decade) | Linear range (mol L−1) | Detection limit (mol L−1) | pH range |
---|---|---|---|---|
15 | 58.0 | 10−1–10−6 | 5.0 × 10−7 | 3.5–9.0 |
16 | 59.2–60.4 | 1 × 10−1–2 × 10−6 | 3 × 10−7 | 2.0–10.0 |
17 | 61 ± 1 | 1.0 × 10−1–3.2 × 10−5 | 2.0 × 10−5 | 4.5–8.5 |
18 | 59.1 ± 0.5 | 10−5–10−1 | 5.0 × 10−6 | 4.0–9.5 |
19 | 59.0 | 7.0 × 10−6–1.0 × 10−1 | 6.0 × 10−6 | 3.0–9.0 |
20 | — | 1.0 × 10−8–1.0 × 10−6 | 2.0 × 10−9 | — |
21 | 61.0 ± 0.9 | 3.0 × 10−2–1.0 × 10−5 | 4.0 × 10−6 | 4.5–7.5 |
22 | 54.53 ± 0.15 | 1.0 × 10−7–1.0 × 10−1 | 7.1 × 10−8 | — |
23 | 64.4 ± 0.4 | 1.0 × 10−1–1.0 × 10−5 | 8.0 × 10−6 | 6.0–12.0 |
26 | — | 1.0 × 10−1–1.0 × 10−4 | 2.2 × 10−5 | 4.0–10 |
24 | 63 ± 1 | 1.0 × 10−1–1.0 × 10−6 | 1.6 × 10−5 | 2.0–11.0 |
CoC7 | 59.4 | 6.0 × 10−7–1.0 × 10−2 | 5.5 × 10−7 | 4.0–9.0 |
CrC7 | 59.2 | 8.7 × 10−7–1.0×10−2 | 6.5 × 10−7 | 4.0–9.0 |
The original contributions presented in the study are included in the article material; further inquiries can be directed to the corresponding author.
This journal is © The Royal Society of Chemistry 2025 |