Ahmed Galal Eldina,
Ekram H. Mohamed
b,
Said M. El-Sheikh
c,
Sheta M. Sheta
d,
Ahmed O. Youssefa,
Esraa Elshahate,
Mohamed S. Attia
*a and
Mona N. Abou-Omar*f
aChemistry Department, Faculty of Science, Ain Shams University, Abbassia, Cairo, 11566, Egypt. E-mail: mohd_mostafa@sci.asu.edu.eg
bAnalytical Chemistry Department, The British University in Egypt, El Sherouk city, Cairo, 11378, Egypt
cDepartment of Nanomaterials and Nanotechnology, Central Metallurgical R & D Institute, Cairo, 11421, Egypt
dDepartment of Inorganic Chemistry, National Research Centre, 33 El-Behouth St., Dokki, Giza, 12622, Egypt
eClinical Pathology Department, Faculty of Medicine, Ain Sham University, Abbassia, Cairo, 11566, Egypt
fDepartment of Chemistry, Faculty of Science, University of Sadat City, Sadat City, 32897, Egypt. E-mail: mona.na66@yahoo.com
First published on 18th September 2025
Metal–organic frameworks (MOFs) integrated with noble metal nanoparticles offer transformative potential in biosensing, yet bimetal composites face persistent challenges, including inhomogeneous nanoparticle distribution and framework destabilization. Salen-based MOFs, prized for their thermal stability and tunable electronic properties, remain underexplored for bimetal systems despite their catalytic and plasmonic synergies. This study addresses the critical need for a synthesis strategy that ensures the uniform dispersion of plasmonic silver nanoparticles (Ag NPs) within a robust Zn-salen metal–organic framework (MOF) matrix while preserving structural integrity for biomarker detection applications. Conventional one-pot methods for bimetal MOFs often result in Ag NP aggregation (>200 nm) and pore blockage, compromising porosity and active site accessibility—key limitations for biosensor development. A hierarchical Ag@Zn-SalenMOF composite was engineered via a novel two-step hydrothermal approach: (1) Zn-salen framework assembly, followed by (2) DMF-mediated in situ reduction of AgNO3 to submicron Ag NPs (150–200 nm). The composite was characterized by FE-SEM, HR-TEM, XPS, and XRD, and functionalized with anti-CA15-3 antibodies for biosensor fabrication. The composite exhibited exceptional thermal stability (>300 °C), uniform Ag NP distribution (189 nm avg.), and strong interfacial electronic coupling. The biosensor achieved a CA15-3 detection limit of 0.12 U mL−1 (Stern–Volmer constant: 0.004 U mL−1) with 93.3% sensitivity and 91.6% specificity in clinical serum samples (n = 20), outperforming conventional immunoassays. Stability studies confirmed less than 5% signal drift over 2 months, enabled by sodium azide preservation. This work addresses long-standing synthesis challenges in bimetallic metal–organic frameworks (MOFs), providing a scalable platform for ultrasensitive biomarker detection. The Ag@Zn-Salen MOF biosensor's precision, robustness, and clinical validity position it as a transformative tool for breast cancer monitoring and early diagnosis.
In recent years, various techniques have been utilized to measure CA 15-3 levels in serum samples, including enzyme-linked immunosorbent assays (ELISAs), electrochemiluminescence immunoassays (ECLIAs), and radioimmunoassays (RIAs).3 While these conventional methods offer reasonable accuracy, they often suffer from limitations such as long analysis times, high costs, and the need for sophisticated instrumentation. In contrast, emerging fluorescent biosensing platforms, particularly those based on metal–organic frameworks (MOFs) and lanthanide complexes, have gained attention due to their high sensitivity, rapid detection capabilities, and potential for point-of-care applications.4
Over the past five years, comparative studies have highlighted the advantages and disadvantages of traditional analytical methods versus fluorescent biosensors for CA 15-3 detection. Conventional immunoassays provide well-standardized protocols but may lack the sensitivity required for early-stage cancer detection.5 On the other hand, MOF-based and lanthanide-doped fluorescent biosensors exhibit enhanced luminescence properties, low detection limits, and multiplexing capabilities, making them promising alternatives.4,6 However, challenges such as matrix interference, reproducibility, and the need for extensive validation remain significant hurdles for clinical adoption.
Metal–organic frameworks (MOFs) have emerged as one of the most promising classes of porous crystalline materials in contemporary materials science, offering unprecedented opportunities for designing functional materials with tailored properties through the strategic combination of organic linkers and metallic nodes.7,8 Among the diverse array of organic linkers employed in MOF synthesis, salen-based ligands have garnered significant attention due to their exceptional coordination versatility, robust chelating capabilities, and unique electronic properties that stem from their extended π-conjugated aromatic systems.9,10 The incorporation of salen (N,N′-bis(salicylidene)ethylenediamine) derivatives as organic building blocks in MOF architectures offers distinct advantages, including enhanced thermal stability, tunable electronic properties, and the ability to accommodate multiple metal coordination sites within a single framework structure. These characteristics make salen-based MOFs particularly attractive for applications requiring precise control over electronic, optical, and catalytic properties.11
Integrating noble metal nanoparticles, particularly silver, into MOF matrices represents a cutting-edge approach to developing hybrid composite materials that synergistically combine the unique properties of both components.12,13 Silver nanoparticles are renowned for their exceptional plasmonic properties, broad-spectrum antimicrobial activity, high electrical conductivity, and catalytic performance in various organic transformations. When incorporated into MOF structures, these nanoparticles can create unique interfacial interactions that significantly modify the resulting composite materials' electronic structure, optical response, and functional capabilities.14,15 The challenge lies in achieving uniform distribution and intimate integration of metal nanoparticles within the MOF matrix while maintaining the structural integrity and porosity of the framework, which requires careful control of synthesis conditions and mechanistic understanding of the formation processes.16
Hydrothermal synthesis has proven to be an exceptionally effective method for preparing high-quality MOF-based composites, offering precise control over crystallization kinetics, particle size distribution, and phase purity under relatively mild conditions.17,18 This synthetic approach enables the simultaneous formation of the MOF framework and incorporation of metal nanoparticles through carefully orchestrated reaction sequences that promote both nucleation and growth processes in a controlled manner. The hydrothermal environment provides optimal conditions for achieving homogeneous mixing of precursor materials, facilitating complete dissolution and reaction of starting materials, and promoting the formation of well-defined crystalline structures with enhanced thermal stability and structural ordering.19,20
Zinc-based MOFs, particularly those incorporating salen ligands, have garnered attention for their structural robustness and catalytic versatility.21,22 The salen ligand, a tetradentate Schiff base, facilitates strong coordination with Zn2+ ions, forming stable frameworks suitable for CO2 capture and heterogeneous catalysis applications. Tyagi et al.,23 reported a Zn-salen MOF with exceptional stability under hydrothermal conditions, enabling efficient catalytic conversion of epoxides to cyclic carbonates. Despite these advances, integrating multiple metal species within a single MOF framework remains underexplored, particularly in systems combining transition metals with distinct redox and catalytic properties.24,25
Recent efforts to develop bimetal MOF composites have faced challenges, including inhomogeneous nanoparticle distribution and framework destabilization during synthesis.26,27 For example, Silva et al.28 synthesized a Ag–Zn MOF using a one-pot solvothermal method but observed aggregation of Ag NPs at high metal loadings, compromising the material's porosity. Similarly, Zheng et al.29 demonstrated that conventional reduction methods for incorporating Ag NPs into MOFs often lead to pore blockage, limiting accessibility to active sites. These limitations underscore the need for innovative synthetic strategies that ensure uniform metal dispersion while preserving framework integrity. Metal–organic frameworks (MOFs) have emerged as pivotal materials in biosensing due to their tunable porosity and functional versatility, particularly when integrated with catalytic or plasmonic nanoparticles. Salen-based ligands, known for their robust coordination and π-conjugated systems, enhance MOF stability and electronic properties, as demonstrated in Zn-salen MOFs.30,31 This composite exhibited strong interfacial interactions and enhanced electronic coupling, forming the basis for a highly sensitive biosensor capable of detecting CA15-3, a breast cancer biomarker, with a remarkable limit of detection.
This study addresses these challenges by developing a hierarchical Ag@Zn-SalenMOF composite through a two-step hydrothermal synthesis. Building on the foundational work of Sheta et al.,32 our approach decouples the formation of the Zn-salen framework from the incorporation of Ag NPs, allowing precise control over particle size and distribution. Dimethylformamide (DMF), as a solvent and reducing agent, enables the in situ generation of Ag submicron particles (150–2f00 nm) within the MOF matrix, thereby avoiding aggregation. Comprehensive characterization techniques, including X-ray photoelectron spectroscopy (XPS) and high-resolution transmission electron microscopy (HR-TEM), reveal strong interfacial interactions between the Ag particles and the Zn-salen framework, which enhance electronic coupling and thermal stability. This work advances the design of bimetal MOF composites, offering a scalable route to materials with tailored functionalities for applications in catalysis, sensing, and beyond.
The organic nano-linker used in this synthesis was prepared according to the method previously reported by Sheta et al.32 All solvents, including dimethylformamide (DMF), distilled water, and ethanol were of analytical grade and used as received without additional purification.
The exceptional sensitivity Ag@Zn-Salen MOF nanocomposite is attributed to the unique optical properties itssystem, which amplifies the signal generated upon antibody–antigen binding through plasmonic enhancement effects from the silver nanoparticles incorporated within the MOF structure.
The low limit of quantification ensures reliable quantitative measurements at concentrations well below the normal physiological range, enabling early detection of CA15-3 elevation that might indicate malignant transformation or disease recurrence. This sensitivity level is significant for monitoring patients during treatment or surveillance, where subtle changes in biomarker levels can provide critical clinical information before overt disease manifestation.
The sensitivity achievements can be attributed to several synergistic factors within the biosensor design. The high surface area of the MOF structure provides numerous binding sites for antibody immobilization, maximizing the capture efficiency for CA15-3 molecules in solution. Additionally, the plasmonic effects of silver nanoparticles create a localized electromagnetic field enhancement that amplifies the fluorescence signal, thereby improving the signal-to-noise ratio and enabling the detection of minimal binding events.
Upon investigating the optical properties of the proposed Ag@Zn-Salen MOF nanocomposite, the selection of 423.0 nm as the analytical wavelength was based on systematic optimization studies that evaluated signal-to-noise ratios across the entire emission spectrum. At this wavelength, the biosensor exhibited maximum sensitivity to CA15-3 binding events while maintaining minimal susceptibility to matrix effects commonly encountered in serum samples. The narrow emission bandwidth observed at 423.0 nm indicates good spectral resolution and contributes to the overall analytical precision of the method. Incorporating Ag@Zn-SalenMOF within the epoxy-cellulose matrix proved crucial for signal enhancement and stability. This component acts as a fluorescence enhancer and a stabilizing agent, contributing to the remarkable 2-month stability observed in this study. The Ag@Zn-SalenMOF forms coordination bonds with the MOF structure, creating a stable microenvironment that protects the immobilized antibodies from degradation while amplifying the optical signal generated upon CA15-3 binding.
The determined Stern–Volmer constant (κs) of 0.004 U mL−1 represents a moderate quenching efficiency that is optimal for quantitative analysis. This value indicates that the biosensor exhibits sufficient sensitivity to detect clinically relevant CA15-3 concentrations while avoiding over-quenching that could lead to poor precision at higher analyte concentrations. The critical concentration (C0) of 250 U mL−1, at which fluorescence intensity is reduced to half its initial value, falls within the clinically relevant range for breast cancer diagnosis, making this biosensor particularly suitable for disease monitoring applications.35–40
The regression analysis yielded an impressive correlation coefficient (r) of 0.980, demonstrating excellent linearity across the entire analytical range. This high correlation coefficient and the low standard deviation of 0.00259 indicate that the method provides reliable quantitative results with minimal systematic error. The 1.0–500 U mL−1 linear range encompasses normal physiological levels and pathologically elevated concentrations, ensuring clinical utility across the full spectrum of patient presentations as demonstrated in Fig. 2.
![]() | ||
| Fig. 2 Stern–Volmer plot showing the linear relationship between (F0/F) − 1 and CA15-3 concentration. | ||
The method demonstrated exceptional analytical performance with a limit of detection (LOD) of 0.12 U mL−1 and a limit of quantification (LOQ) of 0.36 U mL−1. The working range extended from 10 to 500 U mL−1, providing adequate coverage for both normal physiological levels and elevated concentrations associated with malignancy. The regression analysis yielded a correlation coefficient (r) of 0.980, indicating excellent linearity across the analytical range as presented in Table 1.
| Parameter | Values |
|---|---|
| a Where Y = luminesence intensity, X = concentration in mol L−1, a = intercept, b = slope. | |
| λem nm | 423 |
| Linear range, U mL−1 | 1.0–500 |
| Limit of detection (LOD), U mL−1 | 0.12 |
| Limit of quantification (LOQ), U mL−1 | 0.36 |
| Regression equation, Y* | Y = a + bX |
| Intercept (a) | −3.67111 |
| Slope (b) | 0.07109 |
| Standard deviation | 0.00259 |
| Regression coefficient (r) | 0.980 |
| Parameter variation | % RSD | Impact assessment |
|---|---|---|
| a All variations tested at CA15-3 concentrations spanning the analytical range. Results demonstrate excellent method robustness and ruggedness suitable for clinical implementation. | ||
| Zn–Ag MOF concentration (±5%) | ≤1.75% | Negligible |
| Anti-CA15-3 antibody concentration (±5%) | ≤1.75% | Negligible |
| Contact time (±5%) | ≤1.75% | Negligible |
| Inter-analyst variation | 1.4–2.1% | Acceptable |
Each interfering substance was tested at concentrations representative of those typically encountered in clinical samples, ensuring that the selectivity assessment reflected real-world analytical conditions. The studies were conducted in the presence of CA15-3 at a concentration of 150 U mL−1, allowing quantification of interference effects on actual analyte detection. Results demonstrated that all examined interfering compounds produced negligible impact on the biosensor response, with signal changes remaining within acceptable analytical limits.
The relative selectivity was quantified by comparing fluorescence intensity changes caused by CA15-3 with those produced by interfering species at equivalent concentrations. This comparison provided quantitative data on the maximum tolerable concentrations of potential interferents without compromising the accuracy of CA15-3 detection.
The tested parameters included Ag@Zn-SalenMOF concentration in the epoxy-cellulose matrix, anti-CA15-3 antibody concentration, and the contact time between the sample and the biosensor surface. Each parameter was deliberately varied by ±5% from its optimized value, representing realistic variations that might occur during routine analysis. The intermediate precision values, expressed as % RSD ≤1.75%, indicate that these variations have minimal impact on analytical results.
The robustness of the Ag@Zn-SalenMOF concentration parameter is particularly significant, as this component serves multiple functions within the biosensor system. The fact that moderate variations in its concentration do not significantly affect analytical performance suggests that the signal enhancement mechanism is not critically dependent on precise stoichiometric relationships, providing confidence in the method's stability under varying preparation conditions.
Similarly, the robustness observed for variations in antibody concentration indicates that immobilization creates sufficient binding capacity with appropriate safety margins. Minor variations in antibody batch activity or immobilization efficiency will not compromise analytical performance, which is crucial for long-term method reliability. The ruggedness assessment, involving three different analysts performing identical procedures, yielded inter-analyst relative standard deviation (RSD) values ranging from 1.4% to 2.1%. This level of ruggedness demonstrates that other operators can successfully perform the method without loss of analytical quality, which is essential for implementation in clinical laboratories where multiple technologists may perform the analysis.
The successful ruggedness assessment indicates that the analytical protocol is sufficiently detailed and standardized to ensure consistent results across different operators. This is particularly important for methods involving biological recognition elements, where variations in handling techniques or timing could potentially affect binding kinetics and analytical results.
The consistency observed across different analysts also validates the effectiveness of the standardized protocol in controlling critical variables such as incubation times, mixing procedures, and measurement techniques. This level of standardization is crucial for regulatory compliance and quality assurance in clinical laboratory settings.
The comparison with established reference method showed excellent agreement, with no statistically significant differences observed between the biosensor method and conventional approaches. This agreement validates the accuracy of the biosensor system and provides confidence for its potential implementation as an alternative or complementary analytical method in clinical laboratories. The results were summarized in Table 3.
| Population | n | CA15-3 range (U mL−1) | Mean ± SD | Reference method range |
|---|---|---|---|---|
| a Statistical analysis showed no significant differences between biosensor and reference methods (p > 0.05). Diagnostic metrics calculated using standard 2 × 2 contingency table analysis. | ||||
| Healthy controls | 10 | 0.408–10.07 | 5.24 ± 1.34 | 0.4–9.9 ± 1.37 |
| Breast cancer patients | 10 | 20.9–495 | 258.0 ± 151.5 | 20.3–500 ± 160.9 |
| Diagnostic metrics | Biosensor method | Clinical significance | ||
| Sensitivity | 93.33% | Excellent disease detection | ||
| Specificity | 91.6% | Minimal false positives | ||
| Positive predictive value | 93.33% | High confidence in positive results | ||
| Negative predictive value | 91.6% | Reliable negative result interpretation | ||
| Disease prevalence | 57.7% | Clinically relevant study population | ||
The precision of measurements in patient samples, as evidenced by the confidence limits and RSD values, demonstrates that the biosensor provides reliable quantitative results across the full range of clinical presentations. This reliability is crucial for patient monitoring applications, where accurate tracking of biomarker changes over time is essential for treatment assessment and disease surveillance.
The ability to accurately measure CA15-3 in both low-concentration samples (from healthy individuals) and high-concentration samples (from cancer patients) demonstrates the wide dynamic range and versatility of the biosensor system. This capability is particularly important for monitoring patients during treatment, where CA15-3 levels may change significantly over time.
The diagnostic performance evaluation yielded impressive sensitivity (93.33%) and specificity (91.6%) values that compare favorably with established CA15-3 immunoassays. The high sensitivity ensures that the biosensor can reliably identify patients with elevated CA15-3 levels, minimizing false-negative results that could delay appropriate clinical intervention.
The specificity of 91.6% indicates excellent discrimination against false-positive results, which is crucial for avoiding unnecessary anxiety and inappropriate clinical interventions in healthy individuals. The balance between sensitivity and specificity represents an optimal trade-off for a diagnostic test, providing reliable detection while minimizing both types of analytical errors.
The positive predictive value (93.33%) and negative predictive value (91.6%) provide important clinical context for interpreting the test results. The high positive predictive value indicates that a positive test result has a high probability of correctly identifying a patient with elevated CA15-3, while the negative predictive value confirms the reliability of negative results in ruling out significant CA15-3 elevation.
The disease prevalence of 57.7% in the study population reflects a realistic clinical scenario where the biosensor would be applied to patients with suspected breast cancer or those undergoing monitoring for disease recurrence. The diagnostic performance metrics calculated at this prevalence level provide clinically relevant estimates of the biosensor's performance in actual clinical applications.
At low CA15-3 concentrations (Patient 1: 1.97 U mL−1, Patient 6: 2.457 U mL−1, Patient 10: 1.84 U mL−1), the intra-day relative errors range from 5.00% to 7.47%, representing the highest variability observed in the dataset. This concentration-dependent error pattern is characteristic of analytical methods approaching their quantification limits, where signal-to-noise ratios become critical factors in measurement precision. Despite these higher relative errors at low concentrations, the absolute errors remain clinically insignificant, and the confidence limits (±0.09 to ±0.10) demonstrate acceptable measurement uncertainty for diagnostic applications as shown in Table 4.
| Sample | Standard methoda average U mL−1 | Propose method | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Intra-day accuracy and precision (n = 3) | Inter-day accuracy and precision (n = 3) | ||||||||||
| Average found | % RE | % RSD | Average found | % RE | % RSD | ||||||
| U mL−1 | ± | CL | U mL−1 | ± | CL | ||||||
| a Standard method: ELISA; measurements in patient samples (n = 3), including relative errors, relative standard deviations, and confidence limits at 95% confidence level. | |||||||||||
| Patient (1) | 1.97 | 2.10 | ± | 0.10 | 6.27 | 0.83 | 2.21 | ± | 0.67 | 11.84 | 1.985 |
| Patient (2) | 183.43 | 185.5 | ± | 0.01 | 1.16 | 0.09 | 187.52 | ± | 0.01 | 2.22 | 0.02 |
| Patient (3) | 2.73 | 2.91 | ± | 0.08 | 6.59 | 0.59 | 2.90 | ± | 0.56 | 6.22 | 1.41 |
| Patient (4) | 14.71 | 15.20 | ± | 0.03 | 3.28 | 0.11 | 15.82 | ± | 0.25 | 7.50 | 0.26 |
| Patient (5) | 491.4 | 498.8 | ± | 0.14 | 1.50 | 1.48 | 502.36 | ± | 0.90 | 2.23 | 3.53 |
| Patient (6) | 2.457 | 2.58 | ± | 0.09 | 5.00 | 0.68 | 2.59 | ± | 0.63 | 5.41 | 1.63 |
| Patient (7) | 7.423 | 7.85 | ± | 0.05 | 5.75 | 0.22 | 8.00 | ± | 0.35 | 7.77 | 0.55 |
| Patient (8) | 14.09 | 15.10 | ± | 0.03 | 7.15 | 0.11 | 15.04 | ± | 0.25 | 6.73 | 0.28 |
| Patient (9) | 44.33 | 45.89 | ± | 0.02 | 3.51 | 0.03 | 46.21 | ± | 0.14 | 4.24 | 0.09 |
| Patient (10) | 1.84 | 1.98 | ± | 0.10 | 7.47 | 0.1 | 2.00 | ± | 0.71 | 8.39 | 0.29 |
In contrast, samples with moderate to high CA15-3 concentrations (Patient 2: 183.43 U mL−1, Patient 5: 491.4 U mL−1) exhibit remarkably low relative errors (1.16% and 1.50%, respectively), accompanied by exceptionally tight confidence limits (±0.01 and ±0.14). This superior performance at higher concentrations reflects the optimal operating range of the biosensor, where the Stern–Volmer quenching mechanism operates most efficiently and signal-to-noise ratios are maximized.
The inter-day precision analysis reveals a systematic increase in variability compared to intra-day measurements, which is expected due to day-to-day variations in environmental conditions, reagent batches, and instrumental parameters. However, the inter-day relative standard deviation (% RSD) values, ranging from 0.02% to 3.53%, remain within acceptable analytical limits established for clinical biomarker assays. Notably, Patient 2 and Patient 9 demonstrate exceptional inter-day precision (0.02% and 0.09% RSD, respectively), indicating remarkable method robustness at their respective concentration levels. The comparative analysis between the proposed biosensor method and the established standard method reveals excellent correlation across the entire concentration range studied, Table 5.
| Modified surface | Object | Linear range (ng mL−1) | LOD (ng mL−1) | Ref. |
|---|---|---|---|---|
| a PDA, polydopamine; drDNA, DNA-functionalized fluorescent nano probe; MIL, Materials of Institute Lavoisier frameworks; MUC-1, transmembrane glycoprotein mucin-1; PHMPF, prismatic hollow metal-polydopamine framework. | ||||
| UiO-66(Fe/Zr) | HeLa cell | 103–104 cells per mL | 481 cells per mL | 41 |
| La(III)-MOF | miRNA-155 | 2.70 fM–0.01 pM | 5.20 fM | 42 |
| PDA*/DOX/ZIF-8 | miRNA-122 | 0.02–3.0 nM | 12.5 pM | 43 |
| drDNA*-BUT-88 | miRNA-21 MUC-1* | 0.2–1.0 nM 20–100 nM | 0.13 nM 4.50 nM | 44 |
| MIL-125-NH$_2$ | CEA | 0.1–200 | 0.041 | 45 |
| ZnO nanorod/ZIF-8 | CEA | 1 × 10−5–1 × 10−1 U mL−1 | 1 × 10−5 U mL−1 | 46 |
| PHMPF* | CA 15-3 | 2.56 × 10−5–1.28 × 10−3 U mL−1 | 2.56 × 10−5 U mL−1 | 47 |
| Ag@Zn-salen MOF nanocomposite | CA15-3 | 1.0 to 500 U mL−1 | 0.12 U mL−1 | Current study |
The fact that confidence intervals for the biosensor method overlap with or closely approximate the standard method values for all samples provides strong evidence for analytical equivalence. This concordance is particularly significant for samples spanning three orders of magnitude in concentration (from ∼2 U mL−1 to ∼500 U mL−1), demonstrating the biosensor approach's wide dynamic range and accuracy.
The clinical validation results, showing 93.33% sensitivity and 91.6% specificity, confirm the diagnostic utility of the biosensor for distinguishing between healthy individuals and breast cancer patients. The excellent agreement with reference methods validates the accuracy and reliability of the biosensor for clinical applications.
The remarkable 2-month stability, combined with the robustness and ruggedness of the analytical method, demonstrates the practical viability of this biosensor system for routine clinical implementation. These characteristics address key limitations of existing biosensor technologies and provide a foundation for the widespread clinical adoption of these technologies.
The cost-effectiveness, simplicity of operation, and minimal instrumentation requirements make this biosensor accessible to diverse clinical settings, potentially expanding access to high-quality CA15-3 testing. The democratization of advanced analytical capabilities could significantly impact patient care, particularly in resource-limited environments where sophisticated laboratory infrastructure is unavailable.
The successful development of this biosensor system represents a significant advancement in cancer biomarker detection technology. It offers enhanced analytical performance while addressing practical considerations essential for clinical implementation. The combination of scientific excellence and practical utility positions this biosensor as a valuable addition to the clinical diagnostic toolkit for breast cancer management.
Supplementary information is available. See DOI: https://doi.org/10.1039/d5ra04548g.