Cystatin C (CysC), a protein, has replaced creatinine as a biomarker of kidney function and other diseases and has led to a surge in the research on the development of efficient CysC biosensors. The current CysC sensing technologies are remarkable in terms of selectivity and reproducibility. However, the complexity, cost, and space requirements of these methods render them unsuitable for real-time monitoring or point-of-care (PoC) implementations in healthcare settings. This review discusses the most recent developments in the field of CysC biosensing and to the best of our knowledge, this is the first focused review exclusively on CysC biosensing modalities. Our goal is to provide a thorough overview of the current state of CysC biosensors, and presenting mechanisms related to biosensor recognition and transduction. The review starts with clinical significance of CysC detection followed by detailed analysis of different CysC biosensing methods with emphasis on the necessity of PoC monitoring of CysC. We have also highlighted current challenges and an outlook on future perspectives. We anticipate that this study will play a key role in the understanding the working principle of CysC sensors and will aid in the designing of new efficient sensing modalities for the detection of CysC.
Human CysC is a stable basic 13 kDa protein composed of a single polypeptide chain with 120 amino acid residues and two disulfide bonds. CysC consist of a short alpha helix and a long alpha helix, which lies across five-stranded beta sheet. About 120 residues and two disulfide bridges define Family II, which includes cystatin C and S, and their derivatives. It is a critical cysteine protease inhibitor routinely generated by nearly every nucleated cell in the body and the production of CysC is irrespective of one's gender, age, or muscle mass. Because of its low molecular weight, it can easily cross the glomerular membrane and undertake near-complete metabolism within the tubular cells at the proximal end of the kidneys.6 The level of CysC in the blood will be in the normal range when kidneys functions well and the level of CysC varies if kidneys are not functioning well. These distinct characteristics of CysC make them as the best biomarker for estimating the GFR, capable of detecting even modest changes in GFR which can be used to detect early kidney disease and aid in the diagnosis of CKD.7 In addition to estimating the GFR, the CysC levels have been shown to be a prognostic marker for a variety of illnesses, such as tumors, polycystic ovarian syndrome (PCOS), brain diseases, cardiovascular problems, and endocrine disorders.
It has also been shown that fetal urine CysC correlates with postnatal kidney function at birth and serves as a biomarker of fetal renal tubular damage. The range of reported CysC levels in amniotic fluid is 0.06 to 2.68 mg L−1 in healthy full term pregnant women.13 The amniotic fluid samples from fetal uropathies and those with normal kidney functions display significant differences in the CysC values while creatinine levels were found to be very similar between these two groups.14 The CysC based diagnosis shows 96% diagnostic accuracy when it comes to distinguishing between fetuses with obstructive uropathies and those with normal kidney function, while creatinine has only 62% screening precision.15 Patients with hypertension and coronary heart disease show elevated CysC levels, which correlate with the severity of the illness.16 Recent research has also revealed that elevated CysC in the patients with acute heart failure and chronic kidney disease acts as an independent risk factor for cardiovascular death.17
The CysC levels in the urine are typically extremely small, between 0.03 to 0.30 mg L−1 and the urine CysC concentrations in healthy individuals do not significantly change with the age or muscle mass.18 An increase in the concentration of CysC in urine might be due to either a disturbed tubular reabsorption mechanism or a higher serum level over its reabsorptive threshold. Urinary CysC concentrations in such individuals can increase up to 200 times. The monitoring of CysC also plays a significant role in the prognosis of type 2 diabetic nephropathy as serum CysC gives more accurate values than serum creatinine.19 Additionally, the CysC levels are also linked with the thyroid glands functioning as mild thyroid dysfunction significantly alters CysC levels.20 CysC levels exhibit a more pronounced reduction in cases of hypothyroidism and it is higher in hyperthyroidism. Consequently, this implies that GFR is likely to be underestimated in hyperthyroidism and overestimated in hypothyroidism when considering CysC levels.21 Therefore, thyroid function parameters have to be considered when CysC is used to evaluate the kidney function.
It has long been known that the CysC is the main cysteine protease inhibitor in the cerebral spinal fluid (CSF). The CysC levels in the CSF of healthy adults are roughly 5.8 mg L−1, which is 5.6 fold greater than in plasma.22 Patients with disorders like meningitis, encephalitis, chorioretinitis, and myelitis have been found to have higher concentrations of CysC in their CSF, whereas patients with multiple sclerosis, acute Guillain–Barré syndrome, and chronic inflammatory demyelinating polyneuropathy had significantly lower levels in their CSF.23 This drop in concentration could be due to decreased CysC synthesis or an increased degradation in inflammatory lesions of the central nervous system. Another study evaluated the possibility of CysC as a CSF biomarker for pain in humans based on the hypothesis that increased expression of CysC in the dorsal horn leads to its increased secretion in CSF. They confirmed that pain induces the production and discharge of CysC in the dorsal spinal cord and it overflows into the CSF so that the level will be high during active pain states.24
Proteases and their inhibitors are in equilibrium, and this balance governs most normal and pathological processes.25 It was discovered that CysC might be involved in regulating the expression of cathepsin B, a cysteine protease primarily present in fast spreading cancer cells.26 The less common cathepsin B/CysC complexes in the sera of cancer patients compared to healthy controls or patients with benign diseases suggested an imbalance between the enzyme and its inhibitor in these patients.27
It was discovered that seminal plasma had a mean CysC concentration of 51.0 mg L−1. This is approximately 36 times greater than that found in healthy adult blood plasma and is due to the larger synthesis of CysC by the male genital tract glands.28 Similarly, the CysC level is upregulated in inflammatory conditions, such as periodontal disease and gingivitis.29,30 Compared to periodontally healthy individuals, those with gingivitis or periodontitis show approximately 1.3 times greater amounts of CysC in the saliva.31 In cerebral amyloid angiopathy, a disease characterized by amyloid beta-peptide deposits within blood vessels of brain, hemorrhage is highly predicted by the degree of CysC accumulation.32 Also, increased serum levels of CysC are associated with possible future stroke events in the elder patients especially with carotid atherosclerosis.33 Similarly, the CysC level is higher for amyotrophic lateral sclerosis, which is a fatal neurologic disease characterized by progressive motor neuron degeneration.34 The higher value of CysC in comparison with healthy controls highlights the severity of the disease in terms of faster progression rate and shorter survival. The overall clinical significance of CysC is summarized in Fig. 1.
![]() | ||
Fig. 1 Schematics showing the clinical significance of CysC in various disorders. The authors have drawn the figure using https://www.BioRender.com. |
Based on these reports, it is clear that CysC concentrations in body fluid are strongly correlated with a variety of pathologies. The majority of the reports highlight the importance of CysC for monitoring renal impairment along with other mentioned diseases. A number of techniques including conventional methods and newly introduced methods are used in various clinical settings to monitor CysC in various body fluids. The following section discusses these conventional and newly introduced methods for the sensing of CysC.
PETIA and PENIA methods have been widely explored and studies show that PENIA methods for the detection of CysC are more sensitive.38 PENIA methods have been reported to exhibit optimum performance in dilute solutions, thus making it suitable for pediatric population where small volumes of the samples are preferred. The detection of CysC in the same blood samples with the PETIA assay and the PENIA method have reported differences in values,35,39 which could be attributed to difference in reactivity to CysC antibodies, or difference in standards or substrates.39 Most of the PETIA and PENIA studies use rabbit anti-human antibodies for the analysis. Jonkisz et al. compared the CysC values obtained by PETIA and PENIA assays in the dog by using 40 dogs in four groups with different stages of chronic kidney disease based on serum creatinine values.40 The serum CysC was measured using PETIA and PENIA, and the correlations with the serum urea and creatinine values were studied. The results confirmed the higher sensitivity and correlation with other accepted indicators of renal function in the values obtained in PENIA when compared to the PETIA. Another study was conducted to evaluate the performance of the analytical technique using two reagents for the detection of traceable urinary CysC by Bargnoux et al.38 Siemens reagents were used for PENIA and DiaSys reagents were used for PETIA for assessing their clinical relevance in healthy and individuals with tubular dysfunction. They found that results were comparable in both methods.
ELISA is a widely accepted method for measuring CysC levels due to its sensitivity and specificity.41 ELISA has been used as a method of measurement to prove the effect of non-renal factors like corticosteroids, which could lead to an increase in the plasma CysC levels without any effects on kidney function. The hypothesis was tested in rats by studying the effect of corticosteroids on the levels of CysC in their tissues and its effect on GFR.42 The data showed that corticosteroids may increase CysC levels in the plasma by promoting its production, without any decrease in GFR. Wilson et al. conducted quantitative ELISA to assess the plasma CysC levels in 104 patients diagnosed with ALS.34 They found that the CysC levels in amyotrophic lateral sclerosis patients were significantly elevated in plasma and diminished in CSF compared to healthy controls. Additionally, the CysC level in CSF is correlated to the rate of ALS disease progression, which suggests it can be used as a surrogate marker of disease progression and survival.
It is difficult to get efficient paired monoclonal antibodies against small molecules such as CysC, which can be used in clinical set ups and this led to use of hybridoma and phage display techniques to obtain monoclonal antibodies against CysC. Jiang et al. introduced efficient paired monoclonal antibodies against CysC by establishing a new double-antibody-sandwich ELISA (DAS-ELISA)-based measurements.43 Mommaerts et al. developed a double indirect sandwich ELISA to estimate the total CysC present in the biological fluids (both cleaved and uncleaved forms).44 The ratio of cleaved and uncleaved concentrations was used to evaluate the extent of CysC cleavage using this novel method.
Despite being sensitive and rapid tests, automated PETIA, PENIA and ELISA are limited in their application in ordinary clinical practice due to costly procedures and use of specific immunoassay apparatus. The requirement for relatively expensive test kits and clunky plate readers, on the other hand, restrict their applicability for point-of-care (PoC) diagnostics. To overcome these constraints, PoC biosensors based on new biosensing methods can be utilized for CysC monitoring.
A competitive Enzyme-Linked Aptamer Sorbent Assay (ELASA) employing single-stranded DNA aptamers was developed to measure the levels of CysC in serum samples.55 The chosen aptamer exhibited a low dissociation constant (Kd) of 65.5 ± 0.007 nM and limit of detection (LOD) of the assay was found to be 216.077 pg mL−1. The analytical outcomes of this method were in line with those from widely used commercial kits for the CysC detection. In another work, Lopez-Silva et al. assessed the use of SOMAscan technology, which is an aptamer-based technology, for the detection of CysC from kidney disease patients using immunoassays as the gold standard.56 The study suggested that the SOMA scan is an efficient and relatively reliable technique for quantifying CysC and to correlate clinical outcomes. Recently, Dubin et al. also studied the analytical and biological variability of the SomaScan V4 aptamer proteomic assay in plasma samples from patients with moderate to severe CKD.57 The assay's robust performance was demonstrated by low intra-assay and within-subject variability, as well as high Rho correlations between aptamer and traditional CysC assays. These correlations were comparable to those reported in individuals without CKD.
Papain is a cysteine protease, which consists of a polypeptide chain with three disulfide bridges and a sulfhydryl group which are necessary for activity of the enzyme.60 A major group of the cysteine proteases are structurally related to papain and are therefore named as papain-like cysteine proteases. The proteins, which block papain-like cysteine proteases have been recognized in 1946 by D. Grob and CysC is one among them.61 When papain mixes with an excess of human CysC or chicken cystatin, an enzyme–inhibitor complex was formed immediately.62 Abrahamson et al. investigated the formation of inhibitor complex and identified that the peptidyl bond of the conserved glycine residue in human CysC and chicken cystatin might be the substrate-like inhibitory reactive site.62 They also proposed that the inhibition of cysteine proteinases might be due to the simultaneous interactions with the inhibitory reactive sites of the protease and cystatins. Lindahl et al. studied the interaction between recombinant human CysC and the cysteine proteinases papain by spectroscopic, kinetic and equilibrium methods.63 The dissociation equilibrium constant for binding CysC to papain was approx. 11 fM. The affinity between papain and CysC decreased with increasing size of the inactivating group, which is similar to earlier observations with the chicken cystatins. The CysC–papain interactions were extensively used for the designing of biosensors for the detection of CysC in conjunction with most of the new biosensing based transduction mechanisms which are discussed in the following sections.
![]() | ||
Fig. 2 (a). Schematic representation of the sandwich-immunoassay with labeled enzyme alkaline phosphatase (AP) for catalyzing silver deposition reaction for the CysC detection. Reprinted with permission from ref. 64. Copyright © 2019 Elsevier B.V. (b) Electrochemical detection of CysC by protein converting strategy in combination with T7 Exo-assisted protein cyclic enzymatic amplification. Reprinted with permission from ref. 69. Copyright © 2016, American Chemical Society. |
Another label-free electrochemical immunobiosensor for the rapid detection CysC has been developed by Krishnan et al. Here, anti-CysC antibody was immobilized on a carbon working electrode that had been modified with a graphene oxide-chitosan nanocomposite.65 The CysC binding to the antibody increased the charge transfer resistance of the electrode surface and this was evaluated by using ferricyanide as a redox mediator. The sensor detected CysC ranging from 1 to 10 mg L−1 with a detection limit of 7.8 μg L−1. The sensor was able to produce the response in just 420 s, which was shorter than conventional ELISA and other reported electrochemical sensors. The sensor response was evaluated with patient samples and found that the sensor can be employed used to predict the risk of developing diabetic retinopathy. In addition, the prediction accuracy of the sensor was marginally greater than that of the data acquired using the conventional ELISA approach. Similarly, a label-free electrochemical immunosensor for CysC was introduced by using ferrocene functionalized-graphene platform on a gold electrode.66 Here, the ferrocene acts as the redox active indicator of immunoassay production on the electrode surface. The complete sensor platform was created by drop-casting polyethyleneimine film followed by the addition of CysC. The addition of CysC decreases the current due to the electron transfer hindering by these adsorbed molecule and this was used for the detection of CysC. The sensor platform exhibited a response with a range of 0.1 to 1000 ng mL−1 and a LOD of 0.03 ng mL−1.
In another work, Chakraborty et al. prepared a novel nanostructure composite of ZIF-8-Cu1−xNix(OH)2@Cu, which was made by combining Cu shell on zeolitic imidazolate frameworks (ZIF-8) and Cu1−xNix(OH)2.67 In order to develop the CysC sensor, the poly(3,4-ethylenedioxythiophene) polystyrene sulfonate (PEDOT:PSS) coated indium tin oxide (ITO) electrode had been modified using the synthesized ZIF-8-Cu1−xNix(OH)2@Cu nanocomposite. Streptococcal protein G (SPG) was then used to functionalize the electrode, enabling it to bind the Fc region of anti-cystatin C. A concentration increase of CysC from 0.1 ng mL−1 to 1000 ng mL−1 increased the sensor's differential pulse voltammetry (DPV) current response, and the LOD was determined to be 33 pg mL−1. The developed sensor's excellent sensor performance may be attributed to its high surface coverage, excellent redox properties, improved electroactive sites, and synergistic catalytic activities. The sensor's clinical viability was confirmed through the measurement of CysC derived from human serum.
It is well known that papain is a cysteine protease, which can bind to CysC and this interaction was used for the fabrication of an electrochemical sensor for CysC.68 A screen printed electrode (SPE) with carboxyl activated multiwalled carbon nanotube electrode was used to modify the papain through binding between amino group of papain and carboxyl group on electrode surface by 1-ethyl-3-(-3-dimethylaminopropyl) carbodiimide (EDC)-N-hydroxysuccinimide (NHS) linking. The CysC–papain interaction mediated transduction of electrons by the sensor electrode was used to get the quantitative detection of CysC by using potassium ferricyanide as an indicator. The selectivity of the sensor was evaluated by using other urine-based biomarkers and found that there is no significant sensor response, which confirms the superior selectivity. The LOD of the sensor was found to be 60 ng L−1.
In a different approach, Chai et al. developed a novel immunoassay method for the CysC detection by using immunoreaction-induced DNA strand displacement and T7 Exonuclease (T7 Exo)-assisted protein cyclic enzymatic amplification.69 They have used Au@Fe3O4 as magnetic separator, which was labeled by conjugated DNA (DNA1) and the DNA substrate of T7 Exo (DNA3). This was followed addition of output DNA, antibody 1 (Ab1) and hexanethiol and incubation for 16 h. This resulted in the formation of DNA3-output DNA complex and DNA–Ab1 complex on the surface of Au@Fe3O4. This bio complex was incubated with DNA2–Ab2 complex and T7 Exo and CysC. The schematics showing the complete steps of sensor fabrication are given in Fig. 2(b). In the presence of CysC, the sandwich immunoreaction between DNA1–Ab1 and DNA2–Ab2 induces the proximity hybridization between DNA2 and DNA3 and thus displace output DNA from the output DNA/DNA3 duplex and leads to the formation of DNA2/DNA3 duplex. The T7 Exo digest the DNA2/DNA3 duplex with the release of DNA2, which acts as competing DNA to displace output DNA further. This output DNA strand was captured using an electrode modified with hairpin DNA, and a hybridization chain reaction was initiated on the electrode surface. The quantification of output DNA, which is the indirect measurement of CysC was done by using thionine as electron mediator. The linear range of the developed sensor was significantly larger, ranging from 0.01 pg mL−1 to 30 ng mL−1, with a low detection limit of 3 fg mL−1.
![]() | ||
Fig. 3 The schematics showing the stepwise modifications of the surface of the IDE immunosensor for the detection of CysC by measuring the double layer capacitance by the changes in phase angle. Reprinted with permission from.70 Copyright (2020), with permission from Elsevier. |
![]() | ||
Fig. 4 The schematics of the sensor working principle of FET based CysC sensor by using extended-gate field-effect transistor (EG-FET). Reproduced with permission from ref. 73. Copyright © 2024 Elsevier B.V. |
An innovative dual detection method for CysC in the blood plasma by surface-enhanced Raman spectroscopy (SERS) and electrochemistry was presented by Hassanain et al.74 Here, the gold-coated silicon nanopillar electrode was modified with thiol and CysC antibody to recognize the CysC specifically and was followed by buffer wash for the quick isolation of CysC. The isolated CysC was then subjected to rapid reduction of disulfide bonds, and the reduced protein could be easily immobilized on conductive gold coated silicon nanopillar substrate using strong gold–sulfur bonds. Following that, the electrochemical DPV and SERS measurements were used to quantify the reduced protein detection. A reproducible surface-enhanced Raman spectroscopy measurement is made possible by the oriented immobilization of CysC. The LOD of the developed sensor for SERS and DPV was 1 pM and 62.5 nM, respectively. These values satisfy the requirements for tracking renal failure and Alzheimer's disease.
In another work, the interdigitated electrodes were modified with carbon nanowire and papain and this was used as the detection probe for CysC.75 The immobilization of papain was done by covalent bonding between the carboxyl group of carbon nanowire and the amine group of papain. By measuring the voltage vs. current responses using a picoammeter, this papain-modified electrode surface was able to detect CysC concentration in the range of 100 pg mL−1 to 3.2 ng mL−1. The current response gradually increased with increase in the CysC concentration from 100 to 3200 pg mL−1 and the LOD was found to be 200 pg mL−1. The developed sensor was highly specific as other interfering molecules such as creatinine and gliadin are not properly binding to the immobilized papain.
In a different approach, an effective photo electrochemical immuno sensor was developed for the sensitive detection of CysC using titanium dioxide nanotube arrays (TNAs) after modifying with the CysC-specific nanobody.76 From a library of phage display nanobodies, they have isolated CysC-specific nanobodies. Under irradiation the TNAs produced by electrochemical anodization showed a large and stable photocurrent response. CysC was quantitatively detected in the range of 0.72 pM to 3.6 nM using the change in photocurrent that occurs when CysC binds to the CysC-specific nanobody attached to TNA. The created immunosensor demonstrated assay accuracy and a detection limit of 0.14 pM, indicating selectivity and stability. A comparison of all the electrochemically based sensors reported for CysC detection is presented in Table 1.
Method | Electrode | Reaction | Signal | LOD | Detection range | Selectivity | Real sample | Ref. |
---|---|---|---|---|---|---|---|---|
Voltammetry-DPV | Sandwich assay by using a capture antibody modified SPCE | Alkaline phosphatase label in anti-CysC antibody induces a chemical reaction in addition of mixture of 3-IP and silver nitrate in the assay | Current increases | 6.0 ng mL−1 | 10–100 ng mL−1 | — | Serum | 64 |
Voltammetry-DPV | Antibody modified graphene oxide-chitosan (GO-Chit) nanocomposite/GCE | Ferricyanide [Fe(CN)6]3− mediated electron transfer | Current decreases | 0.0078 mg L−1 | 1–10 mg L−1 | Ascorbic acid, uric acid, citric acid, alanine, asparagine, tyrosine, nicotinamide adenine dinucleotide, cysteine, creatinine, and cystine (Cys) | Serum | 65 |
Voltammetry-square wave voltammetry | Antibody modified polyethyleneimine-graphene oxide-ferrocene nanofilm/GCE | Ferrocene mediated the electron transfer studies | Current decreases | 0.03 ng mL−1 | 0.1–1000 ng mL−1 | — | Serum | 66 |
Voltammetry-DPV | ZIF-8-Cu1−xNix(OH)2@Cu nanocomposite modified on PEDOT:PSS coated ITO electrode | CysC immobilization with antibodies | Current increases | 33 pg mL−1 | 0.1–1000 ng mL−1 | Ascorbic acid, creatinine, lipocalin 2, human serum albumin, hemoglobin | Serum | 67 |
Voltammetry-cyclic voltammetry (CV), DPV | Papain immobilized on screen printed MWCNT electrode | Papain–CysC interaction | Current decreases | 0.58 ng L−1 | 0.6 × 10−5–6.6 × 10−3 ng μL−1 | Creatinine, albumin and gliadin | — | 68 |
Voltammetry-DPV | AuNP/C60/GCE | Immunoreaction-driven DNA strand displacement | Current increases | 3 fg mL−1 | 0.01 pg mL−1 to 30 ng mL−1 | Influenza, thrombin, hemoglobin, and matrix metalloproteinase-2 | Serum | 69 |
Capacitance | Monoclonal antibodies modified polypyrrole/carbon nanotube nanohybrid film/Gold interdigitated electrode | Change in capacitance on CysC binding | Difference of the phase angle-increases | 28 ng mL−1 | 0–300 ng mL−1 | — | Serum | 70 |
Voltammetry-CV, DPV | MIP based SPCE electropolymerizing pyrrole (Py) with carboxylatedPy (Py-COOH) in the presence of Cys-C and MWCNT and selective removal | Selective recognition of CysC with MIP | Current decreases | 0.5 ng mL−1 | 0.5–20.0 ng mL−1 | Creatine kinase-MB, ascorbic acid, creatinine and bovine serum albumin (BSA) | Serum | 71 |
FET | CysC antibodies modified SiNW FET | CysC immobilization with antibodies | Current decreases | 0.2529 ag mL−1 | 1 ag mL−1 to 1 ng mL−1 | Recombinant human retinol binding protein | — | 72 |
FET | BSA/Papain/Au NPs/adhesive/laser-induced graphene electrode | Papain–CysC interaction | Current decreases | 0.05 ag μL−1 | 5 ag μL−1 to 50 ng μL−1 | BSA, papain, uric acid, creatinine, urea, gliadin, glucose | Urine | 73 |
Spectroelectrochemical-SERS, DPV | Gold nanopillar electrode | Thiol chemistry used to develop target-specific and recyclable extractor chip | Raman band intensity of reduced, DPV current increases | 62.5 nM by DPV and 1 pM by SERS | 0.0625 μm to 1 μm for DPV 100 nM to 1 pM for SERS | — | Plasma | 74 |
Current-volt measurement | Carbon nanowire-modified surface with interdigitated electrode | Papain–CysC interaction | Current increases | 200 pg mL−1 | 200 pg mL−1 to 3.2 ng mL−1 | Creatinine and gliadin | — | 75 |
Photoelectrochemical | CysC-specific Nb modified TiO2 nanotube arrays | CysC immobilization with nanobodies | Photocurrent decrease | 0.14 pM | 0.72 pM to 3.6 nM | Prostate-specific antigen, human chorionic gonadotropin, and α-fetoprotein | Serum | 76 |
Lateral flow immunoassays (LFIA) have been explored as PoC devices as they allow fast, simple and cost-effective features along with high specificity, sensitivity, and low LOD. Zhang et al. reported quantitative LFIAs for the colorimetric detection of CysC by using two Au nanostructures: AuNPs and Au nanorods (AuNR).78 The conjugation of Au nanostructure-monoclonal antibodies was confirmed by UV-Vis spectroscopy and the authors have arrived at the optimum composition for the best LIFA performance by measuring the hydrodynamic radius of monoclonal antibodies-Au nanostructure conjugate through dynamic light scattering.
The LIFA signal intensity at various concentrations of CysC was monitored at optimum conditions and was used for the quantitative detection of CysC. The detection range was enhanced by the wash step to remove the unbound analyte to free up the space so that monoclonal antibodies–Au nanostructures conjugate could efficiently bind to the capture antibody-immobilized CysC. The authors found that AuNRs conjugates were more sensitive than AuNPs, despite AuNPs having greater test line signal intensities. The LOD of AuNRs based conventional and step-wash LFIA in dipstick format were found to be 0.08 mg L−1 and 0.35 mg L−1, respectively.
Recently, a colloidal gold-based LFIA device was developed for the quantitative detection of CysC with a minimum assay time of 15 min.79 The AuNP-CysC antibody conjugate was synthesized by covalent binding between carboxyl functionalized AuNP and EDC modified antibody by EDC-NHS chemistry. The binding between the CysC and antibody was measured colorimetrically. The schematic of the experimental procedure for the detection of CysC by using the developed LFIA strips is given in Fig. 5. The developed device is able to detect CysC in the range from 0.5 to 7.5 μg mL−1, with a LOD at 0.18 μg mL−1. The developed LFIA device exhibited satisfactory stability of more than 30 day when stored at 50 °C thus indicating the long shelf life at room temperature. In addition, the developed device did not require expensive readers or complex conjugation protocols, which helps in fast device development towards easy commercialization. Furthermore, the device has the potential to provide home-based diagnostics as it can be connected to a mobile phone.
![]() | ||
Fig. 5 The schematic of the experimental procedure for the detection of CysC by using the developed LFIA strip. The red line represents the baseline correction and shaded area represents the tolerance. Reprinted with permission from ref. 79. Copyright © 2024 by the authors. |
In another work, a arginine liposome based colorimetric assay was created for the detection of CysC.80 Here, arginine (Arg)-loaded liposomes and magnetic beads (MBs) were used as surfaces for the conjugation of CysC antibodies (Ab1 and Ab2), resulting in Ab1-MBs and Ab2-Arg-liposomes, accordingly. The schematic illustration of the principle and operation steps of CysC sensor made by a arginine liposome-amplified colorimetric immunoassay is given in Fig. 6. When CysC is present, it will be captured by Ab1-MBs and this was followed by the addition of Ab2-Arg-liposomes to form the Ab1-MBs-CysC-Ab2-Arg-liposomes immuno-sandwich complex. Triton X-100 was added as a last step to break the liposomes and release Arg, which could cause the aggregation of gold nanoparticles. Utilizing this discoloration, which has a linear range of 10–100 μg L−1, CysC can be detected quantitatively and visually.
![]() | ||
Fig. 6 Schematic illustration of the principle and operation steps of CysC sensor made by a arginine liposome-amplified colorimetric immunoassay. Reprinted with permission from.80 Copyright © 2022 Elsevier B.V. |
Lin et al. reported a simple, immune-independent and label-free fluorescent detection for CysC using BSA-stabilized Au nanoclusters (AuNCs).82 This method relied on the BSA scaffold degradation caused by the specific inhibition of papain activity by CysC and cysteine protease activity of papain. The fluorescence of BSA-AuNCs could be effectively quenched by papain, and restored by the presence of CysC.The concentration of CysC was determined by observing the turn-on fluorescence signal of the AuNCs–papain combination. The schematics showing the working principle of the sensor and the fluorescence responses of BSA-AuNCs with different concentrations of CysC as well as its calibration plot are given in Fig. 7. This novel protein-templated AuNCs based protease inhibitor assay was able to detect CysC concentrations in the range of 25 ng mL−1 to 2.0 μg mL−1 with high selectivity. The detection limit of the sensor was found to be 4.0 ng mL−1 and this was around 40 times better than commercial sensing approaches.
![]() | ||
Fig. 7 (a) Schematics of the BSA-AuNCs-based fluorescent sensor for the detection of Cys C (b) fluorescence responses of BSA-AuNCs in presence of papain and different concentrations of CysC. (c) Plot showing the relation between fluorescence intensity of BSA-AuNCs and concentration of CysC. Inset shows the calibration plot of the normalized intensity versus the concentrations of CysC in the range of 00.25 to 2.0 μg mL−1. Reprinted with permission from.82 Copyright © 2012 Elsevier B.V. |
Tao et al. employed a similar method of hydrolyzing the BSA template with papain to create a label-free CysC sensor that used PbS@BSA quantum dots (QDs) that emit near-infrared (NIR) light.83 Pb2+ in PbS@BSA QDs was primarily bound to the amino and carboxyl groups of BSA, and the papain digestion process reduced the QDs' fluorescence. An analysis and optimization of the relationship between papain concentration and NIR fluorescent intensity was done. Before papain was combined with QDs, CysC could inhibit its protease activity while preserving the intensity of the fluorescence. This was used in the development of the CysC sensor, and it was discovered that the LOD was 0.02 μg mL−1. They have also found CysC in 50% serum, and they suggest that this is a good way to detect CysC quantitatively in comparison to other optical detection techniques, as well as for qualitative bioimaging.
In a similar approach of hydrolyzing the BSA, a novel fluorescence on-off sensor was made by combining using aggregation-induced emission (AIE) feature of 2-(4-bromophenyl)-3-(4-(4-(diphenylamino)styryl)phenyl) fumaronitrile (TPABDFN) and BSA.84 The BSA loaded with TPABDFN shows “On” state due to the restricted intramolecular motions. The BSA could be hydrolyzed by papain and this causes a decrease in the fluorescence. The presence of CysC prevents the template protein from being hydrolyzed by proteases, thus maintaining the fluorescence and the fluorescence intensity was utilized to successfully to quantify CysC. The developed sensor was able to detect CysC with a linear range of 12.5 ng mL−1 to 800 ng mL−1 and a LOD of 7.10 ng mL−1. In another work, an immune-pillars chips were used for fluorescent-based detection of CysC by using fluorophore labeled antibody on the micro beads.85 The micro beads assembled in the nanopillars provide the binding site for CysC and the fluorescence intensity was measured. The fluorescence intensity increased with increasing CysC concentration in the immuno-pillar chips and the sensor was able to detect CysC from 0 to 50 ng mL−1 with LOD of 3 ng mL−1. The developed immunochip showed a significant decrease in the reaction time (20 min) when compared to the current ELISA technique (240 min) and also needs only minimum quantity reagent for the sensing.
Fluorescence-based LFIA sensors have also been reported for the CysC detection towards PoC applications. An aptamer-antibody pair-based sandwich lateral flow immunoassay is proposed by Natarajan et al.86 They have fabricated the sensor by conjugating CysC selective aptamers to the organic dye Alexa fluor-647. In presence of CysC, a complex is formed between the aptamer and CysC, provided the LFIA strip's test zone is immobilized with CysC antibody. The developed assay was able to achieve the LOD of 0.013 μg μL−1, which is better than any antibody-based kit. In summary, the developed LFIA aptamers-based sensor provides a fast, sensitive, low cost POC sensor for CysC detection from human samples. A similar fluorescent-based LFIA sensor was developed in a sandwich format by using a fluorescent organic dye conjugated monoclonal antibody.87 By monitoring the fluorescence intensity with respect to the analyte content, the CysC level in the sample could be quantified. The calibration of the kit was linear in nature between 0.023–32 μg mL−1 and the LOD was found to be 0.023 μg mL−1.
Utilizing a specific interaction between CysC and immobilized papain in a solution, a SPR sensor was fabricated for the detection of CysC.90 An amino modified gold surface was used to immobilize papain by activating the surface with NHS and EDC. Amino modification of gold surface was achived by cysteamine so that it interacts better with the immobilized papain and makes it easier for CysC to bind. This sensor had a sensitive LOD of 0.09 μg mL−1 and a responsive range of 0 to 0.6 μg mL−1. A comparison examination using the PETIA method was carried out in order to validate these results and it showed good agreement. This biosensor could precisely measure the quantities of CysC in a range of biological fluids such as blood plasma, urine, and saliva. Another SPR-based sensor was developed by same research group by using bromelain or chymopapain or ficin for the specific CysC detection.91 The enzymes bromelain, ficin, and chymopapain are categorized as cysteine proteases, sharing a similar mechanism of action. In presence of CysC, the formation of an enzyme–inhibitor complex was formed by immobilized bromelain or chymopapain or ficin on the biosensor surface. As part of the optimization process, the effects of interaction pH and the influence of enzyme concentration on the SPR signal were examined. It is important to note that there was an outstanding degree of similarity between the calibration curves for the three different enzyme-based sensors. The developed sensor was able to detect CysC in the range of 0 to 0.6 μg mL−1 with a LOD of 0.1 μg mL−1. In addition, the developed sensor could accurately detect CysC levels in the blood plasma, urine, and saliva.
Another chemiluminescence sensor for CysC detection was developed by using the perylenetetracarboxylic acid (PTCA) based donor and the ZnO@L-CysNPs as the acceptor.93 In order to create PTCA@Fe(III)-MIL-88B-NH2@Au/Ab1, PTCA was first mixed with a Fe containing metal–organic framework (Fe(III)-MIL-88B-NH2). This was followed by the in situ growth of AuNPs and antibody binding. This complex was drop casted on GCE so that CysC could bind to the Ab1. ZnO@L-CysNPs and Ab2 of CysC and BSA were combined in parallel to create the ZnO@L-CysNPs/Ab2 complex. Fig. 8 provides the schematics for the synthesis of PTCA@Fe(III)-MIL-88B-NH2@Au/Ab1 and ZnO@L-CysNPs/Ab2, as well as the preparation of the ECL-FRET immunosensor. The specific interaction of the antigen with the antibody links the ZnO@L-CysNPs/Ab2 to the PTCA@Fe(III)-MIL-88B-NH2@Au/Ab1 when CysC is present. Through a subsequent resonance energy transfer between the PTCA@Fe(III)-MIL-88B-NH2@Au donor and the ZnO@L-CysNPs acceptor, this binding process may be able to quench the electrochemiluminescence (ECL) of PTCA. The developed ECL-resonance energy transfer (RET) based biosensor exhibited a linear range between 0.01 pg mL−1 and 100 ng mL−1 with a detection limit of 2.2 fg mL−1.
![]() | ||
Fig. 8 (A) Schematic showing the preparation of ZnO@L-CysNPs/Ab2 (A) and synthesis of PTCA@Fe(III)-MIL-88B-NH2@Au/Ab1. (B) The preparation of ECL-FRET immunosensor as well as (C) the detection of CysC. Reprinted with permission from ref. 93. Copyright © 2023 Elsevier B.V. |
The sensor chip incorporates a glass transducer with a sensitive layer produced on its surface by immobilizing CysC, resulting in a stable platform for the detection. The second step involves incubating the sample with constant antibody concentrations and injecting the incubated sample onto the surface of the sensing platform. Using RifS, the antibody's binding to the immobilized CysC was identified. The amount of CysC in the sample has an inverse relationship with the binding response. The developed sensor was able to accurately quantified CysC in serum samples in the medically relevant range of 0.53–1.02 mg L−1. A comparison showing the characteristics of optical sensors developed for the detection of CysC is given in Table 2.
Method | Sensing strategy/element | Reaction | Signal | LOD | Detection range | Selectivity | Real sample | Ref. |
---|---|---|---|---|---|---|---|---|
Absorbance | Papain modified MWCNT in solid support and casein beads | Protease assay and its inhibition by CysC | Increases | 0.5–5 mg L−1 | — | 77 | ||
Colorimetric | Antibody conjugates of AuNPs and AuNRs | Lateral flow assay | Increases | AuNPs – 0.42 mg L−1 AuNRs – 0.35 mg L−1 | Up to 12.5 mg L−1 | Serum | 78 | |
Colorimetric | Antibody conjugates of AuNPs | Lateral flow assay | Increases | 0.18 μg mL−1 | 0.5–7.5 μg mL−1 | Hemoglobin, bilirubin, triglycerides, rheumatoid factor | Serum, blood | 79 |
Colorimetric | Ab1-MBs and Ab2-Arg-liposomes | Arginine-liposome-amplified colorimetric immunoassay | Discoloration | 4.32 μg L−1 | 10–100 μg L−1 | BSA, FBS, Cr, blood urea nitrogen | Serum | 80 |
Fluorescence | BSA protected AuNCs | Fluorescence quenching by papain, and restoration by the presence of CysC | Increases | 4.0 ng mL−1 | 25 ng mL−1 to 2.0 μg mL−1 | Thrombin, hemoglobin, glucose oxidase and trypsin | Urine | 82 |
Fluorescence | PbS@BSA QDs | Fluorescence quenching by papain, and restoration by the presence of CysC | Increases | 0.02 μg mL−1 | 0.05–2 μg mL−1 | K+, Mg2+, Zn2+, Ca2+, Mn2+, uric acid, dopamine, glucose, fructose, lactose, lysine, cysteine, glutamic acid, tyrosine, tryptophan, alanine, urea, haemoglobin, DNA | Serum | 83 |
Fluorescence | TPABDFN co loaded BSA | Aggregation-induced emission TPABDFN and BSA quenched by papain and restored by the coexistence of CysC | Increases | 7.10 ng mL−1 | 12.5–800 ng mL−1 | lipase, pancrelipase,neuraminidase, thrombin,IL-6,IL-10,TNF-α,and TGF-β | Urine | 84 |
Fluorescence | Sandwich format by using a fluorescent organic dye conjugated monoclonal antibody | Sandwich lmmuno assay | Increases | 3 ng mL−1 | 0–50 ng mL−1 | — | — | 85 |
Fluorescence | CysC selective aptamers conjugated to dye Alexafluor-647 | Lateral flow assay | Increases | 0.013 μg mL−1 | 0–120 μg mL−1 | — | — | 86 |
Fluorescence | Sandwich format by using a fluorescent organic dye conjugated monoclonal antibody | Lateral flow assay | Increases | 0.023 μg mL−1 | 0.023–32 μg mL−1 | BSA | Urine | 87 |
Fluorescence resonance energy transfer | Graphene oxide-fluorophore-labelled aptamer and DNase I | Resonance energy transfer | Increases | 0.16 ng mL−1 | 0.625–20 ng mL−1 | BSA, kidney injury molecule-1, neutrophil gelatinase-associated lipocalin | Urine | 88 |
Surface plasmon resonance | Nitrocellulose paper coated with AuNPs | CysC interaction with AuNPs | Red shifts in the LSPR wavelength | 0.1 μg mL−1 | 0.1 to 3 μg mL−1 | N-terminal pro B-type natriuretic peptide and BSA | Blood | 89 |
Surface plasmon resonance | Papain modified gold surface | Papain–CysC interaction | Red shifts in the LSPR wavelength | 0.09 μg mL−1 | 0–6 μg mL−1 | — | Plasma, urine and saliva | 90 |
Surface plasmon resonance | Bromelain or chymopapain or ficin modified gold surface | Bromelain or chymopapain or ficin–CysC interaction | Red shifts in the wavelength of the SPR | 0.1 μg mL−1 | 0–0.6 μg mL−1 | — | Plasma, urine and saliva | 91 |
Chemiluminescence | Functionalized mesoporous silica nanoparticles | Sandwich immunoassay | Luminescence increases | 0.0029 ng mL−1 | 0.0035–0.5 ng mL−1 | — | Urine | 92 |
Electrochemiluminescence resonance energy transfer | GCE modified with PTCA@Fe(III)-MIL-88B-NH2@Au/Ab1 | Resonance energy transfer | Luminescence decreases | 2.2 fg mL−1 | 0.01 pg mL−1 to 100 ng mL−1 | Prostate-specific antigen, carcinoembryonic antigen, bull serum albumin, carbohydrate antigen 125 | Serum | 93 |
Reflectometric interference spectroscopy | CysC coated glass plate | Binding inhibition assay with CysC antibodies | Optical thickness increases | — | 0.53–1.02 mg L−1 | — | Serum | 94 |
Chung et al. developed a technique named micro-urine nanoparticle detection (μUNPD) approach to detect soluble protein markers KIM-1 and CysC in urine by using miniaturized nuclear magnetic resonance device.95 To begin with, polystyrene microbeads coated with capture antibodies are mixed with CysC containing sample. Following incubation, all unattached molecular components were removed by a washing method. After that, the detecting antibody conjugate after combing with trans-cyclooctene (TCO) was added to the mixture. After another round of washing, amino modified Fe2O3 magnetic nanoparticles are added for labeling. The amino modified Fe2O3 magnetic nanoparticles were bound to TCO-modified detection antibody conjugate. Then the tagged beads were tested for T2 relaxation using micro NMR device. The schematics of the μUNPD assay for CysC detection is given in Fig. 9. This method ensures that the CysC was recognized and measured properly in the urine sample. The method was used to detect CysC in the concentration rage of 300–10000 ng mL−1 with a LOD of 20 ng mL−1. In another study, Wu et al. utilized non-contrast enhanced native T1 mapping of the renal cortex towards assessing renal fibrosis and found that a favorable correlation between T1 value and the concentration of CysC.96
![]() | ||
Fig. 9 The schematics of the μUNPD assay for CysC detection. Firstly, capture antibody conjugated microbeads, followed by TCO conjugated detection antibodies, and finally amino modified magnetic nanoparticles, are added for the CysC detection from the urine sample on the microchip. Reprinted with permission from ref. 95. Copyright: © 2015 Chung et al. |
The conventional methods used for the CysC monitoring are PETIA, PENIA and ELISA while the newly introduced methods include mainly electrochemical, optical and NMR based techniques. The research on the development of biosensor for CysC detection is not fully explored as only few articles are published compared to other biosensors developed for other important biomarkers. The major bio recognition methods used in the development of these sensors are either antibody based, aptamer based or cysteine proteases based recognition with CysC. Among these, aptamer-based recognition methods are mainly used in conventional methods. Hence, there are ample opportunities for developing aptamer-based sensors considering their strong binding affinity with the target. The electrochemical and optical biosensors for CysC detection mainly use the binding between CysC and their specific antibodies or binding between CysC and cysteine proteases (CysC–papain binding).
The LOD of the developed sensors are promising as the normal range of blood CysC is 0.8–1.2 mg L−1. However, the performance should be maintained for commercialized PoC sensors for CysC. Additionally, the sensor response time should be as minimum as possible. Even though LFIA-based sensors are providing the sensor response within 15 min, it should be maintained for the real sample analysis. The sensors based on binding CysC with antibody or cysteine proteases have the potential to upgrade to device development and easy commercialization as it does not require complex conjugation protocols or expensive readers.
To overcome the challenges especially in producing deliverable and reproducible response without any interferences, there is a need to modify the sensor development strategy for optimizing the immobilization process of antibodies or cysteine proteases. This process will help to enhance the storage and operational stability of the sensor devices. Additionally, there is need to avoid the pre-treatment of biological samples for fast screening and ease-of-use. Furthermore, the sensor devices should have the potential to be connected with a mobile phone, which helps in providing home-based diagnostics. These developments are expected to revolutionize and commercialize the sensor devices for CysC diagnostics ideal for practical POC applications.