Open Access Article
Mark A. Brzezinski
*ab,
Stephen F. Rablenc,
Janice L. Jones
a,
Ivia Clossetad and
Julien T. Middletona
aMarine Science Institute, University of California Santa Barbara, Santa Barbara, CA, USA
bDepartment of Ecology Evolution and Marine Biology, University of California Santa Barbara, Santa Barbara, CA, USA
cNu Instruments Ltd, PO Box 36, 2 New Star Road, Leicester, LE4 9JQ, UK
dFinnish Meteorological Institute, Helsinki, Uusimaa, Finland
First published on 20th October 2025
The performance of a new inlet system, the Nu Sil, for the automated determination of the isotopic abundances of silicon in SiF4 gas generated from the thermal decomposition of BaSiF6 is reported. The inlet system is coupled to an isotope ratio mass spectrometer through a conventional dual inlet system. The method uses straightforward and proven sample preparation chemistry that is suitable for converting the silicon in biogenic and lithogenic solids, or that dissolved in fresh or salt waters, to BaSiF6. Yields of silicon tetrafluoride are 99.8 ± 0.16%. δ30Si values obtained with the method for a variety of natural and synthetic materials agree with published values to better than 0.05‰. External long-term average δ30Si values of the solid standard NBS28, the secondary standards Big Batch and diatomite, and the reference seawater ALOHA1000 are all within 0.03‰ of their respective consensus values. Typical loading of BaSiF6 represents 10 to 40 μg of Si. The analytical rate is 30 unattended analyses daily.
The most common instruments presently employed in Si isotope measurement are multi-collector inductively coupled plasma mass spectrometers (MC-ICPMS).24 These instruments and associated preparatory chemistry can analyze small samples and have reasonable analytical throughput, but they require frequent, if not daily, tuning and they are sensitive to matrix effects, mass discrimination and temporal drift in isotope ratios. Less common are methods employing isotope ratio mass spectrometers (IRMS). For IRMS, the analyte gas, SiF4, is prepared either by high temperature fluorination by the reaction of Si compounds with F2 (ref. 10 and 25) or BrF5,26 through the acid decomposition of Cs2SiF6 (ref. 27) or through the thermal decomposition of BaSiF6.28,29 Fluorination with F2 or BrF5 has the advantage of allowing the simultaneous determination of silicon and oxygen isotope values in silicates.10,30 IRMS methods are largely free of matrix effects and mass discrimination is less than in MC-ICPMS, but they often employ hazardous chemicals and typically require larger samples than do those employing MC-ICPMS. They have historically required custom-built vacuum lines (fluorination) or customized inlet systems (acid decomposition), hampering adoption.
Here we evaluate a new commercial inlet system, the Nu Sil, manufactured by Nu Instruments Ltd, that automates the analysis of Si isotopes using IRMS with SiF4 generated through the thermal decomposition of BaSiF6 according to:
| BaSiF6 (s) → SiF4 (g) + BaF2 (s) | (1) |
The SiF4 is ionized via electron bombardment producing SiF3+ ions as the main fragment. As fluorine consists of a single isotope of mass nineteen, 28SiF3+, 29SiF3+ and 30SiF3+ produce ion beams at m/z 85, 86 and 87, respectively. The thermal decomposition of BaSiF6 was among the first methods employed to measure stable isotopes of Si in natural materials.31,32 More recently, it has been employed to make highly accurate absolute Si isotope abundance measurements in support of refining the Avogadro constant.28,33,34 The requisite sample preparation chemistry and the thermal decomposition of the compound are both known to not fractionate isotopes of Si;28 however, BaSiF6 methods have been criticized for being slow and for requiring large sample mass.35 The new instrumentation and sample preparation methods presented here are straightforward to implement, have relatively high sample throughput (30 unattended samples per day) and are capable of analyzing small samples (≥10 μg Si as BaSiF6) with accuracy and precision comparable to MC-ICPMS.
![]() | ||
| Fig. 1 Schematic showing a side view of the Nu Sil sample heater and valve block assembly shown as a photograph in Fig. 2. In A the sample carousel (A) rotates each sample vial (B) into position beneath the tube furnace (C) using the carousel optical alignment device (D). In B the lifting mechanism (E) raises the sample vial through the tube furnace (C) to position the sample within the furnace pressing the vial opening against the BTO seal on sample port (F). Evacuation is accomplished through valve block G that is connected to the CF PREP cold finger (not shown) through stainless tube H. A sheet metal heat shield (I) protects the valve block from radiant heat. A threaded plug (J) seals off a port formerly used for acid introduction the Nu Carb and can be opened to clean the internals of the sample port. | ||
![]() | ||
| Fig. 2 Photograph of the inside of the Nu Sil sample chamber showing components illustrated in Fig. 1: the sample carousel with samples vials (A & B), the Nu Sil heater (C), the heat shield (I), vacuum valves (G), cleaning port (J) and outlet to the CF PREP cold finger (H). Other components shown in Fig. 1 are obscured in the view shown. | ||
The furnace is 1.5 cm long and is resistively heated with temperature feedback via a thermocouple inserted into the body of the furnace tube. A control circuit achieves temperatures from 100 to 600 °C to within 3 °C. The furnace is mounted so that the open central bore is aligned with the vertical axis of the shaft that lifts glass sample vials up to the sample inlet port (Fig. 1A). When a sample vial is raised through the inner bore of the tube furnace, the flat vial lip is pressed against a BTO seal and the lower half of the vial aligns with the center of the tube furnace with the upper 1.5 cm of the vial protruding above the top of the furnace (Fig. 1B). This configuration allows the BaSiF6 that is located in the conical depression in the bottom of the vial to be heated to the reaction temperature of 590 °C without the lip of the vial exceeding the thermal tolerance of the BTO seal (∼400 °C). A lightweight sheet-metal heat shield protects the vacuum valves and the vial detection system that are positioned above the sample port from radiant heat (Fig. 1 and 2). Because sample vials are pushed entirely out of the carousel into the furnace tube, the sample carousel is modified from the Nu Carb design, being taller with chamfered vial receptacles to allow vials to relocate easily after sampling. All other hardware is unmodified from that on the Nu Carb.
The Nu Sil is connected to an unmodified Nu perspective IRMS equipped with a dual inlet. The preamplifiers on the Faraday collectors are equipped with gain resistors of 1 GΩ, 100 GΩ and 100 GΩ for measurement at m/z 85, 86 and 87, respectively. The control software is customized to operate the Nu Sil and to execute a custom control sequence for valve and cold finger operation on the dual inlet and the Nu perspective IRMS as described in section 2.3.
:
Na2O. The TEA-moly procedure is not necessary for higher purity solid samples or solid standards.
Solid SiO2 from the TEA-moly procedure or from naturally occurring solids is converted to BaSiF6 as follows: up to 2.4 mg of SiO2 is placed in a 5 mL polypropylene or polyethylene snap-cap centrifuge tube and 1 mL of 7.5 M HF is added to dissolve the sample and produce SiF6− ions. (The final concentration of dissolved Si must be <40 mM to prevent Si volatilization25). After allowing 24 to 48 h for sample dissolution (24 for amorphous silica, 48 h for crystalline forms), 0.5 mL of 3 M CsCl is added to precipitate Cs2SiF6. After 12 hours, the precipitate is isolated by centrifugation and rinsed 3 times with ethanol. The final product is dried in the centrifuge tube at 65 °C. Conversion of Cs2SiF6 to BaSiF6 takes advantage of the greater solubility of Cs2SiF6 (0.60 g per 100 mL H20) compared to BaSiF6 (0.026 g 100 mL H20) whereby 2.4 mL of an aqueous solution that is 17 mM in BaCl and 2 mM in HF is added to the dry Cs2SiF6 to dissolve the Cs2SiF6 and precipitate BaSiF6. After 24 h the BaSiF6 is isolated by centrifugation, rinsed with alcohol and dried as described for Cs2SiF6. Samples are stored sealed in the centrifuge tubes at room temperature and are stable indefinitely.
![]() | ||
| Fig. 3 The Nu Sil and Nu perspective gas handling systems. Valves are labeled following the labeling scheme in the Nu Sil software. | ||
The initial filling of an empty reference bellows from a full reference gas tank proceeds as follows: the refill capillary is pressurized behind valve GB (closed) by opening all 3 valves on the reference gas canister and waiting ∼10 minutes for isotopic equilibrium. The lower 2 valves on the refill tank are closed so that SiF4 is constrained to the volume of the refill capillary and the 2 cm piece of 6 mm OD tubing located between the last valve on the refill cylinder and the capillary. This reduced volume is necessary to slow the filling of the bellows so that the following manual fill sequence is manageable: first, the changeover valve is set to allow reference gas into the ionization source of the mass spectrometer. Then the reference bellows is expanded to 100%. With valves SP, GB and RM closed, valves RI, RP, RV and LV are opened to evacuate the reference bellows and the tubing between RM and GB under low vacuum for 3 minutes. Then LV is closed and HV opened to evacuate under high vacuum for another 3 minutes. Valves RF, RV and HV are then closed and LV opened. Then, while still actively pumping under low vacuum, valve GB is opened for 5 s to establish flow from the reference gas capillary after which valve RP is closed. Valves RF, RV and RM are opened and the signal intensity at m/z 85 is monitored. When the signal reaches ∼1.9 × 10−9 amperes (∼20% of signal saturation) valve GB is closed and the signal intensity at m/z 85 is allowed to stabilize for 3 minutes after which a reading of ∼2.2 × 10−9 amperes at m/z 85 is typically reached. On subsequent refills of the bellows it is not necessary to evacuate the remaining SiF4 from the bellows and valve RV can be closed during the initial evacuation of space between valves RF, RV and RI.
Prior to isotopic analysis, the SiF4 generation subsystem is prepared for the next sample. By this time the tube furnace has cooled to <150 °C and valve CO is closed and the sample vial is detached from the sample port by introducing high purity (99.998%) nitrogen gas by opening valves CA and CV to relieve vacuum. The sample vial lowered back into the sample carousel and valves CA and CV are then closed. The next sample vial is rotated into position and attached to the sample port, brought under vacuum and the prebake at 250 °C initiated as above. The PREP cold finger is heated to 100 °C under high vacuum with valves WP, WI, PF and CO open. After 30 s at 100 °C the cold finger is allowed to passively cool and evacuate during the rest of the sample prebake at 250 °C.
Analysis of the sample begins with both the reference and sample gas streams directed to waste in the changeover valve. The changeover valve is then set to introduce reference gas into the ionization source, and a peak-centering routine is performed at m/z 85 to determine the magnet setting for measuring beam intensity at the center of the peaks. Then both the sample and reference gas streams are directed back to waste and the signal at m/z 85 is monitored every 7 s until its value has fallen by no more than 1 × 10−14 amperes between readings. The background ion beam intensities at m/z 85, 86 and 87 are then each measured for 8 s and the means of these readings are subsequently subtracted from all sample readings at each respective m/z. The reference-gas bellows is adjusted so that the reference gas ion beam intensity at m/z 85 matches that of the sample after which the bellows and sample beam intensities are typically within 1% or better of one another. Then the reference-valve configuration is set to match that of the sample's with valve RM open and RF is closed for samples that did not require expansion or chopping or with valves RM and RF open (valves RV, RI and RP are closed) for samples that were expanded or chopped. Sample and reference gas are now at the same pressure, confined to identical volumes and continuously pumped either to the ion source of the Nu perspective or to waste through matched crimped capillaries so that the sample and reference gas deplete at the same rate during sample analysis. Sample measurement proceeds using 8 sample/reference cycles of the changeover valve with each reading of the sample or reference gas lasting 20 s at 10 Hz.
When the analysis is complete the remaining sample gas is pumped to waste by opening valves SF and SP (PH is open). The INLET cold finger is heated to 100 °C for 30 s and the sample side of the dual inlet is pumped for 6 minutes at high vacuum. Total sample analysis time is 65–70 minutes depending on the time required for sample chopping and balancing sample and reference gas pressures. As preparation of the next sample begins before the previous sample is fully analyzed, the effective analytical throughout after the first sample averages 48 minutes to obtain the raw uncalibrated delta values for a sample.
| δ30Si = slope × δ30Simeasured + intercept | (2) |
As part of QC/QA procedures, potential interfering masses can be monitored for each analysis by measuring the signal at m/z 83 (one of the larger peaks given off by the BTO seal) with acceptable readings indicated by a peak height that does not exceed background. Deviations in the δ29Si
:
δ30Si ratio also proved to be a sensitive indicator of the presence of interfering masses. The interfering signals biased the δ29Si
:
δ30Si ratio to be more negative than the theoretical value of 0.5110 (ref. 27, 38 and 40) for samples with positive δ30Si values and to be more positive than expectation for samples with negative δ30Si values. This issue is essentially eliminated with BTO seals where the average δ29Si
:
δ30Si ratio for Big Batch with its negative δ30Si value is +0.5112 ± 0.0012 (1σSD) while diatomite with its positive δ30Si value has an average ratio of +0.5085 ± 0.0023 (1σSD) (Table 1).
| Standard | 29δSi (‰) | 95% CI | 1σSD | 30δSi (‰) | 95% CI | 1σSD | δ29Si : δ30Sib |
95% CI | 1σSD | N | Consensus values, 30δSi (‰)a |
|---|---|---|---|---|---|---|---|---|---|---|---|
a Published consensus values for diatomite,38 Big Batch38 and Aloha1000.24b δ29Si : δ30Si calculated from data normalized to the reference gas before calibration (eqn (2)). |
|||||||||||
| NBS | +0.010 | 0.004 | 0.020 | +0.026 | 0.006 | 0.032 | 0.0507 | 0.0003 | 0.0020 | 119 | 0.000 |
| Diatomite | +0.629 | 0.004 | 0.020 | +1.236 | 0.005 | 0.030 | 0.5085 | 0.0004 | 0.0022 | 117 | +1.260 |
| Big batch | −5.334 | 0.004 | 0.030 | −10.450 | 0.007 | 0.057 | 0.5112 | 0.0002 | 0.0013 | 226 | −10.480 |
| ALOHA1000 | +0.647 | 0.012 | 0.028 | +1.274 | 0.012 | 0.029 | 0.5086 | 0.0005 | 0.0011 | 22 | +1.240 |
| Multipoint calibration of δ30Si with big batch, NBS and diatomite | ||||
|---|---|---|---|---|
| Mean | 95% CI | 1σSD | N | |
| Slope | 0.9822 | 0.0018 | 0.0040 | 20 |
| Intercept | −3.192 | 0.035 | 0.008 | 20 |
| R2 | 0.99998 | 0.00001 | 0.00003 | 20 |
Multiple precautions ensure the final BaSiF6 product is free of both water and HF as water vapor decomposes the analyte gas, SiF4, and, in a moist environment, HF can react with the borosilicate glass sample vials producing contaminant SiF4 and H2O. First, samples are preheated to 110 °C prior to being loaded into the Nu Sil. Then the sample chamber of the Nu Sil is kept at 70 °C to maintain sample dryness, and finally each sample is heated to 250 °C for 20 min under vacuum after being attached to the sample port of the Nu Sil. Although the initial heating to 110 °C should effectively remove HF and water, the additional precautions ensure that the samples remain dry prior to analysis. The Nu Sil carousel holds 50 sample vials, which at our analytical rate of 48 minutes per analysis, requires 40 hours to analyze an entire carousel. Maintaining the Nu Sil sample chamber at 70 °C minimizes absorption of water by samples during this time. Finally, any water that is absorbed while vials are in the Nu Sil sample chamber is eliminated by the prebaking of each sample at 250 °C under vacuum before analysis. The pre-bake does not add to the analysis time (apart from the first sample) as it starts at nearly the same time as does the isotopic analysis of the previous sample and the time required to analyze a sample and then prepare the dual inlet for the next is >20 minutes. The prebake procedure does reduce the time required to achieve the 590 °C sample reaction temperature.
![]() | ||
| Fig. 5 Yield of SiF4 from the thermal decomposition of BaSiF6 as a function of thermocouple temperature. Error bars are standard deviations of triplicate measurements at each temperature. | ||
| δ30Si = (δ30Simeasured + δ30Sioffset) + (δ30Simeasured × δ30Sioffset)/1000 | (3) |
The approach used wherein differs in that we use a regression approach to calculate δ30Si from the relationship between δ30Simeasured for a series of standards regressed against their consensus values. The regression approach produces a smaller normalization error than the single-point normalization to NBS28 as it is more robust against random error in the analysis of any one standard.39 A random selection of 20 calibration curves generated on the Nu Sil gave a mean slope of 0.9822 ± 0.0018 (95% CI) with an average intercept of −3.192 ± 0.035 (95% CI) and a mean R2 of 0.99998 ± 0.00001 (95% CI) (Table 1). The correlation coefficient of the calibration curve is very close to unity because standards are well calibrated and the systematic error introduced during mass spectrometric analyses is linear within a limited dynamic range.42 The deviation of the slope from unity is related to the mass bias within the mass spectrometer with the non-zero intercept reflecting the δ30Si value of the reference gas.42
As most δ30Si values published to date were produced using a single point normalization to NBS28, it is important to evaluate whether that single point calibration and the multipoint calibration recommended here produce significantly different results. In the multipoint regression approach, the value obtained for NBS28 is dictated by the regression rather than being assumed to be zero as done in the single-standard approach. The regression approach for the same 20 calibrations as above predicts a value of NBS28 of +0.025 ± 0.033‰ (1σSD) (Table 1) compared to the assumed value of zero, with the difference being within the analytical precision of current δ30Si measurements.24,38,43
The Nu perspective is capable of analyzing samples that are at least two orders of magnitude smaller than recommended here; however, at least two factors make this impractical. First is the difficulty of consistently loading a very small mass of BaSiF6 into a sample vial. Consistency is required so that the reference gas balancing procedure is successful for all samples in a sample carousel as there is a limit to the range of sample sizes that can be balanced at a given gas pressure in the reference bellows. Moreover, SiF4 has a strong tendency to adsorb onto metal surfaces where it remains even after prolonged pumping under high vacuum. The adsorbed gas does exchange when additional SiF4 gas is introduced, creating a memory effect for sequential samples. The effect is inconsequential for sample sizes of 25–50 μg Si but would be more pronounced with very small samples. This limitation might be overcome by heating the entire analytical train while analyzing samples, but the challenge of consistently loading extremely small samples remains.
![]() | ||
| Fig. 6 Whisker plot of the δ30Si value of diatomite as a function of sample gas analytical volume. Microvolume refers to the volume between valves SF and SM in Fig. 2. Expansion is the volume between valves SI and SM. Chopped refers to discarding the gas between valves Si and SF after expansion as described in the text. Long dashed line is consensus value of +1.26‰ for diatomite with the two long dotted lines reflecting the reported uncertainty of ± 0.10‰ (1σSD, Table 1).38 For the whisker plots the top and bottom of the shaded box indicate the 25th percentile, the solid line within the box marks the median, the dotted line marks the mean and the boundary of the box farthest from zero indicates the 75th percentile. Error bars above and below the box indicate the 90th and 10th percentiles. Symbols are outliers. | ||
While chopping with an appropriate evacuation time produces quality data, avoiding chopping speeds analysis. Sufficient consistency in mass across samples is achieved with the Wiretrol II® samplers described above with the number of times the sampler is inserted into the BaSiF6 powder serving as a measure of sample packing. It is not necessary to weight each sample precisely. Sample mass can be routinely calculated from the transducer pressure reading for a sample using the ideal gas law and the volume between valves PF and SI of 5.53 mL (factory calibration). In practiced hands the risk of loading a sample too large for the microvolume is extremely rare and typically <2% of samples require expansion or chopping. Unintentional loading of samples that are too small is also extremely rare when using this sample loading method.
The long-term reproducibility for measurements on secondary standards and reference materials made over several months are given in Table 1. Mean δ30Si values for the standards NBS28, Big Batch and diatomite obtained by the thermal decomposition of BaSiF6 are all within 0.03‰ or better relative to their consensus values38 and the average δ30Si of the reference seawater ALOHA1000 (ref. 24) is within 0.034‰ (Table 1) indicating excellent performance of the Nu Sil and Nu perspective. External reproducibility of each of the standards is ≤0.02‰ (95% CI) comparable to the precision obtained by MC-ICPMS.24,38
The secondary standard diatomite was used in a more detailed assessment of performance. The consensus value for diatomite, 1.26 ± 0.20 (2σSD), is derived from measurements from six laboratories whose results were reported in Reynolds et al. (2007).38 Between 2007 and January 2024, 128 additional studies have reported additional statistics for diatomite, 124 of which were performed on MC-ICP mass spectrometers (Table S8). The mean δ30Si values from each of those studies and the original data from Reynolds et al. 2007 (ref. 38) are plotted as a histogram along with the corresponding probability density function (PDF) in Fig. 7a. The PDF is unimodal with a strong peak at +1.238‰ representing the mode of the PDF. The overall mean of the measurements is +1.248 ± 0.109‰ (2σSD, n = 137). Those values differ from the mean for diatomite from the Nu Sil (+1.236 ± 0.060‰, 2σSD, n = 117) by 0.002‰ for the PDF peak and by 0.012‰ for the overall mean of measurements. A similar analysis for measurement uncertainty (2σSD) reported in these same studies reveals a PDF that is also unimodal but skewed to the right (Fig. 7b) with a mode of 0.095‰. The mean 2σSD across studies is 0.14‰ compared to the uncertainty in the current consensus value for diatomite of 0.20‰ (Fig. 7b). The average 2σSD obtained using the Nu Sil (0.060‰, Table S2) is lower than that for all but seven studies (5.1%) with the PDF predicting that 10.7% of studies would have a lower value for 2σSD. The reported level of precision for measurements on the Nu Sil can be routinely obtained with quadruplicate analyses of samples with each replicate analyzed within a separate carousel of samples.
![]() | ||
| Fig. 7 (A) Histogram of mean δ30Si values for diatomite reported in the literature including and following38 through January 2025. Red bars are the original data from38 used to define the current consensus value for this material. Black line is the probability density function. Vertical dashed lines are the average from38 (dark red) and from the Nu Sil (blue). (B) As in A but for the uncertainty in the mean isotope value for diatomite (2σSD) reported in each study. | ||
Several types of previously analyzed materials were obtained and analyzed for their silicon isotopic composition to illustrate the versatility of the sample preparation chemistry and that capabilities of the Nu Sil. The materials range from dissolved Si in rivers and oceans, soils, clays and sponge spicules representing a wide range of sample composition and matrices. The method is designed to systematically eliminate potential matrix interferences by converting all samples to SiO2 prior to their conversion to BaSiF6. The results show excellent agreement with published values for all of these materials (Table 2), indicating that the method is indeed insensitive to sample matrix effects. With the exception of the one river water sample measurements from the Nu Sil agree with previously published results to within 0.08‰ or better (range 0‰ to 0.08‰, mean absolute difference 0.05 ± 0.06‰, n = 11) irrespective of material type. We note that the results in Table 1 for the standards NBS28 (RM8546, a quartz), Big Batch (cristobalite) and diatomite (calcined diatomite) add to the range of material types successfully analyzed.
| Sample type | Station or sample ID | Silicic acid (μM Si) | 30δSi (‰) MC-ICP-MS | 1σSD | N | δ30Si (‰) Nu-Sil | 1σSD | N |
|---|---|---|---|---|---|---|---|---|
| a Seawater samples from the Arctic Ocean. Numbers are sample number form the GEOTRACES cruise GN01.45b NuPlasma 1700 at IGMR ETH Zürich, Switzerland.c Analyses at Bristol, UK via Thermo-Fisher Neptune MC-ICPMS by Kate Hendry and Jade Hatton.d Source: Texas, USA.e Source: Natural Pigments, LLC, USA.f Original measurements reported in Delvigne et al. (2021).46g Mean from 11 laboratories with 3–10 independent replicates each.24h Further details published in Hendry et al. (2019)47 data available at https://doi.pangea.de/10.1594/PANGEA.898676. | ||||||||
| River water44 | Mungo | 324 | +0.91 | NA | 1 | +1.1 | NA | 1 |
| Seawater silicic acida | GEOTRACES 11667 | 39.9 | +1.66b | 0.01 | 2 | +1.66 | 0.01 | 3 |
| GEOTRACES 11915 | 29.7 | +1.71b | 0.06 | 2 | +1.71 | 0.09 | 3 | |
| ALOHA1000 | 112.8 | +1.24g | 0.10 | 11 | +1.29 | 0.01 | 44 | |
| Soilsf | GBW-07401 | 62.60 | −0.27 | 0.03 | 78 | −0.22 | ||
| GBW-07404 | 50.95 | −0.76 | 0.06 | 79 | −0.84 | |||
| GBW-07407 | 32.67 | −1.82 | 0.09 | 72 | −1.62 | |||
| Clay minerals | Wgt% SiO2 | |||||||
| Kaolinited | 46.55 | −2.16c | 0.08 | 15 | −2.13 | 0.06 | 7 | |
| Calcined Kaolinitee | 46.55 | −1.44c | 0.09 | 16 | −1.43 | 0.06 | 6 | |
| Marine sponge spiculesh | Vazella pourtalesi | Depth (m) | ||||||
| B0027 | 199 | −1.61 | 0.07 | nr | −1.65 | 0.08 | 12 | |
| B0044 | 199 | −1.73 | 0.06 | nr | −1.75 | 0.05 | 12 | |
| B0227 | 206 | −1.75 | 0.06 | nr | −1.71 | 0.06 | 12 | |
![]() | ||
| Fig. 8 δ30Si value of diatomite as a function of sample size expressed as micrograms of Si for a typical source tuning. Horizontal dashed line is the consensus δ30Si value for diatomite of +1.26‰ with the dotted lines representing the reported uncertainty of the consensus value, ± 0.10‰ (1σSD, Table 1).38 Horizontal solid line is the mean for the samples. | ||
| This journal is © The Royal Society of Chemistry 2025 |