Elias
Wölfer
*,
Christoph
Burkhardt
and
Thorsten
Kleine
Max Planck Institute for Solar System Research, Justus-von-Liebig-Weg 3, 37077 Göttingen, Germany. E-mail: woelfer@mps.mpg.de
First published on 27th February 2025
We present analytical procedures for the measurement of mass-dependent Ge isotope compositions using a 70Ge–73Ge double spike and multiple-collector inductively coupled plasma mass spectrometry (MC-ICP-MS). Two different mass spectrometers (ThermoScientific Neptune Plus and Neoma) and two different sample introduction systems (Teledyne Cetac Technologies HGX-200 hydride generator and, for the first time in Ge isotope analyses, a Cetac Technologies Aridus II desolvator) were used. A series of analytical tests demonstrate that our analytical procedure efficiently separates Ge from the sample matrix and provides accurate and precise Ge concentration and isotope data for both instruments and sample introduction methods. The external reproducibility (2 s.d.) of the entire analytical procedure is ±0.09‰ for δ74/70Ge (the permil deviation from the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 3120a). Using the new methods, we obtained Ge isotope data for an Alfa Aesar Ge standard solution, the NIST SRM metals 129c and 261, two terrestrial basalts (BHVO-2 and BCR-2), and eight iron meteorites. We find excellent agreement between the data obtained using the hydride generator and using the Aridus II desolvator. Due to the overall easier use of the latter, the desolvator may be the sample introduction system of choice for most Ge isotope applications. Overall, the samples of this study show δ74/70Ge variations up to ∼2‰, and in agreement with literature data, indicate that mass-dependent Ge isotope variations may be used to study a wide range of geochemical and cosmochemical processes.
In terrestrial settings, mass-dependent Ge isotope variations have been investigated in a number of studies.9,10,15–21 These studies have revealed that the silicate rocks of the Earth's crust and mantle are depleted in Ge as a result of core formation and show rather homogeneous δ74/70Ge values (i.e., the permil deviation of the 74Ge/70Ge ratio from a standard), indicating there is only limited Ge isotope fractionation during melting and fractional crystallization processes within the silicate Earth. By contrast, Ge is highly enriched and isotopically fractionated in sphalerite and other sulfide deposits, most likely as a result of kinetic isotope fractionation during sulfide formation.
In the field of cosmochemistry, Ge isotopes have not yet attracted major interest, despite their broad potential range of use. With the exception of ordinary chondrites,22 the unusual CB chondrites,23 and some iron meteorites groups,16,17,24 no mass-dependent Ge isotope data have been published. Nevertheless, the available data alone already demonstrate that Ge isotopes are fractionated in meteorites by more than 2‰, highlighting the potential of Ge stable isotopes for examining processes associated with the formation and subsequent evolution of meteorites and their parent bodies. The scarcity of Ge isotope data for meteorites at least in part reflects analytical difficulties associated with Ge isotope measurements by multicollector inductively coupled plasma mass spectrometry (MC-ICPMS). For instance, earlier studies on meteorites used a cyclonic spray chamber as the sample introduction system, but this method requires relatively large amounts of Ge, thereby limiting the samples that can be analyzed to some Ge-rich iron meteorites.16,17 More recent studies overcame this problem by using a hydride generator for the sample introduction, which improves sensitivity and thus allows for measurements of Ge-poorer samples.21–23 However, until now only a small number of meteorite samples have been analyzed using this technique.
To facilitate Ge isotope measurements for a comprehensive set of extraterrestrial samples irrespective of their Ge concentration and modal mineralogy (i.e., silicate- and metal-dominated samples), we developed analytical procedures for precise Ge isotope measurements by 70Ge–73Ge double spike MC-ICPMS, using either a Cetac Technologies Aridus II desolvator or a Teledyne Cetac Technologies HGX-200 Hydride Generator as the sample introduction systems. Compared to prior studies, our analytical procedures include several important novel aspects. First and foremost, a desolvator as a sample introduction system has so far not been used for Ge isotope measurements, but we will show below that this may be the method of choice for most Ge isotope applications. Second, the application of a Ge double spike has yet mostly been limited to geothermal fluids and (sea)water samples.21,25–27 With the exception of a single study on terrestrial rocks18 and a single study using thermal ionization mass spectrometry,28 no Ge double spike has been used for precise Ge isotope measurements in planetary sciences. Third, direct comparison of Ge isotope data obtained using different introduction systems is very limited. Finally, our methods, which are tailored towards the analyses of meteorites, will allow precise Ge isotope measurements of strongly Ge-depleted iron meteorites, which have not been possible using previously used techniques.16,17 Together, the new analytical procedures presented here will, therefore, help to unlock the full potential of Ge isotopes as a tracer in geochemistry and cosmochemistry.
The reduction of double spike data relies on the assumption that any isotopic difference between the sample and the standard is purely mass-dependent.29,32 However, especially among extraterrestrial samples, mass-independent isotope anomalies are widespread (such as e.g. nucleosynthetic anomalies), and the presence of such anomalies might lead to spurious results, if not corrected for.33,34 However, until now no such anomalies have been reported for Ge. This is consistent with our finding that all meteorite samples analyzed in this study using a Ge double spike show indistinguishable mass-dependent Ge isotope compositions compared to previous measurements on the same samples obtained without the use of a Ge double spike (see Fig. 5 and 8 and Section 4.4 for details). Therefore, mass-independent Ge isotope anomalies either do not exist or, if present, are small and as such have no measurable effect on the mass-dependent Ge isotope data reported here.
Traditional double spike measurements use four different isotopes, two of which are the spike isotopes, which together with the other two unspiked isotopes are used in the double spike inversion to determine the natural mass fractionation factor α.29,35 Germanium has five naturally occurring stable isotopes (70Ge, 72Ge, 73Ge, 74Ge, and 76Ge), and there are several options for the setup of a Ge double spike. However, 76Ge should be avoided because of Ar-based polyatomic interferences (e.g., 36Ar40Ar and 36S40Ar) at mass 76, which naturally occur in MC-ICPMS measurements using an Ar plasma. Similar to prior studies using a Ge double spike,18,21,25–27 of the remaining four Ge isotopes, we selected the two less-abundant isotopes 70Ge (20.57%) and 73Ge (7.75%) for the double spike, which was optimized using 70Ge, 72Ge, 73Ge, and 74Ge in the inversion. The predicted intrinsic error on α of the 70Ge–73Ge double spike setup is small [∼39 ppm per amu; 1 s.d.],29 whereas the range of permissible double spike–sample ratios is relatively large (Fig. 1 and 2). All other possible double spike designs using 70Ge, 72Ge, 73Ge, and 74Ge would result in both larger errors on α and a smaller range of permissible double spike–sample ratios.
![]() | ||
Fig. 1 Contour plot of the error propagation on the natural fractionation coefficient α (1 s.d.) in the 70Ge, 72Ge, 73Ge, and 74Ge inversion depending on the proportion of double spike in the double spike–sample mixture and the proportion of 70Ge spike in the 70Ge–73Ge double spike as calculated using the Double Spike Toolbox.29 The calculated optimum is marked with a black square (i.e., 73.18% 70Ge spike in the 70Ge–73Ge double spike; 52.15% double spike in the double spike–sample mixture). The contour lines represent a 1% increase in the error on α relative to the minimum error. |
![]() | ||
Fig. 2 Propagation of the theoretical uncertainties (error) on the natural fractionation coefficient α as a function of the proportion of Ge double spike in the spike–sample mixture calculated using the Double Spike Toolbox and using 70Ge, 72Ge, 73Ge, and 74Ge for the inversion.29 |
The 70Ge and 73Ge single spikes (batch no. 198640 and 198841, respectively) were obtained as ∼10 mg fine grained metal powder from the Oak Ridge National Laboratory. The powders were weighed into 15 ml Savillex Teflon perfluoroalkoxy (PFA) vials and digested via table-top digestion using 10 ml 8 M HNO3–0.1 M HF at 120 °C for 10 days. After complete digestion, the solutions were transferred into larger Savillex PFA vials and diluted with MQ. Small aliquots of the two single spikes were then used for determining their Ge concentrations and isotopic compositions by MC-ICPMS. The two single spikes were then combined into the final 70Ge–73Ge double spike, where the mixing proportions were optimized to minimize the error propagation on α (73.18% 70Ge spike and 26.82% 73Ge spike, Fig. 1). The double spike was dried down, re-dissolved in 0.78 M HNO3–0.01 M HF, and equilibrated for 48 h at 120 °C on a hotplate. Finally, small aliquots of the double spike were used to determine its isotopic composition and concentration by MC-ICPMS (see below).
The National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 3120a Ge solution standard (Lot no. 080429; 10.015 ± 0.021 mg per g Ge) is the recommended standard reference solution for stable Ge isotope measurements,17,18 and was used as the reference standard throughout this study. We routinely also measured an Alfa Aesar Specpure 1000 μg per ml Ge plasma standard solution in 5% HNO3/traces of HF (Lot no. 12-12328C) along with each sample set to monitor the accuracy and reproducibility of our analytical method.
To investigate the magnitude and direction of Ge stable isotope variations in extraterrestrial samples, we analyzed three variably Ge-depleted magmatic iron meteorites from both the non-carbonaceous (NC) and carbonaceous (CC) meteorite reservoirs.36,37 These include the CC-type iron meteorites Kumerina (group IIC; ∼100 μg per g Ge; Ge/NiCI ∼0.3) and Needles (group IID; ∼100 μg per g Ge; Ge/NiCI ∼0.3), as well as the NC-type iron meteorite Henbury (group IIIAB; ∼35 μg per g Ge; Ge/NiCI ∼0.15). The variable degrees of volatile element depletion (e.g., Ge-depletion) and/or the distinct provenances among these iron meteorites could potentially have affected the Ge stable isotope composition of the samples, making them promising candidates for evaluating natural Ge isotope variations.
In addition to the magmatic iron meteorites, we also analyzed five non-magmatic iron meteorites. Of these, two belong to the IIE irons (NC-type) and three belong to the IAB irons (NC-type). The origin of non-magmatic iron meteorites is often linked to small(er)-scale impact events on planetary surfaces, rather than to the global melting events that characterize the formation of the magmatic irons.38,39 Evaporation and (re-)condensation processes during impact-heating related to these local incidents, therefore, might have led to Ge stable isotope variations among these iron meteorite samples.
Finally, some of the samples mentioned above have already been investigated in prior studies.16,17 This allows for a direct comparison of the Ge concentrations and stable isotope compositions obtained here to the results of previous studies.
The chemical separation of Ge from metal samples via ion exchange chemistry followed previously established analytical protocols.16,17,22,23 The metal samples were loaded in 0.5 M HNO3 onto Bio-Rad columns filled with 2 ml of pre-cleaned and conditioned Bio-Rad AG 50W-X8 cation exchange resin (200–400 mesh, hydrogen form). While Ge was directly eluted with an additional 8 ml 0.5 M HNO3, the matrix elements remained on the resin. Small aliquots (1%) were removed from the Ge cuts and cross-checked for purity using a Thermo Scientific iCAP TQ quadrupole inductively coupled plasma mass spectrometer (ICP-MS) at the Max Planck Institute for Solar System Research (MPS) in Göttingen. The samples were introduced into the mass spectrometer in a 0.5 M HNO3 solution, and the Ge concentrations were measured in SQ mode, relative to an in-house standard solution. If sufficiently clean for Ge isotope measurements (i.e., Zn/Ge < 0.01, see section 4.3), the Ge cuts were dried down at 120 °C, treated with cHNO3 to destroy organic compounds from the cation resin, and re-dissolved in 2 ml 0.5 M HNO3. Otherwise, the ion exchange chemistry was repeated once.
The chemical separation of Ge from the silicate matrices via a two-stage ion exchange chemistry was adapted from previously established protocols16,17,22,23 and calibrated using the rhyolite JR-1 obtained from the Geological Survey of Japan (Fig. 3). In the first stage, the samples were loaded in 1 ml 1 M HF onto Bio-Rad columns filled with 2 ml of pre-cleaned and conditioned Bio-Rad AG 1-X8 anion exchange resin (200–400 mesh, chloride form), and most matrix elements were directly washed off with an additional 13 ml 1 M HF, whereas Ge together with Ti and some Ca, Al, Mg, and V remained on the column (Fig. 3). After rinsing with 2 ml ultrapure water and 4 ml 0.2 M HNO3, Ge was eluted with 6 ml 0.2 M HNO3 together with some remaining Al and Ti. The Ge cuts were dried down at 120 °C and re-dissolved in 2 ml 0.5 M HNO3. The purification of Ge from Al, Ti, and other remaining matrix elements (especially Zn) was achieved with a second ion chromatography step, which uses Bio-Rad columns filled with 2 ml of pre-cleaned and conditioned Bio-Rad AG 50W-X8 cation exchange resin (200–400 mesh, hydrogen form). The samples were loaded in 2 ml 0.5 M HNO3 and Ge was directly eluted with an additional 8 ml 0.5 M HNO3, whereas Al, Ti, Zn, and other remaining matrix elements remained on the column. The final Ge cuts were dried down at 120 °C, treated with cHNO3 to destroy organic compounds from the cation resin, and then re-dissolved in 2 ml 0.5 M HNO3 for Ge isotope measurements.
The yields for each of the two ion exchange procedures were typically >90%. As the Ge double spike is added prior to sample digestion, these slightly incomplete yields are inconsequential for the Ge isotope measurements. Total procedural blanks as measured by isotope dilution are typically less than 1 ng and negligible for all samples, given that >1000 ng Ge was processed for each sample.
Instrument operating conditions | Desolvator (Neoma) | Hydride generator (Neoma) | Desolvator (Neptune) | Hydride generator (Neptune) |
---|---|---|---|---|
a All Faraday detectors are connected to 1011 Ω feedback resistors. b The both outer Faraday detectors (L4 and H4) are connected to 1012 Ω feedback resistors, while the seven inner Faraday detectors (L3 to H3 as well as the center cup) are connected to 1011 Ω feedback resistors. c Nebulizer gas flow rate. | ||||
RF plasma power | 1400 W | 1400 W | 1200 W | 1200 W |
Plasma gas flow rate | 15 l min−1 | 15 l min−1 | 16 l min−1 | 16 l min−1 |
Sampler cone | Standard nickel | Standard nickel | Standard nickel | Standard nickel |
Skimmer cone | X nickel | X nickel | X nickel | X nickel |
Acceleration voltage | ∼10![]() |
∼10![]() |
∼10![]() |
∼10![]() |
Instrumental resolution | Low resolution | Low resolution | Low resolution | Low resolution |
Mass analyzer pressure | ∼2 × 10−9 mbar | ∼2 × 10−9 mbar | ∼4 × 10−8 mbar | ∼4 × 10−8 mbar |
Detector | 11 Faraday detectorsa | 11 Faraday detectorsa | 9 Faraday detectorsb | 9 Faraday detectorsb |
Sample uptake rate | ∼70 μl min−1 | ∼210 μl min−1 | ∼70 μl min−1 | ∼320 μl min−1 |
![]() |
||||
Measurement parameters | ||||
Sample gas flow rate | ∼0.75–0.80 l min−1 | ∼0.645 l min−1c | ∼0.75 l min−1 | ∼0.130 l min−1c |
Auxiliary gas flow rate | ∼0.75–0.85 l min−1 | ∼0.80 l min−1 | ∼0.70 l min−1 | ∼1.50 l min−1 |
Solution concentration | 100 ppb | 20 ppb | 100 ppb | 20 ppb |
Typical sensitivity | 100 V ppm−1 | 300 V ppm−1 | 60 V ppm−1 | 250 V ppm−1 |
Sample measurement time | 50 × 8 s integrations | 50 × 8 s integrations | 40 × 8.4 s integrations | 40 × 8.4 s integrations |
Washout time | 3.5 min | 15 min | 3.5 min | 15 min |
Background measurement time | 20 × 8 s integrations | 20 × 8 s integrations | 20 × 4.2 s integrations | 20 × 4.2 s integrations |
Mass spectrometer | L5 | L4 | L3 | L2 | L1 | C | H1 | H2 | H3 | H4 | H5 |
---|---|---|---|---|---|---|---|---|---|---|---|
a Note that the Neptune in Münster only had 9 Faraday cups instead of the 11 Faraday cups of the Neoma in Göttingen. | |||||||||||
Neptune | — | 68Zn | 69Ga | 70Ge | 71Ga | 72Ge | 73Ge | 74Ge | 76Ge | 77Se | — |
Neoma | 66Zn | 67Zn | 69Ga | 70Ge | 71Ga | 72Ge | 73Ge | 74Ge | 75As | 76Ge | 77Se |
For both instruments, sample solutions were introduced into the mass spectrometer using either a Savillex C-Flow PFA nebulizer (∼70 μl min−1 uptake rate) connected to a Cetac Technologies Aridus II desolvator (hereafter AR) or a Teledyne Cetac Technologies HGX-200 Hydride Generator (hereafter HG). For the AR measurements on the Neoma, the oxide formation was minimized by the addition of N2 to the sample gases, resulting in oxide formation rates below 1–2% (measured as Ce/CeO). On the Neptune, however, N2 could not be used to minimize oxide production because large quantities of 40ArN2 would form, which directly interferes with 68Zn. Given the Faraday cup configuration on the Neptune, 68Zn is the only monitor available to correct for isobaric 70Zn interferences on 70Ge (note that this is not a problem for the Neoma, where the cup configuration allows monitoring 66Zn and 67Zn). To reduce oxide production on the Neptune, other parameters such as the auxiliary gas flow rate and torch position were adjusted. For the HG measurements, a 0.25 M NaOH–0.27 M NaBH4 solution was freshly prepared before each measurement sequence and mixed with the sample solutions in the reaction chamber of the HG to produce Ge hydrides, which were then introduced into the mass spectrometer.
On both instruments, a combination of standard sampler and X skimmer cones was used and measurements were performed in low resolution mode. Using the AR setup, a signal intensity of ∼10 V on 70Ge (70Ge is the most abundant Ge isotope in the optimal ∼50:
50 double spike–sample mixture) was obtained for a ∼100 ppb Ge solution at a ∼70 μl min−1 uptake rate. For the HG setup, ∼6 V on 70Ge was obtained for a ∼20 ppb Ge solution at a ∼210 μl min−1 uptake rate. Ion beams were collected simultaneously using Faraday cups (9 cups covering the mass range 68Zn to 77Se on the Neptune, 11 cups covering the mass range 66Zn to 77Se on the Neoma) connected to 1011 Ω feedback resistors for all five Ge isotopes (70Ge, 72Ge, 73Ge, 74Ge, and 76Ge) and the two interference monitors 66Zn (Neoma) or 68Zn (Neptune) and 77Se. On the Neoma, each analysis consisted of on-peak background measurements with 20 × 8 s integrations on a solution blank, followed by 50 cycles of isotope measurements of 8 s each. On the Neptune, on-peak background measurements consisted of 20 × 4.2 s integrations on a solution blank, followed by 40 cycles of isotope measurements of 8.4 s each.
δ74/70Gesample = −1000 × (αsample − αBracketing NISTSRM3210a) × ln(m74/m70), | (1) |
![]() | (2) |
The results for the standards and samples are reported as the mean of replicate measurements and the corresponding errors are Student-t 95% confidence intervals (95% CI) for samples with N > 3, or the long-term external reproducibility (2 s.d.) of the Ge Alfa Aesar standard solution measurements for samples with N ≤ 3.
Mass spectrometer | 70Ge | 72Ge | 73Ge | 74Ge | 76Ge |
---|---|---|---|---|---|
a The 70Ge and 73Ge single spikes were mixed together in a ratio of 73![]() ![]() |
|||||
70Ge single spike | 98.45 (11) | 0.562 (9) | 0.173 (66) | 0.592 (21) | 0.131 (11) |
73Ge single spike | 0.68 (17) | 1.99 (27) | 94.62 (97) | 2.36 (56) | 0.34 (3) |
70Ge–73Ge double spike | 71.5028 (24) | 0.9920 (11) | 26.2724 (23) | 1.0559 (16) | 0.1769 (12) |
The accuracy of the double spike calibration was tested by measuring twelve double spike–standard mixtures with double spike fractions varying between ∼10 and ∼90% (the optimal spike proportion is 52.15%) using both the Neptune and Neoma, and using both the AR and HG setups on both instruments. The double spike–standard mixtures were equilibrated in closed Savillex PFA vials with HNO3–HF on a hotplate at 120 °C. For the final double spike calibration, the isotopic compositions of the double spike and the NIST SRM 3120a Ge solution standard were iteratively adapted within their measured uncertainties, to allow for optimal agreement among the Ge isotopic compositions determined for the various double spike–standard mixtures. As shown in Fig. 4, the δ74/70Ge values obtained for the different spike–standard mixtures are indistinguishable within 0.05–0.07‰ (2 s.d.; depending on the measurement setup) for all four setups (i.e., AR-Neptune, AR-Neoma, HG-Neptune, and HG-Neoma). This together with the constancy of the δ74/70Ge values over a wide range of spike-to-sample ratios indicates that the double spike calibration is accurate.
![]() | ||
Fig. 4 Diagrams of the proportion of double-spike in variable double spike–sample mixtures vs. the measured Ge isotopic composition using the Cetac Technologies Aridus II desolvator (AR; panels A and B) or the Teledyne Cetac Technologies HGX-200 Hydride Generator (HG; panels C and D) as sample introduction systems to the Neptune Plus (A and C) or Neoma (B and D) MC-ICP-MS, respectively. In each case the different double spike–sample mixtures define a near-horizontal array, illustrating that (moderate) under- or overspiking has no effect on the analyses and that precise Ge isotope measurements are obtained for the different solution introduction systems and mass spectrometer instruments used. Data points represent the mean of five repeated measurements of each double spike–sample mixture, and error bars correspond to the Student-t 95% confidence intervals (CI). The bright grey bands refer to the external reproducibility of each measurement routine (2 s.d.). The dashed vertical lines depict the optimal spike–sample mixing ratio (f = 0.5204) as calculated using the Double Spike Toolbox.29 |
Second, we analyzed two terrestrial basalts (BHVO-2 and BCR-2) and several iron meteorite samples, which have already been analyzed in prior studies.9,15,17,18 These previous studies did not use a double spike, so these data also provide a useful comparison of different analytical methods for Ge isotope measurements. The results of this study are in very good agreement with previously reported results (Table 3, Fig. 5 and 8), demonstrating good overall agreement among the different analytical methods.
![]() | ||
Fig. 5 Comparison of the δ74/70Ge data obtained for different iron meteorite samples, terrestrial basalts, metal reference materials, and the Alfa Aesar Ge solution standard analyzed in this study using (A) either the AR desolvator or the HG setup, or (B) either the Neptune or Neoma MC-ICP-MS. In both cases, all samples align along a 1![]() ![]() ![]() ![]() |
The typical external reproducibility (2 s.d.) of the NIST SRM 3120a standard measurements performed during a given measurement session is ±0.08‰ (2 s.d.) for δ74/70Ge. This uncertainty is determined via standard–sample bracketing, where the average composition of the NIST SRM 3120a standard measurements during one session, αMeanSRM, is subtracted from each individual standard measurement in the same session:
δ74/70Gesample = −1000 × (αSRM − αMeanSRM) × ln(m74/m70) | (3) |
Repeated analyses of the Ge Alfa Aesar solution standard measured during the course of this study yielded a mean δ74/70Ge value of −0.76 ± 0.09‰ (N = 194; 2 s.d.; Fig. 6), defining a similar external reproducibility as the individual NIST SRM 3120a measurements. To account for potential effects induced by the chemical separation of Ge and by the presence of matrix elements in the analyzed Ge cuts, it is useful to compare the external reproducibility (2 s.d.) obtained for pure standard solutions to that of natural samples. The IID iron meteorite Needles has been measured 24 times during the course of this study using both the AR and HG setups on the Neptune and Neoma instruments. All four different measurement setups yield indistinguishable δ74/70Ge values and similar external reproducibilities, and the total external reproducibility of the combined measurements is ±0.10‰ for δ74/70Ge (2 s.d., Table 4), similar to the long-term external reproducibilities obtained for the pure solution standards. This good agreement indicates that sample processing and different sample matrices have no significant effect on the precision of the Ge isotope measurements, and that a value of ±0.09‰ (2 s.d.) provides a good estimate for the external precision of our δ74/70Ge measurements.
![]() | ||
Fig. 6 (A) Long-term reproducibility (2 s.d.) of the δ74/70Ge measurements of the optimally-spiked Ge Alfa Aesar standard solution over a period of two years (N = 194). The data points are arranged in the temporal order and include all measurements using the AR-Neptune, HG-Neptune, AR-Neoma, and HG-Neoma setups, respectively. Due to the temporally variable use of either of these setups, the data are not shown individually per setup. However, as seen in Table 4, the mean δ74/70Ge and corresponding external reproducibility of each measurement setup are indistinguishable from each other. Error bars on the individual data points represent the internal error (2 s.e.) of each measurement (e.g., typically ±0.045). (B) External reproducibility of repeated δ74/70Ge measurements of the IID iron meteorite Needles (N = 24) using the AR-Neptune, HG-Neptune, AR-Neoma, and HG-Neoma setups, respectively, indicating that potential interfering matrix effects as well as sample processing through our analytical protocol do not compromise the accuracy and reproducibility of the Ge isotope measurements. |
Sample | Classification | Gea (μg g−1) (±2σ) | N | δ 74/70Ge (±2σ) | N | δ 74/70Ge (±2σ) | N | δ 74/70Ge (±2σ) | N | δ 74/70Ge (±2σ) | N | δ 74/70Ge (±2σ) | δ 74/70Ge (±2σ) | References |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Aridus II @ Neptune | Hydride Generator @ Neptune | Aridus II @ Neoma | Hydride Generator @ Neoma | Average (this study) | Literature data | |||||||||
a The δ74/70Ge data of individual samples are reported as the mean of pooled measurements and the corresponding uncertainties reflect Student-t 95% confidence intervals (95% CI) for samples with N > 3 or the long-term external reproducibility (2 s.d.) of the Ge Alfa Aesar standard solution measurements of 0.09‰ for samples with N ≤ 3. Stated uncertainties on the δ74/70Ge data for the Ge Alfa Aesar standard solution measurements represent two times the standard deviation of the respective measurements. N: Number of Ge isotope analyses. Germanium concentrations as determined by isotope dilution (e.g., Stracke and Scherer, 2014;30 see the main text for details). References: (1) Luais et al., 2007;16 (2) Escoube et al., 2012;18 (3) Rouxel and Luais, 2017;9 (4) Rouxel et al., 2006;15 (5) Luais et al., 2012.17 b Re-normalized to the Ge SRM 3120a solution standard by Luais et al. (2012). c This Alfa Aesar solution has been processed together with the samples through the entire analytical protocol, including sample digestion, chemical separation of Ge, and isotope measurements. | ||||||||||||||
Iron meteorites | ||||||||||||||
C. Diablo | IAB | 301.8 ± 3.3 | 7 | 0.87 ± 0.07 | 1 | 0.77 ± 0.09 | 8 | 0.86 ± 0.07 | 0.88 ± 0.19 | 1 | ||||
Odessa | IAB | 244.3 ± 2.8 | 6 | 0.98 ± 0.05 | 1 | 0.92 ± 0.09 | 7 | 0.97 ± 0.04 | 0.90 ± 0.17 | 1 | ||||
Odessa | IAB | 0.96 ± 0.07 | 2 | |||||||||||
Landes | IAB | 366.6 ± 4.0 | 6 | 0.92 ± 0.04 | 1 | 0.97 ± 0.09 | 7 | 0.93 ± 0.04 | 0.83 ± 0.20 | 1 | ||||
Kumerina | IIC | 97.0 ± 1.1 | 6 | 1.16 ± 0.06 | 1 | 1.17 ± 0.09 | 7 | 1.16 ± 0.05 | 1.31 ± 0.18 | 1 | ||||
Needles | IID | 96.4 ± 0.9 | 7 | 1.09 ± 0.04 | 1 | 1.12 ± 0.09 | 6 | 1.09 ± 0.06 | 10 | 1.06 ± 0.04 | 24 | 1.08 ± 0.02 | ||
Miles | IIE | 69.1 ± 0.8 | 6 | −0.39 ± 0.06 | 1 | −0.47 ± 0.09 | 1 | −0.45 ± 0.09 | 1 | −0.42 ± 0.09 | 9 | −0.41 ± 0.04 | −0.59 ± 0.16 | 1 |
W. Station | IIE | 68.1 ± 0.8 | 6 | −0.50 ± 0.09 | 1 | −0.55 ± 0.09 | 7 | −0.51 ± 0.07 | −0.63 ± 0.20 | 1 | ||||
Henbury | IIIAB | 35.4 ± 0.4 | 5 | 0.94 ± 0.01 | 2 | 0.92 ± 0.09 | 1 | 1.02 ± 0.09 | 1 | 0.95 ± 0.09 | 9 | 0.95 ± 0.02 | ||
![]() |
||||||||||||||
Terrestrial standards | ||||||||||||||
BHVO-2 | Basalt | 1.63 ± 0.02 | 1 | 0.51 ± 0.09 | 5 | 0.54 ± 0.03 | 6 | 0.53 ± 0.03 | 0.53 ± 0.12 | 2 and 3 | ||||
BCR-2 | Basalt | 1.57 ± 0.02 | 3 | 0.58 ± 0.09 | 1 | 0.58 ± 0.09 | 4 | 0.58 ± 0.04 | 0.55 ± 0.13 | 2, 4 and 5 | ||||
![]() |
||||||||||||||
Solution standard | ||||||||||||||
Alfa Aesar | 67 | −0.77 ± 0.09 | 9 | −0.76 ± 0.08 | 104 | −0.76 ± 0.09 | 14 | −0.76 ± 0.08 | 194 | −0.76 ± 0.09 | ||||
Alfa Aesar (proc.)c | 5 | −0.75 ± 0.02 | 5 | −0.75 ± 0.02 | ||||||||||
![]() |
||||||||||||||
Metal reference materials | ||||||||||||||
NIST 129c | High-S steel | 26.3 ± 0.3 | 7 | −0.90 ± 0.04 | 7 | −0.90 ± 0.04 | ||||||||
NIST 361 | Steel | 32.0 ± 0.4 | 2 | 0.25 ± 0.09 | 5 | 0.24 ± 0.05 | 7 | 0.24 ± 0.03 |
Similarly, for the HG setup, Zn is efficiently removed because it does not form hydrides and, consequently, does not affect the Ge isotope measurements. However, for the AR setup isobaric Zn interferences may be relevant. We, therefore, tested the tolerable amount of Zn in the measured Ge solutions for reliable correction and accurate measurements using the AR setup by a series of Zn doping tests, in which variable amounts of Zn were added to the spiked NIST SRM 3210a Ge standard solution. The final Ge cuts after the chemical purification of Ge had Zn/Ge < 0.005, hence, we performed a Zn doping test in the Zn/Ge range between 0.00025 and 0.11. The results of the doping tests show that the presence of Zn can be accurately corrected for up to Zn/Ge ratios of 0.01, which is much higher than the range of Zn/Ge observed for the final Ge cuts of the samples.
Further isobaric interferences may arise from oxides and other polyatomic compounds. However, oxide production is irrelevant for HG measurements as all forming oxides of elements that have been hydrogenated before introduction into the mass spectrometer would have higher masses than the Ge isotopes used in the double spike inversion.41–43 For the AR setup, oxides might affect the quality of the Ge isotope measurements, but care was taken to minimize oxide production rates. The excellent agreement of measured δ74/70Ge values using either the AR or HG setup illustrates that the potential production of oxides has no significant effect on the Ge isotope measurements using the AR setup (Fig. 7).
![]() | ||
Fig. 7 Plot of δ74/70Ge vs. Zn/Ge, illustrating the effect of isobaric Zn interferences (on 70Ge) on the measured δ74/70Ge composition of variably Zn-doped, optimally spiked NIST SRM 3120a Ge solution standard mixtures using the AR as the sample introduction system to the Neoma MC-ICP-MS. The horizontal grey band represents the long-term external reproducibility of our measurement routine of 0.09‰ (Fig. 6), and the vertical green band represents the total range of Zn/Ge of the processed standard reference materials, natural samples, and standard solutions analyzed in this study. In the range of relevant Zn/Ge, isobaric Zn interferences are either insignificant or can be corrected accurately. Of note, doping tests using the AR as the sample introduction system to the Neptune MC-ICP-MS yield very similar results. By contrast, Zn doping tests cannot be made using a HG setup, as Zn does not form hydrides and, thus, is efficiently removed using the HG as the sample introduction system. Therefore, the general very good accordance of AR and HG measurements gives further proof for the accurate correction of isobaric Zn interferences using the AR setup. |
The major advantage of the HG setup over the AR setup is that volatile aqueous species (e.g., GeOH4) are converted to gaseous hydride species (e.g., GeH4gas). This effect not only removes isobaric interferences from elements that do not form hydrides (e.g., Zn, FeO, and NiO),22 but also results in a strong increase in Ge sensitivity.9 Moreover, as noted above, the oxide formation has no effect on the Ge isotope measurements, because Ge is the lightest hydride-forming element in classical HG setups. The major disadvantage of the HG setup over the AR setup is that it is more time-consuming and more complex in handling the measurement setup (e.g., rinse times have to be increased to 15 min to ensure an efficient wash-out of any Ge from the HG tubing; the high uptake rate consumes more chemicals needed in the preparation of the measurement solution; reducing agents (NaOH or NaBH4) are needed). Thus, while the HG setup may be the best choice for Ge isotope measurements in samples with low Ge abundances or complex sample matrices, for most samples the AR setup may be the easier and more efficient method for Ge isotope measurements.
![]() | ||
Fig. 8 Caltech diagram of the δ74/70Ge data of the eight iron meteorites (NCs in red and CCs in blue), two terrestrial basalt standards (green), the Ge Alfa Aesar solution standard (yellow), and two metal reference materials (grey) analyzed in this study. Uncertainties of the individual samples are Student-t 95% confidence intervals (95% CI). Light colored symbols represent literature data; references are listed in Table 4. |
All samples analyzed in this study are isotopically fractionated relative to the NIST SRM 3120a standard and display overall δ74/70Ge variations of up to ∼2‰. These variations are well-resolved using our Ge double spike method, given that the external uncertainty of our method is 0.09‰ (2 s.d.). The Alfa Aesar Ge solution standard and NIST 129c metal have negative δ74/70Ge values and, thus, are isotopically lighter than the NIST SRM 3120a standard, while the NIST 361 metal has positive δ74/70Ge values and, thus, is isotopically heavier than the NIST SRM 3120a standard. These isotopic variations among different reference materials may either be inherited from different starting raw materials (e.g., ores) or are the result of isotope fractionation induced during the standard production. The terrestrial basalts BHVO-2 and BCR-2 show distinctly more positive δ74/70Ge values compared to all other reference materials mentioned above. These more positive δ74/70Ge values are indistinguishable from values reported for other terrestrial igneous rocks,9,10 suggesting that high-temperature magmatic processes induce no significant Ge isotope fractionation.
The three magmatic iron meteorites investigated here (i.e., groups IIC, IID, and IIIAB) display even heavier Ge isotope compositions than the terrestrial basalts, in line with observations from previous studies.16,17 Moreover, despite having variable Ge concentrations and Ge/Ni ratios, these three samples have very similar δ74/70Ge values, suggesting that the variable volatile element depletions of their parent bodies did not induce large Ge isotope fractionations. By contrast, the non-magmatic iron meteorites investigated here (i.e., groups IAB and IIE) display variable Ge isotope compositions, including both light and heavy compositions. These samples might have formed by local impacts on planetary surfaces and, thus, might have lost significant amounts of volatile Ge in the course of impact heating.38,39 As such, these samples provide evidence for Ge isotope fractionation during localized heating and degassing. Overall, the data of this study demonstrate that large Ge isotope fractionation occur among terrestrial and extraterrestrial samples, making Ge isotopes a powerful tool for a wide range of geochemical and cosmochemical applications.
This journal is © The Royal Society of Chemistry 2025 |