Anja Ileničac,
Radmila Milačič Ščančar
bc,
Marija Đurić
a,
Alenka Mauko Pranjića and
Janez Ščančar
*bc
aDepartment of Materials, Slovenian National Building and Civil Engineering Institute, Ljubljana, Slovenia. E-mail: janez.scancar@ijs.si
bDepartment of Environmental Sciences, Institute Jožef Stefan, Ljubljana, Slovenia
cJožef Stefan International Postgraduate School, Ljubljana, Slovenia
First published on 29th August 2025
A sampling device was developed for collecting and size-fractionating airborne particulate matter (PM). A low-volume cascade system with polytetrafluoroethylene membrane filters (PM10, PM2.5, and PM0.1) connected to an ultrapure-water trap was used to retain the PM<0.1 fraction that passed through the filters. In the collected samples, metal(oid)s and platinum group elements (PGEs) were determined by inductively coupled plasma mass spectrometry after microwave-assisted digestion using a mixture of acids. Polycyclic aromatic hydrocarbons (PAHs) were extracted with a solvent mixture of acetone and petroleum ether, assisted by mechanical shaking and determined by gas chromatography mass spectrometry. The analytical methods were optimised and validated using the urban PM Standard Reference Material 1648a. Measurement repeatability and accuracy were 6% and 2% for metal(oid)s and PGEs, and 3% and 6% for PAHs, respectively. Field emission scanning electron microscopy analysis confirmed that the sampling device efficiently fractionated airborne PM to nano-sized particles. The sampling device was successfully applied to collect PM, enabling effective air pollution monitoring in urban areas. Analysis revealed that metal(oid)s were most abundant in PM10, while those originating from traffic were also elevated in PM<0.1. High-molecular-weight PAHs were distributed between PM10 and PM2.5 fractions. The sampling device provides valuable insights into the chemical composition of (ultra)fine particles, overcoming the limitations of filter-based PM fractionation and providing improved air pollution control for nanoparticle-bound pollutants.
Numerous epidemiological and toxicological studies have linked air pollution to adverse health effects and related health risks.6–8 Even limited exposure can lead to oxidative stress at the air-lung interface, contributing to acute and chronic cardiopulmonary diseases, as well as neurodegenerative and reproductive disorders.9–14 These effects are more pronounced with fine and ultrafine particles, as 50% of these particles remain in the lung parenchyma.15 Additionally, fine and ultrafine particles have longer atmospheric residence times, ranging from a few hours to weeks, until they are removed by either wet or dry deposition.16 Studies have shown that toxicity of PM per mass unit increases as particle size decreases, indicating that fine particles are more toxic than coarser ones.17,18 This increased toxicity is primarily due to the larger total porous surface area of finer particles, which enables them to more easily adsorb and retain toxic substances.15
Understanding the relationship between chemical composition and sizes of particles is crucial, but published data indicate that this remains limited. Most research on airborne pollutants involves the use of high-volume samplers (>100 L min−1) to collect larger sample quantities, enabling the detection of compounds with lower limits of detections (LODs).19–22 However, increased air flow rates can potentially reduce the adsorption capacity of filters and increase the likelihood of filter clogging, making them less suitable for accurately assessing human health risks during daily commuting. People typically inhale 10 to 15 times per minute, equating to about 6 L min−1, making low-volume samplers (10 L min−1) more suitable alternatives for human health risk assessments.23–28
Adequate sampling techniques are particularly important for detecting pollutants present at very low concentrations, such as metal(oid)s and polycyclic aromatic hydrocarbons (PAHs), which pose significant health and environmental risks. The United States Environmental Protection Agency (US EPA) has identified 16 priority PAHs for analysis in various environmental samples.29 Among these, benzo[a]pyrene (BaP) is recognised as one of the most potent carcinogens. Other PAHs, such as benzo[a]anthracene (BaA), benzo[b]fluoranthene (BbF), benzo[k]fluoranthene (BkF), chrysene (Ch) and indeno[1,2,3-cd]pyrene (IP) are classified as possibly carcinogenic to humans.30,31 Among metal(oid)s, arsenic (As), hexavalent chromium (Cr(VI)), cadmium (Cd), nickel (Ni), and lead (Pb) are of particular concern as carcinogenic airborne contaminants.32 Accurate chemical analysis of these PAH compounds and metal(oid)s is essential across all inhalable PM sizes, including nanoparticles, which are often overlooked.
Previous studies have reported PAHs in size-fractioned PM10 and PM2.5 (ref. 33–35) and metals in size-fractioned PM10 and PM2.5.36–38 However, the simultaneous determination of both types of pollutants in different PM sizes is rarely reported.39,40 Very few studies performed characterisation of PM of smaller fractions.41–44 Physio-chemical analyses of samples collected on filters by low-volume cascade samplers present significant challenges, mainly due to the small amount of the PM collected. Therefore, highly sensitive analytical techniques such as inductively coupled plasma mass spectrometry (ICP-MS) and gas chromatography mass spectrometry (GC-MS) are used for chemical analysis of samples. Analytical methods with high selectivity, resolution and sensitivity must be applied to provide data for reliable assessment of health and environmental hazards. However, only a few atmospheric studies report the use of certified reference materials (CRMs). These materials are analysed to check the accuracy of results and validate the analytical procedures.26,45–47
Given that pollutants bound to airborne (ultra)fine particles have the most harmful effects on human health, it is essential to improve pollution control measures and enhance risk assessment strategies by taking into account various PM fractions, including the nano-sized PM particles. Therefore, the objectives of our work were (i) to develop a low-volume cascade sampling device for collecting and size-fractionating airborne PM, consisting of polytetrafluoroethylene (PTFE) membrane filters (PM10, PM2.5, and PM0.1) connected to an ultrapure-water trap designed to retain the PM<0.1 fraction; (ii) to verify the size of PM particles collected on filters using field emission scanning electron microscopy (SEM) analysis; (iii) to optimise and validate analytical methods for the determination of metal(oid)s, including platinum group elements (PGEs) using ICP-MS, and PAHs in the collected sample fractions using GC-MS; and (iv) to analyse the concentrations of metal(oid)s and PAHs in collected samples at different urban locations in Slovenia using the developed sampling device, placed within one metre of bicycle lanes, in order to assess the air pollution-related health risks for daily commuters.
PAH determination was carried out by GC-MS, using an Agilent HP6890 GC (Agilent Technologies, Tokyo, Japan) with a 5973 mass selective detector (MSD) and equipped with an Agilent 6890 Series autosampler injector. PAHs were separated on an Agilent capillary column HP-5MS (30 m × 0.25 mm, film thickness 0.25 μm) coated with 5% phenylmethylpolysiloxane. The GC-MS operating parameters are presented in Table S2.
A Milestone ETHOS UP Microwave Digestion System (Sorisole, Italy) was used for sample digestion in the analysis of metal(oid)s and PGEs. For PAH analysis, the samples were mechanically shaken using a Vibromix 40 orbital shaker (Tehtnica, Železniki, Slovenia). The sample extracts were concentrated using a sample concentrator Dri-Block Heater DB100/3 (Techne Inc., Vernon Hills, IL, USA). A Mettler AE 163 analytical balance (Zürich, Switzerland) was used for weighing.
Morphological analysis of airborne PM was conducted using SEM with a Zeiss ULTRA plus scanning electron microscope (Carl Zeiss AG, Oberkochen, Germany) equipped with an energy dispersive X-ray spectrometer (EDS) system manufactured by Oxford Instruments (High Wycombe, UK).
A Vacuubrand (Wertheim, Germany) low-volume diaphragm pump ME 1C (100 mbar, 0.8 m3 h−1) was used for sampling of real airborne PM, while air flow through the system was measured using a G1.6 gas meter (Dadolab, Cinisello Balsamo, Italy). The air flow rate was controlled using a variable area flow meter DK46/800 (Krohne, Wellingborough, UK). PM was collected using a three-stage cascade NILU polycarbonate filter holder system (Innovation NILU, Kjeller, Norway) equipped with hydrophobic Whatman PTFE membrane filters (pore sizes of 10 μm, 2.5 μm and 0.1 μm and diameters of 47 mm) obtained from Merck (Darmstadt, Germany).
For metal(oid)s analysis, Suprapur nitric acid (67–70% HNO3) was obtained from Carlo Erba Reagents (Val-de-Reuil, Normandie, France), while Suprapur hydrochloric acid (30% HCl) and Suprapur hydrofluoric acid (40% HF) were purchased from Merck (Darmstadt, Germany). ICP multi-element standard solution XVI (21 elements in diluted nitric acid, 100 mg L−1, part no. 1.09487.0100) was used to prepare calibration curves for the metal(oid)s arsenic (As), cadmium (Cd), chromium (Cr), cobalt (Co), copper (Cu), molybdenum (Mo), nickel (Ni), lead (Pb), thallium (Tl), vanadium(V) and zinc (Zn). For barium (Ba), a single standard (1000 ± 2 mg L−1 in 2–3% HNO3, part No. 1.70304.0100) was used. Single standard solutions of palladium (Pd) (1000 ± 2 mg L−1 in 0.5 mol L−1 HNO3, part no. 1.14282.0100) and platinum (Pt) (1000 ± 2 mg L−1 in 2 mol L−1 HCl, part no. 1.102645.0100) were used to prepare the calibration curve for PGEs. Stock standard solutions of germanium (Ge), rhodium (Rh) (1000 ± 2 mg L−1 in 2–3% HNO3, part Nos. 1.70320.0100 and 1.02650.0100, respectively), and iridium (Ir) (1000 ± 2 mg L−1 in 7% HCl, part no. 1.70325.0100) were used to prepare internal standards in ICP-MS analysis. All standard solutions were purchased from Merck.
For PAH analysis, acetone (C3H6O) and petroleum ether, both purchased from Merck, were used for extraction of the high-molecular weight PAHs benzo[a]anthracene (BaA), benzo[a]pyrene (BaP), benzo[b]fluoranthene (BbF), benzo[ghi]perylene (BghiP), chrysene (Ch), fluoranthene (Fl), indeno[1,2,3-cd]pyrene (IP) and pyrene (P). Anhydrous sodium sulphate (Na2SO4) from Merck was used for water removal from the organic phase. Strata FL-PR Florisil (activated magnesium silicate, MgO3Si, 170 μm, 80 Å, 500 mg3 mL−1; Phenomenex, Inc., Torrance, CA, USA) was used for the cleaning of sample extracts. A standard stock solution PAH-Mix 9 (10 μg mL−1, part no. DRE-XA20950009CY) in cyclohexane, containing 16 analytes, was used to create calibration curves for the determination of PAHs by GC-MS. A deuterated PAH-Mix 9 (10 μg mL−1, part no. DRE-L20950902CY) in cyclohexane, also containing 16 analytes, was used as an internal standard. Both the PAH-Mix standard and internal standard were purchased from LGC (Teddington, UK). Isooctane (2,2,4-trimethylpentane) obtained from Merck was used as a reference standard to calibrate the GC-MS system. The retention times of PAHs in the samples were compared to the retention time of isooctane, aiding in the identification of PAHs during the GC-MS analysis. A typical chromatogram illustrating the separation of the PAH-Mix 9 standard solution (16 PAHs, 500 ng mL−1) and the deuterated PAH-Mix 9 (16 PAHs-D, 500 ng mL−1) used as an internal standard, applying the GC-MS procedure, with PAHs considered for analysis marked in red, is presented in Fig. S1. Retention times of the separated PAHs and deuterated PAH compounds are presented in Table S3. The certified standard reference material SPS-SW1 (reference material for measuring elements in surface waters), obtained from Spectrapure Standards (Oslo, Norway), was used daily during ICP-MS analysis to verify the accuracy of the calibration curves and measurement procedures for metal(oid)s. Standard Reference Material (SRM) 1648a (Urban Particulate Matter) from the National Institute of Standards and Technology (NIST) (Gaithersburg, MD, USA), certified for selected elements and PAHs, was employed to verify the accuracy of metal(oid) determinations in digested PM, as well as PAHs in the extracted PM from both the filters and the aqueous phase. This SRM was also used to optimise and validate the analytical procedures.
The sampling unit consisted of a three-stage cascade filter holding system incorporating hydrophobic PTFE membrane filters with pore sizes of 10 μm, 2.5 μm and 0.1 μm (Fig. S3). This cascade filter system was connected to the water trap inside a 2000 mL amber glass Pyrex burette bottle in order to retain the nano-sized particles (PM<0.1). The air volume passing through the system was controlled by using a low-volume diaphragm pump connected to a gas meter. Prior to the sample analysis, the stability of the airflow rate of a low-volume diaphragm pump was monitored continuously over one month using a flow meter.
After the sampling, the nano-sized PM<0.1 fraction (captured in a water trap) and the PTFE membrane filters were stored at −15 °C until analysis. Each filter and the aqueous phase of the nano-sized PM<0.1 fraction represented the total monthly amount of PM. The volume of air pumped through the sampling unit was monitored throughout the entire sampling period, allowing for the reporting of metal(oid) and PAH concentrations in ng m−3 of air.
After sampling, filters (PM10, PM2.5, and PM0.1) were weighed, cut in half with iron scissors, and aliquots of the filters were weighed again to accurately determine the proportion of the whole PM collected in each aliquot. The nano-sized PM<0.1 collected in the 1 L water trap was acidified with 1 mL of HNO3, evaporated to 20 mL in a sand bath at 200 °C, and then divided into two 10 mL aliquots. One half of the filters (PM10, PM2.5, and PM0.1) and one aliquot of the nano-sized PM<0.1 fraction were used for the analysis of metal(oid)s, while the other half was used for the analysis of PGEs.
For the determination of metal(oid)s, 3 mL of HNO3, 1 mL of HCl and 0.2 mL of HF were used in microwave-assisted digestion, while for the determination of PGEs, aqua regia (3 mL of HCl and 1 mL of HNO3) with the addition of 0.2 mL of HF was used. Samples were subjected to closed-vessel microwave-assisted digestion, using the following temperature programme: ramp to 140 °C for 15 min, hold at 140 °C for 5 min, ramp to 200 °C for 20 min, hold at 200 °C for 60 min, and cool down for 30 min. The clear digests of the filters and the nano-sized PM<0.1 aqueous phase fraction were quantitatively transferred to 30 mL polypropylene graduated tubes and filled to the mark with ultra-pure water. Concentrations of metal(oid)s in digested samples were determined by ICP-MS using external calibration for quantification. The same analytical procedures, with no samples added, were applied to determine procedural blanks. Concentrations of metal(oid)s and PGEs in the procedural blanks were below the LODs, or did not exceed 3% of the measured elemental concentrations.
After sampling, PAHs collected on the filters (PM10, PM2.5, and PM0.1) and the nano-sized PM<0.1 fraction retained in the aqueous phase were extracted using organic solvents. The filters were transferred to Erlenmeyer flasks, and a mixture of 15 mL of acetone and petroleum ether (1:
1, v/v) and 100 μL of isooctane was added. The samples were then spiked with 50 μL of an internal standard solution (10 μg mL−1) and shaken on a mechanical shaker for 16 h at 180 rpm.
For the nano-sized PM<0.1 fraction, a liquid–liquid extraction (LLE) was performed. A mixture of 100 mL of acetone and petroleum ether (1:
1, v/v) along with 100 μL of isooctane was added to 1 L of the sample. After spiking with 50 μL of the internal standard solution (10 μg mL−1), the mixture was shaken for 16 h at 180 rpm. The content was quantitatively transferred into a glass separatory funnel to separate the organic phase from the aqueous phase. Traces of water in the organic phase were removed using anhydrous Na2SO4, and pre-heated at 550 °C for 6 h.
The organic phase from both the filter extracts and the nano-sized PM<0.1 fraction was passed through a 3 mL column filled with Florisil for sample clean-up. The organic phase was concentrated to approximately 1 mL under a nitrogen stream at 40 °C using a sample concentrator. The concentrated samples were then transferred to 2 mL amber vials with a Pasteur pipette and sealed. Finally, 1 μL of the sample was injected onto the column for the determination of PAHs by GC-MS, using external calibration for quantification. The same analytical procedure, with no samples added, was applied to determine procedural blanks. Concentrations of PAHs in the procedural blanks were below the LODs.
A flow chart illustrating the optimised analytical procedures for the determination of metal(oid)s, PGEs and PAHs is presented in Fig. 1.
![]() | ||
Fig. 1 Flow chart of the optimised analytical procedures for the determination of metal(oid)s, PGEs and PAHs in size-fractionated airborne PM. |
The studied metal(oid)s and PGEs were selected due to their association with traffic emissions from the wear of engine parts, lubricants and additives used in gasoline (Ba, Cr, Cu, Mo, Pb, V, and Zn),38,49 and emissions from diesel fuels (Ba, Cd, Cr, Mo, Ni, V, and Zn),49,50 brake pads (Ba, Cu, and Mo) and tyre wear (Ba, Cd, and Zn),38 as well as their use as catalysts to control car exhaust emissions (Pd and Pt).51
Element | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Average (mg kg−1) | RSD (%) |
---|---|---|---|---|---|---|---|---|
As | 127 | 129 | 123 | 120 | 118 | 109 | 119 | 5.4 |
Ba | 801 | 812 | 807 | 806 | 805 | 815 | 808 | 0.6 |
Cd | 75.0 | 72.8 | 74.5 | 71.3 | 73.4 | 73.6 | 73.4 | 1.6 |
Co | 18.7 | 19.1 | 18.6 | 18.4 | 17.3 | 19.0 | 18.5 | 3.3 |
Cr | 390 | 400 | 391 | 367 | 384 | 426 | 393 | 4.5 |
Cu | 738 | 748 | 748 | 749 | 700 | 658 | 724 | 4.7 |
Mo | 22.6 | 21.1 | 22.9 | 21.4 | 22.9 | 21.3 | 22.0 | 3.5 |
Ni | 85.3 | 86.4 | 87.5 | 85.7 | 82.3 | 76.5 | 84.0 | 4.4 |
Pb | 6527 | 6374 | 6455 | 6544 | 6359 | 6441 | 6450 | 1.1 |
Pd | 0.69 | 0.49 | 0.52 | 0.62 | 0.50 | 0.41 | 0.54 | 17 |
Pt | 0.041 | 0.027 | 0.040 | 0.022 | 0.030 | 0.033 | 0–032 | 21 |
Tl | 1.87 | 1.91 | 1.96 | 1.81 | 1.86 | 1.66 | 1.85 | 5.6 |
V | 137 | 139 | 153 | 150 | 140 | 131 | 142 | 5.3 |
Zn | 5241 | 5260 | 4744 | 4561 | 4610 | 4263 | 4780 | 7.6 |
Good repeatability of measurement was obtained for metal(oid)s, with an average of RSD 6% and accuracy of 2%. However, the repeatability for Pt and Pd was lower, with RSDs of 21% and 17%, respectively, due to their low concentrations in the urban dust samples analysed.
The LODs and LOQs for the determination of individual metal(oid) and PGE concentrations in the sample digests were calculated as the concentration that provided signals equal to 3s or 10s of the blank sample, respectively. To calculate the LODs and LOQs, 8 blank samples (acids used for digestion) were analysed by ICP-MS. The LODs and LOQs for metal(oid)s and PGEs, expressed in ng mL−1, and for urban dust samples expressed in mg kg−1, are provided in Table S4. Values expressed in ng m−3 are presented in Table S5. Low LODs, ranging from 0.001 to 0.02 ng mL−1 were achieved, enabling the quantification of metal(oid)s and PGEs in urban dust and airborne PM using the optimised analytical procedures.
The linearity of measurement for metal(oid)s and PGEs ranged from the LOQ to 1000 ng mL−1, with correlation coefficients (R2) for the calibration curves better than 0.998.
The accuracy of the optimised analytical procedures was verified by analysing SRM 1648a and comparing the determined concentrations with the certified concentrations. For elements not certified in SRM 1648a, a spike recovery test was applied. The sample was spiked with an appropriate amount of selected elements before microwave-assisted digestion. Nitric acid digestion was used for Ba, Mo and Tl, and aqua regia digestion for Pd and Pt. The relative recovery, expressed as a percentage, was calculated as the ratio between the found and expected concentrations (the sum of the element concentration in urban dust and the added concentration) multiplied by 100. The results are presented in Table 2.
Element | Determined concentration (mg kg−1) | Certified concentration (mg kg−1) | Added concentration (mg kg−1) | Found concentration (mg kg−1) | Relative recovery (%) |
---|---|---|---|---|---|
As | 121 ± 2.9 | 115.5 ± 3.9 | — | — | — |
Ba | 808 ± 2.0 | — | 472 | 1321 | 103 |
Cd | 73.4 ± 0.5 | 73.7 ± 2.3 | — | — | — |
Co | 18.5 ± 0.3 | 17.93 ± 0.68 | — | — | — |
Cr | 393 ± 7.9 | 402 ± 13 | — | — | — |
Cu | 724 ± 15.2 | 610 ± 70 | — | — | — |
Mo | 22.0 ± 0.3 | — | 18.9 | 41.3 | 101 |
Ni | 84.0 ± 1.6 | 81.1 ± 6.8 | — | — | — |
Pb | 6450 ± 31.1 | 6550 ± 330 | — | — | — |
Pd | 0.54 ± 0.04 | — | 9.5 | 10.24 | 98.5 |
Pt | 0.032 ± 0.003 | — | 14.35 | 13.83 | 96.2 |
Tl | 1.85 ± 0.04 | — | 1.9 | 3.6 | 97.3 |
V | 142 ± 3.3 | 127 ± 11 | — | — | — |
Zn | 4780 ± 162 | 4800 ± 270 | — | — | — |
It is evident that the determined concentrations are in good agreement with the certified values. The relative recoveries of the spiked elements ranged between 96% and 103%, indicating quantitative determination of Ba, Mo, Pd, Pt and Tl in urban dust samples. Therefore, it can be concluded that the optimised analytical procedures provide accurate results for the analysis of the studied metal(oid)s and PGEs.
Initially, extraction was performed using a 100 mL mixture of acetone and petroleum ether (1:
1, v/v). However, due to the small quantity of the sample, it was experimentally found that the volume of the extracting agent can be significantly reduced. For a 40 mg urban dust SRM sample, a mixture of 15 mL of acetone and petroleum ether (1
:
1, v/v), along with 100 μL of isooctane, was added. The samples were spiked with 50 μL of an internal standard solution (10 μg mL−1) and subjected to microwave-assisted extraction (40 °C, 600 W) for 1 h, or shaken on a mechanical shaker for 16 h or 24 h (25 °C, 180 rpm).
The organic phase was passed through a 3 mL column filled with Florisil for sample clean-up and concentrated to approximately 1 mL under a nitrogen stream at 40 °C using a sample concentrator. GC-MS analysis indicated that samples shaken for 16 h achieved better recovery rates than those shaken for 24 h (recoveries greater than 7% and 11%, respectively), while the recovery rate for microwave-assisted extraction was lower (14%). Thus, mechanical shaking for a duration of 16 h was chosen as the optimal extraction procedure, followed by the clean-up step. Additionally, the results indicated that a single cleaning of the organic phase with Florisil yielded better recoveries compared to a double cleaning (recoveries of 7% and 11%, respectively). Consequently, a single cleaning procedure was employed for the petroleum ether extract containing SRM urban dust.
The optimised procedure was then applied to the determination of PAHs in the PM10, PM2.5 and PM0.1 fractions collected on filters and for the PM<0.1 fraction retained in the aqueous phase, as described in the section Optimised analytical procedure for the analysis of PAHs.
In this study, PAHs were investigated due to their common presence as organic contaminants in urban dust and PM, originating from vehicle exhaust, combustion processes and industrial activities.56,57 The following 8 PAHs were selected for monitoring due to their high toxicity and prevalence in the environment: BaA, BaP, BbF, BghiP, Ch, Fl, IP, and P.31
PAH | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Concentration (mg kg−1) | Average (mg kg−1) | RSD (%) |
---|---|---|---|---|---|---|---|---|
BaA | 2.63 | 2.56 | 2.52 | 2.56 | 2.59 | 2.59 | 2.58 | 1.3 |
BaP | 2.39 | 2.27 | 2.30 | 2.20 | 2.37 | 2.32 | 2.31 | 2.8 |
BbF | 7.70 | 8.88 | 8.48 | 8.33 | 7.90 | 8.04 | 8.22 | 4.8 |
BghiP | 4.32 | 4.03 | 4.23 | 4.18 | 4.17 | 4.14 | 4.18 | 2.1 |
Ch | 6.75 | 6.70 | 6.81 | 6.66 | 6.85 | 6.81 | 6.76 | 1.0 |
Fl | 7.59 | 7.65 | 7.57 | 7.40 | 7.52 | 7.96 | 7.62 | 2.3 |
IP | 4.08 | 4.55 | 4.15 | 4.16 | 3.95 | 3.78 | 4.11 | 5.7 |
P | 6.03 | 6.00 | 5.92 | 5.86 | 5.98 | 6.16 | 5.99 | 1.6 |
The results in Table 3 demonstrate good repeatability of measurement, with an average RSD of 3% and accuracy of 6%.
The LODs and LOQs for the determination of concentrations of PAHs in sample extracts were calculated as the concentration that provided a signal equal to 3s or 10s of the blank sample, respectively. To calculate the LODs and LOQs, 8 blank samples (petroleum ether extract) were analysed by GC-MS. The LODs and LOQs for PAHs, expressed in ng mL−1, and for urban dust samples expressed in mg kg−1, are provided in Table S6, and expressed in ng m−3 in Table S7. The LODs for the PAHs measured (4.5 ng mL−1) were adequate for the quantification of PAHs in urban dust and airborne PM using the optimised analytical procedures.
The linearity of measurement for PAHs ranged from the LOQ to 1000 ng mL−1, with correlation coefficients (R2) for the calibration curves better than 0.997.
The accuracy of the optimised analytical procedures was verified by analysing SRM 1648a and comparing the determined concentrations with the certified values. The results are presented in Table 4.
PAH | Determined concentration (mg kg−1) | Certified concentration (mg kg−1) |
---|---|---|
BaA | 2.58 ± 0.02 | 2.71 ± 0.15 |
BaP | 2.31 ± 0.03 | 2.57 ± 0.10 |
BbF | 8.22 ± 0.18 | 8.89 ± 0.05 |
BghiP | 4.18 ± 0.04 | 5.00 ± 0.18 |
Ch | 6.76 ± 0.03 | 6.12 ± 0.06 |
Fl | 7.62 ± 0.08 | 8.07 ± 0.14 |
IP | 4.11 ± 0.11 | 4.17 ± 0.17 |
P | 5.99 ± 0.04 | 5.88 ± 0.07 |
Data from Table 4 show good agreement between the determined PAH concentrations. Deviations from the certified value did not exceed 10%, and the exception was only BghiP, for which the determined concentration was 16% lower than the certified value. These data confirm that the optimised analytical procedure provides accurate results for the analysis of the studied PAHs.
The identified particles were classified into six morphological clusters (Fig. 2): spherules (P1, P7, P12, and P13), floccules (P2), biogenic particles (P3), tabular particles (P5 and P10), irregular shaped grains (P6 and P8) and agglomerate particles (P9 and P11). Most of the identified particles were non-spherical, thereby providing a higher surface area for potential toxic compound adsorption and enhancement of their interaction with lung tissues.61 The most common particles found in this study include alumosilicates (P1), magnesium silicates (P7), and carbonate particles (P6), along with sulphates (P8 and P10) and carbon-enriched particles (P5) (Table S9).
Anthropogenic particles include spherules, floccules and agglomerates. Spherules are typically smaller, with an average diameter of less than 3 μm,62–65 and generally exhibit a Si–Al–Fe elemental composition. Floccules consist of small spherical particles (<1 μm) with a loose structure, while agglomerates are characterised by flexuous, fractal edge branches or crinkles randomly distributed on the surface, forming a lump-like morphology. Agglomerate and spherical particles are typically derived from coal combustion, while floccules result from vehicle emissions.65
Geogenic particles are mostly characterised by irregularly shaped grains, which exhibit smooth, flat surfaces (P5 and P8) or semi-coarse to coarse, holed surfaces (P6), with an average diameter of around 10 μm. These particles result from mechanical abrasion and originate from soil and surface deposits.65 Tabular particles appear more elongated, either rectangular or elliptical, and are generally emitted through geogenic processes and biological activity.65
The SEM-EDS images show that the particles collected on filters align with the pore sizes of the filters used, confirming the reliability of the low-volume cascade PTFE membrane filter system for fractionating PM10, PM2.5 and PM0.1 particle size fractions.
![]() | ||
Fig. 3 The distribution of metal(oid)s in size-fractionated airborne PM in Maribor (MB), Ljubljana (LJ) and Dobrova (DB). |
![]() | ||
Fig. 4 The distribution of PAHs in size-fractionated airborne PM in Maribor (MB), Ljubljana (LJ) and Dobrova (DB). |
Fig. 3 (continued) shows the distribution of metal(oid)s in size-fractionated airborne PM in Maribor (MB), Ljubljana (LJ) and Dobrova (DB).
As shown in Fig. 3, the highest concentrations of metal(oid)s are primarily found in PM10 particles, with lower concentrations observed in PM2.5 and PM0.1. In the PM<0.1 fraction, the concentrations of metal(oid)s (Ba, Cd, Cu, Ni, Pb, Pd, Pt, V and Zn) originating from traffic-related and coal combustion as well biomass sources, increased again to levels similar to those observed in PM10. This indicates that the PM<0.1 fraction, which penetrates most deeply into lung tissue and is potentially the most harmful to health, contributes significantly to the overall exposure to metal(oid)s.
The origin of As, Cr, Ni, Pb and V is primarily linked to high-temperature processes and fossil fuel combustion.37,40,69,70 Ba, Cr, Cu, Mo, Pb, V and Zn are associated with gasoline engines,38,49 while Ba, Cu and Mo arre related to lubricants and additives used in brake pads. Ba, Cd and Zn are linked to tyre wear38 and Pd and Pt originate from catalyst emissions in car exhausts.51
The data in Fig. 3 further show that the highest concentrations of metal(oid)s were recorded at MB and LJ, as these sites are impacted by relatively dense traffic and are moderately populated, with populations of approximately 100000 and 350
000, respectively. Lower metal(oid) concentrations were found at DB, which is a less densely populated residential area with less traffic. Additionally, LJ and MB also have a stronger presence of industrial facilities, including energy production plants, chemical manufacturing sites, and metal processing and waste management facilities.
Among metal(oid)s, Ba, Cr, Cu, Ni and Zn were measured in the highest concentrations in all size-fractioned PM, being elements typically associated with vehicular emissions. In MB, the PM10 fraction exhibited higher concentrations of As, Ba, Cr, Cu, Ni, Pb, and Zn compared to other fractions and locations, suggesting a more pronounced resuspension process and dust redistribution from both natural and anthropogenic sources.
In the PM10 fraction, Zn concentrations reached approximately 70 ng m−3 in MB and up to 15 ng m−3 in LJ, while Ba concentrations were around 30 ng m−3 in MB and 7 ng m−3 in LJ. Concentrations were significantly lower than those reported in larger cities, such as Fez in Morocco (with a population of about 1.1 million) or Changsha, China, where concentrations reached up to 250 ng m−3 for Zn and 70 ng m−3 for Ba.66,67 The concentrations of Zn and Ba determined in DB in the PM10 fraction (Fig. 3) reached up to 10 ng m−3 for Zn and 5 ng m−3 for Ba, showing slightly lower concentrations than those reported by Jandacka et al.41 for PM10 in the smaller city of Žilina, Slovakia (approximately 80000 inhabitants). Across all observed PM fractions from this study (2.8–61.5 ng m−3 for Zn; 0.45–28.0 ng m−3 for Ba), concentrations were similar to those in Turin, Italy, where Zn varied between 5 and 90 ng m−3 for 0.54–11 μm PM fractions,44 and to those reported in the Los Angeles area,38 where Zn varied between 6 ng m−3 and 8.7 ng m−3 for PM10 and PM2.5 respectively, and Ba between 16.5 ng m−3 and 14.9 ng m−3. Ba and Zn, most abundant in coarse PM fractions, are typical of tyre and brake wear. Since their emissions are largely unaffected by fuel type or emission standards, this contributes to the observed similarities in their concentrations across different cities.
The highest concentrations of Cu in MB in the PM10 fraction (around 30 ng m−3) were similar to those reported in Fez, Morocco,67 while Ni concentrations in MB (around 2.5 ng m−3) were significantly lower than those reported in Fez (95 ng m−3). At the DB and LJ sampling sites, Cu concentrations in the PM10 fraction ranged from 2.8 and 12 ng m−3, which is lower than the levels reported for Žilina (20 ng m−3).41 Ni concentrations in the PM10 fraction ranging from 0.7 to 2.7 ng m−3 were similar to those reported for Žilina (0.6 ng m−3).41 Additionally, Cu (0.15–28.6 ng m−3) and Ni (0.28–2.7 ng m−3) values across all PM fractions were similar to those reported in Leipzig (0.14–50 ng m−3 for Cu; 0.02–2.3 ng m−3 for Ni)36 and Los Angeles (7.5 ng m−3 for Cu; 0.4–0.5 ng m−3 for Ni).38 In Changsha, China,66 Cu values across PM > 9 μm and PM < 0.4 μm varied between 10 and 20 ng m−3. In Turin, Italy, Cu and Ni across 0.54–11.0 μm PM fractions varied from 0.5–3.3 ng m−3 and 0.5–6.0 ng m−3, respectively.44 Similarly to Ba and Zn, Cu concentrations are mainly influenced by non-exhaust traffic emissions, resulting in similar Cu emission profiles. In contrast, Ni is more characteristic of gasoline and diesel vehicles.
The levels of Pb in MB in PM10 (4 ng m−3) and PM2.5 as well as PM<0.1 (0.4 ng m−3) were similar to those observed across all PM fractions in Turin, Italy (0.25–4.5 ng m−3).44 In Los Angeles,38 average Pb concentrations in PM2.5 and PM10 varied between 0.7 and 1.9 ng m−3, significantly lower than those found in Fez (around 40 ng m−3)67 and Changsha, China (30−120 ng m−3),66 but higher than Pb concentrations at other sampling sites in this study (0.9 to 2.6 ng m−3) and in those reported for Žilina (1.7 ng m−3).41 The concentrations of Cr in the PM10 fraction were also the highest in MB (around 10 ng m−3) and in LJ in PM10 and PM0.1 (around 4 ng m−3). In PM<0.1, Cr varied between 0.12 and 0.88 ng m−3. Concentrations were lower than those reported for Fez (35 ng m−3)67 and Changsha (15−20 ng m−3).66 However, they were similar to those measured in Turin (0.3–2.5 ng m−3) and Los Angeles (0.8–1.4 ng m−3).38,44 At other sampling sites in this study, Cr in the PM10 fraction was considerably lower (between 2.3 and 4 ng m−3) and similar to the concentration level of Cr in Žilina (1.7 ng m−3).41 Higher concentrations of Pb in MB can be attributed to industrial activities such as metal processing and battery manufacturing, along with exhaust emissions. Similarly elevated concentrations of Cd, Cu and Zn were also attributed to the anthropogenic activities in the local area.67
In the PM10 fraction, As concentrations in MB were 1.2 ng m−3, while at other sampling sites, the levels reached 0.5 ng m−3, which is similar to the reported As concentration in Žilina.41 As in PM0.1 and PM<0.1 across all observed locations varied between 0.02 and 0.11 ng m−3. As in Turin, Italy44 in the PM fraction 0.54–11 μm varied between 0.04 and 0.16 ng m−3. Cd concentrations at all sampling sites were very low, ranging from 0.05 to 0.2 ng m−3. In Los Angeles,38 Cd concentrations in PM2.5 and PM10 varied between 0.013 and 0.047 ng m−3. Co, Mo, V and Tl were also found at low levels, with concentrations not exceeding 2 ng m−3.
As expected, PGEs (Pd and Pt) related to the traffic emissions were present in low concentrations at all sampling sites. Across all size fractions, Pd concentrations were generally higher than those of Pt. Pt concentrations varied between 0.002 and 0.02 ng m−3, while Pd ranged from 0.01 to 0.09 ng m−3. These concentrations are similar to those reported for side streets, non-urban and rural areas with low traffic near Frankfurt, Germany,43,69 and in aerosols from cities around the world, except for metropolitan areas such as Beijing, where Pt and Pd concentrations were significantly higher (around 0.2 and 0.8 ng m−3, respectively).51 In Changsha, China,66 Pd concentrations in PM10 and PM2.5 averaged around 0.02 ng m−3.
Despite the differences in city size between the urban areas selected in this study and larger cities like Turin or Los Angeles, the similar chemical fingerprinting of size-fractionated PM (PM10 and PM2.5) suggests comparable pollution level profiles for selected metal(oid)s. This may be attributed to more localised emission sources in Slovenian cities, higher traffic density per unit volume of air, and limited dispersion of air pollutants. Additionally, the timing and location of sampling also plays a significant role in comparability of detected metal(oid)s; for instance, PM sampling in Turin and Fez was conducted on rooftops, reflecting a greater influence of meteorological and regional emission factors rather than direct local exposure to traffic-related emissions.
High-molecular-weight PAHs (BaA, BaP, BbF, BghiP, Ch, Fl, IP and P), typically bound to particles, were found in the highest concentrations in the PM10 fraction. They were also present in the PM2.5 fraction at the MB and LJ sampling sites, but were not detected in the PM0.1 and PM<0.1 fractions. BbF, BghiP and IP, 5- and 6-ring PAHs typically associated with traffic-related sources exhibited higher concentrations in the PM2.5 fraction. Their concentrations were low, ranging from 0.07 to 0.2 ng m−3. BbF, BghiP and IP values in PM10, PM2.5 and PM1 observed in Zagreb, Croatia (population about one million)71 were higher, ranging between 0.4 and 3.4 ng m−3. Higher PAH concentrations (with the sum of PAHs in the PM10 fraction ranging from 7 to 10 ng m−3) were reported for the large city of Fez.68 Comparable concentrations of PAHs to those reported in our study were found by Xing et al.72 in Kanazawa, Japan (population of about 450000). They observed that the sum of 9 PAHs (Fl, P, BaA, Ch, BbF, BkF, BaP BghiP and IP) in the PM2.5 fraction ranged from 0.2 ng m−3 in summer to 1 ng m−3 in winter. A similarly low sum of PAHs (1–2.1 ng m−3) was reported by Jedynska et al.73 in PM10 and PM2.5 fractions observed in Rome, Copenhagen, London and Oslo. For comparison, in our study conducted in MB in December 2024, the sum of 8 PAHs in PM10 was 0.8 ng m−3 and in PM2.5 was 0.7 ng m−3. In LJ, the sum of 8 PAHs in the PM10 fraction was 0.7 ng m−3, while in the PM2.5 fraction, it was 0.8 ng m−3. The presence of high-molecular-weight PAHs can be attributed to traffic emissions as well as biomass burning.33–35
BaP, characterised as one of the most carcinogenic PAHs, varied between <0.02 ng m−3 and 0.1 ng m−3, with the highest concentrations observed in the PM10 fraction in MB. The data from the literature show that the concentration of BaP in Porto varied between 0.01 ng m−3 and 2.7 ng m−3, with an average value of 0.37 ng m−3; while in Athens it averaged at 0.25 ng m−3 and 0.21 ng m−3 in Copenhagen.73 In Zagreb, Croatia,71 BaP in PM10, PM2.5 and PM1 varied between 0.8 ng m−3 and 2 ng m−3.
Certain compounds, such as As, Cd, Ni and BaP, are identified as human genotoxic carcinogens. The European Union Directive (Directive 2004/107/EC 2004)74 has established threshold levels to mitigate their harmful effects on human health, with annual target values set for PM10 at 6 ng m−3 for As, 5 ng m−3 for Cd, 20 ng m−3 for Ni and 1 ng m−3 for BaP. In this study, none of the target values were exceeded in any size-fractioned PM samples. Threshold limits for other metal(oid)s, PGEs and PAHs for finer PM fractions have not yet been established.
The results of this study revealed that metal(oid)s, PGEs and PAHs are primarily attributed to anthropogenic activities, particularly traffic emissions and biomass burning. However, their concentrations at the selected sampling sites, located within one metre of bicycle lanes, were low, indicating that the pollution-related health risks for daily commuters are minimal.
The developed sampling device for size-fractionated PM, in combination with optimised analytical procedures, enables reliable and sensitive analysis of metal(loid)s, PGEs and PAHs at very low concentrations and is also applicable in areas with high levels of air pollution. It makes a significant contribution to improving air pollution control and legislation, identifying the sources of air pollution and assessing the associated health and environmental risks.
The applicability of the sampling device was demonstrated through a one-month study conducted at three different locations in Slovenia, which differ in size, population and traffic density. The results revealed that the highest concentrations of metal(oid)s and PGEs were predominantly found in the PM10 fraction, largely influenced by resuspension processes and dust redistribution. Trace elements linked to traffic related emissions, both exhaust and non-exhaust, as well as coal and biomass combustion (Ba, Cd, Cu, Ni, Pb, Pd, Pt, V, and Zn) were significantly elevated in PM<0.1 across all sampling locations, indicating the importance of characterising this size range. High-molecular-weight PAHs were distributed between PM10 and PM2.5 fractions, with 5-ring (BaP and BbF) and 6-ring (BghiP and IP) PAH compounds, typically associated with traffic emissions, found in higher concentrations in the PM2.5 fraction. Contaminant concentrations were low, indicating the minimal pollution-related health risks for daily commuters using bicycle lanes.
This one-month study, incorporating a newly developed sampling device, will be further expanded into a systematic one-year study to analyse size-fractionated PM samples across urban areas in Slovenia.