Open Access Article
This Open Access Article is licensed under a
Creative Commons Attribution 3.0 Unported Licence

Highly sensitive and accurate detection of cholesterol based on a single red upconversion biosensor

Pang Tingyuana, Liu Xiaoruia, Li Jiaa, Song Qia, Li Junrena, Han Lingb, Shu Wenyinga, Jian Xiaoshuna and Zhang Meimei*a
aDepartment of Pharmacy, Affiliated Cancer Hospital, Institute of Guangzhou Medical University, Guangzhou, China
bIntegrated Traditional Chinese and Western Medicine Department (Internal Medicine Section 5), Affiliated Cancer Hospital & Institute of Guangzhou Medical University, Guangzhou, China

Received 28th October 2023 , Accepted 2nd December 2023

First published on 6th March 2024


Abstract

Cholesterol (CHOL) is an important clinical biochemical indicator that plays an important role in the regulation of the fluidity, permeability, and microstructure of cell membranes. Therefore, it is necessary to accurately monitor CHOL levels in biological samples for the early prevention and diagnosis of various diseases. The single-band red upconversion nanoparticle (UCNP) emits light within the optical transmission window of biological tissues, and can penetrate deeper biological tissues and cause less energy loss due to scattering and thus have higher sensitivity and accuracy. Here, using the nontoxic, sensitive, and photochemically stable 3,3′,5,5′-tetramethylbenzidine (TMB) as the quenching agent and single red UCNP as the fluorescent donor, a dual-readout colorimetric and fluorescent sensor was developed to detect CHOL. The detection mechanism and feasibility were discussed in detail, and experimental conditions such as Fe2+ concentration, TMB concentration and reaction time were explored. Under optimal conditions, the limits of CHOL detection by colorimetry and fluorescence were 0.85 μM and 0.63 μM. The sensing system was used to measure CHOL in serum samples and the values obtained by these two modes were close, and the spiked recoveries were 97.2–102.2% and 97.1–103.7%, respectively, which holds great potential in clinical diagnosis and health management.


1 Introduction

As an essential lipid, cholesterol (CHOL) is the structural component of the hormone system and the only precursor of steroid hormones, and is essential to maintain the integrity of the biofilm and cell structure and the fluidity of cell membranes.1,2 Abnormal levels of CHOL in serum are associated with many diseases, for example, excessive levels of CHOL have a connection with the occurrence and development of coronary heart disease, cardiovascular disease, hypertension, and myocardial infarction.3,4 Low CHOL levels may also lead to hypolipoproteinemia, sepsis, malnutrition, hyperthyroidism, and liver disease.5 Therefore, it is significant to accurately monitor CHOL levels in biological samples for early prevention and diagnosis of related diseases.

In recent years, the reported CHOL detection methods have mainly included electrochemical biosensors,6 colorimetry,7 chemiluminescence8 and fluorescence biosensing.9 Among them, the fluorescence method has attracted much attention due to some advantages of sensitivity, stability, and simple operation, but when applied to biological samples, it is limited owing to interference from the spontaneous fluorescence of biological tissues and scattered light. Unlike traditional fluorescent materials, upconversion nanoparticle (UCNP) is inorganic luminescent materials that could transform near-infrared (NIR) excited light into visible or ultraviolet light, with rich energy levels and long-life emission, good photochemical stability, continuous emission capacity, low potential toxicity and relatively narrow bandwidth.10–12 Because most substances in the organism do not emit light under the excitation of NIR light, the light scattering caused by biological tissues is greatly decreased and the self-background fluorescence can be ignored. At the same time, the penetration depth of biological tissues becomes greater and can effectively reduce the damage to biological organs and tissues, which makes UCNP capable of being applied to the complex biological internal environment.13 More interestingly, red UCNP luminescence falls in the optical transmission window of biological tissue, can penetrate deeper biological tissue, and less energy is lost by scattering, making them ideal sensor luminescent materials. Rakov14 prepared red UCNP by doping Er3+ and Tm3+ in powders of YOF and Y2O3. However, the synthesized samples had poor morphology and relatively weak red luminescence because of the combustion preparation method. In addition, many researchers directly use KMnF3 as a lattice host and uniformly doping Er3+ in KMnF3 to achieve single-red upconversion emission through energy transfer between Mn2+ and Er3+ ions. However, the formed cubic phase and the presence of electron vacancies are not conducive to upconversion emission.15 In order to overcome the above shortcomings of the preparation of single-band bright red UCNP, Xu16 prepared a novel and highly efficient active core–shell UCNP (NaErF4[thin space (1/6-em)]:[thin space (1/6-em)]Tm@NaGdF4[thin space (1/6-em)]:[thin space (1/6-em)]Yb) using Er3+ ions as both sensitizer and activator, which emits ultra-bright red by 980 nm excitation. To alleviate the fluorescence quenching induced by energy transfer between Er3+ ions and internal lattice defects, a Tm3+ capture center is doped in the core to limit the energy of Er3+ through energy return (Er3+ → Tm3+ → Er3+). Subsequently, the active shell is grown in the core region, and the high efficiency energy transfer of near infrared photons in the up-conversion region is realized, which inhibits the fluorescence quenching induced by surface defects and surface-related ligands. This research provides a convenient method for preparing single red UCNP, however, the application of biosensors based on red UCNP in disease diagnosis still needs to be developed, especially in the study of CHOL detection in complex biological samples.

Herein, a red UCNP-based sensor platform was developed for the detection of CHOL through coupling the cycle signal amplification strategy and the mechanism of detecting CHOL is shown in Scheme 1. First, CHOL is catalyzed by cholesterol oxidase (CHOD) to produce hydrogen peroxide (H2O2), which oxidizes Fe2+ to generate Fe3+ and the hydroxyl radical (·OH). Then 3,3′,5,5′-tetramethylbenzidine (TMB) is oxidized by the Fe3+ and ·OH produced to TMBox, during which Fe2+ is generated and can be recycled for producing TMBox. Finally, TMBox can effectively quench the red light emitted by UCNP through the internal filtration effect (IFE). The practicability and analysis principle of the developed assay for detecting CHOL were explored in detail and the effect of some detection condition on the sensing system was also studied. Finally, the proposed sensing platform was used to measure CHOL in actual serum samples, and the detection data obtained by colorimetry and fluorescence were compared.


image file: d3ra07354h-s1.tif
Scheme 1 Detection mechanism diagram of CHOL.

2 Materials and methods

2.1 Materials

Gadolinium chloride hexahydrate (GdCl3·6H2O), ytterbium chloride (YbCl3·6H2O), thulium chloride hexahydrate (TmCl3·6H2O), Erbium chloride hexahydrate (ErCl3·6H2O), ammonium fluoride (NH4F), oleic acid (OA) and 1-octadecene (1-ODE) were obtained from Shanghai Aladdin Biochemical Technology Co., Ltd. (Shanghai, China) (Shanghai, China). 3,3′,5,5′-Tetramethylbenzidine (TMB), cholesterol (CHOL) and cholesterol oxidase (CHOD) were gotten from Aladdin Biochemical Reagent Co. Ltd. (Shanghai, China). Sodium hydroxide, hydrochloric acid, methyl alcohol, ethyl alcohol, cyclohexane, phosphate, FeCl2 and hydrogen peroxide (H2O2) were bought from Sinopharm Chemical Reagent Co. Ltd (Shanghai, China). Triton X-100 was purchased from Sigma-Aldrich (Shanghai) Trading Co., Ltd. Prepare a cholesterol stock solution by adding Triton X-100 to dissolve the cholesterol solid and dilute it with phosphate buffer saline (PBS) (10 mM, pH 7.4).

2.2 Measurements

The morphology of UCNP is tested by transmission electron microscopy using a Talos F200i apparatus (Carl Zeiss AG, Germany). UV-vis absorption and fluorescence spectrums of different samples were determined on a Lambda 950 UV spectrophotometer (PerkinElmer, USA) and a FluoroMax-4 spectrophotometer (Horiba Jobin Yvon, France) with laser excitation at 365 or 975 nm (Changchun New Industries, China), respectively. The crystalline phase of the prepared material was tested by an Ultima III X-ray diffractometer (Rigaku, Japan). Fourier transmission infrared spectra of UCNP were tested using a Nicolet-5700 infrared spectrometer (Thermo Elemental, USA).

2.3 Preparation of single red UCNP

The preparation of core–shell NaErF4[thin space (1/6-em)]:[thin space (1/6-em)]0.5% Tm@NaGdF4[thin space (1/6-em)]:[thin space (1/6-em)]40% Yb UCNP was referred to the reported literature.17 As UCNPs is used in the biological detection system in this work, water-soluble core–shell UCNP is obtained by removing OA from the surface of core–shell UCNP with hydrochloric acid. The detailed steps are as follows. The OA-UCNP dispersion is added to an equal amount of 2 M HCl solution and ultrasound is performed for 30 min. After centrifugation and washing, the ligand-free hydrophilic UCNP is obtained.

2.4 Fluorescent and colorimetric assay for the detection of H2O2

For H2O2 recognition by the fluorescent method, 20 μL of UCNP suspension (1 mg mL−1), 100 μL of HAc–NaAc buffer solution (100 mM, pH 3.6), 200 μL of H2O2 solution of different concentrations (0.5, 1.0, 5.0, 10, 50, 80 μM), 50 μL of 180 μM Fe2+, and 30 μL of TMB (4 mM) were blended. After reaction for 15 min and diluting to 500 μL with water, the luminous intensity was determined under 980 nm excitation.

For H2O2 recognition by the colorimetric method, 50 μL of UCNP suspension (1 mg mL−1), 200 μL of HAc–NaAc buffer solution (100 mM, pH 3.6), 400 μL of H2O2 solution with different concentrations (1, 5, 10, 35, 50, 80 μM), 100 μL of 180 μM Fe2+, and 60 μL of TMB (4 mM) were mixed. After incubation for 15 min and diluting to 2000 μL with water, the UV absorption intensity was determined.

2.5 Fluorescent and colorimetric assay for CHOL detection

The fluorescent assay for CHOL was realized under the following procedures. 200 μL of PBS (10 mM, pH 7.4) containing different concentrations of cholesterol (1, 15, 25, 60, 110, 160, 195 and 225 μM) was mixed with 50 μL of CHOD solution (1 mg mL−1). After reaction for 45 min, add 100 μL of HAc–NaAc buffer solution (100 mM, pH 3.6), 50 μL of Fe2+ (180 μM), 30 μL of TMB (4 mM) and 20 μL of UCNP (1 mg mL−1). After reaction for 15 min and diluting to 500 μL with water, the luminous intensity was determined under 980 nm excitation.

The colorimetric assay for CHOL was realized under the following procedures. 400 μL of PBS (10 mM, pH 7.4) containing different concentrations of cholesterol (1, 3, 5, 10, 20, 25, 55, 115, 175 and 220 μM), and 100 μL of CHOD solution (1 mg mL−1) were blended and reacted for 45 min. Then add 200 μL of HAc–NaAc buffer solution (100 mM, pH 3.6), 100 μL of Fe2+ (180 μM), 60 μL of TMB (4 mM) and 40 μL of UCNP (1 mg mL−1). After incubation for 15 min and diluting to 2000 μL with water, the intensity of UV absorption was determined.

2.6 Detection of CHOL in actual samples

At room temperature, the blood coagulated naturally for 10–20 min, centrifuged at 10[thin space (1/6-em)]000 rpm for 10 min, collected the supernatant and diluted 100 times with phosphate buffer solution (10 mM, pH 6.7). The follow-up procedure for the detection of CHOL levels in serum is the same as in Section 2.5.

3 Results and discussions

3.1 Characteristics of UCNP

The morphological structure of the resultant UCNP was observed by transmission electron microscopy (TEM), energy dispersive spectrometer (EDS), X-ray diffractometry (XRD) and spectral properties. As shown in Fig. 1A and B, the TEM images clearly show that the prepared UCNP nanocrystals are uniformly dispersed. EDS analysis (Fig. 1C) indicates that the nanoparticles are mainly composed of Na, F, Er, Tm, Yb and Gd. Compared to the XRD pattern of the NaErF4 standard card, it can be seen in Fig. 1D that the crystal structure of UCNP is the same as that of NaErF4, which is a standard pure hexagonal phase with high crystallinity. As shown in Fig. 1E, the prepared UCNP emits strong red light at 654 nm under 980 nm excitation (black line), which can be attributed to the 4F9/24I15/2 transitions of Er3+.18 While TMBox has a wide UV-vis absorption spectrum at 550–750 nm (red line), which overlaps well with the red light emitted by UCNP and can effectively quench the fluorescence of UCNP. Therefore, a sensing platform for detecting CHOL is constructed using TMBox as the energy receptor and UCNP as the energy donor.
image file: d3ra07354h-f1.tif
Fig. 1 Morphological characterization and spectrum of UCNP. TEM (A and B), EDS (C), XRD (D) and fluorescence spectrum (E) of the prepared UCNP.

3.2 Feasibility and detection mechanism

To explore the feasibility of the proposed method, the fluorescent and UV-vis spectroscopy of the different samples were performed. When CHOL exists, it reacts with CHOD to generate H2O2, converting Fe2+ to Fe3+ and ·OH. TMB is oxidized to TMBox by Fe3+ and ·OH. The generated TMBox can effectively quench the fluorescence of UCNP, resulting in a significant quenching of the 654 nm fluorescence emission peak and a corresponding increase in the UV absorption peak around 654 nm (as shown in Fig. 2A and B). Therefore, the proposed assay for detecting CHOL through a fluorescent and colorimetric dual signal is feasible.
image file: d3ra07354h-f2.tif
Fig. 2 Fluorescent (A) and UV-vis spectra (B) of the different samples, the fluorescent lifetime of UCNP before and after TMBox (C) and the zeta potential of UCNP and TMBox (D). Experimental conditions: [UCNP]: 1 mg mL−1; [Fe2+]: 250 μM; [TMB]: 5 mM; [H2O2]: 200 μM; [Fe3+]: 250 μM; [HAc–NaAc buffer]: 100 mM, pH 3.6.

The quenching mechanism may be attributed to fluorescence resonance energy transfer (FRET) or IFE.19 The difference between FRET and IFE is manifested mainly in two aspects: (1) the fluorescence lifetime in FRET decreases, while the fluorescence lifetime in IFE does not change20 and (2) the distance between the energy donors and receptors in FRET should be short enough (usually less than 10 nm) to allow the energy transfer process to occur, while the distance between the two materials in IFE is not relevant.21 To investigate the luminescence quenching mechanism of UCNP by TMBox, the luminescence lifetime of UCNP was investigated with and without TMBox, respectively. Fig. 2C shows that the fluorescence lifetime of a single UCNP solution is approximately 183.90 μs and the fluorescent lifetime of the mixed solution of UCNP and oxTMB is approximately 180.45 μs, which is almost the same. Therefore, IFE is the quenching mechanism of TMBox on UCNP. To further confirm this mechanism, the distance between energy donors and receptors can be explained by measuring the surface charges of UCNP and TMBox using the zeta potential. Fig. 2D shows that the zeta potentials of UCNP and TMBox are +35.9 mV and +22.2 mV, respectively. Due to electrostatic repulsion, the space between TMBox and UCNP could be greater than 10 nm,22 which is not conducive to the occurrence of FRET.

3.3 Optimization of the determination conditions

According to the analysis mechanism described above, when CHOL exists, it reacts with CHOD to generate H2O2, converting Fe2+ to Fe3+ and ·OH. Under acidic conditions, Fe3+ and ·OH by the Fenton reaction can oxidize TMB to TMBox, effectively quenching the luminescence of UCNP. To achieve the best detection effect, the relevant detection conditions were studied. Under the following experimental conditions, the relative luminescence quenching rate ((F0F)/F0) of the sensing system (F0 and F express the fluorescent intensity values before and after CHOL addition, respectively) attains the optimal value: pH value is 3.6 (Fig. 3A), Fe2+ concentration is 0.18 mM (Fig. 3B); TMB concentration is 4 mM (Fig. 3C); the reaction time is 15 min (Fig. 3D).
image file: d3ra07354h-f3.tif
Fig. 3 Effects of pH (A), Fe2+ concentration (B), TMB concentration (C) and reaction time (D) on ((F0F)/F0) of the detection system.

3.4 Detection of H2O2

Fig. 4A shows that the fluorescence intensity of the detection system continuously decreases at 654 nm when the H2O2 concentration gradually increases. As shown in Fig. 4B, if the H2O2 concentration increases from 0.5 to 80 μM, the value of (F0F)/F0 of the detection system continuously increases, furthermore, there is a good linear relationship between the value of (F0F)/F0 and the H2O2 concentration. The linear regression equation is Y = 0.0070X + 0.0595 with a correlation coefficient of 0.9994. According to the formula LOD = 3σ/K (where s is the slope of the standard curve, σ is the standard deviation), the detection limit (LOD) of H2O2 is calculated to be 0.33 μM.
image file: d3ra07354h-f4.tif
Fig. 4 Fluorescence spectra of the detection system with increasing concentration of H2O2 (A), relationship between the (F0F)/F0 value and the concentration of H2O2 (B), absorbance spectra of the detection system with increasing concentration of H2O2 (C) and relationship between relative absorption intensity and concentration of H2O2 (D). Experimental conditions: [UCNPs]: 1 mg mL−1; [Fe2+]: 180 μM; [TMB]: 4 mM; [HAc–NaAc buffer]: 100 mM, pH 3.6.

Under optimal conditions, the sensing platform can also detect H2O2 by colorimetry. Fig. 4C shows that the absorbance value of the detection system at 654 nm gradually increases as the H2O2 concentration increases. Within the range of 1 to 80 μM, there is a good linear relationship between the H2O2 concentration and the absorbance value of the sensing platform. The linear equation is Y = 0.0040X + 0.0442 with a correlation coefficient of R2 = 0.9965 (Fig. 4D) and an LOD of 0.66 μM.

3.5 Detection of CHOL

Under the best conditions, the different concentrations of CHOL were detected by the fluorescence-based sensing platform. Fig. 5A shows that the fluorescence intensity of the detection system continuously decreases at 654 nm as the concentration of CHOL gradually increases. As shown in Fig. 5B, as the concentration of CHOL gradually increased from 1 to 225 μM, the value of (F0F)/F0 of the detection system continues to increase and there is a good linear relationship between the value of (F0F)/F0 and the CHOL concentration, according to the linear regression equation Y = 0.0037X + 0.0164 with a correlation coefficient of 0.9994. By formula LOD = 3σ/K, the LOD of CHOL is calculated to be 0.63 μM.
image file: d3ra07354h-f5.tif
Fig. 5 Fluorescence of the detection system with increasing concentration of CHOL (A), relationship between relative (F0F)/F0 value and concentration of CHOL (B), absorbance spectra of the detection system with increasing concentration of CHOL (C) and relationship between relative absorption intensity and concentration of CHOL (D). Experimental conditions: [UCNPs]: 1 mg mL−1; [Fe2+]: 180 μM; [TMB]: 4 mM; [HAc–NaAc buffer]: 100 mM, pH 3.6.

Similarly, the response of the colorimetry-based sensing platform to different concentrations of CHOL was also measured. Fig. 5C shows that the absorbance value of the detection system at 654 nm gradually increases as the CHOL concentration increases. There is a good linear relationship between CHOL concentration in the range of 1–220 μM and the absorbance value of the detection system. The linear equation is Y = 0.0031X + 0.0438 with a correlation coefficient of 0.9998 (Fig. 5D) and the LOD is 0.85 μM. A comparison of the analytical performance of this method with previously reported sensors based on various nanomaterials was conducted to detect CHOL. The results show that compared to other CHOL sensors reported in the literature, the LOD of the sensing platform proposed in this work is lower or equivalent (Table 1).

Table 1 Comparison of reported strategies and this work detection for CHOLa
Material Detection method Linear range (μM) Limit of detection (μM) References
a Boron nitride nanosheet and copper sulfide nanohybrids: BNNS@CuS, Prussian blue: PB, multi-walled carbon nanotubes: MWCNT.
BNNS@CuS Colorimetry 10–100 2.9 23
PB/MWCNT Colorimetry 4–100 3.01 24
AuNCs Fluorescence 10–100 5.77 25
Grp/b-CD/rhodamine 6G Fluorescence 5–30 5 26
β-CD-CQD Fluorescence 0–110 0.7 27
UCNPs Fluorescence 1–225 0.63 This work
Colorimetry 1–220 0.85


3.6 Specific verification

To verify the specificity of the sensing platform for CHOL detection, other possible interfering substances in the serum such as NaCl, KCl, cysteine, glycine, glucose, uric acid, glycerin, ascorbic acid, and urea were analyzed as controls. Fig. 6 shows that there is no obvious change in the values of (F0F)/F0 in the presence of other interfering substances, while the values of (F0F)/F0 significantly increases in the presence of CHOL. The above results indicate that the proposed assay has good selectivity for detecting CHOL and is expected to be used for efficient detection of CHOL in actual samples.
image file: d3ra07354h-f6.tif
Fig. 6 Effects of interfering substances on the (F0F)/F0 value of the detection system. Selectivity of the proposed sensing platform. Experimental conditions: [CHOL]: 200 μM, [NaCl]: 1 mM, [KCl]: 1 mM, [cysteine]: 200 μM, [glycine]: 1 mM, [glucose]: 1 mM, [uric acid]: 200 μM, [glycerin]: 1 mM, [ascorbic acid]: 200 μM, [urea]: 1 mM.

3.7 Detection of CHOL in human serum

To evaluate the practicality of the proposed detection system, CHOL was detected in human serum samples. First, the actual concentration of CHOL was measured in diluted human serum samples. Table 2 shows that the measured CHOL values in the diluted serum samples of (19.90 and 35.41 μM, fluorescence method), (19.73 and 35.15 μM, colorimetric method). To further investigate the accuracy, stability, and actual detection performance of the sensing platform, the CHOL spiked recovery experiment was conducted. After adding different concentrations of standard CHOL samples (1, 50 and 100 μM) to human serum samples, the measured recoveries range from (97.2% to 102.2%, fluorescence method), (97.1% to 103.7%, colorimetric method) and the relative standard deviation (RSD) (n = 3) range from (1.23% to 3.21%, fluorescence method) (1.22% to 3.76%, colorimetric method). These results indicate that the proposed sensing platform can be used for the analysis of CHOL in actual biological samples and provide the possibility for real-time and accurate detection of CHOL in actual biological samples.
Table 2 Detection results of CHOL in human serum samples
Samples Added (μM) Fluorescence Colorimetry
Found (μM) Recovery (%) RSD (%) Found (μM) Recovery (%) RSD (%)
Serum 1 0 19.90 1.59 19.73 2.10
1 20.88 97.4 3.21 20.73 99.9 1.50
50 70.91 102.0 1.23 71.58 103.7 1.34
100 119.40 98.5 2.65 121.40 101.7 2.21
Serum 2 0 35.41 2.25 35.15 1.22
1 36.41 99.9 2.32 36.16 100.8 3.76
50 84.03 97.2 1.93 84.69 97.1 2.37
100 137.66 102.2 1.63 136.29 101.1 2.87


4 Conclusions

A dual mode signal-sensing platform based on fluorescence and colorimetry has been constructed for the quantitative detection of CHOL in biological samples using UCNP with strong single-infrared emission as the luminescent material and TMBox as the quenching agent. The practicability and analysis principle of the developed assay for detecting CHOL were explored in detail and the effect of some detection condition on the sensing system was also studied. Compared to traditional biosensors, the proposed sensing platform has a better response through the coupling IFE and cyclic amplification strategy. Under the best experimental conditions, the LOD of CHOL by the fluorescence and colorimetry-based sensing platform are 0.63 and 0.85 μM, respectively. The quantitative detection of CHOL in human serum shows that the developed assay has good accuracy and stability. The sensing platform constructed in this work is expected to offer new ideas for the development of CHOL clinical detection methods that are easy to operate, highly sensitive and selective.

Ethical statement

Human serum samples were collected from Affiliated Cancer Hospital of Guangzhou Medical University. Before laboratory study, written informed consent was signed by all volunteers. All experiments were performed in accordance with the guidelines approved by Affiliated Cancer Hospital of Guangzhou Medical University. The research protocol (no. GYZL-ZN-2023(053)) was approved by Medical Ethics Committee of Affiliated Cancer Hospital of Guangzhou Medical University.

Author contributions

Pang Tingyuan, Liu Xiaorui and Zhang Meimei designed research. Pang Tingyuan, Liu Xiaorui, Li Jia and Song Qi performed the experiments. Li Junren, Han Ling, Shu Wenying and Jian Xiaoshun analyzed data. All author wrote and revised the manuscript.

Conflicts of interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgements

This work was supported by plan on enhancing scientific research in GMU, Guangzhou Traditional Chinese Medicine and Integrated Traditional Chinese and Western Medicine Science and Technology Project (20222A010049, 20182A011026); National Natural Science Foundation of China (No. 82204372); Science and Technology Program of Guangzhou (No. 2023A04J0601); Guangdong Basic and Applied Basic Research Foundation (2022A1515110525). General Guidance Project of Guangzhou Municipal Commission of Health and Family Planning (20171A011318).

References

  1. L. Goicoechea, L. C. de la Rosa and S. Torres, et al., Mitochondrial cholesterol: metabolism and impact on redox biology and disease, Redox Biol., 2023, 102643,  DOI:10.1016/j.redox.2023.102643.
  2. N. Motomura, Y. Yamazaki and D. Koga, et al., The Association of Cholesterol Uptake and Synthesis with Histology and Genotype in Cortisol-Producing Adenoma (CPA), Int. J. Mol. Sci., 2022, 23(4), 2174,  DOI:10.3390/ijms23042174.
  3. R. D. Santos, Complexity of Triglyceride-rich Lipoproteins Remnant Cholesterol With Atherosclerotic Cardiovascular Disease Risk, Eur. J. Prev. Cardiol., 2023, zwad064,  DOI:10.1093/eurjpc/zwad064.
  4. L. Guo, S. Chen and Y. L. Yu, et al., A smartphone optical device for point-of-care testing of glucose and cholesterol using Ag NPs/UiO-66-NH2-based ratiometric fluorescent probe, Anal. Chem., 2021, 93(48), 16240–16247,  DOI:10.1021/acs.analchem.1c04126.
  5. S. Nantaphol, O. Chailapakul and W. Siangproh, Sensitive and selective electrochemical sensor using silver nanoparticles modified glassy carbon electrode for determination of cholesterol in bovine serum, Sens. Actuators, B, 2015, 207, 193–198,  DOI:10.1016/j.snb.2014.10.041.
  6. R. S. Dey and C. R. Raj, Development of an amperometric cholesterol biosensor based on graphene–Pt nanoparticle hybrid material, J. Phys. Chem. C, 2010, 114(49), 21427–21433,  DOI:10.1021/jp105895a.
  7. H. Peng, J. Zhang and C. Zeng, et al., One-dimensional synergistic core–shell nanozymes with superior peroxidase-like activity for ultrasensitive colorimetric detection of blood cholesterol, ACS Appl. Bio Mater., 2020, 3(8), 5111–5119,  DOI:10.1021/acsabm.0c00588.
  8. M. Amirzehni, H. Eskandari and B. Vahid, et al., An efficient chemiluminescence system based on mimic CuMOF/Co3O4 nanoparticles composite for the measurement of glucose and cholesterol, Sens. Actuators, B, 2021, 348, 130690,  DOI:10.1016/j.snb.2021.130690.
  9. H. Abdolmohammad-Zadeh and F. Ahmadian, A fluorescent biosensor based on graphene quantum dots/zirconium-based metal-organic framework nanocomposite as a peroxidase mimic for cholesterol monitoring in human serum, Microchem. J., 2021, 164, 106001,  DOI:10.1016/j.microc.2021.106001.
  10. K. Malhotra, D. Hrovat and B. Kumar, et al., Lanthanide-doped upconversion nanoparticles: exploring a treasure trove of nir-mediated emerging applications, ACS Appl. Mater. Interfaces, 2023, 15(2), 2499–2528,  DOI:10.1021/acsami.2c12370.
  11. M. K. Mahata, R. De and K. T. Lee, Near-infrared-triggered upconverting nanoparticles for biomedicine applications, Biomedicines, 2021, 9(7), 756,  DOI:10.3390/biomedicines9070756.
  12. T. H. Ho, C. H. Yang and Z. E. Jiang, et al., NIR-Triggered Generation of Reactive Oxygen Species and Photodynamic Therapy Based on Mesoporous Silica-Coated LiYF4 Upconverting Nanoparticles, Int. J. Mol. Sci., 2022, 23(15), 8757,  DOI:10.3390/ijms23158757.
  13. C. Gao, P. Zheng and Q. Liu, et al., Recent advances of upconversion nanomaterials in the biological field, . Nanomaterials, 2021, 11(10), 2474,  DOI:10.3390/nano11102474.
  14. N. Rakov and G. S. Maciel, Comparative study of Er3+ and Tm3+ co-doped YOF and Y2O3 powders as red spectrally pure upconverters, Opt. Mater., 2013, 35(12), 2372–2375,  DOI:10.1016/j.optmat.2013.06.037.
  15. Y. Zhang, F. Wang and Y. Lang, et al., KMnF3: Yb3+, Er 3+@KMnF3: Yb3+ active-core–active-shell nanoparticles with enhanced red up-conversion fluorescence for polymer-based waveguide amplifiers operating at 650 nm, J. Mater. Chem. C, 2015, 3(38), 9827–9832,  10.1039/C5TC01838B.
  16. J. Xu, D. Yang and W. Han, et al., A novel strategy for markedly enhancing the red upconversion emission in Er3+/Tm3+ cooperated nanoparticles, J. Mater. Chem. C, 2018, 6(28), 7533–7540,  10.1039/C8TC02370K.
  17. D. Giust, M. I. Lucío and A. H. El-Sagheer, et al., Graphene oxide–upconversion nanoparticle based portable sensors for assessing nutritional deficiencies in crops, ACS Nano, 2018, 12(6), 6273–6279,  DOI:10.1021/acsnano.8b03261.
  18. J. Wang, C. Li and Y. Cui, et al., Optimized core-shell lanthanide nanoparticles with ultrabright Ce3+-modulated second near-infrared emission for “lighting” plant, Nanoscale, 2023, 15, 11026–11037,  10.1039/D3NR01970E.
  19. M. S. Arai and A. S. S. de Camargo, Exploring the use of upconversion nanoparticles in chemical and biological sensors: from surface modifications to point-of-care devices, Nanoscale Adv., 2021, 3(18), 5135–5165,  10.1039/D1NA00327E.
  20. X. Li, S. Zhao and B. Li, et al., Advances and perspectives in carbon dot-based fluorescent probes: mechanism, and application, Coord. Chem. Rev., 2021, 431, 213686,  DOI:10.1016/j.ccr.2020.213686.
  21. J. Zhang, R. Zhou and D. Tang, et al., Optically-active nanocrystals for inner filter effect-based fluorescence sensing: achieving better spectral overlap, TrAC, Trends Anal. Chem., 2019, 110, 183–190,  DOI:10.1016/j.trac.2018.11.002.
  22. Q. Chen, R. Sheng and P. Wang, et al., Ultra-sensitive detection of malathion residues using FRET-based upconversion fluorescence sensor in food, Spectrochim. Acta, Part A, 2020, 241, 118654,  DOI:10.1016/j.saa.2020.118654.
  23. Y. Zhang, Y. N. Wang and X. T. Sun, et al., Boron nitride nanosheet/CuS nanocomposites as mimetic peroxidase for sensitive colorimetric detection of cholesterol, Sens. Actuators, B, 2017, 246, 118–126,  DOI:10.1016/j.snb.2017.02.059.
  24. Y. He, X. Niu and L. Shi, et al., Photometric determination of free cholesterol via cholesterol oxidase and carbon nanotube supported Prussian blue as a peroxidase mimic, Microchim. Acta, 2017, 184, 2181–2189,  DOI:10.1007/s00604-017-2235-y.
  25. W. Xiao, Z. Yang and J. Liu, et al., Sensitive cholesterol determination by β-cyclodextrin recognition based on fluorescence enhancement of gold nanoclusters, Microchem. J., 2022, 175, 107125,  DOI:10.1016/j.microc.2021.107125.
  26. A. Mondal and N. R. Jana, Fluorescent detection of cholesterol using β-cyclodextrin functionalized graphene, Chem. Commun., 2012, 48(58), 7316–7318,  10.1039/C2CC33410K.
  27. Q. Sun, S. Fang and Y. Fang, et al., Fluorometric detection of cholesterol based on β-cyclodextrin functionalized carbon quantum dots via competitive host-guest recognition, Talanta, 2017, 167, 513–519,  DOI:10.1016/j.talanta.2017.02.060.

Footnote

Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ra07354h

This journal is © The Royal Society of Chemistry 2024