An alternative method of rice sample preparation for measuring the total content of selected elements, i.e., Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn, by ICP OES was developed. The proposed approach is based on the ultrasound-assisted extraction (USAE) of rice samples in the presence of a small amount of concentrated HNO3. The optimal operating parameters were found using the design of experiments (DOE) approach, and the studied experimental factors were the temperature of the ultrasonic bath (A), the sonication time (B), and the volume of concentrated HNO3 added per 0.5 g of a rice sample (C). Under the optimal conditions of the USAE procedure, i.e., A = 60 °C, B = 16 min and C = 4.0 mL, the rice samples were readily solubilized, and the obtained sample solutions could be analyzed by ICP OES with the simple standard solution calibration (without matrix matching). The analysis of the certified reference material (rice flour, NIST SRM 1568b) confirmed the satisfactory trueness of the USAE-ICP OES method. Additionally, no statistically significant differences between the results obtained for the samples prepared by USAE and open-vessel wet digestion (WD, the reference method) were found. In comparison to the routinely used microwave-assisted digestion and open-vessel digestion, the USAE approach offers lower acid consumption, lower detection limits (LODs) of elements, ranging from 4.0 ng g−1 for Mn to 2.7 µg g−1 for K, and a much shorter time of sample preparation.
Apparently from the literature devoted to the element analysis of rice, two spectrometric methods are commonly applied, i.e., inductively coupled plasma mass spectrometry (ICP MS)2,4–10 and inductively coupled plasma optical emission spectrometry (ICP OES).2,5,8–21 Both the mentioned methods need the rice samples to be digested before the measurements to decompose their organic matrix and quantitatively transfer the elements into the prepared sample solutions. This is commonly made with the aid of various microwave-assisted closed-vessel systems.2,4–9,13–17 In this case, the rice samples are digested under the conditions of very high temperature (up to 200–280 °C) and very high pressure (up to 160–200 bars) and using oxidizing reagents such as concentrated HNO3 alone,6,11 or mixtures containing concentrated HNO3 with concentrated (30%) H2O2,4,5,7,12,15,17 or concentrated HCl (3:
1).16 In addition to this, concentrated H2O2 alone,9 diluted (4.5 mol L−1) HNO3,8 or diluted HNO3, i.e., at 0.1,2 1 or 8 mol L−1,13,14 with concentrated H2O2 were also applied to decrease the usage of HNO3. A mixture of diluted (2%) HClO4 with diluted (4%) H2O2 was also used for this purpose.15 Although quite popular and efficient, this type of wet digestion (WD) requires, however, a quite high economic outlay for the purchase of an appropriate microwave oven and appropriate accessories and spare parts. Moreover, although the microwave-assisted decomposition of rice samples is fast (taking up to a max. 30 min), the vessels used for this process need to be washed every time. The washing procedure should be carried out once or twice, using conditions similar to those employed during the sample decomposition step. This is just to prevent possible memory effects and any uncontrolled contamination of the subsequent samples. In consequence, appropriate washing programs are additionally run, which, however, extends the sample preparation time.
Traditional open-vessel WD is also willingly applied at the step of the sample preparation of rice before the element analysis by ICP OES and ICP MS.10,18–21 In this case, the treatment of the samples at lower temperatures 110–130 °C is, however, longer, i.e., 2 h and more. Considering the reaction mixtures applied to treat the samples, it can be concentrated HNO3 with concentrated HCl (3:
1)18 or concentrated HNO3 with concentrated H2O2,20,21 as well as more complex mixtures of the concentrated reagents, e.g., HNO3, HCl with H2SO4 (ref. 19) or HNO3, HF, HClO4 with HCl.10 Unfortunately, the longer preparation time and the use of the open-vessel system increase the risk of contamination of the analyzed samples and/or the loss of the analytes.
Both of the aforementioned cases result in the necessity of searching for entirely new procedures for sample preparation. The ideal approach would be to make these procedures more environmentally friendly (greenish) by reducing the amount of concentrated and toxic reagents used, as well as by using energy-efficient equipment, or significantly reducing the time required for sample preparation. All these efforts certainly reduce the costs of the analysis itself. In the case of rice samples, due to the sample matrix, such searches for a new alternative to WD procedures are rather uncommon. The studies in which concentrated HNO3 is replaced with its diluted solutions2,8,13,14 or by other reagents (concentrated9 or diluted15) can be partially considered as such, because in all these cases very high temperatures and pressures are needed, which are only ensured by the application of closed-vessel microwave-assisted systems. Therefore, appropriate mineralization devices are required and their use, apart from the generally known benefit – quick decomposition, has its disadvantages such as preparing polytetrafluoroethylene (PTFE) vessels and their covers each time before use.
The literature reports sample preparation procedures based on ultrasound-assisted extraction (USAE), but this is extremely exceptional and rare. The USAE procedure for rice samples with an alkaline (pH > 10) EDTA solution was used to determine the concentration of P, K, Mg, Ca, Zn, Mn, Cu and Mo in rice using ICP OES.16 In another study, an alkaline (pH = 8) solution of pancreatin was taken to treat the rice sample before the determination of As, Cd, Mn and Zn by ICP MS.17 The experiments were carried out at room temperature, and the procedure lasted 20 min. Finally, USAE was also applied to prepare rice samples with concentrated HNO3 before the determination of the concentrations of Cd, As, Pb and Se by GF-AAS.22 However, although the treatment was carried out at 80 °C, the procedure was very long (up to 2 h) and, hence, the authors recommended at the end WD in a reflux system in a digestion block (for 90 min and at 120 °C with concentrated HNO3).22
For all these reasons, the present work was aimed at developing a new sample preparation procedure for rice, omitting completely WD in open- or closed-vessel microwave-assisted systems. Instead, the USAE of rice samples was proposed while the optimal working conditions of this procedure, i.e., the temperature of water filling an ultrasonic bath (A, in °), the sonication time (B, in min), and the volume of a concentrated HNO3 solution added per 0.5 g of rice (C, in mL), were selected based on the design of experiments (DOE) approach with the Box–Behnken response surface design along with the desirability functions (delements) and the overall desirability (D) approach. The multiresponse of the system was the concentration of 11 elements (Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn) determined in the prepared sample solutions. The newly developed ICP OES method with the USAE sample preparation was validated and applied for the analysis of different types of rice.
Considering the experimental material, rice samples, commercially available in local stores and distributed by leading retailers in Poland, were taken. All of them were sold in bulk packages containing 4 to 6 individual 100 g bags. In the case of each rice material, including white rice (WR1, WR2, WR3, WR4, and WR5), basmati rice (BR1 and BR2), jasmine rice (JR1 and JR2), parboiled rice (PBR1 and PBR2) and red rice (RR1 and RR1), 5 individual bags from different bulk packages were taken and combined. 50 g of such prepared laboratory samples of WR, BR, JR, PBR and RR were ground using a Fritsch planetary micro mill, model Pulverisette 7 premium line, with agate grinding balls (fi 15 mm) and an 80 mL grinding bowl. The resulting ground rice material was kept in screwed polypropylene (PP) containers in the dark. In addition, for the whole optimization study of the USAE sample preparation procedure, which was carried out using the response surface design, another WR (WR0) was selected and prepared in the same way as the other rice samples.
To validate the ICP OES methods with the sample preparation by open-vessel WD in a digestion block (the reference method) as well as with USAE (the new, alternative method), a NIST certified reference material (CRM) of rice flour, i.e., SRM 1568b, was applied.
For the WD of the selected rice samples, a SCP Science (Canada) digestion block, model DigiPREP Jr, was used. It was equipped with a timer and a temperature controller.
In the case of the rice samples subjected to the USAE procedure, a Polsonic (Poland), model Sonic 36, 48 L ultrasonic bath with an ultrasonic power of 2 × 900 W at 40 kHz, a temperature controller (0–80 °C) for heating water filling its tank, and a timer (0–30 min) was used. The resulting sample solutions were centrifuged, if necessary, to separate any insoluble remnants of the rice samples, and a Medical Instruments (Poland) centrifuge, model MPW-352, was used for that.
Accordingly, the 0.5 g analytical samples of the test material WR0 were taken and treated with concentrated HNO3 (1.0, 2.5, or 4.0 mL). Then, these sample mixtures were sonicated for 10, 20, or 30 min using the ultrasonic bath with water inside the tank, which was unheated (20 °C) or heated to 40 or 60 °C. The resulting sample mixtures or solutions (in the case of complete solubilization) were topped with water to prepare 20.0 g sample solutions. In the case of incomplete solubilization of the samples, the resulting mixtures were additionally centrifuged for 10 min at 11000 rpm. The collected solutions or supernatants were subjected to the ICP OES analysis on the content of Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn versus the simple working standards solutions (no matching according to the HNO3 concentration in the resulting sample solutions was considered). The corresponding procedural blank solutions were also prepared in 3 replicates for each treatment, analyzed, and considered in the final results.
The blank-corrected mean concentrations of the studied elements were the multiresponse of the Box–Behnken response surface design and were fitted with full quadratic functions that comprised the following terms: linear (A, B, C), square (A2, B2, C2), and 2-way interactions (AB, AC, BC). A general form of the regression model equation, elucidating the concentration of the studied elements versus the parameters of the USAE procedure, was as follows: Celement = β0 + β1A + β2B + β3C + β11A2 + β22B2 + β33C2 + β12AB + β13AC + β23BC, where β0 is the intercept, while β1 − β33 – the coefficients of each term included in this equation. The mentioned intercept, the terms, and their coefficients were established by using the stepwise-selection-of-terms algorithm with α to enter and remove the term equal to 0.15, following the recommendations and tips of the software used, i.e., Minitab 17. The analysis of variance (ANOVA) was used to test the statistical and practical significance of the established response surface regression models as well as their adequacy. The p-values for the regression models and the terms included in these models, as well as the p-values for the lack-of-fit test were considered in this case. In addition, the coefficients of determination (R2, %) were also regarded. The statistically significant regression models, illustrating how the concentration of each element was changed versus the parameters A, B and C of the USAE procedure, were used to find the settings of these parameters that enabled achieving the concentrations of all studied elements that unbiasedly corresponded to their CREFs as obtained by the WD procedure of the WR0 samples (the reference method).
Finally, the individual delement values for each element and the D value for all studied elements enabled us to find the global settings of the parameters A, B and C that provided the concentrations of these elements that did not statistically differ from the respective CREFs. For a given combination of the parameters A, B and C, the delement value was calculated as follows:
Al | Ba | Ca | Cu | Fe | K | Mg | Mn | Na | Sr | Zn | |
---|---|---|---|---|---|---|---|---|---|---|---|
a C CRM: the certified concentration given in the attest. SD: the standard deviation. Fcritical: the critical value of the F-test at α = 0.05, i.e., Fcritical (df1 = 2, df2 = 2) = 19.00. tcritical: the critical value of the t-test at α = 0.05, i.e., tcritical (df = 4) = 2.776. b Since the Fcalculated value was higher than the Fcritical value, the C-test was applied. c The critical value (Ccritical) of the C test was calculated. | |||||||||||
Certified concentrations = CCRM | |||||||||||
Mean, n = 3 | 4.21 | 118 | 2.35 | 7.42 | 1282 | 559 | 19.2 | 19.42 | |||
SD | 0.17 | 2 | 0.08 | 0.22 | 6 | 5 | 0.9 | 0.13 | |||
![]() |
|||||||||||
Concentrations determined using WD | |||||||||||
Mean, n = 3 | 4.27 | 0.151 | 125 | 2.40 | 7.36 | 1250 | 546 | 19.5 | 10.2 | 0.141 | 19.6 |
SD | 0.12 | 0.011 | 4 | 0.07 | 0.32 | 30 | 9 | 0.4 | 0.4 | 0.008 | 0.2 |
![]() |
|||||||||||
Concentrations determined using microwave-assisted WD | |||||||||||
Mean, n = 3 | 4.11 | 0.153 | 116 | 2.39 | 7.43 | 1269 | 531 | 19.3 | 10.6 | 0.140 | 18.5 |
SD | 0.09 | 0.003 | 6 | 0.03 | 0.21 | 11 | 18 | 0.3 | 0.3 | 0.001 | 0.6 |
![]() |
|||||||||||
Concentrations determined using USAE | |||||||||||
Mean, n = 3 | 4.29 | 0.156 | 126 | 2.47 | 7.66 | 1260 | 544 | 19.3 | 10.8 | 0.139 | 19.3 |
SD | 0.09 | 0.005 | 5 | 0.04 | 0.19 | 20 | 11 | 0.5 | 0.4 | 0.002 | 0.5 |
![]() |
|||||||||||
Comparison of the concentrations obtained with WD and the certified concentrations | |||||||||||
F calculated | 2.01 | 6.66 | 1.31 | 2.12 | 29.75 | 3.24 | 5.06 | 2.37 | |||
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | |||
t calculated | +0.499 | +2.665 | +0.815 | −0.268 | −1.376b | −2.187 | +0.528 | +1.307 | |||
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 4.303c | 2.776 | 2.776 | 2.776 | |||
![]() |
|||||||||||
Comparison of the concentrations obtained with microwave-assisted WD and the certified concentrations | |||||||||||
F calculated | 3.57 | 9.00 | 7.11 | 1.10 | 3.36 | 12.96 | 9.00 | 21.30 | |||
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | |||
t calculated | −0.900 | −0.548 | +0.811 | +0.057 | −1.797 | −2.596 | +0.183 | −2.119 | |||
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 4.303c | |||
![]() |
|||||||||||
Comparison of the concentrations obtained with USAE and the certified concentrations | |||||||||||
F calculated | 3.57 | 10.41 | 4.00 | 1.34 | 13.22 | 4.84 | 3.24 | 14.79 | |||
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | |||
t calculated | +0.720 | +2.515 | +2.324 | +1.430 | −1.837 | −2.150 | +0.168 | −0.402 | |||
t calculated | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | ||||
![]() |
|||||||||||
Comparison of the concentrations obtained with USAE and concentrations obtained with WD | |||||||||||
F calculated | 1.78 | 4.84 | 1.56 | 3.06 | 2.84 | 2.25 | 1.49 | 1.56 | 1.00 | 14.91 | 6.25 |
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 |
t calculated | +0.231 | +0.717 | +0.297 | +1.504 | +1.396 | +0.480 | −0.244 | −0.541 | +1.837 | −0.866 | −0.965 |
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 |
![]() |
|||||||||||
Comparison of the concentrations obtained with USAE and concentrations obtained with microwave-assisted WD | |||||||||||
F calculated | 1.00 | 2.78 | 1.44 | 1.78 | 1.22 | 3.31 | 2.68 | 2.78 | 1.78 | 4.00 | 1.44 |
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 |
t calculated | +2.449 | +0.891 | +2.218 | +2.771 | +1.407 | 0.683 | +1.067 | 0.000 | −0.693 | −0.775 | +1.774 |
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 |
In addition, WD in a microwave-assisted closed-vessel system was carried out and accurate results were obtained in reference to the CCRMs of Al, Ca, Cu, Fe, K, Mg, Mn and Zn (see Table 1 and the calculated values of the t- and F-tests applied for statistical comparison).
To find out the working conditions of the USAE procedure, the effect of the selected parameters of this sample preparation procedure, i.e., A, B and C, on the concentrations of Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn determined in rice (the WR0 material was used) was studied using the Box–Behnken response surface design (see the description in Section 2.4). When all planned experiments were carried out and the resulting sample solutions were analyzed by ICP OES, the created data matrix, i.e., the concentrations of the studied elements obtained under different conditions according to the treatments within the Box–Behnken design, was statistically analyzed. At first, it was checked for each element if the SD value of the determined concentrations between the treatments, carried out under different conditions, was higher than the SD values of the concentrations determined for 3 parallel samples prepared at a given treatment. This condition was met for all studied elements; therefore, it was concluded that the multiresponse (=the concentrations of the studied elements) of the Box–Behnken response surface design was influenced by the changes in the parameters selected to develop the USAE sample preparation procedure. In addition, the relative standard deviations (RSDs), evaluated for the concentrations of the elements determined after running the center points treatment, showed that the selected response surface design functioned quite well. The mentioned RSD values were as follows: 6.0% for Al, 3.0% for Ca, 0.7% for Ba, K, Mn and Sr, 1.1% for Cu, 1.7% for Fe, 1.9% for Mg, 4.4% for Na and 1.3% for Zn.
The equations of the established response surface regression models for all studied elements are given in Table 3S.† The p-values of the models were lower than the α = 0.15 level of the stepwise-selection-of-terms algorithm that was used to establish the linear, square, and 2-way interaction terms in the regression equations (p-values were changed from 0.000 for Al and K to 0.061 for Ca). Next to this, it was found that the p-values of the lack-of-fit test were higher than α = 0.15 and varied from 0.166 for Ba to 0.889 for Al, which proved the statistical significance of the regression models. In the case of Mg, the p-value of the regression model established for this element, equal to 0.387, was higher than the mentioned α level. This might indicate that the model for this element is statistically insignificant. However, the p-value of the lack-of-fit test was really high, i.e., 0.448, which did not allow this model to be rejected either. Additionally, considering the R2 values that were changed in the following ranges: 100–80% for Al, K, Na and Zn, 79–60% for Ba, Fe and Mn, and <60% for Ca, Cu, Mg and Sr, it was concluded that the response surface regression models given in Table 3S† reasonably well fitted the collected data. As such, they properly described the variance of the concentrations of the studied elements versus the examined parameters A, B and C of the USAE procedure and, hence, could be used for the development of this procedure based on the optimization of the multiresponse with the aid of the individual delements and the D approach.
Analyzing the terms included in the equations of the response surface regression models, it was noted that the parameter C (the volume of concentrated HNO3 added to the 0.5 g rice sample) was statistically significant for all studied elements and presented as linear terms (for Al, Ba, Fe, K, Mn, Na, Sr and Zn), square terms (Al, Cu, K, Na, Sr and Zn), and 2-way interaction terms (Al, Ba, Ca, Fe and Na). The exception was Mg, for which the only statistically significant term was the interaction between the parameters A and B.
Finally, it was also verified how the residuals between the measured concentrations of the studied elements and the predicted concentrations of these elements based on the response surface regression models behaved. No serious deviations in the normal probability plots were observed for all studied elements. In addition, no trends were confirmed to exist in the scatter plots of the normalized residuals versus the run order. The residuals accompanying the established response surface regression models for the concentrations of the studied elements were uncorrelated and followed the normal distribution.
Al | Ba | Ca | Cu | Fe | K | Mg | Mn | Na | Sr | Zn | |
---|---|---|---|---|---|---|---|---|---|---|---|
a C REF: the concentration obtained by using the reference method. SD: the standard deviation. Fcritical: the critical values of the F-test at α = 0.05, i.e., Fcritical (df1 = 14, df2 = 2) = 19.42, Fcritical (df1 = 2, df2 = 14) = 3.74, and Fcritical (df1 = 2, df2 = 2) = 19.00. tcritical: the critical values of the t-test at α = 0.05, i.e., tcritical (df = 16) = 2.121, and tcritical (df = 4) = 2.776. b Since the Fcalculated value was higher than the Fcritical value, the C-test was applied. c The critical value (Ccritical) of the C test was calculated. | |||||||||||
Target values = CREF | |||||||||||
Mean, n = 3 | 0.498 | 0.451 | 48.5 | 2.04 | 1.35 | 543 | 184 | 9.09 | 4.80 | 0.190 | 14.8 |
SD | 0.038 | 0.020 | 3.3 | 0.14 | 0.14 | 28 | 5 | 0.10 | 0.35 | 0.006 | 0.3 |
![]() |
|||||||||||
Fitted values | |||||||||||
Mean, n = 15 | 0.433 | 0.449 | 45.3 | 2.03 | 1.31 | 558 | 180 | 8.91 | 4.85 | 0.192 | 14.4 |
SD | 0.159 | 0.019 | 3.9 | 0.58 | 0.35 | 15 | 8 | 0.19 | 1.47 | 0.004 | 0.4 |
D | 0.869 | 0.949 | 0.490 | 0.946 | 0.938 | 0.558 | 0.588 | 0.528 | 0.932 | 0.596 | 0.754 |
![]() |
|||||||||||
Model validation values | |||||||||||
Mean, n = 3 | 0.425 | 0.457 | 42.6 | 2.17 | 1.55 | 570 | 186 | 9.04 | 4.50 | 0.195 | 14.5 |
SD | 0.035 | 0.008 | 1.3 | 0.06 | 0.17 | 6 | 2 | 0.08 | 0.25 | 0.004 | 0.2 |
![]() |
|||||||||||
Comparison of fitted values and target values | |||||||||||
F calculated | 12.51 | 1.55 | 1.00 | 12.26 | 4.46 | 4.88 | 1.83 | 2.58 | 12.60 | 3.15 | 1.27 |
F critical | 19.42 | 3.74 | 19.42 | 19.42 | 19.42 | 3.74 | 19.42 | 19.42 | 19.42 | 3.74 | 19.42 |
t calculated | −0.688 | −0.165 | −1.321 | −0.029 | −0.191 | +0.743b | −0.823 | −1.571 | +0.057 | +0.735 | +1.626 |
t critical | 2.121 | 2.121 | 2.121 | 2.121 | 2.121 | 4.218c | 2.121 | 2.121 | 2.121 | 2.121 | 2.121 |
![]() |
|||||||||||
Comparison of model validation values and fitted values | |||||||||||
F calculated | 14.74 | 4.03 | 6.43 | 66.75 | 3.03 | 1.98 | 11.43 | 4.03 | 24.70 | 1.40 | 2.86 |
F critical | 19.42 | 19.42 | 19.42 | 19.42 | 19.42 | 19.42 | 19.42 | 19.42 | 19.422 | 3.74 | 19.42 |
t calculated | +0.111 | −0.920 | −0.732 | −0.534b | −1.492 | +1.151 | −1.652 | −1.495 | +0.812b | −1.553 | −0.544 |
t critical | 2.121 | 2.121 | 2.121 | 2.295c | 2.121 | 2.121 | 2.121 | 2.121 | 2.508c | 2.121 | 2.121 |
![]() |
|||||||||||
Comparison of model validation values and target values | |||||||||||
F calculated | 1.18 | 6.25 | 6.44 | 5.44 | 1.47 | 9.68 | 6.25 | 1.56 | 1.96 | 2.25 | 2.25 |
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 |
t calculated | −2.447 | +0.071 | −1.747 | +0.577 | +0.738 | +2.236 | +1.493 | −0.242 | −0.792 | +0.102 | −0.865 |
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 |
To confirm the correctness of the response surface regression models and the established set of settings of the parameters A, B and C for all studied elements, an independent validation experiment was carried out, in which the WR0 samples were prepared using the USAE procedure and run using the established optimal parameters, i.e., A = 60 °C, B = 16 min and C = 4.0 mL. The prepared sample solutions were then analyzed by ICP OES versus the simple standard solutions, while the determined concentrations of Al, Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn were statistically compared with the fitted concentrations and the CREFs. The results of these two comparisons are given in Table 2. As can be seen, the determined concentrations of the studied elements achieved in the validation experiments did not differ from the fitted concentrations, as predicted using the response surface regression models, and from the CREFs, as obtained by using the reference method. In both cases, the tcalculated and Ccalculated values were lower than the respective critical values of these tests. This confirmed that the developed, based on the response surface design along with the optimization by the individual delelements and D values approach, sample preparation procedure gave accurate (true and precise) results of the multielement analysis of rice by ICP OES. This new analytical method could be used for the faster (∼14-fold) and cheaper (lower electricity consumption) multielement analysis of rice as compared to the method with WD using a hot plate or a digestion block. Considering the microwave-assisted WD procedure, the new USAE sample preparation procedure could also be attractive because it does not require any sophisticated and expensive device such as a microwave oven. It is also easier to use it because dispensable digestion tubes are available, saving the time of analysis and high purity reagents necessary to eventually clean them. In contrast, microwave-assisted WD requires digestion vessels to be especially pre-cleaned using an appropriate temperature/digestion program.
Finally, the new USAE sample preparation procedure along with the ICP OES measurements was also applied to the analysis of the CRM. In this case, the determined concentrations of the studied elements obtained using the new alternative method were compared with the CCRMs included in the certificate (see Table 1). According to the tcalculated values, the new method provided true results; the bias between the determined concentrations and the CCRMs was in the range from −2.7% for Mg to +6.8% for Ca. Similarly, as before, in the case of Ba, Sr and Na, the CRM samples were spiked with known amounts of their single-element standards solutions and subjected to the newly developed method with USAE sample preparation. The determined recoveries of these elements were as follows: 101 ± 1% for Ba, 98.2 ± 1.3% for Sr, and 100 ± 1% for Na; hence, the bias ranged from −1.8% to +1.0%. Finally, when comparing the concentrations of the studied elements in the CRM determined with the new method with the CREFs obtained using the reference method (with the open-vessel as well as the microwave-assisted closed-vessel WD for the sample preparation), the differences between them were also statistically insignificant (see Table 1); the differences in the concentrations of the elements spanned the range from −1.5% for Zn to +5.9% for Na in the case of the WD with the digestion block and from −4.6% for Zn to +1.7 for Cu in the case of the WD with the microwave-assisted closed-vessel system.
Both comparisons made and the obtained statistical measures univocally confirmed that the newly developed ICP OES based method with the alternative to the WD sample preparation by USAE was reliable and could be routinely used for the analysis of rice. As compared to the USAE procedure reported for the determination of Cd, As, Pb and Se by GFAAS,22 the method proposed in the present contribution was much faster, i.e., 16 min versus 120 min, and used lower temperature of water in a tank of the ultrasonic bath, i.e., 60 °C versus 80 °C. In addition, the resulting sample solutions were acidic, which is more convenient for the ICP OES measurements using glass/quartz concentric nebulizers mounted on Scott-type or cyclonic spray chambers. This certainly favorably distinguishes the newly proposed sample preparation procedure in the present work from the studies in which USAE was actually used but the rice samples prior to the ICP OES analyses were treated with an alkaline (pH = 8) pancreatin solution (the determination of As, Cd, Mn and Zn)17 or alkaline (pH > 10) EDTA solution (the determination of P, K, Mg, Ca, Zn, Mn, Cu and Mo).16
Al | Ba | Ca | Cu | Fe | K | Mg | Mn | Na | Sr | Zn | |
---|---|---|---|---|---|---|---|---|---|---|---|
a RSD: the relative standard deviation. Fcritical: the critical value of the F-test at α = 0.05, i.e., Fcritical (df1 = 2, df2 = 2) = 19.00. tcritical: the critical value of the t-test at α = 0.05, i.e., tcritical (df = 4) = 2.776. b Since the Fcalculated value was higher than the Fcritical value, the C-test was applied. c The critical value (Ccritical) of the C-test was calculated. | |||||||||||
WR1 | |||||||||||
Mean, n = 3 | 3.52 | 0.334 | 500 | 1.47 | 2.97 | 1620 | 375 | 11.5 | 14.8 | 0.436 | 8.02 |
%RSD | 3.3 | 0.9 | 1.9 | 1.8 | 7.0 | 1.2 | 2.3 | 3.2 | 1.4 | 1.1 | 1.0 |
![]() |
|||||||||||
WR2 | |||||||||||
Mean, n = 3 | 0.254 | 0.238 | 33.5 | 2.84 | 0.738 | 875 | 109 | 8.31 | 3.62 | 0.119 | 13.6 |
%RSD | 10.6 | 1.7 | 2.1 | 4.1 | 3.4 | 1.5 | 2.5 | 2.8 | 5.6 | 0.8 | 1.4 |
![]() |
|||||||||||
WR3 | |||||||||||
Mean, n = 3 | 0.381 | 0.126 | 40.5 | 2.15 | 1.29 | 712 | 174 | 8.25 | 5.21 | 0.159 | 16.7 |
%RSD | 13.9 | 1.1 | 3.4 | 1.9 | 4.9 | 0.9 | 1.7 | 0.8 | 1.5 | 1.1 | 2.3 |
![]() |
|||||||||||
WR3 subjected to WD | |||||||||||
Mean, n = 3 | 0.361 | 0.117 | 39.2 | 2.09 | 1.40 | 688 | 174 | 8.50 | 4.92 | 0.161 | 16.1 |
%RSD | 10.2 | 6.8 | 3.3 | 2.9 | 8.6 | 2.6 | 2.3 | 2.4 | 4.7 | 3.7 | 1.9 |
F calculated | 2.05 | 64.00 | 1.16 | 2.25 | 4.00 | 6.61 | 1.78 | 11.11 | 8.27 | 9.00 | 1.78 |
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 |
t calculated | +0.536 | +1.579b | +1.179 | +1.441 | −1.420 | +2.152 | 0.000 | −2.055 | +2.063 | −0.548 | +2.078 |
t critical | 2.776 | 4.303c | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 |
![]() |
|||||||||||
WR4 | |||||||||||
Mean, n = 3 | 0.316 | 0.084 | 44.4 | 1.62 | 1.30 | 596 | 111 | 9.35 | 5.39 | 0.238 | 15.5 |
%RSD | 14.0 | 1.0 | 4.5 | 3.5 | 3.9 | 1.4 | 1.5 | 1.9 | 1.6 | 1.0 | 1.3 |
![]() |
|||||||||||
WR5 | |||||||||||
Mean, n = 3 | 0.425 | 0.457 | 42.6 | 2.17 | 1.55 | 574 | 190 | 9.12 | 4.50 | 0.197 | 14.5 |
%RSD | 8.2 | 1.7 | 3.0 | 1.8 | 5.2 | 1.0 | 1.1 | 0.9 | 5.7 | 1.7 | 1.6 |
![]() |
|||||||||||
BR1 | |||||||||||
Mean, n = 3 | 0.535 | 0.083 | 51.1 | 2.58 | 2.28 | 771 | 148 | 7.80 | 17.7 | 0.303 | 18.0 |
%RSD | 13.2 | 1.3 | 3.4 | 2.4 | 2.6 | 2.3 | 3.3 | 1.5 | 4.6 | 3.6 | 1.1 |
![]() |
|||||||||||
BR2 | |||||||||||
Mean, n = 3 | 0.491 | 0.093 | 57.9 | 2.23 | 1.74 | 760 | 129 | 8.01 | 26.9 | 0.364 | 12.0 |
%RSD | 1.0 | 1.1 | 4.8 | 3.1 | 4.0 | 3.0 | 3.1 | 1.7 | 1.5 | 3.8 | 1.7 |
![]() |
|||||||||||
JR1 | |||||||||||
Mean, n = 3 | 0.317 | 0.335 | 38.7 | 1.75 | 0.907 | 581 | 96.4 | 9.65 | 8.43 | 0.156 | 12.5 |
%RSD | 14.3 | 0.8 | 1.8 | 2.9 | 2.0 | 0.9 | 2.6 | 2.3 | 1.9 | 1.3 | 2.7 |
![]() |
|||||||||||
JR2 | |||||||||||
Mean, n = 3 | 0.276 | 0.342 | 40.5 | 1.62 | 1.57 | 557 | 136 | 9.21 | 8.28 | 0.137 | 13.8 |
%RSD | 1.1 | 5.6 | 3.5 | 4.9 | 0.6 | 1.8 | 3.7 | 1.4 | 3.0 | 1.5 | 2.9 |
![]() |
|||||||||||
PBR1 | |||||||||||
Mean, n = 3 | 0.316 | 0.092 | 52.7 | 2.52 | 3.35 | 1970 | 424 | 15.6 | 5.32 | 0.136 | 10.1 |
%RSD | 2.5 | 2.2 | 4.7 | 3.6 | 2.7 | 1.0 | 1.9 | 4.5 | 3.8 | 2.2 | 4.0 |
![]() |
|||||||||||
PBR2 | |||||||||||
Mean, n = 3 | 0.344 | 0.132 | 11.2 | 2.73 | 1.21 | 1620 | 180 | 1.97 | 6.81 | 0.092 | 3.99 |
%RSD | 1.2 | 0.8 | 0.9 | 4.4 | 2.5 | 1.2 | 2.8 | 3.6 | 4.7 | 8.7 | 1.0 |
![]() |
|||||||||||
RR1 | |||||||||||
Mean, n = 3 | 0.560 | 2.64 | 93.6 | 3.01 | 8.17 | 2350 | 756 | 23.9 | 14.2 | 0.526 | 19.9 |
%RSD | 0.9 | 2.3 | 1.3 | 0.7 | 2.6 | 2.1 | 1.2 | 1.7 | 2.1 | 0.6 | 1.5 |
![]() |
|||||||||||
RR2 | |||||||||||
Mean, n = 3 | 0.395 | 2.66 | 103 | 3.01 | 7.79 | 2400 | 769 | 31.1 | 11.0 | 0.571 | 20.6 |
%RSD | 5.1 | 0.8 | 1.0 | 4.7 | 1.8 | 0.8 | 1.3 | 1.3 | 2.7 | 1.4 | 1.5 |
![]() |
|||||||||||
RR2 subjected to WD | |||||||||||
Mean, n = 3 | 0.405 | 2.58 | 104 | 3.08 | 7.68 | 2440 | 760 | 31.4 | 11.2 | 0.566 | 20.4 |
%RSD | 9.1 | 2.8 | 2.1 | 5.8 | 4.7 | 2.0 | 3.2 | 1.9 | 5.4 | 4.1 | 3.9 |
F calculated | 2.72 | 12.96 | 6.25 | 1.65 | 3.59 | 6.25 | 5.76 | 2.25 | 4.00 | 7.62 | 7.11 |
F critical | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 | 19.00 |
t calculated | −0.449 | +1.854 | −0.775 | −0.532 | +0.493 | −1.285 | +0.600 | −0.721 | −0.516 | +0.350 | +0.405 |
t critical | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 | 2.776 |
Max. mean | 3.52 | 2.66 | 500 | 3.01 | 8.17 | 2400 | 769 | 31.1 | 26.9 | 0.571 | 20.6 |
Min. mean | 0.254 | 0.083 | 11.2 | 1.47 | 0.738 | 557 | 96.4 | 1.97 | 3.62 | 0.092 | 3.99 |
Average | 0.625 | 0.586 | 85.4 | 2.28 | 2.68 | 1180 | 277 | 11.8 | 10.2 | 0.264 | 13.8 |
%CV | 140 | 158 | 149 | 24 | 92 | 60 | 86 | 65 | 66 | 61 | 34 |
Max. %RSD | 14.3 | 5.6 | 4.8 | 4.9 | 7.0 | 3.0 | 3.7 | 4.5 | 5.7 | 8.7 | 4.0 |
Min. %RSD | 0.9 | 0.8 | 0.9 | 0.7 | 0.6 | 0.8 | 1.1 | 0.8 | 1.4 | 0.6 | 1.0 |
As can be seen, no statistically significant differences were found between the determined concentrations of all elements in WR3 and RR2 and the respective CREFs. This proved the good trueness of the results obtained with the new method; the bias of the determined concentrations of the studied elements in reference to their CREFs was changed from −7.9% for Fe to +7.7% for Ba in the case of WR3 and from −2.5% for Al to +3.1% for Ba in the case of RR2. The precision of the results, expressed as the RSD, was also acceptable and changed in the following ranges: 0.9–14.3% for Al, 0.8–5.6% for Ba, 0.9–4.8% for Ca, 0.7–4.9% for Cu, 0.6–7.0% for Fe, 0.8–3.0% for K, 1.1–3.7% for Mg, 0.8–4.5% for Mn, 1.4–5.7% for Na and 0.6–8.7% for Sr and 1.0–4.0% for Zn.
The limits of detection (LODs) of the studied elements, assessed for rice (the mass of the samples and the final mass of the sample solutions were considered), and based on the 3SDblank criterion, where SDblank is the SD for the repeatedly prepared and measured procedural blank sample solutions (n = 5), were as follows: 76 ng g−1 for Al, 10 ng g−1 for Ba, 1.8 µg g−1 for Ca, 28 ng g−1 for Cu, 200 ng g−1 for Fe, 2.7 µg g−1 for K, 64 ng g−1 for Mg, 4.0 ng g−1 for Mn, 1.7 µg g−1 for Na, 10 ng g−1 for Sr, and 120 ng g−1 for Zn. These LODs were 1.8–1.9-fold (Al, Ba, Ca, and Zn), 2.1-fold (Fe and Sr), 2.6-fold (Cu), 3.4-fold (Mg), and 4.0-fold (Mn) better than those assessed for the reference method, mostly because the SDblank values obtained for the latter method were higher as a result of higher fluctuations of the background for these elements. Concurrently, the values of the LODs of K and Na obtained with the new method were 30 and 20% higher than the respective LODs of these elements achievable with the reference method. Comparing the LODs obtained with the sample preparation procedure proposed here with those reported by Oliveira et al. (2012),16 who used the USAE treatment of rice samples with an alkaline EDTA solution, except for Ca, Fe and Zn, they were better from 2 times (for Cu, Mg and Mn) to 11 (for Al) and even 18 times (for K).
Considering all rice samples analyzed in this work it was observed that the highest variance in the mean concentrations of the studied elements, expressed as the coefficient of variance (CV), was established for Al (the CV of 140%), Ba (the CV of 158%), Ca (the CV of 149%), Fe (the CV of 92%) and Mg (the CV of 86%). The lowest variance of the mean concentrations of the studied elements in the studied rice samples was noted for Cu (the CV of 24%) and Zn (the CV of 34%). The average concentrations of the studied elements, i.e., 0.625 µg g−1 for Al, 0.586 µg g−1 for Ba, 85.4 µg g−1 for Ca, 2.28 µg g−1 for Cu, 2.68 µg g−1 for Fe, 1180 µg g−1 for K, 277 µg g−1 for Mg, 11.8 µg g−1 for Mn, 10.2 µg g−1 for Na, 0.264 µg g−1 for Sr and 13.8 µg g−1 for Zn, corresponded to the concentrations of these elements reported by other authors.2,15–17 Nevertheless, it was noted that the concentrations of Ba, Ca, Fe, Mg, Mn, Sr and Zn determined in RR were much higher than the mean concentrations of these elements in other types of rice.
The two-side one-way analysis of variance (ANOVA) for the independent groups was applied to check if the concentrations of the elements could be used for the classification/discrimination of the rice samples. The 95% significance level (α = 0.05) was considered in this analysis. Since the variance within five groups of rice, i.e., WR, BR, JR, PBR and RR, was unequal, the ANOVA with Welch correction was selected. Next, the post hoc least-significant difference (LSD) Fisher test was used to look for differences between all possible pairs of the rice groups due to the mean concentration of the studied elements. This comparison gave the possibility of finding 10 such differences for each element. The results of this comparison are given in Table 5S.† It was established that the concentration of Fe provided 6 pairs of rice differing from each other, while the concentrations of Ba and Sr were responsible for 5 such pairs. In addition, RR was established to be the most distinguishable due to the concentration of the studied elements out of all types of rice compared. In this case, 26 differences between RR and other types of rice were found due to the content of the selected elements, i.e., Ba, Fe, K, Mg, Mn, Na, Sr and Zn.