Amanda D.
French
*,
Kirby P.
Hobbs
,
Richard M
Cox
and
Isaac J.
Arnquist
Pacific Northwest National Laboratory, Richland, WA 99352, USA. E-mail: amanda.french@pnnl.gov
First published on 16th October 2024
Interference removal in inductively coupled plasma tandem mass spectrometry (ICP-MS/MS) is strongly dependent on the gas selected for use within the collision/reaction cell. There has been little investigation on the effects that reaction gas impurities may have on the resulting spectra. The reactivity of 60 elements was evaluated using nitric oxide (NO 99.5%) with and without a gas purifier to reduce H2O impurities to <100 pptV. Experiments were performed using V, Ce, Tl and Th to investigate the effects of purified NO at various flowrates (0.22–1.49 mL min−1). Purified NO was shown to significantly mitigate oxy-hydride interferences, improve total ion sensitivity (notable at high gas flows), and shift product distributions advantageously. The reduction in oxy-hydride species results in a product distribution favoring the major expected products, where signals were shown to increase by an order of magnitude. Reduced background and increased signal for the major expected products provides avenues for improving various analytical applications of ICP-MS/MS.
The purity of gases utilized within the CRC is not often discussed beyond a mention in the methods section of ICP-MS related publications. Typically, instrument manufacturers recommend using a minimum gas purity of 99.999% (5N) to achieve the best results when using the CRC.4 However, overall purity may not capture the effects of individual contaminants present at trace levels. For example, a 5N pure gas may behave differently if the 0.001% impurity is O2 as compared to H2O. In high purity gases (e.g., O2, CO2, He, etc.) H2O content can range from 0.2–2.0 ppm depending on the gas and the source. Further, the 5N recommendation, if followed, limits the gases that could be used within the CRC as some gases that may provide advantageous separations cannot be readily sourced at 5N purity (e.g., NO, OCS, SO2, etc.).
There have been limited studies focusing on the effects (both positive and negative) of gas impurities in collision/reaction gases despite early mentions of its significance.5,6 Single quadrupole ICP-MS instruments do not mass filter ions created in the plasma prior to the CRC, which may result in undesirable reactions that produce interferences in the mass spectrum. On CRC single quadrupole instruments, Yamada et al., reported that gas impurities in lower purity H2 (99.99% compared to 99.999%) had a negligible effect on the reduction of argide (e.g. ArH+, ArN+, ArO+) species and suggested that the most prevalent impurity was H2O at concentrations < 100 μg g−1.5 Dexter et al., reported that H2O impurities present in collision gases (He) may be advantageous for the removal of some interferences; thus, controlling the H2O impurities is necessary for reproducible results. The authors also report that H2O impurities in a reactive gas mixture of H2 and He (1.7 mL min−1 H2 + 1.0 mL min−1 He) does not significantly alter the M+ signal intensities at the flow rate used for the 22 elements assessed.6 Consequently, the type of gas used (collision versus reaction) may dictate how gas impurities impact a given analysis.
A comprehensive review on reaction and collision cells by Tanner et al., mentions that gas impurities potentially aid or hinder analyses using reaction gases. The authors state that the amount of impurity present within a gas has an increasingly significant effect as the gas flow increases that ultimately results in an increased number of reactant ion (M+) collisions with the gas and impurities.2 The added collisions slow the ions down within the CRC and ultimately increases the probability that M+ will collide with the gas and the impurity(ies). Added collisions from gas impurities could allow for improved interference removal with collision gases as the ion kinetic energy of the polyatomic ions will be reduced further than the primary M+ ion. Conversely, the reaction gas impurities could also cause a competitive or additional reaction to occur with the impurities resulting in a reduced sensitivity for a given reaction product.
In 2012, commercial inductively coupled plasma tandem mass spectrometers (ICP-MS/MS), with the CRC between two quadrupoles, were released. This provided the ability to mass filter ions prior to entering the CRC using the first quadrupole (Q1) and mass filter ions exiting the CRC with a second quadrupole mass filter (Q2) to better deconvolute ion/gas interactions in the CRC. The added mass filtering limits additional, unwanted reactions of other ions from occurring in the CRC. This capability has provided a significant advancement in the field of ICP-MS analyses, allowing additional interferences to be resolved.
To the authors best knowledge, investigations into collision/reaction gas purity have not been reported using ICP-MS/MS. The additional mass filtering available with ICP-MS/MS allows for conditions within the CRC to be limited to a single m/z interaction with the collision/reaction gas being used and any associated impurities. To address this knowledge gap, we have analyzed a lower purity gas (nitric oxide; NO 99.5%) for use in the CRC and the effects of using a gas purifier to reduce H2O impurities to <100 pptV. A total of 60 elements were assessed for their reactivity with NO with and without the gas purifier in use. The full Q2 mass spectra were collected to determine common interferences. Additional experiments were performed using V, Ce, Tl and Th to investigate the effect of flow rate for the purified gas compared to the non-purified gas. The use of the purifier was found to eliminate several interferences, improve total ion sensitivity, and shift product distributions to favor the primary expected products (M+, MN+, MO+ and MO2+). The use of a gas purifier provides an avenue to utilize gases that are limited to lower purities within the CRC to realize applications that were not previously feasible without the purifier.
All work was carried out using an Agilent 8900 ICP-MS/MS system equipped with an octopole-based CRC. Nitric oxide (NO; 99.5% purity; Advanced Specialty Gases; Reno, NV; Table S1†) was used as the reaction gas. Instrument parameters relating to the CRC are reported in Table S2† for each experiment. A general tuning scheme was used to ensure products would be detected across the entire mass range (6–238 amu). Experiments were performed initially without a gas purifier for the “un-purified” experiments. The specifications sheet for NO from the manufacturer list the H2O content < 20 ppmV. “Purified” NO experiments were performed using an Entegris GateKeeper GPU WX Media gas purifier (GPU 70; Billerica, MA) in-line with the fourth cell gas line. The primary function of this purifier is to remove contaminant H2O (<100 pptV; Table S1†) from the source gas. Similar to previous studies,7–10 Q1 was fixed on a single ion m/z (the reactant ion, M+) and Q2 was scanned from 2 to 275 m/z to observe all possible product species formed within the CRC and acquisition times were set to 0.1 s for each m/z. Multiple isotopes of each element were analyzed, where feasible, to ensure there are no additional spectra interferences for the products analyzed. The percent product formation was calculated by dividing the product produced by the sum of all Q2 species greater than 1000 cps.
To investigate the effect of flow rate, a subset of elements was selected ranging in mass and reactivity (V, Ce, Tl, and Th).10 Based on previous work, these four elements show different levels of reactivity; Tl being unreactive, V being slightly reactive and Ce/Th being more reactive with NO.10 Additionally, these 4 elements span a large mass range to ensure there were no mass bias effects. Each M+ was analyzed in a 5 and 10 ng g−1 single-element standard with a 0.1 s acquisition time. The full Q2 mass spectra (amu 2–275) were collected to determine the change in hydride and oxy-hydride species along with any product distribution shifts. A total of five gas flows (0.00, 0.22, 0.45, 0.89, and 1.49 mL min−1), including no gas, were assessed. These gas flows equate to approximated pressures of 0.00, 0.35, 0.71, 1.4, and 2.3 Pa. The octopole bias voltage (Voct), used to control ion kinetic energy, and deflect voltage (focusing voltage applied after the octopole and prior to Q2) tuning parameters were adjusted with each gas flow (Table S2†) during tuning where Q1 → Q2 were monitored at m/z 205 → 205, 140 → 140 and 140 → 156 for optimal sensitivity. The Voct and deflect settings were kept the same when analyzing a given gas flow with and without the purifier; these parameters were only changed between gas flows for maximum sensitivity.
For all experiments, product percentage was calculated by dividing the product signal by the sum of all major product signals observed (e.g., MO+/∑(M+ + MO+ + MO2+), etc.). Similarly, total ion sensitivity was calculated as the sum of all Q2 species observed, excluding the NO+ signal (a product of a charge transfer reaction).
Lighter elements, Li, Be, and B, show a greater level of sensitivity enhancement when the purifier is used compared to other elements analyzed (Fig. 1). This mass dependent enhancement may be caused by to increased scattering occurring from gas escaping the CRC. Typically, lighter elements are scattered more than heavy elements and a larger amount of gas impurities appears to increase the amount of scattering observed. Assuming the primary impurity is H2O, this effect could be primarily due to collisions outside of the octopole ion guide. Notably, H2O is a notoriously difficult gas to remove from vacuum background, so removal from the reaction gas would lower the overall vacuum background decreasing the probability of a collision outside of the ion guide, ultimately resulting in higher sensitivity.
For 232Th and 238U, it is likely that there are missing products in the spectra because m/z = 275 is the maximum m/z on Q2 for the Agilent 8900. Previous guided ion beam mass spectrometry (GIBMS) studies show the hydrolysis of Th+ to form Th(OD)3+ (m/z = 286) that likely supports the hypothesis that Th+ is forming higher mass products that we cannot detect with current instrument limitations.11 Furthermore, 140Ce+, the lanthanide analog for 232Th+, formed 12.5% 140Ce16O31H3+ at 0.89 mL min−1 (Fig. 2) without the purifier. It is suspected that this shift to higher order products (e.g., ThO3+)12 is greater with the purified gas, as more Th+ collisions can occur with NO rather than the impurities. Further, the actinides have often showed greater reactivity and increased higher order products than other M+.10 While an increase in total ion transmission was observed for both Th and U, the total ion sensitivity (cps ppb−1) for both elements is lower than expected when compared other high mass M+ (e.g., Tl, Pb or Bi).
Several elements (Be, V, Cr, Zr, Mo, Nb, Ru, Ce, La, Pr, and Re) had a larger percentage of MN+ product in the purified NO as compared to the non-purified gas. V, Cr, Mo, and Ru had no detectable MN+ product in the non-purified NO, but MN+ was formed (<2%) when using the purified NO. The presence of impurities (presumably H2O as this is the only impurity removed with the purifier used) may decrease the production of MN+ by providing a more favorable reaction pathway for M+, ultimately allowing the M+ to react with the impurity(ies) rather than form the MN+ product. Additionally, the impurities present in the non-purified gas can cause greater dispersion of the reactant ion kinetic energy, as compared to the purified gas, which can result in a shift of the product distribution. In this case, M+ may retain more kinetic energy (or there is less dispersion in ion energy) which can result in increased MN+ formation.
Several elements show a large shift in the product distribution when using the purified NO that have potential utility for a number of applications. For example, Au+ has relatively low reactivity;7–10 however, with the purifier added, additional AuO+ is formed (Fig. 3). This suggests that Au+ reactivity is suppressed by the impurities within the NO. Alternatively, PtO+ was detectable in the non-purified NO but not with purified NO. Thus, the resulting increase in Pt on-mass sensitivity (M+) was primarily due to the removal of the PtO+ production, likely resulting from Pt+ reacting with the impurities within the NO.
Increased sensitivity on a product used for analysis and a reduction in unwanted products may result in a novel analytical method previously determined unviable. For example, the on-mas sensitivity for Co increased by 12× when using the purifier. When examining the full spectra, it was found that a charge transfer reaction (M+ + NO → NO+ + M) was occurring in the non-purified NO and, when using the purifier, the signal at NO+ decreased by 7× along with a slight decrease in the CoO+ product (∼20%). Thus, purified NO reduced the number of unwanted products (NO+ and CoO+) and on-mass sensitivity was retained. This combination of the increased total ion sensitivity and product distribution shift is not unique to the analytes mentioned here and can provides improved detection limits for many analytical methods (Fig. 3).
There are limited reports for the use of NO in ICP-MS/MS applications,10,13,14 with one paper specifically stating that NO (99.5%) produced excess hydrides that were problematic for the analysis of Pu isotopes.14 Previous work from our group showed the utility of NO (99.5%) for Pu isotopes, despite the potential purity issues.10 Variation in impurities in NO employed across studies may be reason for variations seen in analytical outcomes. Such analytical discrepancies may also be true for other gases of various source purity. Based on the product distribution shifts and the decreased production in hydride species observed when using purified NO, the 239Pu measurement is likely to be improved by using NO with a purifier, which we will investigate in a future study.
When assessing the product distribution for ICP-MS/MS analyses, the largest gain is the improvement in total ion sensitivity and the product distribution shift that can increase sensitivity on a given product. For high resolution CRC-based instruments (e.g., multi-collector (MC)-ICP-MS variations such as the Thermo Neoma or the Nu Sapphire), the impurities in a reaction gas can produce unfavorable products which can cause added interferences in the mass spectrum and greatly reduce the utility of reaction gases. Unlike the quadrupole-based instruments, the MC-ICP-MS variations do not mass filter at 1 amu resolution prior to the CRC. Thus, an increased background can be observed at unexpected masses when there are reactions occurring with gas impurities. Additionally, the higher extraction voltages used in MC-ICP-MS instrument results in more ions being extracted into the mass spectrometer and the use of non-purified gas may result in additional side reactions causing an increase in spectral interferences. Based on the purified NO data collected on the quadrupole ICP-MS/MS, it is predicted that spectral interferences on CRC-containing MC-ICP-MS may suffer from increased backgrounds resulting from reaction gas impurities. Thus, using purified reaction gases with MC-ICP-MS may be vitally important.
For NO specifically, the backgrounds are greatly reduced in the spectra across the entire mass range when using the purified form. The reduction in oxy-hydride species results in a product distribution favoring the major expected products (e.g., M+, MN+, MO+ and MO2+), which alone can increase signal by as much as 40% and significantly improve detection limits. The use of purified NO enables analyses that were not previously obtainable by removing additional background interferences and improving product sensitivity as a result of fewer products being produced.
At low flow rates there are subtle differences observed in product formation which become more pronounced as gas flow/pressure increases (Fig. 4). For example, CeO+ formation is 16% greater with the purified gas compared to non-purified at 0.22 mL min−1, while at the higher flow rate (1.49 mL min−1) CeO+ formation is 38% greater for the purified gas (Fig. 4a). A greater quantity of MOxHy species is formed as gas flow increases both with and without using the purifier (Fig. 4b). Similarly, previous work using selection-ion flow tube mass spectrometry (SIFT-MS), where total pressure in the flow tube is 0.3–0.5 Torr of He with trace reaction gas, has shown that Ce readily reacts with D2O to form both CeO+ and higher order products (CeO+(D2O)1–5) in subsequent collisions.15 The conditions in SIFT-MS are similar to the work presented here where H2O is a trace impurity within the NO gas, except NO is more reactive than He. Nevertheless, at higher flow rates the ions in ICP-MS/MS are expected to thermalize so the reaction with NO likely only proceeds when the reaction is exothermic. Under these conditions, ICP-MS/MS and SIFT-MS are likely very similar, so reactions with a trace impurity do occur. For other M+ that are still reactive with NO, the increased oxy-hydrides at higher flow rates suggest that even at <100 pptV H2O remaining in the NO reaction gas. For some elements (e.g. Ce and Th), the reaction kinetics of M+ with the impurities (primarily expected to be H2O) can still be favored over NO because of the strong H2O dipole moment. Nitric oxide has a weaker dipole moment that H2O; thus, the collision radius of H2O will be much larger than that of NO. Further discussion of the effect of the dipole moment on collision rates can be found elsewhere.16,17 Furthermore, H2O is also favored thermodynamically by ∼1.5 eV for oxidation reactions.
Non-reactive elements like Tl (>99% remains at M+) retain total ion sensitivity across all gas flows used; while the more reactive elements (V, Ce, and Th) lose sensitivity as the gas flow increases (Fig. S1†). When comparing the total ion sensitivity for the lowest (0.22 mL min−1) and highest (1.49 mL min−1) pressure, Tl sensitivity only decreased by ∼1.5× regardless of whether the purifier is used, while Th sensitivity decreased by ∼22× with the purifier and by ∼130× without the purifier (Fig. S1†). The purifier had a greater impact on total ion sensitivity at 1.49 mL min−1 (Fig. 5), while at 0.22 mL min−1 limited improvement was observed. This is in agreement with Tanner et al., 2002 who reported that under multiple-collision conditions (i.e., higher gas flows/pressures) gas purity becomes exceedingly important as trace impurities can impact the observed reactions if the reaction pathways favor the impurity reaction over the reaction gas.2
The number of collisions, and thus the gas pressure, is an important driver of the reactions occurring within the CRC. Previous work with NO (0.22 mL min−1) has shown the collisional probability distribution using the hard sphere model, which does not account for polarizability or dipole moment.10 With each (non-reactive) collision, the kinetic energy (KE) of the reactant ion decreases which effects the probability a given reaction will occur. At low flow rates (fewer collisions) in the ICP-MS/MS the hard sphere model provides a reasonable estimate of the number of collisions and the subsequent energy of each ion after collision. Previous work with CO2 has shown that the Langevin Gioumousis Stevenson (LGS) model (which accounts for polarizability) provides a more accurate explanation of the expected chemistry, especially at lower energies (<1 eV).18 Specifically, Cox et al., found that the collisional probability distribution is bimodal with the LGS model which leads to a higher energy ion population with few collisions and a thermalized ion population from many collisions. As the gas pressure increases the distribution shifts to favors a thermalized distribution. Note that CO2 is not polar, but both H2O and NO are. The Su–Chesnavich trajectory model incorporates the dipole into the LGS model. The polarizability of NO and H2O are 1.7 and 1.45 Å3, respectively. The dipole moments are 0.16 and 1.84 D.16,19 The theoretical probability of collision for H2O will increase at a greater rate than that of NO due to the size and polarizability of H2O. For reference, the trajectory cross sections at 300 K (i.e. thermal distribution) for NO and H2O are 120 and 390 Å2. Therefore, as the NO pressure increases, additional collisions with H2O impurities are expected to be greatly increased. When weighted against prevalence, the advantage of H2O is diminished, but reactions with H2O are ∼1.4 eV more favorable thermodynamically based on differences in neutral bond energies. Furthermore, as seen in the SIFT-ICP-MS reactions, H2O tends to form clusters that further increase contact and subsequent collisions. This is clearly observed in Fig. 4, where the impurity-reacted products exceed that of the NO-reacted products.
Additionally in Fig. 4, there is a large increase in the MO2H+ product for Ce and Th in the purified gas at 1.49 mL min−1, which was not anticipated. Upon closer investigation, there is a potential reaction pathway for the removal of an H2O adduct from the higher order product (MO3H3+) upon additional collisions, resulting in increased MO2H+. Due to the increased number of collisions there would be a higher probability of removing this H2O adduct at the higher flow rates. Another potential pathway could be through the increased formation of MO2+ when the purifier is used, which collides with a H-containing gas impurity, resulting in the MO2H+ formation. Understanding the fundamental chemistry and how the gas impurities effect the chemistry occurring within the CRC is important to interpreting the resulting spectra and ultimately finding ways to mitigate unwanted side reactions and exploit advantageous chemistry.
The gas purifier used in this experiment is anticipated to reduce H2O to <100 pptV, while typical 5N pure gases are specified to have <1 ng g−1 H2O. This difference in specifications may mean that even gases with 5N purity can benefit from using a gas purifier. Further, it is important to note that the reproducibility of ICP-MS/MS results across different systems/labs may vary depending on the purity of the gas being used. This can be especially true when looking at general reactivity experiments and methods comparing separation factors amongst different labs.
Gas purity may be more important in single quadrupole instruments since additional interferences are being directed into the CRC, which suggests that higher resolution ICP-MS instruments that are beginning to utilize CRC technology (e.g., Thermo Neoma, Nu Sapphire) may also be greatly affected by collision/reaction gas purity. These higher resolution instruments are more sensitive to the impurity reactions that may occur within the CRC as ions from the CRC are extracted at a high voltage prior to mass separation.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4an01227e |
This journal is © The Royal Society of Chemistry 2024 |