Esther Pinilla-Peñalverab,
Adrián Esteban-Arranzcd,
Ana M. Contentoa and
Ángel Ríos*a
aDepartment of Analytical Chemistry and Food Technology, University of Castilla-La Mancha, Avenue Camilo José Cela, s/n, 13071 Ciudad Real, Spain. E-mail: Angel.Rios@uclm.es; Esther.Pinilla@uclm.es; AnaMaria.Contento@uclm.es; Tel: +34926295232
bRegional Institute for Applied Chemistry Research, IRICA, Avenue Camilo José Cela, 1, 13071 Ciudad Real, Spain
cDepartment of Chemical Engineering, University of Castilla La-Mancha, Avenue Camilo José Cela, 12, 13071, Ciudad Real, Spain. E-mail: adrian.esteban@ictp.csic.es
dDepartment of Polymeric Nanomaterials and Biomaterials, Polymer Science and Technology Institute (CSIC), Juan de la Cierva 3, 28006, Madrid, Spain
First published on 10th November 2023
The novel fluorescent sensor is proposed in this work based on the competitive interactions between the known bioactive compounds, riboflavin and catechin, which act as guests, and graphene oxide (GO) material that acts as host. Specifically, this proposal is based on an indicator displacement assay which allows the detection of GO and catechin (fluorescence quenching of riboflavin signal by GO and increase in fluorescence by catechin on the GO–riboflavin system). Three GO structures with different lateral dimensions and thickness were synthesized and tested, being able to be the three differentiated depending on the attenuation capacity of the fluorescent signal that each one possesses. The environmental analytical control of GO is more and more important, being this method sensitive and selective in the presence of other potential interferent substances. On the other hand, the other sensing capacity of the sensor also allows the determination of catechin in food samples based on the formation of riboflavin–GO complex. It is a rapid, simple and non-expensive procedure. Thus, the same 2D nanomaterial (GO) is seen to play a double role in this sensing strategy (analyte and analytical tool for the determination of another compound).
Intensive production of nanomaterials at industrial scale, as well as their commercialization in consumer products, such as electronic devices, cosmetics, food and drugs will increase the human and environmental exposure to these nanomaterials. Their release into the environment could presents several risks to human health and ecosystem,3,10 so, it is necessary to evaluate their possible effects. Specifically, GO nanostructures could reach the environment through atmospheric emissions, transport pollution and waste from production and research facilities, contaminating the soil, surface and groundwater, even affecting the organisms present there.11 Therefore, analytical methods are increasingly needed for the characterization and determination of engineered nanomaterials in complex matrices.12 Despite this, some approaches have been developed for the quantification of metallic nanoparticles,13,14 carbon-based nanomaterials15,16 and nanostructured organic materials.17,18 However, to the best of our knowledge, GO-based sensors found in the literature in which GO is used as a quencher do not quantitatively determine the nanomaterial.
Furthermore, catechin belongs to the representative family of flavonoids, specifically flavan-3-ols, and involves four diastereoisomers, being the most common isomer of catechin the (+)-catechin. Favan-3-ol compounds present diverse beneficial effects related to human health based on antioxidative, anti-inflammatory, anti-microbial, anti-cancer, cardioprotective and neuroprotective properties, among many others.19,20 All the properties of catechin arises from its polyphenolic structure, which enables to exhibit a strong antioxidant activity. Some of the natural sources where catechins are commonly found are fresh tea leaves, red wine, cocoa beans, grapes and raspberries.21 As a biologically active compound, this type of flavanol contributes in many of the characteristics that make these natural products essential for the human diet. The content of catechins is an important parameter for the evaluation of many food products quality such as tea or cocoa.22 To date, the determination of catechin content in food samples was mostly carried out based on high-performance liquid chromatography-UV/fluorescent detection (HPLC-UV/FL), gas chromatography-mass spectrometry (GC-MS) and capillary electrophoresis-UV (CE-UV).23–27 However, HPLC-DAD/FL, GC-MS or CE-DAD require complex separation and determination systems and long analysis times. Compared to the above methods, fluorescence technology is highly efficient and easy to operate and it has become an effective method for routine analysis.22
For both reasons mentioned above, a novel turn-off-on fluorescence approach based on the affinity of a fluorescent essential vitamin (B2), riboflavin, for GO surface in comparison to catechin was developed. GO showed to act both as an effective nanoquencher and as a highly selective nanoadsorbent. This fact allowed its quantification, being of great relevance since at present the development of techniques and methodologies for the determination and quantification of nanomaterials is still necessary. Moreover, this study demonstrates the competitive interactions between GO and riboflavin or catechin allowing the flavonoid detection. The applicability of this strategy for both GO and catechin determination in environmental and food matrices was successfully carried out.
To catechin determination, the concentration of riboflavin–lGO nanoplatform was constant and fixed to 1 μM of riboflavin and 126.1 μg mL−1 of lGO (corresponding to maximal quenched riboflavin signal). Specifically, to carry out these experiments 15 μL of riboflavin solution (7.5 μg mL−1, 20 μM), 90 μL of lGO solution (0.42 mg mL−1) and 45 μL MES (0.1 M, pH 6.5) were mixed with aliquots of 150 μL samples-containing different concentrations of catechin, to complete a total volume of 300 μL. Then, the mixtures were allowed to interact for 1 min at room temperature before measuring the emission signals at the same wavelength, exciting at 445 nm. The fluorescence recovery efficiency of free riboflavin, expressed as the ratio between the fluorescence of riboflavin–lGO complex in the presence (I) and absence (I0) of catechin, was used as analytical signal.
Control experiments to verify that the switch-off-on of riboflavin emission was produced by the different affinities between the competitors (riboflavin and catechin) for GO were following the same protocol, replacing GO by deionized water.
Previously, a calibration curve of the free riboflavin emission signal at 524 nm (λexc 368 nm) was prepared between 0.05 and 1.00 μM. Once the previous samples were prepared, a control of them was followed by fluorescent measurements to quantify the free riboflavin. Excess free riboflavin in solution was removed from the complex by the necessary centrifugation processes at 10000 rpm for 7 min. The supernatant was removed and the complex was washed with deionized water. Depending on the complex, this process was repeated so many times until no free riboflavin was detected in the solution.
Finally, the samples were frozen by liquid nitrogen at −77 K and sublimated at 0.005 mbar for 3 days and were subjected to the characterization techniques.
The determination of catechin was carried out in several food matrices, such as beverages (red wine and green tea) and cocoa beans. Different concentrations of catechin between 10.0 and 36.6 μg mL−1 were added to a liquid green tea sample, already prepared and bought in a local supermarket. This sample was previously 1:2 diluted with deionized water. 5 mL of spiked green tea were extracted with 2 mL diethyl ether to get rid of some unknown food additives. This extraction procedure was repeated three times. Finally, aqueous samples were subjected to the analytical process by triplicate.
The red wine sample was spiked at the same concentrations levels as the green tea. The treatment of this sample followed the sample procedure as for the tea samples, except that this sample was previously diluted 3 times.
On the other hand, roasted cocoa sample (Ocumare de la Costa, Venezuela) was treated as follows: the beans were crushed with a mortar until a uniform powder was obtained to achieve a better contact with the extractant mixture. 0.5 g of this powder were treated with 5 mL of an ethanol:water mixture (50:50, v/v), then sonicated for 20 min and, subsequently, the suspension was filtered through a cellulose membrane. The remaining cocoa powder was treated again with 5 mL of the extractant mixture, the suspension was then sonicated and finally filtered. Each one of the obtained extracts was directly subjected to the proposed analytical method and electrophoretic one (see the experimental conditions in ESM.S1† “Quantification of catechin by capillary electrophoresis”) to estimate the catechin concentration in the sample. In addition, the cocoa sample was fortified at 2 concentration levels of catechin (30.24 and 69.72 μg g−1) and the procedure was followed.
Fig. S1B† compiles Raman results of the different materials. Typical and similar spectra have been detected. Two bands are distinguished between 1000 and 2000 cm−1, the D-band at 1351 cm−1, which describes the hybridized carbon disorder that is present in the GO, and the G-band located at 1598 cm−1 that is designated to sp2-carbon atoms. The ratio between their intensities (ID/IG) has already been designated as a parameter to establish their degree of defects.31 Based on D. López-Díaz classification,32 these GOs can be considered as low-defected graphene based-materials because they exhibit ID/IG values between 0.90–0.91, ID/IG-lGO (0.91) = ID/IG-sGO (0.91) ≈ ID/IG-mlGO (0.90). In this sense, lGO and sGO display very similar results, being in agreement with F. Rodrigues et al. outcomes.29 On the other hand, mlGO is characterized by a slightly less defected structure in comparison to the other two GOs.
Scanning transmission electron microscopy (STEM) experiments were carried out to define the morphology and size of the different GOs. Micrographs acquired are compiled in Fig. S1C.† Materials show defined flat structures typical of this kind of materials. The freeze-drying methodology used to dry the GO solution has permitted to obtain a less compacted material. Regarding the size of the materials, sGO flakes displays smaller (400–1200 nm) lateral dimensions than lGO (5–20 μm) and mlGO (5–25 μm). It has been seen that these materials have different thicknesses where that of sGO is 1.1–2.4 nm, that of lGO is 1.2–4.5 nm and that of mlGO is 0.6–0.9 nm.30 This suggests the monolayer structure of mlGO and multilayer (n ≥ 2) of s- and lGO.
Fourier-transform infrared spectroscopy (FT-IR) was used to determine the surface chemistry of these materials. All of them present the same type of oxygen functional groups. Results are depicted in Fig. S1D.† They show a band centered at ∼1725 cm−1 that is related to the stretching vibration of –CO groups typically assigned to the existence of carboxylic groups. Fact that is verified by the presence of a shoulder at ∼3612 cm−1 assigned to free hydroxyl groups.33 In addition, a broad band from 2920 to 3620 cm−1 designated to –OH groups is also found. Additionally, the band centred at ∼1620 cm−1 is related to the aromatic domain, sp2 carbon, of the material that is connected with hydroxyl groups in acidic surfaces.34 On the other hand, the change of hybridization is corroborated by the existence of a sharp band at 1065 cm−1.35 Finally, –C–H groups in aromatic structures can be established because of the existence of bands below 950 cm−1.36
Fluorescence emission and UV-Vis absorbance results of the synthesized materials are plotted in Fig. S1E and S1F,† respectively. The main band of absorbance is located at 230 nm, which is characteristic of π → π* transitions from sp2 carbon domains. Additionally, a shoulder associated with n → π* transitions of free electrons present in the oxygen atoms of their carbonyl and carboxyl surface functional groups, is visible at 300 nm.37 Regarding the fluorescence emission results ranging from 370 to 570 nm, it can be seen that lGO and sGO display similar optical properties, however a bathochromic shift is found for mlGO.
Furthermore, the effect of catechin on the emission of riboflavin was investigated (see ESM.S2 “Evaluation of catechin effect over riboflavin fluorescence” and Fig. S4†). Results revelated that there was a negligible effect on the riboflavin fluorescence intensity as the catechin concentration increases, demonstrating that there was no interaction between both compounds in the absence of GO flakes. Thus, it can be stated that the riboflavin fluorescence recovery was attributed to the replacement of riboflavin molecules for catechin ones on the surface of lGO to form a new complex catechin–GO, leaving riboflavin free in the solution and regaining its strong fluorescence as there is more catechin in the medium.
A kinetic study between riboflavin and each GO structure was also carried out to test the interaction time needed to reach adsorption equilibrium (Fig. S6†). The maximal and constant analytical signal (I0/I) was obtained in less than 3 min.
The interaction time required for the adsorption of catechin on lGO to occur was also investigated. As it can be observed in Fig. S7,† after 1 min of interaction the I/I0 signal remains constant.
GO flakes mostly consist in hexagonal ring-based carbon system bearing sp2 carbon atoms and many sp3 carbons that hold the oxygen-containing functional moieties.29 Therefore, GO structure would enable to accept electrons from donor compounds42 being adsorbed on its surface probably via π–π stacking interactions. Steady-state emission profiles have demonstrated a strong quenching of riboflavin emission when they interact with GO. Furthermore, Fig. 2A shows a slight red shift in the maximum emission of the probe, between 524 and 528 nm. This effect is more promoted with the increment of lGO concentration from 0 to 126.1 μg mL−1, showing that riboflavin molecules are strongly adsorbed at the lGO surface via π–π interactions.6
The fluorescence quenching of the excited-state riboflavin could take place by different possible processes, such as energy transfer or electron transfer. Fig. 2B shows that the emission spectrum of the fluorophore (riboflavin) does not overlap with the absorption band of the acceptor (GO), excluding the possibility of the energy transfer process between both compounds to occur.43 Therefore, the quenching effect that is plausible to occur follows an electronic transfer process.44 The same behaviour on naphthylamide dye with GO was previously reported by Seraj and Rouhani6 and on rhodamine B, eosin and methylene blue with graphite oxide and graphene was reported by Liu et al.44
To confirm the possible interaction between riboflavin dye and the lGO surface, Raman and FT-IR analyses of the different riboflavin–GO complexes (1–3) at different concentrations (126.1, 42.0 and 14.0 of lGO μg mL−1, respectively) were carried out.
ID/IG results show the greatest disorder degree for the complex with more riboflavin adsorbed (ID/IG = 0.98). Earlier publications have also demonstrated the use of Raman spectroscopy as a tool to define the way that different molecules and carbon-based materials interact.45,46 Thus, the Raman shift values of their G band are compiled and shown in Fig. 2C and Table S1.† In this sense, results obtained from these riboflavin–lGO complexes corroborate our previous fluorescence outcomes, since the graphitic-sp2 (G) band of the different complexes shows a visible shift, being maximum for the highest ID/IG value. In contrast, an alteration in the D-band shift between complexes has not been observed.
FT-IR technique has previously been used as a tool to help elucidating the possible interactions between organic molecules and solid surfaces. In this case, Fig. 2D shows the spectrum of the riboflavin dye and its composites formed with the lGO solutions (riboflavin–lGO). The riboflavin spectrum displays two main contributions of –OH functional groups at 3315 and 1240 cm−1. Aliphatic groups are detected by bands at 2932 and 2850 cm−1. The band centred at 1727 cm−1 is attributed to –CO stretching vibration of the isoalloxazine ring. Also, the band assigned to the ring –CN stretching vibration is found at 1644 cm−1 and the band at 1534 cm−1 can be assigned to –CC– group. Additionally, the band at 1062 cm−1 is correlated with –CO stretching vibrations. The lGO complexes present representative bands of the dye, proving its presence on the surface of the material. These bands are more pronounced for the riboflavin–lGO (3) complex. The complexes do not show any new band in respect to the initial lGO spectrum (Fig. S1D†), revealing that covalent interactions can be ruled out. These findings support previous Raman and fluorescence spectroscopy results, showing that hydrophobic or π–π interactions could be the main driving forces for lGO–riboflavin complexes.
The species used in this study are in solution, which means that there is a high possibility of interactions to be produced between the fluorophore and the fluorescence extinctor. The processes by which the observed quenching effect can occur are diverse, such as reactions of excited states, energy transfer, complex formation and collision quenching.47 The molecular contact between the fluorophore and quencher is the responsible for the deactivation of the fluorescent emission. The quenching constant obtained by the Stern–Volmer equation (KSV) indicates the sensitivity of the fluorophore to the deactivator. Greater KSV values promote higher accessibility of one specie to the other.47 As it will be discussed below, KSV values obtained for riboflavin–lGO system are considerably higher than riboflavin–sGO and riboflavin–mlGO systems. This fact confirms that favourable interactions between lGO sheets and riboflavin molecules are obtained, enhancing the quenching efficiency of GO.38 To evaluate the type of quenching mechanism that the developed procedure follows, the measurements of lifetime for riboflavin and riboflavin–lGO system were defined.48,49
Generally, fluorescence decay can give information related to the excited-state of the fluorophore, being very useful to study the energy transfer- or electron transfer-induced fluorescence quenching. The lifetime values of the riboflavin and riboflavin–lGO complex were calculated by fitting the decay profiles with one-exponential, resulting in 5.70 ± 0.06 ns (100% ± 1.1) and 5.23 ± 0.05 ns (100% ± 1.6), respectively. The obtained values of χ2 for both decays were 1.01 and 1.07, close to 1, which indicates the perfection fitting of the curve. These short lifetimes differentiate the fluorescence process from another well-known luminous phenomenon, such as phosphorescence. The molecular contact between species can be produced by two processes: (i) a static quenching, which is based on the formation of a non-fluorescent ground-state complex and (ii) the dynamic quenching, which is a result from the collisions between the quencher agent and fluorophore during the excited state. This last one produces a rapid recovery to its ground-state without the emission of a photon. In the case of a static quenching, the fluorescence lifetime of the fluorophore should not change in the presence of the quencher.50 As shown in Fig. 2E, the fluorescence lifetime obtained for riboflavin was almost unchanged in the presence of lGO suggesting that the fluorescence change of riboflavin did not follow the to the excited state perturbation.51 Thus, it can be concluded that the occurred effect is mainly referred to as static quenching mechanism in which the dye and GO forms a non-fluorescent complex at ground-state,47,52 being the dynamic quenching little significant in the conditions in which the study was carried out.
The value of the binding constant (Ks) for the formation of the riboflavin–lGO complex and the number of binding sites (n) of the nanomaterial are 1.51 × 10−3 mL μg−1 and 1.79 (∼2), respectively.48,49 These results were obtained by the steady-state fluorescence measurements.
As revealed by fluorescence studies, there is greater affinity of catechin for the adsorbent material (lGO) compared to riboflavin, which is attributed to the size difference and the hydrophilic character. Catechin molecules display smaller size than those of riboflavin, being more accessible to interact with the active centers of GO. In addition, catechin has a lower hydrophilic character than riboflavin due to its higher aromaticity, being this characteristic more favorable for the interaction with GO.
As previously mentioned, the fluorescence extinction of riboflavin as dye indicator with increasing concentrations of the different GO structures can be used for quantitative determination of GO. Fig. 3A shows a clear linear relationship between the riboflavin fluorescence ratio in the absence (I0) and presence (I) of lGO concentrations from 0.6 to 35.4 μg mL−1. Stern–Volmer plot of this phenomenon resulted to be I0/I = 0.843 + 0.026 × [lGO] (μg mL−1) with a determination coefficient of 0.996. For sGO and mlGO, a good linear dependence was also found when plotting the analytical signal (I0/I) and the GO concentration. The precision of the proposed method for GO determination was evaluated in terms of repeatability and reproducibility expressed as relative standard deviation (RSD), both as function of concentration and (I0/I) signal. For determining the repeatability, ten independent measurements of lGO standard solutions were carried out consecutively at 14 μg mL−1. RSD values (intra-day conditions, n = 10) were 5.2% as function of concentration and 1.6% as function of I0/I signal. For reproducibility test (inter-day conditions, n = 5), RSD values of 6.7% and 2.1% were found for lGO as function of concentration and I0/I signal, respectively. The sensitivity of the proposed analytical procedure was assessed in terms of limit of detection (LOD) and quantification (LOQ). The LOD value found for lGO was 0.2 μg mL−1, which was experimentally calculated. Table 1 also summarizes the details of the analytical figures of merits for the determination of sGO and mlGO materials. According to the slopes of each of the linear fits (KSV), the higher affinity of riboflavin for lGO (0.026 mL μg−1) than for sGO (0.023 mL μg−1) and for mlGO (0.018 mL μg−1) is noteworthy, confirming all the above. The quenching magnitude (the slope of the fitting curves) is higher than those previously reported for similar species involved in this effect.53
Analyte | Regression equation | Linear range (μg mL−1) | LOD (μg mL−1) | LOQ (μg mL−1) | R2 | Sy/x | RSD% intra-daya | RSD% inter-dayb | ||
---|---|---|---|---|---|---|---|---|---|---|
Concentration | I0/I | Concentration | I0/I | |||||||
a Measurements under intra-day conditions with n = 10 for GO materials and n = 15 for catechin. To evaluate RSD of the method, 14 μg mL−1 were used in the case of lGO and sGO; 30 μg mL−1 for mlGO and 14 μg mL−1 for catechin.b Measurements under inter-day conditions with n = 5 both for GO materials and catechin. | ||||||||||
lGO | I0/I = (0.026 ± 0.001)[lGO] + (0.843 ± 0.009) | 0.6–35.4 | 0.2 | 0.6 | 0.996 | 0.020 | 5.2 | 1.6 | 6.7 | 2.1 |
sGO | I0/I = (0.023 ± 0.001)[sGO] + (0.967 ± 0.013) | 1.2–33.2 | 0.4 | 1.2 | 0.994 | 0.023 | 6.5 | 1.5 | 8.4 | 1.9 |
mlGO | I0/I = (0.018 ± 0.001)[mlGO] + (0.946 ± 0.013) | 3.8–60.1 | 1.3 | 3.8 | 0.997 | 0.023 | 5.6 | 1.9 | 8.8 | 2.5 |
Catechin | I/I0 = (0.172 ± 0.003)[catechin] + (1.384 ± 0.030) | 1.0–20.8 | 0.2 | 1.0 | 0.998 | 0.056 | 3.1 | 1.2 | 5.8 | 2.3 |
To further evaluate the performance of the riboflavin-lGO probe system for catechin detection, the fluorescence signal of free riboflavin (displaced from the complex) was investigated at different catechin concentrations under the optimized conditions. As shown in Fig. 3B, the probe system gradually begins to fluoresce again as the catechin concentration increased. In the wide range of catechin concentrations evaluated (0.2–125.1 μg mL−1), a first linear range with a steeper slope that comprised concentrations between 1.0 and 20.8 μg mL−1 and a coefficient of determination of 0.998 was found. The fluorescence recovery efficiency of riboflavin in the presence and absence of catechin (I/I0) against the target analyte concentrations is represented in Fig. 3B inset. Meanwhile, for concentrations that exceeded 20.8 μg mL−1 and up to 108.3 μg mL−1, a second linear response with a lower slope was observed with a value of 0.993 as determination coefficient. Being a stronger competitor, initially at lower catechin concentrations it should be easier for riboflavin to be released from the complex, hence the difference in slopes. Concentrations higher than 108.3 μg mL−1 did not produce further fluorescence recovery. As is recommended to work in the first linear range of concentrations, the quantification of the samples was carried out in this region. The regression equation of the first linear fit was I/I0 = 1.384 + 0.172 × [catechin] (μg mL−1). The sensing platform begins to recognize catechin at 0.2 μg mL−1, therefore this value was considered as the experimentally calculated LOD. Concentrations between 0.2 and 1.0 μg mL−1 were perfectly detected but not quantifiable. In this case, the repeatability (intra-day conditions) was evaluated though fifteen analyses of 14 μg mL−1 standard solution and an RSD value of 3.1% in concentration terms was found. Reproducibility (inter-day conditions) was evaluated by quintuplicate at the same catechin concentration, obtaining an RSD value of 5.8%. These results suggest that the fluorescence probe system can be easily applied for the quantitative determination of catechin and with good sensitivity.
The details corresponding to the analytical figures of merits for both sensing systems are summarized in Table 1. It should be noted that the proposed sensing systems, both for the detection of GO and catechin, have attractive qualities such as their sensitivity, the simplicity of the analysis in a very short time and are relatively environmentally friendly sensors (employing a vitamin as sensor probe). This analysis strategy allows simultaneous detection and offers a new approach to sensing other compounds of the same nature.
The selectivity study of the proposed method was carried out by means of an interference test whose results appear in ESM.S2 and Fig. S8.†
The usefulness of the developed sensing strategy for GO analysis was proved by its application to different environmental samples (waters from different nature and a soil) and in the case of catechin, food samples (cocoa beans and beverages) were used. Each sample was treated according to the procedure already described in section “Samples treatment” and subsequently submitted to the analytical procedure detailed in section “Analytical procedure for the analysis of GO and catechin”. All samples were analysed in triplicate by the proposed fluorometric assay.
The different water samples were fortified at four concentration levels of lGO between 1.1 and 4.0 μg mL−1 and the soil sample was also doped with the nanomaterial in the range 23.8–88.2 μg g−1. Previously, all the matrices were checked not showing interference in the analytical signal. The concentration of lGO in each sample was estimated by external calibration and the obtained results are summarized in Table 2. As can be observed, satisfactory results were obtained for all of them with recovery values of the added GO between 96 and 103% and from 97 to 99% for water and soil samples, respectively. Furthermore, RSD values (n = 3) were lower than 4% for all environmental samples evaluated. The fluorescence turn-off effect produced by lGO in well water matrix can be seen in Fig. 4A and that produced in the other environmental samples can be observed in Fig. S9.† In Fig. S9C,† it can be appreciated a lower fluorescence of the probe in the presence of seawater matrix (blank assay) compared to that observed for other water matrixes, probably due its higher salinity. However, the high ionic strength of the sample does not affect the gradual response of the analytical signal in the presence of increasing concentrations of the analyte to be detected. Therefore, this aspect is not an issue and this sample was not been pre-treated before analysis.
Determination of lGO | |||||
---|---|---|---|---|---|
Type of sample | Sample | Spiked concentration | Found concentrationa | Recovery (n = 3, %) | RSD (n = 3, %) |
a Average of 3 independent spiked samples ± standard deviation. | |||||
River water (μg mL−1) | 1 | 1.05 | 1.02 ± 0.03 | 97.1 ± 2.9 | 3.0 |
2 | 1.89 | 1.82 ± 0.05 | 96.0 ± 2.7 | 2.8 | |
3 | 2.94 | 2.92 ± 0.09 | 99.2 ± 3.0 | 3.0 | |
4 | 3.99 | 4.01 ± 0.12 | 100.5 ± 2.9 | 3.1 | |
Well water (μg mL−1) | 1 | 1.05 | 1.05 ± 0.03 | 99.8 ± 3.2 | 3.2 |
2 | 1.89 | 1.82 ± 0.06 | 96.2 ± 3.0 | 3.1 | |
3 | 2.94 | 2.85 ± 0.08 | 97.1 ± 2.7 | 2.8 | |
4 | 3.99 | 3.86 ± 0.13 | 96.8 ± 3.2 | 3.3 | |
Drinking water (μg mL−1) | 1 | 1.05 | 1.07 ± 0.03 | 101.8 ± 2.8 | 2.7 |
2 | 1.89 | 1.84 ± 0.05 | 97.3 ± 2.5 | 2.5 | |
3 | 2.94 | 2.91 ± 0.08 | 99.1 ± 2.7 | 2.7 | |
4 | 3.99 | 3.93 ± 0.11 | 98.5 ± 2.8 | 2.9 | |
Seawater (μg mL−1) | 1 | 1.05 | 1.04 ± 0.03 | 98.8 ± 3.1 | 3.2 |
2 | 1.89 | 1.84 ± 0.05 | 97.1 ± 2.5 | 2.6 | |
3 | 2.94 | 3.02 ± 0.08 | 102.6 ± 2.8 | 2.7 | |
4 | 3.99 | 4.00 ± 0.12 | 100.2 ± 3.1 | 3.1 | |
Soil (μg g−1) | 1 | 23.80 | 23.54 ± 0.73 | 98.9 ± 3.0 | 3.1 |
2 | 40.25 | 39.97 ± 1.01 | 98.4 ± 2.5 | 2.5 | |
3 | 64.40 | 62.65 ± 2.29 | 97.3 ± 3.6 | 3.7 | |
4 | 88.20 | 86.17 ± 2.68 | 97.7 ± 3.0 | 3.1 |
On the other hand, the determination of catechin in liquid food samples was carried out by adding the target analyte to a green tea and a red wine sample in the concentration range 10.0–36.6 μg mL−1. In this case and for both types of samples, an alteration in the analytical signal was observed by the corresponding matrix. As can be observed in Table 3, the recovery values obtained were quite favorable for both types of samples, especially for those with lower concentrations of catechin. At higher concentrations of this analyte in the sample, the catechin concentrations found are increasingly lower than the added ones. This may be since other compounds present in the matrix have a slight affinity to the riboflavin–GO complex and can be adsorbed, competing with catechin molecules (possible interferers). Even so, recovery values between 95 and 99% were found for red wine samples and for green tea ones were between 94 and 102% with low RSD values (≤4%) for all of them. Fig. S10† shows the turn-on effect produced in the red wine and green tea samples.
Determination of (+)-catechin | |||||
---|---|---|---|---|---|
Type of sample | Sample | Spiked concentration | Found concentrationa | Recovery (n = 3, %) | RSD (n = 3, %) |
a Average of 3 independent spiked samples ± standard deviation. | |||||
Red wine (μg mL−1) | 1 | 10.00 | 9.92 ± 0.32 | 99.2 ± 3.2 | 3.2 |
2 | 18.32 | 18.15 ± 0.55 | 99.1 ± 3.0 | 3.0 | |
3 | 25.00 | 24.48 ± 0.68 | 97.9 ± 2.7 | 2.8 | |
4 | 36.60 | 34.97 ± 1.05 | 95.6 ± 2.9 | 3.0 | |
Green tea (μg mL−1) | 1 | 10.00 | 10.19 ± 0.34 | 101.9 ± 3.4 | 3.3 |
2 | 18.32 | 18.25 ± 0.51 | 99.6 ± 2.8 | 2.8 | |
3 | 25.00 | 23.76 ± 0.77 | 95.0 ± 3.1 | 3.2 | |
4 | 36.60 | 34.47 ± 1.28 | 94.2 ± 3.5 | 3.7 |
In both GO and catechin sensing (Fig. S9 and S10†), a blank assay was always performed to evaluate the effect of the matrix on the fluorescent signal obtained. Thus, if there is a matrix effect, it can be easily subtracted.
The determination of catechin was also performed in a cocoa beans sample, which endogenously contains the analyte. Moreover, cocoa beans sample was spiked at two concentration levels of catechin. Recovery values were around 98% as summarized in Table 4, which ensures the accuracy of the proposed method. Fig. 4B presents the results found for the determination of catechin in the cocoa matrix by the proposed procedure. As a further validation study and with the aim to prove the reliability of the proposed catechin sensor, the cocoa sample was also submitted to a previously validated electrophoretic method,27 slightly modified, to estimate the flavonoid concentration. The analyses were performed by five-fold. The experimental conditions of the modified electrophoretic method appear in section ESM.S1† “Quantification of catechin by capillary electrophoresis”. The analyses were performed by five-fold and Table 4 summarizes the obtained concentration values of catechin in cocoa sample by the proposed fluorometric procedure and the capillary electrophoresis (CE) method. Once the cocoa results were obtained, Student-t test was performed between these two sets of data to verify the statistical accordance between the results found by both methodologies.54 The values for texp and tcrit were 1.0 and 2.8, respectively. As the value of texp is lower than that of tcrit, the absence of significant differences between the results obtained by both methodologies can be guaranteed, which would prove its comparative accuracy at the checked level of confidence (95%). These results demonstrated the potential applicability of the fluorometric probing system for the detection of catechin in cocoa samples, even in the presence of other co-existing compounds in the same matrix. The difference between the analysis time required for each of the procedures should be highlighted, being much shorter, in the order of 10 times, for the fluorometric sensor.
Sample | Original amount | Spiked concentration | Found concentration | Recovery (n = 3, %) | RSD (n = 3, %) | |
---|---|---|---|---|---|---|
CE-DAD (n = 5) | This method (n = 5) | |||||
Cocoa beans (μg g−1) | 92.4 ± 4.3 | 103.5 ± 3.3 | 30.24 | 131.37 ± 4.20 | 98.3 ± 3.1 | 3.2 |
69.72 | 170.07 ± 5.96 | 98.2 ± 3.4 | 3.5 |
Herein, riboflavin emission was gradually turned-off as the fluorophore interacted with GO flakes by π–π stacking interactions. The strongest interaction was produced between the dye and lGO, followed by sGO and lastly with mlGO. Therefore, the former was selected based on the least amount of nanomaterial needed for the complex production and the highest quenching effect. In this sense, it has been stablished that the lateral size and the thickness of these carbon-based nanostructures are very important physicochemical parameters for the good performance of this type of novel sensors. This effect occurred for the three GO types evaluated, allowing the quantitative analysis of this type of nanomaterial. Detection limits of GO structures were 0.2, 0.4 and 1.3 μg mL−1, for lGO, sGO and mlGO, respectively. Once the riboflavin-GO complex was formed at the maximum fluorescence extinction, the presence of catechin (as stronger competitor) was capable to replace riboflavin (as weaker competitor) from the complex, switching-on the fluorescence signal due to the competitive adsorption between catechin and riboflavin for the GO active sites. Catechin quantification range was wide from 1.0 to 20.8 μg mL−1 and the limit of detection resulted to be 0.2 μg mL−1. Compared to traditional instrumentation (HPLC, GC, or CE) for catechin analysis, the advantage of this developed procedure was based on short analysis times, economy, easy to operate, no tedious sample pre-treatment required, and to be an eco-friendly strategy. The novel developed sensor was also applied to real samples with satisfactory results, demonstrating its potential applicability. This dual determination significantly contributes to the first (GO as analyte) and second (GO as nanotool) way of Analytical Nanoscience and Nanotechnology.
Furthermore, the detection platform is a potential assay principle for the detection of other interesting organic molecules and biomolecules due to the different adsorption capacity of multiple fluorophores and analytes with different molecular structures.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ra04726a |