Aku
Lampinen
a,
Erich
See
a,
Aleksei
Emelianov
a,
Pasi
Myllyperkiö
a,
Andreas
Johansson
ab and
Mika
Pettersson
*a
aNanoscience centre, Department of Chemistry, University of Jyväskylä, Survontie 9, Jyväskylä 40500, Finland. E-mail: mika.j.pettersson@jyu.fi
bNanoscience centre, Department of Physics, University of Jyväskylä, Survontie 9, Jyväskylä 40500, Finland
First published on 29th March 2023
Here we demonstrate, using pulsed femtosecond laser-induced two-photon oxidation (2PO), a novel method of locally tuning the sensitivity of solution gated graphene field-effect transistors (GFETs) without sacrificing the integrity of the carbon network of chemical vapor deposition (CVD) grown graphene. The achieved sensitivity with 2PO was (25 ± 2) mV pH−1 in BIS-TRIS propane HCl (BTPH) buffer solution, when the oxidation level corresponded to the Raman peak intensity ratio I(D)/I(G) of 3.58. Sensitivity of non-oxidized, residual PMMA contaminated GFETs was 20–22 mV pH−1. The sensitivity decreased initially by 2PO to (19 ± 2) mV pH−1 (I(D)/I(G) = 0.64), presumably due to PMMA residue removal by laser irradiation. 2PO results in local control of functionalization of the CVD-grown graphene with oxygen-containing chemical groups enhancing the performance of the GFET devices. The GFET devices were made HDMI compatible to enable easy coupling with external devices for enhancing their applicability.
Traditional potentiometric pH meters (e.g., a glass pH electrode) are quite large, rigid, must frequently be refilled with solution, and must be stored wet. Failure to properly maintain the solution and wet-storage of these devices can lead to permanent damage. In contrast, graphene field-effect transistor (GFET) pH sensors can be stored dry, be manufactured at the micron scale, and be flexible, allowing them to be used in more complex or difficult-to-reach situations, such as biological or microfluidic systems, and also allowing for measuring local pH with micrometre spatial resolution. Additionally, the lack of a wet-storage requirement means they can be more efficiently packed, shipped, stored, and disposed of after use, making them more viable for tasks such as wound monitoring,7 where re-use of the device with another patient is not desired due to close contact with biological materials.
One of the most commonly used methods of measuring pH with GFETs is using the Dirac point (UDirac).14 The Dirac point is at a gate voltage where the number of electrons and holes of the conduction channel of a FET is balanced and the conductance is at its minimum (i.e. the resistance is at its maximum). The pH change i.e., the change in the concentration of H3O+ and OH− ions in the solution alters the doping of the graphene channel.15 This is seen as a shift of the UDirac, which is typically linearly dependent3,9,10,12,16 on the pH and is reported in the unit of how much the point shifts as a function of the pH (mV pH−1).14
In general, pristine graphene has a relatively poor pH response in comparison to the large commercial pH meters. The reported values vary a lot, but some reported values include 4.2 mV pH−13 6 mV pH−117 22 mV pH−118 and 24 mV pH−15,19 all for nominally-pristine chemical vapor deposition (CVD) grown and reportedly “clean” graphene. Still, it is also often stated that truly clean and pristine monolayer GFETs should not be sensitive to pH change at all.14,17 This is, in part, due to the fact that pristine graphene is hydrophobic, and also lacks functional groups that allow interaction with the oxonium and hydroxyl ions in a solution.14 To address this, several modifications have been made to improve the sensitivity of GFET-based pH sensors, including surface functionalization (e.g., oxygen groups,7 phenol,13 or polyaniline20), utilizing anodized graphene, and altering the physical structure of the GFET.3,4 Many of these techniques have resulted in the creation of oxygenated functional groups, which improve the pH sensitivity of the material.
Tan et al.3 discussed a method of plasma-etching of an existing graphene sheet to create strips of graphene nanoribbons with graphene oxide (GO) edges. This resulted in a significant increase in the sensitivity, boosting it to 24.6 mV pH−1 from the original 4.2 mV pH−1. Other methods include using reduced graphene oxide (rGO), three-dimensional few-to-multilayer graphene or the deposition of various functionalizing layers onto graphene (e.g. Al-oxide17).
These methods of modifying the graphene surface to improve pH sensitivity have various drawbacks. The method outlined in Tan et al.,3 for example, requires removing large portions of graphene from the device to create the graphene ribbons, greatly reducing its structural integrity. This could reduce the viability of the device and its lifetime in applications where it may be subject to physical deformation and stress, such as in-vivo implants. Other methods of oxidation (or reduction of GO to create rGO) can be difficult to control precisely and incrementally or utilize methods that involve by-products whose residues can alter pH sensitivity temporarily.14 Additionally, methods that utilize UV-light21 or plasma treatment22 to achieve incremental control, apply the oxidation to the whole substrate or require a mask.
Our group has previously developed a method for oxidizing pristine graphene in a controlled, precise manner via femtosecond laser-induced two-photon oxidation (2PO).23,24 Our previously published23 X-ray photoelectron spectroscopy (XPS) characterization of the laser-oxidized areas has shown the functionalization by primarily hydroxyl (–OH) and epoxide (C–O–C) groups selectively in these areas,23 which have been shown to improve pH sensitivity in graphene oxide.3,14 Therefore, laser-oxidation is potentially a good method for enhancement of the sensitivity of GFET devices in pH measurements. The largest shift in the Dirac point location and therefore the highest sensitivity should be achieved in pH around the pKa value of the hydroxyl group and the opening/closing of the epoxide group. In literature the reported pKa value is 9.32 ± 0.02 for the hydroxyl groups25 and the epoxide opens between pH 7.0 and 11.5 in GO.26 Therefore, the best sensitivity should be found in the pH range above 7.
In this work, we utilized 2PO to functionalize graphene in GFET-based pH sensors with –OH and C–O–C groups for improving the sensitivity compared to pristine graphene in a controlled manner. With the results presented here, we demonstrate a simple method of tuning the sensitivity of GFET pH sensors without subjecting the graphene or sensor to destructive techniques or methods that compromise the structural integrity.
A self-written LabView program was used to control the measurement devices. During measurements, a bias voltage (UDS) of around 0.2 V was used in order to limit the current through the graphene to non-destructive levels. This was calculated using a series resistor and the actual sample as a potential divider. The gate electrode used was a commercial flexible Ag/AgCl dri-ref reference electrode FLEXREF (WPI) and the leakage current through it was monitored so that it was possible to subtract it from the measured drain-source current. Usually, the leakage current was in the range of nA. The gate electrode was held in place by the solution chamber so that the distance from the gate to the GFET was a constant, approximately 1.5 cm. A schematic and illustration of the SG-FET measurement geometry are presented in Fig. 2(a) and (b).
Fig. 2 Illustrations of (a) the used measurement setup, (b) GFET measurement geometry (c) the in-house built solution chamber, and (d) an optical image of the said chamber. |
The pH dependency of the devices was determined by placing the sample (with 5 devices) into an in-house built solution chamber with a volume of roughly 2 ml (Fig. 2(c) and (d)). The sample was left to soak in a buffer solution for approximately 25 hours before starting the measurements, as it has been reported28 to reduce the amount of drift when there are PMMA residues present on the graphene by countering the doping caused by the residues. The solution was changed by pumping the buffer from a separate container. This way the sample was never dry during measurements, so the number of wet-dry cycles the sample experienced was reduced. This in turn reduced the risk of breaking the sample and prevented the deposition of salt on the graphene due to evaporation. The pump used was an in-house built peristaltic pump, connected to the chamber via standard silicone tubing. The measurements were always started from basic conditions (pH 9.50), going to neutral (pH 6.88) and then coming back to the basic conditions to see how stable the sample was and whether there was any drift or hysteresis present in the response.
The pH measurement data was processed automatically with a custom Python script. From the raw drain-source current (IDS) and potential (UDS) data, the resistance (RDS) was calculated. To automatically find the Dirac point location (UDirac), a Savitzky-Golay filter smoothening function was used. After the smoothening, the maximum resistance value location was used to determine the UDirac.
After Raman characterization, the devices had their pH response measured. Representative measurement data for a pristine and an oxidized GFET sample are presented in Fig. 4. See the ESI† for a summary of all of the measurement data used in this paper (Fig. S4, ESI†) and all the corresponding raw data (Table S3, ESI†). With the shown pristine graphene device, two random current spikes were detected, but they do not affect the analysis. They were most likely due to movement of wires in the measurement setup during the measurement. Also, the initial measurement in pH 9.50 has been left out of the data as it was clear that the device had not yet stabilized, and the single measurement point was an outlier. The data shown here is for the exact same device before and after laser-oxidation. Some of the samples were characterized in both pristine and oxidized state, and this sample history is shown in Table S2 (ESI†).
Fig. 4 Representative measurement data for a single pH measurement for (a) pristine (I(D)/I(G) = 0.06) and (b) highly oxidized (I(D)/I(G) = 1.45) samples. Both include the measured drain-source current that has been normalized by removing the detected leakage current (IDS–IG, top left), measured drain-source potential (UDS, top right), smoothened drain-source resistance that is calculated from IDS and UDS (bottom left) and the found locations of Dirac points (UDirac loc. at UG) for each measurement in each pH measured (bottom right). The errors of the bottom right panels of both figures are of the magnitude of the size of the dots. See ESI† (Fig. S5) for more precise fitting data and directionality of the of the pH measurements. |
When looking at the location of the Dirac point for the pristine device, we can see that it was not around 0 V. This was most likely due to the final EBL step leaving residual PMMA onto the graphene and the SiO2 substrate. It has been reported28 that PMMA residues and SiO2 substrate cause p-type doping and therefore shift the Dirac point. Others have observed this to sometimes be a continuous, approximately 1 nm thick layer that had significant effect on the sensory function of graphene.30
It can be seen from Fig. 4 that the 2PO treatment does not always cause significant additional p-type doping compared to nominally pristine graphene, as the Dirac point is almost at the same location with both non-oxidized (I(D)/I(G) = 0.06) and oxidized (I(D)/I(G) = 1.45) devices when the solution is around pH 7. This lack of shifting could be due to the initial p-type doping being high because of the PMMA residues on the sample. However, with the sample of I(D)/I(G) ratio 0.10 2PO seems to increase amount of p-type doping on the sample. When it was oxidized to 3.46 I(D)/I(G) ratio, the Dirac point location shifts approximately 7–8 mV when in neutral pH. Therefore, the additional doping caused by 2PO seems to have been dependent on the initial doping level. In addition to the shift, some broadening of the transfer curve is visible, when the oxidation level increased. The transconductance of the GFETs increases with the oxidation level, but after an initial increase the charge carrier mobility decreases. (See ESI† for plots, Fig. S2) This would indicate that we are somehow removing impurities from the graphene with low levels of 2PO. Based on the I(D)/I(D′) ratio, introduction of sp3 groups caused the broadening when the oxidation level was low (see Fig. S3b, ESI†)).31–33 At high doses the broadening was caused by the formation of vacancies.31–33 We did not observe a significant degree of hysteresis when measuring the response to gate voltage sweeps at a single pH with pristine or oxidized devices.
The sensitivity dependence on the I(D)/I(G) ratio is shown in Fig. 5. The data shows that 2PO changes the pH sensitivity of GFET devices and, in particular, the sensitivity increases for higher oxidation levels. Alternate plots for correlating the sensitivity with other Raman parameters are shown in the ESI† (Fig. S3). The sensitivity increased as the oxidation level rose. However, the behaviour was not monotonic. Starting from the nominally pristine samples, the sensitivity first decreased. We believe that this was because the devices had some PMMA residues from the final EBL step, similarly to previous reports.30 The presence of PMMA had an effect similar to the laser-functionalization, making graphene more sensitive to pH. When the PMMA contaminated graphene was irradiated with the laser, it was possible that the polymer is broken and therefore the graphene was partially cleaned.34 The lowest sensitivity was reached at the irradiation level corresponding to the I(D)/I(G) of 0.64, after which the sensitivity started increasing with increasing irradiation dose. The sensitivity reached a plateau around I(D)/I(G) of 1.5 and did not change much after that. In this range, the sensitivity increased from ∼19 mV pH−1 to ∼25 mV pH−1, i.e., by about 32%. The fit presented in Fig. 5 is calculated starting from the sample with the lowest oxidation dose used to illustrate the trend of the effect that the oxidation has on the sensitivity. The functional form of the fit had no physical significance. When comparing these results to other published methods, it is important to note that it has been reported35 that the absolute value of the sensitivity achieved is dependent on the ionic species present in the solution. Additionally, it is known that comparing GFET pH sensitivities between publications is difficult.14,17,35
The main result of Fig. 5 is that 2PO could be used to tune and improve the sensitivity of GFETs. Although the improvement in the sensitivity compared to nominally pristine graphene was modest, about 14–25%, it would presumably be much larger for truly clean graphene, as shown by the 32% increase between mildly oxidized and strongly oxidized graphene. One of the advantages of the 2PO method for sensitivity tuning is that it enables optimization of the conditions for not only pH sensing but also simultaneously for other properties, such as functionalization of the GFET for immobilization of molecules on it. For example, 2PO has been shown to promote protein functionalization by non-covalent bonding with high local selectivity.36 Therefore, it should be possible to use 2PO to fabricate a biosensor with proteins binding to very specific areas. As 2PO offers very high control on where the oxidized patterns are, a reference device could be fabricated right next to the actual measurement device with only some μm of distance between them. Finally, a significant advantage of the 2PO method is that it is very simple, it works in ambient air, and it does not require any additional chemicals or tedious and expensive fabrication steps. Thus, it has high potential for becoming a manufacturing technology for applications, as the scale of the oxidation is essentially hampered only by the size of the moving stage used during oxidation.
2PO provides a high level of control in both the oxidation level and the location of the oxidized patterns. It is an easy and fast oxidation method, and we believe it has a lot of potential to advance future sensor applications e.g., in the field of biosensors.
Footnote |
† Electronic supplementary information (ESI) available: Full Raman data fitting parameters, alternative sensitivity correlation plots, sample history, and full sensitivity fitting data. See DOI: https://doi.org/10.1039/d3cp00359k |
This journal is © the Owner Societies 2023 |