Cedric
Wüthrich
a,
Zhiyuan
Fan
a,
Guy
Vergères
b,
Fabian
Wahl
b,
Renato
Zenobi
*a and
Stamatios
Giannoukos
*a
aDepartment of Chemistry and Applied Biosciences, ETHZ, Zurich, Switzerland. E-mail: stamatios.giannoukos@org.chem.ethz.ch; renato.zenobi@org.chem.ethz.ch
bFood Microbial Systems Research Division, Agroscope, Bern, Switzerland
First published on 23rd December 2022
Quantification of metabolites present within exhaled breath is a major challenge for on-line breath analysis. It is also important for gauging the analytical performance, accuracy, reproducibility, reliability, and stability of the measuring technology. Short-chain fatty acids (SCFAs) are of high interest for nutrition and health. Their quantification enables a deep mechanistic understanding of a wide range of biological processes and metabolic pathways, while their high volatility makes them an attractive target for breath analysis. This article reports, for the first time, the development and testing of a modular, dynamic vapor generator for the qualitative and quantitative analysis of volatile SCFAs in the gaseous phase using a secondary electrospray ionization (SESI) source coupled to a high-resolution mass spectrometer. Representative compounds tested included acetic acid, propionic acid, butyric acid, pentanoic acid and hexanoic acid. Gas-phase experiments were performed both in dry and humid (95% relative humidity) conditions from ppt to low ppb concentrations. The results obtained exhibited excellent linearity within the examined concentration range, low limits of detection and quantification down to the lower ppt area. Mixture effects were also investigated and are presented.
Recent advances in chemical sensing tools have facilitated the analysis, screening and decoding of volatile metabolites that are produced by complex and interacting biochemical processes that evolve continuously within the human body. Existing standard technologies for the qualitative investigation and quantification of SCFAs in fecal samples,8 plasma,9 or serum10 include gas chromatography (GC) or liquid chromatography (LC) coupled with mass spectrometry (MS). These methods are well-established; however, they often require invasive sampling, are not on-line, and include demanding sample preparation steps.
On the contrary, exhaled breath analysis is inherently non-invasive, requires minimum or no sample preparation, and provides on-line measurements in real-time.11 Due to these advantages, the research on breath metabolomics has gained momentum in recent years. The presence of SCFAs in human breath has been confirmed,12 but quantification studies are limited, mainly due to the lack of standardization and limited repeatability.11 A recent study13 reported the measurement of carboxylic acids in human breath up to C14 using atmospheric pressure ionization mass spectrometry (API-MS) and estimated the breath concentrations of C3–C6 fatty acids (in a semi-quantitative way) to be on the 100 ppt level which agrees with values reported in the human blood. Lee et al.14 optimized the operational parameters of secondary electrospray ionization (SESI) coupled to a mass spectrometer to measure volatile SCFAs in the headspace gas of gut microbial cultures. Recently, Meurs et al.15 calibrated proton transfer reaction mass spectrometry (PTR-MS) with gas standards of SCFA in the concentration range from 4 to 100 ppb and benchmarked the performance of the PTR-MS against a GC-MS obtaining satisfactory linearity and repeatability. Based on their measurements, the C2–C4 SCFAs exist in the low ppb range in the human breath.
In this article, the development, testing and integration of a dynamic vapor generator with an on-line SESI-Orbitrap mass spectrometer system was demonstrated. Test compounds included volatile C2 to C6 SCFAs. The described vapor generator is a modular system that operates at various levels of relative humidity (RH) and generates gaseous standards of either a single compound or mixtures in the concentration range from low ppt to high ppm in a controllable, repeatable and consistent way. The vapor generator achieves similar conditions to exhaled breath (output total flow per min, humidity, temperature) allowing quantitative investigations of breath metabolites. This paper reports and compares the analytical characteristics (i.e., linear dynamic range, limit of detection and limit of quantification) and performance of this on-line SESI-high resolution mass spectrometry system for the examined SCFA at 0% and 95% RH levels in the positive ion mode.
Compound | Formula | CAS number | Molecular weight (g mol−1) | Vapour pressure (kPa) at 20 °C | H cp (mol (m3 Pa)) at 37 °C |
---|---|---|---|---|---|
Acetic acid | C2H4O2 | 64-19-7 | 60.052 | 1.55 | 17.90 |
Propionic acid | C3H6O2 | 79-09-4 | 74.079 | 0.32 | 23.30 |
Butyric acid | C4H8O2 | 107-92-6 | 88.106 | 0.11 | 18.08 |
Pentanoic acid | C5H10O2 | 109-52-4 | 102.133 | 0.02 | 9.06 |
Hexanoic acid | C6H12O2 | 142-62-1 | 116.160 | 0.018 | 5.78 |
The main components of the vapor generator are (a) a mixing chamber, (b) three individual temperature-controlled evaporation chambers, in which liquid analytes are introduced through a side injection port closed by a septum, (c) four mass flow controllers, (d) an automation platform controlled by a LabVIEW software. The vapor generator is an automated modular system allowing the controllable production of single or multi-component gas standards in either a periodic or a dynamic way. When measurements require humid conditions, a bubbler (indicated by the dashed line at the upper part of Fig. 1) kept at 37 °C can be used in conjunction with the vapor generator to produce a stable humid carrier gas flow. A schematic diagram of the modular vapor generator including only one evaporation chamber is shown in Fig. 1. For the work described in this paper, two different relative humidity levels were investigated (0% and 95%).
The mixing chamber and evaporation chambers were manufactured using grade 303 stainless steel. The mixing chamber is a cylindrically shaped chamber with a length of 80 mm, an inner diameter of 20 mm, and an outer diameter of 25 mm. The input and output ports at the ends are two 6.35 mm Swagelok stainless steel fitting unions. On the peripheral side of the mixing chamber, three 3.18 mm Swagelok fittings are positioned which could be individually connected with a separate evaporation chamber depending on measurement requirements (single or multiple analyte measurements). As such, the mixing chamber of the current version of the vapor generator has connections for three evaporation chambers. The evaporation chambers are cylindrical chambers with an inner diameter of 12 mm, an inner length of 20 mm whereas the outer diameter of the chamber is 30 mm. On the top side of the chamber, there is a 3.18 Swagelok fitting union, and on the peripheral side a Swagelok union tee (1 port connects with a mass flow controller and the other port is sealed with a thermoresistant septum (Sigma Aldrich) and is used for the introduction of the liquid analyte in the evaporation chamber), and on the opposite side, there is an input for a heating cartridge purchased from RS components International. The heater is controlled by a temperature controller with a type K sensor. The temperature range can be set from 0 to 150 °C with 0.1 °C increments and is monitored continuously by a thermocouple. During measurements, all evaporation chambers were kept at 37 °C to simulate human body temperature and the liquid analytes under investigation were loaded into the chamber using a Hamilton syringe. The carrier gas flows passing through all chambers were controlled by mass flow controllers (type GE50A) purchased by MKS Instruments UK Ltd., United Kingdom. The controller connected to the mixing chamber had a flow range between 0.5 and 10 L min−1 and the ones connected to the evaporation chambers had a flow range between 0.1 and 10 mL min−1. The controllers were driven by an automation platform operating through a custom-made software on LabView, which enabled their simultaneous control and the sequence generation of time steps and gas flows for each individual controller. All controllers were calibrated for N2 gas and their input pressure was kept stable at 1.5 bar constantly. The relative humidity was measured at two points (at both ends of the mixing chamber) using two separate sensors at the ends. The first one was an ALMEMO 2590-2A/-4AS (Ahlborn Mess-und Regelungstechnik GmbH, Germany) and the second was a digital humidity and temperature sensor (model SHT45) purchased from Sensirion, Switzerland.
On-line analysis of the produced gas standards was done using a commercial SESI source (Fossil Ion Tech, Spain) coupled to a Q-Exactive Plus Orbitrap (Thermo Fischer, Germany) mass spectrometer. A flow meter (EXHALION, Fossil Ion Tech, Spain) connected at the front-end part of the sample transfer line was used to measure the sample flow rate and volume. All experiments were done at 8 L min−1 corresponding to the average human exhaled breath flow. Gas samples were introduced into the SESI source using a custom-made adaptor attached to the sampling line of the SESI source. The sampling line was constantly heated at 130 °C and the ionization chamber was kept at 90 °C. The electrospray solution was a 0.1% aqueous formic acid solution passing through a nano-electrospray capillary (inner diameter of 20 μm and an outer diameter of 365 μm, Fossil Ion Tech, Spain) with an overpressure of 0.8 bar. In the ionization chamber, the charged electrospray droplets were interacting with the gaseous sample for charge transfer prior to introduction into the mass spectrometer for analysis. Sheath and auxiliary gas values were set at 15 psi and 2 a.u. respectively. The electrospray solution voltage was +3.5 kV and the inlet capillary was heated to 250 °C. The automatic gain control (AGC) target was adjusted to 106 and the maximum injection time was set to 500 ms. Scans were recorded with a m/z range of 50–500 and with a mass resolution of 140000.
![]() | ||
Fig. 2 (a) Stepwise decrease, (b) stepwise increase, (c) random 1 and (d) random 2 flow profiles passing through the evaporation chambers tested for each individual SCFA. |
All experiments were conducted under two conditions, dry and humid (95% RH) in the positive ion mode allowing a comparison of the analytical performance of instrumentation under these two different conditions. The gas phase concentration of each tested SCFA in the evaporation chambers was calculated based on Henry's law.18 The Henry's law solubility constants of the tested SCFAs were obtained by Johnson et al.19 for C2, and the values measured by Khan et al.20 for C3–C6. The Henry's solubility constants for C2, C5 and C6 were measured in the range from 5 to 35 °C and were extrapolated to 37 °C. For C3 and C4, Henry's constants were measured at 25 °C, and were extrapolated to 37 °C using the enthalpy of dissolution reported by Abraham.21 The output of the evaporation chambers was then diluted by the N2 carrier gas flow passing through the mixing chamber.
To examine mixture effects in the gas phase, individual liquid stock solutions for C2, C3 and C4 were prepared at the following concentrations: 0.035, 0.027, and 0.022 mol L−1 respectively. 20 μL of each stock solution were then introduced into three separate evaporation chambers and a fully randomized N2 flow program per chamber was applied: 6, 2, 4, 10, 1, 8 mL min−1 for C2, 1, 8, 2, 10, 4, 6 mL min−1 for C3 and 2, 1, 8, 6, 10, 4 mL min−1 for C4. The N2 flow passing through the mixing chamber was kept again at 8 L min−1 throughout the whole duration of the measurement.
To obtain a calibration curve, the time traces of each compound were further processed. In each time trace, the scans corresponding to flow through the evaporation chamber were identified using the ruptures package,24 which helps to identify change points using the L2 regularization model. Within each pulse in the time trace corresponding to the evaporation chamber activity, the last 15 scans were averaged, thus obtaining the response to the applied concentration. Each pulse was then compared to the limit of detection (LOD) and limit of quantification (LOQ), which were calculated from the baseline scans. The criteria for a pulse being higher than the limit of detection was that the intensity of the pulse had a higher intensity than the average of the baseline plus three times the standard error of the baseline. For the limit of quantification, ten times the standard error was used. It is important to note that the concentrations obtained for these limits were calculated by applying the inverse calibration curve on the corresponding intensity values of these limits. To calculate the regression curves, the baseline subtracted averages of the different flow program measurements were taken. To achieve optimal coverage over the different concentration ranges, a linear regression was drawn through the double logarithmically transformed data.
Accuracy estimation for the individual compounds and RH condition was determined by the exclusion of one data point per technical replicate for each stock solution. A linear regression was fitted over the log-transformed remaining data and the expected value for the omitted data point was compared with the actual value. Reproducibility was assessed through the determination of the coefficients of variation calculated from the technical replicates.
Compound | Exact m/z [M + H]+ | Measured m/z [M + H]+ | Mass error (ppm) |
---|---|---|---|
Acetic acid | 61.0284 | 61.0285 | 1.6 |
Propionic acid | 75.0441 | 75.0440 | −1.3 |
Butyric acid | 89.0597 | 89.0595 | −2.3 |
Pentanoic acid | 103.0754 | 103.0751 | −2.9 |
Hexanoic acid | 117.0910 | 117.0907 | −2.5 |
Calibration curves were obtained over three orders of magnitude concentration for acetic, propionic, butyric, and pentanoic acid, except for hexanoic acid which showed a linear regression over two orders of magnitude. No stable signal was observed for a lower stock concentration for this compound compared to the other compounds. In terms of sensitivity (slope of the linear fit), hexanoic acid showed the steepest slope, followed by acetic acid. For the other three acids, a similar regression slope was obtained. The R2 values obtained were in the range from 0.97 to 0.99. The corresponding LODs and LOQs are reported in Table 3 with average values ranging from low to a few hundreds of ppt's respectively. The calibration curves and the corresponding linearities (R2 values) obtained for each stock solution for the individual analytes under dry conditions are shown in the ESI, Fig. S2.†
Compound | 0% RH | 95% RH | ||||
---|---|---|---|---|---|---|
R 2 | LOD (ppt) | LOQ (ppt) | R 2 | LOD (ppt) | LOQ (ppt) | |
Acetic acid | 0.9969 | 420.63 | 507.36 | 0.9386 | 157.42 | 261.25 |
Propionic acid | 0.9789 | 5.78 | 7.29 | 0.9915 | 3.40 | 5.27 |
Butyric acid | 0.9721 | 1.85 | 2.32 | 0.9838 | 0.71 | 1.00 |
Pentanoic acid | 0.9702 | 2.53 | 3.16 | 0.9890 | 21.89 | 28.13 |
Hexanoic acid | 0.9971 | 40.63 | 47.42 | 0.9802 | 53.67 | 66.81 |
In terms of analytical performance, the fits of the double logarithmically transformed data show satisfactory linearity with the R2 of the fits, possessing values from 0.9802 to 0.9915. A notable exception was the fit obtained for acetic acid, for which an R2 of 0.9386 was obtained. The lower performance relates back to the high variance and stability of the time trace over the pulse experiments. The calibration curves and the corresponding linearities (R2 values) obtained for each stock solution for the individual analytes under humid conditions are shown in the ESI, Fig. S3.†
![]() | ||
Fig. 5 (a) Time traces of acetic acid, propanoic acid and butyric acid, (b) calibration curves of the individual fatty acids, when a mixture of all was used. The curves are given on a linear scale. |
Compared to the individual calibration curves reported in Fig. S2,† the slopes are different, but the magnitudes are on the same level. The sensitivity behavior of the SCFAs (Fig. 5b) compared to each other is the same in the mixture as in a single pure compound calibration (butyric acid > propionic acid > acetic acid). The individual slope differences likely arise from instrumental drift; therefore, a proper calibration must be performed on a daily basis.
In Table 4, the accuracy and reproducibility estimates are given for each compound and RH condition. The estimated accuracy remains for most compounds within ±15% of the expected value apart from butanoic acid, which shows a slightly higher deviation under dry and acetic acid under humid conditions. The variation between technical replicates is low with the trend of humid conditions being less precise, although almost with a negligible difference.
Compound | 0% RH | 95% RH | ||
---|---|---|---|---|
Accuracy (%) | Reproducibility (%) | Accuracy (%) | Reproducibility (%) | |
Acetic acid | 94–107 | 4.35 | 99–118 | 7.78 |
Propanoic acid | 90–109 | 2.34 | 96–107 | 4.36 |
Butanoic acid | 88–116 | 3.93 | 97–108 | 3.72 |
Pentanoic acid | 97–112 | 4.50 | 95–106 | 4.49 |
Hexanoic acid | 94–105 | 5.26 | 90–103 | 6.06 |
The methodology reported here can be used in human studies to quantitatively evaluate how diet quality and the health status of the consumer modulate the production of SCFAs by the human organism.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d2ay01778d |
This journal is © The Royal Society of Chemistry 2023 |