TLC-smartphone in antibiotics determination and low-quality pharmaceuticals detection

Asmaa G. Gada, Yasmin Mohammed Fayezb, Khadiga M. Kelaniab and Amr M. Mahmoud*b
aAnalytical Chemistry Department, Faculty of Pharmacy, Modern University for Technology and Information, Cairo, Egypt
bAnalytical Chemistry Department, Faculty of Pharmacy, Cairo University, El-Kasr El-Aini Street, 11562 Cairo, Egypt. E-mail: amr.bekhet@pharma.cu.edu.eg

Received 19th February 2021 , Accepted 18th May 2021

First published on 26th May 2021


Abstract

Thin layer chromatography (TLC) is a powerful and simple technique for screening and quantifying low quality and counterfeit pharmaceutical products. The detection methods used to detect and quantify separate analytes in TLC ranges from the densitometric method to mass spectrometric or Raman spectroscopic methods. This work describes the development and optimization of a simple and sensitive TLC method utilizing a smartphone CCD camera for verification of both identity and quantity of antibiotics in dosage form, namely ofloxacin and ornidazole. Mixtures of ofloxacin and ornidazole were chromatographed on a silica gel 60 F254 plate as a stationary phase. The optimized mobile phase is n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]1.5 by volume). Iodine vapor has been used as a “universal stain” to visualize the spots on the TLC plates in order to obtain a visual image using the smartphone camera and a desk lamp as an illumination source, thus eliminating the need for a UV illumination source. The recorded images were processed to calculate the Rf values (Rf values for ofloxacin and ornidazole were 0.12 and 0.76, respectively) which provide identity of the drugs while spot intensity was calculated using a commercially available smartphone app and employed for quantitative analysis of the antibiotics and “acetaminophen” as an example of a counterfeit substance. The smartphone TLC method yielded a linearity of ofloxacin and ornidazole in the range of 12.5–62.5 μg/band and 500–1000 μg/band, respectively. The limit of detection was found to be 1.6 μg/spot for ofloxacin and 97.8 μg/spot for ornidazole. The proposed method was compared with the bench top densitometric method for verification using a Camag TLC Scanner 3, the spot areas were scanned at 320 nm. The Rf value of ofloxacin and ornidazole was calculated to be 0.12 and 0.76, respectively. The densitometric method yielded a linearity of ofloxacin and ornidazole in the range of 5–40 μg/band and 5–50 μg/band, respectively. The limit of detection was found to be 0.8 μg/spot for ofloxacin and 1.1 μg/spot for ornidazole. The proposed method has been successfully applied for the determination of ofloxacin and ornidazole present in more than one pharmaceutical dosage form and was comparable to the densitometric method.


1. Introduction

Low-quality medicine is a major problem for patients in developing countries, and it is estimated that 1 in 10 medical products in low- and middle-income countries is substandard or falsified.1–3 Counterfeit medications may contain falsified medication where no active pharmaceutical ingredient (API) is added or substandard APIs (wrong amount of the correct active ingredient), substituted by other cheaper and incorrect APIs (for example acetaminophen)4 or even toxic substances.5 Low-quality pharmaceuticals have a negative impact on the patient's health as the incorrect drug might make the patient's condition worse, or even kill them. Moreover, for antibiotic dosage form, the absence or low active ingredient content may result in the emergence of drug-resistant bacterial strains.6 Antimicrobial resistance (AMR) is the ability of a microbe to resist the effects of medication that once could successfully treat the microbe.7 The term antibiotic resistance is a subset of AMR, as it applies only to bacteria becoming resistant to antibiotics.8,9 Resistant microbes are more difficult to treat, requiring alternative medication, and combination or higher doses of antimicrobials. These approaches may be more expensive, more toxic or both.10

For the developing countries there is an urge for developing rapid, robust and low-cost techniques for identification of low-quality pharmaceuticals. Ideally, such detection technique can be able to detect falsified or substandard medicines. Thin layer chromatographic (TLC) method is ideal for this task and has been employed for detecting counterfeit medication, recent advances have been recently reviewed.11 TLC (and HPTLC) can be regarded as a simple but, in many cases, effective technique with similar performance to HPLC. TLC detection methods enable both qualitative information by utilizing the characteristic Rf values; and quantitative data by various techniques.12 Recently, cheap and widely available detection methods such as scanner, smartphones CCD camera,13 etc. have been employed. The detectors capture the color produced on the TLC plate and gauge the intensity of color using image-processing software. The intensity is then computed in regression equation obtained from calibration curve to get the concentration of the drug.14 CCD smartphone cameras in particular are gaining more interest as detectors for colorimetric and TLC plates. Smartphone has been used for reading TLC plates to detect counterfeit drugs and even detection of illicit substance (i.e. cocaine) under UV lamp as source of illumination to visualize the analyte of interest.15,16 Using UV lamp as a visualization method for TLC adds extra cost for detection in developing and resource-limited areas; moreover, it is only limited to aromatic and conjugated compounds. Iodine has been used as universal method for TLC plate visualization, with the advantages of being widely available, cheap and “semi-destructive” as complexation is reversible and I2 will eventually evaporate. Moreover, there are indicators for visualizing the TLC plate that are specific for certain functional groups, such as nitro group like (acidified potassium permanganate, methanolic potassium hydroxide). It's worth mentioning that an innovative approach has been recently developed to visualize HPTLC plates using multiple illumination sources (in both UV and visible range) and images collected by smartphone CCD camera followed by data manipulation utilizing chemometrics, thus enabled HPTLC clustering/fingerprint.17

Ofloxacin is a broad spectrum antibiotic belonging to the fluoroquinolone class,18,19 its structure is shown in Fig. 1a. It inhibits bacterial cell division by inhibiting DNA gyrase, a type II topoisomerase, and topoisomerase IV, which is an enzyme necessary to separate replicated DNA.20 Ornidazole is an antibiotic of the imidazole class;21 its structure is shown in Fig. 1b. Ornidazole enters the cell by diffusion where the nitro group is reduced by redox proteins present only in anaerobic organisms to reactive nitro radical which exerts cytotoxic action by damaging DNA and other critical biomolecules.22,23 DNA helix destabilization and strand breakage has been observed. The binary mixture of ofloxacin and ornidazole is used for treatment of bacterial & parasitic infections. It is used to treat gastrointestinal infections such as acute diarrhea or dysentery, gynecological infections, lung infections and urinary infections.24 There are many reported RP-HPLC,25–27 HPTLC,26,28 capillary zone electrophoresis,29,30 voltammetry,31 HPLC32,33 UPLC tandem mass spectrometry34 and spectrophotometry24,35 methods for the estimation of ofloxacin and ornidazole from binary pharmaceutical preparations or biological fluids.


image file: d1ra01346g-f1.tif
Fig. 1 (a) Chemical structure of ofloxacin and (b) chemical structure of ornidazole.

In this contribution, we describe the development of easy, cost-effective, fast, green and convenient method ‘to identify the presence of the correct active pharmaceutical ingredients with the stated concentration of each antibiotic drug in the dosage form’ in resource limited areas. Hence, TLC method was employed for separating the antibiotics, visualization of TLC plates was adopted and CCD smartphone camera was employed as detector and finally processing the image with commercially available app software. The proposed method has been applied and compared to standard bench top densitometric method.

2. Experimental

2.1. Apparatus

- Samsung smartphone S5 CCD camera was used to collect the images.

- Camag Linomat 5 autosampler with Camag micro syringe (100 μL); CAMAG, Muttenz, Switzerland.

- Camag TLC scanner 3 densitometer model 3 S/N 130319 equipped with wincats software for densitometric evaluation; CAMAG, Muttenz, Switzerland.

- Thin-layer chromatographic plates; pre-coated with silica gel 60 F254, 20 × 20 cm2, 0.25 mm thickness (E. Merck, Darmstadt, Germany).

- UV lamp-short wavelength 254.0 nm, Spectro line®, model CM-10 (Westbury, New York, USA).

2.2. Chemicals and reagents

All chemicals and reagent used were of analytical grade. Methanol and n-butanol were obtained from Alfa chemical company. NH4OH was obtained from PioChem Company. Standards of OFL and ORN were obtained from National Organization for Drug & Control Research. ORNI-O™ tablet was manufactured by India Acme Lite Tech-LLP. Batch number (ALT19317) was purchased from India. Each tablet is claimed to contain 500 mg ORN and 200 mg OFL as active ingredients.

2.3. Prepared solutions

2.3.1. Stock standard solutions. A weight equivalent to 10.0 mg of either OFL or ORN were accurately weighted and transferred to 10 mL volumetric flask, dissolved and completed to the mark with the methanol to prepare final stock solution of concentration (1 mg mL−1) of the corresponding drugs.
2.3.2. Laboratory prepared mixtures. Different aliquots of OFL and ORN were accurately taken from their standard stock solutions to prepare mixtures containing different ratios of the two drugs.
2.3.3. Chromatographic conditions.
2.3.3.1 Mobile phase. Different developing systems were tried, initially, (1) chloroform[thin space (1/6-em)]:[thin space (1/6-em)]ethyl acetate (6[thin space (1/6-em)]:[thin space (1/6-em)]4 v/v), (2) chloroform[thin space (1/6-em)]:[thin space (1/6-em)]ethyl acetate[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (7[thin space (1/6-em)]:[thin space (1/6-em)]3[thin space (1/6-em)]:[thin space (1/6-em)]0.1, by volume), and (3) n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]0.1, by volume), OFL didn't move from the base line using the previous mentioned systems. By increasing the amount of ammonia to 1.5 mL in the (n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia) system, OFL moved from the base line. Sharp and symmetric spots were obtained using (n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia) (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]1.5, by volume) as a developing system where good separation between OFL and ORN with sufficient difference in their (Rf) values without tailing of the separated bands.
2.3.3.2 Visualization of TLC plate. An “iodine chamber” was used for visualization by adding few crystals of solid iodine with powdered silica in a screw-capped TLC chamber. Then, the developed TLC plate (5 × 10 cm2) was placed in the iodine chamber for 8 minutes until the spots appeared yellow-brown. After that we used, within 5 minutes, a smartphone camera to detect the intensity of each spot color appeared on the TLC plate using “Color Picker” freely available software application version is 5.0.6 (https://play.google.com/store/apps/details?id=gmikhail.colorpicker) and calculated intensity was utilized for the calibration curve construction for quantitative analysis and also, we made qualitative analysis for the detection of adulterant acetaminophen appearance. It's worth mentioning that iodine is known to be volatile as it sublimes easily, therefore, all the images were collected within 5 minutes once we removed the TLC plates out of the developing jar. The smartphone's rear-facing camera is aligned with a plate guide that places the TLC plate into focus and into the camera field of view. The distance from the camera to the plate is 10 cm.36 The TLC plate background was white in color and we capture the image under desk lamp as a source of illumination.

TLC plates also were visualized by certain reagents that are selective for some functional groups, which represent another identification element besides using Rf. To demonstrate that, on a separate plate we employed a selective method for detecting ORN based on visualization by dipping into acidified KMnO4 which gives light brown spots after drying, then the “Color Picker” application used to detect the intensity of each spot and calculated intensity was utilized for the calibration curve construction for quantitative analysis.

2.4. Pharmaceutical dosage form

In case of TLC-smartphone method, ten tablets were weighed, powdered and the average weight of one tablet was calculated. An accurately weighed powder equivalent to one tablet of ORNI-O™ contains (200 and 500 mg of OFL and ORN, respectively) was transferred to a 100 mL volumetric flask; dissolved in 60 mL methanol as a solvent and sonicated for 15 minutes, the volume was completed with methanol to obtain final concentration of 5000 μg mL−1 of ORN and 2000 μg mL−1 of OFL. On the other hand, 1 mL was taken from previous flask and was diluted into 10 mL volumetric flask for analyzing ORN to obtain the volume of 500 μg mL−1 of ORN. Then 1 mL was taken from the last flask and was diluted into 10 mL volumetric flask for analyzing OFL to obtain the volume of 20 μg mL−1 of OFL.

In case of TLC-densitometric method, ten tablets were weighed, powdered and the average weight of one tablet was calculated. An accurately weighed powder equivalent to one tablet had been taken and then 0.01 g of ORNI-O™ tablet was weighted which equivalent to (2000 μg and 5000 μg of OFL and ORN, respectively) and was transferred to 25 mL volumetric flask; dissolved in 10 mL methanol as a solvent and sonicated for 15 minutes, the volume was completed with methanol to obtain final concentration of 80 μg mL−1 OFL and 200 μg mL−1 ORN. Then 1.25 mL was taken from the previous flask was diluted into 10 mL volumetric flask for analyzing OFL and ORN to obtain the volume of 10 μg mL−1 of OFL and 25 μg mL−1 of ORN.

3. Results and discussion

The TLC visualization technique has many advantages as it is simple, low cost, rapid, available, portable and convenient in resources limited countries, no need for instrumentation or skilled technician, we used conventional TLC plates which is cheaper and we achieved TLC visualization using smartphone for detection and we were able to achieve comparable results by rationale optimization of the developing system, these advantages are on contrary of previous studies which use the HPTLC technique ‘which is expensive and not readily available in the limited resource countries’28 and which use smartphones with UV visualization.36 It is very convenient to use this technique in the resources limited countries to check if the drug product is adulterated or not without the need of using sophisticated instrumentation, just an image of a plate representing the chromatographic results with the detected spots for visual comparison of Rf values (identity) and intensities (drug content) to ensure that content in not sub-therapeutic dose.

TLC plate shows 3 spots of ofloxacin, ornidazole, and acetaminophen (commonly used adulterant) visualized with I2 after a run with the mobile phase (n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia) (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]1.5, by volume) to determine the adulteration of the mixture OFL and ORN with acetaminophen rapidly as shown in Fig. 2. Both OFL and acetaminophen appear as brown spots, while ornidazole appears as a light brown spot. It worth nothing that specific and selective visualization of ORN can be performed with acidified KMnO4 reagent which produces yellow changes into light brown color, after drying, which is selective for nitro functional group.


image file: d1ra01346g-f2.tif
Fig. 2 TLC plate showing the three separated compounds; OFL, ORN, acetaminophen visualized with iodine.

The calculated Rf for OFL, ORN, acetaminophen is (0.12, 0.76, 0.7), respectively so it's so easy and rapid determining the appearance of adulterant such as acetaminophen (qualitative analysis) only by making visualization with iodine and determine the Rf of each compound.

To test the capability of the proposed method for quantitative analysis of the antibiotics, we spotted five concentrations of ofloxacin (12.5, 25, 37.5, 50, 62.5 μg/band) on a TLC plate and visualized the separated spots with iodine and on the recorded image we determined the luminance of each spot using the “color picker” freely available software application as shown in Table 1.

Table 1 The relation between luminance and different concentrations of OFL and ORN (12.5, 37.5, 50, 62.5 μg/band), (500, 600, 700, 900, 1000 μg/band), respectively
Drug Concentrations (μg/band) Average luminance (average of 3 replicates of the same spot) Standard deviation (SD) Average luminance (average of three replicates of 3 different spots) Standard deviation (SD)
OFL 12.5 14.3 0.5 14.4 0.6
25 12.7 0.5 12.5 0.5
37.5 10.7 0.5 10.73 0.5
50 9 0.8 8.9 0.5
62.5 7 0.6 7.3 0.6
ORN 500 34.3 0.5 34.7 0.6
600 33.3 0.6 33.4 0.4
700 31.3 0.5 31.4 0.6
900 29.3 0.5 28.4 0.5
1000 26.3 0.5 26.4 0.6


Fig. S1a represents the relation between calculated luminance and OFL concentration. The relation between concentration of the drug and luminance is linear with r = 0.9999 and we spotted five concentrations of ornidazole (500, 600, 700, 900, 1000 μg/band) on a TLC plate and visualized the separated spots with acidified KMnO4 and on the recorded image we determined the luminance of each spot using the “color picker” software application as shown in Table 1. Fig. S1b represents the relation between calculated luminance and ORN concentration. The relation between concentration of the drug and luminance is linear with r = 0.9991. This method is sensitive for determination of the antibiotics in their tablet as the concentration in tablet is high (200 mg OFL & 500 mg ORN) but it would be challenging to determine these antibiotics in plasma as the concentration is smaller than TLC-smartphone CCD camera detection limit. Blank was performed on a TLC plates that have been exposed to iodine under the same conditions and the average luminance was 40.0 and the S. D was 0.6.

The validity of the proposed TLC-visualization method was achieved by means of LOD, LOQ, accuracy and precision as shown in Table 2.

Table 2 Comparison between validation parameters of the proposed TLC-visualization method and TLC-densitometric method for determination of ofloxacin and ornidazole in their binary mixture
  TLC-visualization method TLC-densitometric method
a Average of three determinations.b Accuracy (the mean of 5 different concentrations of each OFL and OR).c Recovery of different laboratory prepared mixtures containing different ratios of OFL and OR.d Limit of detection is determined via calculations, LOD = (SD of response/slope) × 3.3; LOQ = (SD of response/slope) × 10.e Intraday precision (the RSD of 3 different concentrations) (10, 20, 25 μg/band for OFL) & (5, 10, 50 μg/band for OR) in (TLC-densitometric) and (12.5, 25, 37.5 μg/band for OFL) & (500, 700, 900 μg/band for OR) in TLC-smartphone, 3 replicates each, within the same day.f Interday precision (the RSD of 3 different concentrations) (10, 20, 25 μg/band for OFL) & (5, 10, 50 μg/band for OR) in (TLC-densitometric) and (12.5, 25, 37.5 μg/band for OFL) & (500, 700, 900 μg/band for OR) in TLC-smartphone, 3 replicates each, on 3 successive days.
Parameters OFL ORN OFL ORN
Wave length     320.0 nm 320.0 nm
Range (μg/band) 12.5–62.5 500–1000 5–40 5–50
Linearity (regression equation) Y = −0.1466x + 16.183 Y = −0.0171x + 43.553 Y = 434.5x + 10[thin space (1/6-em)]455 Y = 516.98x + 23[thin space (1/6-em)]107
Slopea 0.15 0.02 434.5 516.9
Intercepta 16.2 43.6 10[thin space (1/6-em)]455 23[thin space (1/6-em)]107
Correlation coefficienta 0.9999 0.9991 0.9999 0.9999
Accuracyb (mean ± SD) 100.0 ± 0.6 100.1 ± 0.7 99.9 ± 0.7 99.7 ± 0.9
Specificityc     98.6 ± 0.4 100.5 ± 1.5
LOD (μg/band)d 1.6 97.8 0.8 1.1
LOQ (μg/bandd 4.9 296.3 2.3 3.3
Precision (±RSD%)
(a) Repeatabilitye ±0.8 ±0.7 ±0.4 ±0.8
(b) Intermediate precisionf ±0.4 ±0.5 ±0.3 ±0.7


The TLC-smartphone method was successfully applied for the determination of OFL and ORN in their combined pharmaceutical formulation (ORNI-O™ tablet) and the recovery results was 99.1 ± 0.7 for ofloxacin and 98.4 ± 0.6 for ornidazole as shown in Table 3.

Table 3 Determination of OFL and ORN in ORNI-O™ tablet by TLC-smartphone method and the TLC-densitometric
Pharmaceutical formulation aRecovery% ± SD (TLC-smartphone method) aRecovery% ± SD (TLC-densitometric method)
a Average of three determinations.b Batch no. ALT19317 (labeled to contain 200 mg OFL and 500 mg ORN).
bORNI-O™ OFL ORN OFL ORN
99.1 ± 0.7 98.4 ± 0.6 98.3 ± 0.5 99.9 ± 0.6


The proposed method has been compared to the bench top TLC-densitometry method which is a useful technique for the resolution and in turn for the quantitative determination of drug mixtures. This method offers high sensitivity and selectivity for the analysis of OFL and ORN in the presence of each other without any interference in their pure form and in their pharmaceutical dosage form. Different developing systems were tried, chloroform[thin space (1/6-em)]:[thin space (1/6-em)]ethyl acetate (6[thin space (1/6-em)]:[thin space (1/6-em)]4 V/V), chloroform[thin space (1/6-em)]:[thin space (1/6-em)]ethyl acetate[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (7[thin space (1/6-em)]:[thin space (1/6-em)]3[thin space (1/6-em)]:[thin space (1/6-em)]0.1, by volume), and n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]0.1, by volume), OFL didn't move from the base line using the previous mentioned systems. Increasing the amount of ammonia to 1.5 mL made the OFL moves from the base line. Sharp and symmetric spots were obtained using (n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia) (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]1.5, by volume) as a developing system where good separation between OFL and ORN with sufficient difference in their (Rf) values without tailing of the separated bands as shown in Fig. 3 was obtained. Well defined bands were obtained when the chromatographic tank was previously saturated with the mobile phase for 20 minutes at room temperature. The instrumental conditions such as slit dimension and detection wavelengths were optimized. Detection at λ 320 nm for both OFL and OR was suitable providing good sensitivity for determination of OFL and OR with minimal noise. TLC-densitometric scanning chromatograms of different concentrations of OFL and ORN were performed. Calibration curves were constructed representing the relationship between the obtained integrated peak areas and the corresponding concentrations in the range of (5–40 μg/band) for OFL and (5–50 μg/band) for OR. Linear relationships were obtained as shown in Fig. S2a and b. The regression equations were computed and found to be:

PAOFL 320.0 nm = 434.5x + 10455 r = 0.9999

PAOR 320.0 nm = Y = 516.98x + 23107 r = 0.9999
where PA is the integrated peak area at 320.0 nm for OFL and OR, x is the corresponding concentration in μg/band and r is the correlation coefficient.


image file: d1ra01346g-f3.tif
Fig. 3 2D TLC chromatogram of (1) ofloxacin (Rf = 0.12) and (2) ornidazole (Rf = 0.76), at 320.0 nm using n-butanol[thin space (1/6-em)]:[thin space (1/6-em)]methanol[thin space (1/6-em)]:[thin space (1/6-em)]ammonia (8[thin space (1/6-em)]:[thin space (1/6-em)]1[thin space (1/6-em)]:[thin space (1/6-em)]1.5, by volume) as a developing system.

The validity of the TLC densitometric method was achieved by means of LOD, LOQ, accuracy and precision as shown in Table 2. The results obtained from the laboratory prepared mixtures containing different ratios of OFL and ORN separated at specified conditions were presented in Table 4. The TLC-densitometric method was successfully applied for the determination of OFL and ORN in their combined pharmaceutical formulation (ORNI-O™ tablet) and the recovery results was 98.3 ± 0.5 for ofloxacin and 99.9 ± 0.6 for ornidazole as shown in Table 3.

Table 4 Determination of ofloxacin and ornidazole in their laboratory prepared mixtures by the TLC-densitometric method
Mix. ratio aRecovery%
OFL[thin space (1/6-em)]:[thin space (1/6-em)]OR OFL ORN
a Average of three determinations.
10[thin space (1/6-em)]:[thin space (1/6-em)]10 98.1 100.8
10[thin space (1/6-em)]:[thin space (1/6-em)]20 98.5 98
20[thin space (1/6-em)]:[thin space (1/6-em)]10 98.5 100.6
10[thin space (1/6-em)]:[thin space (1/6-em)]25 98.6 102.3
40[thin space (1/6-em)]:[thin space (1/6-em)]10 99.3 100.7
Mean 98.6 100.5
SD 0.4 1.5


Statistical comparison of the results obtained by the proposed method and those by applying the “official” and “reported” methods showed that there is no significant difference with respect to accuracy and precision as represented in Table 5. Where the proposed method is the TLC-smartphone method. In case of OFL there is an “official” method (Pharmacopeial method), therefore it has been adopted for comparison, whereas, in case of ORN there is no official method so the “reported” densitometric method had been used to compare the ORN results with TLC-smartphone method.

Table 5 Statistical analysis of the results obtained by the proposed method and the official method for the determination of OFL and ORN in pure powder form
Item OFL ORN
Proposed methoda Official methodb Proposed method Reported methodc
a Proposed method is the TLC-smartphone method.b Potentiometric method by dissolving 0.3 g of OFL in 100 mL of anhydrous acetic acid. Titrate with 0.1 M perchloric acid and determine the end point.37 (pharmacopeial method which readings of OFL were compared with).c Reported method is the method which ORN readings were compared with, as ORN has no official method.38d Numbers between parentheses represents the corresponding tabulated values of t and F at P = 0.05.
Mean 100.1 100.4 100.7 99.7
SD 0.9 0.8 0.8 0.9
Variance 0.9 0.7 0.6 0.8
N 5 6 5 5
Student's t-testd 0.39 (2.3)   0.72 (2.36)  
F-valued 1.2 (5.41)   1.34 (9.12)  


4. Conclusion

Low quality medication in low-income countries represent a major challenge for healthcare system, therefore, developing simple, robust method for detecting low quality and counterfeit medications will have a major impact on population health. TLC separation method combined with simple I2 visualization method and smartphone detection for capturing and analysis of the image has been proposed for both qualitative and quantitative determination of antibiotics drugs in bulk and pharmaceutical formulation. Furthermore, based on the specific functional group such as nitro, ornidazole can be independently identified in presence of other co-formulated drug and possible adulterant such as “acetaminophen”. The method was compared to the benchmark densitometric method and no statistical difference was reported. This TLC separation plus smartphone detection method is very simple, cost-effective (no UV source is required) and more available so it can be extended to detect other antibiotics in pharmaceutical formulations.

Conflicts of interest

There are no conflicts of interest to declare.

References

  1. A. Petersen, N. Held and L. Heide, PLoS One, 2017, 12(9), e0184165,  DOI:10.1371/journal.pone.0184165.
  2. S. M. Beargie, C. R. Higgins, D. R. Evans, S. K. Laing, D. Erim and S. Ozawa, PLoS One, 2019, 14(8), e0217910,  DOI:10.1371/journal.pone.0217910.
  3. E. Pisani, A. L. Nistor, A. Hasnida, K. Parmaksiz, J. Xu and M. O. Kok, Wellcome Open Res., 2019, 4, 70,  DOI:10.12688/wellcomeopenres.15236.1.
  4. A. A. Weaver, H. Reiser, T. Barstis, M. Benvenuti, D. Ghosh, M. Hunckler, B. Joy, L. Koenig, K. Raddell and M. Lieberman, Anal. Chem., 2013, 85(13), 6453,  DOI:10.1021/ac400989p.
  5. P. N. Newton, M. D. Green and F. M. Fernández, Trends Pharmacol. Sci., 2010, 31, 99–101 CrossRef CAS PubMed.
  6. A. Johnston and D. W. Holt, Br. J. Clin. Pharmacol., 2014, 78(2), 218,  DOI:10.1111/bcp.12298.
  7. J. Ali, Q. A. Rafiq and E. Ratcliffe, Future Sci. OA, 2018, 4, FSO290 CrossRef CAS PubMed.
  8. S. Metz-Gercek, R. Strauß, C. Rambaud, A. A. Adegoke, W. Adamus-Bialek and L. M. Durso, Advances in Antibiotic Resistance, 2019, 327, 1–28 Search PubMed.
  9. N. Kardos, SOJ Microbiology & Infectious Diseases, 2017, 5, 1–21 Search PubMed.
  10. M. L. Cohen, Science, 1992, 257, 1050–1055 CrossRef CAS PubMed.
  11. J. Sherma and F. Rabel, J. Liq. Chromatogr. Relat. Technol., 2019, 42, 367–379,  DOI:10.1080/10826076.2019.1610772.
  12. R. J. Ekiert, J. Krzek and W. Rzeszutko, Chromatographia, 2008, 67, 995 CrossRef CAS.
  13. I. Hussain, K. Ahamad and P. Nath, RSC Adv., 2016, 6, 22374–22382,  10.1039/c6ra02483a.
  14. L. F. Capitán-Vallvey, N. Lopez-Ruiz, A. Martinez-Olmos, M. M. Erenas and A. J. Palma, Anal. Chim. Acta, 2015, 899, 23–56 CrossRef.
  15. F. Tosato, T. R. Rosa, C. L. M. Morais, A. O. Maldaner, R. S. Ortiz, P. R. Filgueiras, K. M. Gomes Lima and W. Romão, Anal. Methods, 2016, 8, 7632–7637,  10.1039/c6ay02126c.
  16. H. Yu, H. Le, S. Lumetta, B. T. Cunningham, E. Kaale and T. Layloff, Proc. IEEE Sens., 2017, 125, 85–93,  DOI:10.1109/ICSENS.2016.7808847.
  17. L. Pieszczek and M. Daszykowski, Talanta, 2021, 221, 121599,  DOI:10.1016/j.talanta.2020.121599.
  18. W. J. Munckhof, in Kucers the Use of Antibiotics: A Clinical Review of Antibacterial, Antifungal, Antiparasitic, and Antiviral Drugs, 7th edn, 2017,  DOI:10.1201/9781315152110.
  19. X. Si, Y. Wei, C. Wang, L. Li and Y. Ding, Anal. Methods, 2018, 10, 1961–1967 RSC.
  20. M. Iftekhar Hussain, A. Sarker, T. Begum, F. Hasin, F. Binte Farid Himu and M. Riazul Haque Shahin, J. Res. Pharmaceut. Sci., 2016 Search PubMed.
  21. P. Rutgeerts, G. Van Assche, S. Vermeire, G. D'Haens, F. Baert, M. Noman, I. Aerden, G. De Hertogh, K. Geboes and M. Hiele, Gastroenterology, 2005, 128, 856–861 CrossRef CAS PubMed.
  22. W. Raether and H. Hänel, Parasitol. Res., 2003, 90, S19–S39,  DOI:10.1007/s00436-002-0754-9.
  23. A. Dublanchet and R. Durieux, Ann. Microbiol., 1980, 131, 45–59 CAS.
  24. D. M. Patel, J. A. Soneji, P. B. Patel and C. N. Patel, Pharm. Methods, 2012, 3, 102–105 CrossRef PubMed.
  25. T. D. F. B. Y. RP-HPLC, Methods, 2010, 1, 78–83 Search PubMed.
  26. M. Puranik, D. V Bhawsar, P. Rathi and P. G. Yeole, Indian J. Pharm. Sci., 2010, 72, 513 CrossRef CAS PubMed.
  27. R. Singh, M. Maithani, S. K. Saraf, S. Saraf and R. C. Gupta, Eurasian J. Anal. Chem., 2009, 4(2), 161–167 CAS.
  28. T. K. Ravi and N. Shukla, Indian J. Pharm. Sci., 2006, 68(6), 838–840,  DOI:10.4103/0250-474X.31033.
  29. K. L. See, A. A. Elbashir, B. Saad, A. S. Mohamed Ali and H. Y. Aboul-Enein, Biomed. Chromatogr., 2009, 23, 1283–1290,  DOI:10.1002/bmc.1251.
  30. B. Koppenhoefer, A. Jakob, X. Zhu and B. Lin, HRC J. High Resolut. Chromatogr., 2000, 23, 413–429,  DOI:10.1002/1521-4168(20000601)23:6<413::AID-JHRC413>3.0.CO;2-E.
  31. S. A. Özkan, B. Uslu and H. Y. Aboul-Enein, Crit. Rev. Anal. Chem., 2003, 33(3), 155–181,  DOI:10.1080/713609162.
  32. K. Lu and Q. Tong, Fenxi Ceshi Xuebao, 2011 Search PubMed.
  33. A. P. Dewani, R. L. Bakal, P. G. Kokate, A. V. Chandewar and S. Patra, J. AOAC Int., 2015, 98(4), 913–920,  DOI:10.5740/jaoacint.14-189.
  34. F. Tamtam, F. Mercier, J. Eurin, M. Chevreuil and B. Le Bot, Anal. Bioanal. Chem., 2009, 393, 1709–1718,  DOI:10.1007/s00216-008-2576-9.
  35. K. P. Bhusari and D. R. Chaple, Asian J. Res. Chem., 2009, 2, 60–62 CAS.
  36. H. Yu, H. M. Le, E. Kaale, K. D. Long, T. Layloff, S. S. Lumetta and B. T. Cunningham, J. Pharm. Biomed. Anal., 2016, 125, 85–93,  DOI:10.1016/j.jpba.2016.03.018.
  37. United States Pharmacopeial Convention, 2003 USP NF and USP NF 21, Rockville, MD.
  38. P. N. Ranjane, S. V. Gandhi, S. S. Kadukar and K. G. Bothara, J. Chromatogr. Sci., 2010, 48(1), 26–28,  DOI:10.1093/chromsci/48.1.26.