Pascal
Becker
a,
Christoph
Neff
a,
Sabine
Hess
b,
Peter
Weis
c and
Detlef
Günther
*a
aDepartment of Chemistry and Applied Biosciences, Laboratory of Inorganic Chemistry, ETH Zurich, Vladimir-Prelog-Weg 1, 8093 Zürich, Switzerland. E-mail: detlef.guenther@sl.ethz.ch
bZurich Forensic Science Institute, Zeughausstrasse 11, 8004 Zurich, Switzerland
cFederal Criminal Police Office, Forensic Science Institute, KT42 Inorganic Materials and Micro Traces, Coatings, Äppelallee 45, 65203 Wiesbaden, Germany
First published on 23rd July 2020
The most discriminating method that is currently applied in routine forensic elemental analysis of glass is laser ablation in combination with quadrupole or sector field based inductively coupled plasma mass spectrometry (LA-ICPMS) following the standard method ASTM E2927-16E1. Due to the sequential measurement method, sample size constraints of approximately 400 μm × 200 μm × 100 μm with six individual measurements have been reported to provide optimum parameters for glass discrimination. The necessary sample size is given by the spot size, number of laser pulses per replicate and number of replicate measurements. In this study, a single-pulse laser ablation inductively coupled plasma time of flight mass spectrometry (Single-Pulse LA-ICP-TOFMS) method was developed and applied for the matching and mismatching of forensic float glass evidence. 110 laser pulses with an ablation spot diameter of 90 μm were applied to float glass fragments of a size below 400 μm in diameter. 18 elements were quantified from each individual laser pulse in order to compare the corresponding concentrations between various glass fragments. In contrast to previous work, a modified 5-sigma criterion was used to successfully match fragments from the same source and mismatch fragments from differing sources. The data set generated within this study demonstrates new capabilities when using quasi simultaneous signal detection and reduced the necessary sample volume from 400 μm × 200 μm × 100 μm to 100 μm × 100 μm × 33 μm, which corresponds to a reduction in sample material from 20 μg to 0.8 μg. The proof of concept is shown for the application of a single-pulse based method for glass fragment analysis in a forensic context, which would allow the measurement of smaller fragments than previously possible.
Analysis of glass samples using LA-ICPMS has been studied since the 1990's but gained further popularity in a forensic context due to the “natural isotopes and trace elements in criminalistics and environmental forensics” (NITE-CRIME) initiative, with the goal of using trace-elements and isotope ratios in forensic sciences.2,10–12 As a result of this project, a standard method for LA-ICPMS has been developed which is now used routinely in forensic institutes around the world.2,13 The method includes quantification of approximately 18 elements (typically Li, Na, Mg, Al, K, Ca, Ti, Mn, Fe, Rb, Sr, Zr, Ba, La, Ce, Nd, Hf and Pb) and the comparison of their concentrations. The standard method requires at least 3, optimally 6, measurements of each glass fragment, using an ablation spot size of 50–100 μm in diameter. A constant ablation time of 50–60 seconds at a laser repetition rate of 10 Hz is applied. This results in a required sample volume of approximately 400 μm × 200 μm × 100 μm per glass fragment. However, the majority of samples retrieved are below that size, making it impossible to measure them according to the standard method.
Additionally, due to the low number of measurements, many statistical models fail in dealing with the multidimensional data acquired by quantifying multiple elements due to the so called “Curse of Dimensionality”, described by Bellman in 1961.14 It describes the increase of distance between data points in “nearest neighbour” statistical approaches when the dimensionality of the data increases. This results in data points being too far away from each other to be clustered when the sample size is too low. As such the amount of required data points often increases exponentially with every additional dimension, which is a general challenge in applied analytical methods. Therefore, there is an interest in methods that can provide similar discrimination power with less sample material and a larger number of measurements.
Quadrupole, sector field and multi-collector instruments rely on sequential measurements to cover the entire mass spectrum. This results in a spectral acquisition time that is dependent on the amount of measured isotopes. An increase in the number of isotopes will increase the measurement time per spectral acquisition, resulting in longer ablation times and increasing sample amount requirements. ICP-TOFMS instruments allow for quasi-simultaneous measurements of the entire mass range, which opens the possibility to quantify more elements at once.15 This allows for further discrimination factors between samples and could possibly increase the discrimination power of the method. When the standard method was developed, first generation ICP-TOFMS instruments showed rather low sensitivity and were not suited for forensic glass examinations. Recent instrumental developments have allowed ICP-TOFMS-instruments to reach sensitivities which are comparable to quadrupole instruments and were therefore reconsidered to extend the range of applications.16,17
Single-pulse laser ablation was first described for ICP-TOFMS in 2001 by Leach et al.18 by resolving the signal of each individual laser pulse in the transient signal. Every laser pulse was quantified individually, increasing the amount of information per ablated material significantly. This increases the amount of measurements from one for 600 laser pulses to one for every laser pulse. With the development of ICP Time-of-Flight Mass Spectrometry (ICP-TOFMS), it is possible to measure the entire mass spectrum at a frequency of more than 30000 Hz.19 The quasi-simultaneous detection of all ICPMS accessible elements results in a representative measurement of the ion cloud for every data point. A fast washout of the ablated aerosol is required in Single-Pulse-LA-ICPMS to increase the signal-to-noise ratio and thus lower the limits of detection. In this work, a fast aerosol washout is achieved using a low dispersion tube cell as described by Wang et al. in combination with ICP-TOFMS.20 With the use of this low-dispersion LA cell, it is possible to measure 99% of the total signal within 5–100 ms (5–90 μm spot diameter).
In this study, a reduced sample consumption method was developed and evaluated to match forensic glass fragments by means of Single-Pulse-LA-ICP-TOFMS. It relies on the fast and “quasi” simultaneous detection of single laser ablation events using an ICP-TOFMS and results in a reduction of required sample amount. This sampling strategy allows the analysis of significantly smaller glass fragments compared to the standard method currently used. Additionally, it increases the amount of individual measurements which opens the possibilities to utilize different statistical methods.
The method proposed in this study was used to quantify a wide range of elements covering concentrations from 1 to 100000 μg g−1. The major goal was focused on matching glass fragments from the same source, while mismatching fragments from different sources through comparison of the element concentrations.
Laser ablation | |
Repetition rate | 2 Hz |
Laser fluence | ∼15 J cm−2 |
Spot size | 90 μm |
Carrier gas (Ar) | 0.75–0.85 L min−1 |
Carrier gas (He) | 1.2–1.3 L min−1 |
Number of pulses per spot | 110 |
![]() |
|
ICPMS | |
RF power | 1550 W |
Cooling gas (Ar) | 16 L min−1 |
Auxiliary gas (Ar) | 0.8 L min−1 |
Reaction cell (H2) | 2 ml min−1 |
Sampling depth | 5.5–6 mm |
Time resolution | 10 ms |
An in-house built modified low dispersion ablation cell, based on the tube cell design,21 was used with mixed argon and helium (99.999%, PanGas AG, Dagmersellen, Switzerland) gas flows to carry the aerosol into the ICP.17,20
The ICP-TOFMS instrument was operated in reaction cell mode, using hydrogen (99.9999%, PanGas AG, Dagmersellen, Switzerland) as reaction gas.22
Measuring FGS 2, the LODs were calculated and compared to the mean concentrations measured in the float glass fragments. It is worth noting, that a notch filter was applied at m/z 23 and 28 to prevent saturation of the detector from Na and Si, which are the main elements in such glasses. This results in higher LODs for low mass elements than one would expect, since it also affects nearby masses.
Measurements consisting of 110 pulses per sampling position were carried out and 100 individual pulses were quantified as a larger number of data points resulted in better reproducibility of the quantitative results. The 10 pulses at the beginning of the analysis were not considered to avoid analysis being affected by surface effects, caused by impurities, crater formation and uneven surfaces (Fig. 3). Furthermore, the uneven shape of glass fragments is suspected to cause local gas turbulences within the ablation cell which could result in incomplete or non-stoichiometric aerosol transport. This becomes evident since a change in signal duration and signal shape was observed for certain samples with uneven surfaces (Fig. 4).
RSDs for the quality control (FGS1 in this case) measurements were calculated by taking the RSDs over all medians for each element. RSDs for samples were determined by calculating the RSDs of the medians of all 10 fragments per glass source and taking the mean RSDs over all 10 glass sources for every element. Optimal RSDs were determined by optimizing the value for every element for the lowest achievable amount of false positives, identifying 2 different fragments as a match, and negatives, identifying 2 equal fragments as a mismatch (see Table 2).
Element | RSD quality control | RSD samples | RSD optimal | RSD standard method |
---|---|---|---|---|
23Na | 0.04 | 0.05 | 0.05 | 0.03 |
25Mg | 0.04 | 0.04 | 0.05 | 0.03 |
27Al | 0.06 | 0.06 | 0.06 | 0.03 |
39K | 0.07 | 0.07 | 0.09 | 0.03 |
44Ca | 0.04 | 0.04 | 0.04 | 0.03 |
49Ti | 0.08 | 0.04 | 0.04 | 0.04 |
55Mn | 0.03 | 0.04 | 0.05 | 0.04 |
57Fe | 0.04 | 0.04 | 0.05 | 0.03 |
85Rb | 0.05 | 0.06 | 0.07 | 0.04 |
88Sr | 0.05 | 0.04 | 0.03 | 0.04 |
90Zr | 0.05 | 0.05 | 0.05 | 0.04 |
118Sn | 0.05 | 0.09 | 0.07 | 0.03 |
138Ba | 0.06 | 0.08 | 0.09 | 0.04 |
139La | 0.05 | 0.05 | 0.04 | 0.04 |
140Ce | 0.05 | 0.06 | 0.06 | 0.04 |
146Nd | 0.06 | 0.06 | 0.05 | 0.05 |
178Hf | 0.05 | 0.08 | 0.07 | 0.03 |
208Pb | 0.05 | 0.13 | 0.12 | 0.05 |
The quantification of the quality control standard showed an increase in the RSD between the individual laser pulses compared to the standard method. This increase in the RSD when comparing the single-pulse method to the commonly applied low spatial resolution method found in the standard method was expected due to the approximately 25 times lower amount of sampled material entering the plasma and the significantly shorter integration times.
The RSDs for samples were found to be higher than the RSDs for the quality control. This can be partially explained by a higher inhomogeneity of trace elements in real samples in comparison to specifically produced glass standards. Another possible explanation is fractionation induced by gas turbulences caused by the uneven surfaces found in natural glass fragments. A detailed knowledge about the origin of the increased RSDs will be required since the currently achievable precision makes it difficult to directly apply the currently established matching criteria of the standard method. However, the next goal is to develop a statistical method that reduces false negatives and false positives.
With 10 different sample origins and 10 fragments each, a total of 100 fragments were compared. This resulted in a total of 450 possible comparisons of matching fragments and 4500 possible comparisons of mismatching fragments. The rate of false negatives is thus the amount of false mismatches divided by the total amount of comparisons of matching fragments, whereas the rate of false positives is the amount of false matches divided by the amount of comparisons of mismatching fragments. The three different groups of RSDs can then be compared as matching criteria (Table 3).
Matching criterion | 5 × RSDQC | 5 × RSDsamples | 5 × RSDoptimal | Standard method |
---|---|---|---|---|
False negatives | 17.1% | 4.00% | 0.89% | 1.04% |
False positives | 0.00% | 0.73% | 0.37% | 0.11% |
Using the RSDs of a quality control (RSDQC) underestimates the concentration intervals significantly, resulting in a very high rate of false negatives. Using instead the average RSDs of all the measured real samples (RSDsamples), effects caused by the interaction of the carrier gas and the shape of the individual fragment can be compensated, which results in an acceptable range of false negatives closer to the standard method. Optimizing the interval for each element to reach the lowest possible amount of false negatives while maintaining a low amount of false positives, by adapting the RSD in increments, allows for a low rate of 0.89% false negatives. This is an optimized criterion for the specific data set, and would not show the same success for a different data set, however it can be used as a proof of concept and an indication of what can be possible when using further improved statistical methods.
The increase of measured data points from 6 to more than 100 would allow working with distributions rather than absolute values for statistical evaluation. This allows for matching criteria based on elemental distributions, which would work better in a high dimensional space. It is possible to further increase the amount of data points by adding a second ablation spot to each fragment, which would double the required sample area but also the amount of data points. The proposed method in this work offers adaptability to the shape of the sample and for example, working around limitations in sample thickness but not in sample area. Doing so may decrease the amount of false negatives and false positives, as precision increases with amount of measured sample material.
This journal is © The Royal Society of Chemistry 2020 |