Element imaging aims to provide quantitative data on multi-element distributions from major to trace elements with high lateral resolution. Here, we describe a control system for laser ablation inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) imaging with complex integration of translational stage, laser, and mass spectrometer data acquisition which was tested on glass and mineral samples. In particular, besides single pulse imaging at a laser repetition rate of 100 Hz using a 5 μm laser spot size, a hole drilling imaging approach provided higher pixel sensitivity and lower limits of detection (<1 mg kg−1 for most heavy elements) while maintaining the same lateral resolution. An optional surface cleaning pulse can be applied without additional recording of data. Furthermore, the ablation area can be adapted to specific object structures of interest and leads to significant shorter imaging times. Triggering the data acquisition for every ablation position led to binned pixel data in relation to the sample position. This simplifies the data evaluation and allows a more automated image generation. The approach presented in this study enables flexible adjustments of distinct ablation modes to a specific analytical task and provides the basis for fully automated element imaging. To test the applicability of our approach, two complex geological samples containing crystalline solids were imaged to gain insights into the distribution of trace elements that occur typically in the low mg kg−1 range. We show that both single pulse and hole drilling ablation modes allow the determination of a large number of trace elements. However, the hole drilling mode shows a superior sensitivity per pixel, which in turn provides more detailed information about the formation of geological samples.
Technical advances in the determination of major and trace element distributions in key rock minerals improved the definition of the large-scale geological processes significantly. For instance, refined definition of growth conditions and stability intervals of the garnets and zircons groups of silicate minerals allowed improved constraints on rock metamorphic processes6,7 and the discrimination of the genesis of ore minerals in economic deposits.8 These advances demonstrated invariably that high resolution techniques are at the heart of detailed investigations of geological samples. This calls for high spatial resolution, which is essential to distinguish small geometrical features within samples, and for a combination of high sensitivity with low limits of detection (LOD), and simultaneous multi-element detection capability from major to trace elements.
Since the advent of laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS),9,10 methods were developed to determine the bulk composition of particular phases by spot analysis,11 the element distribution in one dimension (line scan),12–17 of two-dimensional areas,18–22 and even spatial spread (3D).23,24 The recent revival of ICP-TOFMS and low dispersion laser ablation cells have led to a rise of interest in fast, multi-element, and high spatial resolution LA-ICP-MS imaging for geological samples.20,24–26 ICP-TOFMS mass analyzers provide quasi-simultaneous detection of the full mass spectrum with fast acquisition rate.27 Altogether, small laser spot sizes, different ablation cells and fast, simultaneous multi-element detection with ICP-TOFMS offer a unique and promising combination to tackle the analytical challenges of element imaging.
High lateral resolution can be achieved by a small laser ablation spot size. Imaging has been achieved at sub-cellular resolution using a laser spot size of ≤1 μm.22,28,29 In some cases, simple laser scan distance in relation to the data acquisition frequency was misunderstood as lateral resolution(e.g.ref. 30) However, while sensitivities still need to be sufficient to yield signals for individual laser pulses, smaller pixel sizes were also accomplished with overlapping scans and signal deconvolution.31–35
To achieve high data acquisition speed while maintaining the lateral resolution, low dispersion ablation cells have to be used to transport the aerosol in a dense plume without pulse-to-pulse aerosol mixing into the ICP-MS.36 Shorter signal durations increase the signal-to-noise ratio and allows a faster data acquisition. In recent years, signal duration of <10 ms (full width at 1% of the maximum, FW0.01M) for LA-ICP-MS imaging were achieved.20,28,37,38 These ablation cells deliver signal durations between 4 and 9 ms (FW0.01M). As an early benchmark, Tanner and Günther39 achieved signal durations in the range of 1–3 ms for an in-torch ablation approach, although without the capability to perform imaging experiments. Similar short signals were most recently reported by Van Malderen et al.29 for imaging applications.
As noted earlier by Paul et al.,40 software development for LA-ICP-MS imaging is lagging behind compared to other imaging techniques. A detailed discussion of this issue can be found in a review by Weiskirchen et al.41 Until now, the major focus was on software solutions for data analysis.40–46 Nevertheless, the more LA-ICP-TOFMS imaging becomes routine, the more need there is for an automation of the data acquisition process. In addition, there is hitherto mostly unused potential for utilizing the capabilities of low dispersion ablation cells in combination with LA-ICP-TOFMS for convenient data acquisition and processing, which can support different aims for imaging beyond mere line scan approaches. For instance, in contrast to quadrupole MS and sector field MS, TOFMS provides complete mass spectra and the full information on isotope signals can be accessed for each laser pulse. Once data is stored in bins representing each ablation position the cumbersome isolation and sectioning of the signals can be avoided, accelerates data processing24 and offers exceptional opportunities for data analysis, e.g. phase specific pixel clustering and thereby improved quantitative data even for small-size sample features.25,26 Moreover, continuous line scans with overlapping laser spots distort spatial information since the ablated material from individual laser pulses originate from different depths along the scan direction. The TOFMS timing can be adjusted so that data for a single or multiple laser pulses directed at the same sample position without sample stage movement can be stored in a single data bin. Hence, software which integrates control of sample translation stage, laser, and ICP-TOFMS is required to further foster LA-ICP-TOFMS imaging with different ablation modes fitting to different study requirements.
Here, we present an integral control system for flexible LA-ICP-TOFMS imaging and show how such a system can be applied to the determination of element distributions in geological samples. We report a leap towards LA-ICP-TOFMS imaging automation and applied different ablation modes by synchronization of translational stage, the laser, and mass spectrometer data acquisition with a custom-built imaging control system. In particular, the setup enables to perform a single pulse imaging mode with high scanning speeds or a hole drilling imaging mode that provides lower LODs. In the former case, laser pulses are applied in quick succession in an edge-to-edge raster over the area of interest. For analysis of elements at lower mass fractions, multiple laser pulses per sample position were applied and thereby the pixel LOD improved while the same lateral resolution was maintained. Due to synchronized and selective triggering in both modes, data is stored for each pixel individually, which assist analysis, data processing, and image generation. The application of these imaging modes to two geological samples demonstrate that they can be employed strategically for the analysis of mineral samples.
Laser fluence | 15 J cm−2 |
Laser spot diameter | 5 μm |
Ablation gas flow (He) | 1.25–1.45 L min−1 |
Carrier gas flow (Ar) | 0.7–0.9 L min−1 |
Auxiliary gas flow (Ar) | 0.8 L min−1 |
Cooling gas flow (Ar) | 16 L min−1 |
RF power | 1400–1550 W |
Notch filter | 40Ar+, 40Ar2+, 56Fe+, 75As+ |
232Th+/238U+ | 0.9–1.1 |
ThO+/Th+ | <1% |
To generate consecutive line scans in single pulse (SP) mode, a laser repetition rate of 100 Hz and a continuous translational sample moving speed of 500 μm s−1 were applied, leading to an edge-to-edge placement of the ablation spots. The overlapping line scan (OL) mode was carried out using a laser repetition rate of 100 Hz and a continuous translational sample moving speed of 20 μm s−1. This led to 25 laser pulses per 5 μm sample movement. Consecutive lines were aligned edge-to-edge. For the hole drilling (HD) mode, 25 laser pulses were fired on the same sample position before the sample was moved to the next position. The signal of 25 laser pulses was summed and displayed as a square pixel of 5 μm × 5 μm for the OL and HD mode. In the hole drilling mode with cleaning pulse (CP + HD), a cleaning pulse was employed at each sample position prior to the start of the hole drilling. There was no ICP-MS data recorded for the cleaning pulse. The triggering pattern and the corresponding TOFMS data acquired of the four ablation modes SP, OL, HD, and CP + HD are illustrated in Fig. 1. The imaging of the minerals ilvaite, hedenbergite, quartz and the analysis of the corresponding external reference material were conducted with additional notch filters for the major elements 56Fe and 75As to avoid detector saturation. Gas blank signals were acquired as separate files.
The laser system was controlled using two separate communication channels. Using an ASCII protocol over a serial interface (RS-232) all parameters of the laser system are controlled (high voltage, power supply mode, discharge voltage, internal or external triggering mode, readout of status, and error messages). The linear piezo-electrical actuators of the sample translation stage were controlled via a MCS 3-axis controller (SmarAct, Oldenburg, Germany) using the integrated USB port with a proprietary binary protocol. A Python C Foreign Function Interface (CFFI) wrapper around the supplied C library was written for integration into the main program. For triggering the laser and TOFMS, a microcontroller (Arduino Uno, Arduino AG, Switzerland) was used and placed in-between. This avoids additional jitter of the trigger signals and ensures sufficient task frequency stability. Communication with the trigger controller was implemented using an ASCII based serial protocol.
The imaging software dovetails different hardware components for LA-ICP-TOFMS imaging experiments, namely control of the movement of the sample translation stage, triggering of laser pulses, and ICP-TOFMS data acquisition in a highly coordinated manner. It further allows for different ablation modes based on specific parameters set by the user, which included laser spot size, laser pulse frequency, and size dimensions of the region of interest (ROI), while the latter may be divided into subsections. Additional parameters included the application of cleaning pulses and time delays.
The specific use of the ICP-TOFMS data for imaging, the data structure, the data acquisition and the need of dividing the analyzed areas into subsection are described in detail in the ESI.† In brief, the different dimensions of the TOFMS data are used to trigger the TOFMS data acquisition for individual pixels and accordingly store data separately and avoid part of the previous cumbersome data processing.24,35 Since the software prompts triggers and controls for the sample translation stage, laser, and data acquisition independently, flexible stage movements and ablation modes are available. Irrespective of ablation mode and sample movement, the imaging control system enables to acquire the full mass spectrum and a transient signal for each sample position results in one pixel. Additionally, the distinct transient signal acquisition allowed to monitor element variations as a function of ablation depth. Notably, the core software and procedures are vendor agnostic and not limited to the LA-ICP-TOFMS hardware employed in this study.
Other laser systems or ICP-MS can be employed in the imaging control system whenever an external trigger input is available to which the trigger board can be connected. Additional development effort has to be made only if data communication with the laser or ICP-MS is necessary. Successful experiments have been conducted using an EX5 laser (GAM LASER, Inc., Orlando, USA) as well as the CyTOF system (Fluidigm, San Francisco, USA). Every stage system which supports a minimal set of features (mainly setting position and movement speed) and has a programming interface which can be called from Python, can be integrated. The “TEMAimaging” software of the imaging control system is provided under an open-source license (http://doi.org/10.5905/ethz-1007-269) and is hosted at https://github.com/ethz-tema/TEMAimaging.
The sample RM3 is an aggregate of ilvaite (formula: CaFe3Si2O8(OH)), hedenbergite (formula: CaFeSi2O6) and pyrite (formula: FeS2). The other sample RM2 is an aggregate of crystallized ilvaite and quartz (formula: SiO2). Prior to our study, the mineralogical and chemical compositions of the two samples have been determined using ancillary optical and electron probe microanalytical techniques (ESI Fig. S1†). The NIST SRM 610 was used as external reference.49,50
The TOFMS data structure of “bufs” and “writes” was used as a relative coordinate system to create quantified element distribution maps (ESI†). Data processing was automated for both Python scripts. LODs were calculated and plotted according to Poisson counting statistics52,53 using an in-house written script running on R.54,55
The TOFMS data acquisition is conducted without loss of information on mineral composition, and the simultaneous detection using a TOFMS allows measuring an extensive mass range (m/z 23–238) in a given mineral without preselection of elements of interest. In order to carry out data analysis and interpretation, the determined element mass fractions can be exported as comma-separated values from the data analysis software and used with other software packages such as XMapTool.56
As an example, we used this mode to image the ilvaite–hedenbergite assemblage of sample RM3 (Fig. 2). Sample RM3 was imaged over an area of 3000 μm × 3000 μm using the SP mode. The data was acquired in 36 subsections, each one covering an area of 500 μm × 500 μm (100 × 100 pixel). 18 subsections were acquired automatically and consecutively, and every subsection was triggered in addition to the triggering of each pixel. Reference material NIST SRM 610 was measured for external calibration at the start, after 18 subsections, and after 36 subsections, respectively. The second half of the image (18 subsections) was also acquired automatically. The 36 subsections were combined during the data processing stage.
The measurement time of the area of 3000 μm × 3000 μm was 77 min, resulting in an acquisition speed of approx. 7 mm2 h−1 at 5 μm lateral resolution. Considering some communication overhead and the stage movement to the next line start position the effective pixel acquisition rate was 78 pixel per s.
The mass fractions of the majority of rare earth elements (REEs) in the ilvaite crystal such as Sc, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Tm, Yb and Lu were below the LOD. For some REEs such as Y, Dy, Ho, and Er at least small areas containing elevated mass fractions were detected. This shows that while the SP mode cannot provide LODs that are sufficiently low to measure complete distribution maps, but it has ability to determine the distribution of a broad range of components, from major to trace elements which can be well suited for a fast screening of samples.
![]() | ||
Fig. 3 Full mass spectrum (a) and transient signal (b) are acquired for each sample position. As an example, the signal of a hole drilling imaging measurement of one sample position on ilvaite is displayed (Fig. 4, x = 670 μm, y = 395 μm). The transient signal shows homogenous distribution of the elements Ca, Fe, In, and Tb in depth. Due to the lower data acquisition rate for the hole drilling mode, transient signals from individual laser pulses are not distinguishable here (compare with ESI†). |
The ablation depth per laser pulse depends largely on the sample composition and is in most cases significantly smaller than 1 μm (e.g. NIST SRM 610 ≈ 0.2 μm per pulse at 15 J cm−2).59 The burst of 25 laser pulses per sample position resulted in surface sampling diameter of 5 μm and a depth of approximately 5 μm. A larger number of laser pulses per position leads in total to a higher signal intensity per pixel, and therefore to higher pixel sensitivity while maintaining the lateral resolution. The number of laser pulses and time duration for data acquisition can be adjusted for different laser spot sizes or sample composition.
We applied this approach to the ilvaite–quartz assemblage of sample RM2 (Fig. 4). This sample was imaged over an area of 600 μm × 1300 μm. The data was acquired in two subsections of 600 μm × 650 μm (120 × 130 pixel). The measurement of reference material NIST SRM 610 was carried out for external calibration at the beginning and at the end of each subsection acquisition. The two subsections were combined later on, during the data processing stage.
![]() | ||
Fig. 4 Element imaging of the ilvaite (Ilv)–quartz (Qtz) assemblage of sample RM2 generated with HD mode. The panels include a representative selection of major (a and b), minor (c–e) and trace elements (f–l) of these two minerals, whose lateral distributions mark a distinct crystalline texture. Similar to what is displayed for RM3 (Fig. 2), also this sample shows a homogeneous distribution of Si (a), but this is not the case for most major, minor, and trace elements. Within ilvaite, a large number of elements (Al, Mn, Sn, In, REEs, others not shown) mark bands of high and low mass fractions in the elongated prisms that may be 5–10 μm in width. In contrast, quartz does not show a systematic distribution of minor and trace elements with the exception for Al (b), alkali metals (c and f), and Sb (h). |
The 600 μm × 1300 μm area measured of sample RM2 took 213 min, resulted in an acquisition speed of approx. 0.22 mm2 h−1 at 5 μm lateral resolution. Considering some communication overhead and the stage movement to the next line start position, the effective pixel acquisition rate was 2.4 pixel per s. Further development of the HD imaging procedure allowed to speed up the effective pixel acquisition rate to 2.9–3.0 pixel per s which was subsequently used for imaging of a shape adapted ablation area. 250 ms per sample position are required for 25 laser pulses at 100 Hz and signal detection. In addition, after firing 25 laser pulses the imaging control system required an average of 80–90 ms to move the stage to the next sample position and to receive the feedback. Compared to the SP mode, the HD mode was 27 times slower.
According to the different applications it is important to note that despite of the slower speed of the HD mode, its lower LOD allowed the determination of trace elements which occur typically in the low mg kg−1 range. The HD mode enabled quantification of most of the REEs, as well as areas containing elevated mass fractions of Sc, Y, La, Ce, Pr, Nd, Sm, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, Pb, Th and U which enhance the understanding of the crystal formation processes as reported in [e.g., ref. 60 and 61]. Eu was the only REE whose mass fraction was below the LOD (0.4 mg kg−1).
![]() | ||
Fig. 5 Quantified maps of Fe and In of an ilvaite–quartz assemblage of sample RM2 comparing the ablation modes SP (a), HD (b), CP + HD (c), and OL (d). Notice that this sample area analyzed is smaller compared to the imaging displayed in Fig. 2 and 4. |
![]() | ||
Fig. 6 Pixel LODs for the ablation modes HD, OL and SP using NIST SRM 610 according to Table S1.† The notch filter was applied at m/z 56 and m/z 75 leading to higher LODs for elements with close m/z isotopes such as Mn, Fe, Ga, Ge, As, and Se. Data points of OL are covered by them of HD as the LODs for HD and OL are similar. |
For the same element the pixel sensitivity and consequently the LOD largely depended on the amount of material ablated as laser ablation rates varied depending on the composition of the material of interest.59 To compare the ablation rate of NIST SRM 610, ilvaite, hedenbergite, and quartz the 29Si pixel sensitivity was used. The composition of each mineral was calculated based on its stoichiometric formula. The pixel sensitivity for ilvaite and hedenbergite was in a range of ±15% compared to NIST SRM 610. Therefore, we assumed a similar ablation rate for ilvaite, hedenbergite and NIST SRM 610 as well as similar LODs. However, the pixel sensitivity for quartz was in average 60–70% lower than that of NIST SRM 610 resulting in approximately three times higher LODs compared to NIST SRM 610. This is assumed to be caused by a three times lower ablation rate of quartz.
The lower LODs obtained from the HD mode allowed measuring the composition of different minerals in great detail. For example, HD applied to ilvaite permitted to detect the distribution of In, whereas the SP mode creates a blurred distribution to the point that the edges of the ilvaite crystals were barely visible (LOD In: SP 4 mg kg−1, HD 0.3 mg kg−1, OL 0.3 mg kg−1).
The HD mode can also be used with an additional cleaning pulse (CP + HD) at each sampling position prior to hole drilling and data acquisition, which is specifically important when crossing phases or when re-deposition of ablated material occurs.48 To this end, a single cleaning pulse was applied by triggering the laser without ICP-TOFMS data acquisition. For this sample, an influence of surface contamination and the resulting element map was not observed. The CP + HD mode allowed an effective acquisition speed of 1.9–2.0 pixel per s. Further optimization of the CP + HD imaging procedure similar to the improvements in the HD mode allowed to speed up the effective pixel acquisition speed to 2.8–2.9 pixel per s.
The imaging control system allows also an overlapping line scan (OL) mode where each laser pulse can be acquired separately or as reported here summing the signal of 25 laser pulses to lower the LOD. The OL mode delivered LODs that were indistinguishable from those of the HD mode (Fig. 6) and an effective acquisition speed of 3.8–3.9 pixel per s, but was different in the effective resolution compared to the HD mode. The resolution of the OL mode in the line scan direction was slightly lower than that of the HD mode. This was caused by the overlap of the ablation spots of consecutive pixels. However, we stress that no data deconvolution was applied in our maps.31,33
The RM2 sample analyzed with the HD mode (Fig. 4) discloses lateral distributions of major, minor, and trace elements that are variable at the scale of 10 μm or less (1–2 pixels) and that were captured for mass fraction differences in the order of few mg kg−1 in neighboring areas of the crystals. This was shown by the relatively subtle changes of Al, Mn, Sn, In, and REEs mass fractions within ilvaite, which form alternating high and low mass fraction bands of 5–10 μm in width (Fig. 4). Notably, the distributions of Al, Sb, and alkali metals mass fractions within the quartz of RM2 indicate also relatively subtle changes. There were also thin and irregular bright bands within ilvaite corresponding to high Al, Mn, Na, Sn, Rb, In, and Ce mass fractions.
In summary, the element distribution maps constructed with the HD mode were able to capture multiple stages of mineral growth in geological samples, with which the lateral distribution of a large number of major, minor, and trace elements could be determined at low LODs. This gamut of very detailed compositional variations reflects the superior pixel sensitivity of the HD approach, which provided deeper insights into the crystallization histories of the solids.
To determine the effective area for analysis, a photomicrograph of the ROI of the sample is recorded and the size of the desired ablation area is measured in two orthogonal x–y directions. A black and white overlay is created using the photomicrograph. On a later step the actual ROI is dimensioned using the overlay, and its image is extracted from the surrounding area. Using the black and white overlay, the imaging control software calculates the ablation positions of the identified area, the laser spot size, and the image dimensions in the x and y directions. The data evaluation program uses this virtual mask to link the pixel data with the relative ablation positions, and generates the element distribution maps (Fig. 7). The shape adapted sample ablation was applied in a stepped sampling mode, during which the sample stage movement is controlled to enable precise positioning and is limited by the speed of the stage moving and by the feedback cycle. Therefore, a comparison was only made for the HD mode.
The ablation area fitted to the mineral shape enabled a reduced data acquisition time in this example by a factor of 2 compared to what would have been necessary by the imaging of a rectangular area of the same crystal in HD mode (605 pixels compared to 1332 pixels). The presented imaging of shape adapted ablation areas turned out particularly useful for small, non-rectangular or non-circular and even vein-like ROI, which could then be analyzed fit-to-shape.
The generation of composite spectra and further quantification of elements based on those spectra, and specific mineral assignments, are possible similar to the ‘inspector’ functionalities presented by Petrus et al. in the ‘Monocle’ interrogation software tool.44
Our data demonstrates that the single pulse ablation mode is well suited for fast imaging of major, minor, and a limited set of trace elements in mineral samples. Hole drilling imaging provides an order of magnitude improved LODs for most elements at constant lateral resolution. Therefore, the HD mode provides a more representative sampling result and can be considered as the method of choice for accurate quantification of a large number of major to trace elements in minerals. Furthermore, the fast data acquisition allows monitoring into depth to reveal multi-phase sampling.
We show further that a significant time saving step for elemental imaging could be the use of shape adapted ablation areas, which reduce the effective ablated area and therefore the duration of data acquisition. Triggering and binning of each ablation position allows to link the pixel data with the relative sample position and simplifies data evaluation and image generation.