Open Access Article
This Open Access Article is licensed under a Creative Commons Attribution-Non Commercial 3.0 Unported Licence

Recent advances in portable heavy metal electrochemical sensing platforms

Alejandro García-Miranda Ferrari a, Paul Carrington b, Samuel J. Rowley-Neale *a and Craig E. Banks *a
aFaculty of Science and Engineering, Manchester Metropolitan University, Chester Street, Manchester M1 5GD, UK. E-mail: s.rowley-neale@mmu.ac.uk; c.banks@mmu.ac.uk; Fax: +44(0)1612476831; Tel: +44(0)1612471196
bAquacheck Engineering Ltd, Royle Barn Road, Castleton, Rochdale, Lancashire, OL11 3DT, UK

Received 28th April 2020 , Accepted 10th July 2020

First published on 16th July 2020


Abstract

In 2020, the new European Drinking Water Directive will be updated to include new drinking-water safety parameters. These will introduce reduced thresholds for the presence of heavy metals, pesticides and other contaminants in domestic drinking water. The aim of these reduced thresholds is to impose a higher level of consumer protection as well as provide guidance for effective monitoring, planning and water management. An example of a newly proposed threshold is the concentration of lead ions (Pb2+) within drinking water, which has been reduced from 10 to 5 parts per billion (ppb). The strict timeframe for water suppliers to meet the new thresholds has created an impetus to develop in situ detection techniques to quantify the levels of chemical contamination on-site in drinking water, allowing them to make prompt interventions. This review explores and discusses the current advances related to electrochemical methods, electrode materials and modifications that have the potential to be the foundations of a new generation of portable electrochemical sensors towards trace-level ion in situ heavy metal sensors.



Water impact

The 2020 Drinking Water Directive (DWD) has legislated new lower thresholds for the heavy metal (HM) contamination within consumable/tap water. This has created an impetus for novel in situ HM sensors to be developed. This review highlights the new DWD HM detection limits and describes the parameters that have to be considered in order to build a suitable in situ electrochemical sensor.

1. Introduction

As the global population is increasingly found within urban conurbations, there is an ever-growing demand to resource these population centres with safe drinkable water. However, the aging/inadequate water infrastructure present within many cities results in severe problems, arising from contaminants (particularly heavy metals) entering the domestic water supply.1 These heavy metal contaminants not only occur due to leaching from the existing water infrastructure but are also a result of anthropogenic, agricultural, mining and industrial activities, which have drastically increased the levels of such metals within the environment.2,3 Heavy metals (HM) that commonly occur as contaminants within domestic water supplies are Hg2+, Pb2+, Cu2+ and Cd2+, although copper for example is considered nutritionally essential for human beings,4 heavy metals are toxic at high concentrations.

This toxicity of these heavy metals arises from their ability to bind with protein sites, displacing the original essential metals, bio-accumulating and ultimately becoming harmful. The detrimental effects that occur (post biological safe concentrations) typically affect the nervous system, kidney/liver function and on hard tissue such as bones and teeth.5–9 According to EU water quality reports, the main sources of heavy metal presence in drinking water are: its natural presence, pollution, waste water treatment and water distribution systems.10 HMs present in the water will enter small organisms and bio-accumulate within the food chain; in the case of humans there is also risk due to direct exposure/intake of HM.

The first attempt to limit the negative effect of HM water contamination was with water quality standards. These were introduced by the EU in 1980 when the Water Framework Directive (WFD) was established to ensure the protection of rivers, lakes, ground-waters, bathing waters, environmental organisations, nature and water-using sectors.11 In 1998, the Drinking Water Directive (DWD)12 was enacted to regulate the quality of water intended for human consumption. The DWD was amended in 2015 to include monitoring programs, parameters and sampling methods.13 As a result of the popular ‘Right2Water’ campaign, which gathered 1.8 million citizens' signatures, in 2020 the EU is renewing the DWD, aiming to update and control 48 parameters that must be monitored by water suppliers and regulators.14 These parameters include chemicals, such as heavy metals, pesticides, acrylamides etc. and microbiologicals (i.e. Clostridium perfringens spores, E. coli, somatic coliphages etc.).14Table 1 provides a comparison of heavy metal presence in drinking water between the current US Environmental Protection Agency (EPA)15 and European Union (EU)16 limits and the newly updated EU limits.14 Note that the permissible concentration of lead (Pb2+) and chromium (Cr3+) in drinking water, is set to be lowered from 10 down to 5 μg L−1 (ppb) and from 50 to 25 PPB respectively. These values are to be met no longer than 10 years after the initiation of this directive.14

Table 1 Comparison of current EPA, EU, and newly approved EU drinking water guidelines for selected parameters
Parameter EPA limit15 EU limit16 New EU limit 2020 (ref. 14) Unit Note
Antimony 6 5 5 μg L−1
Arsenic 10 10 10 μg L−1
Cadmium 5 5 5 μg L−1
Chromium 100 50 25 μg L−1 The value shall be met, at the latest, by [10 years after the entry into force of this directive]. The parametric value for chromium until that date is 50 μg L−1
Copper 1.3 2 2 mg L−1
Lead 15 10 5 μg L−1 The value shall be met at the latest by [10 years after the entry into force of this directive]. The parametric value for lead until that date is 10 μg L−1
After the transitional period, the value of 5 μg L−1 shall be met at least at the point of supply to the domestic distribution system
Mercury 2 1 1 μg L−1 For inorganic mercury
Nickel 100 20 20 μg L−1
Selenium 50 40 10 μg L−1 Parametric value of 30 μg L−1 shall be applied for regions where geological conditions may lead to high levels in ground water
Uranium 30 30 30 μg L−1


In order for countries to comply with the newly proposed contaminant thresholds, there is a requirement to employ the latest scientific methods/knowledge to create technological solutions that are able to reduce costs, preparation/analysis time and to improve the performance of in situ contaminants monitoring. As an example of HM drinking water contamination, copper (Cu2+) is an important parameter to monitor as it can leach from copper pipes due to the corrosive action of the oxygen present in the water. This can vary greatly depending on the pH, hardness and Cu2+ availability in the water distribution system,17 meaning that there is a need to identify areas, which are particularly at risk. High Cu2+ concentrations can lead to overt signs of contamination such as stains on sanitary appliances at levels above 1 parts per million (ppm), and a bitter taste to water at 5 ppm or higher.17

Lead (Pb2+), is recognised as an exceptional case because it arises in drinking water from the plumbing system. In the UK, the ubiquitous use of lead solder18 and fittings19 in properties built before 1970 (after which it was banned) presents a considerable challenge as the replacement and re-fitting of the systems would be cost/time prohibitive (the total lead fitting replacement was estimated to be around £7 billion only for UK properties20). The presence of Pb2+ in drinking water also depends on the pH, alkalinity and concentration of orthophosphates in the sample,21,22 reaffirming the need to quickly identify areas that are most at risk of high exposure (i.e. schools and hospitals) using affordable, mass-producible, in situ and accurate methods.

Typically, in order to accurately assess the levels of HM contamination with drinking water, large, sophisticated and expensive lab-based techniques such as atomic absorption spectrometry (AAS),23,24 mass spectrometry (MS),25 inductively coupled plasma MS (ICP-MS),25 atomic emission spectrometry (AES), X-ray fluorescence (XRF)26 and optical27 techniques are required in order to perform trace metal analysis. These techniques might also apply preconcentration and separation techniques (such as solid-phase micro- and normal extraction, dispersive liquid–liquid microextraction or cloud point extraction28), for the highly qualified technician, to be able to run the sample. Unfortunately, chemical, biological and physical reactions can occur during transportation, handling, pre-treatment and analysis of the water samples, which can affect the sample's composition and potentially result in inaccurate results.29 The development of new state-of-the-art in situ sensors is needed to reduce the time, cost and sampling of such monitoring studies in order to mitigate the problems highlighted above.27 Currently commercially available HM detection kits offer high sensitivities, but require image analysis (colorimetric analysis using microfluidic paper-based devices30) or the use of molecular recognition probes.31 The complexity and price of these commercial kits makes them impractical for large scale real time in situ environmental monitoring. Electrochemical methods on the other hand, can be considered one of the most suitable methods for in situ analysis of HM,32–34 mainly due to the small size of equipment, easy installation, low cost, simple sample preparation and the ability to perform multi-elemental detection. As shown in Fig. 1A, electrochemical detection uses a transducer to generate a measurable current, potential, charge, phase or frequency changes made by the electrochemical reaction taking place at the working electrode (WE)–electrolyte interface. Fig. 1B depicts a classic electrochemical cell comprised by the working, reference and counter electrodes (WE, RE and CE), respectively, where an electrochemical process called anodic stripping voltammetry is shown (Fig. 1B). Pad-printing, roll-to-roll and screen-printing electrode circuit fabrication methods have been applied to electrode manufacturing as a response to classic solid electrodes, by offering a reduction in manufacturing costs and the ability of performing laboratory-in-field experiments.35–38 In Fig. 2, one can observe the main differences between classic electrode platforms against those of screen-printed electrodes (SPEs), where the latter can be used as single-shot, disposable, reproducible and ready to use electrode. On the other hand, classic (solid) electrodes such as glassy carbon (GC), edge plane and basal plane pyrolytic graphite (EPPG and BPPG respectively) or highly ordered pyrolytic graphite (HOPG) (and others) need to be rigorously polished and cleaned before undertaking every measurement and require the presence of external RE and CE. Through offering a disposable, low-cost, reproducible and yet easily modified platform39,40 SPEs are one of the most used platforms for in situ electrochemical methods due to their versatility and can be applied towards a multitude of sensing devices towards a variety of targets such as gases,41,42 food quality,43,44 biomolecules45 or drugs of abuse,46 to name a few. SPEs have previously been played a pivotal role within the multi-billion dollar industry of the biological glucose sensing where they are used vital components of point-of-care devices. There disposable nature allows them to be cheaply used on a daily basis enabling individuals with diabetes to accurately check their blood glucose levels without having to visit a clinic, hospital or pharmacy.35,47


image file: d0ew00407c-f1.tif
Fig. 1 Schematic representation of electrochemical sensing (A) and a classic electrochemical cell/stripping voltammetric method (B).

image file: d0ew00407c-f2.tif
Fig. 2 Schematic overview comparing a classic electrochemical cell vs. a screen-printed electrode platform performing the analysis within a single drop of analyte.

2. Portable devices

There are several essential electronic components required to manufacture portable sensors, which thanks to the advances in computing power, electronics energy efficiency and manufacturing techniques have been able to be manufactured at a low-cost and within a small/portable package; thus making widespread small, fast and easy to use point-of-care (POC) testing more viable. A potentiostat is the most common electronic hardware used to control and measure electroanalytical experiments (adjusting/monitoring the signal, electrical current, voltage, resistance, etc.). Portable potentiostats are now widely available commercially as small, portable and low-cost devices that are applicable to in situ/on-line/point-of-care analysis. Open-source format technology has also reached electrochemists in the form of electric circuits, components, circuit boards, micro-controllers, software etc.48–52 In recent years, there has been a lot of development the miniaturization and increased portability of potentiostat devices. Researchers such as Moussy et al.53 have designed an implantable and dynamically configurable potentiostat for remote monitoring. Ainla et al.48 have recently simplified potentiostats, designing an of open-source universal battery powered and bluetooth connected potentiostat (UWED)48 capable of offering enough potential ranges for most electrochemical analysis in aqueous solutions.

3. Electroanalytical methods towards the detection of heavy metals in drinking water

Electroanalytical methods study the relationship between chemistry and electricity by measuring the resultant signal outputs (current, potential, charge, phase, frequency etc.) associated with a chemical reaction.54 The use of electrochemistry for analytical purposes includes quality control,55 environmental,56,57 food43,58 and biomedical45,59 applications to name a few. These are typically dynamic processes that occur at the electrode–electrolyte interface.60,61 Potentiostatic and potentiometric are the main two types of electroanalytical methods.

Potentiostatic methods study the transfer of charge at the electrode–electrolyte interface by measuring oxidation/reduction of species through the analysis of the observable current. This is undertaken by changing the potential between WE and RE electrodes, respectively. The main advantages of potentiostatic measurements over other detection techniques are high sensitivities, selectivity, microliter-size sample volumes, low limit of detections (LODs) and wide linear ranges.60,61 A widely use potentiostatic approach for HM ions determination is the use of stripping voltammetry.5,54,62 Stripping voltammetry can be mainly divided into anodic, cathodic or adsorptive stripping voltammetry (ASV, CSV or AdSV respectively).61 As shown in Fig. 3A, stripping analysis is based on a pre-concentration step on the surface of the WE followed by a stripping step, inducing Faradaic loss of the accumulated HM from the electrode (the HM ions go back into solution).61 Lastly, it measures the associated current, which is directly proportional to its concentration and pre-concentration time which can be generally expressed by the following equation:62

i = KCtpre-concentration
where the term K, is a constant that includes the electrode (geometric) area (A), scan rate (ν), diffusion coefficient (D), number of electrons transferred in the electrochemical process (n) and convection rate (usually applied using the deposition step). Electrochemical stripping metal detection offers low limits of detection within the ppb-range, multi-element determination, an associated low-cost and the ability to perform HM on-site/in situ determination.62 Stripping voltammetry generally detects the concentration of free HMs at the pH of the sample, unlike many other detection techniques, such as ICP-MS, which require to be strongly acidified to force all the HM ions into free state.63 It is also essential to note that an adequate WE is required for effective accumulation step and a favourable redox reaction of the target HM. Ideally the WE should be highly reproducible, display low background currents and have wide potential ranges.62


image file: d0ew00407c-f3.tif
Fig. 3 Schematic diagram of a potentiostatic (anodic stripping voltammetry; A) and a potentiometric (ion-selective electrode (ISE); B) method respectively.

Potentiometric methods are also an applicable methods, they are zero-current techniques in which the recorded response is the voltage across a membrane (electromotive force) that is directly related to the sample composition.54 In this case, different types of membranes, depending on their ion recognition capabilities have been commonly applied towards the detection of ionic species such as H+, Ca2+, F and K+ ions in complex matrixes.54,64,65Fig. 3B shows a typical schematic of an ion-selective electrode (ISE) performing a potentiometric method. ISEs are placed between two aqueous phases (sample and an inner electrolyte), a reference electrode (RE) as well as a salt bridge.66 ISEs offer rapid selective measurements, low energy consumption and low-cost, although LODs and multielement detection remain a challenge.67 The membrane, which can be polymeric, polycrystalline or made from glass,68 is the most fundamental part of the sensor and depending on the chosen membrane's properties, the sensor will have specific selectivity and performance. However, they exhibit ion interference and potential drift over time plus they need delicate handling, pre-measurement calibration.68 Pretsch et al.69 reported direct potentiometry at trace levels, a method capable of sub-femtomole LODs for Ca2+, Pb2+ and Ag2+, however its application to real samples with unknown composition remains a challenge, due to the difficulties in predicting the optimal applied current required.70 Cui et al.71 recently reported a μISE, that by taking advantage of microfabrication techniques, cleverly fabricating a multiplexed micro array comprised of different ionophores targeting Pb2+, Cd2+, Hg2+ and AsO2 capable of reaching ppb levels in drinking water.

4. Electrode materials

The electroanalytical output of the electrode is critically associated with the material of which the WE is comprised/made. The selection of an appropriate WE for a given analyte is therefore vital. Typically for electrochemical sensors one would choose a WE that has: an efficient electron transporter, high surface-to-volume-ratio, wide potential window, low background current, chemical stability and (electrochemically) interference-free.72,73 Historically speaking, hanging mercury drop electrodes (HMDE) were the most widely used WE material from the mid-20th century.62 HMDEs were followed by precious metals (such as polycrystalline Au, Ag and Pt) and carbon electrodes in the last three decades due to their low background current, wide potential range and chemical inertness.74,75

4.1 Metallic electrodes

Platinum (Pt) and gold (Au) are traditional electrode materials and have been utilised in a plethora of electrochemical reactions due to their excellent stability over a wide range of potentials and electrolytes whilst offering very favourable electron transfer kinetics.54,76 Au electrodes, display wide anodic polarization ranges77 and have conventionally been the optimal choice for mercury and arsenic determination.78,79 Copper (Cu), nickel (Ni) or silver (Ag) have also been used as electrode materials for sensing applications, however they are less applicable than Au in many electrochemical detection methodologies due to them forming high-valence oxy-hydroxide species acting as redox mediators.54,80,81 Cu is easily oxidised and offers poor long-term stability,82 therefore is not typically used as an WE. Cu is however low cost and widely used within industrial manufacturing and micro-fabrication methods, thereby making it application to niche applications.83 Last, bismuth film electrodes (BiFE) have been presented as a “green” alternative to mercury (Hg) films at the beginning of the century due to their similar performance to Hg but lower toxicity. BiFE and other metal film electrodes (PbFE, SnFE and SbFE) introduction to electrochemical detection was considered a landmark in the research of “green” electrode materials for stripping analysis towards sensing Cd2+, Pb2+, Tl2+, Cu2+ and Zn2+,84 although they need polishing and surface renewal steps prior to its use, which makes them highly dependent on the experimentalist.85,86 Banks et al.57 deposited Bi, Sb, Sn films on SPEs as well as classic electrode platforms (BDD and GC) and tested the platforms towards the detection of Cd2+ and Pb2+. Interestingly, they found that unmodified SPEs are capable of detecting both analytes at WHO levels, without any electrode pre-treatment (prior use, polishing and/or electrochemical treatment/surface- or bulk-modification).

4.2 Graphitic electrodes

Graphite, GC, nano-graphite, carbon black as well as a plethora of other graphitic materials have been widely reported as electrode materials for electrochemical applications.87–93 They offer excellent electrical conductivity, chemical stability, structural versatility, broad potential windows, abundant surface groups and robust carbon-to-carbon bonds (both internally and externally).94 In order to move away from large immobile testing equipment, research has focused on developing miniaturised, simple and low-cost electrodes. SPEs can be modified with chosen catalysts to make them applicable for niche applications.95 SPEs allow for increased portability, decentralised analysis and a reduction in the cost of sensors. They can also be easily modified with selected additives enabling them to display the beneficial properties of other materials. Graphitic materials such as carbon nanotubes (CNTs), graphene, graphene oxide (GO), carbon nanofibers (CNFs) and carbon nano-powders (CNPs) all have been successfully applied to SPEs, as modifications and/or inks for HM analysis in drinking water,96–101 river-waters102–105 and also food.106–108 However, as suggested by Ping et al., the synthesis and application of some 2D-nanomaterials to large-scale production might be sometimes impractical.109

4.3 Electrode modifications

The electrochemical sensitivity of some WE's can be enhanced not only by using 2D nanomaterials with enhanced properties, but also by using surface modifications to take advantage of the previously described beneficial properties of different materials. There are four typical categories of surface modifications, these being: adsorption, covalent bonding, electrochemical polymerization and electrochemical deposition.34,110 Adsorption methods consist of a non-covalent bond between the surface and the modified suspension, and can be sub-divided into: coatings, chemical adsorption and self-assembled monolayers. The most common adsorption method used is coating (i.e. drop-casting,111 spin-coating112 and dip-coating113), consists in covering the bare/unmodified WE's surface with a solvent-modifier mixture, the solvent then evaporates leaving a coat on the surface.34 Electro-polymerisation and deposition methods consist of passing a voltage thought the WE and the electrolyte, which contains the chosen modifier, to create stable polymers114 or metal115 films onto the WE's surface. Bulk modification of the SPE's ink with the same materials as described above, is a convenient mass-producible approach that is commonly explored within the literature to overcome the limitation of surface modification, such as poor cycling stability and uneven distribution by adding the chosen modification to the WE's ink. Mass-producible ink modification of SPE's has been widely reported towards sensing Cu2+,116 Pb2+[thin space (1/6-em)]56,117 Cd2+,56 and Zinc2+[thin space (1/6-em)]56 to name a few.

Molecular imprinted polymers (MIPs), or plastic antibodies, are bio-mimetic receptors based on cross-linking and functional monomers that are co-polymerised in the presence of the target analyte (the imprint molecule), acting as a template.118,119 Molecular imprinted polymers (IPs), as shown in Fig. 4, can be synthesised by a combination of monomers, cross-linkers, initiators and solvents, where the monomers polymerise around the target/imprint molecule with a high cross-link polymeric structure.120 Next, the removal of the imprint molecule reveals the scaffolds with exclusive non-covalent binding sites (molecular memory) to the target analyte.121 Imprinted polymer's main advantages are simple and low-cost preparation, high stability, affinity, selectivity, versatility of template/target molecule and ease of application as transducer in assays/sensors.122 They are currently applied to HM sensors,123 enzyme mimics,124 solid phase extractions,125 antibody mimics,126,127 biomarkers etc.128 Those MIPs that can specifically recognise and remove ions are called ion imprinted polymers (IIPs), these benefit from being broadly applicable, highly selective and water-compatible.129 IIP-based electrochemical sensors have been reported towards Co2+,130 Cd2+[thin space (1/6-em)]131,132 and others133 exhibiting ultra-trace levels in water samples.


image file: d0ew00407c-f4.tif
Fig. 4 Schematic representing MIPs applied to electrochemical sensor platforms.

Metal dopants/modifications (in the form of nanoparticles, films etc.) upon the surface of SPEs have also been widely studied concerning their application as sensors. Bismuth film electrodes (Bi-modified) SPEs has been reported to offer a simple, portable, mass-produce and yet sensitive and with and multi-element detection capabilities of HM such as Cd2+ and Pb2+.56,57,134 Bastos-Arrieta et al.135 recently reported that the main factor to enhance SPE's electrocatalytic performance is the shape of the nanoparticles, rather than the mere incorporation of them, when detecting Cd2+ or Pb2+ with Ag-NP SPEs. Barton et al.27 elegantly reviewed the use of Au-coated SPEs to detect HM, showing the determination of Pb2+, Cd2+, Hg2+, As2+, Cu2+ and Sb2+ in a variety of modifications such as Au bulk-modified inks, Au-coated, Au-sputters and nano-deposited, towards matrixes such as ground-,136 river-,137–139 rain-,140 waste-,138 sea-waters141–143 and soil extracts.144 Gold nanoparticles (AuNPs) modified graphitic SPEs have shown to be excellent alternatives to mercury electrodes as they exhibit high sensitivity and selectivity towards the simultaneous detection of Hg and Pb.145 For example Bernalte et al.146 reported LODs of 2.2, 1.5 and 1.3 ppb for the simultaneous determination of Pb2+, Cu2+ and Hg2+ respectively in in situ Amazonian waters by using bulk-film modified Au-SPEs. Cu-SPEs have also been demonstrated as point-of-care (POC) disposable sensors for Pb2+, Cd2+ and Zn2+,83 however the sensor cannot detect metals such as As and Ag due potential oxidation of Cu.83

The functionalization of an electrodes surface using ether crowns is also widely reported within the literature for Cu2+ (ref. 147) and Pb2+ (ref. 148–150) and other HMs151–153 sensing applications. Crown ethers are cyclic compounds with remarkable ability to form selective strong complexes with metal and organic cations.154,155 Del Valle et al.156 reported the simultaneous use of CB-15-crown-5 and CB-18-crown-6 for the Cd2+, Pb2+ and Cu2+ multi-elemental determination exhibiting ultra-trace sensing in environmental samples, when combined with chemometric tools to deconvolute the signals.

With the recent interest in applications for 2D materials, numerous studies have explored the utilisation of hybrid layered transition metal dichalcogenides (LTMDs) (such as MoS2, MoSe2, WS2etc.) with graphene/graphitic nanomaterials towards HM sensing applications.157,158 The combination of 2D materials with LTMDs allows the construction of heterojunctions that can be adjusted by changing the energy band of each semiconductor, which offers strong adsorption and fast electron transfer.158,159Fig. 5 shows the electrochemical steps that occur when TMDs are used within an electrochemical sensor. The HM ions (HMn+) freely flow in solution (A); a chosen deposition potential is applied (B) the HM pre-accumulates onto the highly adsorption sites (as HM0) of the TMDs (C); the stripping step applies a voltage re-dissolving (as HMn+) and the stripping signal will be recorded. Chen et al.158 recently reported a tailored 3D Ni/NiO/MoO3/chitosan heterojunction for the direct electrochemical detection of Cu2+. In this NiO and MoO2 both offer large band gaps and the chitosan (CS) exhibits exceptional Cu2+ adsorption capacity due to its abundance of hydroxyl and carboxyl reactive functional groups.160,161 The reported technique is a promising solution for the direct selective detection of HM in real water samples as the application of an interfacial barrier filters the solution and improves the electrochemical signal output. Another interesting study was performed by Huang et al.162 who reported an electrochemical sensitivity of 133.90 μA μM−1 when detecting Pb2+ using a Mn–MoS2 modified GC. This technique takes advantage of the interaction between Pb2+ and S atoms caused by tailored phases and defects at the nanosheet to offer high conductivity. Huang et al.162 reported that Pb–S bonding occurs due to the stimulated activity of S atoms, facilitating the electron transfer and promoting the in situ Pb reduction/re-oxidation at the Mn–MoS2 nanosheet.


image file: d0ew00407c-f5.tif
Fig. 5 Schematic representing the electrochemical steps when TMDS are used within an electrochemical sensor: pre-measurement (A), deposition potential application (B), pre-accumulation step (C) and stripping step (D).

4.4 Screen-printed electrode (SPEs) designs

The capability to shape an electrode's design has strong electrochemical performance implications. As represented in Fig. 6, changing the design from a macro- to a micro-electrode can change the profile of the diffusion regime for basic electron transfer (from hemi-spherical to planar and/or from non- to over-lapping), which can increase current densities, reduce ohmic drops and improve the ratio of faradaic-capacitive currents.35,40,163 Micro-electrode array sensors have been reported to exhibit lower LODs and greater sensitivities in comparison to macro-electrodes with equivalent geometrical areas, when single micro-electrodes are placed in a parallel configuration.35,164 This is due to the creation of diffusional independent voltammetric responses.163 The configuration of multiple single electrodes in parallel to each other takes advantage of radial diffusion dominating the mass transport, which results in a larger combined diffusional area than a single macro-electrode yielding improved electroanalytical sensing.165 Using multiple distinct also allows for simultaneous determination of analytes by using different electrode modifications on each WE, which can also save time and money in the case of critical situations.166–168 An elegant application of this was made by Rusling,169 who by modifying each electrode differently (different antibodies immobilised on each WE of the array) has reported the performance enhancement of microfluidic immuno-sensors,169,170 widening the application of these SPE platforms to a plethora of different target analytes at the same time; such an approach has the potential to be applied to HMs.
image file: d0ew00407c-f6.tif
Fig. 6 Schematic represensing the catergories of diffusion profiles at different micro-electrode array configurations. Planar diffusion layers at macro-electrodes (A), individual non-planar diffusion layers at micro-electrodes (B), non-overlapping diffusion layers at micro-electrodes (C) and planar overlapping diffusion layers at micro-electrodes (D).

Micro-electrode arrays and designs can be obtained via a wide range of processes such as sono-chemical/laser ablation, which can encounter challenges to seal the patterned material leading to leakages.40 Other methods include screen-printing technologies using small defined geometries.163 Further approaches include the fabrication of co-planar,171 micro-band electrodes,172 flexible micro-electrodes,173 ultra-micro-electrodes (UME)174,175 and micro-tubes176 to name a few. Zufelt et al.177 reported the fabrication of a photolithography fabricated array, which consisted of 18 WE's each with a geometric area of 0.0025 mm2. This array was ideal for ion sensing in aqueous solutions as it had the ability of multi-elemental monitoring conducting different electrochemical measurements in each single WE. The geometry and configuration of the WE plays an important role in sensors. Macro-electrode array/multiplexed designs have also been explored in the literature, offering the possibility of modifying each WE in a different way to detect different target analytes.178–183 The presence of multiple WEs using the same SPE also brings new challenges, such as an increase in internal circuit's resistance. This has been reported to contribute to frequency and severity of imprecisions and errors when electrode designs are not symmetrical.184Fig. 7 highlights the essential aspects that need to be considered when producing a sensor for a specific role such as heavy metal sensing in water. The plethora of modifications include the alteration and/or inclusion of active materials, catalyst, bulk modified recognition systems (added into screen-printing ink), plus the electrode/sensor design, and substrate of the electrochemical sensing platform. Considering the discussed research, we can conclude that SPE platforms offer a simple, customisable, cheap and highly sensitive approach to detect HMs. However, they still require improvements in terms of their reproducibility and interference avoidance from other metal, organic or inorganic soluble molecules present in the electrolyte at the same time.


image file: d0ew00407c-f7.tif
Fig. 7 Schematic representation of possible options for tailoring screen-printed electrochemical sensing platforms.

5. Conclusions

In situ monitoring of water quality at the source point of human consumption is vital in order to comply with the new lower permissible contaminant levels imposed by European directives. This review comprehensively introduces the parameters that one has to consider when developing HM electrochemical sensors capable of meeting the newly imposed LODs. The majority of the studies described within this manuscript offer sensing platforms capable of meeting these LODs. However, they typically require expensive electrode materials that can only be used within a lab, need reagents/chemicals and suffer from multi-elemental interferences. There are a limited number of studies that demonstrate a realistic approach of meeting the new EU regulations in a cost effective manner. It is therefore essential that research focuses on the development of portable/cost effective HM detection technologies in order to allow for facile, cheap and rapid in-field HM detection devices. The combination of electrochemical techniques, made possible by miniaturised and portable potentiostats, with cheap, calibration-less and tailorable screen-printed electrodes are promising candidates to support this transition. Further work is required regarding determination methods, composition/design of WE etc., in order to produce in situ sensor devices capable of meeting the new EU regulation requirements, quality targets and effectively monitor a wide variety of global health parameters that affect us all.

Conflicts of interest

The authors declare no competing financial and/or non-financial interests in relation to the work described herein.

Acknowledgements

This study is funded by Innovate UK (Reference: 11606) in cooperation with Aquacheck Ltd. Funding is also from the Engineering and Physical Sciences Research Council (Reference: EP/P007767/1 and EP/N0011877/1).

References

  1. R. I. McDonald, K. Weber, J. Padowski, M. Flörke, C. Schneider and P. A. Green, et al., Water on an urban planet: Urbanization and the reach of urban water infrastructure, Glob. Environ. Change, 2014, 27, 96–105 CrossRef.
  2. H. Kim, G. Jang and Y. Yoon, Specific heavy metal/metalloid sensors: current state and perspectives, Appl. Microbiol. Biotechnol., 2020, 104(3), 907–914 CrossRef CAS.
  3. P. B. Tchounwou, C. G. Yedjou, A. K. Patlolla and D. J. Sutton, Heavy metal toxicity and the environment, Exper. Suppl., 2012, 101, 133–164 Search PubMed.
  4. E. J. Underwood, Factors influencing trace element needs and tolerances in man, Mar. Pollut. Bull., 1974, 5(6), 86–88 CrossRef CAS.
  5. G. Aragay, J. Pons and A. Merkoçi, Recent Trends in Macro-, Micro-, and Nanomaterial-Based Tools and Strategies for Heavy-Metal Detection, Chem. Rev., 2011, 111(5), 3433–3458 CrossRef CAS PubMed.
  6. J. W. Hamilton, R. C. Kaltreider, O. V. Bajenova, M. A. Ihnat, J. McCaffrey and B. W. Turpie, et al., Molecular basis for effects of carcinogenic heavy metals on inducible gene expression, Environ. Health Perspect., 1998, 106(Suppl 4), 1005–1015 CrossRef CAS PubMed.
  7. B. L. Vallee and D. D. Ulmer, Biochemical effects of mercury, cadmium, and lead, Annu. Rev. Biochem., 1972, 41(10), 91–128 CrossRef CAS PubMed.
  8. T. Partanen, P. Heikkila, S. Hernberg, T. Kauppinen, G. Moneta and A. Ojajarvi, Renal cell cancer and occupational exposure to chemical agents, Scand. J. Work, Environ. Health, 1991, 17(4), 231–239 CrossRef CAS PubMed.
  9. M. Jaishankar, T. Tseten, N. Anbalagan, B. B. Mathew and K. N. Beeregowda, Toxicity, mechanism and health effects of some heavy metals, Interdiscip. Toxicol., 2014, 7(2), 60–72 Search PubMed.
  10. Commission E., Proposal for a Directive of the European Parliament and of the Council on the quality of water intended for human consumption, in Environment D-Gf, 2018, p. 134, https://eur-lex.europa.eu Search PubMed.
  11. Exploitation and marketing of natural mineral waters, 80/777/EEC, 1980 Search PubMed.
  12. Quality of water intended for human consumption, 98/83/EC, 1998 Search PubMed.
  13. Annexes on the quality of water intended for human consumption, 2015/1787, 2015 Search PubMed.
  14. ECI, ECI. Right2Water, 2020, Available from: https://www.right2water.eu/ Search PubMed.
  15. Drinking Water Standards and Regulations, EPA 822-F-18-001 Search PubMed.
  16. COUNCIL, DIRECTIVE 98/83/EC of 3 November 1998 on the quality of water intended for human consumption, 1998 Search PubMed.
  17. Organization WHO, in Guidelines for drinking-water quality, ed. W. Press, World Health Organization, 4th edn, 2011, p. 564 Search PubMed.
  18. T. D. B. Lyon and J. M. A. Lenihan, Corrosion in Solder Jointed Copper Tubes Resulting in Lead Contamination of Drinking Water, Br. Corros. J., 1977, 12(1), 41–45 CrossRef CAS.
  19. M. Zabochnicka-Świątek, Best Practice Guide on the Control of Lead in Drinking Water, 2010 Search PubMed.
  20. S. Potter, in Lead in Drinking Water, ed. Section SaE, House of Commons Library, 1997 Search PubMed.
  21. G. Korshin, J. Ferguson and A. Lancaster, Influence of natural organic matter on the corrosion of leaded brass in potable water, Corros. Sci., 2000, 42, 53–66 CrossRef CAS.
  22. E. J. Kim, J. E. Herrera, D. Huggins, J. Braam and S. Koshowski, Effect of pH on the concentrations of lead and trace contaminants in drinking water: a combined batch, pipe loop and sentinel home study, Water Res., 2011, 45(9), 2763–2774 CrossRef CAS PubMed.
  23. P. Pohl, Determination of metal content in honey by atomic absorption and emission spectrometries, TrAC, Trends Anal. Chem., 2009, 28, 117–128 CrossRef CAS.
  24. J. Gasparik, D. Vladarova, M. Capcarova, P. Smehyl, J. Slamecka and P. Garaj, et al., Concentration of lead, cadmium, mercury and arsenic in leg skeletal muscles of three species of wild birds, J. Environ. Sci. Health, Part A: Toxic/Hazard. Subst. Environ. Eng., 2010, 45(7), 818–823 CrossRef CAS PubMed.
  25. R. Flamini and A. Panighel, Mass spectrometry in grape and wine chemistry. Part II: The consumer protection, Mass Spectrom. Rev., 2006, 25(5), 741–774 CrossRef CAS PubMed.
  26. F. E. Mcneil and J. M. O'meara, The In Viva Measurement of Trace Heavy Metals by K X-Ray Fluorescence, 1999 Search PubMed.
  27. J. Barton, M. B. G. García, D. H. Santos, P. Fanjul-Bolado, A. Ribotti and M. McCaul, et al., Screen-printed electrodes for environmental monitoring of heavy metal ions: a review, Microchim. Acta, 2016, 183(2), 503–517 CrossRef CAS.
  28. G. Peng, Q. He, G. Zhou, Y. Li, X. Su and M. Liu, et al., Determination of heavy metals in water samples using dual-cloud point extraction coupled with inductively coupled plasma mass spectrometry, Anal. Methods, 2015, 7(16), 6732–6739 RSC.
  29. Z. Zou, A. Jang, E. T. MacKnight, P. Wu, J. Do and J. S. Shim, et al., An On-Site Heavy Metal Analyzer With Polymer Lab-on-a-Chips for Continuous Sampling and Monitoring, IEEE Sens. J., 2009, 9(5), 586–594 CAS.
  30. G. G. Morbioli, T. Mazzu-Nascimento, A. M. Stockton and E. Carrilho, Technical aspects and challenges of colorimetric detection with microfluidic paper-based analytical devices (μPADs) – A review, Anal. Chim. Acta, 2017, 970, 1–22 CrossRef CAS PubMed.
  31. N. Ullah, M. Mansha, I. Khan and A. Qurashi, Nanomaterial-based optical chemical sensors for the detection of heavy metals in water: Recent advances and challenges, TrAC, Trends Anal. Chem., 2018, 100, 155–166 CrossRef CAS.
  32. O. J. Guerra, J. Eichman, J. Kurtz and B.-M. Hodge, Cost Competitiveness of Electrolytic Hydrogen, Joule, 2019, 3(10), 2425–2443 CrossRef.
  33. G. E. Batley, Electroanalytical techniques for the determination of heavy metals in seawater, Mar. Chem., 1983, 12(2), 107–117 CrossRef CAS.
  34. Y. Lu, X. Liang, C. Niyungeko, J. Zhou, J. Xu and G. Tian, A review of the identification and detection of heavy metal ions in the environment by voltammetry, Talanta, 2018, 178, 324–338 CrossRef CAS PubMed.
  35. C. Banks, C. Foster and R. Kadara, Screen-Printing Electrochemical Architectures, 2016 Search PubMed.
  36. J. Willmann, D. Stocker and E. Dörsam, Characteristics and evaluation criteria of substrate-based manufacturing. Is roll-to-roll the best solution for printed electronics?, Org. Electron., 2014, 15(7), 1631–1640 CrossRef CAS.
  37. S. Bae, H. Kim, Y. Lee, X. Xu, J. S. Park and Y. Zheng, et al., Roll-to-roll production of 30-inch graphene films for transparent electrodes, Nat. Nanotechnol., 2010, 5(8), 574–578 CrossRef CAS PubMed.
  38. M. Hösel, R. R. Søndergaard, D. Angmo and F. C. Krebs, Comparison of Fast Roll-to-Roll Flexographic, Inkjet, Flatbed, and Rotary Screen Printing of Metal Back Electrodes for Polymer Solar Cells, Adv. Eng. Mater., 2013, 15(10), 995–1001 Search PubMed.
  39. A. García-Miranda Ferrari, C. W. Foster, P. Kelly, D. C. Brownson and C. E. Banks, Determination of the Electrochemical Area of Screen-Printed Electrochemical Sensing Platforms, Biosensors, 2018, 8(2), 53 CrossRef.
  40. J. P. Metters, R. Kadara and C. Banks, New directions in screen printed electroanalytical sensors: An overview of recent developments, Analyst, 2011, 136, 1067–1076 RSC.
  41. A. Morata, J. P. Viricelle, A. Tarancón, G. Dezanneau, C. Pijolat and F. Peiro, et al., Development and characterisation of a screen-printed mixed potential gas sensor, Sens. Actuators, B, 2008, 130(1), 561–566 CAS.
  42. J. Lee, G. Hussain, C. Banks and D. Silvester, Screen-Printed Graphite Electrodes as Low-Cost Devices for Oxygen Gas Detection in Room-Temperature Ionic Liquids, Sensors, 2017, 17(12), 2734 CrossRef PubMed.
  43. K. Ngamchuea, P. Hurst, C. Batchelor-McAuley and R. G. Compton, Handheld electrochemical device for the determination of the strength of garlic, Sens. Actuators, B, 2016, 232, 138–142 CrossRef CAS.
  44. R. Kachoosangi, G. G. Wildgoose and G. R. Compton, Carbon Nanotube-Based Electrochemical Sensors for Quantifying the ‘heat’ of Chilli Peppers: The Adsorptive Stripping Voltammetric Determination of Capsaicin, Analyst, 2008, 133, 888–895 RSC.
  45. N. Hernández-Ibáñez, L. García-Cruz, V. Montiel, C. W. Foster, C. E. Banks and J. Iniesta, Electrochemical lactate biosensor based upon chitosan/carbon nanotubes modified screen-printed graphite electrodes for the determination of lactate in embryonic cell cultures, Biosens. Bioelectron., 2016, 77, 1168–1174 CrossRef PubMed.
  46. H. M. Elbardisy, A. García-Miranda Ferarri, C. W. Foster, O. B. Sutcliffe, D. A. C. Brownson and T. S. Belal, et al. Forensic Electrochemistry: The Electroanalytical Sensing of Mephedrone Metabolites, ACS Omega, 2019, 4(1), 1947–1954 CrossRef CAS.
  47. A. Heller and B. Feldman, Electrochemical Glucose Sensors and Their Applications in Diabetes Management, Chem. Rev., 2008, 108(7), 2482–2505 CrossRef CAS PubMed.
  48. A. Ainla, M. P. S. Mousavi, M.-N. Tsaloglou, J. Redston, J. G. Bell and M. T. Fernández-Abedul, et al., Open-Source Potentiostat for Wireless Electrochemical Detection with Smartphones, Anal. Chem., 2018, 90(10), 6240–6246 CrossRef CAS PubMed.
  49. M. D. M. Dryden and A. R. Wheeler, DStat: A Versatile, Open-Source Potentiostat for Electroanalysis and Integration, PLoS One, 2015, 10(10), e0140349 CrossRef PubMed.
  50. A. A. Rowe, A. J. Bonham, R. J. White, M. P. Zimmer, R. J. Yadgar and T. M. Hobza, et al., CheapStat: An Open-Source, “Do-It-Yourself” Potentiostat for Analytical and Educational Applications, PLoS One, 2011, 6(9), e23783 CrossRef CAS PubMed.
  51. T. Dobbelaere, P. M. Vereecken and C. Detavernier, A USB-controlled potentiostat/galvanostat for thin-film battery characterization, HardwareX, 2017, 2, 34–49 CrossRef.
  52. P. Lopin and K. V. Lopin, PSoC-Stat: A single chip open source potentiostat based on a Programmable System on a Chip, PLoS One, 2018, 13(7), e0201353 CrossRef PubMed.
  53. R. Beach, R. Conlan, M. Godwin and F. Moussy, Towards a Miniature Implantable In Vivo Telemetry Monitoring System Dynamically Configurable as a Potentiostat or Galvanostat for Two- and Three-Electrode Biosensors, IEEE Trans. Instrum. Meas., 2005, 54, 61–72 CrossRef.
  54. J. Wang, Analytical electrochemistry, 2006 Search PubMed.
  55. R. Pemberton, A. Amine and J. Hart, Voltammetric Behavior of Chlorophyll a at a Screen-Printed Carbon Electrode and Its Potential Role as a Biomarker for Monitoring Fecal Contamination, Anal. Lett., 2007, 37, 1625–1643 CrossRef.
  56. R. O. Kadara, N. Jenkinson and C. E. Banks, Disposable Bismuth Oxide Screen Printed Electrodes for the High Throughput Screening of Heavy Metals, Electroanalysis, 2009, 21(22), 2410–2414 CAS.
  57. C. W. Foster, A. P. de Souza, J. P. Metters, M. Bertotti and C. E. Banks, Metallic modified (bismuth, antimony, tin and combinations thereof) film carbon electrodes, Analyst, 2015, 140(22), 7598–7612 RSC.
  58. E. P. Randviir, J. P. Metters, J. Stainton and C. E. Banks, Electrochemical impedance spectroscopy versus cyclic voltammetry for the electroanalytical sensing of capsaicin utilising screen printed carbon nanotube electrodes, Analyst, 2013, 138(10), 2970–2981 RSC.
  59. N. Hernández-Ibáñez, I. Sanjuán, M. Á. Montiel, C. W. Foster, C. E. Banks and J. Iniesta, l-Cysteine determination in embryo cell culture media using Co (II)-phthalocyanine modified disposable screen-printed electrodes, J. Electroanal. Chem., 2016, 780, 303–310 CrossRef.
  60. A. Bard and L. Faulkner, Electrochemical Methods: Fundamentals and Applications, John Wiley & Sons, Inc, 2001 Search PubMed.
  61. R. G. Compton and C. E. Banks, Understanding Voltammetry, Imperial College Press, 2nd edn, 2010 Search PubMed.
  62. J. Wang, Stripping Analysis, Encyclopedia of Electrochemistry, 2007 Search PubMed.
  63. A. J. Borrill, N. E. Reily and J. V. Macpherson, Addressing the practicalities of anodic stripping voltammetry for heavy metal detection: a tutorial review, Analyst, 2019, 144(23), 6834–6849 RSC.
  64. C. De Stefano, D. Milea, N. Porcino and S. Sammartano, Speciation of Phytate Ion in Aqueous Solution. Sequestering Ability toward Mercury(II) Cation in NaClaq at Different Ionic Strengths, J. Agric. Food Chem., 2006, 54(4), 1459–1466 CrossRef CAS PubMed.
  65. R. W. Cattrall, C. G. Gregorio, R. D. Webster, A. M. Bond and K. B. Oldham, Interaction between Silver and Other Heavy Metal Ions and the Ionophore S,S-Dipropyl Pyridine-2,6-dicarbothioate, Anal. Chem., 1997, 69(16), 3353–3359 CrossRef CAS.
  66. S. Makarychev-Mikhailov, A. Shvarev and E. Bakker, New trends in ion-selective electrodes, in Electrochemical Sensors, Biosensors and their Biomedical Applications, ed. X. Zhang, H. Ju and J. Wang, Academic Press, San Diego, 2008, ch. 4, pp. 71–114 Search PubMed.
  67. E. Malinowska, Z. Brzózka, K. Kasiura, R. J. M. Egberink and D. N. Reinhoudt, Lead selective electrodes based on thioamide functionalized calix[4]arenes as ionophores, Anal. Chim. Acta, 1994, 298(2), 253–258 CrossRef CAS.
  68. X. Tang, P.-Y. Wang and G. Buchter, Ion-Selective Electrodes for Detection of Lead (II) in Drinking Water: A Mini-Review, Environments, 2018, 5(9), 95 CrossRef.
  69. A. Malon, T. Vigassy, E. Bakker and E. Pretsch, Potentiometry at Trace Levels in Confined Samples: Ion-Selective Electrodes with Subfemtomole Detection Limits, J. Am. Chem. Soc., 2006, 128(25), 8154–8155 CrossRef CAS PubMed.
  70. L. Höfler, I. Bedlechowicz, T. Vigassy, R. E. Gyurcsányi, E. Bakker and E. Pretsch, Limitations of Current Polarization for Lowering the Detection Limit of Potentiometric Polymeric Membrane Sensors, Anal. Chem., 2009, 81(9), 3592–3599 CrossRef PubMed.
  71. R. You, P. Li, G. Jing and T. Cui, Ultrasensitive micro ion selective sensor arrays for multiplex heavy metal ions detection, Microsyst. Technol., 2019, 25(3), 845–849 CrossRef CAS.
  72. U. Yogeswaran and S.-M. Chen, A Review on the Electrochemical Sensors and Biosensors Composed of Nanowires as Sensing Material, Sensors, 2008, 8(1), 290–313 CrossRef CAS PubMed.
  73. I. Moldoveanu, R.-I. van Stefan-Staden and J. F. van Staden, Electrochemical Sensors Based on Nanostructured Materials, in Handbook of Nanoelectrochemistry: Electrochemical Synthesis Methods, Properties and Characterization Techniques, ed. M. Aliofkhazraei and A. S. H. Makhlouf, Springer International Publishing, Cham, 2016, pp. 1–15 Search PubMed.
  74. A. Hayat and J. L. Marty, Disposable Screen Printed Electrochemical Sensors: Tools for Environmental Monitoring, Sensors, 2014, 14(6), 10432–10453 CrossRef CAS PubMed.
  75. R. W. Murray, A. G. Ewing and R. A. Durst, Chemically modified electrodes. Molecular design for electroanalysis, Anal. Chem., 1987, 59(5), 379A–390A CAS.
  76. V. S. Bagotsky, Fundamentals of Electrochemistry, Wiley, 2005 Search PubMed.
  77. A. Economou, Screen-Printed Electrodes Modified with "Green" Metals for Electrochemical Stripping Analysis of Toxic Elements, Sensors, 2018, 18(4), 1032 CrossRef PubMed.
  78. D. Martín-Yerga, M. B. González-García and A. Costa-García, Electrochemical determination of mercury: A review, Talanta, 2013, 116, 1091–1104 CrossRef PubMed.
  79. D. E. Mays and A. Hussam, Voltammetric methods for determination and speciation of inorganic arsenic in the environment--a review, Anal. Chim. Acta, 2009, 646(1–2), 6–16 CrossRef CAS PubMed.
  80. L. A. Colon, R. Dadoo and R. N. Zare, Determination of carbohydrates by capillary zone electrophoresis with amperometric detection at a copper microelectrode, Anal. Chem., 1993, 65(4), 476–481 CrossRef CAS.
  81. I. G. Casella, T. R. I. Cataldi, A. M. Salvi and E. Desimoni, Electrocatalytic oxidation and liquid chromatographic detection of aliphatic alcohols at a nickel-based glassy carbon modified electrode, Anal. Chem., 1993, 65(21), 3143–3150 CrossRef CAS.
  82. A. J. de Bethune, Reference electrodes: Theory and practice, J. Chem. Educ., 1962, 39(2), A134 CrossRef.
  83. X. Pei, W. Kang, W. Yue, A. Bange, W. R. Heineman and I. Papautsky, Disposable Copper-Based Electrochemical Sensor for Anodic Stripping Voltammetry, Anal. Chem., 2014, 86(10), 4893–4900 CrossRef CAS PubMed.
  84. J. Wang, J. Lu, S. B. Hocevar, P. A. M. Farias and B. Ogorevc, Bismuth-Coated Carbon Electrodes for Anodic Stripping Voltammetry, Anal. Chem., 2000, 72(14), 3218–3222 CrossRef CAS PubMed.
  85. K. Bedin, E. Mitsuyasu, A. Ronix, A. L. Cazetta, O. Pezoti and V. Almeida, Inexpensive Bismuth-Film Electrode Supported on Pencil-Lead Graphite for Determination of Pb(II) and Cd(II) Ions by Anodic Stripping Voltammetry, Int. J. Anal. Chem., 2018, 2018, 1–9 CrossRef PubMed.
  86. A. Królicka and A. Bobrowski, Bismuth film electrode for adsorptive stripping voltammetry - Electrochemical and microscopic study, Electrochem. Commun., 2004, 99–104 CrossRef.
  87. D. A. C. Brownson, D. K. Kampouris and C. E. Banks, An overview of graphene in energy production and storage applications, J. Power Sources, 2011, 196(11), 4873–4885 CrossRef CAS.
  88. D. A. C. Brownson and C. E. Banks, Graphene electrochemistry: an overview of potential applications, Analyst, 2010, 135(11), 2768–2778 RSC.
  89. M. Pumera, Electrochemistry of graphene: new horizons for sensing and energy storage, Chem. Rec., 2009, 9(4), 211–223 CrossRef CAS PubMed.
  90. D. A. C. Brownson, P. J. Kelly and C. E. Banks, In situ electrochemical characterisation of graphene and various carbon-based electrode materials: an internal standard approach, RSC Adv., 2015, 5(47), 37281–37286 RSC.
  91. K. Zeng and D. Zhang, Recent progress in alkaline water electrolysis for hydrogen production and applications, Prog. Energy Combust. Sci., 2010, 36(3), 307–326 CrossRef CAS.
  92. W. M. Singh, T. Baine, S. Kudo, S. Tian, X. A. N. Ma and H. Zhou, et al., Electrocatalytic and Photocatalytic Hydrogen Production in Aqueous Solution by a Molecular Cobalt Complex, Angew. Chem., Int. Ed., 2012, 51(24), 5941–5944 CrossRef CAS PubMed.
  93. S. Srinivasan and F. J. Salzano, Prospects for hydrogen production by water electrolysis to be competitive with conventional methods, Int. J. Hydrogen Energy, 1977, 2(1), 53–59 CrossRef.
  94. R. L. McCreery, Advanced Carbon Electrode Materials for Molecular Electrochemistry, Chem. Rev., 2008, 108(7), 2646–2687 CrossRef CAS PubMed.
  95. A. García-Miranda Ferrari, O. Amor-Gutiérrez, E. Costa-Rama and M. T. Fernández-Abedul, Batch injection electroanalysis with stainless-steel pins as electrodes in single and multiplexed configurations, Sens. Actuators, B, 2017, 253, 1207–1213 CrossRef.
  96. L. Fu, X. Li, J. Yu and J. Ye, Facile and Simultaneous Stripping Determination of Zinc, Cadmium and Lead on Disposable Multiwalled Carbon Nanotubes Modified Screen-Printed Electrode, Electroanalysis, 2013, 25(2), 567–572 CrossRef CAS.
  97. S. Chaiyo, E. Mehmeti, K. Žagar, W. Siangproh, O. Chailapakul and K. Kalcher, Electrochemical sensors for the simultaneous determination of zinc, cadmium and lead using a Nafion/ionic liquid/graphene composite modified screen-printed carbon electrode, Anal. Chim. Acta, 2016, 918, 26–34 CrossRef CAS PubMed.
  98. C. Huangfu, L. Fu, Y. Li, X. Li, H. Du and J. Ye, Sensitive Stripping Determination of Cadmium(II) and Lead(II) on Disposable Graphene Modified Screen-Printed Electrode, Electroanalysis, 2013, 25(9), 2238–2243 CrossRef.
  99. S.-M. Choi, D.-M. Kim, O.-S. Jung and Y.-B. Shim, A disposable chronocoulometric sensor for heavy metal ions using a diaminoterthiophene-modified electrode doped with graphene oxide, Anal. Chim. Acta, 2015, 892, 77–84 CrossRef CAS PubMed.
  100. C. Pérez-Ràfols, N. Serrano, J. M. Díaz-Cruz, C. Ariño and M. Esteban, Simultaneous determination of Tl(I) and In(III) using a voltammetric sensor array, Sens. Actuators, B, 2017, 245, 18–24 CrossRef.
  101. C. Pérez-Ràfols, N. Serrano, J. M. Díaz-Cruz, C. Ariño and M. Esteban, A screen-printed voltammetric electronic tongue for the analysis of complex mixtures of metal ions, Sens. Actuators, B, 2017, 250, 393–401 CrossRef.
  102. H. Shuai, Graphene Ink Fabricated Screen Printed Electrode for Cd2+ and Pd2+ Determination in Xiangjiang River, Int. J. Electrochem. Sci., 2016, 7430–7439 CrossRef CAS.
  103. W. Feng, L. Hong-Wei, Y. Xin and C. Di-Zhao, Gs-nafion-Au nanocomposite film modified SPCEs for simultaneous determination of trace Pb2+ and Cd2+ by DPSV, Int. J. Electrochem. Sci., 2013, 8, 7702–7712 Search PubMed.
  104. N. Promphet, P. Rattanarat, R. Rangkupan, O. Chailapakul and N. Rodthongkum, An electrochemical sensor based on graphene/polyaniline/polystyrene nanoporous fibers modified electrode for simultaneous determination of lead and cadmium, Sens. Actuators, B, 2015, 207, 526–534 CrossRef CAS.
  105. R. María-Hormigos, M. J. Gismera, J. R. Procopio and M. T. Sevilla, Disposable screen-printed electrode modified with bismuth–PSS composites as high sensitive sensor for cadmium and lead determination, J. Electroanal. Chem., 2016, 767, 114–122 CrossRef.
  106. J. Ping, Y. Wang, J. Wu and Y. Ying, Development of an electrochemically reduced graphene oxide modified disposable bismuth film electrode and its application for stripping analysis of heavy metals in milk, Food Chem., 2014, 151, 65–71 CrossRef CAS PubMed.
  107. Y. Yao, H. Wu and J. Ping, Simultaneous determination of Cd(II) and Pb(II) ions in honey and milk samples using a single-walled carbon nanohorns modified screen-printed electrochemical sensor, Food Chem., 2019, 274, 8–15 CrossRef CAS PubMed.
  108. J.-M. Jian, Y.-Y. Liu, Y.-L. Zhang, X.-S. Guo and Q. Cai, Fast and Sensitive Detection of Pb2+ in Foods Using Disposable Screen-Printed Electrode Modified by Reduced Graphene Oxide, Sensors, 2013, 13(10), 13063–13075 CrossRef CAS PubMed.
  109. X. Liu, Y. Yao, Y. Ying and J. Ping, Recent advances in nanomaterial-enabled screen-printed electrochemical sensors for heavy metal detection, TrAC, Trends Anal. Chem., 2019, 115, 187–202 CrossRef CAS.
  110. A. J. Bard, Chemical modification of electrodes, J. Chem. Educ., 1983, 60(4), 302 CrossRef CAS.
  111. Y. R. Kim, S. Bong, Y. J. Kang, Y. Yang, R. K. Mahajan and J. S. Kim, et al., Electrochemical detection of dopamine in the presence of ascorbic acid using graphene modified electrodes, Biosens. Bioelectron., 2010, 25(10), 2366–2369 CrossRef CAS PubMed.
  112. M. Petit-Dominguez, H. Shen, W. Heineman and C. Seliskar, Electrochemical Behavior of Graphite Electrodes Modified by Spin-Coating with Sol−Gel-Entrapped Ionomers, Anal. Chem., 1997, 69(4), 703–710 CrossRef CAS.
  113. Y. Lu, R. Ganguli, C. A. Drewien, M. T. Anderson, C. Jeffrey Brinker and W. Gong, et al., Continuous formation of supported cubic and hexagonal mesoporous films by sol-gel dip-coating, Nature, 1997, 389(6649), 364–368 CrossRef CAS.
  114. M. A. Rahman, M.-S. Won and Y.-B. Shim, Characterization of an EDTA Bonded Conducting Polymer Modified Electrode: Its Application for the Simultaneous Determination of Heavy Metal Ions, Anal. Chem., 2003, 75(5), 1123–1129 CrossRef CAS PubMed.
  115. P. J. Mafa, A. O. Idris, N. Mabuba and O. A. Arotiba, Electrochemical co-detection of As(III), Hg(II) and Pb(II) on a bismuth modified exfoliated graphite electrode, Talanta, 2016, 153, 99–106 CrossRef CAS PubMed.
  116. M. Somasundrum and J. V. Bannister, Amperometric determination of copper using screen-printed electrodes, Sens. Actuators, B, 1993, 15(1), 203–208 CrossRef CAS.
  117. I. Palchetti, C. Upjohn, A. P. F. Turner and M. Mascini, Disposable Screen-Printed Electrodes (Spe) Mercury-Free for Lead Detection, Anal. Lett., 2000, 33(7), 1231–1246 CrossRef CAS.
  118. K. Haupt and K. Mosbach, Molecularly Imprinted Polymers and Their Use in Biomimetic Sensors, Chem. Rev., 2000, 100(7), 2495–2504 CrossRef CAS PubMed.
  119. A. G. Mayes and M. J. Whitcombe, Synthetic strategies for the generation of molecularly imprinted organic polymers, Adv. Drug Delivery Rev., 2005, 57(12), 1742–1778 CrossRef CAS PubMed.
  120. G. Wulff, Molecular Imprinting in Cross-Linked Materials with the Aid of Molecular Templates— A Way towards Artificial Antibodies, Angew. Chem., Int. Ed. Engl., 1995, 34(17), 1812–1832 CrossRef CAS.
  121. J. Wackerlig and R. Schirhagl, Applications of Molecularly Imprinted Polymer Nanoparticles and Their Advances toward Industrial Use: A Review, Anal. Chem., 2016, 88(1), 250–261 CrossRef CAS PubMed.
  122. O. S. Ahmad, T. S. Bedwell, C. Esen, A. Garcia-Cruz and S. A. Piletsky, Molecularly Imprinted Polymers in Electrochemical and Optical Sensors, Trends Biotechnol., 2019, 37(3), 294–309 CrossRef CAS PubMed.
  123. X. Luo, W. Huang, Q. Shi, W. Xu, Y. Luan and Y. Yang, et al., Electrochemical sensor based on lead ion-imprinted polymer particles for ultra-trace determination of lead ions in different real samples, RSC Adv., 2017, 7(26), 16033–16040 RSC.
  124. K. Severin, Applications of Molecularly Imprinted Materials as Enzyme Mimics, in Molecularly Imprinted Materials: Science and Technology, 2004 Search PubMed.
  125. L.-X. Yi, R. Fang and G.-H. Chen, Molecularly Imprinted Solid-Phase Extraction in the Analysis of Agrochemicals, J. Chromatogr. Sci., 2013, 51(7), 608–618 CAS.
  126. J. Pan, W. Chen, Y. Ma and G. Pan, Molecularly imprinted polymers as receptor mimics for selective cell recognition, Chem. Soc. Rev., 2018, 47(15), 5574–5587 RSC.
  127. Y. Md, Molecularly imprinted polymers as antibody mimics: Applications in immunoassays and recent developments, J. Clin. Ligand Assay, 2002, 25, 234–236 Search PubMed.
  128. Y. Saylan, S. Akgönüllü, H. Yavuz, S. Ünal and A. Denizli, Molecularly Imprinted Polymer Based Sensors for Medical Applications, Sensors, 2019, 19(6), 1279 CrossRef CAS.
  129. X. Dai, S. Wu and S. Li, Progress on electrochemical sensors for the determination of heavy metal ions from contaminated water, J. Chin. Adv. Mater. Soc., 2018, 6(2), 91–111 CrossRef CAS.
  130. M. Torkashvand, M. B. Gholivand and R. Azizi, Synthesis, characterization and application of a novel ion-imprinted polymer based voltammetric sensor for selective extraction and trace determination of cobalt (II) ions, Sens. Actuators, B, 2017, 243, 283–291 CrossRef CAS.
  131. K. Huang, Y. Chen, F. Zhou, X. Zhao, J. Liu and S. Mei, et al., Integrated ion imprinted polymers-paper composites for selective and sensitive detection of Cd(II) ions, J. Hazard. Mater., 2017, 333, 137–143 CrossRef CAS PubMed.
  132. S. A. Rezvani Ivari, A. Darroudi, M. H. Arbab Zavar, G. Zohuri and N. Ashraf, Ion imprinted polymer based potentiometric sensor for the trace determination of Cadmium (II) ions, Arabian J. Chem., 2017, 10, S864–S869 CrossRef CAS.
  133. S. Hu, X. Xiong, S. Huang and X. Lai, Preparation of Pb(II) Ion Imprinted Polymer and Its Application as the Interface of an Electrochemical Sensor for Trace Lead Determination, Anal. Sci., 2016, 32(9), 975–980 CrossRef CAS PubMed.
  134. H. Wang, G. Zhao, Y. Yin, Z. Wang and G. Liu, Screen-Printed Electrode Modified by Bismuth/Fe3O4 Nanoparticle/Ionic Liquid Composite Using Internal Standard Normalization for Accurate Determination of Cd(II) in Soil, Sensors, 2018, 18(1), 6 CrossRef PubMed.
  135. K. Torres-Rivero, L. Torralba-Cadena, A. Espriu-Gascon, I. Casas, J. Bastos-Arrieta and A. Florido, Strategies for Surface Modification with Ag-Shaped Nanoparticles: Electrocatalytic Enhancement of Screen-Printed Electrodes for the Detection of Heavy Metals, Sensors, 2019, 19(19), 4249 CrossRef CAS PubMed.
  136. Y.-S. Song, G. Muthuraman, Y.-Z. Chen, C.-C. Lin and J.-M. Zen, Screen Printed Carbon Electrode Modified with Poly(L-Lactide) Stabilized Gold Nanoparticles for Sensitive As(III) Detection, Electroanalysis, 2006, 18(18), 1763–1770 CrossRef CAS.
  137. S. Laschi, I. Palchetti and M. Mascini, Gold-based screen-printed sensor for detection of trace lead, Sens. Actuators, B, 2006, 114(1), 460–465 CrossRef CAS.
  138. X. Niu, C. Chen, Y. Teng, H. Zhao and M. Lan, Novel Screen-Printed Gold Nano Film Electrode for Trace Mercury(II) Determination Using Anodic Stripping Voltammetry, Anal. Lett., 2012, 45(7), 764–773 CrossRef CAS.
  139. L. Zhang, D. Li, W. Song, L. Shi, Y. Li and Y. Long, High Sensitive On-Site Cadmium Sensor Based on AuNPs Amalgam Modified Screen-Printed Carbon Electrodes, IEEE Sens. J., 2010, 10(10), 1583–1588 CAS.
  140. E. Bernalte, C. M. Sánchez and E. P. Gil, Determination of mercury in ambient water samples by anodic stripping voltammetry on screen-printed gold electrodes, Anal. Chim. Acta, 2011, 689(1), 60–64 CrossRef CAS PubMed.
  141. E. Punrat, S. Chuanuwatanakul, T. Kaneta, S. Motomizu and O. Chailapakul, Method development for the determination of mercury(II) by sequential injection/anodic stripping voltammetry using an in situ gold-film screen-printed carbon electrode, J. Electroanal. Chem., 2014, 727, 78–83 CrossRef CAS.
  142. O. Domínguez Renedo and M. J. Arcos Martínez, Anodic stripping voltammetry of antimony using gold nanoparticle-modified carbon screen-printed electrodes, Anal. Chim. Acta, 2007, 589(2), 255–260 CrossRef PubMed.
  143. K. Malzahn, J. R. Windmiller, G. Valdés-Ramírez, M. J. Schöning and J. Wang, Wearable electrochemical sensors for in situ analysis in marine environments, Analyst, 2011, 136(14), 2912–2917 RSC.
  144. M. F. Noh and I. E. Tothill, Development and characterisation of disposable gold electrodes, and their use for lead(II) analysis, Anal. Bioanal. Chem., 2006, 386(7–8), 2095–2106 CrossRef CAS PubMed.
  145. I. Ait Tayeb and K. Razak, Development of Gold Nanoparticles Modified Electrodes for the Detection of Heavy Metal Ions, J. Phys.: Conf. Ser., 2018, 1083, 012044 CrossRef.
  146. E. Bernalte, S. Arévalo, J. Pérez-Taborda, J. Wenk, P. Estrela and A. Avila, et al., Rapid and on-site simultaneous electrochemical detection of copper, lead and mercury in the Amazon river, Sens. Actuators, B, 2020, 307, 127620 CrossRef.
  147. V. S. Ijeri and A. K. Srivastava, Voltammetric determination of copper at chemically modified electrodes based on crown ethers, Fresenius' J. Anal. Chem., 2000, 367(4), 373–377 CrossRef CAS PubMed.
  148. V. S. Ijeri and A. K. Srivastava, Voltammetric Determination of Lead at Chemically Modified Electrodes Based on Crown Ethers, Anal. Sci., 2001, 17(5), 605–608 CrossRef CAS PubMed.
  149. S. V. Prabhu, R. P. Baldwin and L. Kryger, Preconcentration and determination of lead(II) at crown ether and cryptand containing chemically modified electrodes, Electroanalysis, 1989, 1(1), 13–21 CrossRef CAS.
  150. I.-M. Simionca, A. Arvinte and M. Pinteala, Crown ether-based structures for sensitive electrochemical detection, High Perform. Polym., 2015, 27(5), 669–675 CrossRef CAS.
  151. G. Zhiqiang, L. Peibiao and Z. Zaofan, Anodic stripping voltammetric determination of traces of mercury with a chemically modified electrode, Microchem. J., 1991, 43(2), 121–132 CrossRef.
  152. M. V. Tsymbal, I. Y. Tuijan, Z. A. Temerdashev and K. Z. Brainina, Crown ethers in stripping voltammetry of palladium, Electroanalysis, 1994, 6(2), 113–117 CrossRef CAS.
  153. S. Tanaka and H. Yoshida, Stripping voltammetry of silver(I) with a carbon-paste electrode modified with thiacrown compounds, Talanta, 1989, 36(10), 1044–1046 CrossRef CAS PubMed.
  154. Z. Huang and S. Chang, Synthesis of Unique Ionophores of penta-Crown Ethers, Synlett, 2005, 2005, 2257–2259 Search PubMed.
  155. H. An, J. S. Bradshaw, R. M. Izatt and Z. Yan, Bis- and Oligo(benzocrown ether)s, Chem. Rev., 1994, 94(4), 939–991 CrossRef CAS.
  156. N. Serrano, A. González-Calabuig and M. del Valle, Crown ether-modified electrodes for the simultaneous stripping voltammetric determination of Cd(II), Pb(II) and Cu(II), Talanta, 2015, 138, 130–137 CrossRef CAS PubMed.
  157. Y.-F. Sun, J.-H. Sun, J. Wang, Z.-X. Pi, L.-C. Wang and M. Yang, et al., Sensitive and anti-interference stripping voltammetry analysis of Pb(II) in water using flower-like MoS2/rGO composite with ultra-thin nanosheets, Anal. Chim. Acta, 2019, 1063, 64–74 CrossRef CAS PubMed.
  158. J. Shang, M. Zhao, H. Qu, H. Li, R. Gao and S. Chen, New application of p-n junction in electrochemical detection: The detection of heavy metal ions, J. Electroanal. Chem., 2019, 855, 113624 CrossRef CAS.
  159. N. Baghban, E. Yilmaz and M. Soylak, A magnetic MoS2-Fe3O4 nanocomposite as an effective adsorbent for dispersive solid-phase microextraction of lead(II) and copper(II) prior to their determination by FAAS, Microchim. Acta, 2017, 184(10), 3969–3976 CrossRef CAS.
  160. B. Wang, Z. Bai, H. Jiang, P. Prinsen, R. Luque and S. Zhao, et al., Selective heavy metal removal and water purification by microfluidically-generated chitosan microspheres: Characteristics, modeling and application, J. Hazard. Mater., 2019, 364, 192–205 CrossRef CAS PubMed.
  161. Y. Fu, L. Qin, D. Huang, G. Zeng, C. Lai and B. Li, et al., Chitosan functionalized activated coke for Au nanoparticles anchoring: Green synthesis and catalytic activities in hydrogenation of nitrophenols and azo dyes, Appl. Catal., B, 2019, 255, 117740 CrossRef CAS.
  162. W.-Y. Zhou, S.-S. Li, X.-Y. Xiao, S.-H. Chen, J.-H. Liu and X.-J. Huang, Defect- and phase-engineering of Mn-mediated MoS2 nanosheets for ultrahigh electrochemical sensing of heavy metal ions: chemical interaction-driven in situ catalytic redox reactions, Chem. Commun., 2018, 54(67), 9329–9332 RSC.
  163. R. O. Kadara, N. Jenkinson and C. E. Banks, Screen printed recessed microelectrode arrays, Sens. Actuators, B, 2009, 142(1), 342–346 CrossRef CAS.
  164. T. J. Davies, S. Ward-Jones, C. E. Banks, J. del Campo, R. Mas and F. X. Muñoz, et al. The cyclic and linear sweep voltammetry of regular arrays of microdisc electrodes: Fitting of experimental data, J. Electroanal. Chem., 2005, 585(1), 51–62 CrossRef CAS.
  165. P. N. Bartlett and S. L. Taylor, An accurate microdisc simulation model for recessed microdisc electrodes, J. Electroanal. Chem., 1998, 453(1), 49–60 CrossRef CAS.
  166. D. Ortiz-Aguayo, M. Bonet-San-Emeterio and M. del Valle, Simultaneous Voltammetric Determination of Acetaminophen, Ascorbic Acid and Uric Acid by Use of Integrated Array of Screen-Printed Electrodes and Chemometric Tools, Sensors, 2019, 19(15), 3286 CrossRef CAS PubMed.
  167. V. Ruiz-Valdepenas Montiel, E. Povedano, S. Benede, L. Mata, P. Galan-Malo and M. Gamella, et al., Disposable Amperometric Immunosensor for the Detection of Adulteration in Milk through Single or Multiplexed Determination of Bovine, Ovine, or Caprine Immunoglobulins G, Anal. Chem., 2019, 91(17), 11266–11274 CrossRef CAS PubMed.
  168. E. C. Rama and A. Costa-García, Screen-printed Electrochemical Immunosensors for the Detection of Cancer and Cardiovascular Biomarkers, Electroanalysis, 2016, 28(8), 1700–1715 CrossRef CAS.
  169. M. Sharafeldin, K. McCaffrey and J. F. Rusling, Influence of antibody immobilization strategy on carbon electrode immunoarrays, Analyst, 2019, 144(17), 5108–5116 RSC.
  170. F. Crespilho, Nanobioelectrochemistry, 2013 Search PubMed.
  171. J. P. Metters, R. O. Kadara and C. E. Banks, Fabrication of co-planar screen printed microband electrodes, Analyst, 2013, 138(9), 2516–2521 RSC.
  172. D. H. Craston, C. P. Jones, D. E. Williams and N. El Murr, Microband electrodes fabricated by screen printing processes: Applications in electroanalysis, Talanta, 1991, 38(1), 17–26 CrossRef CAS PubMed.
  173. R. O. Kadara, N. Jenkinson and C. E. Banks, Characterization and fabrication of disposable screen printed microelectrodes, Electrochem. Commun., 2009, 11(7), 1377–1380 CrossRef CAS.
  174. W. C. Chen, Y. L. Hsu, S. Venkatesan and J. M. Zen, Disposable Screen-Printed Edge Band Ultramicroelectrodes for Use as Nitric Oxide Gas Sensor in Designing an Easily Applicable Method for Real Sample Analysis of Nitrite with Superior Selectivity and Sensitivity, Electroanalysis, 2014, 26, 565–572 CrossRef CAS.
  175. J. Wang, J. Lu, B. Tian and C. Yarnitzky, Screen-printed ultramicroelectrode arrays for on-site stripping measurements of trace metals, J. Electroanal. Chem., 1993, 361(1), 77–83 CrossRef CAS.
  176. N. G. Karousos, R. Carrington, S. J. Wilkins, C. Livingstone and J. Davis, Molecular sieving of anti-oxidants at a copper–carbon laminate assembly, Electrochem. Commun., 2005, 7(5), 500–504 CrossRef CAS.
  177. R. D. Gardner, A. Zhou and N. A. Zufelt, Development of a Microelectrode Array Sensing Platform for Combination Electrochemical and Spectrochemical Aqueous Ion Testing, Sens. Actuators, B, 2009, 136(1), 177–185 CrossRef CAS PubMed.
  178. E. Sánchez-Tirado, C. Salvo, A. González-Cortés, P. Yáñez-Sedeño, F. Langa and J. M. Pingarrón, Electrochemical immunosensor for simultaneous determination of interleukin-1 beta and tumor necrosis factor alpha in serum and saliva using dual screen printed electrodes modified with functionalized double–walled carbon nanotubes, Anal. Chim. Acta, 2017, 959, 66–73 CrossRef PubMed.
  179. V. Escamilla-Gómez, D. Hernández-Santos, M. B. González-García, J. M. Pingarrón-Carrazón and A. Costa-García, Simultaneous detection of free and total prostate specific antigen on a screen-printed electrochemical dual sensor, Biosens. Bioelectron., 2009, 24(8), 2678–2683 CrossRef PubMed.
  180. F. Conzuelo, S. Campuzano, M. Gamella, D. G. Pinacho, A. J. Reviejo and M. P. Marco, et al., Integrated disposable electrochemical immunosensors for the simultaneous determination of sulfonamide and tetracycline antibiotics residues in milk, Biosens. Bioelectron., 2013, 50, 100–105 CrossRef CAS PubMed.
  181. J. Pilas, T. Selmer, M. Keusgen and M. J. Schöning, Screen-Printed Carbon Electrodes Modified with Graphene Oxide for the Design of a Reagent-Free NAD+-Dependent Biosensor Array, Anal. Chem., 2019, 91(23), 15293–15299 CrossRef CAS PubMed.
  182. S. K. Tuteja, C. Ormsby and S. Neethirajan, Noninvasive Label-Free Detection of Cortisol and Lactate Using Graphene Embedded Screen-Printed Electrode, Nano-Micro Lett., 2018, 10(3), 41 CrossRef PubMed.
  183. D. Xie, C. Li, L. Shangguan, H. Qi, D. Xue and Q. Gao, et al., Click chemistry-assisted self-assembly of DNA aptamer on gold nanoparticles-modified screen-printed carbon electrodes for label-free electrochemical aptasensor, Sens. Actuators, B, 2014, 192, 558–564 CrossRef CAS.
  184. R. García-González, A. Costa-García and M. T. Fernández-Abedul, Dual screen-printed electrodes with elliptic working electrodes arranged in parallel or perpendicular to the strip, Sens. Actuators, B, 2014, 198, 302–308 CrossRef.

This journal is © The Royal Society of Chemistry 2020