James H.
Bannock
ab,
Tsz Yin (Martin)
Lui
a,
Simon T.
Turner
a and
John C.
deMello
*ab
aDepartment of Chemistry, Imperial College London, Exhibition Road, London SW7 2AY, UK. E-mail: j.demello@imperial.ac.uk
bCentre for Organic Electronic Materials, Department of Chemistry, NTNU, N-7491 Trondheim, Norway. E-mail: john.demello@ntnu.no
First published on 16th May 2018
We report an automated procedure for the inline separation of two immiscible liquids based on a porous polytetrafluoroethylene (PTFE) capillary and a small number of inexpensive electronic components. By monitoring the light transmitted through fluid streams at the two outlets of the separator and iteratively adjusting a needle-valve located at one outlet until smooth time-invariant signals are observed at both outlets, the separator is capable of establishing complete liquid/liquid separation within minutes. Using mixtures of water and toluene as a test system, near quantitative recovery of the two liquids was achieved over a wide range of flow conditions without detectable cross contamination at either outlet. In a twenty-four hour test run, departures from complete separation occurred just three times, and on each occasion complete separation was automatically restored within ninety seconds. Further tests on other liquid/liquid mixtures showed that the automated separator is capable of rapidly and reliably inducing the separation of aqueous–organic, aqueous–fluorous and organic–fluorous mixtures, making it a versatile tool for numerous applications in fluidic analysis, synthesis and purification.
In many cases, it subsequently becomes necessary to separate the immiscible liquids, while keeping at least one of the liquids flowing in a stable and controlled manner. Key applications of liquid–liquid separation include: (i) multistep chemical processing, where an intermediate purification or quench is required prior to further downstream processing;8 (ii) inline analysis, where switching to continuous flow can greatly simplify detection by removing the need for sophisticated detectors synchronised to the segmented solvent flow;9 and (iii) physical removal of the unwanted phase (and any impurities contained therein) prior to collection of the product.10
In microscale systems, phase separation is most effectively achieved using wetting-based methods that exploit differences in the tendency of the two liquids to wet a surface or membrane. One approach is to use micro-engineered structures to induce phase separation and coerce the two liquids into following separate exit paths as a result of one liquid maximizing and the other minimizing its contact with an exposed surface.11–19 Another approach, exploited in some commercial systems, is to use porous, planar membranes that can be selectively permeated by one of the two phases.5,20–24 When such membranes are used in conjunction with additional componentry to balance the pressure either side of the membrane, separation can be achieved at relatively high flow rates of up to 10 mL min−124 (please see ref. 9 and 25 for recent reviews of methods for separating immiscible liquids on the microscale).
We recently reported an alternative method for inline separation of immiscible liquids using a commercially sourced porous polytetrafluoroethylene (PTFE) capillary.9,26 Injection of a two-phase flow into the porous capillary causes one phase to preferentially permeate the capillary wall, leaving a continuous stream of the other phase to pass out of the capillary outlet. Insertion at the capillary outlet of a narrow flow restriction of appropriately chosen length and diameter allows the back pressure to be tuned for different flow conditions and liquid–liquid combinations, including aqueous–organic, aqueous–fluorous and organic–fluorous mixtures.
In this paper we demonstrate how a porous capillary can form the basis of a fully automated system for liquid/liquid separation that operates by optically monitoring the fluid streams extracted at the two outlets and iteratively modifying the back pressure downstream of the capillary to establish and subsequently maintain complete separation. Although simple in design, the automated separator induces rapid and reliable separation of a broad range of immiscible liquids, and should prove effective for many applications in microscale synthesis and analysis.
The principle of a porous capillary based separator is straightforward. The two-phase fluid stream enters the inlet of the porous capillary. The carrier preferentially wets and subsequently permeates the porous wall, accumulating on the exterior until it is of sufficient weight to drip from the capillary into a collection vial. This process repeats, with new carrier liquid collecting on the exterior of the porous capillary until the next drip occurs, thereby allowing the carrier to be extracted continuously from the channel without any drop in separation efficiency. A continuous (single-phase) stream of solvent is left flowing through the porous capillary and emerges at the outlet. The outflowing solvent may then be transferred to a vial for collection or passed into the next stage of a multistep chemical process as required. By encasing the porous capillary in an outer jacket, the carrier may be coerced into a continuous stream, allowing it too to be subjected to further flow processing. In this way, continuous separated streams of the two phases may be obtained at the separator outlets, see Methods and Fig. 1a.
![]() | ||
Fig. 1 Principle of capillary-based liquid–liquid separation. (a) The segmented flow is passed into a short length of porous PTFE tubing, causing the carrier to preferentially wet and pass through the capillary wall into a jacket surrounding the capillary, where it is coerced into a continuous stream and extracted via a side-channel (S); the solvent passes through the porous capillary unaffected and emerges as a continuous stream via the through-channel (T); (b) automation is achieved by monitoring each outgoing fluid stream with a constant intensity light-emitting diode and a photodiode, and iteratively adjusting a motorized needle-valve at T until smooth non-fluctuating signals are obtained at both outlets (see scheme in Fig. 4). |
In the following discussion, we use the term through-channel (T) to describe the outlet at the end of the porous capillary, and the term side-channel (S) to describe the outlet connected to the wall of the porous capillary. To achieve complete separation of the two liquids, the back pressure at T must be carefully controlled: if it is too high, a fraction of the solvent will be forced through the capillary walls, leading to its incomplete recovery at T; if it is too low, a fraction of the carrier will pass through the entire length of the porous tubing without being depleted through the walls, causing a mixture of carrier and solvent to emerge at T. In previous work9,26 we achieved a suitable back pressure by inserting a narrow flow restriction of carefully selected length and diameter immediately after the through-channel outlet. For the work described here we use the same general approach of placing a flow restriction after T but, for ease of automation, the restriction takes the form of a needle-valve coupled to a stepper motor (M). Closing the valve via a clockwise turn of the stepper motor narrows the flow path and so increases the back pressure at T, making it easier for the liquids to pass through the porous capillary wall; opening the valve via an anticlockwise turn of the stepper motor has the opposite effect, making it easier for the liquids to pass straight through the capillary.
To automate the separation procedure, it is necessary to monitor the exiting fluid streams and iteratively modify the valve position until a single phase emerges at each outlet. The flow is monitored at each outlet using a light-emitting diode (LED) and a photodiode, located on opposing sides of the flow channel (see Fig. 1b). Each LED is driven by a constant current source to provide a near-constant light intensity. Hence, if the liquid/liquid separation is perfect and a single phase is flowing uniformly through each exit channel, the photodiodes at T and S will both generate stable, time-invariant photocurrents. On the other hand, if the separation is imperfect, at least one of the photodiodes will generate a fluctuating photocurrent due to scattering of incident light by the two-phase flow (scattering is caused by the different refractive indices of the two liquids, and so will occur even with colourless solutions). In-built transimpedance amplifiers within the casing of the photodiodes convert the photocurrents to voltages that are read at a rate of several hundred samples per second by a dual channel analogue to digital converter (ADC) coupled to a microcontroller. The microcontroller transfers the sampled data from the ADC to a personal computer (PC) for analysis and plotting, and periodically receives back from the PC an integer that specifies the number of steps through which the motor should move in the next iteration. The microcontroller in turn communicates with a stepper motor driver that delivers to the coils of the stepper motor the current waveforms required to execute the desired number of steps. For the needle-valve selected, just over one complete anticlockwise turn (equal to 400 full steps of the motor) is required to move the valve from fully closed to open. Full details regarding the construction and operation of the separator are provided in the Methods section.
For initial testing, the flow rates of the solvent and carrier were set to equal values of 0.5 mL min−1. The valve was set to the fully closed position, and the system was allowed to stabilise for 90 s before acquiring data. Simultaneous thirty-second traces were then recorded for each channel (see bottom row of Fig. 2a). With the valve fully closed, the entire fluid stream was forced to pass through the walls of the porous capillary into the side channel, leading to a broadly static signal at T and a strongly fluctuating signal at S. The static nature of the T signal reflects the absence of liquid flow in the (closed) through-channel, while the irregular comb-like appearance of the S signal is characteristic of a disordered two-phase flow, in which alternating slugs of the two liquids are constantly crossing the detection zone.
To determine the influence of the valve position V on the thirty-second traces, the valve was rotated anticlockwise – in fifty-step increments of the motor – from zero (closed) to five hundred (432°). Following each change of valve position, the system was allowed to stabilise for 90 s before recording the traces. For small rotations away from zero (V ≤ 350), the signals in the two channels resembled those seen for V = 0, with the side-channel signal fluctuating substantially and the through-channel signal remaining broadly static. At V = 400, the comb-like appearance of the S-channel went away, with both channels now exhibiting static signals, consistent with a single liquid phase being present in each channel. Inspection of the liquids collected at S and T revealed them to be toluene and water, respectively (as expected from the poorer wettability of water to PTFE). Hence, with the valve position set to V = 400, complete separation of the two liquids was achieved. Opening the valve further to V = 450 caused fluctuations to appear in T, indicating the valve had been opened too far and was now allowing some toluene to exit via the through-channel, alongside the water. Opening the valve still further, caused the majority of the injected two-phase stream to pass directly through the porous capillary into the T-channel.
The signal fluctuations in the S- and T-channels may be conveniently quantified in terms of the relative standard deviations, γS and γT. (Note, the relative standard deviation (RSD) γ of a signal is obtained by dividing its standard deviation σ by its mean μ: γ = σ/μ). Fig. 2b shows the variation of γS and γT with step position based on the transient data shown in Fig. 2a. To allow direct comparison with the parameter Δγ = γS − γT (introduced below), the RSD value of the T-channel is shown inverted in the diagram, i.e. we have plotted −γTversus valve position V. At V = 0, with the entire two-phase flow passing through the S-channel, the RSD was high in the S-channel (γS ≈ 14.1%) and low in the T-channel (γT ≈ 0.18%). The RSD values remained approximately constant until the valve position reached V = 350, at which point there was a slight reduction in the RSD value of the S-channel to 13.98%, followed by a much larger drop to 0.12% at V = 400. In the latter position complete separation of the two liquids was achieved, with the S- and T-channels both exhibiting low RSD values of 0.12% and 0.14%, respectively. Increasing V to 450 and beyond resulted in a substantial increase in the RSD value of the T-channel to 14.2% without substantially affecting the RSD value of the S-channel, consistent with a mixture of water and toluene passing through the T-channel and pure toluene passing through the S-channel.
To better understand the behaviour of the separator, a series of additional thirty-second traces was recorded in the vicinity of V = 400, using smaller ten-step increments (Fig. 3a). In moving from V = 350 to V = 370, lengthening periods of constant signal appeared in the S-channel trace, while the T-channel trace remained broadly static. This behaviour is consistent with a small amount of water – the liquid with poorer wettability to PTFE – passing into the T-channel (a reduction in the water content in the S-channel causes toluene to form longer slugs, leading to prolonged periods of constant signal in the S-channel trace). Inspection of the liquid extracted at T confirmed that it was indeed water, with no evidence of any contamination by toluene. Complete separation of the two liquids was achieved between V = 380 and V = 400, with both channels exhibiting static signals within this range. We refer to the range of valve positions that result in complete separation as the “separation window”. Increasing V to 410 caused occasional fluctuations to appear in the T-channel trace due to small amounts of toluene leaking into the through-channel as a result of insufficient back pressure at T. The T-channel fluctuations became more frequent as V was increased further from 420 to 450, consistent with a steadily rising toluene content in the through-channel. The RSD values extracted from the traces of Fig. 3a are shown in Fig. 3b. Inside the separation window 380 ≤ V ≤ 400, γS and γT had similar small values of <0.2%. To the left of the separation window (V < 380), γS rose rapidly to ∼14% while γT remained approximately constant at <0.2%. To the right of the separation window (V > 400), γT rose rapidly to ∼15% while γS remained approximately constant at <0.2%.
From the discussion above, it is clear that the RSD values of the signals in the two channels provide useful diagnostic information about the instantaneous quality of the separation. Four scenarios can occur if the chosen solvent/carrier combination is separable: (i) the RSD may be high in the S-channel and low in the T-channel, indicating too much liquid is passing through the walls of the porous capillary and the valve should be loosened; (ii) the RSD may be low in the S-channel and high in the T-channel, indicating too much liquid is passing straight through the porous capillary and the valve should be tightened; (iii) the RSD may be low in both channels, indicating good separation; (iv) the RSD may be moderately high in both channels, indicating the valve position is close to the separation window but still requires some further tuning to achieve complete separation.
The direction of valve rotation needed to reach the separation window from a given starting position may be readily identified by subtracting the RSD value of the T-channel from the RSD value of the S-channel to obtain a differential RSD, Δγ = γS − γT. If Δγ is large and positive, too much liquid is passing through the walls of the porous capillary and the valve should be opened to reduce the back pressure at the end of the capillary; conversely, if Δγ is large and negative, too much liquid is passing straight through the porous capillary and the valve should be tightened. The variation of Δγ with valve position is indicated by the black squares in Fig. 2b and 3b. To the left of the separation window (V < 380) Δγ followed the behaviour of γS, while to the right of the separation window (V > 400), it followed the behaviour of −γT. Within the separation window |Δγ| was <0.1%.
High values of |Δγ| occur when the valve position is far from the separation window, meaning a large number of corrective steps are required to bring the system into a state of complete separation. Close to the separation window |Δγ| is smaller and fewer corrective steps are required. This suggests the use of a simple iterative algorithm, in which the number ΔV of corrective steps made within a single iteration scales linearly with the magnitude of Δγ from a value of zero when |Δγ| < Δγ0 to a maximum value of ΔV* when |Δγ| ≥ Δγ*, where Δγ0, Δγ* and ΔV* are user-defined tuning parameters for the optimisation routine. ΔV may be expressed as a piecewise function of Δγ:
![]() | (1) |
![]() | ||
Fig. 4 Details of optimisation procedure used to control the automated separator. (a) Graph showing the adjustment ΔV in the valve position versus the differential RSD Δγ, as defined by eqn (1); the valve adjustment, expressed in terms of discrete steps of the stepper motor, increases linearly from zero when |Δγ| = Δγ0 to ±ΔV* when |Δγ| = Δγ* and is always rounded to the nearest integer; it is set equal to zero for |Δγ| < Δγ0 and to ±ΔV* for |Δγ| > Δγ*; the inset graph shows ΔV versus Δγ for small values of Δγ. (b) Flow chart summarising the automated procedure used to bring the system into a state of complete separation: (i) the valve is closed and variables are initialized; (ii) five-second traces are recorded at the T- and S-channel outlets, and quantified in terms of their respective RSD values, γT and γS; (iii) the required adjustment ΔV of the valve position is determined from Δγ = γS − γT, using eqn (1); in the event ΔV is of opposite sign to the previous adjustment ΔVold then an upper limit of ΔV = ΔVrev is enforced; (iv) variables are updated and steps (ii) and (iii) are repeated until the end of the run. |
For the purposes of selecting Δγ0 and Δγ* it is useful to obtain approximate lower and upper limits on Δγ, which we denote Δγmin and Δγmax, respectively. Δγmin may be estimated by recording traces when the two channels are empty (i.e. when no liquid is present in the system, and any fluctuations in the signals are therefore entirely due to intrinsic noise in the detection system), while Δγmax may be estimated by blocking one of the exit channels and hence forcing the entire two-phase fluid stream through the other channel. Following this approach, we obtained values of 0.03% and ∼15% for Δγmin and Δγmax respectively.
The value of Δγ0 is chosen to avoid unnecessary movement of the valve when the system is operating in a state of complete separation, and so is typically set to be a small multiple of |Δγmin|. Δγ* is typically set to be slightly smaller than |Δγmax|, while ΔV* – the maximum permitted valve adjustment within a single iteration – is set empirically. Larger values of ΔV* ensure the separation window is approached more quickly, but excessively high values risk overshooting the separation window and may cause indefinite oscillation about the window (due to repeated overshooting), preventing convergence. In practice, a suitable value for ΔV* that is able to ensure rapid and reliable convergence may be easily determined by first setting the maximum angular rotation to a moderate value (e.g. 72°), and then increasing it if convergence is too slow or decreasing it if excessive overshooting occurs. In practice, once a satisfactory set of Δγ0, Δγ* and ΔV* values has been found for one particular liquid/liquid combination under a specific choice of flow condition, the same set of parameter values may be applied without adjustment to many other flow conditions and liquid/liquid combinations.
The optimisation routine continues to run even after complete separation has been achieved. Hence, in the event of an external disturbance that disrupts the separation, the optimisation routine will make whatever adjustments are needed from the current valve position V to restore the separation (there is no need to wind the valve back to the zero position and start again). The setting of a minimum threshold for |Δγ| – below which no corrective action is taken – is beneficial as it prevents the valve position from constantly “twitching” after a position in the separation window has been found, thereby avoiding needless perturbations to the pressure in the system. Only when a significant increase in |Δγ| occurs will the valve move to restore separation. Importantly, the optimisation procedure is robust against occasional anomalous signals (“blips”) in the measured signals due e.g. to gas bubbles passing through one of the detection zones: minor blips are ignored due to the minimum threshold for |Δγ|, while larger blips cause only a momentary adjustment to the valve position that is rapidly corrected once the anomaly has passed.
Preliminary testing of the algorithm in the form described above revealed that executing a large number of clockwise steps (ΔV ≪ 0) immediately after a series of anticlockwise steps (ΔV > 0) could cause a brief interruption of the flow in the side channel (until the pressure in the system had built up to the level needed to sustain flow), substantially increasing the time required to reach complete separation. To avoid this problem, for the first iteration only after a directional change, it was found necessary to cap the value of ΔV* at a small value ΔVrev (note, a change of direction may be readily detected by comparing the latest value of ΔV with its previous value ΔVold). The complete logic flow of the algorithm – including the correction for a change of direction – is summarised in Fig. 4b. All measurements reported below were obtained using the formulation described in the flow chart with the parameter values summarised in Table 1.
Paramater | Value | Definition |
---|---|---|
S | 40 | Samples per second |
Δt | 5 s | Duration of each trace |
ΔV* | 80 steps | Maximum valve movement, equivalent to 72° |
ΔVrev | 20 steps | Maximum valve movement after a directional change |
Δγ0 | 0.1% | Minimum threshold for valve movement |
Δγ* | 10% | Effective cap on Δγ. Values of Δγ ≥ Δγ* result in a valve movement of ΔV = ΔV* |
![]() | ||
Fig. 5 Behaviour of automated separator during initial convergence to complete separation. Plots showing valve adjustment ΔV (blue line, upper plot), valve position V (blue line, lower plot) and differential RSD value Δγ (red line, lower plot) versus time, using equal injection rates of 0.5 mL min−1 for water and toluene. Due to the mapping depicted in Fig. 4a, ΔV can be seen to have a similar profile to Δγ. Starting from an initial position of V = 0 in which the entire two-phase fluid stream was forced through S, a series of six large, positive Δγ values in excess of Δγ* were obtained at the start of the run, causing the valve to make six sequential positive adjustments at the maximum allowed adjustment of +80 steps. Δγ fell abruptly to <10−3% at the seventh iteration, causing the valve to hold its current position of V = 480; a series of large and negative Δγ values were then obtained, causing the valve to make five sequential negative adjustments. Owing to the change in direction, the first of these adjustments was capped at −20 steps (see main text). The remaining four steps occurred at the maximum permitted value of −80 steps, bringing the differential RSD to <10−1% at t = 1 min. Δγ remained at a similar low value for a further iteration, before turning positive. A series of four positive adjustments of the valve followed, bringing the system – on the nineteenth iteration (t = 1.5 min) – into a state of complete separation that persisted for the remainder of the ten-minute run (first two minutes shown only). |
Initially at V = 0, with the valve fully closed, the entire two-phase fluid stream was forced to pass through the side channel, leading to a high positive value of Δγ (17.3%) in excess of +Δγ* (10%). Consequently, the motor driving the valve made its maximum permitted adjustment of ΔV* = +80 steps. After the first valve adjustment, the differential RSD still exceeded +Δγ*, resulting in another adjustment of +80 steps. This behaviour continued until 0.5 min, at which point the differential RSD fell abruptly to an extremely small value of 10−3%, which lies far below the threshold Δγ0 (= 0.1%) for moving the valve. In the next iteration the valve therefore maintained its current position. Despite doing so, the next measured Δγ value was large and negative (−13.5%), indicating the optimiser had overshot the usable range and a two-phase fluid stream was now flowing in the through-channel. The significant change in Δγ despite the valve maintaining the same position is a consequence of the algorithm updating the valve position before the flow has stabilised.
Owing to the change in the sign of Δγ, the next valve adjustment occurred in the opposite (clockwise) direction and was consequently capped at a value of −ΔVrev = −20 steps. A series of −80 step adjustments followed until t = 1 min, when Δγ fell to a small value of ∼0.02%. Δγ remained at ∼0.02% for one further step – causing the motor to hold its current position – but then jumped to a value of 4.3%. During the next four iterations the valve advanced by +29, +80, +80 and +61 steps, finishing at a final position of V = 390 which was then maintained until the end of the optimisation run (t = 5 min). Δγ values during the last four iterations before convergence were +15.5, +13.6, +7.8 and −0.05% – the latter value falling below the threshold for a valve movement. The final valve position was reached just 1.5 min after beginning the optimisation run, with |Δγ| remaining below 0.02% for the remainder of the run and the system operating in a state of complete separation throughout this time.
To determine whether the separation was indeed complete – with pure toluene passing through the S-channel and pure water passing through the T-channel – the optimisation was repeated using a small amount of orange dye (Sudan) in the toluene and a small amount of blue dye (methylene blue) in the water. The system was allowed to converge, and samples of each liquid were then collected in separate vials over a ten minute period. Visual inspection of the collected solutions indicated complete separation of the two liquids, with no cross contamination evident in either channel, see inset photograph in Fig. 6. The optimisation was then repeated once again (without the dyes present), using mass balances to record – after convergence – the amount of liquid collected in each vial over a thirty minute period. The mass of liquid collected at each outlet increased linearly with time, indicating stable flow in both channels (see Fig. 6). By comparing the mass collection rate at each outlet to the mass injection rate from the corresponding syringe, the recovery rates were determined to be ∼99.5% and ∼98.5% for toluene and water, respectively (assuming fluid densities of 867 and 1000 g L−1). The slight shortfalls from 100% recovery are attributable to evaporative losses and/or systematic inaccuracies in the syringe specifications.
The ability of the separator to maintain complete separation over an extended period of time was investigated by carrying out a twenty-four hour run under the same flow conditions of 0.5 mL min−1 each for toluene and water. Following an initial 90 s convergence period, the two traces remained broadly static over the full duration of the run, with only three brief periods in which significant fluctuations appeared in the exit channel signals (see Fig. S1 and S2†) due to momentary departures from full separation or scattering of probe light by dust or air bubbles. In each case, the fluctuations lasted no more than ninety seconds, with the system then reverting to complete separation. Following the initial convergence period, 99.9% of the five-second traces recorded during the twenty-four-hour run had differential RSD values of 0.1% or below, corresponding to stable time-invariant signals in both channels, i.e. complete separation. Changes in the valve position were infrequent, with just thirteen adjustments being made during the course of the run, ranging in size from a single step to eighty steps, see Fig. S2.†
The performance of the automated separator under different flow conditions was determined by recording a series of optimisation runs at total flow rates in the range 0.2 to 10 mL min−1, using equal injection rates of water and toluene. The results are shown in Fig. 7a where, as before, blue lines indicate the variation of step position with time, while the red lines indicate the variation of the differential RSD with time. In all cases convergence was rapid, with complete separation being achieved in two and a half minutes or less. At the highest flow rates of 4 and 10 mL min−1, the system converged rapidly to complete separation in less than 1 min, with no overshoot of the separation window. Convergence also occurred within 1 min at 2 mL min−1, although there was a slight initial overshoot of the separation window that required subsequent correction. At slower flow rates ≤1 mL min−1 convergence to complete separation was less direct, with the step position adjusting repeatedly before complete separation was achieved. The complex convergence behaviour at low flow rates is a consequence of the physical design of the separator, which allows a small amount of water to become trapped in the side-channel jacket at the start of the run (when the complete fluid stream is passing through the S-channel). Later in the run, when fluid is flowing through both channels, the trapped water escapes intermittently into the side-channel outlet, resulting in anomalously high γS values that can cause the valve to make spurious movements in the positive direction, which must subsequently be corrected. Fortunately, owing to the low (<100 μL) dead volume of the jacket, the trapped water is soon depleted, causing the anomalous signals to cease. Hence convergence is still attained rapidly (<2.5 min), albeit by a complicated sequence of valve movements.
The above results were obtained using equal flow rates for water and toluene, but the separator may also be applied in situations where the flow rates are imbalanced. In Fig. 7b, we show results for a series of runs using water-to-toluene flow-rate ratios in the range 1:
10 to 10
:
1, keeping the total flow rate fixed at 1 mL min−1. Convergence was rapid in all cases, with complete separation again being achieved in two and a half minutes or less. Visual inspection of the collected solutions indicated complete separation of the two liquids in all cases, with no cross contamination evident in either channel, see Fig. S3.† Finally we note that, for maximum synthetic and analytic versatility, it is desirable that a separator should be able to handle a broad range of liquid/liquid combinations. In Fig. S4† we present additional data demonstrating the successful application of the separator to a number of commonly used organic/aqueous, fluorous/aqueous and fluorous/organic fluid streams, namely: dichloromethane/water, chloroform/water, hexane/water, cyclohexane/water, PFPE/water and PFPE/tetrahydrofuran, where PFPE denotes perfluorinated polyether. In all cases, complete separation of the two liquids was achieved with convergence times of a few minutes, confirming the broad applicability of the separator.
In summary we have described an autonomous inline liquid/liquid separator based on a porous PTFE capillary and a small number of inexpensive electronic components. By monitoring the light transmitted through fluid streams at the two outlets of the separator and iteratively adjusting a needle-valve located downstream of the porous capillary until smooth time-invariant signals are observed at both outlets, the separator is capable of establishing complete liquid/liquid separation within a few minutes. The simple construction of the separator, its ease of use, its low cost, and its good performance using a broad range of liquid combinations make it a promising tool for numerous applications in fluidic analysis, synthesis and purification.
Datasets generated during the current study and MATLAB plotting scripts are available in the Imperial College Box repository at: https://imperialcollegelondon.box.com/v/automated-liquid-separator.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c8re00023a |
This journal is © The Royal Society of Chemistry 2018 |