S. Alex
Rautu
a,
Davide
Orsi
b,
Lorenzo
Di Michele
c,
George
Rowlands
a,
Pietro
Cicuta
*c and
Matthew S.
Turner
*a
aDepartment of Physics, University of Warwick, Coventry, CV4 7AL, UK. E-mail: m.s.turner@warwick.ac.uk; Tel: +44 (0)24 7652 2257
bDepartment of Mathematics, Physics and Computer Sciences, University of Parma, Parma, Italy
cCavendish Laboratory, University of Cambridge, Cambridge, CB3 0HE, UK. E-mail: pc245@cam.ac.uk; Tel: +44 (0)1223 337462
First published on 25th April 2017
The spectral analysis of thermal fluctuations, or flickering, is a simple and non-invasive method widely used to determine the mechanical properties of artificial and biological lipid membranes. In its most common implementation, the position of the edge of a cell or vesicle is tracked from optical microscopy videos. However, a systematic disagreement with X-ray scattering and micromechanical manipulation data has brought into question the validity of the method. We present an improved analysis protocol that resolves these discrepancies by accounting for the finite vertical resolution of the optics used to image fluctuations.
Experimental method | κ/kBT |
---|---|
Flicker spectroscopy of GUVs | |
Literature values10,11 | 28 ± 3 |
Present work – no optical projection | 27 ± 1 |
Present work – optical projection | 19 ± 1 |
Active driving of GUVs10 | 21 ± 3 |
Pulling membrane tethers12,13 | 20 ± 4 |
X-ray scattering on bilayer stacks14–17 | 18 ± 2 |
Micropipette aspiration of GUVs18–20 | 20 ± 2 |
A fluctuating GUV can be described as a quasi-spherical shell S ≡ R[1 + u(θ,φ)], where (θ,φ) are the spherical angular coordinates, u(θ,φ) is a small deviation about a sphere of radius R, and
is the radial unit vector (see Fig. 1a). The free-energy of the vesicles is described by the Helfrich Hamiltonian, which depends on κ and σ.5 By expanding this Hamiltonian to second order in u, written in the basis of the spherical harmonics Ymn as
, with
mn the mode amplitude, one can express the time-averaged mean square fluctuation amplitudes as follows:6
![]() | (1) |
![]() | (2) |
Membrane fluctuations are recorded by microscopy imaging in 2D, normally around the equatorial plane. Thus, experimental data have been typically compared with mode amplitudes obtained by projecting the modes in eqn (1) onto the plane θ = π/2; see ref. 7 and the ESI† for a full derivation. However, due to finite depth of focus, the experiments cannot isolate the signal from the equatorial plane alone. What is actually observed is a projection over the strip of membrane that lies within a focal region near the equator, as shown in Fig. 1(a); this strip can support a spectrum of surface modes that are partially averaged out in projection. This averaging effect is expected to be particularly strong for modes q ≳ 1/Δ, where Δ is a non-dimensional parameter given by the ratio of the focal depth of the microscope to the vesicle radius R.
In the case of fluorescence microscopy, where the membrane is assumed to be uniformly labelled with fluorophores that emit isotropically (see Fig. 1b), we idealise the acquired optical signal as a convolution of the membrane shape with a Gaussian of width equal to the focal depth. Namely, light arriving from height z above (or below) the focal plane has intensity scaled by . Thus, the projected intensity field on the equatorial plane is
![]() | (3) |
![]() | (4) |
![]() | (5) |
The experiments are performed on GUVs prepared by means of electroformation as in ref. 9, with DOPC and the fluorescent lipid Texas Red 1,2-dihexadecanoyl-sn-glycero-3-phosphoethanolamine (DHPE) in proportions of 99.2% and 0.8%, respectively. Videos of the equatorial fluctuations of GUVs are collected on a Leica TCS SP5 II scanning confocal microscope, using a HCX-PL-APO-CS 40.0× oil immersion objective with numerical aperture NA = 1.30. The focal depth is defined by the pinhole size of the microscope, which allows us to experimentally obtain the value of Δ by fitting the intensity profile along the z-direction to a Gaussian form of standard deviation RΔ. The position of the equatorial contours in every frame is determined with sub-pixel precision by correlating the radial intensity profile with a template. By Fourier transforming the contours and averaging over ∼2000 frames, we extract the mean-squared amplitude of the equatorial modes Fq(Δ), which can be directly compared with the model (eqn (4)); see ESI† for full experimental details. Note that a cutoff condition in the mode spectrum, q ≲ Q, arises due the finite temporal resolution of the microscope, particularly the rate at which the illumination spot is raster-scanned across the field of view. With our setup we typically obtain Q ≈ 20 (see ESI†). The accessible wave number is also limited by the spatial resolution of the setup to q ≲ qw = R/, where
is the lateral width of the diffraction-limited illumination spot.
Using eqn (4), the best-fit values of κ and to the experimental spectrum Fq(Δ) are found by means of a maximum posterior estimate,21 assuming a uniform prior and that the measurement errors are independent and Gaussian; namely, we seek to minimise
![]() | (6) |
![]() | (7) |
By imaging three GUVs with radii between 6–17 μm at various pinhole sizes, the fluctuation spectrum associated with each Δ yields an individual estimate for κ and σ. A systematic decrease in the inferred value of κ is found when the data is fitted with the model in eqn (4) in comparison with the “standard” model that considers only equatorial fluctuations, i.e. Δ = 0 (see Fig. 2a). Using a maximum posterior estimate based on the data of all the vesicles imaged at different Δ, we find κ = 19 ± 1kBT (Fig. 2b), in perfect agreement with literature values obtained with X-rays and micromanipulation techniques (see Table 1). In contrast, by fitting all spectra with the conventional model (Δ = 0), we find that κ = 27 ± 1kBT, a value compatible with the literature estimates of κ obtained by conventional flickering analysis (see Table 1).
![]() | ||
Fig. 2 (a) Fluctuation spectrum for a GUV (R ≈ 10.2 μm) imaged via confocal fluorescence microscopy, with Δ = 0.07 and τ ≈ 1.2 ms, on a log–log scale. Lines are the best-fit for the “standard” model (i.e. incorrectly assuming Δ = 0, red), and eqn (4) with Δ = 0.07 (blue). Both fits are of similar quality, but their best-fit values are significantly different, see the inset table (same colors). Dashed lines give their extrapolation outside the fitting range. The error-bars are standard errors in the mean, scaled up by a factor 10 to improve visibility. Inset plot shows the residuals Rε(q), normalised to the standard deviation. (b) Values of κ from the analysis of three GUVs of radii 6.4 μm (purple), 10.2 μm (black), and 16.7 μm (red), observed at different values of Δ. The blue dashed line is the estimate of κ from the entire dataset ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
Large discrepancies in the inferred surface tension values are also to be expected. To illustrate the dependence of the inferred values of κ with the focal depth, the conventional fitting procedure is repeated at arbitrary non-zero values of Δ for all of the spectra in , yielding an interpolated curve depicted by the green line in Fig. 2b. This shows how increasing the focal depth can result in an apparent increase in κ if its effects are not properly analysed.
In conclusion, a new methodology is developed for the analysis of flickering data that can accurately account for the finite focal depth of the optical imaging setups. Thus, this resolves a systematic inconsistency in the estimated values of bending rigidity, bringing flickering into full quantitative agreement with other methods such as X-ray scattering and micromanipulation techniques. Neglecting this correction leads to a systematic and size-dependent error which is unfortunately present in all (hundreds) published papers. Using our approach, flickering analysis is now competitive in terms of accuracy with the aforementioned methods, allowing the user to exploit its several advantages in terms of easiness of use and simplicity, as it relies on general purpose and easily accessible equipment, it is non-invasive, and can be integrated into microfluidic devices.
Footnotes |
† Electronic supplementary information (ESI) available: We provide details on the preparation of GUVs, on the confocal microscopy setup, on the code used to extract the fluctuation spectrum from a video of a GUV, and on the data analysis procedure. We report the complete theoretical derivation of the model at the basis of our Method. We also provide the Matlab routine used to analyse fluctuations of GUVs with example and the Mathematica routine used to fit the fluctuation spectrum and retrieve the mechanical properties of the membrane. See DOI: 10.1039/c7sm00108h |
‡ This crossover q-mode separates the regimes in which the membrane is dominated by the surface tension term (q ≲ qc) and the bending rigidity term (q ≳ qc). Thus, we require its value to lie within the fitting range. |
This journal is © The Royal Society of Chemistry 2017 |