Front-face fluorescence spectroscopy of tryptophan and fluorescein using laser induced fluorescence and excitation emission matrix fluorescence

Stéphane Mounier*ab, Roland Redona, Gustavo Nicolodellib and Debora Milorib
aLaboratoire PROTEE, Université de Toulon, CS 60584, 83041 Toulon Cedex 9, France. E-mail: stephane.mounier@univ-tln.fr; Fax: +33 4 9414 2157; Tel: +33 4 9414 2829
bEMBRAPA Instrumentação, Rua XV de Novembro, 1452-Centro, São Carlos, SP, Brazil

Received 8th August 2017 , Accepted 27th November 2017

First published on 12th December 2017


Abstract

Herein we study the responses of two standard molecules, tryptophan and fluorescein, over a large concentration range in boric acid via two front-face fluorescence spectroscopies: excitation emission matrix fluorescence and laser-induced fluorescence. The responses in relation to the tryptophan content were measured and were found to depend on the measurement technique. For fluorescein, no response was found for the excitation emission matrix, and an inverse relationship with the solid content was observed for the measured laser-induced fluorescence. Aiming to enhance the very weak fluorescence response of fluorescein, a novel protocol using a gel interface was established. These methods represent a future premise to investigate molecules with low fluorescence efficiency on surfaces using a logarithmic response.


Introduction

Front-face fluorescence spectroscopy (FFFS) is a powerful method to detect and determine fluorescent molecules in solid matrices without contact and sample preparation. It allows rapid and efficient characterization and quantification in pharmaceutical or alimentary domains. In environmental or agricultural applications, this technique could be a powerful tool for pollutant or nutrient quantification. However, FFFS was developed essentially in the agronomic domain using the fluorescence of the chlorophyll family to determine plant health,1 for lake sediment surveys2 and for red wine monitoring.3 Very few applications were based on other samples such as soils,4 sewage or compost.5 The application of FFFS in the environmental domain is under development because the samples present complex mixtures of several mineral matrices as well as complex and simple organic molecules. Hence, before using FFFS as a routine technique for heterogeneous surfaces in environmental systems, it is necessary to study several key parameters including diffusive effects, color or absorbance, the chemical and physical states of surfaces, and theoretical models. These parameters cannot be treated simultaneously, and simpler systems must be studied first to progressively increase the complexity. This work addresses the relationship between the FFFS responses and the molecule content in a solid. Compacted boric acid or simple powder mixture matrices were studied. Tryptophan, a well-known and well-studied molecule, and fluorescein, a molecule used as a fluorescence standard for solutions, were chosen to investigate two FFFS techniques: laser induced fluorescence (LIF) and emission excitation matrix fluorescence (EEM). Tryptophan is a fluorescent amino acid and is present in all proteins and consequently, in every biological material. Its fluorescence properties depend on its chemical environment and its structural location.6 Whatever its state, solid or in solution, the more efficient excitation wavelength domain is 250 to 280 nm, and the emission wavelength domain ranges from 310 to 380 nm,6–8 depending on the chemical conditions. Pure tryptophan is a white yellowish solid. Fluorescein fluorescence in solution occurs in the excitation wavelength domain of 440 to 500 nm and its emission occurs from 510 to 516 nm without an emission wavelength shift except under extreme conditions.9,10 Solid fluorescein is a dark red powder. No previous report was found on the solid-state fluorescence of fluorescein. In this work, tryptophan and fluorescein were studied in the solid state over a wide range of content, from 0.0002% to 7.618% and from 0.002% to 77.92% for fluorescein and tryptophan, respectively. Fluorescence emission was measured by FFFS with LIF and EEM to establish the relationship between the molecule content and fluorescence response in a solid boric acid matrix.

Materials and methods

Sample manufacturing

Boric acid powder (BA, Fisher Chemical) was used to prepare solid pellets with different tryptophan (TRP) and fluorescein (FLU) content. An initial solid mixture was created by finely grinding a boric acid mass with a TRP or a FLU mass, giving the highest solid content of C0-TRP and C0-FLU, respectively (Table 1). After this, known masses of C0-TRP and C0-FLU were added to another mass of BA to prepare the next set of samples C1-TRP and C1-FLU, respectively, and so on, to obtain a near-logarithmic decrease of TRP and FLU content in the sample set (Table 1). The zero content pellet was obtained using only the BA solid. For each solid mixture, one part of the powder was pressed at 8 tons per cm2 to obtain two solid pellets. Eventually, measurements were performed on the pellets and powders for 13 and 14 different content samples ranging from 2 × 10−3 to 78% for TRP and 2 × 10−4 to 7.6% for FLU, respectively. The content in percentage of mass for TRP and FLU is denoted as % TRP and % FLU, respectively. All solids were conserved in a dark, dry, enclosed plastic bag to avoid contamination and degradation. All amounts and references are given in Table 1.
Table 1 Tryptophan (TRP) and fluorescein (FLU) content in % of mass and code
Code FLU (% of mass) Code TRP (% of mass)
C0-FLU 7.618 C0-TRP 17.32
C1-FLU 3.766 C1-TRP 8.518
C2-FLU 2.505 C2-TRP 5.545
C3-FLU 1.364 C3-TRP 3.043
C4-FLU 0.7949 C4-TRP 1.306
C5-FLU 0.5929 C5-TRP 0.4205
C6-FLU 0.3790 C6-TRP 0.2726
C7-FLU 0.2845 C7-TRP 0.1812
C8-FLU 0.2044 C8-TRP 0.0923
C9-FLU 0.1017
CA-FLU 0.0158 CA-TRP 77.92
CB-FLU 0.0045 CB-TRP 39.46
CC-FLU 0.0008 CC-TRP 0.0100
CD-FLU 0.0002 CD-TRP 0.0021


EEM acquisition

EEM measurements were taken on a Perkin Elmer LS50B spectrofluorimeter using a solid holder module while measuring the pellet under quartz glass. The excitation range was from 200 to 400 nm varying every 10 nm with the excitation slit fixed at 5 nm. The corresponding emission spectra were acquired from 200 to 600 nm with a scan speed of 250 nm min−1 and a slit of 5 nm. No emission filter was used. The response integration time and the photomultiplicator tension were both set as automatic. The 10 by 10 nm extractions of the excitation emission matrices were obtained via the 3D export tool from the Perkin Elmer program FLWinLab.

PARAFAC analysis

To determine the component’s source contribution to the EEMs, the CP/PARAFAC algorithm was used. First all EEMs were cleaned from the scattering signals: the Rayleigh by cutting the scattering band (20 nm) and the Raman from first and second diffraction orders by applying the Zeep procedure.11 Decomposition was then performed from two to six components on the 28 corrected EEMs, and the higher number of components giving a CORCONDIA test over 60% was selected as the optimal component number.12,13 No outliers were present in the data and 3 components were found. Because the fluorescence gives a response depending on the apparatus, the component contribution was noted by relative contribution to the fluorescence (RCF). The RCF of a component should be related to the molecule content and is comparable from one sample to another.12 The CP/PARAFAC model components were indexed as CXTRP-FLU, where X is the component number and TRP-FLU indicates the name of the database.

LIF measurement

The laser-induced fluorescence was measured with a portable LIF system (lab-made equipment developed by EMBRAPA Instrumentation, Sao Carlos, Brazil) equipped with a 405 nm laser source.14 An optical fiber was coupled with a solid sample holder. Each face of each pellet was measured under the following conditions: emission wavelength range from 450 to 800 nm, integration time 1 s, boxcar 4, and average 5. For each measurement, the integration of the fluorescence emission spectra was denoted as AXLIF expressed in arbitrary units of fluorescence, a.u. by nm, where X is the studied molecule TRP or FLU.

Gel interface measurements

A gel interface was tested to enhance the fluorescence response of the solid state. A hydroxy-gel was synthesized on the same day of the measurement. Tetraethylorthosilicate (TEOS) was added to HCl until the pH = 3. Hydrolysis was performed under gentle stirring for 36 h. After this, ethanol was added to the solution, and the gel was extracted by centrifugation, cleaned with ethanol and conserved in a glass container at 4 °C in the dark. Two grams of the lump gel was placed on the quartz glass of the sample holder. By applying gentle pressure with the counter piece, the gel sliced and became transparent. After removal of the counter piece, the gel remained as a layer, sticking to the quartz windows, with slight cracks. Ten milligrams of powder was then deposited on the gel surface, and the system was pressed again until the gel cracks disappeared. The gel layer width was approximately 1 to 2 mm following pressure application. During the stabilization time, the emission fluorescence (λex/λem = 490/510 nm) was measured every minute to monitor the response evolution. This protocol was performed for the 14 pellets containing fluorescein and the boric acid pellet.

Results and discussion

Direct solid EEM decomposition

The CP/PARAFAC decomposition of the fourteen EEM measurements of the TRP and FLU pellets gives three independent components (CORCONDIA 81.71%) presented in Fig. 1. Component C1TRP-FLU (Fig. 1a) is defined by a peak maximum at (λex/λem) = 250/325 nm with an ellipsoidal shape. Component C2TRP-FLU (Fig. 1b) is slightly shifted compared to C1TRP-FLU with a peak maximum at (λex/λem) = 280/380 nm and a sharper ellipsoidal shape. These two locations are near the fluorescence of tryptophan in solution.6–8,15,16 One can note that the C1TRP-FLU excitation range is narrower than that of C2TRP-FLU. The third component, C3TRP-FLU (Fig. 1c), shows a horizontal ellipsoidal peak centered on (λex/λem) = 210/400 nm with a large emission range from 320 to 500 nm. C3TRP-FLU seems to be related neither to fluorescein nor tryptophan but to a non-specific solid surface response.17 The relationship between components C1TRP-FLU and C2TRP-FLU with the TRP content in percent (Fig. 1d and e) shows a rapid increase of the response from 0 to 1% and a saturation from 1% to 80% for the tryptophan content. The same behavior was observed for C2TRP-FLU with a higher RCF for the powder than for the pellet. Moreover, this difference persists at the low range of tryptophan (0 to 1%), and it is obvious that a linear relationship exists only for the powder state with a good correlation coefficient for both C1TRP-FLU (r2 = 0.93, slope = 2355 RCF% of TRP−1) and C2TRP-FLU (r2 = 0.94, slope = 610 RCF% of TRP−1) as shown in the insets of Fig. 1(d) and (e), respectively. For the pellets, the RCF directly plateaus just below the saturation value (Fig. 1d and e). The only difference between the powder and the pellets is the solid surface texture, and reflection or diffusive effects could explain this difference. Concerning C3TRP-FLU, which was attributed to the solid state, it presents no variation (r2 = 0.38, slope = 103 RCF% of TRP−1) (Fig. 1f), except for the low % of TRP that could be attributed to an incomplete separation of the TRP fluorescence by CP/PARAFAC; i.e., TRP fluorescence emission occurs up to 380 nm and this could influence the decomposition, leading at low content to a linear relationship with the % of TRP.
image file: c7ra08775f-f1.tif
Fig. 1 CP/PARAFAC components from the whole set of pellets and powders in the TRP-FLU data set (a–c) and their RCF related to the TRP content in mass (d–f). PE and PO represent the pellet and powder responses, respectively. The inset graphs focus on low TRP content ranging from 0 to 1% in mass.

Direct solid LIF spectra

LIF spectra measurements are presented in Fig. 2a–d; each individual spectrum is the average of the four measured spectra from the two pellets (two faces for two pellets of the same TRP content) and two measured spectra for the powder (duplicate). There is little difference in the spectrum shapes of the powder and pellet states for the signal of TRP (Fig. 2). The correlation between the pellet and powder responses is very strong; r2 = 0.94 and r2 = 0.95 for the TRP area and maximum intensity, respectively, whereas it is less correlated for FLU, where r2 = 0.83 and r2 = 0.84 for the area and maximum intensity, respectively. This means that for TRP, the solid state does not influence the response, whereas for FLU, it is more significant. Regardless of this, the fluorescein response presents a maximum intensity, IFLULIF,Max, at 520 nm and several slight shoulders at 480 nm, 575 nm, and 620 nm. This emission location, 520 nm, is slightly red shifted compared with the FLU emission in solution, which is at 510 nm. Concerning tryptophan, the maximum intensity, ITRPLIF,Max, is centered at 490 nm, and only one shoulder appears at 520 nm. This emission peak position for TRP was not expected compared to the emission maximum for TRP in solution, which is between 310 nm and 380 nm. The variation of the area under the curve AFLULIF and ATRPLIF versus percent of chemical, and the corresponding log(AFLULIF) and log(ATRPLIF) versus the logarithm of the chemical content, are described in Fig. 2 for both powder and pellet states. For FLU, there is little difference between the pellet and powder responses. Hence, for pellets or powders, AFLULIF decreases exponentially with increasing FLU content; this is unexpected behavior for a specific fluorescence response. This could be because the fluorescein solid state color induces a strong inner filter effect. However, because fluorescence is isotropic, some part of the fluorescence emitted by molecules located at the surface should be observed because it irradiates directly out of the solid. Another hypothesis could be drawn by understanding this phenomenon as a strong quenching effect by the neighbor of the fluorescein molecules, either the boric acid or the fluorescein itself. This hypothesis could be supported by the linear correlation (AFLULIF = 50[thin space (1/6-em)]010 × % of FLU + 408) observed at a very low content of FLU (<0.1%), which is strong (r2 = 0.95). The tryptophan LIF response, ATRPLIF, shows a logarithmic trend curve with a lower response for the powder than for the pellet state in Fig. 2. This saturation phenomenon could be explained by an inner filter effect like in solution18 but must be theoretically demonstrated for solids. However, for the low % of TRP, the relationship is linear until log(% of TRP) = −2. Below this % of TRP, ATRPLIF does not vary when the blank response is reached. By using a linear regression it is possible for the tryptophan to determine a detection limit, in the solid boric acid matrix, of 0.022% tryptophan in mass.
image file: c7ra08775f-f2.tif
Fig. 2 Mean LIF spectra of FLU in pellet (PE-FLU) and powder (PO-FLU) forms and of TRP in pellet (PE-TRP) and powder (PO-TRP) forms. The dotted line spectra represent the fluorescein spectra in solution.

EEM with gel interface results

The emission spectra of fluorescein extracted using a gel are shown in Fig. 3(a), for three different stabilization periods (0, 10 and 35 min) after gel contact and for the fluorescein powder C8 (0.2022% fluorescein in mass). During the diffusion of the fluorescein in the gel, the intensity increases and there is no shift in the peak position. Fig. 3b presents the maximum fluorescence intensity (510 nm) normalized to the initial intensity (t = 0 s), a few seconds after the deposition of the powder in the gel, as a function of time, except for the C3-FLU content (550 nm), owing to the saturation of the detector at the maximum wavelength. For the C3-FLU content, the intensity increases slightly and decreases regularly with time (Fig. 3b, black squares). This is due to the inner filter effect of fluorescein, which becomes highly concentrated in the gel. For lower concentrations (from C8-FLU to CD-FLU), the stabilization follows a logarithmic rule, and the stabilization time is reached more rapidly when the content is low. For the CD-FLU content (Fig. 3b, light-gray triangles), the intensity is very low, and the signal is too noisy to detect a stabilization trend. Considering this stabilization time, the FFFS-EEM fluorescence measurement was performed after 1 h of stabilization to ensure that there was no evolution with time that influenced the results. The system was then stabilized for 1 h, and the FFFS-EEM measurement was taken with the same parameters described previously.
image file: c7ra08775f-f3.tif
Fig. 3 Emission spectra of fluorescein in a gel layer (a) after 0, 10, and 35 min of stabilization. Kinetics of the fluorescence response of fluorescein with powders of different content during the diffusion into the gel at the maximum intensity of fluorescence (b).

The CP/PARAFAC decomposition of all the gel-extracted fluorescein samples gives three components (CORCONDIA 88.51%), which are presented in Fig. 4. Component C1FLU-GEL shows a large domain of excitation with a maximum at 400 nm and a straight emission band centered at 510 nm. It is slightly blue shifted compared to the emission range of the fluorescein in solution. The second component, C2FLU-GEL, presents an excitation maximum at 450 nm and the same emission as C1FLU-GEL. The range of excitation for C1FLU-GEL and C2FLU-GEL of the fluorescein in the gel is not consistent with those found in the literature, 480–500 nm. This blue shift could be explained by conformational influence due to the gel network. However, the redshift excitation position of C2FLU-GEL compared to that of C1FLU-GEL could be attributed to the inner filter effect for the fluorescein with high concentration.18 Indeed, it was not possible to apply an inner filter effect correction to the FFFS measurements because the surface absorbance measurement is difficult and because no theoretical model allows this correction. When the concentration of the fluorescein increases in the gel, the auto-absorption process shifts the excitation peak to higher wavelengths such as those that occur in solution. This redshift of emission wavelength, normally observed when the fluorescein is in solution, does not occur in this experiment. The C3FLU-GEL component fluorescence maximum is located at the (λex/λem) = 200/400 nm position. It seems that this component is similar to C3TRP-FLU and specific to the FFFS-EEM technique and not related to the samples in this work.


image file: c7ra08775f-f4.tif
Fig. 4 CP/PARAFAC components for the EEM fluorescence measurement of fluorescein (FLU) in the presence of the gel layer.

The RCF of C1FLU-GEL, C2FLU-GEL and C3FLU-GEL after 1 h of stabilization is presented in Fig. 5. The RCF of C1FLU-GEL is linear for the low content of fluorescein of <0.02% in mass (Fig. 5, inset, r2 = 0.99) but dramatically decreases after this content limit (Fig. 5a). The C2FLU-GEL component contribution increases with increasing fluorescein content. However, there is a difference of slope between the lower content and the higher content (Fig. 5b) owing to the very low contribution of C2FLU-GEL from 0 to 0.005% of fluorescein in mass. In fact, assuming that C1FLU-GEL and C2FLU-GEL are related to the two responses of fluorescein in the presence of the inner filter effect in the gel, it can be considered as the sum of contributions between C1FLU-GEL and C2FLU-GEL. Hence, the relationship clearly presents two lines of correlation, one under 0.022% of fluorescein (r2 = 0.96, slope = 14[thin space (1/6-em)]825 RCF% of FLU−1) and another one over this limit, with a lower correlation coefficient and slope (r2 = 0.87, slope = 630 RCF% of FLU−1) (Fig. 5c). The C3FLU-GEL component is unrelated to the fluorescein content and presents a non-linear increase for the first point, after which it decreases regularly to reach an average contribution of 500 RCF (Fig. 5d).


image file: c7ra08775f-f5.tif
Fig. 5 CP/PARAFAC relative contribution of CXFLU-GEL versus the FLU content in the solid powder.

Conclusions

The solid state fluorescence responses of tryptophan and fluorescein in boric acid depend on the spectroscopic technique and content. For tryptophan, the FFFS-EEM gives a response that fits well with the fluorescence spectral location of tryptophan in solution and it confirms the possibility of measuring this compound directly in solid samples. The CP/PARAFAC decomposition in a data-set of EEM with a mass content between 0 and 80% of fluorescein or tryptophan gives three components. Two are highly related to the tryptophan fluorescence and give a linear response for low content (0 to 1% of TRP) with a saturation of the contribution over 1%. The relative contribution to the fluorescence response of the components at a low content of TRP clearly depends on the solid state: the powder gives a linear response, whereas the pellets saturated the detector, probably due to the diffusive or reflective effects. The FFFS-LIF response of tryptophan shows a linear trend with different slopes for the powder and pellet forms. However, the LIF fluorescence occurs at an unusual wavelength (λex/λem) = 405/495 nm for tryptophan and further investigations have to be performed to understand which physical processes are involved. For fluorescein, no response was obtained for the FFFS-EEM technique, whereas using FFFS-FIL, a decreasing spectral response was obtained in the emission wavelength range for fluorescein in solution. This behavior could be due to a very strong inner filter effect attributed to the fluorescein color (red-brown), and this hypothesis should be tested.

By interposing a silicon gel between the quartz glass and solid powder, it was possible to obtain a fluorescein response using the FFFS-EEM technique. The CP/PARAFAC components show a fluorescence peak at the same emission position as fluorescein in solution, but with a 90 nm blue shift for the excitation wavelength. A piecewise linear relationship was observed with a steep slope from 0 to 0.02% of fluorescein in mass and a less steep slope over this limit. No saturation of the signal was observed with the gel interface during this experiment. For all FFFS-EEM measurements, a component located at (λex/λem) = 200/400 nm was found after CP/PARAFAC separation, which can be attributed to the solid-state response or to the apparatus response. For fluorescein, it was possible to measure some fluorescence by FFFS-EEM using a silicon gel layer, but further investigations should be performed to obtain quantitative results. However, even without preparing a solid surface measurement, this gel method could be developed to easily and rapidly measure heterogeneous surfaces like soils or coated materials that normally exhibit weak or no fluorescence.

Conflicts of interest

There are no conflicts to declare.

Acknowledgements

This work was supported by Professor Visitante Especial, Ciença sem Fronteira on Fluorescence Response on Heterogeneous Surfaces [processo No 313576/2013-0]. This work was supported by the French government [2015-UTLN-CRCT] and the FUI project BAC-TRACK (AAP-21/DOS0039705/00). We thank Dr Elaine Cristina Paris from EMBRAPA Instrumentation of Sao Carlos, Brazil, who provided the gels for this work.

References

  1. D. M. B. P. Milori, M. Raynaud, P. R. Villas-Boas, A. L. Venancio, S. Mounier, R. B. Bassanezi and R. Redon, Comput. Electron. Agr., 2013, 95, 11–18 CrossRef.
  2. D. Airado-Rodríguez, T. Galeano-Díaz, I. Durán-Merás and J. P. Wold, J. Agric. Food Chem., 2009, 57, 1711–1720 CrossRef PubMed.
  3. M. Woszczyk, A. Bechtel, R. Gratzer, M. J. Kotarba, M. Kokocinski, J. Fiebig and R. Cieslinski, Org. Geochem., 2011, 42, 1025–1038 CrossRef CAS.
  4. M. Muller, D. M. B. P. Milori, S. Déléris, J. P. Steyer and Y. Dudal, Waste Manag., 2011, 31, 1916–1923 CrossRef CAS PubMed.
  5. R. Albrecht, E. Verrecchia and H. R. Pfeifer, Talanta, 2015, 134, 453–459 CrossRef CAS PubMed.
  6. J. T. Vivian and P. R. Callis, Biophys. J., 2001, 80, 2093–2109 CrossRef CAS PubMed.
  7. M. Tabak, G. Sartor and P. Cavatorta, J. Lumin., 1989, 43, 355–361 CrossRef CAS.
  8. C. P. Pan, P. L. Muiño, M. D. Barkley and P. R. Callis, J. Phys. Chem. B, 2011, 115, 3245–3253 CrossRef CAS PubMed.
  9. R. Sjöback, J. Nygren, M. Kubista, R. Sjback, J. Nygren and M. Kubista, Spectrochim. Acta, Part A, 1995, 51, L7–L21 CrossRef.
  10. X. He, G. Guoquan, Z. Hui and L. Hu-lin, Microchem. J., 1997, 331, 327–331 CrossRef.
  11. R. G. Zepp, W. M. Sheldon and M. A. Moran, Mar. Chem., 2004, 89, 15–36 CrossRef CAS.
  12. R. Bro and H. A. L. Kiers, J. Chemom., 2003, 17, 274–286 CrossRef CAS.
  13. S. Mounier, H. Zhao, C. Garnier and R. Redon, Biogeochemistry, 2011, 106, 107–116 CrossRef.
  14. C. H. Santos, G. Nicolodelli, R. A. Romano, A. M. Tadini, P. R. Villas-Boas, C. R. Montes, S. Mounier and D. M. B. P. Milori, J. Braz. Chem. Soc., 2015, 26, 1136–1142 CAS.
  15. M. Tedetti, P. Joffre and M. Goutx, Sens. Actuators, B, 2013, 182, 416–423 CrossRef CAS.
  16. D. M. Reynolds, Water Res., 2003, 37, 3055–3060 CrossRef CAS PubMed.
  17. S. Mounier, G. Nicolodelli, R. Redon and D. Milori, Spectrochim. Acta, Part A, 2017, 177, 79–85 CrossRef CAS PubMed.
  18. X. Luciani, S. Mounier, R. Redon and A. Bois, Chemom. Intell. Lab. Syst., 2009, 96, 227–238 CrossRef CAS.

This journal is © The Royal Society of Chemistry 2017