A graphene-based porous carbon material as a stationary phase for gas chromatographic separations

Xiaohong Yang, Changxia Li, Meiling Qi* and Liangti Qu*
Key Laboratory of Cluster Science, Ministry of Education of China, Beijing Key Laboratory of Photoelectronic/Electrophotonic Conversion Materials, School of Chemistry and Chemical Engineering, Beijing Institute of Technology, Beijing, 100081, China. E-mail: mlqi@bit.edu.cn; lqu@bit.edu.cn

Received 28th April 2017 , Accepted 16th June 2017

First published on 23rd June 2017


Abstract

This work presents the separation performance of a 3D graphene-based porous carbon material (GPCM) as a new stationary phase for capillary gas chromatography (GC). The statically coated GPCM column showed a column efficiency of 3210 plates per m and weak polar nature. Its resolving capability and retention behaviours were investigated by using structural and positional isomers and a complex mixture covering analytes of diverse types. The GPCM column exhibits high resolution for nonpolar to polar isomers and analytes of wide ranging variety. It also displays a different retention mechanism from the commercial polysiloxane column, showing preferential retention of alcohols, aromatics and halogenated compounds. In addition, the GPCM column exhibited good column repeatability and reproducibility with RSD values on retention times in the range of 0.01–0.03% for intra-day, 0.06–0.36% for inter-day and 0.44–2.1% for between-column, respectively. The advantageous separation performance shows the good potential of the GPCM stationary phase for GC analysis.


1 Introduction

Porous carbon materials (PCMs) have high specific surface area, hierarchical porous structures and physicochemical properties such as electric conductivity, thermal conductivity, chemical stability and low density and have attracted broad attention in a variety of fields such as catalyst supports, adsorbents and electrode materials.1,2 In separation science, PCMs have been reported as stationary phases for liquid chromatography3,4 and as adsorbents for benzene5 and dye6 removal and carbon dioxide separation.7 Their preparation has been investigated regarding the effect of the nature of the templates, carbon sources/precursors and carbonization conditions.1,2 Recently, using a metal–organic framework (MOF) as a template has gained wide research attention.1,8–13 A new graphene-based PCM (denoted as GPCM in the following)12 was obtained by preparation of a three-dimensional (3D) graphene–MOF composite followed by thermal carbonization to decompose the MOFs and leave 3D GPCM behind. The as-fabricated GPCM possesses a hierarchical nanoporous structure with the improved surface area and chemical/thermal stability. In addition, it was doped with 3.4% nitrogen originating from the 2-methylimidazole (MeIm) in ZIF-8 (ZIF = zeolitic imidazolate framework).

Over the recent years, graphene-based materials have shown promise in separation science. Currently, they are mostly investigated as adsorbents for sample enrichment,14–17 but their potential as stationary phase in gas chromatography (GC)18–21 and liquid chromatography22,23 is not well explored. Particularly, their separation capability for analytes of high resemblance such as structural/positional isomers is rarely reported,20 which is a challenging task that essentially requires a highly selective stationary phase to fulfill. As previously reported, 3D graphene19,20 and its composite with MOF21 showed good GC separation performance of various types of organic analytes. Different from the reported graphene materials, the GPCM described above has hierarchical nanoporous structure doped with nitrogen. The basic 3D graphene structure may maintain its GC selectivity and hierarchical nanopores may facilitate quick mass transfer and diffusion of analytes in a chromatographic process. These features endow GPCM a promising candidate as GC stationary phase. It could be expected that GPCM may achieve high resolving capability for analytes of interest in short run time, which is quite worth exploration. To the best of our knowledge, however, the graphene-based porous materials have not been reported as stationary phases so far.

Herein, we present the investigation of the GPCM for its GC separation performance. After statically coated onto a fused-silica capillary column, the GPCM column was investigated for its column efficiency, polarity, separation performance and column repeatability and thermal stability. Its separation performance was evaluated with respect to its selectivity, resolving capability and chromatographic retention behaviors for nonpolar to polar isomers including alkanes, alcohols and phenols. In addition, its comprehensive resolving ability and retention behaviours for a wide range of analytes covering alkanes, alcohols, esters, aldehydes, halogenated and aromatic compounds was also evaluated by a more complex mixture in comparison with 5% phenyl methylpolysiloxane phase. This work demonstrates the great potential of the GPCM as a new GC stationary phase and also provides a research basis for applying diverse types of GPCMs as stationary phases in this field.

2 Experimental

2.1 Materials and equipment

All the chemicals used in this work were of analytical grade. Benzene, 2-pentanone, pyridine, 1-butanol and 1-nitropropane were purchased from Alfa Chemical Co., Ltd (Tianjin, China). n-Heptane, n-octane, 2,2,3-trimethylbutane, 2,2,4-trimethylpentane, 2,3-dimethylpentane, 2,2-dimethylhexane and 3-methylheptane were purchased from Beijing Chemical Reagent Company (Beijing, China). The rest of the chemicals were purchased from Aladdin Chemistry Co. Ltd. (Shanghai, China). Untreated fused-silica capillary tubing (0.25 mm i.d.) was purchased from Yongnian Ruifeng Chromatogram Apparatus Co. Ltd. (Hebei, China). A commercial HP-5MS capillary column (10 m × 0.25 mm, i.d., film thickness 0.25 m) coated with 5% phenyl methylpolysiloxane from Agilent Technologies Co. Ltd. (Palo Alto, CA, USA) was employed for reference.

An Agilent 7890A gas chromatograph (Palo Alto, CA, USA) equipped with a split/splitless injector, a flame ionization detector (FID) and an autosampler was used for GC separations. All the separations were performed under the following GC conditions unless otherwise specified: nitrogen of high purity (99.999%) as carrier gas at a flow rate of 1 mL min−1, injection port at 250 °C, split injection mode at a split ratio of 50[thin space (1/6-em)]:[thin space (1/6-em)]1 and FID as detector at 300 °C. Temperature programs for the separations were provided under the figure legends. Thermogravimetric analysis (TGA) was performed on a DTG-60AH Thermogravimetric Analyzer (Shimadzu, Tokyo) from 30 °C to 800 °C at a ramp rate of 10 °C min−1 under nitrogen atmosphere. The scanning electron microscopy (SEM) micrographs were recorded on a Hitachi S4800 microscope (Tokyo, Japan).

2.2 Synthesis and characterization of the GPCM stationary phase

The 3D graphene-based N-doped porous carbon material (GPCM) was synthesized and characterized by the methods described in our previous work.12 First, 3D graphene was prepared by hydrothermal reduction of graphene oxide (GO) aqueous dispersion. Briefly, 15 mL GO aqueous dispersion (2 mg mL−1) was sealed in a 25 mL Teflonlined autoclave. After heating at 180 °C for several hours, the autoclave was cooled down to room temperature and then the 3D graphene was obtained by freeze-drying. Afterwards, the 3D graphene sheets were soaked with methanol solution of Zn2+ ion and then placed in a ligand solution of 2-methylimidazole (MeIm). Upon molecules diffusion and solvothermal growth, uniform deposition of ZIF-8 crystals on 3D graphene sheets can be obtained. Afterwards, the prepared G-ZIF8 were transferred into a temperature programmed furnace under an Ar flow and were heat-treated at 800 °C for 4 h at a heating rate of 3 °C min−1 for carbonization. After the materials were cooled to the room temperature, the resultant black lumps were collected and then washed with HCl aqueous solution. Finally, the carbonized sample was rinsed with a copious amount of distilled water and freeze-dried, and then the GPCM product was obtained.

2.3 Fabrication of the GPCM capillary column

A GPCM capillary column (10 m × 0.25 mm, i.d.) was prepared by static coating method. Prior to coating, the column was purged with nitrogen at 200 °C for 3 h and pretreated with a saturated solution of sodium chloride in methanol for the surface roughing to facilitate the uniform coating of the stationary phase on the inner wall surface.24 Briefly, a suspension of sodium chloride, which was obtained under stirring condition by addition of saturated sodium chloride solution in methanol into trichloromethane, was passed through the capillary column. After the solvent was expelled from the column, the column stayed at 200 °C for 3 h under nitrogen atmosphere. Then, the column was statically coated with the solution of GPCM in dichloromethane (0.4 mg mL−1) at 40 °C. One end of the column was sealed and the other end was connected to a vacuum system in order to gradually remove the solvent under vacuum. The coated column was then conditioned from 40 °C to 180 °C at 1 °C min−1 and held at the high-end temperature for 7 h under a constant flow of nitrogen at 1 mL min−1.

3 Results and discussion

3.1 Characteristic parameters and column efficiency

The GPCM stationary phase was evaluated for its inherent thermal stability and polarity by TGA and McReynolds constants, respectively. As described in Fig. 1A, GPCM remains thermally stable up to 350 °C, suggesting its good thermal stability as stationary phase for GC separations. Its McReynolds constants were determined at 120 °C by the five probe compounds, namely benzene (X′), 1-butanol (Y′), 2-pentanone (Z′), pyridine (S′) and 1-nitropropane (U′). Table 1 lists the general polarity and average polarity of the GPCM stationary phase as well as those of 5% phenyl methylpolysiloxane stationary phase for comparison, suggesting the weak polarity of GPCM as the stationary phase for GC separations. In addition, it shows relatively larger Y′, U′ and S′ values, suggesting its stronger H-bonding and dipole–dipole interactions with specific analytes. Moreover, its column efficiency was examined at different flow rates. Thus, the Golay curve of the GPCM capillary column, which relates the height equivalent to a theoretical plate (HETP) with flow rate, was determined by n-dodecane at 120 °C. As shown in Fig. 1B, it attained the minimum HETP of 0.311 mm at 0.40 mL min−1, corresponding to the column efficiency of 3210 plates per meter. In addition, the SEM images of the cross section and the inner coating of the GPCM capillary column are shown in Fig. 1C and D, respectively, showing the uniform coating of the GPCM stationary phase on the capillary column. The above results show that in comparison with the reported graphene and graphene oxide columns,18,19 the GPCM column achieved higher column efficiency mainly due to the good dispersibility and uniform coating of the GPCM stationary phase on the inner wall of the capillary column.
image file: c7ra04774f-f1.tif
Fig. 1 (A) TGA curve of the GPCM stationary phase under nitrogen atmosphere; (B) Golay curve of the GPCM capillary column determined by n-dodecane at 120 °C; (C) and (D) SEM images showing the cross section and the inner coating of the GPCM capillary column, respectively.
Table 1 McReynolds constants of the GPCM and HP-5MS capillary columnsa
Stationary phase X Y Z U S General polarity Average polarity
a X′, benzene; Y′,1-butanol; Z′, 2-pentanone; U′, 1-nitropropane; S′, pyridine. Temperature: 120 °C.
I for GPCM 707 743 715 793 841    
I for squalane 653 590 627 652 699    
ΔI for GPCM 54 153 88 141 142 578 116
ΔI for HP-5MS 30 72 62 96 65 325 65


3.2 Separation of isomers of alkanes, alcohols and phenols

Separation of linear and branched alkanes is of significant importance in petroleum industry25 and is also a challenging work, particularly for the branched isomers due to their similar physicochemical properties. On the basis of its unique structural features, we made investigations on the resolving capability of the GPCM column for alkane isomers. Fig. 2A and B show its isothermal separations of C7 and C8 alkane isomers, respectively, at 30 °C. The baseline separation of these isomers within short time demonstrates the high selectivity and resolving capability of the GPCM stationary phase for analytes of high resemblance. In addition, it can be noted that the more branched alkanes exhibited earlier elution. Its high capability for achieving the high resolution may result from the cooperative effect of pore size restriction and van der Waals interactions. The nanoporous structure may selectively retain the less branched analytes since the steric hindrance of highly branched molecules restrict their close contact with the graphite surface of GPCM, leading to the weaker interaction and shorter retention.26 Moreover, failing in close contact with the stationary phase also weakens their strength of van der Waals interactions with the phase.
image file: c7ra04774f-f2.tif
Fig. 2 Isothermal separations of (A) heptane isomers and (B) octane isomers at 30 °C on the GPCM capillary column. Flow rate at 0.5 mL min−1 for (A) and at 1 mL min−1 for (B).

Fig. 3A–C present the isomer separations of alcohols and phenols, respectively, on the GPCM column. For the polar butanol isomers in Fig. 3A, the GPCM column also exhibited nearly baseline separation, suggesting its high resolving ability for polar analytes. The noticeable tailing of the alcohols, especially the ones retained longer, may originate from the cooperative effect of their stronger H-bonding and dipole–dipole interactions with the stationary phase and the size selectivity of the porous structure. Also, the GPCM column achieved complete separation of the two unsaturated alcohol isomers, i.e., nerol and geraniol (Fig. 3B). Fig. 3C shows its high resolution for the phenol isomers, musk phenol and carvacrol. The adjacent isopropyl to the phenol group in musk phenol sterically hinder its H-bonding interaction with the stationary phase, leading to its earlier elution than its counterpart. The above results demonstrated the high selectivity and resolving ability of the GPCM column for nonpolar to polar analytes of high similarity, showing advantages over the graphene and graphene oxide columns18,19 owing to the cooperative effect of multiple molecular interactions and pore selectivity of the GPCM stationary phase. These results inspired us to further explore its separation performance for analytes of wide variety and its potential for separation of complex samples.


image file: c7ra04774f-f3.tif
Fig. 3 Isothermal separations of alcohol and phenol isomers on the GPCM capillary column. (A) butanol isomers at 40 °C, (B) nerol and geraniol at 110 °C and (C) musk phenol and carvacrol at 100 °C. Flow rate at 0.3 mL min−1 for (A) and at 1 mL min−1 for (B) and (C).

3.3 Separation of a complex mixture of analytes of wide variety

To comprehensively evaluate the separation performance of the GPCM column for analytes of diverse variety, a complex mixture consisting of 22 analytes of different types was employed. Meanwhile, a widely-used HP-5MS capillary column was used as reference for the evaluation. As shown in Fig. 4, the GPCM column achieved good resolution of all the analytes whereas the commercial column partially overlapped two analyte pairs, namely methyl octanoate/1,6-dichlorohexane (peaks 11/12, R = 1.09) and ethyl benzoate/1,2,4-trichlorobenzene (peaks 13/14, R = 0.68). Clearly, the GPCM stationary phase exhibits high resolving ability of halogenated analytes from esters, which may be attributed to the stronger dipole–dipole interaction and halogen-bonding interaction (N⋯Cl).27,28 The cooperative effect of these interactions may enhance the distinguishing ability of the GPCM stationary phase for these analytes. Moreover, the GPCM stationary phase shows reversal elution sequence for three analyte pairs, namely cyclohexanone/1-hexanol (peaks 4/5), 1-bromononane/m-nitrochlorobenzene (peaks 15/16) and n-tridecane/1-decanol (peaks 17/18). The later elution of the alcohols and halogenated analytes further evidence the different retention mechanism of the GPCM stationary phase from the polysiloxane phase. Specifically, the later elution of hexanol and decanol may be attributed to their stronger H-bonding interaction (O–H⋯N) with the nitrogen atoms in different forms (pyridinic, pyrrolic and graphitic N) of the GPCM phase. In addition, the later elution of m-chloronitrobenzene than bromononane on the GPCM phase can be mainly ascribed to the comprehensive result of dipole–dipole, π–π and halogen-bonding interactions. The above results demonstrated the high resolving capability and different retention mechanism of the GPCM stationary phase in contrast to the polysiloxane thanks to the cooperative contribution of its multiple types of molecular interactions with specific analytes, which may mainly involve H-bonding, van der Waals, π–π and halogen-bonding interactions as well as shape-fitting effect.
image file: c7ra04774f-f4.tif
Fig. 4 Separations of the complex mixture consisting of analytes of different types on the GPCM and HP-5MS capillary columns. Peaks: (1) n-octane, (2) 1-bromopentane, (3) ethylbenzene, (4) cyclohexanone, (5) 1-hexanol, (6) 2,5-hexanedione, (7) benzaldehyde, (8) benzyl chloride, (9) o-dichlorobenzene, (10) acetophenone, (11) methyl octanoate, (12) 1,6-dichlorohexane, (13) ethyl benzoate, (14) 1,2,4-trichlorobenzene, (15) 1-bromononane, (16) m-nitrochlorobenzene, (17) n-tridecane, (18) 1-decanol, (19) 1-methyl naphthalene, (20) biphenyl, (21) n-pentadecane and (22) methyl dodecanoate. Temperature program: 40 °C for 1 min to 160 °C at 10 °C min−1. Flow rate at 1.0 mL min−1 for both of the columns.

3.4 Column repeatability and thermal stability

For the GPCM columns, its column repeatability (intra-day, inter-day) and reproducibility (between-column) were investigated by separations of the indicated mixture and evaluated by relative standard deviation (RSD%) on the analyte retention times. The obtained results are listed in Table 2, showing the RSD values of 0.01–0.03% for intra-day, 0.06–0.36% for inter-day and 0.44–2.1% for between-column, respectively. Notably, the column also achieved good reproducibility for aldehydes and alcohols that are sensitive to the column variations with the RSD values in the range of 0.44–0.74%. The above results demonstrate its good column repeatability and reproducibility for diverse types of analytes.
Table 2 Repeatability and reproducibility of the GPCM capillary column on the retention times (tR, min) for separations of the indicated mixture (n = 4)
Analytes Intra-day Inter-day Between-column
tR RSD (%) tR RSD (%) tR RSD (%)
Salicylaldehyde 2.597 0.01 2.594 0.29 2.594 0.74
Methyl octanoate 2.925 0.03 2.921 0.25 2.925 1.1
n-Dodecane 3.850 0.02 3.847 0.06 3.749 2.1
Capraldehyde 4.004 0.02 4.001 0.06 3.984 0.44
1-Nonanol 4.211 0.03 4.205 0.12 4.179 0.71
n-Tridecane 5.110 0.01 5.109 0.17 4.990 1.8
Methyl decanoate 5.379 0.01 5.378 0.20 5.349 0.51
n-Tetradecane 6.377 0.01 6.376 0.29 6.246 1.5
1-Undecanol 6.664 0.01 6.661 0.32 6.612 0.74
n-Pentadecane 7.604 0.01 7.605 0.36 7.466 1.4
Methyl dodecanoate 7.858 0.01 7.858 0.36 7.805 0.55


Its column thermal stability was evaluated by separation of the indicated mixture after the column was conditioned from 40 °C up to each of the temperatures (180–260 °C in 20 °C increments) and held for 2 h, respectively. As illustrated in Fig. 5, over the range of conditioning temperature, the retention times remained almost unchanged for most of the analytes with RSD values in the range of 0.60–3.3%. In addition, the selectivity factors of the analytes over the temperature range are also provided in Table 3, indicating that the column maintained its high selectivity above 1.05 for all of the analytes with RSD values of 0.51–3.5%. The above results demonstrated the good thermal stability of the GPCM column.


image file: c7ra04774f-f5.tif
Fig. 5 Effect of column conditioning temperatures on the retention times of analytes in the indicated mixture on the GPCM capillary column.
Table 3 Effect of column conditioning temperatures on the selectivity factors of the analytes in the indicated mixture on the GPCM capillary column
Selectivity factor (α) 180 °C 200 °C 220 °C 240 °C 260 °C RSD%
Bromobenzene/methyl hexanoate 1.18 1.18 1.19 1.19 1.19 0.51
n-Decane/1-bromobenzene 1.22 1.19 1.18 1.16 1.13 2.9
p-Dichlorobenzene/decane 1.24 1.27 1.30 1.32 1.35 3.5
n-Undecane/p-dichlorobenzene 1.25 1.22 1.21 1.19 1.17 2.3
Methyl octanoate/n-undecane 1.11 1.13 1.14 1.15 1.17 2.1
1-Octanol/methyl octanoate 1.05 1.05 1.07 1.08 1.10 2.1
n-Dodecane/1-octanol 1.18 1.16 1.14 1.12 1.08 3.3
Naphthalene/n-dodecane 1.08 1.09 1.11 1.12 1.14 2.3
n-Tridecane/naphthalene 1.18 1.16 1.15 1.14 1.12 1.9
1,6-Dibromohexane/n-tridecane 1.07 1.09 1.10 1.11 1.12 2.0
n-Tetradecane/1,6-dibromohexane 1.13 1.11 1.10 1.09 1.08 1.7
Biphenyl/n-tetradecane 1.04 1.05 1.06 1.07 1.08 1.6
n-Pentadecane/biphenyl 1.12 1.10 1.09 1.09 1.08 1.4
Acenaphthene/n-pentadecane 1.05 1.06 1.07 1.07 1.08 1.3


4 Conclusions

This work describes the separation performance of the GPCM stationary phase for GC separations, showing high selectivity and resolving capability for nonpolar to polar isomers (alkanes, alcohols and phenols) and general analytes of diverse variety. Compared with the widely-used HP-5MS column, the GPCM column shows high resolution and different elution order for certain analytes. Particularly, it delayed the elution of alcohols and halogenated analytes probably due to its stronger H-bonding and halogen-bonding interactions with these analytes. Besides, the GPCM column also exhibited good separation for a wide range of analytes thanks to its multiple molecular interactions and size selectivity. It achieves high selectivity for different types of analytes through its unique interactions with specific analytes. This work demonstrates the promising future of the graphene-based porous carbon material in chromatographic separations and may also interest more researchers in utilizing porous carbon materials in separation science.

Acknowledgements

The authors are grateful for the financial support by the National Natural Science Foundation of China (21575013) and the 111 Project B07012 in China.

References

  1. C. D. Li, Z. J. Li and S. Dai, Angew. Chem., Int. Ed., 2008, 47, 3696–3717 CrossRef PubMed .
  2. P. F. Zhang, H. Y. Zhu and S. Dai, ChemCatChem, 2015, 7, 2788–2805 CrossRef CAS .
  3. C. West, C. Elfakir and M. Lafosse, J. Chromatogr. A, 2010, 1217, 3201–3216 CrossRef CAS PubMed .
  4. Y. Zhao, S. S. W. Szeto, R. P. W. Kong, C. H. Law, G. Li, Q. Quan, Z. J. Zhang, Y. Q. Wang and I. K. Chu, Anal. Chem., 2014, 86, 12172–12179 CrossRef CAS PubMed .
  5. J. J. Li, R. J. Lu, B. J. Dou, C. Y. Ma, Q. H. Hu, Y. Liang, F. Wu, S. Z. Qiao and Z. P. Hao, Environ. Sci. Technol., 2012, 46, 12648–12654 CrossRef CAS PubMed .
  6. L. Chen, T. Ji, L. Brisbin and J. H. Zhu, ACS Appl. Mater. Interfaces, 2015, 7, 12230–12237 CAS .
  7. X. F. Bing, Y. J. Wei, M. Wang, S. Xu, D. H. Long, J. T. Wang, W. M. Qiao and L. C. Ling, J. Colloid Interface Sci., 2017, 488, 207–217 CrossRef CAS PubMed .
  8. B. Liu, H. Shioyama, T. Akita and Q. Xu, J. Am. Chem. Soc., 2008, 130, 5390–5391 CrossRef CAS PubMed .
  9. B. Liu, H. Shioyama, H. Jiang, X. Zhang and Q. Xu, Carbon, 2010, 48, 456–463 CrossRef CAS .
  10. L. Radhakrishnan, J. Reboul, S. Furukawa, P. Srinivasu, S. Kitagawa and Y. Yamauchi, Chem. Mater., 2011, 23, 1225–1231 CrossRef CAS .
  11. H. Jiang, B. Liu, Y. Lan, K. Kuratani, T. Akita and H. Shioyama, J. Am. Chem. Soc., 2011, 133, 11854–11857 CrossRef CAS PubMed .
  12. C. X. Li, C. G. Hu, Y. Zhao, L. Song, J. Zhang, R. D. Huang and L. T. Qu, Carbon, 2014, 78, 231–242 CrossRef CAS .
  13. H. X. Zhong, J. Wang, Y. W. Zhang, W. L. Xu, W. Xing, D. Xu, Y. F. Zhang and X. B. Zhang, Angew. Chem., Int. Ed., 2014, 53, 14235–14239 CrossRef CAS PubMed .
  14. Q. Liu, J. B. Shi and G. B. Jiang, Trends Anal. Chem., 2012, 37, 1–11 CrossRef CAS .
  15. J. Fan, Z. L. Dong, M. L. Qi, R. N. Fu and L. T. Qu, J. Chromatogr. A, 2013, 1320, 27–32 CrossRef CAS PubMed .
  16. X. Wang, B. Liu, Q. P. Lu and Q. S. Qu, J. Chromatogr. A, 2014, 1362, 1–15 CrossRef CAS PubMed .
  17. H. Amanzadeh, Y. Yamini, M. Moradi and Y. A. Asl, J. Chromatogr. A, 2016, 1465, 38–46 CrossRef CAS PubMed .
  18. Y. Feng, C. G. Hu, M. L. Qi, R. N. Fu and L. T. Qu, Chin. Chem. Lett., 2015, 26, 47–49 CrossRef CAS .
  19. J. Fan, M. L. Qi, R. N. Fu and L. T. Qu, J. Chromatogr. A, 2015, 1399, 74–79 CrossRef CAS PubMed .
  20. N. Han, M. Y. Ye, M. L. Qi, R. N. Fu and L. T. Qu, RSC Adv., 2015, 5, 74040–74045 RSC .
  21. X. H. Yang, C. X. Li, M. L. Qi and L. T. Qu, J. Chromatogr. A, 2016, 1460, 173–180 CrossRef CAS PubMed .
  22. X. Q. Zhang, S. Chen, Q. Han and M. Y. Ding, J. Chromatogr. A, 2013, 1307, 135–143 CrossRef CAS PubMed .
  23. X. J. Liang, X. S. Wang, H. X. Ren, S. X. Jiang, L. C. Wang and S. J. Liu, J. Sep. Sci., 2014, 37, 1371–1379 CrossRef CAS PubMed .
  24. A. J. Huang, N. Nan, M. Chen, X. Pu, H. Tang and Y. L. Shun, Acta Sci. Nat. Univ. Pekin., 1988, 24, 425–429 CAS .
  25. Z. R. Herm, E. D. Bloch and J. R. Long, Chem. Mater., 2014, 26, 323–338 CrossRef CAS .
  26. J. L. Gundersen, J. Chromatogr. A, 2001, 914, 161–166 CrossRef CAS PubMed .
  27. S. H. Jungbauer, D. Bulfield, F. Kniep, C. W. Lehmann, E. Herdtweck and S. M. Huber, J. Am. Chem. Soc., 2014, 136, 16740–16743 CrossRef CAS PubMed .
  28. M. H. Kolář and P. Hobza, Chem. Rev., 2016, 116, 5155–5187 CrossRef PubMed .

This journal is © The Royal Society of Chemistry 2017