DOI:
10.1039/C6RA24185A
(Paper)
RSC Adv., 2016,
6, 114685-114689
Highly sensitive and selective fluorescence detection of Hg(II) ions based on R-phycoerythrin from Porphyra yezoensis
Received
29th September 2016
, Accepted 5th December 2016
First published on 6th December 2016
Abstract
R-Phycoerythrin (R-PE) is a kind of natural fluorescent protein from marine Porphyra yezoensis. A highly sensitive and selective fluorimetric method has been developed to detect Hg2+ ions utilizing R-PE as a fluorescent probe. R-PE could respond selectively to Hg2+ ions among a series of metal ions. The reaction mechanism was investigated through UV-vis, fluorescence microscope imaging and fluorescence measurements. It was demonstrated that the interaction between the R-PE fluorescent probe and Hg2+ ions would cause fluorescence quenching. Besides, there was a linear relationship between R-PE fluorescence intensities and Hg2+ concentrations. A novel method was thus explored to detect Hg2+ ions with high sensitivity, selectivity, and a broad linear range of 0.0010–25.0 μM, which could allow for Hg2+ ions down to 0.0130 μM with a relative standard deviation of about 1.60%. Additionally, the method was successfully applied to detect Hg2+ ions in diverse water samples, showing the appropriate rates of recovery between 92.0% and 108.0%. This highly sensitive and selective natural green fluorescent probe of R-PE should have huge potential applications for detecting Hg2+ ions in diverse environments.
1 Introduction
There has been an increasing awareness of heavy metal pollution throughout the world because of the negative effects on the environment. Among the heavy metals of concern, mercury has long been recognized as a global pollutant because it can remain in the atmosphere, food and water. It can bioaccumulate in the human body via the food chain to severely damage the human brain, the nervous system, the endocrine system and other biological systems, resulting in brain damage, kidney failure and various cognitive and motion disorders.1
Up to now, many analysis technologies have been developed to detect heavy metal ions including Hg2+ ions, such as atomic absorption spectrometry,2 inductively coupled plasma mass spectrometry (ICP-MS),3 cold vapor atomic absorption spectrometry (CV-AAS),4 electrochemical sensing methods,5 colorimetric assays,6 and fluorescence detection methods.7 All of these methods have their own advantages and have made great contributions for the detection of Hg2+ ions. Among these analysis methods, fluorescence detection methods have prominent advantages including simplicity, sensitivity, short reaction time, whose detection performances are dependent on the used fluorescent probes to some extent.
Many kinds of synthesized fluorescence materials have been extensively applied as fluorescent probes. But the harsh and complicated synthesis procedures with high cost might limit their further applications. Therefore, a variety of novel fluorescent proteins have been rapidly developed and widely used in marine biosciences, biotechnology, and proteomics.8 Up to date, many green fluorescent proteins (GFPs) have been separated from Aequorea. Particularly, blue fluorescent protein (BFP), cyan fluorescent protein (CFP) and yellow fluorescent protein (YFP) were gained and applied. For example, scientists had applied fluorescent proteins for the super-resolution imaging techniques.9,10 Besides, photo-switchable fluorescent probes with distinct colors were also introduced for the multicolor.11
Recently, phycoerythrin (PE), as a natural fluorescent protein, has become a valuable candidate in flow cytometry and immunofluorescence microscopy.12 In addition, it was applied not only as natural colorants for use in food and cosmetics, but also as fluorescent dye. As one of the most important type of PE, R-PE has been applied to make the R-PE-labeled IgG, which can be applied in fluorescence immunoassays as secondary antibody.13 In our previous work, many synthetic fluorescent probes were prepared and extensively applied. For example, bimetallic gold–silver nanoclusters was successfully synthesized to analyze mercury and copper levels in blood.14 Silver nanoclusters (AgNCs) were synthesized and upon the GSH passivation, the specific recognition of the AgNCs was modulated from Hg2+ ions to Cu2+ ions which can apply for ion sensing and biological imaging in the complicated media like blood.15 Currently, many natural and nontoxic fluorescent probes were in high demanded. In the present report, R-PE extracted from marine Porphyra yezoensis was utilized as natural and nontoxic fluorescent probe for detecting Hg2+ ions combining with fluorescent method in diverse water environments with high selectivity and sensitivity.
2 Experimental
2.1 Materials and instrumentation
Porphyra yezoensis was purchased from Yantai, Shandong Province, China. (NH4)2SO4, NaHCO3, Na2HPO4, NaH2PO4, hydroxyapatite (HA), ethylenediaminetetraacetic acid (EDTA), HCl, Coomassie Blue R-250, NaCl, FeSO4, MgSO4·7H2O, CuSO4, Ca(NO3)2, ZnSO4, NiSO4, KNO3, BaCl2, CrCl3, CoCl2, HgCl2, Pb(NO3)2, AgNO3, CdCl2, KI, Na2CO3, KBr, KCl, CaCl2, Na2SO4 and NaHCO3 were purchased from Sinopharm Chemical Reagent Co., Ltd (Beijing, China). All of these reagents were of analytical grade.
Fluorescence spectra were obtained using a fluorescence spectrometer (F-2700, Hitachi, Japan). The sample solutions were excited at 495 nm, with slits for excitation and emission both of 10.0 nm. The characteristic peak was obtained by UV-vis spectrophotometer (UNICO, UV-2000). The photographs were obtained by ultraviolet analyzer (Lichen, ZF-1, China) at 365 nm. The morphological characteristics were obtained by microscope (Olympus, BX53, Japan).
2.2 R-PE extraction, isolation and purification
R-PE was extracted and purified according to the reported literature method with slight modification.16 Porphyra yezoensis was collected and washed in phosphate buffer solution (PBS, 10 mM, pH 6.8). The alga-buffer mixture suffered from freeze–thaw cycles of −25 °C and 4 °C for four times. The resulting slurry was filtrated by gauze, then centrifuged (7500 g, 20 min, 4 °C). The crude extraction was precipitated with ammonium sulfate at concentrations of 10% and 50%, discarding the supernatant. The final precipitant, dissolved in PBS (10 mM, pH 6.8), was dialyzed in PBS (10 mM, pH 6.8). After centrifuged at 7500 g for 20 min, the crude extract was further purified with HA column. After using different concentrations PBS for gradient eluting, the eluent containing R-PE was collected. The R-PE was examined by SDS-PAGE (17% separating and 3% stacking gel). The gel was stained with Coomassie Blue R-250 (0.1%) and destained with double distilled water. The protein content of R-PE was determined using the Bradford method. The absorption peak of R-PE was measured on a spectrophotometer.
2.3 Fluorimetric analysis of metal ions
A series of 1.0 μM metal ions (Zn2+, Cr3+, Cu2+, Ba2+, Co2+, Fe2+, Pb2+, Cd2+, Mg2+, Ag+, Ni2+, Hg2+, K+, Na+ and Ca2+) were separately added to R-PE solution for the fluorescent selective assay. Then the morphological characteristics of R-PE with Hg2+ ions were observed by microscope in bright-field image and fluorescence image, respectively.
2.4 Optimization of the detection conditions
1.0 μM Hg2+ ions was added to R-PE solution at different reaction time (1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15 min), pH (5.0, 6.0, 7.0, 8.0, 9.0), and temperatures (20, 25, 30, 35, 40, 45, 50 °C), respectively, to explore the optimum conditions. In addition, the interference of other negative ions (I−, PO43−, SO42− and CO32−) at 1.0 μM and compounds (0.101 g L−1 KBr, 0.695 g L−1 KCl, 24.530 g L−1 NaCl, 1.537 g L−1 CaCl2, 4.090 g L−1 Na2SO4 and 0.201 g L−1 NaHCO3), which possibly co-exist in sea water samples were evaluated under the optimal conditions.
2.5 Standard curve for detecting Hg2+ ions
Hg2+ ions at different concentrations (0, 0.0010, 0.005, 0.010, 0.10, 0.50, 1.0, 5.0, 25.0 μM) were analyzed under the optimum condition of Tris–HCl (50 mM, pH 8.0) at 40 °C and 10 min reaction time. Then the fluorescence spectra of the reaction mixtures were recorded at the excitation of 495 nm. A calibration curve was thereby prepared to show the relationship between the logarithmic concentrations of Hg2+ ions and fluorescence quenching efficiencies of R-PE. The quenching efficiencies of R-PE caused by Hg2+ ions were calculated in accordance with the following equation reported previously: quenching efficiencies = (F0 − F)/F0, where F0 and F refer to the fluorescence intensities of R-PE (λem = 570 nm) in the absence and presence of Hg2+ ions, respectively.
2.6 Detection of Hg2+ ions in diverse water samples
Tap water and sea water were collected from Weihai, Shandong province, China. The samples were centrifuged at 10
000g for 10 min and then filtered through a 0.22 μm filtered membrane. To further investigate the feasibility and possible application of the developed R-PE fluorescent probe for analysis of Hg2+ ions in the complex water sample, Hg2+ ions standards including 0, 0.050, 0.100, 0.500 μM were initially added into blank tap-water solution and sea water solution. R-PE with Tris–HCl (50 mM, pH 8.0) solution was mixed with diverse water samples containing various concentrations of Hg2+ at 40 °C for 10 min. Then, the fluorescence quenching intensity at λem of 570 nm was measured by a fluorescence spectrometer. The recovery (%) = 100 × (found concentration)/(spiked concentration). Data from the experiment was analyzed by SPSS16.0 statistic software.
3 Results and discussion
3.1 Characterization of purified R-PE
The purified R-PE was determined by SDS-PAGE (Fig. 1A). It was revealed that purified R-PE from the HA column contained three main bands. On the basis of the positions of the protein markers, the molecular weights of three bands could be estimated to be approximately 18.0 kDa, 21.0 kDa and 30.0 kDa. It was reported that SDS-PAGE analysis previously confirmed the presence of two major subunits with 18.0 kDa and 20.0 kDa, respectively, and a minor subunit of 30.0 kDa, which were much smaller and slighter.17,18 The optical properties of R-PE were characterized by UV-vis absorbance and fluorescence spectroscopy. Fig. 1B indicated that there were two adsorption peaks at 495 nm and 560 nm with a shoulder at 540 nm, corresponding to the previous reports (495 nm, 566 nm and 540 nm shoulder; 499 nm, 565 nm, and 545 nm shoulder).19,20 As shown in Fig. 1C, the R-PE showed red color under visible light, while exhibited powerful orange color when irradiated with UV light. When excited at 495 nm, R-PE displayed a maximum emission peak at 570 nm. As being mainly in accordance with a former literature report,21 the features could demonstrate that the substance extracted from Porphyra yezoensis was a kind of R-PE.
 |
| Fig. 1 Characterization of purified R-PE. (A) SDS-PAGE of purified R-PE. Lane 1 and lane 2, purified R-PE; lane 3, molecular mass markers. (B) Absorption spectra of purified R-PE. (C) Fluorescence excitation (red line) and emission (black line) spectra of R-PE after purification with the described procedure. Inset shows R-PE under visible light (left) and UV light (right). Then the relevant fluorescence changes were recorded by the fluorescence spectrometer. | |
3.2 Sensing response of R-PE to metal ions
To explore the sensing selectivity, R-PE was used to investigate the influence of various biologically and environmentally relevant metal ions under the same conditions. Fig. 2 presented the difference in relative fluorescence against the common metal ions including Zn2+, Cr3+, Cu2+, Ba2+, Co2+, Fe2+, Pb2+, Cd2+, Mg2+, Ag+, Ni2+, Hg2+, K+, Na+ and Ca2+ ions at 1.0 μM.
 |
| Fig. 2 Relative fluorescence (F/F0) of R-PE solution in the presence of various metal ions at 1.0 μM. F0 and F correspond to the fluorescence intensity of R-PE in the absence and presence of 1.0 μM of various metal ions. | |
The results showed that only Hg2+ ions could trigger the quenching of the fluorescence obviously, while the other ions displayed weak or even negligible effects on the fluorescence intensities, as shown in their corresponding photographs (Fig. 2, insert). These observations reflected that the prominent sensing response to Hg2+ ions endowed R-PE with high selectivity. In order to assess the quenching reactions between Hg2+ ions and R-PE, the morphological characteristics of R-PE was observed by microscope in Fig. 3. As shown in Fig. 3D, the introduction of Hg2+ ions caused the fluorescence quenching of R-PE compared with Fig. 3B. The fluorescence quenching of R-PE might be partly attributed to the energy transfer between R-PE and Hg2+ ions.15,22
 |
| Fig. 3 (A) Bright-field image of R-PE (0.080 mg mL−1). (B) Fluorescence image of (A). (C) Bright-field image of R-PE (0.080 mg mL−1) treated with Hg2+ (5 μM) for 10 min at 30 °C. (D) Fluorescence image of (C). | |
3.3 Optimization of the detection conditions
In order to obtain a highly-sensitive response to Hg2+ ions, the effects of the experimental conditions, such as the reaction time, pH values and temperature, were investigated respectively. The relative fluorescence of R-PE and Hg2+ ions in various times was measured in Fig. 4A. The figure showed that the fluorescence intensity reduced at first ten minutes and became constant later. So the reaction time at ten minutes was chosen for the following experiments. Compared with some other fluorescent probes such as Pvi,23 long lifetime fluorescence quantum dots and gold nanoparticles,24 colorimetric method25 and flame atomic absorption spectrometry,26 R-PE as a fluorescent probe to Hg2+ ions was more rapid in this work. Fig. 4B illustrated the relative fluorescence of reactions conducted at different pH values. It was obviously found that the highest relative fluorescence was obtained at pH 8.0. As shown in Fig. 4C, the relative fluorescence reached the maximum when temperature reached up to 40 °C. Thus, 40 °C was selected for the optimal detection conditions. Fig. 4D showed that relative fluorescence values of potential interferents (I−, PO43−, SO42−, CO32−, KBr, KCl, NaCl, CaCl2, Na2SO4 and NaHCO3) were around 1.0, indicating that no significant interference occurred when co-existing with Hg2+ ions.
 |
| Fig. 4 Effects of time (A), pH (B) and temperature (C) on the relative fluorescence (F0/F) of R-PE (0.080 mg mL−1) for Hg2+ detection. (D) Effects of potential interferents on relative fluorescence (F/Fx) of R-PE (0.080 mg mL−1) for Hg2+ detection. F and Fx correspond to the fluorescence intensity of R-PE with Hg2+ ions in absence and presence of 1.0 μM potential interferents. | |
3.4 The response of R-PE fluorescent probes to Hg2+ ions
Besides the selectivity, the sensitivity was another important parameter to assess the performance of the sensing system. Therefore, the capability of R-PE fluorescent probe for quantitative detection of Hg2+ ions was evaluated under the optimal conditions. As shown in Fig. 5A, the R-PE was sensitive to Hg2+ ions and the fluorescence intensity decreased gradually with increasing concentrations of Hg2+ ions. When the concentration of Hg2+ ions was increased to 25.0 μM, no further restoring fluorescence could be obtained, meaning that the sensing response had reached to the maximum. Fig. 5B exhibited the relationship between the logarithmic concentrations of Hg2+ ions and fluorescence quenching efficiencies of R-PE (0.080 mg mL−1), corresponding to photographs under UV light (Fig. 5B). The calibration curve was obtained with the concentrations of Hg2+ ions ranging from 0.0010 μM to 25.0 μM. A linear regression equation for Hg2+ ions was thereby acquired. The limit of detection (LOD) was estimated to be 0.0130 μM based on the standard deviation rule (LOD = 3Sd/s), which was much lower than the toxicity level of Hg2+ ions in drinking water (30 nM) defined by the World Health Organization (WHO).27 Besides, the detection limit of Hg2+ ions was much lower than that of other fluorescent sensors such as nitrogen-doped carbon quantum dots (N-CQDs) (230 nM),28 fluorescent carbon dots (FCDs) (20 nM)29 and label-free colorimetric sensor (50 nM),30 which were not natural and required high cost. It was widely recognized that the traditional analytical methods such as colorimetric biosensors,25 flame atomic absorption spectrometry26 and electrochemistry,31 can detect Hg2+ ions with high selectivity and sensitivity, but most of them might suffer from expensive specialized instruments, time-consuming and complex operations. Moreover, some organic probes and quantum dots are generally applied but involve either of complicated synthesis and toxic agent, or cost ineffective and time-consuming analysis.28,29 R-PE, as a kind of natural fluorescent protein, is widely existed and easy to purify from Porphyra yezoensis. By contrast, R-PE used in the current work that was obtained from Porphyra yezoensis as fluorescent probe is natural and nontoxic probe with strong fluorescence. The developed fluorimetric analysis with R-PE is rapid (ten minutes), simple, cost-effective, and highly sensitive for detecting Hg2+ ions.
 |
| Fig. 5 (A) Fluorescence spectra of R-PE (0.080 mg mL−1) in the presence of various concentrations of Hg2+ ions (from bottom to top: 0, 0.0010, 0.005, 0.010, 0.10, 0.50, 1.0, 5.0, 25.0 μM) in the test solutions. (B) Fluorescence quenching efficiencies of R-PE (0.08 mg mL−1) versus the logarithmic concentrations of Hg2+ ions (0.0010, 0.0050, 0.010, 0.10, 0.50, 1.0, 5.0, 25.0 μM). All measurements were taken in Tris–HCl (50 mM, pH 8.0) at 40 °C. The error bar represents the standard deviation of five measurements. Insert: photographic images of the corresponding solutions under UV light. | |
3.5 Detecting Hg2+ ions in diverse water samples
In order to evaluate the feasibility in practical applications, the fluorimetric assay mentioned above was employed to detect Hg2+ ions in tap water and sea water (Weihai) samples. The accuracy of the sample analysis was examined through standard addition method. And the detection results were revealed in Table 1. Recoveries of different added Hg2+ concentrations were acquired between 92.0% and 108.0% with the relative standard deviations (RSDs) lower than 1.60%. The results demonstrated that the fluorescent probe of R-PE possessed high sensitivity and selectivity, which made R-PE become an excellent tool to detect Hg2+ ions in water samples.
Table 1 Detection of Hg2+ ions in two kinds of water samples
Sample |
Added (μM) |
Found (μM) |
RSD (%, n = 3) |
Recovery (%) |
Tap water |
0 |
Not detected |
|
|
0.050 |
0.046 |
1.2 |
92.0 |
0.100 |
0.096 |
0.5 |
96.0 |
0.500 |
0.477 |
0.8 |
95.4 |
Sea water (Weihai) |
0 |
Not detected |
|
|
0.050 |
0.054 |
1.6 |
108.0 |
0.100 |
0.106 |
1.3 |
106.0 |
0.500 |
0.523 |
0.4 |
104.6 |
4 Conclusions
In summary, a rapid, sensitive and selective fluorescence detection method has been successfully established for detecting Hg2+ ions utilizing R-PE from Porphyra yezoensis as fluorescent probe. Under the optimal conditions, there is a good linear relationship for Hg2+ ions in the range of 0.0010–25.0 μM, with the LOD was 0.0130 μM. The developed R-PE-based fluorescence method has presented some outstanding advantages over the traditional detection methodologies, such as natural nontoxic fluorescent probe, short reaction time, mild reactive conditions, simple operation, high selectivity and sensitivity. All these remarkable advantages of this method suggest that the proposal has promising application prospects in the field of environment monitoring.
Acknowledgements
This study was supported by National Natural Science Foundation of China (No. 31100037 and 21375075), the National Science & Technology Pillar Program during the 12th Five-year Plan Period (2015BAD17B03), Subject Construction Fund Guided by HIT (WH20150206, WH20160206), Key Research and Development Plan of Shandong Province (2015GSF115016), Science and Technology Development Project of Weihai (2015DXGJZD002).
References
- G. Q. Chen, Z. Guo, G. M. Zeng and L. Tang, Analyst, 2015, 140, 5400–5443 RSC.
- S. Akram, R. Najam, G. H. Rizwani and S. A. Abbas, Pak. J. Pharm. Sci., 2015, 28, 1781–1787 CAS.
- B. M. Fong, T. S. Siu, J. S. Lee and S. Tam, J. Anal. Toxicol., 2007, 31, 281–287 CrossRef CAS PubMed.
- J. L. Manzoori, M. H. Sorouraddin and A. M. H. Shabani, J. Anal. At. Spectrom., 1998, 13, 305–308 RSC.
- G. Bhanjana, N. Dilbaghi, R. Kumar and S. Kumar, Electrochim. Acta, 2015, 178, 361–367 CrossRef CAS.
- W. Chansuvarn, T. Tuntulani and A. Imyim, TrAC, Trends Anal. Chem., 2014, 65, 83–96 CrossRef.
- L. Liang, M. Horvat, H. Li and P. Pang, J. Anal. At. Spectrom., 2003, 18, 1383–1385 RSC.
- G. Mocz, Mar. Biotechnol., 2007, 9, 305–328 CrossRef CAS PubMed.
- H. Shroff, C. G. Galbraith, J. A. Galbraith and E. Betzig, Nat. Methods, 2008, 5, 417–423 CrossRef CAS PubMed.
- M. Fernández-Suárez and A. Y. Ting, Nat. Rev. Mol. Cell Biol., 2008, 9, 929–943 CrossRef PubMed.
- M. Bates, B. Huang, G. T. Dempsey and X. W. Zhuang, Science, 2007, 317, 1749–1753 CrossRef CAS PubMed.
- D. Isailovic, I. Sultana, G. J. Phillips and E. S. Yeung, Anal. Biochem., 2006, 358, 38–50 CrossRef CAS PubMed.
- M. Zhou, C. Cai, J. Liu, Q. Wang, W. Wu and P. He, J. Appl. Phycol., 2010, 22, 459–464 CrossRef CAS.
- N. Zhang, Y. M. Si, Z. Z. Sun, L. J. Chen, R. Li, Y. C. Qiao and H. Wang, Anal. Chem., 2014, 86, 11714–11721 CrossRef CAS PubMed.
- Z. Z. Sun, S. Y. Li, Y. C. Jiang, Y. Qiao, L. Y. Zhang, L. L. Xu, J. H. Liu, W. Qi and H. Wang, Sci. Rep., 2016, 6, 20553 CrossRef CAS PubMed.
- J. F. Niu, Z. F. Chen, G. C. Wang and B. C. Zhou, J. Appl. Phycol., 2010, 22, 25–31 CrossRef CAS.
- S. N. Tripathi, S. Kapoor and A. Shrivastava, J. Appl. Phycol., 2007, 19, 441–447 CrossRef CAS.
- M. Munier, M. Morançais, J. Dumay, P. Jaouen and J. Fleurence, J. Chromatogr. B: Anal. Technol. Biomed. Life Sci., 2015, 992, 23–29 CrossRef CAS PubMed.
- R. Rossano, N. Ungaro, A. D'Ambrosio, G. M. Liuzzi and P. Riccio, J. Biotechnol., 2003, 101, 289–293 CrossRef CAS PubMed.
- A. V. Galland-Irmouli, L. Pons, M. Luçon, C. Villaume, N. T. Mrabet, J. L. Guéant and J. Fleurence, J. Chromatogr. B: Biomed. Sci. Appl., 2000, 739, 117–123 CrossRef CAS.
- J. F. Niu, G. C. Wang and C. K. Tseng, Protein Expression Purif., 2006, 49, 23–31 CrossRef CAS PubMed.
- S. Li and S. Dong, J. Mater. Chem., 2008, 18, 4636–4640 RSC.
- B. Gu, L. Huang, N. Mi, P. Yin, Y. Zhang, X. Tu, X. Luo, S. Luo and S. Yao, Analyst, 2015, 140, 2778–2784 RSC.
- D. Huang, C. Niu, M. Ruan, X. Wang, G. Zeng and C. Deng, Environ. Sci. Technol., 2013, 47, 4392–4398 CrossRef CAS PubMed.
- Y. Wang, F. Yang and X. Yang, Biosens. Bioelectron., 2010, 25, 1994–1998 CrossRef CAS PubMed.
- T. Madrakian, B. Zadpour, M. Ahmadi and A. Afkhami, J. Iran. Chem. Soc., 2015, 12, 1235–1243 CrossRef CAS.
- L. Zhang, T. Li, B. Li, J. Li and E. Wang, Chem. Commun., 2010, 46, 1476–1478 RSC.
- R. Zhang and W. Chen, Biosens. Bioelectron., 2014, 55, 83–90 CrossRef CAS PubMed.
- Z. H. Gao, Z. Z. Lin, X. M. Chen, Z. Z. Lai and Z. Y. Huang, Sens. Actuators, B, 2016, 222, 965–971 CrossRef CAS.
- T. Li, S. J. Dong and E. K. Wang, Anal. Chem., 2009, 81, 2144–2149 CrossRef CAS PubMed.
- R. M. Kong, X. B. Zhang, L. L. Zhang, X. Y. Jin, S. Y. Huan, G. L. Shen and R. Q. Yu, Chem. Commun., 2009, 37, 5633–5635 RSC.
|