M. V. Russo*a,
P. Avinob and
E. Veschettic
aUniversità degli Studi del Molise, Dipartimento Agricoltura, Ambiente e Alimenti (DiAAA), via De Sanctis, Campobasso 86100, Italy. E-mail: mvrusso@unimol.it; Fax: +39 0874 404652; Tel: +39 0874 404631
bDIPIA-INAIL settore Ricerca, Via IV Novembre 144, Rome 00184, Italy
cIstituto Superiore di Sanità, Dipartimento Ambiente e connessa Prevenzione Primaria, Viale Regina Elena 299, Rome 00161, Italy
First published on 8th January 2015
A rapid analytical screening method allowing simultaneous analysis of few halogenated persistent organic pollutants (POPs) in water samples at ultra-trace levels was developed. Two-metre long capillary traps with an inner diameter of 320 μm, internally coated with 1.2 μm thick 5%-diphenyl-95%-dimethylpolysiloxane or 14%-cyanopropylphenyl-86%-dimethyl-polysiloxane stationary phase, were used to extract some chlorinated pesticides (aldrin, heptachlor, heptachlorepoxide, dieldrin, endrin, 4,4′-DDE, α-endosulfan) and polychlorobiphenyls (PCBs 1, 15, 44, 77, and 180) from aqueous media. Water samples were pushed through the traps at constant velocity using nitrogen. Every trap was installed in a GC oven and connected in series to an analytical column (25 m long, 320 μm I.D. of CP-Sil 8 CB with a 1.2 μm film thickness) installed in a second GC equipped with an ECD detector (GC-GC tandem system). The capillary trap in the first GC was quickly heated from 50 to 280 °C to focus the retained analytes into a narrow zone between the trap outlet and the analytical column inlet, temporarily kept at room temperature outside the first GC oven. After moving this zone inside the oven of the second GC, focused solutes were thermally desorbed and separated into the analytical column by programming the temperature of the second GC. Extraction recovery was always greater than 70–80% and nearly quantitative for most of the analytes. The sorptive properties of the two stationary phases were independent of sample volume and velocity but were slightly influenced by their polarity. The latter effect was used to investigate the possibility of fractionating the two classes of compounds. For this purpose, a capillary trap containing 14%-cyanopropylphenyl-86%-dimethylpolysiloxane was connected in series to a second capillary trap containing 5%-diphenyl-95%-dimethyl-polysiloxane immediately before the trapping step. Adsorbed solutes were fractioned between the two traps by eluting 0.5 mL of water–methanol 40:
60 v/v. Most of the tested compounds were retained exclusively by one of the two stationary phases. Overall, the proposed method proved to be practically insensitive to laboratory contamination, reproducible, and suitable for the determination of halogenated POPs at trace levels (LoD in the range 5–50 pg L−1).
Although their determination in solid matrices such as soils, sediments, and biological tissues have been well established, their quantification in aqueous samples is still required. This is mainly due to the extremely low concentrations at which every compound may be found, that requires enrichment and clean-up steps, being prone to a number of interferences in the laboratory environment.
Much effort has been devoted during the last two decades to the development of faster, safer and more reliable pre-concentration techniques for these pollutants. In particular, solvent-less procedures such as solid phase micro-extraction (SPME) and stir-bar sorptive extraction (SBSE) were applied to reduce contaminations and artefacts introduced using classical well-established methods like liquid–liquid (LLE) and solid-phase (SPE) extractions. SPME uses 1 cm extractive fibres externally coated with not more than 100 μm of polymeric sorbent.3,4 During extraction, the fibre remains immersed in the aqueous sample for a known amount of time, generally less than 30 min, after which it is desorbed into a conventional heated GC injector. SBSE uses small glass-encapsulated magnetic stir-bars usually covered with 500 μm of polymeric sorbent, which are swirled into the samples for some length of time and then thermally desorbed in a dedicated GC injector.5 SPME is less concerned with ghost peaks or carryover than SBSE but detection limits are worse than those obtainable by the latter. In both cases, additional band focusing by cold-trapping with column temperature programming or rather external cryofocusing may be necessary to reduce peak tailing.
In order to avoid artefacts such as ghost peaks and thermal decompositions caused by heated injectors during the desorption step, a complementary approach to SPME and SBSE, called in-tube SPME, was recently proposed by our research group6–9 and other authors:10–21 it is based on the use of capillary extractors made of short pieces (usually 5–30 cm long) of coated capillaries, trimmed from conventional high-resolution GC columns and carrying glass press-fits at their ends. Target analytes are first trapped by the polymeric sorbent inside the capillary extractor while a sample plug is made to flow at constant speed by applying a nitrogen overpressure on the sample surface. After removing water by purging the inert gas at low flow rates, trapped compounds are directly desorbed into the GC column without using any heated injector and subsequently refocused as described previously for SPME and SBSE. By acting on stationary phase thickness, trap length and internal diameter, extraction of analytes from aqueous samples may be non, partially or completely depletive. In the first case the volume of the stationary phase is low and swelling of the stationary phase is needed prior to sorption. Moreover, mixing conditions in the trap are rather poor and sorption equilibrium cannot be obtained in a single sorption cycle; therefore, several sorption cycles are needed to approach the equilibrium. On the contrary, in the other two cases the quantity of stationary phase inside every capillary extractor is usually sufficient to retain quantitatively, or nearly so, target analyets in one single extraction.
The present article explores the performances of this technique in depletive conditions for the extraction, pre-concentration and subsequent GC analysis of 12 chlorinated POPs (7 chlorinated pesticides and 5 polychlorobiphenyls) using short open tubular capillary columns internally coated with non-polar or medium-polar stationary phases. The effect of experimental conditions on the recovery of the target analytes is examined in details.
Stock solutions (0.1 g L−1) of pesticides and PCBs was prepared by dissolving every product in acetone. These standards were then diluted with acetone to prepare intermediate mix solutions used to spike aqueous samples. All standard solutions were kept in a refrigerator at 4 °C and were stable for 1 month, at least. During the preparation of aqueous samples containing 0.2–240 μg L−1 of each target analyte, no more than 10 μL of the appropriate mixture in acetone was introduced per mL of water.
Two different real aqueous matrices (that is, water samples collected from rivers Aniene and Tiber at Rome, Italy) were used to test the matrix effect on analyte recovery. The water samples, respectively containing a total organic carbon (TOC) of 16.7 and 14.6 mg L−1 determined in accordance with the American Public Health Association (APHA)'s standard method,23 were filtered through 25 mm diameter, glass-fibre membranes with a pore size of 0.2 μm (Alltech Deerfield, IL, USA) and spiked with an intermediate standard mixture of target analytes directly before capillary extraction.
Sampling step was performed with a capillary trap, disconnected from the analytical column, using the equipment shown in Fig. 1 unless otherwise indicated. A known volume of aqueous sample was pushed through the trap at room temperature (18–30 °C) by nitrogen humidified with demineralized water to prevent solvent evaporation in the sample reservoir. Sample flow rate was controlled (SD: ±0.1 cm s−1) by acting on course and fine valves while timing liquid front or tail into the capillary trap. After sampling, the trap was washed with 1.0 mL of demineralised water introduced at the maximum speed (near 20 cm s−1) to remove soluble inorganic salts, left by the sample, without significantly altering the profile of adsorbed solutes.7
![]() | ||
Fig. 1 Sampling apparatus. 1: N2 99.999%, 3.0 kg cm−2; V1: coarse needle valve; V2: fine needle valve; D1: nitrogen humidifier (10 mL); D2: sample reservoir (1.5–10.0 mL); CT: capillary trap. |
Finally, the trap was emptied of residual liquid and dried with nitrogen for 5–10 min.
In a following step, the focused solutes were thermally desorbed and separated into another 25 m long, 320 μm I.D. GC column internally coated with 1.2 μm of Chrompack CP-Sil 8 CB. This analytical column was permanently installed in a second gas chromatograph (DANI 6500) equipped with a 63Ni electron capture detector (ECD) connected to a personal computer implementing a Chemstation CSW32 v.1.4 software (Data Apex Ltd. 2002, Prague, The Czech Republic). A deactivated press-fit connector (Carlo Erba, Milan, Italy) was used to interface the outlet of the trap to the analytical column. Thermal desorption of the analytes previously focused at the outlet of the trap and the consequent chromatographic separation were started by moving manually the last 20 ± 1 cm of the trap inside the second GC oven and heating the analytical column as follows: 50 °C for 1 min, 20 °C min−1 from 50 to 170 °C, 3 °C min−1 from 170 to 280 °C, 280 °C for 5 min. At the same time, the first GC oven was set at 50 °C.
The ECD temperature was kept at 300 °C during the whole operating sequence while nitrogen was made to flow at 30 mL min−1 as detector make-up gas. Fig. 2 depicts a schematic representation of the GC-GC-ECD tandem system.
Quantitative results acquired by applying this procedure were compared with the corresponding data obtained after injecting 1.0 μL of a standard solution in the S/SL injector working in splitless mode for 5 min at the temperature of 290 °C. During this calibration step, the tandem system was operated in the same way as described above.
Hydrogen at 40 cm s−1 was chosen as carrier gas. Every gas chromatographic test was replicated three times.
![]() | ||
Fig. 3 Sampling apparatus used to fractionate different classes of chlorinated POPs. CT1: capillary trap containing CP-Sil 19 CB as sorbent film; CT2: capillary trap containing CP-Sil 8 CB as sorbent film; PFC: press-fit connector. Other symbols as in Fig. 1. |
Compound | Added quantity (pg) | Sample volume eluted inside CP-Sil 8 CB traps | Sample volume eluted inside CP-Sil 19 CB traps | ||||||
---|---|---|---|---|---|---|---|---|---|
1.0 | 5.0 | 10.0 | 20.0 | 1.0 | 5.0 | 10.0 | 20.0 | ||
a Sampling velocity: 1.2 cm s−1; standard deviations: 4–6% (three replicates). | |||||||||
Heptachlor | 10 | 100 | 101 | 100 | 100 | 100 | 100 | 100 | 100 |
Aldrin | 10 | 71 | 71 | 76 | 74 | 100 | 100 | 100 | 100 |
Heptachlor epoxide | 10 | 99 | 100 | 99 | 100 | 100 | 100 | 100 | 100 |
α-Endosulfan | 10 | 99 | 101 | 100 | 99 | 100 | 100 | 100 | 100 |
Dieldrin | 10 | 100 | 100 | 100 | 101 | 100 | 100 | 100 | 100 |
Endrin | 10 | 100 | 100 | 100 | 100 | 100 | 101 | 100 | 99 |
4,4′-DDE | 10 | 80 | 77 | 75 | 79 | 100 | 100 | 100 | 100 |
PCB 1 | 240 | 99 | 101 | 100 | 101 | 100 | 100 | 99 | 99 |
PCB 15 | 120 | 101 | 100 | 99 | 101 | 100 | 100 | 99 | 99 |
PCB 44 | 60 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
PCB 77 | 60 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
PCB 180 | 20 | 70 | 69 | 68 | 67 | 82 | 81 | 81 | 78 |
All the tested compounds were not affected by the volume of the aqueous sample introduced in the capillary traps containing one of the two sorbent films examined in the present research. This means that it is possible to sample volumes up to 20 mL, at least, without compromising the retention of the 12 analytes.
Compound | Added quantity (pg) | Sampling velocity | |||
---|---|---|---|---|---|
1.2 | 2.4 | 3.6 | 5.2 | ||
a Sample volume: 5.0 mL; standard deviations: 3–5% (three replicates). | |||||
Heptachlor | 10 | 100 | 100 | 101 | 100 |
Aldrin | 10 | 101 | 100 | 100 | 101 |
Heptachlor epoxide | 10 | 100 | 100 | 100 | 100 |
α-Endosulfan | 10 | 99 | 99 | 100 | 100 |
Dieldrin | 10 | 101 | 99 | 100 | 100 |
Endrin | 10 | 100 | 100 | 100 | 100 |
4,4′-DDE | 10 | 101 | 100 | 100 | 101 |
PCB 1 | 240 | 101 | 100 | 99 | 100 |
PCB 15 | 120 | 101 | 99 | 100 | 100 |
PCB 44 | 60 | 100 | 100 | 100 | 100 |
PCB 77 | 60 | 100 | 100 | 100 | 100 |
PCB 180 | 20 | 82 | 83 | 85 | 82 |
In the same experimental conditions, the limit of detection (LoD) estimated according to Knoll's definition (analyte concentration that produces a chromatographic peak equal to three times the standard deviation of the baseline noise)24 was 9 pg L−1 for heptachlor, 14 pg L−1 for aldrin, 8 pg L−1 for heptachlor epoxide, 9 pg L−1 for α-endosulfan, 12 pg L−1 for dieldrin, 13 pg L−1 for endrin, 14 pg L−1 for 4,4′-DDE, 50 pg L−1 for PCB 1, 32 pg L−1 for PCB 15, 15 pg L−1 for PCB 44, 12 pg L−1 for PCB 77, and 5 pg L−1 for PCB 180. These values are sufficiently low to enable the direct determination of chlorinated POPs at the concentration levels found in most of polluted natural waters (Table 3).
POPs | Aqueous matrix | Individual POP concentration range (ng L−1) | Reference |
---|---|---|---|
a n.d.: lower than detection limit, WPCP: water pollution control plant. | |||
OCPs | Chao river (China) | n.d.–1.8 | 25 |
PCBs | Chao river (China) | n.d.–0.01 | 25 |
OCPs | Selangor river (Malaysia) | 0.6–25 | 26 |
OCPs | Pangani river (Tanzania) | n.d.–4.5 | 27 |
PCBs | Khour-e-Mousa, Persian gulf (Iran) | n.d.–120 | 28 |
OCPs | Densu river (Ghana) | n.d.–260 | 29 |
PCBs | North West Persian gulf (Iran) | n.d.–100 | 30 |
OCPs | Coastal marine environment (Singapore) | n.d.–1.7 | 31 |
PCBs | New York/New Jersey WPCP (USA) | n.d.–29 | 32 |
Compound | Added quantity (pg) | DW volume | RA volume | RT volume | |||
---|---|---|---|---|---|---|---|
1.0 | 5.0 | 1.0 | 5.0 | 1.0 | 5.0 | ||
a DW: demineralized water; RA: aliquot of the river Aniene; RT: aliquot of the river Tiber; standard deviations: 4–6% (three replicates). | |||||||
Heptachlor | 1 | 100 | 100 | 97 | 97 | 98 | 97 |
Aldrin | 1 | 100 | 100 | 97 | 97 | 97 | 97 |
Heptachlor epoxide | 1 | 100 | 100 | 98 | 97 | 98 | 98 |
α-Endosulfan | 1 | 100 | 100 | 99 | 97 | 98 | 98 |
Dieldrin | 1 | 100 | 100 | 97 | 97 | 98 | 99 |
Endrin | 1 | 100 | 101 | 97 | 96 | 97 | 97 |
4,4′-DDE | 1 | 100 | 100 | 97 | 98 | 98 | 98 |
PCB 1 | 240 | 100 | 100 | 99 | 99 | 99 | 98 |
PCB 15 | 120 | 100 | 100 | 99 | 99 | 99 | 99 |
PCB 44 | 60 | 100 | 100 | 100 | 100 | 100 | 100 |
PCB 77 | 60 | 100 | 100 | 100 | 100 | 100 | 100 |
PCB 180 | 20 | 82 | 81 | 79 | 78 | 78 | 78 |
Fig. 4A shows a typical chromatogram produced after sampling 1.0 mL of water collected from the river Tiber, filtered and spiked with 20 ng of Arochlor 1232 immediately before the trapping step. The resulting profile may be compared with the chromatogram obtained by injecting the same amount of the PCB mixture in the GC column through the conventional split/splitless injector (Fig. 4B). Every chromatogram is made up of two signal profiles: the first one (GC1) represents the ECD response recorded during the preliminary focalization of the trapped analytes at the end of the capillary trap while the second one (GC2) represents the signal produced by the following gas chromatographic separation. By comparing the two GC2 profiles it can be seen that the inevitable loss of chromatographic resolution due to the sampling technique was almost efficiently contrasted by the well-known re-focusing action at the end of the trap operated by programming the temperature in the first GC.
Compound | Added quantity (pg) | Recovery from CP-Sil 19 CB | Recovery from CP-Sil 8 CB | Total recovery |
---|---|---|---|---|
a Sample volume eluted during the trapping step: 5.0 mL; sample and fractionating eluent velocity: 1.2 cm s−1; standard deviations: 3–4% (three replicates). | ||||
Heptachlor | 5 | 0 | 100 | 100 |
Aldrin | 5 | 0 | 100 | 100 |
Heptachlor epoxide | 5 | 0 | 100 | 100 |
α-Endosulfan | 5 | 0 | 100 | 100 |
Dieldrin | 5 | 0 | 100 | 100 |
Endrin | 5 | 20 | 80 | 100 |
4,4′-DDE | 5 | 0 | 100 | 100 |
PCB 1 | 150 | 0 | 0 | 0 |
PCB 15 | 100 | 45 | 55 | 100 |
PCB 44 | 35 | 65 | 35 | 100 |
PCB 77 | 30 | 97 | 3 | 100 |
PCB 180 | 10 | 82 | 18 | 100 |
A better separation of the two classes of compounds could be achieved by acting on the percentage of the methanol in water, the overall volume of the fractionating solvent as well as the relative order of two traps with respect to the flow direction. These aspects will be investigated thoroughly in the near future.