An ultrasensitive electrochemical immunosensor for determination of estradiol using coralloid Cu2S nanostructures as labels

Sen Zhangab, Yaoguang Wanga, Yunhui Zhangb, Tao Yana, Liangguo Yanb, Qin Wei*a and Bin Duab
aKey Laboratory of Chemical Sensing & Analysis in Universities of Shandong, School of Chemistry and Chemical Engineering, University of Jinan, Jinan 250022, P. R. China. E-mail: sdjndxwq@163.com; Fax: +86-531-82765969; Tel: +86-531-82767872
bSchool of Resources and Environment, University of Jinan, Jinan 250022, P. R. China

Received 24th October 2014 , Accepted 17th December 2014

First published on 17th December 2014


Abstract

Herein, coralloid Cu2S nanoparticles were prepared through a facile etching method and used as labels for the first time to fabricate an electrochemical immunosensor. A novel competitive immunoassay was then proposed using Cu2S covalent conjugation with bovine serum albumin (BSA)–estradiol (E2) for the sensitive detection of trace E2 concentrations. Without using enzyme-label and acid dissolution, the Cu2S can generate excellent electrochemical signals. The quantitative detection was based on the competitive binding of the E2 antibody with Cu2S-labeled E2 or free E2. The redox signal decreased with increasing concentration of the free E2 as the amount of Cu2S–BSA–E2 labels decreased at the immunosensor probe. Cyclic voltammetry and electrochemical impedance spectroscopy techniques were used to characterize the immunosensor. Square wave voltammetry was used to monitor the electrochemical response. The immunosensor exhibited a wide linear response from 25 to 7500 pg mL−1 with a detection limit of 7.5 pg mL−1. The proposed method showed good precision, broad linear range, and acceptable stability and could be used for the detection of E2 in real samples, which showed promising application in field research.


Introduction

Although estradiol (E2) is endogenously produced by all mammalian species, several studies have indicated that E2 is frequently released into the water environment through human activities.1,2 As a typical endocrine disrupting compound, E2 is toxic to the environment because of its potential effects on animal reproductive efficiency and the immune functions of aquatic systems.3,4 What is more, the presence of low concentrations of E2 in the environment would act on humans, cause abnormal sexual development and decrease the average number of human spermatozoa.5 Therefore, it is very important to establish a simple, sensitive and rapid method for the detection of E2 in the environment to protect the health of animals and human.

Various methods have become powerful for E2 analysis, such as high performance liquid chromatography,6 molecular imprinting technology,7 chemiluminescence8 and surface plasmon resonance sensing.9 Although determination of E2 by these methods is promising because of their high selectivity and sensitivity, these instruments are expensive, the operating processes are complicated and long, and pretreatment of samples is involved. Recently, several studies have focused on fabricating electrochemical sensors for ultrasensitive and fast detection of E2. From the molecular structure, E2 has electrochemical active due to its phenolic hydroxyl group, which can be oxidized on the electrode surface. However, direct electrochemical detection of E2 is very limited. He et al. used nano-Al2O3 film modified glassy carbon electrode (GCE) to detect E2 and got detection limit of 21.79 ng mL.10 By combining immunoassay with electrochemistry, high analytical speed, excellent selectivity, high sensitivity and economics, especially low detection limit can be obtained. Based on the highly specific molecular recognition of the immunoreaction, the E2 detection limit could generally achieve at picogram level. Many groups have fabricated electrochemical immunosensors for the determination of E2. Liu et al. developed an electrochemical immunosensor with underpotential deposition Cu|DTBP-Protein G scaffold to detect E2. The determination of E2 was based on the electrochemical signal produced by square wave voltammetry (SWV) oxidation of ferrocenemethanol in the detection solution. The increased amounts of sample E2 decreased the BSA–E2 conjugation on the immunosensor surface, which resulted in charge transfer resistance decrease of the immunosensor.11 Ojeda et al. investigated an immunosensor which performed by applying a competitive immunoassay with peroxidase-labelled E2 (HRP-E2) and measured the amperometric response at −200 mV using hydroquinone (HQ) as redox mediator.12 Chaisuwan et al. developed a competitive format to determine trace E2 concentrations using CdSe quantumdots (QDs) conjugation with BSA–E2. They used the in situ bismuth-coated carbon electrodes for detecting the cadmium ions (Cd2+) which was released from CdSe QDs during the acid dissolution step.13 The above mentioned methods have obtained favorable results, whereas development of an enzyme-free and without acid dissolution immunosensor is necessary. Cuprous sulfide (Cu2S) nanomaterial has widely utilized in solar cells, optical filters, photoelectron slice and sensors.14 As a novel nanomaterial, Cu2S has been used in sensors to detect hydrogen peroxide,15 glucose,16 polyphenols17 or ammonia gas.18 Cu2S used in fabricating sensors were based on different synthetic methods and showed different morphology. And to the best of our knowledge, there are few reports based on Cu2S for fabrication of immunosensor. In the present work, coralloid Cu2S was prepared by a simple etching method. The process of preparing the Cu2S is cheap, convenient and environmentally friendly. Due to the good electroactivity of Cu2S, it could produce obvious redox signal in moderate environment. Using Cu2S as labels could effectively avoid the shortcoming of enzyme labels. In addition, it does not need to release ions with HCl or HNO3, which is time-consuming and the acid may lead to the denaturizing of the biomolecules.

Herein, we report on a novel electrochemical immunosensor for the determination of E2 by using Cu2S incubated BSA–E2 conjugates modified gold electrode based on competitive immunoassay mode. The electrochemical signal was generated by the redox reaction of Cu2S under SWV method. The concentration of E2 can be reflected directly because the decreased free E2 concentration in sample will increase the binding site of Cu2S–BSA–E2 captured with the E2 antibody. The method has been successfully applied to analyze E2 in water samples and shows good recovery and accuracy, demonstrating its potential use in real sample analysis.

Experimental

Reagent and materials

Mouse monoclonal estradiol antibody (E2 Ab), E2 and BSA–E2 were purchased from Donglinchangsheng Biotechnology Co., Ltd. (Beijing, China). Mercaptoacetic acid, 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide hydrochloride (EDC), N-hydroxysuccinimide (NHS), glutaraldehyde, high purity copper sheet, thiourea, ethylenediamine and other chemicals were obtained from Sinopharm Chemical Reagent Co., Ltd (Beijing, China). Phosphate buffered saline (PBS, 1/15 M Na2HPO4 and KH2PO4) was used as an electrolyte for all electrochemistry measurement. Ultrapure water was used throughout the experiment. All the chemicals were of analytical reagent grade.

All electrochemical behaviors were performed by a traditional three-electrode system. It contained a gold electrode (4 mm diameter) as the working electrode, a platinum wire electrode as the auxiliary electrode and a saturated calomel electrode (SCE) as the reference electrode. All electrochemical measurements were carried out using a CHI 760D electrochemical workstation (Chenhua Instrument Shanghai Co. Ltd, China). Transmission electron microscope (TEM) images were obtained from an H-800 microscope (Hitachi, Japan). Fourier transform infrared spectroscopy (FTIR) spectrum was obtained from VERTEX 70 (Germany). X-ray powder diffraction (XRD) was gathered from a Bruker D8 Focus diffractometer (Germany) using CuKα radiation (40 kV, 30 mA) of wavelength 0.154 nm.

Preparation of coralloid Cu2S

Coralloid Cu2S was synthesized as described by Zhang et al. with slight modifications.14 Briefly, copper sheet (1 cm2) was firstly polished by fine sandpaper. Then a clean copper sheet was prepared from consecutive ultrasonication in ethanol. Subsequently, it was immersed in a 5 mL test tube, which contains 0.8 mL 0.01 M thiourea and 0.1 mL ethylenediamine at room temperature for 3 days. Ethylenediamine has strong basicity and corrosivity. It could regulate the morphologies of nanomaterials. Thiourea could provide S element which constituted Cu2S nanoparticles. Furthermore, ethylenediamine and thiourea both contain amine groups, which may result in the amine groups linking on Cu2S. Finally, the nanoparticles were released from the Cu sheet by ultrasonication and washed with distilled water and ethanol for several times to remove the impurities before characterization.

Cu2S conjugation to BSA–E2

The conjugation procedure of Cu2S to BSA–E2 was as follows: Cu2S was first pretreated by ultrasonication to disperse evenly. Then it was mixed with 0.5 mL of 2% glutaraldehyde solution and stirred for 2 h at room temperature. Subsequently, residual glutaraldehyde was removed by washing three times in PBS. The products were then centrifuged and dispersed in 0.5 mL of pH 7.4 PBS. Finally, 4 μg of BSA–E2 was added into the aforesaid products and shaken for 12 h at 4 °C. The amine groups on the Cu2S surface were activated with glutaraldehyde to give a stable intermediate derivative which is used for covalent binding BSA–E2.19 The reaction mixture was then centrifuged and the supernatant was removed. Cu2S–BSA–E2 bioconjugates were treated with 1 mL 1% BSA solution for 30 min to block non-specific binding. The mixture was centrifuged using PBS to remove any free E2 and BSA. The final mixture was redispersed in 1.0 mL PBS (pH 7.4) and stored at 4 °C before use. The synthesis of Cu2S–BSA–E2 conjugates is schematically summarized in Fig. 1A.
image file: c4ra13066a-f1.tif
Fig. 1 (A) Cu2S incubate with BSA–E2 by covalent immobilization using glutaraldehyde and (B) schematic diagram of the detection principle of E2 with competitive immunoassay mode.

Preparation of competitive immunoassay

A gold electrode was first polished to a mirror-like finish with 0.05 μm alumina slurries to remove adsorbed impurities. After removal of the trace alumina from the electrode surface, the electrode was cleaned by ultrapure water and ethanol in an ultrasonic bath. Then the pretreated electrode was transferred to an electrochemical cell for cleaning by cyclic voltammetry (CV) between −0.4 and +1.5 V versus SCE at 50 mV s−1 in 0.1 M H2SO4 until a stable CV profile was obtained. After rinsing, the mercaptoacetic acid covered electrodes were prepared by immersing the cleaned gold electrodes in an ethanol solution containing 10 mM mercaptoacetic acid in the darkness for 12 h.20 E2 Ab was immobilized onto gold electrode by cross-linking E2 Ab onto the mercaptoacetic acid surface via the EDC/NHS-mediated amine coupling reaction. Briefly, mercaptoacetic acid covered electrodes were soaked in a solution comprised of 0.2 M EDC and 0.1 M NHS for 15 min.21 After cleaning, 4 μL E2 Ab (100 ng mL−1) was dropped on the pretreated gold electrode, and incubated for 2 h at 4 °C. In order to minimize nonspecific binding, 2 μL 1% wt BSA was then added and incubated for 2 h at 4 °C. After drying, the electrodes were washed with pH 7.4 PBS. Then, a solution was prepared by mixing Cu2S–BSA–E2 bioconjugation and an unknown amount of free E2. Finally, 6 μL of aforesaid mixture solution was dropped onto electrode and reaction for another 1 h at 4 °C. The procedures used for the construction of the immunosensor are depicted in Fig. 1B.

Measurement procedure

The PBS (pH 6.6) was used for all the electrochemical measurements. Electrochemical signals were measured using SWV technique. For SWV measurement of the immunosensor, it was scanned from −0.3 V to +0.6 V with a potential step of 5 mV, a frequency of 25 Hz, and amplitude of 25 mV.

Results and discussion

Characterization of Cu2S and Cu2S–BSA–E2 bioconjugates

Under the simple etching method, a layer of dense and black Cu2S was grown onto pure copper sheet (Fig. 2A). A SEM image in Fig. 2B shows the morphology of the prepared Cu2S and predicts that they were coralloid in shape. XRD analysis was used to characterize the crystalline structure of Cu2S. According to Fig. 2C, the strong peaks at 37.43°, 45.77°, 48.48° and 53.67° are originated from (102), (110), (103) and (112) lattice planes, respectively. Moreover, the weak peaks at 26.49°, 29.20°, 54.56° and 55.22° are related to the lattice planes of (002), (101), (004) and (201), respectively.22 All the diffraction peaks matched well with those from the Jade PDF card (26-1116) for Cu2S. Energy dispersive spectrometer (EDS) analysis of Cu2S was further confirmed that the prepared nanomaterial was Cu2S. Cu2S was dropped on the aluminum foil to test. From Fig. 2D, it shows three major elements in the sample: Cu, S and Al. Obviously, Cu and S came from the Cu2S, while Al originated from the aluminum foil. FTIR was used to characterize the functional groups on Cu2S (Fig. 2E). A band at about 3129 cm−1 in the spectrum showed N–H stretching vibration and a band at about 1623 cm−1 showed N–H bending vibration.23 The results indicated that amino groups have been grafted onto Cu2S. The Cu2S–BSA–E2 bioconjugates were characterized by UV-visible spectroscopy (Fig. 2F). All samples were dispersed in ultrapure water and mixed thoroughly. After laying aside for 2 min at room temperature, the UV-visible spectroscopy was monitored from 200 nm to 800 nm. For Cu2S (curve a), there was no obvious absorption in visible region.24 For BSA–E2 solution, there was a series of fluctuation at approximately 220–500 nm (curve b) and the peak at 280 nm was the characteristic absorption peak of protein BSA. When BSA–E2 was bonded onto Cu2S, it was obvious that the spectrum of Cu2S–BSA–E2 bioconjugates (curve c) was a superposition of curve a and curve b. Electrochemical methods are also suitable for investigating the interaction between biological molecules and other materials.25 Due to the proteins are intrinsically unable to act as redox partners, the combined biological molecules will weaken the electrochemical signals obviously. From ESI Fig. S1, the Cu2S shows an enormous peak (curve a). However, after BSA–E2 combined on Cu2S, the Cu2S–BSA–E2 bioconjugation shows a relatively small peak (curve b). All the results show that BSA–E2 was conjugated onto Cu2S successfully.
image file: c4ra13066a-f2.tif
Fig. 2 Photograph (A) of (a) clean copper sheet and (b) covered Cu2S; SEM images (B), XRD spectrum (C), EDS (D) and FTIR (E) spectrum of Cu2S and UV-visible spectrum (F) of (a) Cu2S, (b) E2–BSA and (c) Cu2S–E2–BSA.

Electrochemical behavior of the immunosensor

Some parameters and optimization of conditions were shown in ESI. The electrochemical behavior of Cu2S–BSA–E2 bioconjugates was investigated before such competitive immunoassay mode established. When Cu2S–BSA–E2 bioconjugates without free E2 coupled to the anti-E2 modified electrode, an obvious SWV peak can be seen from Fig. 3A curve b. However, there is no signal generated when only the free E2 coupled to the anti-E2 modified electrode (curve a). Thus, Cu2S exhibited strong redox electrochemical signal and this performance was also detected under CV technique. After Cu2S–BSA–E2 bioconjugates combined on the immunosensor, it shows a pair of strong redox signal (Fig. 3B curve b). But the free E2 does not show any peak (curve a). A reversible system can be seen from the CV. The reaction occurred on the electrode can be speculated according to the Nernst equation:26
image file: c4ra13066a-t1.tif
where R is constant term, R = 8.314 J K−1 mol−1; F is faraday constant, F = 96[thin space (1/6-em)]485 J mol−1 V−1; n is electron transfer number. n can be calculated and equals about 1. Based on that, the redox reaction and the generated electrical signals may be produced by the mutual transformation of cuprous ion and cupric ion. Based on such appearance, Cu2S can be used as a commendable label in fabrication of immunosensor.

image file: c4ra13066a-f3.tif
Fig. 3 SW (A) and cyclic (B) voltammograms of the immunosensor, (a) without Cu2S labels, (b) Cu2S labels.

Electrochemical impedance spectroscopy (EIS) characterization of the immunosensor

EIS is one of the most powerful tools to investigate the interface properties of surface-modified electrodes and it was carried out to illustrate that all fabricating steps of the immunosensor were effective. It is well known that the high frequency region of the impedance plot shows a semicircle related to the redox probe Fe(CN)63−/4−, followed by a Warburg line in the low frequency region which corresponds to the diffusion step of the over all process.27 Curve a in Fig. 4 shows EIS of the bare gold electrode. It presented a straight line, implying that a very low electron transfer resistance in the electrolyte solution. After mercaptoacetic acid and anti-E2 covalently binding, an increased semicircle at high frequency region was observed (curve b). The reason of Ret increasing was that the nonconductive antibody obstructed the electron transfer of the redox probe from solution to electrode surface. The adsorption of BSA at the E2 Ab/gold electrode resulted in a dramatically increased diameter of the semicircle (curve c), indicating a higher electron transfer resistance at the electrode interface. After modified Cu2S–BSA–E2 and 50 pg mL−1 E2 on the electrode, the electron transfer resistance decreased (curve d). This decrease is attributed to the good electrical conductivity of Cu2S, which presumably accelerates electron transfer at the electrochemical probe. The above results clearly confirm that the immunosensor had been fabricated successfully.28
image file: c4ra13066a-f4.tif
Fig. 4 EIS of gold electrode (a), E2 Ab/gold electrode (b), BSA/E2 Ab/gold electrode (c) and Cu2S–BSA–E2 with 50 pg mL−1 E2/BSA/E2 Ab/gold electrode (d) in 0.1 mol L−1 KCl solution containing 10 mmol L−1 [Fe(CN)6]3−/4− (1[thin space (1/6-em)]:[thin space (1/6-em)]1).

Analytical performance characteristics

SWV was used to evaluate the performance of the immunosensor. The peak currents of SWV were decreased with increasing concentration of free E2 under optimized conditions. Fig. 5A respectively shows the SWV response of 25 pg mL−1, 50 pg mL−1, 75 pg mL−1, 100 pg mL−1, 250 pg mL−1, 500 pg mL−1, 750 pg mL−1, 1000 pg mL−1, 2500 pg mL−1, 5000 pg mL−1 and 7500 pg mL−1 (a–k, in order) of free E2 concentrations. The currents changed linearly with the concentrations of E2 in the range from 25 pg mL−1 to 7500 pg mL−1 with a detection limit of 7.5 pg mL−1. The linear regression equations was i (μA) = 28.8843 − 4.9857[thin space (1/6-em)]log[thin space (1/6-em)]c (pg mL−1) with a statistically significant correlation coefficient of 0.9935 (Fig. 5B). The analytical performance of this fabricated immunosensor was compared with previously reported competitive immunoassay for the determination of estradiol.13,29–31 The comparison results are presented in Table S1. It can be seen that the developed immunosensor is more comparable, convenient, inexpensive and even exhibits better analytical performance toward estradiol than previously reported other methods.
image file: c4ra13066a-f5.tif
Fig. 5 SW voltammogram (A) of 25, 50, 75, 100, 250, 500, 750, 1000, 2500, 5000 and 7500 pg mL−1 (a–k, inorder); the calibration curve (B) of the current values vs. the E2 concentration. Error bar = RSD (n = 5).

Moreover, the reproducibility, selectivity, and stability of the immunosensor were also investigated. To evaluate the reproducibility of the immunosensor, a series of five electrodes were prepared for the detection of 50 pg mL−1 E2. The relative standard deviations (RSD) of the measurements for the five electrodes were 4.7%. The results indicated that the precision and reproducibility of the proposed immunosensor was quite well.

A series of small organic chemicals (each 5 ng mL−1) were selected, saccharose, ascorbic acid, bisphenol A and glucose, owing to their structural similarity to E2, as negative controls, in order to assess the selectivity of this immunosensor system for E2 (50 pg mL−1). No remarkable difference of currents was observed in comparison with the result obtained in presence of only E2 detection. The current variation due to the interfering substances was less than 5.0% of that without interferences, indicating that the selectivity of the immunosensor was acceptable (Fig. 6A). Because of the specificity of the antigen–antibody binding complex, the interferents had almost no effect on the determination of analytes.


image file: c4ra13066a-f6.tif
Fig. 6 Selectivity and stability of the immunosensor. Error bar = RSD (n = 5).

Stability of immunosensors is also a key factor in their application and development. The stability of the immunosensor was examined by checking periodically its current response. When the immunosensor was prepared and not in use, it was stored in a refrigerator at 4 °C. The current response of the as-prepared immunosensor decreased 4.2% after 7 day storage. Ten days later, the current response of the immunosensor decreased to about 88.6% of its initial value (Fig. 6B). The slow decrease in the current response may be due to the gradual denaturation of antibodies.

Application in analysis of water sample

In order to investigate the possibility of the immunosensor to be applied for practical analysis, the detection of E2 in water samples was performed using the proposed immunosensor with standard addition methods. 50 pg mL−1, 100 pg mL−1 and 250 pg mL−1 of E2 solution were added into water samples. The recovery of detection E2 was from 94% to 104% and the RSD was in the range of 1.5–3.4% (Table S2). The facts showed that the developed immunoassay methodology could be applied to the determination of E2 in water samples.

Conclusions

In summary, a novel electrochemical immunosensor for rapid determination of E2 by using Cu2S–BSA–E2 bioconjugates was fabricated. Based on a competitive immunoassay strategy and the excellent redox electrochemical signal of CuS2, the electrochemical immunosensor shows a linear response in the range of 25–7500 pg mL−1 with a detection limit of 7.5 pg mL−1. Due to the good precision, high sensitivity, acceptable stability and excellent selectivity, the electrochemical immunosensor has potential application for the rapid and simple measurement of E2 in real samples.

Acknowledgements

This study was supported by the Natural Science Foundation of China (no. 21175057, 21375047 and 21377046), the Science and Technology Development Plan of Shandong Province (no. 2014GSF120004), the Science and Technology Plan Project of Jinan (no. 201307010) and QW thanks the Special Foundation for Taishan Scholar Professorship of Shandong Province and UJN (no. ts20130937).

References

  1. W. Jiang, Y. Yan, M. Ma, D. Wang, Q. Luo, Z. Wang and S. K. Satyanarayanan, J. Environ. Sci., 2012, 24, 320–328 CrossRef CAS.
  2. S. K. Atkinson, V. L. Marlatt, L. E. Kimpe, D. R. Lean, V. L. Trudeau and J. M. Blais, Sci. Total Environ., 2012, 430, 119–125 CrossRef CAS PubMed.
  3. M. Andrew-Priestley, W. O'Connor, R. Dunstan, L. Van Zwieten, T. Tyler, A. Kumar and G. MacFarlane, Aquat. Toxicol., 2012, 120, 99–108 CrossRef PubMed.
  4. L. Sun, X. Shao, X. Hu, J. Chi, Y. Jin, W. Ye and Z. Fu, Aquat. Toxicol., 2011, 105, 629–639 CrossRef CAS PubMed.
  5. H.-S. Chang, K.-H. Choo, B. Lee and S.-J. Choi, J. Hazard. Mater., 2009, 172, 1–12 CrossRef CAS PubMed.
  6. M. Sun, L. Du, S. Gao, Y. Bao and S. Wang, Steroids, 2010, 75, 400–403 CrossRef CAS PubMed.
  7. L. Yuan, J. Zhang, P. Zhou, J. Chen, R. Wang, T. Wen, Y. Li, X. Zhou and H. Jiang, Biosens. Bioelectron., 2011, 29, 29–33 CrossRef CAS PubMed.
  8. L. Zhao, J.-M. Lin, Z. Li and X. Ying, Anal. Chim. Acta, 2006, 558, 290–295 CrossRef CAS PubMed.
  9. M. Miyashita, T. Shimada, H. Miyagawa and M. Akamatsu, Anal. Bioanal. Chem., 2005, 381, 667–673 CrossRef CAS PubMed.
  10. Q. He, S. Yuan, C. Chen and S. Hu, Mater. Sci. Eng. C, 2003, 23, 621–625 CrossRef.
  11. X. Liu, X. Wang, J. Zhang, H. Feng, X. Liu and D. K. Wong, Biosens. Bioelectron., 2012, 35, 56–62 CrossRef CAS PubMed.
  12. I. Ojeda, J. López-Montero, M. Moreno-Guzmán, B. Janegitz, A. González-Cortés, P. Yáñez-Sedeño and J. Pingarrón, Anal. Chim. Acta, 2012, 743, 117–124 CrossRef CAS PubMed.
  13. N. Chaisuwan, H. Xu, G. Wu and J. Liu, Biosens. Bioelectron., 2013, 46, 150–154 CrossRef CAS PubMed.
  14. X. Zhang, L. Wang, R. Ji, L. Yu and G. Wang, Electrochem. Commun., 2012, 24, 53–56 CrossRef PubMed.
  15. X. Bo, J. Bai, L. Wang and L. Guo, Talanta, 2010, 81, 339–345 CrossRef CAS PubMed.
  16. H. Lee, S. W. Yoon, E. J. Kim and J. Park, Nano Lett., 2007, 7, 778–784 CrossRef CAS PubMed.
  17. M. Xu, L. Cui, R. Han and S. Ai, J. Solid State Electrochem., 2012, 16, 2547–2554 CrossRef CAS.
  18. M. Shinde, D. Patil and R. Patil, Indian J. Pure Appl. Phys., 2013, 51, 713–716 CAS.
  19. S. Eissa, C. Tlili, L. L'Hocine and M. Zourob, Biosens. Bioelectron., 2012, 38, 308–313 CrossRef CAS PubMed.
  20. L. Liu, Q. He, F. Zhao, N. Xia, H. Liu, S. Li, R. Liu and H. Zhang, Biosens. Bioelectron., 2014, 51, 208–212 CrossRef CAS PubMed.
  21. L. Liu, D. Deng, Y. Xing, S. Li, B. Yuan, J. Chen and N. Xia, Electrochim. Acta, 2013, 89, 616–622 CrossRef CAS PubMed.
  22. Y. Wang, X. Ai, D. Miller, P. Rice, T. Topuria, L. Krupp, A. Kellock and Q. Song, CrystEngComm, 2012, 14, 7560–7562 RSC.
  23. T. Thongtem, A. Phuruangrat and S. Thongtem, J. Mater. Sci., 2007, 42, 9316–9323 CrossRef CAS.
  24. A. Kassim, H. S. MINa, L. K. Siang and S. Nagalingam, Chalcogenide Lett., 2011, 8, 405–410 CAS.
  25. M. Guo, L. He, P. Strong and H. Wang, Chemosphere, 2014, 112, 472–480 CrossRef CAS PubMed.
  26. A. Bard, Electrochemical Methods, Wiley, New York, 1980 Search PubMed.
  27. M. A. Panagopoulou, D. V. Stergiou, I. G. Roussis and M. I. Prodromidis, Anal. Chem., 2010, 82, 8629–8636 CrossRef CAS PubMed.
  28. R. Feng, Y. Zhang, H. Yu, D. Wu, H. Ma, B. Zhu, C. Xu, H. Li, B. Du and Q. Wei, Biosens. Bioelectron., 2013, 42, 367–372 CrossRef CAS PubMed.
  29. D. Butler and G. G. Guilbault, Sens. Actuators, B, 2006, 113, 692–699 CrossRef CAS PubMed.
  30. X. Liu, P. A. Duckworth and D. K. Wong, Biosens. Bioelectron., 2010, 25, 1467–1473 CrossRef CAS PubMed.
  31. J. Li, S. Liu, J. Yu, W. Lian, M. Cui, W. Xu and J. Huang, Sens. Actuators, B, 2013, 188, 99–105 CrossRef CAS PubMed.

Footnote

Electronic supplementary information (ESI) available. See DOI: 10.1039/c4ra13066a

This journal is © The Royal Society of Chemistry 2015
Click here to see how this site uses Cookies. View our privacy policy here.